K Number
K243555
Date Cleared
2025-02-05

(79 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  • Mode 1 (Red + Infrared light): treatment of full face wrinkles.
  • Mode 2 (Blue light): Treatment of mild to moderate inflammatory acne.
  • Mode 3 (Mixed light): Treatment of mild to moderate inflammatory acne.
Device Description

LED phototherapy masks are over-the-counter light-emitting diode (LED) devices that emit energy and are used in dermatology to treat acne and wrinkles. The device emits red (630nm) and infrared (850nm) light to treat wrinkles and blue (415nm) light to treat mild to moderate acne. All models had to be fitted with an eye mask and strapped to the treatment area, and all had a controller to set the parameters of the device. There are 10 models of the device (model: E43, E32, E49B, E49C, E100B, E100C, E103B, E104B, E106A, E108A), all of which are operated through a controller.

AI/ML Overview

The provided text describes an FDA 510(k) premarket notification for an "LED light therapy mask." This document primarily focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing rather than clinical performance studies.

Therefore, the information required to answer your request regarding acceptance criteria and performance based on clinical studies is largely not present in the provided text. The document explicitly states:

"8.2 Summary of Clinical Performance: Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively."

This means there was no multi-reader, multi-case study, no standalone algorithm performance, no detailed ground truth establishment, and no sample sizes for test or training sets from clinical data because a clinical study was not performed or required for this submission.

However, I can extract the information related to the non-clinical tests performed and common regulatory standards:


Summary of Information from the Provided Document:

1. A table of acceptance criteria and the reported device performance:

  • Clinical Performance: Not applicable/not performed for this 510(k) submission.
  • Non-Clinical Performance (Acceptance Criteria are inferred by compliance with standards):
Acceptance Criteria (Inferred from regulatory standards)Reported Device Performance ("Compliance")
Electrical SafetyComplied with IEC 60601-1 Edition 3.2 2020-08 (General requirements for basic safety and essential performance).
Home Healthcare Environment SafetyComplied with IEC 60601-1-11 Edition 2.1 2020-07 (Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment).
Non-Laser Light Source Equipment SafetyComplied with IEC 60601-2-57 Edition 1.0 2011-01 (Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use).
Electromagnetic Compatibility (EMC)Complied with IEC 60601-1-2 Edition 4.1 2020-09 (Electromagnetic disturbances - Requirements and tests).
Photobiological SafetyComplied with IEC 62471 First edition 2006-07 (Photobiological safety of lamps and lamp systems).
Battery Safety (for models with lithium batteries)Complied with IEC 62133-2 Edition 1.0 2017-02 (Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems).
BiocompatibilityComplied with ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Sensitization), ISO 10993-23 (Skin Irritation).
Software Verification and ValidationConducted and documented as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." (Software was considered "moderate" level concern).
UsabilityComplied with IEC 62366-1 and IEC 60601-1-6.
CybersecurityNot needed as the device has no external interfaces, according to FDA guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".

2. Sample size used for the test set and the data provenance:

  • Clinical Test Set: Not applicable, no clinical test set was used/required.
  • Non-Clinical/Engineering Tests: Sample sizes for hardware, software, and biocompatibility testing are not specified in this summary, which is typical for 510(k) summaries of this nature (they primarily list the standards complied with). Data provenance for these tests would typically be from the manufacturer's internal testing or third-party accredited labs.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable, as no clinical test set requiring expert ground truth establishment was used.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable, as no clinical test set requiring adjudication was used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, a MRMC study was not done. The device is an LED light therapy mask, not an AI diagnostic assistant for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device is a direct treatment device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Not applicable for clinical ground truth. For non-clinical tests, "ground truth" would be the specifications and requirements defined by the relevant engineering and safety standards (e.g., specific wavelength output, power intensity, safety thresholds for electrical current, absence of cytotoxic effects, etc.).

8. The sample size for the training set:

  • Not applicable, as no clinical training set was used for an AI algorithm.

9. How the ground truth for the training set was established:

  • Not applicable.

In summary, this FDA 510(k) submission for the LED light therapy mask relied on non-clinical performance data and compliance with recognized consensus standards to demonstrate substantial equivalence, rather than new clinical trials or studies involving human "readers" or AI performance evaluation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

February 5, 2025

Shenzhen Eyco Technology Co., Ltd % Lee Cassie Official Correspondent Share Info (Guangzhou) Medical Consultant Ltd. No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road Huangpu District, Guangzhou, China Guangdong, Guangzhou 510530 China

Re: K243555

Trade/Device Name: LED light therapy mask (E43. E32. E49B, E49C, E100B, E100C, E103B, E104B, E106A. E108A) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OLP, OHS Dated: November 15, 2024 Received: November 18, 2024

Dear Lee Cassie:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA - TANISHA - TANISHAL. - TANISHAL.
L. HITHE - C. Dae: 20.02.02.02.02.

Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K243555 Device Name

LED light therapy mask (E43, E32, E49B, E49C, E100B, E100C, E103B, E104B, E106A, E108A)

Indications for Use (Describe)

  • Mode 1 (Red + Infrared light): treatment of full face wrinkles.

  • Mode 2 (Blue light): Treatment of mild to moderate inflammatory acne.

  • Mode 3 (Mixed light): Treatment of mild to moderate inflammatory acne.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) Summary for K243555

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92

1. Submitter's Information

Sponsor Name: Shenzhen Eyco Technology Co., Ltd Address: 2 / F, Building 1, Jiashitai Science Park, 164# Fengtang Avenue, Fuhai Street, Baoan District, Shenzhen 518103, China Contact name: Martin Ding (CEO) Tel: +86 15602440991 E-mail: martin@eycobeauty.com

Application Correspondent:

Contact Name: Cassie Lee Company: Share Info (Guangzhou) Medical Consultant Ltd. Address: No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road, Huangpu District, Guangzhou, China Tel: +86 20 8200 6973 E-mail: 382198657@qq.com

2. Subject Device Information:

Trade Name: LED light therapy mask Model: E43, E32, E49B, E49C, E100B, E100C, E103B, E104B, E106A, E108A Medical specialty: General & Plastic Surgery Regulation: 21 CFR 878.4810 Product Code: OLP (Class II) - Over-The-Counter Powered Light Based Laser For Acne Associated Product Code(s): OHS (Class II) - Light Based Over The Counter Wrinkle Reduction

Predicate Device Information 3.

Predicate Device 1:

Sponsor: Guangdong Newdermo Biotech Co.,Ltd Trade Name: LED light therapy mask 510(k) Number: K223544 Medical specialty: General & Plastic Surgery Regulation: 21 CFR 878.4810 Product Code: OLP, OHS

{5}------------------------------------------------

Predicate Device 2:

Sponsor: Marci Beauty Inc Trade Name: Infrared Red Blue LED Light Heat Beauty Machine 510(k) Number: K210545 Medical specialty: General & Plastic Surgery Requlation: 21 CFR 878.4810 Product Code: OLP, OHS

Reference Device:

Sponsor: Ambicare Health Ltd Trade Name: Lustre PRO Light System 510(k) Number: K143713 Medical specialty: General & Plastic Surgery Regulation: 21 CFR 878.4810 Product Code: OLP

Device Description 4.

LED phototherapy masks are over-the-counter light-emitting diode (LED) devices that emit energy and are used in dermatology to treat acne and wrinkles. The device emits red (630nm) and infrared (850nm) light to treat wrinkles and blue (415nm) light to treat mild to moderate acne. All models had to be fitted with an eye mask and strapped to the treatment area, and all had a controller to set the parameters of the device.

There are 10 models of the device (model: E43, E32, E49B, E49C, E100B, E100C, E103B, E104B, E106A, E108A), all of which are operated through a controller. The differences between the 10 models are listed below:

    1. Mode of power supply:
  • For E43 and E32 devices, the device of model E43, E32 are powered by connecting । to the mains, connected to the controller through the adapter, and then connected to the mask through the power cord provided by the manufacturer.
  • -For masks E49B, E49C, E100B, E100C, E103B, E104B, E106A, E108A, power the device and control the operation through the lithium battery inside the controller (Specifications adapter can be used (input AC100-240V 50/60Hz, outputs DC 10V 2A) to charge the controller).
  1. Different manufacturing materials: E32 and E43 are hard cases, and the material is PC. The corresponding equipment of E49B, E49C, E100B, E100C, E103B, E104B, E106A and E108A are soft cases, and the material is silica gel.

{6}------------------------------------------------

  1. The controller is different, the device model (E43, E32) has 4 LED lights, 7 buttons and LED display on the controller. Seven buttons can be used to set treatment time, treatment mode, treatment brightness level, and start or stop treatment. The display screen shows the treatment time, treatment mode and treatment intensity, and is equipped with 4 LED lights to indicate whether the power is connected. When all four lights are on, the current treatment has begun, the controller is directly powered through an adapter.

The controller for the model (E49B, E49C, E100B, E100C, E103B, E104B, E106A) devices has three buttons to adjust the brightness level, switch modes, set time, and display, The display displays the battery level, treatment mode, time, and brightness information of the current controller, the controller is powered by an internal rechargeable lithium battery.

5. Indications for Use

  • -Mode 1 (Red + Infrared light): treatment of full face wrinkles.
  • -Mode 2 (Blue light): Treatment of mild to moderate inflammatory acne.
  • -Mode 3 (Mixed light): Treatment of mild to moderate inflammatory acne.

6. Comparison to Predicate Devices

Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions and the applicable standards. The differences between subject device and predicate device do not raise any new questions of safety or effectiveness.

ElementsofComparisonSubject DevicePredicate Device 1Predicate Device 2ReferenceDeviceVerdict
CompanyShenzhen EycoTechnology Co., LtdGuangdongNewdermo BiotechCo.,LtdMarci BeautyIncAmbicare HealthLtd--
TradeNameLED light therapymaskLED light therapymaskInfrared RedBlue LE D LightHeat BeautyMachineLustre PROLight System--
510(k)NumberK243555K223544K210545K143713--
MedicalSpecialtyGeneral & PlasticSurgery21 CFR 878.4810General &Plastic Surgery21 CFR 878.4810General &Plastic Surgery21 CFR878.4810General &Plastic Surgery21 CFR878.4810Same
ProductcodeOLP, OHSOLP, OHSOLP, OHSOLPSame
RegulationClassIIIIIIIISame
IntendeduseMode 1 (Red +Infrared light):treatment of full facewrinkles.Mode 2 (Blue light):Treatment of mild tomoderateinflammatory acne.Mode 3 (Mixed light):Treatment of mild tomoderateRed light: Treatmentof full-face wrinklesBlue light: Treatmentof mild to moderateinflammatory acne.Infrared light: Providetopical heating forthe purpose ofelevating tissuetemperature; arthritisand muscleThe InfraredRed Blue LEDLight HeatBeauty Machineis an hand-heldover-the-counterphototherapydevice, the redand infraredlight is intendedThe Lustre PROLight system isintended to emitlight in the blueregion of thelight spectrumand is indicatedfor the treatmentof mildtomoderateacne vulgarisSimilarNote 1
inflammatory acne.spasm; relievingstiffness; promotingthe relaxation ofmuscle tissue; andto temporarilyincrease local bloodcirculation.Mixed light:Treatment of mild tomoderateinflammatory acne.for the use intreating wrinkleson the face andthe blue light isintended for thetreatment of themild tomoderateinflammatoryacne.This device isindicated foradults only.
IntendedLocation ofUseFaceFace and bodyFaceFaceSame
WavelengthBlue: 415nm,Red: 630nm,NIR: 850nmMixed:415nm+630nm+850nmRed: 620nmBlue: 460nmInfrared: 850nmMixed: 620nm and850nm and 460nmBlue:465nmRed:620-630nmIR:845-855nmBlue:415nm±15nmSimilarNote 2
EnergyTypeLight emitting diodesLight emitting diodesLight emittingdiodesLight emittingdiodesSame
Intensity(mW/cm2)R+NIR: 6.08.0Blue light:2.04.0Mixed light: 9.0~12.0Red light: 2.03.0mW/cm2Blue light:2.04.0mW/cm2Infrared light:2.04.0 mW/cm2Mixed light:9.012.0 mW/cm2Blue LightMode:5.4Red Light + NIRMode:7.25SimilarNote 3
Treatmentprotocol(Treatmenttime)Recommended10minutes each time.3-4 treatment aweek, reduce to 1-2treatment a weekonce the resultsshown.Manual Mode:15minutes each time,Automatic Mode:10minutes each time.3-4 treatment aweek, reduce to 1-2treatment a weekonce the resultsshown.3 times a weekfor 30 min.4weeksNo publiclyavailableSame
ControlDevice uses a timerand software tocontrol treatmentduration.Device uses a timerand software tocontrol treatmentduration.No publiclyavailableDevice uses atimer andsoftware tocontrol treatmentduration.Same

{7}------------------------------------------------

7. Comparison in Detail(s):

Note 1:

Although the "Indications for Use / Intended use" is a little different from the predicate devices, the function of mode 1 for treating full face wrinkles corresponds to the intended use of predicate device 2, with the description that the function of mode 2 for treating mild to moderate acne is consistent with the predicate device 1 and the intended use of reference device, and the mode 3 corresponding to the mixed light mode is consistent with the intended use of the predicate device 1.

{8}------------------------------------------------

Note 2:

Although the "Wavelengths" is a little different from the predicate devices, but wavelengths of the subject device can be covered by the predicate and reference device. Due to the characteristics of LED lights, the emission parameters are showing the peak wavelength of the lamp, and have been assessed by IEC 62471 and IEC 60601-2-57, So, these slight differences will not raise any safety or effectiveness issues.

Note 3:

Although the "Irradiance", is a little different from the predicate devices, the values of each treatment for these parameters are very close. The irradiance of the subject device is similar to the predicate device 1, predicate device 2 and reference device. The minor differences do not lead to any safety or efficacy issues.

8. Test Summary

8.1 Non-Clinical Tests Performed

Non-clinical tests were performed on the subject device in order to validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety, and electromagnetic compatibility:

IEC 60601-1 Edition 3.2 2020-08 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.

IEC 60601-1-11 Edition 2.1 2020-07 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

IEC 60601-2-57 Edition 1.0 2011-01 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use.

IEC 60601-1-2 Edition 4.1 2020-09 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.

IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems.

IEC 62133-2 Edition 1.0 2017-02 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems

Biocompatibility

The subject device is biocompatible for its intended use. They are complied with biocompatibility standards ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Sensitization), ISO 10993-23 (Skin Irritation).

Software verification and validation testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this

{9}------------------------------------------------

device was considered as a "moderate" level concern, since a malfunction of, or a latent design flaw in, the Software Device leads to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.

Usability Testing

Usability testing was conducted on the LED light therapy mask (Models: E43, E32, E49B, E49C, E100B, E100C, E103B, E104B, E106A, E108A), which complies with IEC 62366-1 and IEC 60601-1-6.

Cybersecurity

The subject device no any external interfaces, according to FDA guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices", no need cybersecurity evaluation.

8.2 Summary of Clinical Performance

Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively.

9. Date of the summary prepared: January 7, 2025

10. Final Conclusion

This difference does not raise different risks or concerns of safety and effectiveness compared to the predicate. The subject device is as safe, as effective, and performs as well as or better than the legally marketed predicated devices K223544 and K210545.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.