(97 days)
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
This product consists of frame, wheels, seat, armrest, lithium battery, motor and controller with a lightweight and compact design. The seat cushion is detachable. The armrest can be flipped backward, which is convenient for the elderly to move. Users can drive the wheelchair by themselves through the control device. The wheelchair uses lithium batteries as its power source. The controller controls the drive left/right motor to realize the wheelchair forward, backward and turn functions. The frame's material of model ZH-W001 and ZH-W002 is aluminium alloy,the model ZH-W003 is carbon steel. The front wheels are driven wheels suitable for rotation, acceleration, retrograde and other actions of the wheelchair. The front wheels movement will be achieved by thrust generated from the rear wheels. The rear wheels are driving wheels to control the speed and direction. The wheels are PU tires. When in use, the operator drives the motor of the rear wheel by operating the joystick to achieve the rear wheels movement. The DC motor and brake system are fixed on the rear wheels. The max loading of the device is 120KG. Only for one person sit.
The provided FDA 510(k) clearance letter and summary discuss the substantial equivalence of an Electric Wheelchair to a predicate device. This document is for a physical medical device, not an AI/ML-driven device. As such, many of the requested criteria related to AI/ML model performance evaluation (e.g., sample size for AI test sets, expert adjudication, MRMC studies, standalone algorithm performance, training set details) are not applicable to this submission.
The acceptance criteria provided focus on fundamental safety and performance attributes typical of medical devices that are not AI-driven, and the "study" referred to is a series of non-clinical laboratory tests adhering to established ISO standards.
Here's the information that can be extracted and a clear statement about the inapplicable sections:
Acceptance Criteria and Device Performance for Electric Wheelchair (K250158)
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally established by compliance with a comprehensive set of ISO 7176 series standards and FDA guidance for Power Wheelchairs. The "reported device performance" refers to the results of non-clinical laboratory tests showing compliance with these standards.
| Acceptance Criteria (Evaluated Standard/Parameter) | Proposed Device Performance | Predicate Device Performance (K113463) | Outcome/Remark |
|---|---|---|---|
| Safety: | |||
| Braking Distance | ZH-W001: ≤1.2m; ZH-W002: ≤1.0m; ZH-W003: ≤1.0m | 1.5m | Shorter braking distance in proposed device is considered safer. |
| Maximum Safe Operational Incline Degree | 10° | 9° | Minor difference, not impacting performance significantly. |
| Max Speed Backward | 0.5m/s (1.8 km/h) | 2.4 mph (3.84 km/h) | Lower backward speed in proposed device is considered safer. |
| Biocompatibility (user-contacting materials) | Compliance with ISO10993-5, ISO10993-10, ISO10993-23 | Compliance with ISO10993-5, ISO10993-10 | Substantially Equivalent (SE); proposed device meets more recent extensive standards. |
| EMC | Compliance with IEC 60601-1-2 & ISO7176-21 | Compliance with IEC 60601-1-2 & ISO7176-21 | SE |
| Performance | Compliance with ISO7176 series (listed below) | Compliance with ISO7176 series (listed below) | SE |
| Label and Labeling | Conforms to FDA Regulatory | Conforms to FDA Regulatory | SE |
| Performance (ISO 7176 Series Compliance - specific values often not detailed in summary, rather confirmation of compliance): | |||
| ISO 7176-1 (Static ability) | Compliant | Compliant | SE |
| ISO 7176-2 (Dynamic stability) | Compliant | Compliant | SE |
| ISO 7176-3 (Brake performance) | Compliant | Compliant | SE |
| ISO 7176-4 (Theoretical distance range) | ZH-W001: 14.3 km; ZH-W002: 13.8 km; ZH-W003: 13.8km | 20 km | Difference due to battery size/wheel size; no new safety concerns. |
| ISO 7176-5 (Dimensions and weight) | Compliant (dimensions vary slightly but shown not to impact safety/performance) | Compliant | SE |
| ISO 7176-6 (Maximum speed) | ZH-W001: 1.8m/s; ZH-W002: 1.6m/s; ZH-W003: 1.6m/s | Up to 6 km/h (1.6 m/s) | Minor speed differences, no new safety concerns. |
| ISO 7176-7 (Measurements of seat and wheel dimensions) | Compliant | Compliant | SE |
| ISO 7176-8 (Static, impact and fatigue strengths) | Compliant | Compliant | SE |
| ISO 7176-9 (Climatic tests) | Compliant | Compliant | SE |
| ISO 7176-10 (Obstacle-climbing ability) | 20mm | 30 mm | Minor difference, no new safety concerns identified. |
| ISO 7176-11 (Test dummies) | Compliant | Compliant | SE |
| ISO 7176-13 (Friction of test surfaces) | Compliant | Compliant | SE |
| ISO 7176-14 (Power and control systems) | Compliant | Compliant | SE |
| ISO 7176-15 (Documentation and labeling) | Compliant | Compliant | SE |
| ISO 7176-21 (Resistance to ignition) | Compliant | Compliant | SE |
| ISO 7176-25 (Batteries and chargers) | Compliant | Compliant | SE |
| IEC 60601-1-2 (Electromagnetic Compatibility) | Compliant | Compliant | SE |
| Electrical Safety | Compliant | Compliant | SE |
| Maximum Loading Weight | 120kg (265lbs) | 114kg (251lbs) | Proposed device handles higher load; validated with max rated dummy. |
| Turning Radius | ZH-W001: 938mm; ZH-W002: 875mm; ZH-W003: 875mm | 800 mm | Minor difference, no new safety concerns. |
Study Proving Device Meets Acceptance Criteria:
The study that proves the device meets the acceptance criteria is a series of non-clinical laboratory tests conducted on the proposed device models (ZH-W001, ZH-W002, ZH-W003) in accordance with the specified international standards (ISO 7176 series, IEC 60601-1-2, and ISO 10993 series for biocompatibility). The 510(k) summary explicitly states: "The non-clinical laboratory data support the safety and performance of the subject device and demonstrate that the subject device should perform as intended in the specified use conditions."
AI/ML Specific Questions (Not Applicable to this Device):
The following questions are not applicable to this 510(k) submission as the device is an Electric Wheelchair, a physical medical device, and not an AI/ML software or system. Therefore, there is no AI/ML model being evaluated.
- Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable, no AI test set.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, no AI ground truth established by experts.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable, no AI test set requiring adjudication.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, no AI component or human reader study.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, no AI algorithm.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable, ground truth related to AI/ML is not relevant here. The "ground truth" for this device's performance is objective measurements against engineering and safety standards.
- The sample size for the training set: Not applicable, no AI training set.
- How the ground truth for the training set was established: Not applicable, no AI training set.
FDA 510(k) Clearance Letter - Electric Wheelchair
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.07.05
April 28, 2025
Zhejiang Cleisman Industry and Trade Co., LTD
℅ Ariel Xiang
Consultant
Shanghai Sungo Management Consulting Co. Ltd.
14th Floor, 1500# Century Avenue
Shanghai, Shanghai 200122
China
Re: K250158
Trade/Device Name: Electric Wheelchair (ZH-W001, ZH-W002, ZH-W003)
Regulation Number: 21 CFR 890.3860
Regulation Name: Powered Wheelchair
Regulatory Class: Class II
Product Code: ITI
Dated: March 14, 2025
Received: March 14, 2025
Dear Ariel Xiang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K250158 - Ariel Xiang Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K250158 - Ariel Xiang Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Julia E. Slocomb -S 2025.04.28 13:29:41 -04'00'
for Heather Dean, PhD
Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and Physical Medicine Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Indications for Use
Submission Number (if known)
K250158
Device Name
Electric Wheelchair (ZH-W001,ZH-W002 ,ZH-W003)
Indications for Use (Describe)
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 5
Zhejiang Cleisman Industry and Trade Co.,LTD
Workshop No.3,Xiaoweiyuan,Dongtuo Block ‖,Lishui High-tech Zone,Xinbi Street,Jinyun County, Lishui,Zhejiang,China
1 / 7
510(K) Summary
Document Prepared Date: 2025/3/11
A. Applicant
Zhejiang Cleisman Industry and Trade Co.,LTD
Workshop No.3,Xiaoweiyuan,Dongtuo,Block ‖,Lishui High-tech Zone,Xinbi Street,Jinyun County, Lishui,Zhejiang,China
Contact Name:Chongjie Zheng
Tel: +86 15989510939
Mail:674751020@qq.com
Submission Correspondent:
Primary contact: Ms.Ariel Xiang
Title: Senior Consultant
Shanghai SUNGO Management Consulting Co., Ltd.
14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China
Tel: +86-21-58817802
Email: shouqiu.xiang@sungoglobal.com
Secondary contact: Mr. Raymond Luo
Title: Technical Director
Shanghai SUNGO Management Consulting Co., Ltd.
14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China
Tel: +86-21-68828050
Email: zxfda@sungoglobal.com
B. Device
Trade Name: Electric Wheelchair
Common Name: Powered wheelchair
Models: ZH-W001,ZH-W002,ZH-W003
Regulatory Information
Classification Name: Powered wheelchair
Classification: Class II.
Product code: ITI
Regulation Number: 890.3860
Review Panel: Physical Medicine
C. Predicate device
510K number: K113463
power wheelchair, PL00l
SUZHOU KD Medical Appliance Co. Ltd.
Page 6
2 / 7
D. Indications for use of the device
The electric wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
E. Device Description:
This product consists of frame, wheels, seat, armrest, lithium battery, motor and controller with a lightweight and compact design. The seat cushion is detachable. The armrest can be flipped backward, which is convenient for the elderly to move. Users can drive the wheelchair by themselves through the control device. The wheelchair uses lithium batteries as its power source. The controller controls the drive left/right motor to realize the wheelchair forward, backward and turn functions. The frame's material of model ZH-W001 and ZH-W002 is aluminium alloy,the model ZH-W003 is carbon steel. The front wheels are driven wheels suitable for rotation, acceleration, retrograde and other actions of the wheelchair. The front wheels movement will be achieved by thrust generated from the rear wheels. The rear wheels are driving wheels to control the speed and direction. The wheels are PU tires. When in use, the operator drives the motor of the rear wheel by operating the joystick to achieve the rear wheels movement. The DC motor and brake system are fixed on the rear wheels. The max loading of the device is 120KG. Only for one person sit.
F. Comparison with predicate Device
Table 1 General Comparison
| Elements of Comparison | Proposed Device | Predicate Device (K230964) | Remark |
|---|---|---|---|
| Manufacturer | Zhejiang Cleisman Industry and Trade Co.,LTD | SUZHOU KD Medical Appliance Co. Ltd. | -- |
| Common or Usual name | Electric Wheelchair | Power Wheelchair | -- |
| Model(s) | ZH-W001,ZH-W002,ZH-W003 | PL00l | -- |
| Classification | Class II | Class II | Same |
| Classification regulation | 21 CFR890.3860 | 21 CFR890.3860 | Same |
| Product code | ITI | ITI | Same |
| Indications for use | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person | S.E. |
Page 7
3 / 7
| limited to a seated position. | |||
|---|---|---|---|
| Intended user | disabled or elderly person limited to a seated position. | disabled or elderly person limited to a seated position. | S.E. |
| Use condition | indoor and outdoor use | indoor and outdoor use | S.E |
| Number of wheels | 4,including two front wheels and two rear Wheels | 4,including two front wheels and two rear Wheels | S.E |
| Function of wheels | Front wheels:driven wheels suitable for rotation, acceleration, retrograde Rear wheels: driving wheels to control the speed and direction | Front wheels:driven wheels suitable for rotation, acceleration, retrograde Rear wheels: driving wheels to control the speed and direction | S.E |
| Movement control method | By Joystick control | By Joystick control | S.E |
| Driving system | Direct drive on the rear wheels | Direct drive on the rear wheels | S.E |
| Brake system | Automatic electromagnetic brake system | Automatic electromagnetic brake system | S.E |
| Braking distance | ZH-W001:≤1.2m ZH-W002:≤1.0m ZH-W003:≤1.0m | 1.5 m | Minor difference on braking distance will not cause different performance. Shorter distance for braking will be more safety. |
| Maximum safe operational incline degree | 10 ° | 9 ° | Minor difference on operational incline degree will not cause different performance. |
| Main frame material | ZH-W001:aluminium alloy ZH-W002:aluminium alloy ZH-W003:carbon steel | Aluminium alloy | S.E ZH:W003 is different.Different material used for frame, that such difference will not |
Page 8
4 / 7
| impact the safety and effectiveness of the subject device as the performance tests are conducted according to ISO 7176 series. | |||
|---|---|---|---|
| Armrest | PUR | PU | S.E |
| Back cushion | Nylon | PU foam covered by nylon fabric cloth | S.E |
| Seat cushion | Nylon | PU foam covered by nylon fabric cloth | S.E |
| Overall length | ZH-W001:1130mm ZH-W002:1020mm ZH-W003:1020mm | 880mm | Minor difference on wheelchair dimension will not cause different performance. All safety and performance have been validated with the maximum rated weight dummy. |
| Overall width | ZH-W001:660mm ZH-W002:600mm ZH-W003:600mm | 570mm | |
| Folded Dimension (lengthwidthheight) | ZH- W001:770mm650mm380mm ZH- W002:760mm600mm330mm ZH- W003:760mm600mm330mm | 750570400mm | |
| Front wheel size/type | ZH-W001:200mm50mm ZH-W002:190mm30mm ZH-W003:177mm*40mm | 6" x 2"/PU Solid tire | Minor difference on dimension of wheels will not cause different performance. |
| Rear wheel size/type | ZH-W001:317mm57mm ZH-W002:320mm35mm ZH-W003:272mm*40mm | 8″x 2.4″/ PU Solid tire | |
| Max speed forward | ZH-W001:1.8m/s ZH-W002:1.6m/s ZH-W003:1.6m/s | Up to 6 km/h (1.6 m/s), adjustable | Minor difference on max speed forward of wheels will not cause different performance. |
| Max speed backward | 0.5m/s(1.8 km/h) | 2.4 mph (3.84 km/h) | lower speed on max. backward speed will be more safety. |
| Max loading weight | 120kg (265lbs) | 114kg (251lbs) | Difference on loading weight will not cause different performance. All safety and performance |
Page 9
5 / 7
| have been validated with the maximum loading weight. | |||
|---|---|---|---|
| Maximum distance of travel on the fully charged battery | ZH-W001:14.3 km ZH-W002:13.8 km ZH-W003: 13.8km | 20 km | It is caused by the size of the wheel, will not cause different performance, The further away the better. |
| Battery | li-ion battery pack 24V,12Ah, li-ion battery pack; rechargeable, | 24 VDC 20Ah | the battery capacity will impact the travel distance, which will not cause new safety and effectiveness concerns raised. |
| Battery charger | Off-board charger Input: 100-240V, 50/60Hz, 1.5A, Output: 24 Vdc, 2A; | Off-board charger Input: 110-240V, 50-60Hz, Output: 24 Vdc, 2A; | S.E |
| Motor | Brushless DC motor, 24V,250W, 2pcs | Brushless DC motor; 24VDC; 180W; 2pcs | Minor difference on motor power. Both meet the require of ISO 7176-14 standard. It will not cause new safety and effectiveness concerns. |
| Electronic controller | Brushless dual-drive rocker controller;40A | Brushless dual- drive rocker controller | Similar controller is used, both the control system, including the joystick controller, the electromagnetic brakes and the user interface are similar. The joystick controls the directions and when the joystick is released, the powered wheelchair will slow down to stop |
Page 10
6 / 7
| and the brakes will automatically re-engage. The controller also provides the battery status displaying and abnormal condition displaying. Both of the control systems are evaluated according to standard ISO 7176-14:2008 and software validation requirement and there are no new safety and effectiveness concerns due to the difference. | |||
|---|---|---|---|
| Turning Radius | ZH-W001:938mm ZH-W002:875mm ZH-W003:875mm | 800 mm | Minor difference on turning radius will not cause new safety and effectiveness concerns. |
| Maximum obstacle climbing | 20mm | 30 mm | Minor difference on obstacle climbing will not cause new safety and effectiveness concerns. |
Table 2 safety comparison
| Item | Proposed Device | Predicate Device | Results |
|---|---|---|---|
| Biocompatibility | All user directly contacting materials are compliance with ISO10993-5、ISO10993-10 and ISO10993-23 requirements. | All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements. | SE |
| EMC | IEC 60601-1-2&ISO7176-21 | IEC 60601-1-2&ISO7176-21 | SE |
| Performance | ISO7176 series | ISO7176 series | SE |
| Label and labeling | Conforms to FDA Regulatory | Conforms to FDA Regulatory | SE |
Page 11
7 / 7
G. Substantial Equivalence Discussion
The proposed device and predicate device are complying to the same ISO standards, ISO 7176-1, ISO 7176-2, ISO 7176-3, ISO 7176-4, ISO 7176-5, ISO 7176-6, ISO 7176-7, ISO 7176-8, ISO 7176-9, ISO 7176-10, ISO 7176-11, ISO 7176-13, ISO 7176-14, ISO 7176-15, ISO 16840-10,ISO 7176-21, ISO 7176-25, and FDA guidance Submission for Power Wheelchair. The proposed device performs in a similar manner to the predicate device. All these tests have corresponding requirements/ control criteria following above mentioned standards. And the test results show that the subject product is substantially equivalent to the predicate device in performance. The performance testing demonstrates that the subject device is substantially equivalent to the predicate devices regarding Static ability (tipping angle), The Dynamic stability (Safe Gradient Maximum Gradient), Brake performance, Theoretical distance range, Dimension and weight, Maximum speed, Dimension of wheel Static, impact and fatigue strengths, Climatic tests, Obstacle-climbing ability, Dummy, friction of test surfaces, Power and control systems, Documentation and labeling, Resistance to ignition, Electromagnetic Compatibility and Electrical Safety, Batteries and chargers. The non-clinical laboratory data support the safety and performance of the subject device and demonstrate that the subject device should perform as intended in the specified use conditions.
H. Product Performance
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate devices, K113463 Power Wheelchair from SUZHOU KD Medical Appliance Co. Ltd.
§ 890.3860 Powered wheelchair.
(a)
Identification. A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.(b)
Classification. Class II (performance standards).