K Number
K243852
Device Name
Elfor-L
Manufacturer
Date Cleared
2025-04-14

(119 days)

Product Code
Regulation Number
870.2100
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Elfor L is intended for measuring micro-vascular perfusion in skin and muscle in humans. It is intended to be used for clinical research applications and pre clinical research applications.

Device Description

The Elfor-L device is making use of Miniaturized Dynamic Light Scattering (mDLS) technology to non-invasively measure micro-vascular perfusion in the skin. The device is comprised of sensor and signal processing station. The device uses Vertical-Cavity Surface-Emitting Lasers (VCSEL) as the light source and two photodetectors for detection.

The Elfor-L design incorporates three major components:

  1. Sensor - the sensor as its predicate includes a laser light source and two photo sensitive elements. The sensor transfers analog signal to the control unit.
  2. Control unit - the control unit is connected to the sensor by cable and includes the control circuit for the sensor, microcontroller and communication interface for the PC.
  3. Dedicated software that processes, displays and record data on the computer. The computer is not part of the system.
AI/ML Overview

The FDA 510(k) clearance letter for the Elfor-L device mentions that performance testing was conducted, including a comparison to its predicate device and an accuracy evaluation. However, the provided document does not contain the specific details required to fully address your request, such as:

  • A table of acceptance criteria and reported device performance values. While it states "The test passed and met the predefined acceptance criteria" for comparison testing and "The test passed" for accuracy, the actual numerical criteria and results are not listed.
  • Sample sizes used for the test set.
  • Data provenance (e.g., country of origin, retrospective/prospective).
  • Number of experts and their qualifications for ground truth.
  • Adjudication method for the test set.
  • Details on Multi-Reader Multi-Case (MRMC) comparative effectiveness study (effect size, human improvement).
  • Details on standalone (algorithm only) performance.
  • Type of ground truth used (expert consensus, pathology, outcomes data, etc.).
  • Sample size for the training set.
  • How ground truth for the training set was established.

The document primarily focuses on demonstrating substantial equivalence to a predicate device based on intended use, technological characteristics, and compliance with general safety and performance standards (biocompatibility, software validation, electrical safety, EMC). It implies that the performance was found to be acceptable but does not provide the granular data of the clinical study itself.

Therefore, for the specific questions related to acceptance criteria and study details (especially for an AI/machine learning context, which this device, being a physical measurement device, doesn't explicitly describe), the provided text is insufficient.

Based on the provided text, here's what can be extracted and what remains unknown:

1. A table of acceptance criteria and the reported device performance:

Acceptance CriterionStated Performance (or outcome)
Comparison Testing to Predicate Device"The test passed and met the predefined acceptance criteria." (Specific numerical criteria and results not provided)
Perfusion Accuracy"The test passed." (Accuracy value: 250±10 PU, as per the comparison table under "Perfusion Accuracy" which acts as the acceptance criterion from the predicate. The "reported device performance" would be that Elfor-L achieved this.)
Biocompatibility"The device was found biocompatible."
Software Validation"Software verification and validation testing... were conducted, and documentation was provided as recommended by FDA's Guidance."
Electrical Safety (IEC 60601-1)"The tests passed."
EMC (IEC 60601-1-2, IEC 60601-4-2)"The tests passed."
Laser Classification (IEC 60825-1)"The tests passed." (and device is Class I)

2. Sample sized used for the test set and the data provenance:

  • Sample Size: Not specified.
  • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not specified. The performance testing section describes "comparison testing of the Elfor-L to its predicate device" and evaluating accuracy using a "standardized Motility Standard solution." This suggests an objective, instrument-based ground truth (e.g., comparison to a validated device/standard) rather than expert human interpretation in the traditional sense of medical imaging.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable/not specified, as the ground truth appears to be objective, instrument-based.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. The Elfor-L is a direct measurement device ("Cardiovascular Blood Flowmeter") not an AI-assisted diagnostic imaging interpretation system. Therefore, MRMC studies are not relevant to its type of performance evaluation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • The device itself is a standalone measurement instrument. Its performance ("Accuracy of the Elfor-L device") was evaluated in comparison to a predicate and against a standard, which fits the concept of standalone performance for this type of device. There isn't an "algorithm" in the sense of AI interpreting an image; rather, it's a signal processing software function for a physical sensor.

7. The type of ground truth used:

  • For "Accuracy of the Elfor-L device," the ground truth was established using a "standardized Motility Standard solution."
  • For comparison testing, the ground truth was implied to be the performance of the predicate device.

8. The sample size for the training set:

  • Not specified. The document doesn't explicitly mention "training sets" in the context of machine learning, as this device measures physical parameters rather than learning from large datasets to perform a classification or detection task. Its "software validation" refers to V&V for software functionality, not machine learning model training.

9. How the ground truth for the training set was established:

  • Not applicable/not specified, as no machine learning training set is described.

FDA 510(k) Clearance Letter - Elfor-L Device

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID# 04017.07.05
Silver Spring, MD 20993
www.fda.gov

April 14, 2025

Elfi-Tech Ltd.
Orly Maor
Regulatory consultant
150 Derech Menachem Begin
Tel Aviv, 6492105
Israel

Re: K243852
Trade/Device Name: Elfor-L
Regulation Number: 21 CFR 870.2100
Regulation Name: Cardiovascular Blood Flowmeter
Regulatory Class: Class II
Product Code: DPW
Dated: December 12, 2024
Received: December 16, 2024

Dear Orly Maor:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K243852 - Orly Maor Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K243852 - Orly Maor Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA L. HITHE -S
Digitally signed by TANISHA L. HITHE -S
Date: 2025.04.14 17:03:18 -04'00'

Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K243852

Device Name
Elfor-L

Indications for Use (Describe)
The Elfor L is intended for measuring micro-vascular perfusion in skin and muscle in humans. It is intended to be used for clinical research applications and pre clinical research applications.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 5

Traditional Premarket Notification Submission – 510(k)

Elfor-L

510(k) Number ___________
Date Prepared: December 12, 2024

I. SUBMITTER

ELFI-TECH Ltd.
150 Derech Menachem Begin Tel-Aviv
Zip code: 6492105, Israel
Tel: +972-52-336-2625
E-mail: erez.herman@elfi-tech.com

Regulatory Correspondent:
Orly Maor
25A Sirkin Street
Kfar Saba 4442157, Israel
Tel: +972-9-7453607
Fax: +972-153-9-7453607
oram.ma@gmail.com

II. DEVICE

Name of Device: Elfor-L
Common or Usual Name: Elfor-L
Classification Name: 21CFR 870.2100- Cardiovascular blood flowmeter
Regulatory Class: II
Product Code: DPW

III. PREDICATE DEVICE

ELFI-TECH Ltd. believes that Elfor-L is substantially equivalent to the following predicate device:

  • Perimed Inc. PeriFlux 5000 cleared under K974285, Product code DPW, Regulation number 21 CFR 870.2100 and in subsequent submission under K152930, PeriFlux 6000, Product code DPW, Regulation number 21 CFR 870.2100.

IV. DEVICE DESCRIPTION

The Elfor-L device is making use of Miniaturized Dynamic Light Scattering (mDLS) technology to non-invasively measure micro-vascular perfusion in the skin. The device is comprised of sensor and signal processing station. The device uses Vertical-Cavity Surface-Emitting Lasers (VCSEL) as the light source and two photodetectors for detection.

The Elfor-L design incorporates three major components.

Page 6

Page 2 of 4

  1. Sensor - the sensor as its predicate includes a laser light source and two photo sensitive elements. The sensor transfers analog signal to the control unit.
  2. Control unit - the control unit is connected to the sensor by cable and includes the control circuit for the sensor, microcontroller and communication interface for the PC.
  3. Dedicated software that processes, displays and record data on the computer. The computer is not part of the system.

V. INDICATIONS FOR USE

The Elfor-L is intended for measuring microvascular perfusion in skin and muscle in humans. It is intended to be used for clinical research applications and pre-clinical research applications.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The Elfor-L has the same intended use as the predicate device. Its indications for use are identical to that of the predicate device.

The Elfor-L has similar technological characteristics as the predicate device as demonstrated in the table below:

ELFOR-LPERIFLUX 5000PERIFLUX 6000 SEJUSTIFICATION
510(k) NumberTBDK974285K152930___
ManufacturerELFI-TECHPerimed ABPerimed AB___
Product CodeDPWDPWDPWSame
CFR870.2100870.2100870.2100Same
Intended UseThe Elfor L is intended for measuring micro-vascular perfusion in skin and muscle in humans. It is intended to be used for clinical research applications and pre-clinical research applications.The PF 5010 LDPM Unit is indicated for use in measuring micro-vascular perfusion in skin and muscle in humans. It is also intended for use in measuring micro-vascular perfusion in all tissues in animals.PeriFlux 6000, equipped with PF 6010 is intended for measuring micro-vascular perfusion in skin and muscle in humans. It is also intended for measuring micro-vascular perfusion in all tissues in animals for research purposes.Same

Page 7

Page 3 of 4

ELFOR-LPERIFLUX 5000PERIFLUX 6000SE JUSTIFICATION
Intended UserThe Elfor-L is intended for professional use in a hospital environment, laboratory environment and in a hospital intensive care unit environment.The PF 5010 LDPM Unit is intended for professional use in a hospital environment, laboratory environment and in a hospital intensive care unit environment.The PeriFlux 6000 instrument is intended for professional use in a hospital environment, laboratory environment and in a hospital intensive care unit environment.Same
Patient PopulationThe Elfor-L can be used for measurements on patients of all ages.The PF 5010 LDPM Unit can be used for measurements on patients of all ages.The PF 6010 LDPM/Temp Unit can be used for measurements on patients of all ages.Same
Body contactThe Elfor-L is intended for measurements on intact skin.The PeriFlux 5000 is intended for measurements on intact skin.The PeriFlux 6000 is intended for measurements on intact skin.Same
Maximum laser current3-5mANot specified.5mAWithin the predicate range
Laser product classificationClass I laser product per IEC 60825-1.Class I laser product per IEC 60825-1.Class I laser product per IEC 60825-1.Same
Laser TypeInvisible, near-infrared, 850nmInvisible, near-infrared, 780 nmInvisible, near-infrared, 785nmSimilar
Perfusion measurement0-2000 PUUpper limit not specified.0–2000 PUSame as PeriFlux 6000
Perfusion Accuracy250±10Motility standard: 250±20 PU Zeroing disc: 0.0±1.0 PU± 2 PU at zero, ±4 % in motility standardSame
Signal Bandwidth5Hz -24KHz20 Hz to 15 kHz5 Hz - 25 kHzSubstantially equivalent
CalibrationRequiredRequired.RequiredSame
List of main components1. VCSEL laser source 2. Two photodetectors 3. Amplifier 4. Current driver 5. A2D convertor 6. Microprocessor Adhesive Tape and Disposable patches1. Laser source 2. Fiber optic probe 3. Photodetector 4. Amplifier 5. A2D convertor 6. Microprocessor Adhesive Tape and Probe holder1. Laser source 2. Fiber optic probe 3. Photodetector 4. Amplifier 5. A2D convertor 6. Microprocessor Adhesive Tape and Probe holderSimilar

Page 8

Page 4 of 4

ELFOR-LPERIFLUX 5000PERIFLUX 6000SE JUSTIFICATION
Power sourceThe Elfor-L receives its power from a USB port (5V; 0.5A)100–240 Volts, 50 or 60 Hz.100–240 Volts, 50 or 60 Hz.Different however, no safety issue
StandardsIEC 60601-1 IEC 60601-1-2 IEC 60825-1IEC 60601-1 IEC 60601-1-2 IEC 60825-1IEC 60601-1 IEC 60601-1-2 IEC 60825-1Same

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination:

Biocompatibility

Biocompatibility evaluation in compliance with ISO 10993-1 was performed. The following tests were conducted:

  • Cytotoxicity Study Using the ISO Elution Method
  • ISO Guinea Pig Maximization Sensitization Test
  • ISO Intracutaneous -irritation Study in Rabbits

The device was found biocompatible.

Performance Testing

Performance testing included comparison testing of the Elfor-L to its predicate device. The main purpose of this test was to verify the Elfor-L performance is similar to that of its predicate.

The test passed and met the predefined acceptance criteria.

Accuracy of the Elfor-L device- the purpose of this test was to evaluate the accuracy of the Elfor-L by repeated measurements using a standardized Motility Standard solution. The test passed.

Software Validation

Software verification and validation testing of the Elfor-L were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions".

Electrical Safety and EMC

Electrical Safety per IEC 60601-1, Electromagnetic compatibility (EMC) per IEC 60601-1-2, IEC 60601-4-2 and IEC 60825-1 were conducted on the Elfor-L. The tests passed.

VIII. CONCLUSION

The Elfor-L was determined to be substantially equivalent to the predicate device.

§ 870.2100 Cardiovascular blood flowmeter.

(a)
Identification. A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.(b)
Classification. Class II (performance standards).