K Number
K243922
Device Name
Revogene
Date Cleared
2025-03-20

(90 days)

Product Code
Regulation Number
862.2570
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Revogene® instrument is intended for in vitro diagnostic (IVD) use in performing nucleic acid testing of specific IVD assays in clinical laboratories. Revogene is capable of automated lysis and dilution of samples originating from various clinical specimen types. Revogene performs automated amplification and detection of target nucleic acid sequences by fluorescence-based real-time PCR.

Device Description

The Revogene is a PCR instrument that automates lysis and dilution of samples, followed by nucleic acid amplification, and detection of target sequences by fluorescence-based real-time PCR. Revogene runs are orchestrated by a combination of software, firmware and instrument control protocol that ensures the adequate combination times and temperatures for sample homogenization and PCR analysis. The Revogene instrument acquires fluorescence signals generated during amplification. The signals are then interpreted by the system using embedded calculation algorithms.

The Revogene requires the use of a 'PIE', i.e., an assay-specific cartridge to which a patient sample is added. The PIE contains the reagents needed to process a sample and to perform a PCR amplification. When the number of assay PIEs to be run is lower than eight, the user fills empty spaces with "MOCK PIE", which are cartridges that simulate the presence of an assay PIE to confer thermal and rotational balance.

The Revogene instrument subject of this Premarket Notification is substantially equivalent to the Revogene instrument cleared under K222779. Meridian is submitting this 510(k) Premarket Notification to implement a photomultiplier tube (PMT) cooling system. This cooling system keeps the PMT environment at a temperature that prevents the appearance of fluorescence glitches, which may stop the Revogene instrument

AI/ML Overview

The provided document is a 510(k) Premarket Notification for a modified medical device, the Revogene instrument. It focuses on the changes made to an existing device (K222779) and its substantial equivalence to the predicate device.

The document does not contain information about acceptance criteria or a detailed study proving the device meets specific acceptance criteria, as one might find in a clinical trial report for an initial device clearance.

Instead, it describes the performance characteristics of functional testing conducted to demonstrate that the modifications (PMT cooling system and Windows 10 upgrade) do not adversely affect the device's performance compared to the predicate. The goal of this submission is to show substantial equivalence, not to establish new performance acceptance criteria.

Therefore, I cannot provide a table of acceptance criteria and reported device performance in the traditional sense, nor can I answer many of your specific questions about study design, sample sizes, ground truth establishment, or expert adjudication, as this information is not present in the provided text.

However, I can extract the available information regarding the functional testing that was performed to support the substantial equivalence claim.


Summary of Available Information on Device Performance and Testing:

1. A table of (implied) acceptance criteria and the reported device performance

The document does not explicitly state quantitative acceptance criteria. Instead, it describes general observations and conclusions from functional testing. The implicit acceptance criterion is "no statistically significant differences" and "operates as expected and yields expected assay results."

Performance CharacteristicImplicit Acceptance Criterion (based on "no statistically significant differences")Reported Device Performance (Modified Device vs. Predicate)
Positivity/Negativity rates of assaysNo statistically significant differencesNo statistically significant differences observed
Unresolved result rates of assaysNo statistically significant differencesNo statistically significant differences observed
Indeterminate result rates of assaysNo statistically significant differencesNo statistically significant differences observed
Mean Ct values of assaysNo statistically significant differencesNo statistically significant differences observed
Occurrence of PMT glitchesReduced occurrence compared to predicateActivation of PMT cooling system resulted in lower glitch amplitude. No run triggered a PMT error signal upon activation.
Amplitude of PMT glitchesReduced amplitude compared to predicateActivation of PMT cooling system resulted in lower glitch amplitude. No run triggered a PMT error signal upon activation.
Overall operation with Windows 10 & upgraded softwareOperates as expected and yields expected assay resultsOperates as expected and yields expected assay results

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: The document states "contrived and negative samples in relevant clinical matrix using the following assays...". However, it does not specify the number of samples or runs used for this functional testing.
  • Data Provenance: Not explicitly stated, but given it's a regulatory submission by a US company, the testing would typically be conducted according to established protocols within their R&D or QA departments. It is retrospective relative to the design changes, but the testing itself is performed to support the new device version. No information on country of origin of data beyond the manufacturer's location.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • This information is not provided as the testing described is functional performance testing of the instrument, not typically involving expert interpretation of patient samples for ground truth establishment. The "ground truth" here is the expected performance of control samples within the assays.

4. Adjudication method for the test set

  • This is not applicable/provided. The testing focuses on the instrument's functional output (e.g., Ct values, glitch occurrence) rather than interpretation of results that would require adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

  • No, this was not done. The device is an instrumentation for clinical multiplex test systems, meaning it processes samples and detects nucleic acids. It does not output images or data that require human readers for interpretation in the way an AI diagnostic imaging device would. Therefore, an MRMC study is not relevant to this type of device or its modifications.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • The device itself is a standalone instrument that performs automated lysis, dilution, amplification, and detection. The "algorithm" here refers to the embedded calculation algorithms within the system that interpret fluorescence signals to determine results. The functional testing described is a form of standalone performance evaluation for the modified instrument. There is no human-in-the-loop component mentioned for the actual nucleic acid detection and interpretation process of the instrument.

7. The type of ground truth used

  • The ground truth for the functional testing appears to be based on the expected outcomes from known contrived and negative samples when run with specific IVD assays (Revogene® Strep A, Revogene® Carba C and Revogene® SARS-CoV-2). Essentially, the "ground truth" is the established performance of the assays themselves on control materials, and the instrument must correctly process these, showing no statistical degradation from the predicate.

8. The sample size for the training set

  • This information is not provided and is generally not applicable in the context of hardware modifications to an existing IVD instrument as described. The "training set" concept is typically relevant for machine learning algorithms, which are not detailed here beyond "embedded calculation algorithms" that likely leverage established PCR physics and signal processing rather than iterative machine learning training.

9. How the ground truth for the training set was established

  • This information is not provided as there is no mention of a traditional "training set" in the machine learning sense. The established performance of the assays with known control materials serves as the reference for evaluating the modified instrument.

{0}------------------------------------------------

March 20, 2025

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Meridian Bioscience, Inc. Whitney Rooker Senior Regulatory Affairs Specialist 3471 River Hills Drive Cincinnati, Ohio 45244

Re: K243922

Trade/Device Name: Revogene Regulation Number: 21 CFR 862.2570 Regulation Name: Instrumentation For Clinical Multiplex Test Systems Regulatory Class: Class II Product Code: OOI Dated: December 19, 2024 Received: December 20, 2024

Dear Whitney Rooker:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve

{1}------------------------------------------------

changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ribhi Shawar -S

Ribhi Shawar, Ph.D. (ABMM) Branch Chief General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics OPEQ: Office of Product Evaluation and Quality CDRH: Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K243922

Device Name Revogene

Indications for Use (Describe)

The Revogene® instrument is intended for in vitro diagnostic (IVD) use in performing nucleic acid testing of specific IVD assays in clinical laboratories. Revogene is capable of automated lysis and dilution of samples originating from various clinical specimen types. Revogene performs automated amplification and detection of target nucleic acid sequences by fluorescence-based real-time PCR.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Meridian Bioscience. The word "meridian" is in a dark blue sans-serif font, with a curved line underneath. To the right of "meridian" is the word "BIOSCIENCE" in a smaller, sans-serif font, with the trademark symbol next to it.

Revogene®

510(k) Summary Meridian Bioscience

A. Applicant Information

Date of preparation:March 12, 2025
Submitter Information:Meridian Bioscience, Inc.3471 River Hills DriveCincinnati, Ohio 45244
Contact Person:Whitney RookerSenior Regulatory Affairs SpecialistMeridian Bioscience, Inc.Tel: 513.310.4134Email: Whitney.Rooker@meridianbioscience.com

B. Proprietary and Established Names

Revogene®

C. Regulatory Information

Trade Name:Revogene®
Common name:Revogene instrument
K number:K243922
Regulation Number:21 CFR 862.2570
Regulation Name:Instrumentation for clinical multiplex systems
Regulatory Classification:Class II
Product Code:OOI - Real-time nucleic acid amplification system
Panel:Clinical Chemistry (75)

D. Purpose of Submission

This 510(k) Premarket Notification is meant to obtain FDA clearance for a modified version of the Revogene instrument cleared under K222779. A cooling system to reduce the temperature around the photomultiplier tube (PMT) and associated firmware updates are added to mitigate the risk of PMT glitches and Revogene blocking. In addition, the Operating System was upgraded to Windows 10. The Revogene System Software was upgraded to a

Life discovered. Life diagnosed.

{4}------------------------------------------------

version compatible with Windows 10. These changes do not affect the device's intended use nor alter the device's fundamental technological characteristics.

E. Intended Use

Intended Use:The Revogene® instrument is intended for in vitro diagnostic(IVD) use in performing nucleic acid testing of specific IVDassays in clinical laboratories. Revogene is capable of automatedlysis and dilution of samples originating from various clinicalspecimen types. Revogene performs automated amplification anddetection of target nucleic acid sequences by fluorescence-basedreal-time PCR.
Indications for Use:See Intended Use statement.
Special Conditions forUse Statement:For prescription use onlyFor in vitro diagnostic use only
Special InstrumentRequirements:MOCK PIE (optional)

F. Device Description

The Revogene is a PCR instrument that automates lysis and dilution of samples, followed by nucleic acid amplification, and detection of target sequences by fluorescence-based real-time PCR. Revogene runs are orchestrated by a combination of software, firmware and instrument control protocol that ensures the adequate combination times and temperatures for sample homogenization and PCR analysis. The Revogene instrument acquires fluorescence signals generated during amplification. The signals are then interpreted by the system using embedded calculation algorithms.

The Revogene requires the use of a 'PIE', i.e., an assay-specific cartridge to which a patient sample is added. The PIE contains the reagents needed to process a sample and to perform a PCR amplification. When the number of assay PIEs to be run is lower than eight, the user fills empty spaces with "MOCK PIE", which are cartridges that simulate the presence of an assay PIE to confer thermal and rotational balance.

The Revogene instrument subject of this Premarket Notification is substantially equivalent to the Revogene instrument cleared under K222779. Meridian is submitting this 510(k) Premarket Notification to implement a photomultiplier tube (PMT) cooling system. This cooling system keeps the PMT environment at a temperature that prevents the appearance of fluorescence glitches, which may stop the Revogene instrument

{5}------------------------------------------------

G. Substantial Equivalence Information

Predicate device:Revogene®
Predicate Device Number:K222779

Comparison with Predicate:

Modified DevicePredicate Device
ItemRevogene®(K243922, subject of 510(k) PremarketNotification)Revogene® (K222779)
Similarities
ClassificationClass IISame
Intended UseThe Revogene® instrument is intended forin vitro diagnostic (IVD) use in performingnucleic acid testing of specific IVD assaysin clinical laboratories. Revogene iscapable of automated lysis and dilution ofsamples originating from various clinicalspecimen types. Revogene performsautomated amplification and detection oftarget nucleic acid sequences byfluorescence-based real-time PCR.Same
Sample PreparationMethodAutomated cell lysis, DNAamplificationand DNA detectionSame
Mode of OperationReal-time Polymerase chain reaction withfluorogenic detection of amplified DNASame
Sample analysis andresult determinationCombination of software, instrumentcontrol protocols and assay definition filesdeveloped and determined by MeridianSame
Automatic AssayYes-result interpretationSame
Sample identificationThe instrument has two barcode readers toidentify reagents and patient specimens. Itprovides traceability of the sampleidentification (ID) to the PIE ID, SampleBuffer Tube ID, and assay ID.Same
Internal Process ControlDNA assaysEach PIE contains an internal processcontrol (PrC) that controls for amplificationinhibition, assay reagents, and sampleprocessing effectiveness.Same
Internal Process ControlRNA assaysEach PIE contains an Internal Control (IC)that controls for amplification inhibition,and assay reagents effectiveness.Same
Modified DevicePredicate Device
ItemRevogene®(K243922, subject of 510(k) PremarketNotification)Sample processing is monitored by aMicrofluidic Control (MFC).Revogene® (K222779)
External ControlMaterials available commercially but notrequired to run the testSame
DNA ExtractionCell lysisSame
Specimens per runProcesses and analyzes up to 8 specimensper run (8 PIEs)Same
Assay CartridgeOne sample per PIESame
Single UsePIE can be used only onceSame
Instrument OpticalChannelsContains 4 optical channelsSame
Instrument CalibrationThe instrument is factory calibrated by themanufacturer and will undergo performancequalification testing on-site during annualpreventive maintenance. If qualificationtesting results determine significant drift,the instrument will be returned to themanufacturer for recalibration.Same
Temperature ControlHeating element/fan assembly and coolingfan for PCR cycling.Same
Differences
Operating SystemWindows® 10 IoT Enterprise (Windows10)- OS image based on Windows 10- Update Revogene System Software toensure:- compatibility with Windows 10.- compatibility with Third Party/Off theShelf Software- Update the Off the Shelf Softwaremanaging the Revogene database.- Deactivation of internet explorer toprevent unintended access to internet.- Addition of a "NoAutoUpdate" key inWindows registry to prevent automaticupdates from being installedWindows® Embedded Standard7 (Windows 7)
ItemModified DevicePredicate Device
Temperature ControlRevogene®(K243922, subject of 510(k) PremarketNotification)Revogene® (K222779)
PMT cooling systemNo PMT cooling system

{6}------------------------------------------------

{7}------------------------------------------------

H. Performance Characteristics

Functional testing has been conducted using contrived and negative samples in relevant clinical matrix using the following assays, which are representative of the Revogene assays currently on the US market: Revogene® Strep A, Revogene® Carba C and Revogene® SARS-CoV-2 assays. No statistically significant differences were observed between the Revogene instruments with and without PMT cooling system for the Positivity/Negativity rates, the Unresolved and Indeterminate results rates and the mean Ct values.

Testing was also carried out to demonstrate that the cooling system reduced the occurrence and amplitude of glitches. The activation of the PMT cooling system resulted in lower glitch amplitude compared to the amplitude documented on instruments run without cooling. No run triggered the prompting of a PMT error signal.

Finally, Revogene equipped with Windows 10 and with upgraded Revogene System Software operates as expected and yields expected assays results.

I. Proposed Labeling

The labeling is sufficient, and it satisfies the requirements of 21 CFR Part 809.

J. Conclusion

The submitted information demonstrates that the Revogene instrument equipped with the PMT cooling system is safe, effective, and substantially equivalent to the legally marketed device.

§ 862.2570 Instrumentation for clinical multiplex test systems.

(a)
Identification. Instrumentation for clinical multiplex test systems is a device intended to measure and sort multiple signals generated by an assay from a clinical sample. This instrumentation is used with a specific assay to measure multiple similar analytes that establish a single indicator to aid in diagnosis. Such instrumentation may be compatible with more than one specific assay. The device includes a signal reader unit, and may also integrate reagent handling, hybridization, washing, dedicated instrument control, and other hardware components, as well as raw data storage mechanisms, data acquisition software, and software to process detected signals.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems.” See § 862.1(d) for the availability of this guidance document.