K Number
K243876
Manufacturer
Date Cleared
2025-04-25

(128 days)

Product Code
Regulation Number
880.2800
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The integrating indicator is designed to chemically react over time with the critical parameters of a steam sterilization cycle within a specified tolerance. The integrating indicator strip is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:

Steam Sterilization Cycles:

  • 250°F/121°C, 30 minutes Gravity
  • 270°F/132°C, 4 minutes Dynamic Air Removal
  • 270°F/132°C, 5 minutes Dynamic Air Removal
  • 270°F/132°C, 6 minutes Dynamic Air Removal
  • 270°F/132°C, 7 minutes Dynamic Air Removal
  • 270°F/132°C, 8 minutes Dynamic Air Removal
  • 270°F/132°C, 9 minutes Dynamic Air Removal
  • 270°F/132°C, 10 minutes Dynamic Air Removal
  • 270°F/132°C, 15 minutes Gravity
  • 273°F/134°C, 4 minutes Dynamic Air Removal
  • 275°F/135°C, 3 minutes Dynamic Air Removal
  • 275°F/135°C, 10 minutes Gravity

Steam Sterilization Cycles (IUSS):

  • 270°F/132°C, 4 minutes Dynamic Air Removal
  • 270°F/132°C, 3 minutes Gravity
  • 270°F/132°C, 10 minutes Gravity
  • 275°F/135°C, 3 minutes Dynamic Air Removal
  • 275°F/135°C, 3 minutes Gravity
  • 275°F/135°C, 10 minutes Gravity
Device Description

The VERIFY STEAM Integrating Indicator is a single use device used by healthcare providers to monitor steam sterilization cycles. The VERIFY STEAM Integrating Indicator is included in a pack, pouch, container, tray or other containment device in a steam sterilizer and the load is processed in accordance with the sterilizer's manufacturer's directions. Prior to the use of the processed items, the integrator is observed. If the dark bar on the device enters the ACCEPT (OK) window, the integrator is read as a PASS to indicate that the steam sterilization criteria for the cycle have been met. If the dark bar on the device does not enter the ACCEPT (OK), the integrator is read as a FAIL, indicating that sufficient steam sterilization criteria has not been met and processed materials should be subjected to another steam sterilization cycle prior to use.

AI/ML Overview

Here's an analysis of the provided FDA 510(k) clearance letter, focusing on the acceptance criteria and the study proving the device meets them:

The VERIFY STEAM Integrating Indicator (K243876) is a chemical indicator designed to monitor steam sterilization cycles. It is intended to be placed within containment devices (packs, pouches, etc.) during sterilization to independently verify that critical parameters have been met. The device is considered to "pass" if a dark bar on the indicator enters the "ACCEPT (OK)" window.

Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Stated Value Testing per ANSI/AAMI/ISO 11140-1:2014 section 11• All integrators processed will demonstrate no visible deterioration of the film• All integrators processed in a dry heat cycle will demonstrate a failing result• The Stated Values for time at 121 °C and 135 °C shall be specified and shall not be less than 16.5 min at 121 °C and 1.2 min at 135 °C• The temperature coefficient shall not be less than 10°C and not more than 27°C.• The correlation coefficient shall not be less than 0.9PASS
Simulated Use Testing• All integrators processed in a full cycle will demonstrate a passing result with no physical deterioration of the film• All integrators processed in a partial cycle will demonstrate a passing result with no physical deterioration of the film• No integrators will exhibit a leak in the viewing windowPASS
Stability Study• All CIs processed in pass/full cycles in the BIER vessel will demonstrate pass results and shall not be under minimum specification limits as per the standard ISO 11140-1• All CIs processed in fail/partial cycles in BIER vessel will demonstrate fail results• All CIs processed for dry heat testing will demonstrate fail results.PASS
Comparison Testing with Biological Indicator• The Chemical Indicator does not reach its endpoint before the biological indicator is inactivated.PASS

Study Details

The provided document describes non-clinical testing performed to evaluate the performance of the VERIFY STEAM Integrating Indicator and demonstrate its substantial equivalence to a predicate device (K213412).

1. Sample sized used for the test set and the data provenance:

  • The document does not specify the exact sample sizes (number of indicators) used for each individual test (Stated Value, Simulated Use, Stability, Comparison with Biological Indicator).
  • The data provenance is not explicitly stated in terms of country of origin or whether it was retrospective or prospective. However, given the context of a 510(k) submission for a medical device, it is highly likely that these were prospective, laboratory-based studies conducted by the manufacturer (STERIS Corporation, Franklin Park, IL, USA) to evaluate the device's performance under controlled conditions.

2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This device is a chemical indicator that produces a visual change (dark bar entering an "ACCEPT (OK)" window) based on exposure to sterilization parameters. The "ground truth" for these tests is established by the controlled sterilization cycle conditions (temperature, time, presence of steam) and the objective physical change of the indicator, rather than subjective expert interpretation.
  • Therefore, no human experts were used to establish the ground truth for the test set in the way radiologists or pathologists would interpret an image. The performance is assessed against defined physical and chemical criteria.

3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • As the "ground truth" is determined by objective physical and chemical changes under controlled conditions, and the outcome (pass/fail) is a direct observation of the indicator's state, no adjudication method involving human experts was necessary for the test set.

4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret results, sometimes with AI assistance. The VERIFY STEAM Integrating Indicator is a simple chemical indicator, not an AI-powered diagnostic tool, and its primary function is to react chemically, not to be "read" in a complex interpretive sense by multiple human readers.

5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • This question is not applicable in the context of a chemical indicator. The device is entirely standalone in its function – it reacts chemically without any human intervention or an algorithm. Its "performance" is its physical color change, which is then observed by a human. There is no separate "algorithm" to evaluate.

6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The ground truth for these tests is based on objective physical and chemical properties and predefined standards. Specifically:
    • Controlled Sterilization Parameters: The cycles are run at precisely defined temperatures and times, and the presence of steam is controlled. The "truth" is whether these parameters were met or not.
    • ANSI/AAMI/ISO 11140-1:2014: This international standard defines the performance requirements for chemical indicators, providing the objective benchmarks against which the device is evaluated.
    • Biological Indicator Inactivation: For comparison testing, the inactivation of a biological indicator (a highly resistant microorganism) serves as an objective "gold standard" for sterilization efficacy.

7. The sample size for the training set:

  • The document does not mention a "training set" in the context of device development. This concept is typically associated with machine learning or AI models. For a chemical indicator, the "development" involves chemical formulation, material science, and engineering to achieve the desired reaction kinetics and visual endpoint. There isn't a "training set" in the computational sense.

8. How the ground truth for the training set was established:

  • As there is no "training set" in the conventional sense for this type of device, this question is not applicable. The development of such an indicator would involve extensive research and development in chemistry and material science, where performance is iteratively optimized against the physical/chemical principles of sterilization and the requirements of relevant standards.

FDA 510(k) Clearance Letter - VERIFY STEAM Integrating Indicator

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

April 25, 2025

STERIS Corporation
Logan Persons
Regulatory Affairs Specialist
5960 Heisley Rd
Mentor, Ohio 44060

Re: K243876
Trade/Device Name: VERIFY STEAM Integrating Indicator
Regulation Number: 21 CFR 880.2800
Regulation Name: Sterilization process indicator
Regulatory Class: Class II
Product Code: JOJ
Dated: December 17, 2024
Received: December 18, 2024

Dear Logan Persons:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K243876 - Logan Persons Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K243876 - Logan Persons Page 3

Sincerely,

Stephen A. Anisko -S Digitally signed by Stephen A. Anisko -S Date: 2025.04.25 14:59:45 -04'00'

for: Christopher K. Dugard
Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K243876

Device Name
VERIFY STEAM Integrating Indicator

Indications for Use (Describe)

The integrating indicator is designed to chemically react over time with the critical parameters of a steam sterilization cycle within a specified tolerance. The integrating indicator strip is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:

Steam Sterilization Cycles:

  • 250°F/121°C, 30 minutes Gravity
  • 270°F/132°C, 4 minutes Dynamic Air Removal
  • 270°F/132°C, 5 minutes Dynamic Air Removal
  • 270°F/132°C, 6 minutes Dynamic Air Removal
  • 270°F/132°C, 7 minutes Dynamic Air Removal
  • 270°F/132°C, 8 minutes Dynamic Air Removal
  • 270°F/132°C, 9 minutes Dynamic Air Removal
  • 270°F/132°C, 10 minutes Dynamic Air Removal
  • 270°F/132°C, 15 minutes Gravity
  • 273°F/134°C, 4 minutes Dynamic Air Removal
  • 275°F/135°C, 3 minutes Dynamic Air Removal
  • 275°F/135°C, 10 minutes Gravity

Steam Sterilization Cycles (IUSS):

  • 270°F/132°C, 4 minutes Dynamic Air Removal
  • 270°F/132°C, 3 minutes Gravity
  • 270°F/132°C, 10 minutes Gravity
  • 275°F/135°C, 3 minutes Dynamic Air Removal
  • 275°F/135°C, 3 minutes Gravity
  • 275°F/135°C, 10 minutes Gravity

Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D) ☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

Page 5

STERIS Corporation5960 Heisley RoadMentor, OH 44060-1834 USA440-354-2600

April 24, 2025 K243876 Page 1 of 7

510(k) Summary (K243876)

For

VERIFY STEAM Integrating Indicator

Sponsor Facility
STERIS Corporation
5960 Heisley Road
Mentor, OH 44060
Phone: (440) 354-2600
Fax: (440) 357-9198

Manufacturing Facility
STERIS Corporation Franklin Park
11457 Melrose Ave.
Franklin Park, IL 60131
Phone: (847) 455-2881

Contact
Logan Persons
Regulatory Affairs Specialist
Phone: (440) 392-7325
Fax: (440) 357-9198
Email: Logan_Persons@steris.com

Updated Date:
April 24, 2025

Page 6

STERIS Corporation5960 Heisley RoadMentor, OH 44060-1834 USA440-354-2600

April 24, 2025 K243876 Page 2 of 7

1. Device Name

Trade Name: VERIFY STEAM Integrating Indicator
Common/usual Name: Chemical Indicator
Device Classification: Class II
Classification Name: Indicator, physical/chemical sterilization process [21 CFR 880.2800(b)]
Product Code: JOJ

2. Predicate Device

Proprietary Name: VERIFY STEAM Integrating Indicator
Common/usual Name: Chemical Indicator
Classification Name: Indicator, physical/chemical sterilization process [21 CFR 880.2800(b)]
Product Code: JOJ
510(k) Submitter/Holder: STERIS Corporation
510(k) Number: K213412

3. Description of Device

The VERIFY STEAM Integrating Indicator is a single use device used by healthcare providers to monitor steam sterilization cycles. The VERIFY STEAM Integrating Indicator is included in a pack, pouch, container, tray or other containment device in a steam sterilizer and the load is processed in accordance with the sterilizer's manufacturer's directions. Prior to the use of the processed items, the integrator is observed. If the dark bar on the device enters the ACCEPT (OK) window, the integrator is read as a PASS to indicate that the steam sterilization criteria for the cycle have been met. If the dark bar on the device does not enter the ACCEPT (OK), the integrator is read as a FAIL, indicating that sufficient steam sterilization criteria has not been met and processed materials should be subjected to another steam sterilization cycle prior to use.

4. Intended Use/Indications for Use

The integrating indicator is designed to chemically react over time with the critical parameters of a steam sterilization cycle within a specified tolerance. The integrating indicator strip is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:

Steam Sterilization Cycles:

  • 250°F/121°C, 30 minutes Gravity
  • 270°F/132°C, 4 minutes Dynamic Air Removal
  • 270°F/132°C, 5 minutes Dynamic Air Removal
  • 270°F/132°C, 6 minutes Dynamic Air Removal
  • 270°F/132°C, 7 minutes Dynamic Air Removal
  • 270°F/132°C, 8 minutes Dynamic Air Removal
  • 270°F/132°C, 9 minutes Dynamic Air Removal
  • 270°F/132°C, 10 minutes Dynamic Air Removal
  • 270°F/132°C, 15 minutes Gravity
  • 273°F/134°C, 4 minutes Dynamic Air Removal

Page 7

STERIS Corporation5960 Heisley RoadMentor, OH 44060-1834 USA440-354-2600

April 24, 2025 K243876 Page 3 of 7

  • 275°F/135°C, 3 minutes Dynamic Air Removal
  • 275°F/135°C, 10 minutes Gravity

Steam Sterilization Cycles (IUSS):

  • 270°F/132°C, 4 minutes Dynamic Air Removal
  • 270°F/132°C, 3 minutes Gravity
  • 270°F/132°C, 10 minutes Gravity
  • 275°F/135°C, 3 minutes Dynamic Air Removal
  • 275°F/135°C, 3 minutes Gravity
  • 275°F/135°C, 10 minutes Gravity

5. Summary of Technological Characteristics

A comparison of technological characteristics are summarized in Table 1.

Table 1. Summary of SCBI Physical Description and Technological Properties

FeatureVERIFY STEAM Integrating Indicator (K243876)VERIFY STEAM Integrating Indicator (K213412)Comparison
Intended UseThe integrating indicator is designed to chemically react over time with the critical parameters of a steam sterilization cycle within a specified tolerance. The integrating indicator strip is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:Steam Sterilization Cycles:• 250°F/121°C, 30 minutes Gravity• 270°F/132°C, 4 minutes Dynamic Air Removal• 270°F/132°C, 5 minutes Dynamic Air Removal• 270°F/132°C, 6 minutes Dynamic Air Removal• 270°F/132°C, 7 minutes Dynamic Air Removal• 270°F/132°C, 8 minutes Dynamic Air RemovalThe integrating indicator is designed to chemically react over time with the critical parameters of a steam sterilization cycle within a specified tolerance. The integrating indicator strip is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:Steam Sterilization Cycles:• 250°F/121°C, 30 minutes Gravity• 270°F/132°C, 4 minutes Dynamic Air Removal• 270°F/132°C, 5 minutes Dynamic Air Removal• 270°F/132°C, 6 minutes Dynamic Air Removal• 270°F/132°C, 7 minutes Dynamic Air Removal• 270°F/132°C, 8 minutes Dynamic Air RemovalIdentical

Page 8

STERIS Corporation5960 Heisley RoadMentor, OH 44060-1834 USA440-354-2600

April 24, 2025 K243876 Page 4 of 7

FeatureVERIFY STEAM Integrating Indicator (K243876)VERIFY STEAM Integrating Indicator (K213412)Comparison
• 270°F/132°C, 9 minutes Dynamic Air Removal• 270°F/132°C, 10 minutes Dynamic Air Removal• 270°F/132°C, 15 minutes Gravity• 273°F/134°C, 4 minutes Dynamic Air Removal• 275°F/135°C, 3 minutes Dynamic Air Removal• 275°F/135°C, 10 minutes GravitySteam Sterilization Cycles (IUSS):• 270°F/132°C, 4 minutes Dynamic Air Removal• 270°F/132°C, 3 minutes Gravity• 270°F/132°C, 10 minutes Gravity• 275°F/135°C, 3 minutes Dynamic Air Removal• 275°F/135°C, 3 minutes Gravity• 275°F/135°C, 10 minutes Gravity• 270°F/132°C, 9 minutes Dynamic Air Removal• 270°F/132°C, 10 minutes Dynamic Air Removal• 270°F/132°C, 15 minutes Gravity• 273°F/134°C, 4 minutes Dynamic Air Removal• 275°F/135°C, 3 minutes Dynamic Air Removal• 275°F/135°C, 10 minutes GravitySteam Sterilization Cycles (IUSS):• 270°F/132°C, 4 minutes Dynamic Air Removal• 270°F/132°C, 3 minutes Gravity• 270°F/132°C, 10 minutes Gravity• 275°F/135°C, 3 minutes Dynamic Air Removal• 275°F/135°C, 3 minutes Gravity• 275°F/135°C, 10 minutes Gravity
Device Design – componentsBacking material with embossed cavity containing temperature sensitive chemical and coloring dye, wicking strip, covered with laminated paper containing labeling and windows.Backing material with embossed cavity containing temperature sensitive chemical and coloring dye, wicking strip, covered with laminated paper containing labeling and windows.Similar. Plastic layer was changed
Indicator agentProprietary formulationProprietary formulationIdentical
Sterilization method and cyclesSteam Sterilization Cycles:• 250°F/121°C, 30 minutes Gravity• 270°F/132°C, 4 minutes Dynamic Air Removal• 270°F/132°C, 5 minutes Dynamic Air RemovalSteam Sterilization Cycles:• 250°F/121°C, 30 minutes Gravity• 270°F/132°C, 4 minutes Dynamic Air Removal• 270°F/132°C, 5 minutes Dynamic Air RemovalIdentical

Page 9

STERIS Corporation5960 Heisley RoadMentor, OH 44060-1834 USA440-354-2600

April 24, 2025 K243876 Page 5 of 7

FeatureVERIFY STEAM Integrating Indicator (K243876)VERIFY STEAM Integrating Indicator (K213412)Comparison
• 270°F/132°C, 6 minutes Dynamic Air Removal• 270°F/132°C, 7 minutes Dynamic Air Removal• 270°F/132°C, 8 minutes Dynamic Air Removal• 270°F/132°C, 9 minutes Dynamic Air Removal• 270°F/132°C, 10 minutes Dynamic Air Removal• 270°F/132°C, 15 minutes Gravity• 273°F/134°C, 4 minutes Dynamic Air Removal• 275°F/135°C, 3 minutes Dynamic Air Removal• 275°F/135°C, 10 minutes GravitySteam Sterilization Cycles (IUSS):• 270°F/132°C, 4 minutes Dynamic Air Removal• 270°F/132°C, 3 minutes Gravity• 270°F/132°C, 10 minutes Gravity• 275°F/135°C, 3 minutes Dynamic Air Removal• 275°F/135°C, 3 minutes Gravity• 275°F/135°C, 10 minutes Gravity• 270°F/132°C, 6 minutes Dynamic Air Removal• 270°F/132°C, 7 minutes Dynamic Air Removal• 270°F/132°C, 8 minutes Dynamic Air Removal• 270°F/132°C, 9 minutes Dynamic Air Removal• 270°F/132°C, 10 minutes Dynamic Air Removal• 270°F/132°C, 15 minutes Gravity• 273°F/134°C, 4 minutes Dynamic Air Removal• 275°F/135°C, 3 minutes Dynamic Air Removal• 275°F/135°C, 10 minutes GravitySteam Sterilization Cycles (IUSS):• 270°F/132°C, 4 minutes Dynamic Air Removal• 270°F/132°C, 3 minutes Gravity• 270°F/132°C, 10 minutes Gravity• 275°F/135°C, 3 minutes Dynamic Air Removal• 275°F/135°C, 3 minutes Gravity• 275°F/135°C, 10 minutes Gravity
Mechanism of actionProprietaryProprietaryIdentical
Endpoint specificationThe end point is determined by the migration of the steam sensitive dye to an area marked ACCEPT (OK) on the indicator. Endpoint is reached at the stated value (SV) for each claimed temperature. Endpoint is notThe end point is determined by the migration of the steam sensitive dye to an area marked ACCEPT (OK) on the indicator. Endpoint is reached at the stated value (SV) for each claimed temperature. Endpoint is notIdentical

Page 10

STERIS Corporation5960 Heisley RoadMentor, OH 44060-1834 USA440-354-2600

April 24, 2025 K243876 Page 6 of 7

FeatureVERIFY STEAM Integrating Indicator (K243876)VERIFY STEAM Integrating Indicator (K213412)Comparison
reached at the stated value -15% time and/or -1°C.reached at the stated value -15% time and/or -1°C.
Comparison of integrator stated values at biological indicator growth-negative cycle conditionsIntegrator does not reach endpoint before the biological indicator is inactivated.Integrator does not reach endpoint before the biological indicator is inactivated.Identical
Shelf life22 Months5 yearsSimilar
Standard/ GuidanceConforms to:• Guidance for Industry and FDA Staff: Premarket Notification [510(k)] Submissions for Chemical Indicators• ANSI/AAMI/ISO 11140-1:2014: Sterilization of Health Care Products – Chemical Indicators – Part 1: General RequirementsConforms to:• Guidance for Industry and FDA Staff: Premarket Notification [510(k)] Submissions for Chemical Indicators• ANSI/AAMI/ISO 11140-1:2014: Sterilization of Health Care Products – Chemical Indicators – Part 1: General RequirementsIdentical

6. Summary of Non-clinical Tests

Testing was performed to evaluate performance and demonstrate substantial equivalence to the predicate as summarized in Table 2.

Table 2. Performance Testing

TestAcceptance CriteriaResult
Stated Value Testing per ANSI/AAMI/ISO 11140-1:2014 section 11• All integrators processed will demonstrate no visible deterioration of the film• All integrators processed in a dry heat cycle will demonstrate a failing result• The Stated Values for time at 121 °C and 135 °C shall be specified and shall not be less than 16.5 min at 121 °C and 1.2 min at 135 °C• The temperature coefficient shall not be less than 10°C and not more than 27°C.• The correlation coefficient shall not be less than 0.9PASS

Page 11

STERIS Corporation5960 Heisley RoadMentor, OH 44060-1834 USA440-354-2600

April 24, 2025 K243876 Page 7 of 7

TestAcceptance CriteriaResult
Simulated Use Testing• All integrators processed in a full cycle will demonstrate a passing result with no physical deterioration of the film• All integrators processed in a partial cycle will demonstrate a passing result with no physical deterioration of the film• No integrators will exhibit a leak in the viewing windowPASS
Stability Study• All CIs processed in pass/full cycles in the BIER vessel will demonstrate pass results and shall not be under minimum specification limits as per the standard ISO 11140-1• All CIs processed in fail/partial cycles in BIER vessel will demonstrate fail results• All CIs processed for dry heat testing will demonstrate fail results.PASS
Comparison Testing with Biological Indicator• The Chemical Indicator does not reach its endpoint before the biological indicator is inactivated.PASS

7. Conclusion

The VERIFY STEAM Integrating Indicator has met the established performance criteria. Based on the intended use, technological characteristics and non-clinical performance data, the subject device (K243876) is as safe, as effective, and performs as well as the legally marketed predicate device (K213412), Class II (21 CFR 880.2800), product code JOJ.

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).