K Number
K244014
Device Name
i-STAT CG4+ cartridge with the i-STAT 1 System
Date Cleared
2025-05-02

(127 days)

Product Code
Regulation Number
862.1120
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The i-STAT CG4+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of pH, partial pressure of oxygen (PO2), and partial pressure of carbon dioxide (PCO2) in arterial, venous, or capillary whole blood in point of care or clinical laboratory settings. The i-STAT CG4+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of lactate in arterial or venous whole blood in point of care or clinical laboratory settings. pH, PO2, and PCO2 measurements are used in the diagnosis, monitoring, and treatment of respiratory, metabolic, and acid-base disturbances. Lactate measurements are used in (1) the diagnosis and treatment of lactic acidosis in conjunction with measurements of blood acid/base status, (2) monitoring tissue hypoxia and strenuous physical exertion, and (3) diagnosis of hyperlactatemia.
Device Description
The i-STAT CG4+ cartridge is used with the i-STAT 1 analyzer as part of the i-STAT 1 System to measure pH, partial pressure of oxygen (PO2), and partial pressure of carbon dioxide (PCO2) in arterial, venous or capillary whole blood and to measure lactate (Lac) in arterial or venous whole blood. The i-STAT 1 System is an in vitro diagnostic (IVD) medical device intended for the quantitative determination of various clinical chemistry tests contained within i-STAT cartridges using whole blood. The i-STAT 1 System consists of a portable blood analyzer (i-STAT 1 analyzer), single-use disposable test cartridges (i-STAT cartridges), liquid quality control and calibration verification materials, and accessories (i-STAT 1 Downloader/Recharger, i-STAT Electronic Simulator and i-STAT 1 Printer). The i-STAT 1 System, including the i-STAT CG4+ cartridge, is designed for use by trained medical professionals in point of care or clinical laboratory settings and is for prescription use only. The i-STAT CG4+ cartridge contains the required sensors, a fluid pack (calibrant pouch), a sample entry well and closure, fluid channels, waste chamber, and the necessary mechanical features for controlled fluid movement within cartridge. The i-STAT cartridge format allows all the tests in the cartridge to be performed simultaneously. All the test steps and fluid movements occur within the i STAT CG4+ cartridge. The i-STAT 1 analyzer interacts with the i-STAT CG4+ cartridge to move fluid across the sensors and generate a quantitative result. Cartridges require two to three drops of whole blood applied to the cartridge using a transfer device, by the trained user before the cartridge is placed within the analyzer. The i-STAT 1 analyzer is a handheld, in vitro diagnostic analytical device designed to run only i-STAT test cartridges. The analyzer functions as the main user interface and the electromechanical interface to the test cartridge. All within-cartridge fluid movements, as well as the timing and heating of the test cycle, are automated and controlled without user intervention by system software embedded within the analyzer.
More Information

No.

The device descriptions mention "system software embedded within the analyzer" but there is no mention of AI, machine learning, or deep learning in the document. The performance studies focus on traditional analytical metrics like precision, linearity, and comparison studies, which do not suggest the use of an AI model for results generation.

No.

This device is an in vitro diagnostic (IVD) medical device used for quantifying substances in blood for diagnosis, monitoring, and treatment purposes, not for providing therapy directly to a patient.

Yes

The intended use explicitly states that the measurements are used in the "diagnosis" of various medical conditions, such as respiratory, metabolic, and acid-base disturbances, and lactic acidosis. The device also quantifies analytes in whole blood, which is a common function of diagnostic devices.

No

The device is not a software-only medical device. The description clearly outlines hardware components such as the "i-STAT CG4+ cartridge," "i-STAT 1 analyzer" (a portable blood analyzer), and other accessories. While system software is mentioned as controlling automated functions within the analyzer, it is embedded within physical hardware that directly interacts with single-use cartridges and blood samples to perform in vitro diagnostic measurements.

Yes.
The device is explicitly identified as an "in vitro diagnostic (IVD) medical device" in the device description section. Additionally, its intended use involves the "in vitro quantification" of various analytes in whole blood for diagnostic, monitoring, and treatment purposes.

N/A

Intended Use / Indications for Use

The i-STAT CG4+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of pH, partial pressure of oxygen (PO2), and partial pressure of carbon dioxide (PCO2) in arterial, venous, or capillary whole blood in point of care or clinical laboratory settings.

The i-STAT CG4+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of lactate in arterial or venous whole blood in point of care or clinical laboratory settings.

pH, PO2, and PCO2 measurements are used in the diagnosis, monitoring, and treatment of respiratory, metabolic, and acid-base disturbances.

Lactate measurements are used in (1) the diagnosis and treatment of lactic acidosis in conjunction with measurements of blood acid/base status, (2) monitoring tissue hypoxia and strenuous physical exertion, and (3) diagnosis of hyperlactatemia.

Product codes (comma separated list FDA assigned to the subject device)

CHL, KHP

Device Description

The i-STAT CG4+ cartridge is used with the i-STAT 1 analyzer as part of the i-STAT 1 System to measure pH, partial pressure of oxygen (PO2), and partial pressure of carbon dioxide (PCO2) in arterial, venous or capillary whole blood and to measure lactate (Lac) in arterial or venous whole blood.

The i-STAT 1 System is an in vitro diagnostic (IVD) medical device intended for the quantitative determination of various clinical chemistry tests contained within i-STAT cartridges using whole blood. The i-STAT 1 System consists of a portable blood analyzer (i-STAT 1 analyzer), single-use disposable test cartridges (i-STAT cartridges), liquid quality control and calibration verification materials, and accessories (i-STAT 1 Downloader/Recharger, i-STAT Electronic Simulator and i-STAT 1 Printer). The i-STAT 1 System, including the i-STAT CG4+ cartridge, is designed for use by trained medical professionals in point of care or clinical laboratory settings and is for prescription use only.

The i-STAT CG4+ cartridge contains the required sensors, a fluid pack (calibrant pouch), a sample entry well and closure, fluid channels, waste chamber, and the necessary mechanical features for controlled fluid movement within cartridge. The i-STAT cartridge format allows all the tests in the cartridge to be performed simultaneously. All the test steps and fluid movements occur within the i STAT CG4+ cartridge. The i-STAT 1 analyzer interacts with the i-STAT CG4+ cartridge to move fluid across the sensors and generate a quantitative result. Cartridges require two to three drops of whole blood applied to the cartridge using a transfer device, by the trained user before the cartridge is placed within the analyzer.

The i-STAT 1 analyzer is a handheld, in vitro diagnostic analytical device designed to run only i-STAT test cartridges. The analyzer functions as the main user interface and the electromechanical interface to the test cartridge. All within-cartridge fluid movements, as well as the timing and heating of the test cycle, are automated and controlled without user intervention by system software embedded within the analyzer.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

trained medical professionals in point of care or clinical laboratory settings

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

A. Analytical Performance

  • Precision/Reproducibility:

    • Precision 20 days (Aqueous Materials): Evaluated precision using five (5) levels of aqueous materials per CLSI document EP05-A3. Conducted over at least 20 days at one site with multiple analyzers and one test cartridge lot. Results for repeatability, between-run, between-day, and within-laboratory precision estimated for each level are provided in Table 2.
    • Multi-site and operator-to-operator precision (Aqueous materials): Performed at three (3) point of care sites using a panel of aqueous material containing five (5) levels of pH, PO2, PCO2, and lactate. Study followed a 3x5x2x3 design based on CLSI EP05-A3. Each level was tested once per day by two (2) operators for five (5) days on six (6) i-STAT 1 analyzers using one (1) lot of i-STAT CG4+ cartridges. Repeatability, between-day, between-operator, within-site, between-site variance components, and reproducibility were calculated for all sites combined and provided in Table 3.
    • Precision (Whole Blood): Evaluated using anticoagulated arterial, venous, and capillary whole blood specimens. Assessed using duplicate test results collected across multiple point of care sites. Specimens grouped into subintervals of the reportable range for analysis. Results summarized in Table 4.
    • Precision (Within-Sample, Capillary Specimens): Evaluated whole blood within sample precision using native and contrived pooled capillary whole blood samples. Native capillary whole blood samples from 30 subjects via fingerstick punctures, tested in duplicate at Abbott Point of Care. Results in Table 5. Contrived capillary whole blood samples from 27 subjects (16 for low PO2, 11 for high PO2), tested in duplicate. Results in Table 6.
  • Linearity/assay reportable range:

    • Linearity: Designed based on CLSI EP06-Ed2. Evaluated by preparing whole blood samples of varying analyte levels. Demonstrated linearity across the reportable range for each test. Regression summary provided in Table 7.
  • Traceability, Calibration and Reference Range:

    • Traceability and Calibration: pH values traceable to NIST SRMs 186-I, 186-II, 185, and 187. PO2 and PCO2 values traceable to NIST SRMs via commercially available certified specialty medical gas standards. Lactate values traceable to i-STAT System working calibrator prepared from sodium L-lactate (Sigma-Aldrich Fluka, >99 % purity), as no international conventional reference measurement procedure is available.
    • Reference Range: Based on literature reference, listed in Table 8.
  • Detection Limit:

    • Limit of Quantitation (LoQ): Based on CLSI EP17-A2. Evaluated using two (2) i-STAT CG4+ cartridge lots and whole blood altered to a low analyte level. LoQ determined to be at or below the lower limit of the reportable range for each test (Table 9).
    • Limit of Blank and Detection (LoB/LoD): Based on CLSI EP17-A2. Evaluated for the Lactate test using two (2) i-STAT CG4+ cartridge lots. Whole blood altered to achieve a blank lactate level for LoB testing and two (2) low lactate levels for LoD testing. LoB and LoD determined based on maximal values obtained for each lot tested (Table 10).
  • Analytical Specificity:

    • Interference: Evaluated using whole blood samples based on CLSI EP07-ED3. Effect of each substance compared between a control sample spiked with blank solvent and a test sample spiked with the potentially interfering substance at toxic/pathological concentration (based on CLSI EP37-ED1). A substance was identified as an interferent if the difference in means (or medians) was outside of the allowed error (±Ea). Dose-response performed for identified interferents. Table 11 lists tested substances and interference results.
    • Other sensitivity studies:
      • Oxygen Sensitivity for the i-STAT Lactate Test: Effect of oxygen evaluated with low and high oxygen levels using whole blood samples altered to four (4) lactate levels. Equivalency determined if 95% CI of difference in means was within the allowable error (±Ea). Demonstrated insensitivity to oxygen levels between 28 and 521 mmHg.
      • Altitude: Performance evaluated at approximately 10,000 feet above sea level using whole blood samples. Results compared to a comparator device (K200492). Passing-Bablok regression analyses performed based on CLSI EP09c-ED3. Correlation coefficient and slope met acceptance criteria, demonstrating equivalent performance (Table 12).

B. Comparison Studies

  • Method Comparison:

    • Demonstrated for arterial, venous, and capillary whole blood specimens per CLSI EP09c-ED3.
    • Heparinized arterial and venous whole blood specimens compared against a comparative method. Passing-Bablok linear regression analysis performed for pH and PO2 (i-STAT 1 Analyzer replicate vs. comparative method singlicate) and PCO2 and lactate (i-STAT 1 Analyzer replicate vs. mean of comparative method).
    • Two (2) capillary specimens collected from skin puncture, one tube for i-STAT 1 analyzer, one for comparative method.
    • Arterial, venous, and capillary data pooled for pH, PO2, and PCO2, and a Passing-Bablok linear regression analysis performed (i-STAT CG4+ singlicate vs. comparative method result). Results in Table 13.
    • Arterial and venous data pooled for Lactate, and a Passing-Bablok linear regression analysis performed (i-STAT CG4+ singlicate vs. comparative method result). Results in Table 14.
    • Method comparison results for capillary whole blood specimens only for pH, PO2, and PCO2 are in Table 15.
    • Bias at medical decision levels for native capillary whole blood specimens only for pH, PO2, and PCO2 are in Table 16.
  • Matrix Equivalence:

    • Evaluated performance using non-anticoagulated arterial and venous whole blood specimens per CLSI EP35.
    • Matrix equivalence assessed by comparing arterial or venous whole blood specimens collected without anticoagulant (candidate specimen type) to samples collected with balanced heparin or lithium heparin anticoagulant (primary specimen type).
    • Each specimen tested in duplicate using two (2) i-STAT CG4+ cartridges with two (2) i-STAT 1 analyzers.
    • Passing-Bablok linear regression analysis performed using the first replicate result from the candidate (y-axis) versus the first replicate result from the primary specimen (x-axis). Results in Table 17.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found (specific metrics such as sensitivity, specificity, PPV, NPV are not provided for this type of device and study)

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K223857

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K200492

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 862.1120 Blood gases (P

CO2 , PO2 ) and blood pH test system.(a)
Identification. A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.(b)
Classification. Class II.

FDA 510(k) Clearance Letter - i-STAT CG4+ Cartridge

Page 1

May 2, 2025

Abbott Point of Care Inc.
Brian Ma, Ph.D.
Principal Specialist, Regulatory Affairs
400 College Road East
Princeton, New Jersey 08540

Re: K244014
Trade/Device Name: i-STAT CG4+ cartridge with the i-STAT 1 System
Regulation Number: 21 CFR 862.1120
Regulation Name: Blood Gases (PCO2, PO2) And Blood pH Test System
Regulatory Class: Class II
Product Code: CHL, KHP
Dated: April 4, 2025
Received: April 4, 2025

Dear Brian Ma:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

Page 3

the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Paula V. Caposino -S

Paula Caposino, Ph.D.
Deputy Director
Division of Chemistry and Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Indications for Use

Submission Number (if known)
K244014

Device Name
i-STAT CG4+ cartridge with the i-STAT 1 System

Indications for Use (Describe)

The i-STAT CG4+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of pH, partial pressure of oxygen (PO2), and partial pressure of carbon dioxide (PCO2) in arterial, venous, or capillary whole blood in point of care or clinical laboratory settings.

The i-STAT CG4+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of lactate in arterial or venous whole blood in point of care or clinical laboratory settings.

pH, PO2, and PCO2 measurements are used in the diagnosis, monitoring, and treatment of respiratory, metabolic, and acid-base disturbances.

Lactate measurements are used in (1) the diagnosis and treatment of lactic acidosis in conjunction with measurements of blood acid/base status, (2) monitoring tissue hypoxia and strenuous physical exertion, and (3) diagnosis of hyperlactatemia.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

510(k) Summary

The information in this 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.

1. Submitter Information

Owner: Abbott Point of Care Inc.
400 College Road East
Princeton, NJ 08540

Contact: Primary: Brian Ma, PhD
Principal Specialist, Regulatory Affairs
Phone: 613-688-5949

Secondary: Mojgan Soleimani
Director, Regulatory Strategy and Design Quality Assurance
Phone: 613-295-0932

Date Prepared: May 2, 2025

2. Device Information

Proprietary Name: i-STAT CG4+ cartridge with the i-STAT 1 System
Common Name: Blood gas test, lactate test, analyzer, handheld
510(k) Number: K244014

Product CodeDevice Classification NameRegulation NumberClassPanel
CHLElectrode measurement, blood gases (PCO2, PO2) and blood pH862.1120IIClinical Chemistry
KHPAcid, lactic, enzymatic method862.1450IClinical Chemistry

3. Predicate Device

Proprietary Name: i-STAT G3+ cartridge with the i-STAT 1 System
510(k) Number: K223857

Page 6

Product CodeDevice Classification NameRegulation NumberClassPanel
CHLElectrode measurement, blood gases (PCO2, PO2) and blood pH862.1120IIClinical Chemistry

4. Device Description

The i-STAT CG4+ cartridge is used with the i-STAT 1 analyzer as part of the i-STAT 1 System to measure pH, partial pressure of oxygen (PO2), and partial pressure of carbon dioxide (PCO2) in arterial, venous or capillary whole blood and to measure lactate (Lac) in arterial or venous whole blood.

The i-STAT 1 System is an in vitro diagnostic (IVD) medical device intended for the quantitative determination of various clinical chemistry tests contained within i-STAT cartridges using whole blood. The i-STAT 1 System consists of a portable blood analyzer (i-STAT 1 analyzer), single-use disposable test cartridges (i-STAT cartridges), liquid quality control and calibration verification materials, and accessories (i-STAT 1 Downloader/Recharger, i-STAT Electronic Simulator and i-STAT 1 Printer). The i-STAT 1 System, including the i-STAT CG4+ cartridge, is designed for use by trained medical professionals in point of care or clinical laboratory settings and is for prescription use only.

The i-STAT CG4+ cartridge contains the required sensors, a fluid pack (calibrant pouch), a sample entry well and closure, fluid channels, waste chamber, and the necessary mechanical features for controlled fluid movement within cartridge. The i-STAT cartridge format allows all the tests in the cartridge to be performed simultaneously. All the test steps and fluid movements occur within the i STAT CG4+ cartridge. The i-STAT 1 analyzer interacts with the i-STAT CG4+ cartridge to move fluid across the sensors and generate a quantitative result. Cartridges require two to three drops of whole blood applied to the cartridge using a transfer device, by the trained user before the cartridge is placed within the analyzer.

The i-STAT 1 analyzer is a handheld, in vitro diagnostic analytical device designed to run only i-STAT test cartridges. The analyzer functions as the main user interface and the electromechanical interface to the test cartridge. All within-cartridge fluid movements, as well as the timing and heating of the test cycle, are automated and controlled without user intervention by system software embedded within the analyzer.

5. Intended Use Statement

The i-STAT CG4+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of pH, partial pressure of oxygen (PO2), and partial pressure of carbon dioxide (PCO2) in arterial, venous, or capillary whole blood in point of care or clinical laboratory settings.

The i-STAT CG4+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of lactate in arterial or venous whole blood in point of care or clinical laboratory settings.

Page 7

pH, PO2, and PCO2 measurements are used in the diagnosis, monitoring, and treatment of respiratory, metabolic, and acid-base disturbances.

Lactate measurements are used in (1) the diagnosis and treatment of lactic acidosis in conjunction with measurements of blood acid/base status, (2) monitoring tissue hypoxia and strenuous physical exertion, and (3) diagnosis of hyperlactatemia.

6. Summary Comparison of Technological Characteristics

Table 1: Similarities and Differences: System (Test and Instrument)

Feature or CharacteristicCandidate Device: i-STAT CG4+ cartridge with the i-STAT 1 SystemPredicate Device: i-STAT G3+ cartridge with the i-STAT 1 System (K223857)
Intended UseThe i-STAT CG4+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of pH, partial pressure of oxygen (PO2), and partial pressure of carbon dioxide (PCO2) in arterial, venous, or capillary whole blood in point of care or clinical laboratory settings.

The i-STAT CG4+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of lactate in arterial or venous whole blood in point of care or clinical laboratory settings.

pH, PO2, and PCO2 measurements are used in the diagnosis, monitoring, and treatment of respiratory, metabolic, and acid-base disturbances.

Lactate measurements are used in (1) the diagnosis and treatment of lactic acidosis in conjunction with measurements of blood acid/base status, (2) monitoring tissue hypoxia and strenuous physical exertion, and (3) diagnosis of hyperlactatemia. | The i-STAT G3+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of pH, partial pressure of oxygen (PO2), and partial pressure of carbon dioxide (PCO2) in arterial, venous, or capillary whole blood in point of care or clinical laboratory settings.

pH, PO2, and PCO2 measurements are used in the diagnosis, monitoring, and treatment of respiratory, metabolic, and acid-base disturbances. |
| Device Classification(s) | Class II - PCO2, PO2, pH
Class I - Lactate | Class II - PCO2, PO2, pH |
| Product Code(s) | CHL, KHP | CHL |
| Regulation Number(s) | 862.1120, 862.1450 | 862.1120 |

Page 8

Feature or CharacteristicCandidate Device: i-STAT CG4+ cartridge with the i-STAT 1 SystemPredicate Device: i-STAT G3+ cartridge with the i-STAT 1 System (K223857)
Reportable RangepH: 6.500 – 7.800
PO2: 5 – 700 mmHg, 0.7 – 93.3 kPa
PCO2: 5 – 130 mmHg, 0.67 – 17.33 kPa
Lactate: 0.30 – 20.00 mmol/L, 2.7 – 180.2 mg/dLpH: Same
PO2: Same
PCO2: Same
Sample TypepH, PO2, PCO2: Arterial, venous or capillary whole blood
lactate: Arterial or venous whole bloodpH, PO2, PCO2: Same
Sample Volume95 µLSame
Sample PreparationReady to UseSame
Sample CollectionpH, PO2, PCO2:
• Without anticoagulant (for arterial and venous whole blood sample types)
• With balanced heparin anticoagulant or lithium heparin anticoagulant (for arterial, venous, and capillary whole blood sample types)

lactate:
• Without anticoagulant (for arterial and venous whole blood sample types)
• With balanced heparin anticoagulant or lithium heparin anticoagulant (for arterial and venous whole blood sample types) | pH, PO2, PCO2: Same |

Page 9

Feature or CharacteristicCandidate Device: i-STAT CG4+ cartridge with the i-STAT 1 SystemPredicate Device: i-STAT G3+ cartridge with the i-STAT 1 System (K223857)
TraceabilitypH: Traceable to NIST SRMs 186-I, 186-II, 185 and 187
PO2, PCO2: Traceable to NIST SRMs via commercially available certified specialty medical gas tanks
Lactate: Certified Standard Reference Material not available at present; traceable to i-STAT manufacturer's working calibrator (MWC) prepared from sodium L-lactate (Sigma-Aldrich Fluka, >99 % purity)pH: Same
PO2, PCO2: Same
Calibration1-point on-board contained within cartridgeSame

Page 10

Feature or CharacteristicCandidate Device: i-STAT CG4+ cartridge with the i-STAT 1 SystemPredicate Device: i-STAT G3+ cartridge with the i-STAT 1 System (K223857)
Time to Test/Sample Stability (Time from collection to sample fill)Without anticoagulant:
pH, PO2, PCO2 (arterial and venous): within 3 minutes
Lactate (arterial and venous): Immediately

With anticoagulant:
pH, PO2, PCO2 (arterial and venous): within 10 minutes
pH, PO2, PCO2 (capillary): within 3 minutes
Lactate (arterial and venous): Immediately | Without anticoagulant:
pH, PO2, PCO2 (arterial and venous): Same

With anticoagulant:
pH, PO2, PCO2 (arterial and venous): Same
pH, PO2, PCO2 (capillary): Same |
| Principle of Measurement | pH, PCO2: Potentiometric measurement between active working sensor and independent reference sensor.
PO2: Amperometric measurement of oxygen reduction current.
Lactate: Amperometric measurement of oxidized hydrogen peroxide produced by lactate oxidase activity. | pH, PCO2: Same
PO2: Same |
| Reagent Format | Cartridge | Same |
| Storage Conditions | Refrigerated at 2 to 8°C (35 to 46°F) until expiration date.
Room Temperature at 18-30°C (64-86°F) for 2 months. | Same |
| Analyzer Type | Handheld | Same |

Page 11

7. Performance Characteristics

A. Analytical Performance

a. Precision/Reproducibility:

i. Precision 20 days (Aqueous Materials)

The precision of the i-STAT pH, PO2, PCO2, and Lactate tests in the i-STAT CG4+ cartridge with the i-STAT 1 System was evaluated using five (5) levels of aqueous materials. The study followed the standard single-site 20x2x2 experimental design based on guidance provided in CLSI document EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition. The study was conducted using multiple analyzers and one (1) test cartridge lot over at least 20 days at one (1) site. Repeatability, between-run, between-day, and within-laboratory precision were estimated for each level. The results of the 20-day precision study for the i-STAT CG4+ cartridge on the i-STAT 1 System are shown in Table 2.

Table 2: 20-Day Precision of i-STAT CG4+ Cartridge on the i-STAT 1 Analyzer

Test (units)Fluid LevelMeanNRepeatabilityBetween-runBetween-dayWithin-Laboratory
SD%CVSD%CVSD%CVSD%CV
pH (pH units)CV L16.5701840.003590.050.002800.040.000340.010.004570.07
CV L27.0259840.001780.030.001050.010.000680.010.002180.03
CV L37.4532830.002310.030.000510.010.000520.010.002420.03
CV L47.6338840.009610.130.003250.040.002660.030.010490.14
CV L57.9653840.002360.030.000810.010.001650.020.002990.04
PO2 (mmHg)CV L171.0841.712.410.650.921.031.452.102.96
CV L282.6841.742.110.580.700.450.541.892.29
CV L3108.9831.861.711.000.910.670.622.222.03
CV L4138.8842.521.820.900.651.120.812.902.09
CV L5372.9844.661.255.291.422.070.567.351.97
PCO2 (mmHg)CV L189.41841.2511.400.6260.700.3540.401.4431.61
CV L256.28840.6211.100.3380.600.1650.290.7261.29
CV L329.37830.3651.240.1280.430.1410.480.4121.40
CV L422.69840.7073.110.1960.860.1910.840.7583.34
CV L512.19840.3773.100.1311.070.1231.010.4183.43
Lactate (mmol/L)CV L119.791840.18270.920.08690.440.02260.110.20351.03
CV L27.874840.05920.750.03020.380.01230.160.06760.86
CV L32.110830.01280.610.00510.240.00460.220.01460.69
CV L40.820840.01351.640.00360.440.00490.600.01481.80
CV L50.410840.01283.130.00280.700.00551.350.01423.47

Page 12

ii. Multi-site and operator-to-operator precision (Aqueous materials)

Multi-day precision testing was performed at three (3) point of care sites using a panel of aqueous material containing five (5) levels of pH, PO2, PCO2, and lactate. The study followed a 3x5x2x3 design based on guidance provided in CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition. At each site, each level was tested once per day by two (2) operators for five (5) days on six (6) i-STAT 1 analyzers using one (1) lot of i-STAT CG4+ cartridges. Repeatability, between-day, between-operator, within-site, between-site variance components, and reproducibility were calculated for all sites combined and provided in the Table 3 below.

Table 3: Multi-Site Multi-Day Precision of i-STAT CG4+ Cartridge on the i-STAT 1 Analyzer

Test (units)Fluid LevelNMeanRepeatabilityBetween-DayBetween-OperatorWithin-SiteBetween-SiteReproducibility
SD%CVSD%CVSD%CVSD%CVSD%CVSD%CV
pH (pH units)CV L1906.57250.004040.060.001110.020.002810.040.005040.080.001470.020.005250.08
CV L2907.02660.002040.030.001330.020.001240.020.002740.040.002010.030.003400.05
CV L3907.45400.001390.020.000620.010.001190.020.001930.030.000800.010.002090.03
CV L4907.63730.001940.030.000460.010.000960.010.002220.030.001420.020.002630.03
CV L5907.96810.001940.020.000620.010.000930.010.002240.030.001070.010.002480.03
PO2 (mmHg)CV L19176.92.803.640.951.232.633.423.965.152.202.874.535.89
CV L29087.42.022.310.860.982.132.443.063.502.562.933.994.57
CV L390113.42.342.060.570.501.571.382.872.533.082.714.213.71
CV L490141.42.631.861.501.061.150.823.242.293.472.454.743.35
CV L590366.06.261.712.920.807.532.0610.212.794.661.2711.233.07
PCO2 (mmHg)CV L19089.411.3651.530.5010.560.7100.791.6181.810.0000.001.6181.81
CV L29057.200.5080.890.2870.500.3110.540.6611.160.0000.000.6611.16
CV L39029.990.2890.960.1690.560.0630.210.3411.140.1400.470.3681.23
CV L49023.030.3361.460.0920.400.0000.000.3481.510.0540.240.3531.53
CV L59012.180.3612.960.0000.000.0000.000.3612.960.0000.000.3612.96
Lactate (mmol/L)CV L19118.9280.19051.010.05730.300.00000.000.1991.050.00000.000.19901.05
CV L2907.6360.04230.550.02160.280.01260.160.04910.640.00000.000.04910.64
CV L3902.0870.00840.400.00330.160.00220.110.00930.440.00610.290.01110.53
CV L4900.8380.00901.080.00881.050.00230.280.01281.530.00530.630.01391.65
CV L5900.4330.01242.870.00561.290.00120.270.01373.160.00531.210.01463.39

Page 13

iii. Precision (Whole Blood)

The precision of the i-STAT pH, PO2, PCO2 and Lactate tests in the i-STAT CG4+ cartridge on the i-STAT 1 System was evaluated using anticoagulated arterial, venous, and capillary¹ whole blood specimens. The whole blood precision was assessed using duplicate test results collected across multiple point of care sites. For each sample type, the specimens were grouped into subintervals of the test reportable range for analysis. The results are summarized in Table 4.

Table 4: Whole blood precision of arterial, venous, and capillary whole blood for i-STAT CG4+ cartridge on the i-STAT 1 analyzer

Test (units)Sample TypeSample RangeNMeanSD%CV
pH (pH units)Venous Whole Blood6.500-7.30097.02650.002350.03
>7.300 – 7.4501547.37450.006680.09
>7.450-7.800127.53550.007320.10
Arterial Whole Blood6.500-7.30097.22290.002620.04
>7.300 – 7.4501247.38160.004860.07
>7.450-7.800437.47630.006900.09
Capillary Whole Blood6.500-7.30037.24770.013550.19
7.300-7.4501217.40990.020840.28
>7.450-7.800327.47810.026090.35
PO2 (mmHg)Venous Whole Blood10-4012526.20.913.49
>40-502143.10.982.26
>50-1002459.31.322.23
>100-2503227.72.651.16
>250-7008508.46.961.37
Arterial Whole Blood>50-10010873.21.291.76
>100-25066135.92.902.13
>250-7003381.38.942.35
Capillary Whole Blood10-401533.84.3012.74
>40-503446.03.848.35
>50-10011263.46.049.52
>100-250*5166.412.987.80
>250-7008489.68.531.74
PCO2 (mmHg)Venous Whole Blood5.0-35.02330.170.3791.26
>35.0-50.011946.480.6391.38
>50.0-62.53157.050.6871.20
>62.5-130.010117.051.6501.41
Arterial Whole Blood5.0-35.04833.240.3931.18
>35.0-50.010544.550.6411.44
>50.0-62.51661.331.1001.79
>62.5-130.0677.181.0801.40

¹ The capillary whole blood clinical precision study design involved collection of capillary blood (from either two fingersticks or a single heelstick) into two (2) anticoagulated capillary tubes and each tube was used to fill a single i-STAT CG4+ cartridge.

Page 14

Test (units)Sample TypeSample RangeNMeanSD%CV
Capillary Whole Blood5.0-35.04732.311.7365.37
>35.0-50.010540.002.2585.64
>50.0-62.5358.351.9673.37
>62.5-130.0168.202.2633.32
Lactate (mmol/L)Venous Whole Blood0.30-1.001000.6390.01271.99
>1.00-5.00811.5490.02061.33
>5.00-20.002012.4760.07560.61
Arterial Whole Blood0.30-1.00550.6530.01382.11
>1.00-5.00761.7710.01841.04
>5.00-20.0038.1200.02520.31

*A potential outlier due to preanalytical handling was identified for one (1) subject; PO2 precision for capillary whole blood in sample range 100-250 mmHg with this subject excluded (N=4) is as follows: Mean = 177.8 mmHg; SD = 5.50, %CV=3.09

iv. Precision (Within-Sample, Capillary Specimens)

Whole blood within sample precision of the i-STAT pH, PO2, and PCO2 tests in the i-STAT CG4+ cartridge on the i-STAT 1 System was evaluated using native and contrived pooled capillary whole blood samples.

Native capillary whole blood samples collected from 30 subjects via fingerstick punctures and tested in duplicate at Abbott Point of Care were assessed for within sample precision. Results are summarized in Table 5.

Table 5: Whole blood precision (within sample) of native capillary whole blood for i-STAT CG4+ cartridge on the i-STAT 1 analyzer

TestUnitsNMeanSD%CV
pHpH units607.48480.016880.23
PO2mmHg60143.14.553.18
PCO2mmHg6030.341.5054.96

Capillary whole blood samples collected from an additional 27 subjects used to prepare two levels of PO2 (16 for low and 11 for high) at Abbott Point of Care were assessed for within sample precision by testing each sample in duplicate. Results for the contrived capillary whole blood samples are summarized in Table 6.

Page 15

Table 6: Whole blood precision (within sample) of contrived capillary whole blood for i-STAT CG4+ cartridge on the i-STAT 1 analyzer

TestUnitsSample LevelNMeanSD%CV
pHpH unitsL1327.31110.009960.14
L2227.31760.015720.21
PO2mmHgL13243.40.511.18
L222427.45.861.37
PCO2mmHgL13240.450.3870.96
L22236.160.1780.49

b. Linearity/assay reportable range:

i. Linearity

The study was designed based on CLSI EP06-Ed2: Evaluation of the Linearity of Quantitative Measurement Procedures – Second Edition. The linearity of the i-STAT pH, PO2, PCO2, and Lactate tests in the i-STAT CG4+ cartridge with the i-STAT 1 System were evaluated by preparing whole blood samples of varying analyte levels for each i-STAT test. The i-STAT pH, PO2, PCO2, and Lactate tests in the i-STAT CG4+ cartridge demonstrated linearity across the reportable range for each test. The regression summary of the results from each i-STAT test versus the expected values of the whole blood samples is provided in Table 7.

Table 7: Regression summary for the i-STAT pH, PO2, PCO2, and Lactate tests in the CG4+ cartridge on the i-STAT 1 analyzer

TestUnitsReportable RangeRange TestedSlopeIntercept
pHpH units6.500 – 7.8006.4509 – 7.95001.012-0.0960.9996
PO2mmHg5 – 7003.5 – 723.40.9900.1760.9970
PCO2mmHg5.0 – 130.02.59 – 145.971.016-0.5130.9986
Lactatemmol/L0.30 – 20.000.276 – 21.5021.0120.0330.9991

c. Traceability, Calibration and Reference Range

i. Traceability and Calibration

The measured analytes in the i-STAT CG4+ cartridge are traceable to the following reference materials or methods. The i-STAT controls and calibration verification materials are validated for use only with the i-STAT System and assigned values may not be commutable with other methods.

pH values assigned to the i-STAT System controls and calibration verification materials are traceable to the U.S. National Institute of Standards and Technology (NIST) standard reference materials SRMs 186-I, 186-II, 185, and 187.

PO2 values assigned to the i‑STAT System controls and calibration verification materials are traceable to U.S. National Institute of Standards and Technology (NIST) standard reference materials via commercially available certified specialty medical gas standards.

PCO2 values assigned to the i‑STAT System controls and calibration verification materials are traceable to U.S. National Institute of Standards and Technology (NIST)

Page 16

standard reference materials via commercially available certified specialty medical gas standards.

Lactate: Presently, no international conventional reference measurement procedure or international conventional calibrator for lactate is available. Lactate values assigned to the i-STAT System controls and calibration verification materials are traceable to i-STAT System working calibrator prepared from sodium L-lactate (Sigma-Aldrich Fluka, >99 % purity).

ii. Reference Range

Reference ranges for the i-STAT CG4+ cartridge are based on literature reference and are listed in Table 8 below.

Table 8: Reference Ranges

TestUnitsReportable RangeReference Range
ArterialVenous
pHpH units6.50 – 7.807.35 – 7.45¹7.31 – 7.41*
PO2mmHg5 – 70080 – 105²**
kPa0.7 – 93.310.7 – 14.0²**
PCO2mmHg5 – 13035 – 45¹41 – 51¹
kPa0.67 – 17.334.67 – 6.005.47 – 6.80
Lactatemmol/L0.30 – 20.000.36 – 1.25³***0.90 – 1.70³***
mg/dL2.7 – 180.23.2 – 11.3³***8.1 – 15.3³***

*Calculated from Siggard-Andersen nomogram.
**The reference ranges shown are for a healthy population. Interpretation of blood gas measurements depend on the underlying condition (e.g., patient temperature, ventilation, posture and circulatory status).
***The i-STAT reference ranges for whole blood listed above are similar to reference ranges derived from serum or plasma measurements with standard laboratory methods.

¹ Painter, P. C., Cope, J. Y., and Smith, J. L. (1994) "Reference Ranges, Table 41-20" In Teitz Textbook of Clinical Chemistry, 2nd Ed. Burtis CA, Ashwood ER, eds. W.B. Saunders, Philadelphia.
² Statland, B. E. (1987) Clinical Decision Levels for Lab Tests, 2nd Ed., Medical Economics Books, Oradell, N.J.
³ Sacks, D. B. (1994) Carbohydrates In Teitz Textbook of Clinical Chemistry, 2nd Ed. Burtis CA, Ashwood ER, eds. W.B. Saunders, Philadelphia.

d. Detection Limit

i. Limit of Quantitation (LoQ)

The study was based on the CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition. The LoQ of the i-STAT pH, PO2, PCO2, and Lactate tests in the i-STAT CG4+ cartridge were evaluated on the i-STAT 1 analyzer using two (2) i-STAT CG4+ cartridge lots, and whole blood that was altered to a low analyte level for each i-STAT test. The LoQ for each of the i-STAT tests was determined to be at or below the lower limit of the reportable range for each of the i-STAT tests as shown in Table 9.

Page 17

Table 9: Summary of LoQ results for the pH, PO2, PCO2, and lactate tests in the i-STAT CG4+ cartridge on the i-STAT 1 analyzer

TestUnitsLower limit of the reportable rangeLoQ
pHpH units6.5006.471
PO2mmHg55
PCO2mmHg5.03.0
Lactatemmol/L0.300.18

ii. Limit of Blank and Detection (LoB/LoD)

The study was based on CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition. The LoB and LoD of the i-STAT Lactate test in the i-STAT CG4+ cartridge were evaluated on the i-STAT 1 analyzer using two (2) i-STAT CG4+ cartridge lots. Whole blood was altered to achieve a blank lactate level for LoB testing and two (2) low lactate levels for LoD testing. The LoB and LoD were determined based on the maximal LoB or LoD value obtained for each lot tested. The determined LoB and LoD for the i-STAT Lactate test on the i-STAT 1 analyzer are shown in the Table 10 below.

Table 10: Summary of LoB and LoD results for the lactate test in the i-STAT CG4+ cartridge on the i-STAT 1 analyzer

TestUnitsLoBLoD
Lactatemmol/L00.026

e. Analytical Specificity

i. Interference

The interference performance of the i-STAT pH, PO2, PCO2, and Lactate tests in the i-STAT CG4+ cartridge on the i STAT 1 analyzer with the i-STAT 1 System was evaluated using whole blood samples based on CLSI EP07-ED3: Interference Testing in Clinical Chemistry, Third Edition. The effect of each substance was evaluated by comparing the performance of a control sample, spiked with blank solvent solution, with the test results from a test sample spiked with the potentially interfering substance at the toxic/pathological concentration based on CLSI EP37-ED1: Supplemental Tables for Interference Testing in Clinical Chemistry, First Edition, as applicable. A substance was identified as an interferent if the difference in means (or medians) between the control and test samples was outside of the allowed error (±Ea) for the i-STAT test. For an identified interferent, a dose-response was performed to determine the degree of interference as a function of the substance concentration. Table 11 contains the lists of potentially interfering substances tested and the interference results for the i-STAT CG4+ cartridge.

Page 18

Table 11: Potentially Interfering Substances and Test Concentrations for the pH, PO2, PCO2, and Lactate Tests in the i-STAT CG4+ Cartridge on the i-STAT 1 Analyzer

Substance²Substance Concentrationi-STAT TestInterference (Yes/No)Comments
mmol/L (unless specified)mg/dL (unless specified)
Acetaldehyde³0.0450.2LactateNo
Acetaminophen1.0315.6pHNo
PO2No
PCO2No
LactateNo
Acetyl Cysteine (N-Acetyl-Cysteine)0.9215.0LactateNo
Ascorbic Acid (L-Ascorbic Acid)0.2985.25LactateNo
Atracurium (Atracurium Besylate)³0.02873.57pHNo
PO2No
PCO2No
β-Hydroxybutyric Acid³6.062.46LactateNo
Bilirubin0.68440pHNo
PO2No
PCO2No
LactateNo
Bromide³ (Lithium Bromide)2.521.7LactateNo
37.5325.7LactateYesUse another method. Decreased results >10.0 mmol/L bromide.
Calcium (Calcium Chloride)5.020pHNo
PO2No
PCO2No
Dopamine (Dopamine Hydrochloride)4.06 µmol/L0.0621LactateNo
Ethanol130600pHNo
PO2No
PCO2No
Formaldehyde³0.1330.399LactateNo
Glycolic Acid³10.076.05LactateYesIncreased results >0.8 mmol/L glycolic acid.

² The compound tested to evaluate the interfering substance is presented in parenthesis.
³ The test concentration for this substance is not included in CLSI guideline EP37 1st edition.

Page 19

Substance²Substance Concentrationi-STAT TestInterference (Yes/No)Comments
mmol/L (unless specified)mg/dL (unless specified)
Hemoglobin10 g/L1000pHNo
PO2No
PCO2No
LactateNo
Hydroxyurea0.4053.08LactateNo
Ibuprofen1.0621.9pHNo
PO2No
PCO2No
Intralipid 20%N/A2684pHNo
PO2No
PCO2No
3579LactateNo
Morphine (Morphine Sodium Salt)0.02730.78pHNo
PO2No
PCO2No
Potassium (Potassium Chloride)859.6pHNo
PO2No
PCO2No
Pyruvate (Lithium Pyruvate)0.5705LactateNo
Salicylate (Lithium Salicylate)0.2072.86LactateNo
Sodium (Sodium Chloride)170993.48pHNo
PO2No
PCO2No
Thiocyanate (Lithium Thiocyanate)0.8985.22LactateNo
Thiopental1.6640.2pHNo
PO2No
PCO2No
Triglyceride16.941500pHNo
PO2No
PCO2No
LactateNo
Uric Acid1.423.5LactateNo

Page 20

ii. Other sensitivity studies

1) Oxygen Sensitivity for the i-STAT Lactate Test

The effect of oxygen on the i-STAT Lactate test in the i-STAT CG4+ cartridge on the i-STAT 1 System was evaluated with low and high oxygen levels using whole blood samples altered to four (4) lactate levels across the test reportable range. The equivalency between the high and low oxygen conditions was determined if the 95% confidence interval (CI) of the difference in means (or medians) was within the allowable error (±Ea). The study demonstrated that the i-STAT Lactate test in the i-STAT CG4+ cartridge on the i-STAT 1 System perform is insensitive to oxygen levels between 28 and 521 mmHg.

2) Altitude

The performance of the i-STAT pH, PO2, PCO2, and Lactate tests in the i-STAT CG4+ cartridge on the i STAT 1 analyzer at an altitude of approximately 10,000 feet above sea level was evaluated using whole blood samples at relevant analyte levels across the reportable range for each test. The pH, PO2, PCO2 and lactate results obtained from the i-STAT CG4+ cartridges (candidate device) were compared to the results obtained from the i-STAT CG4+ cartridges (blue) on the i-STAT 1 analyzer (K200492; comparator device) condition. Passing-Bablok regression analyses between the first replicate of the candidate device (y-axis) and mean of the comparator device (x-axis) were performed based on the CLSI EP09c-ED3: Measurement Procedure Comparison and Bias Estimation using Patient Samples– Third Edition. The results of the correlation coefficient and slope met acceptance criteria and demonstrate equivalent performance between the candidate and comparator condition at approximately 10,000 feet above sea level. The results are summarized in Table 12 below.

Table 12: Summary of Altitude Study Results

TestCorrelation Coefficient (r)Slope
r95% CISlope95% CI
pH1.000.999 to 1.0001.000.995 to 1.005
PO21.000.997 to 0.9991.031.014 to 1.049
PCO21.000.999 to 0.9990.960.949 to 0.967
Lactate1.000.999 to 1.0001.000.992 to 1.003

B. Comparison Studies

a. Method Comparison

Method comparison for arterial, venous, and capillary whole blood specimens on the i-STAT CG4+ cartridge with the i-STAT 1 System was demonstrated in studies based on CLSI EP09c-ED3: Measurement Procedure Comparison and Bias Estimation Using Patient Samples – Third Edition.

Heparinized arterial and venous whole blood specimens collected across multiple point of care sites were evaluated using i-STAT CG4+ cartridges on the i-STAT 1 analyzer against whole blood specimens tested on a comparative method. For pH and PO2, a Passing-Bablok linear regression analysis was performed using the first replicate result from the i-STAT 1 analyzer versus the singlicate result from the comparative method. For PCO2 and lactate, the

Page 21

first replicate result from the i-STAT 1 analyzer was compared to the mean result from the comparative method.

Two (2) capillary specimens collected from skin puncture with balanced heparin capillary tubes from each study subject across multiple point of care sites. One (1) tube was used to test in singlicate on the i‑STAT 1 analyzer and the other tube was used to test in singlicate on the comparative method.

The arterial, venous, and capillary data were pooled, and a Passing-Bablok linear regression analysis for pH, PO2, and PCO2 was performed using the singlicate result from the i-STAT CG4+ cartridge versus the result from the comparative method.

The method comparison results for arterial, venous, and capillary whole blood specimens are shown in Table 13. In the table, N is the number of specimens in the data set, and r is the correlation coefficient.

Table 13: Method Comparison Results for the pH, PO2, and PCO2 tests on the i-STAT CG4+ Cartridge with i-STAT 1 System

Test (units)Comparative MethodNSlopeInterceptrMedical Decision LevelBias at Medical Decision Level
Arterial/VenousCapillary
pH (pH units)RAPIDPoint 500/500eRAPIDPoint 500/500e5511.00-0.010.987.30-0.0080
7.35-0.0080
7.40-0.0080
PO2 (mmHg)RAPIDPoint 500/500eRAPIDPoint 500/500e5571.01-1.290.9930-0.9
45-0.7
60-0.6
PCO2 (mmHg)i-STAT G3+i-STAT G3+4751.03-0.300.9935.00.67
45.00.95
50.01.08
70.01.64

The arterial and venous data were pooled, and a Passing-Bablok linear regression analysis for Lactate was performed using the singlicate result from the i-STAT CG4+ cartridge versus the result from the comparative method. The method comparison results for arterial and venous whole blood specimens are shown in Table 14.

Page 22

Table 14: Method Comparison Results for the Lactate test on the i-STAT CG4+ Cartridge with i-STAT 1 System

Test (units)Comparative MethodNSlopeInterceptrMedical Decision LevelBias at Medical Decision Level
Lactate (mmol/L)i-STAT CG4+ (K200492)3451.000.001.005.00-0.140

The method comparison results for capillary whole blood specimens only for pH, PO2, and PCO2 are shown in Table 15.

Table 15: Results for i-STAT CG4+ Cartridge with i-STAT 1 System- Native and Contrived Capillary Specimens

Test (units)NSlopeInterceptrXminXmax
pH (pH units)1931.01-0.080.976.6077.709
PO2 (mmHg)1921.08-5.470.9914.0554.0
PCO2 (mmHg)1841.05-0.540.988.9125.8

Bias at the medical decision levels for native capillary whole blood specimens only for pH, PO2, and PCO2 are shown in Table 16.

Table 16: Results for i-STAT CG4+ Cartridge with i-STAT 1 System- Native Capillary Specimens Bias at Medical Decision Levels

Test (units)NRange MinRange MaxMedical Decision LevelBias
Estimate95% CI
pH (pH units)1787.2597.5317.30-0.0166(-0.0341, 0.0007)
7.35-0.0104(-0.0207, 0.0000)
7.40-0.0041(-0.0095, 0.0013)
PO2 (mmHg)1783113930-3.3(-6.1, -0.4)
45-2.1(-3.7, -0.4)
60-0.9(-2.1, 0.0)
PCO2 (mmHg)17523.178.635.01.18(0.48, 1.75)
45.01.84(1.04, 2.46)
50.02.17(1.15, 3.03)
70.03.49(1.40, 5.49)

b. Matrix Equivalence

A matrix equivalence study was conducted to evaluate the performance of the i-STAT pH, PO2, PCO2 and Lactate tests in the i STAT CG4+ cartridge on the i STAT 1 System using non-

Page 23

anticoagulated arterial and venous whole blood specimens. The study design and analysis method were based on recommendations from CLSI EP35: Assessment of Equivalence or Suitability of Specimen Types for Medical Laboratory Measurement Procedures, First Edition. The matrix equivalence of each test in the i-STAT CG4+ cartridge was assessed by comparing arterial or venous whole blood specimens collected without anticoagulant (candidate specimen type) to samples collected with balanced heparin or lithium heparin anticoagulant (primary specimen type). Each specimen was tested in duplicate using two (2) i-STAT CG4+ cartridges with two (2) i-STAT 1 analyzers. A Passing-Bablok linear regression analysis was performed using the first replicate result from the candidate (y-axis) versus the first replicate result from the primary specimen (x-axis). The regression analysis results are summarized in Table 17. In the table, N is the number of specimens in the data set, and r is the correlation coefficient.

Table 17: Matrix Equivalence Results

Test (units)NCandidate Specimen RangePrimary Specimen RangerSlopeIntercept
pH (pH units)2287.176-7.5677.192-7.5490.961.000.00
PO2 (mmHg)22815-19715-1970.991.01-0.53
PCO2 (mmHg)22826.3-77.426.8-77.00.971.03-1.58
Lactate (mmol/L)2890.31-19.420.30-19.431.001.00-0.02

8. Conclusion

The results of these studies demonstrate that performance of the i-STAT pH, PO2, PCO2, and Lactate tests in the i-STAT CG4+ cartridge with the i-STAT 1 System are substantially equivalent to the predicate device.