(49 days)
The Healgen® Accurate Multi-Drug Urine Screening Cup is a rapid lateral flow immunoassay for the qualitative detection of 6-Monoacetylmorphine, d-Amphetamine, Benzoylecgonine, Buprenorphine, EDDP, Norfentanyl, Methadone, d-Methamphetamine, d/l-Methylenedioxymethamine, Mortriptyline, Oxazepam, Oxycodone, Phencyclidine, d-Propoxyphene, Secobarbital, THC-COOH and Tramadol in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
| Test | Calibrator | Cut-off (ng/mL) |
|---|---|---|
| 6-MAM | 6-Monoacetylmorphine | 10 |
| AMP | d-Amphetamine | 500 or 1000 |
| BAR | Secobarbital | 300 |
| BUP | Buprenorphine | 10 |
| BZO | Oxazepam | 300 |
| COC | Benzoylecgonine | 150 or 300 |
| EDDP | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrolidine | 300 |
| FEN or FYL | Norfentanyl | 5 |
| MDMA | Methylenedioxymethamphetamine | 500 |
| MET | d-Methamphetamine | 500 or 1000 |
| MTD | Methadone | 300 |
| OPI | Morphine | 300 or 2000 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | d-Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | 11-nor-Δ9-THC-COOH | 50 |
| TRA or TMI | Tramadol | 100 |
The single or multi-test cups can consist of up to eighteen (18) of the above listed analytes in any combination with or without on-board adulteration/specimen validity tests (SVT).
The tests provide only a preliminary result. A more specific alternative chemical must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
The Healgen® Accurate Home Multi-Drug Urine Test Cup is a rapid qualitative immunoassay. The device provides preliminary results for the detection of one or more of the following drugs
| CODE | SUBSTANCE | Cut-off (ng/mL) |
|---|---|---|
| AMP | Amphetamine | 1000 or 500 |
| BUP | Buprenorphine | 10 |
| BAR | Secobarbital | 300 |
| BZO | Oxazepam | 300 |
| COC | Cocaine | 300 or 150 |
| EDDP | EDDP | 300 |
| FYL | Norfentanyl | 5 |
| MET/mAMP | Methamphetamine | 1000 or 500 |
| MDMA | Ecstasy | 500 |
| OPI | Morphine | 2000 or 300 |
| MTD | Methadone | 300 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | Marijuana | 50 |
| TML | Tramadol | 100 |
| 6-MAM | 6-Monoacetylmorphine | 10 |
This drug test cup may contain any combination of the drug tests listed in the table above.
This test provides only preliminary result. A more specific alternative chemical must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
Healgen® Accurate Home Muti-Drug Urine Test Cup and Healgen® Accurate Muti-Drug Urine Drug Screen Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The device is a cup format. Each test device is sealed with two sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.
Acceptance Criteria and Device Performance for Healgen® Accurate Multi-Drug Urine Drug Screen Cup
This document outlines the acceptance criteria and the evidence provided to demonstrate that the Healgen® Accurate Multi-Drug Urine Drug Screen Cup meets these criteria. The information is extracted from the provided FDA 510(k) summary (K243365).
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this qualitative immunoassay are primarily demonstrated through a combination of precision, analytical specificity/interference, and a method comparison study. The precision study evaluates the device's ability to consistently produce correct results at various concentrations relative to the cutoff, while the method comparison study assesses its agreement with a gold standard (LC-MS/MS) using clinical samples. The layperson study demonstrates ease of use for over-the-counter applications.
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
|---|---|---|
| Precision/Reproducibility | Consistent qualitative results (+/- at least 95% at the cutoff concentration, and 100% agreement for samples far from the cutoff) across multiple lots and runs, at various concentrations relative to the cutoff (-100%, -75%, -50%, -25%, Cutoff, +25%, +50%, +75%, +100%). | For all drugs tested (AMP 1000/500, BAR 300, BUP 10, BZO 300, COC 300/150, EDDP 300, FYL 5, MDMA 500, MET 1000/500, OPI 300/2000, OXY 100, PCP 25, PPX 300, TCA 1000, THC 50, TML 100, 6-MAM 10) across 3 lots: - 100% agreement for concentrations at -100%, -75%, -50%, +50%, +75%, and +100% cutoff (e.g., 0-/50+ or 50-/0+ indicating no false negatives or positives respectively). - At the cutoff concentration, results consistently showed a mix of positive and negative interpretations, typically near 50%/50% (e.g., 23-/27+, 25-/25+), demonstrating sensitivity around the cutoff. - At -25% cutoff, the device predominantly showed negative results (e.g., 50-/0+ or 49-/1+), with a few false positives. - At +25% cutoff, the device predominantly showed positive results (e.g., 0-/50+ or 1-/49+), with a few false negatives. This indicates good precision around the cutoff with expected variability at borderline concentrations. |
| Analytical Specificity | No significant cross-reactivity with commonly encountered substances (drug metabolites, prescription medications, endogenous compounds) at relevant concentrations, and no interference from variations in pH and specific gravity within a physiological range. | Diverse lists of compounds were tested for cross-reactivity. The results show that: - Many cross-reacting compounds are identified with their minimum concentration needed to yield a positive result and their % Cross-Reactivity (e.g., Hydroxyamphetamine with AMP 1000, Amobarbital with BAR 300). - Numerous compounds showed no detection or were negative at high concentrations (e.g., 100,000 ng/mL), indicating no cross-reactivity for those substances. - pH levels of 4 to 9 and specific gravity levels of 1.000 to 1.035 were demonstrated to not affect the assay results. |
| Method Comparison | A high degree of concordance between the device's qualitative results and confirmatory Gas Chromatography-Mass Spectrometry (GC-MS) or Liquid Chromatography-Mass Spectrometry (LC-MS) results for clinical urine samples, particularly for samples at or near the cutoff. | For each drug and its cutoff configuration: - 100% agreement was observed for drug-free samples and "low negative" samples (less than -50% of cutoff). - 100% agreement was primarily observed for "high positive" samples (greater than +50% of cutoff). - For "near cutoff negative" (between -50% and cutoff) and "near cutoff positive" (between cutoff and +50%) samples, there was expected variability, with some samples interpreted differently by the device compared to LC-MS/MS, indicating the device's expected performance around the cutoff. The discordant results table details specific samples where device results differed from LC-MS/MS. Generally, the majority of samples near the cutoff were correctly identified by the device. |
| Lay Person Usability | The device should be easy to understand and use by laypersons as indicated by a high agreement in understanding instructions and performance on prepared samples across different drug classes. A reading level analysis of the instructions for use should demonstrate appropriate readability. | A study with 280 laypersons (mix of male/female, aged 21 to >50, diverse backgrounds) demonstrated: - 100% agreement for interpretation of drug-free and concentrations at -100%, -75%, -50% below cutoff. - 100% agreement for concentrations at +50% and +75% above cutoff. - Expected variability in agreement around the -25% and +25% cutoff concentrations, mostly ranging from 90% to 95%. - All participants indicated the instructions were easy to understand and follow. - The Flesch-Kincaid reading analysis yielded a Grade Level of 7, which is suitable for layperson use. |
| Stability | The device should maintain its performance characteristics for a specified duration under defined storage conditions. | The device is stable at 2-30°C for 36 months based on real-time stability studies. |
2. Sample Size Used for the Test Set and Data Provenance
Precision/Reproducibility Study:
- Sample Size: For each drug and each concentration level (total 9 levels per drug), 50 individual tests were performed (2 runs per day for 25 days). Given there are 17 distinct drug analytes, this amounts to 17 drugs * 9 concentrations * 50 tests/concentration = 7,650 tests.
- Data Provenance: Samples were prepared by spiking target drugs into drug-free urine samples. These samples were likely prepared in a laboratory setting, making the data provenance prospective and controlled. The document does not specify a country of origin for the samples/urine, but the study was conducted "in-house."
Method Comparison Study:
- Sample Size: For each drug, 80 unaltered urine clinical samples were used. This consisted of 40 negative and 40 positive samples. Given there are 17 distinct drug analytes, this amounts to 17 drugs * 80 samples/drug = 1,360 total clinical samples tested across all drugs.
- Data Provenance: Unaltered clinical urine samples. The document does not specify the country of origin but states the study was performed "in-house." The nature of "unaltered clinical samples" suggests these were collected from patients, and their retrospective or prospective nature is not explicitly stated, though being "blind labeled" for comparison suggests they were handled specifically for this study.
Lay Person Study:
- Sample Size: 280 lay persons participated. Urine samples were prepared at 7 concentration levels (-100%, +/-75%, +/-50%, +/-25% of cutoff). Each participant tested 1 blind labeled sample. While it states urine samples were prepared at 7 concentrations, the agreement tables show 7 concentrations and 20 tests per concentration for each drug. If we assume each of the 280 participants tested one sample for one drug, the sample sizes per concentration shown in the table (20 tests) indicate that multiple participants contributed data for each drug and concentration, but a single participant did not test all drugs or all concentrations.
- Data Provenance: Urine samples were prepared by spiking drugs into drug-free pooled urine specimens. The samples were prepared in a laboratory environment, indicating prospective and controlled data provenance. The lay persons were recruited for the study, making their involvement prospective. The study was performed at "three intended user sites."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Precision/Reproducibility Study:
- Ground Truth: The "ground truth" for the spiked samples was established by the known concentrations of the spiked drugs, which were confirmed by LC-MS/MS.
- Experts/Qualifications: The document states that "Each drug concentration was confirmed by LC-MS/MS." LC-MS/MS is a highly accurate analytical method, and while specific "experts" (e.g., toxicologists, clinical chemists) are not explicitly named, the use of this method implies analysis by qualified laboratory personnel specializing in mass spectrometry. "One operator per lot" (for 3 lots) performed the tests on the device, but the LC-MS/MS confirmation implies separate, expert analysis for ground truth.
Method Comparison Study:
- Ground Truth: The ground truth for the clinical samples was established by LC-MS/MS results.
- Experts/Qualifications: Similar to the precision study, the ground truth was determined by LC-MS/MS, implying analysis by qualified laboratory professionals. No specific number or qualification of "experts" is provided beyond the analytical method itself.
Lay Person Study:
- Ground Truth: The "ground truth" for the spiked samples was established by the known concentrations of the spiked drugs, which were confirmed by LC-MS/MS.
- Experts/Qualifications: As above, the ground truth was derived from LC-MS/MS analysis, indicating reliance on qualified laboratory personnel.
4. Adjudication Method for the Test Set
Precision/Reproducibility Study:
- Adjudication Method: Not applicable in the traditional sense of expert consensus. The ground truth was based on the quantitative LC-MS/MS results of the spiked samples. The device's qualitative results were compared against these known concentrations relative to the cutoff. Discrepancies were noted directly.
Method Comparison Study:
- Adjudication Method: Not applicable/Not explicitly stated as an adjudication process involving multiple human readers. The device results were directly compared to the LC-MS/MS results, which served as the definitive "true" value (ground truth). The tables show the device's output (positive/negative) versus the LC-MS/MS quantification, which then categorizes samples as drug-free, low negative, near cutoff negative, near cutoff positive, or high positive.
Lay Person Study:
- Adjudication Method: Not applicable in terms of expert consensus. The laypersons' interpretations of the device results were compared against the established ground truth (LC-MS/MS confirmed concentrations of spiked samples). The study aimed to assess if laypersons could correctly interpret the device results against this objective ground truth, not for them to establish ground truth or adjudicate.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done in the context of comparing human readers with and without AI assistance, as this is an in-vitro diagnostic device (a test cup) not an AI-powered diagnostic imaging or interpretation system. The "operators" in the method comparison study refer to individuals performing the device test, not medical interpreters assisted by AI. The "lay person study" evaluated if laypersons could correctly interpret the device, not an AI system.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This is an in-vitro diagnostic (IVD) device (a lateral flow immunoassay test cup), not an AI algorithm. Therefore, the concept of "standalone (algorithm only)" performance without human-in-the-loop is not applicable in the typical sense for AI/software devices. The device itself provides a visual qualitative result (lines appearing or not appearing), which then requires human observation and interpretation. The performance characteristics described (precision, specificity, method comparison) are inherently "standalone" in that they assess the device's ability to produce correct visual results, which a human then reads. The layperson study specifically evaluates the human (layperson) interpretation of these visual results.
7. The Type of Ground Truth Used
The primary type of ground truth used across all analytical performance studies (Precision, Analytical Specificity, Method Comparison, and Lay Person Study where applicable) is Gas Chromatography-Mass Spectrometry (GC-MS) and Liquid Chromatography-Mass Spectrometry (LC-MS) and their tandem mass-spectrometer versions (LC-MS/MS).
These are considered confirmatory analytical methods and are widely accepted as the gold standard for quantitative drug detection in toxicology.
8. The Sample Size for the Training Set
As this device is a lateral flow immunoassay (a chemical/biological test kit), and not an AI/machine learning algorithm, there is no "training set" in the computational sense. The device's performance is based on its inherent physical and chemical properties and reagent formulations, which are developed and optimized through laboratory research and development, rather than machine learning training.
9. How the Ground Truth for the Training Set Was Established
Since there is no "training set" for this type of device, this question is not applicable. The development and optimization of the immunoassay undoubtedly involved extensive experimentation with samples of known drug concentrations, likely confirmed by GC-MS/LC-MS, but this is part of the product's research and development process, not analogous to establishing ground truth for a machine learning training set.
{0}------------------------------------------------
December 17, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Healgen Scientific LLC % Jenny Xia Director LSI International Inc 504E Diamond Ave., Suite H Gaithersburg, Maryland 20877
Re: K243365
Trade/Device Name: Healgen® Accurate Muti-Drug Urine Drug Screen Cup; Healgen® Accurate Home Muti-Drug Urine Test Cup Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: Class II Product Code: NGL,NFT,PTH,NFY,NFV,PTG,NGG,NGM,QBF,QAW,NFW Dated: October 26, 2024 Received: October 29, 2024
Dear Jenny Xia:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Picture/3 description: The image shows the name Joseph A. Kotarek -S. It also shows that the document was digitally signed by Joseph A. Kotarek -S on December 17, 2024. The time stamp on the document is 10:21:47 -05'00'.
Joseph Kotarek Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K243365
Device Name
Healgen® Accurate Muti-Drug Urine Drug Screen Cup; Healgen® Accurate Home Muti-Drug Urine Test Cup
Indications for Use (Describe)
The Healgen® Accurate Multi-Drug Urine Screening Cup is a rapid lateral flow immunoassay for the qualitative detection of 6-Monoacetylmorphine, d-Amphetamine, Benzoylecgonine, Buprenorphine, EDDP, Norfentanyl, Methadone, d-Methamphetamine, d/l-Methylenedioxymethamine, Mortriptyline, Oxazepam, Oxycodone, Phencyclidine, d-Propoxyphene, Secobarbital, THC-COOH and Tramadol in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
| Test | Calibrator | Cut-off (ng/mL) |
|---|---|---|
| 6-MAM | 6-Monoacetylmorphine | 10 |
| AMP | d-Amphetamine | 500 or 1000 |
| BAR | Secobarbital | 300 |
| BUP | Buprenorphine | 10 |
| BZO | Oxazepam | 300 |
| COC | Benzoylecgonine | 150 or 300 |
| EDDP | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrolidine | 300 |
| FEN or FYL | Norfentanyl | 5 |
| MDMA | Methylenedioxymethamphetamine | 500 |
| MET | d-Methamphetamine | 500 or 1000 |
| MTD | Methadone | 300 |
| OPI | Morphine | 300 or 2000 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | d-Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | 11-nor-Δ9-THC-COOH | 50 |
| TRA or TMI | Tramadol | 100 |
The single or multi-test cups can consist of up to eighteen (18) of the above listed analytes in any combination with or without on-board adulteration/specimen validity tests (SVT).
The tests provide only a preliminary result. A more specific alternative chemical must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
The Healgen® Accurate Home Multi-Drug Urine Test Cup is a rapid qualitative immunoassay. The device provides preliminary results for the detection of one or more of the following drugs
| CODE | SUBSTANCE | Cut-off (ng/mL) |
|---|---|---|
| AMP | Amphetamine | 1000 or 500 |
| BUP | Buprenorphine | 10 |
| BAR | Secobarbital | 300 |
| BZO | Oxazepam | 300 |
| COC | Cocaine | 300 or 150 |
| EDDP | EDDP | 300 |
{4}------------------------------------------------
| FYL | Norfentanyl | 5 |
|---|---|---|
| MET/mAMP | Methamphetamine | 1000 or 500 |
| MDMA | Ecstasy | 500 |
| OPI | Morphine | 2000 or 300 |
| MTD | Methadone | 300 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | Marijuana | 50 |
| TML | Tramadol | 100 |
| 6-MAM | 6-Monoacetylmorphine | 10 |
This drug test cup may contain any combination of the drug tests listed in the table above.
This test provides only preliminary result. A more specific alternative chemical must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
510(k) SUMMARY K243365
| 1. Date: | October 29, 2024 |
|---|---|
| 2. Submitter: | Healgen Scientific LLC.3818 Fuqua StreetHouston, TX 77047 |
| 3. Contact person: | Jenny XiaLSI International Inc.504 East Diamond Ave., Suite HGaithersburg, MD 20877Telephone: 301-525-6856Email: jxia@lsi-consulting.org |
| 4. Device Name: | Healgen® Accurate Muti-Drug Urine Drug Screen CupHealgen® Accurate Home Muti-Drug Urine Test Cup |
5. Classification: Class II
| Product CodeTarget Drug | Regulation Section | Panel |
|---|---|---|
| NFTAmphetamine (AMP) | 862.3100, Amphetamine Test System | Toxicology |
| PTHSecobarbital (BAR) | 862.3150, Barbiturate Test System | Toxicology |
| NGLBuprenorphine (BUP)Morphine (OPI)Oxycodone (OXY)6-Monoacetylmorphine(6-MAM)Norfentanyl (FEN)Tramadol (TML) | 862.3650, Opiate Test System | Toxicology |
| NFVOxazepam (BZO) | 862.3170, Benzodiazepine Test System | Toxicology |
| NFYCocaine (COC) | 862.3250, Cocaine and cocainemetabolite test system | Toxicology |
| PTG2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP)Methadone (MTD) | 862.3620, Methadone Test System | Toxicology |
| NGG | 862.3610,Methamphetamine Test System | Toxicology |
{6}------------------------------------------------
| Methylenedioxymethamphetamine(MDMA)Methamphetamine (MET) | ||
|---|---|---|
| NGMPhencyclidine (PCP) | Unclassified | Toxicology |
| QBFPropoxyphene(PPX) | 862.3700 Propoxyphene test system. | Toxicology |
| QAWNortriptyline (TCA) | 862.3910 Tricyclic antidepressant drugstest system | Toxicology |
| NFWCannabinoids (THC 50) | 862.3870, Cannabinoids Test System | Toxicology |
Predicate Devices: 6.
Healgen Accurate Urine Drug Screen Dip Card (K240686)
7. Intended Use
The Healgen® Accurate Multi-Drug Urine Screening Cup is a rapid lateral flow immunoassay for the qualitative detection of 6-Monoacetylmorphine, d-Amphetamine, Benzoylecgonine, Buprenorphine, EDDP, Norfentanyl, Methadone, d-Methamphetamine, d/l-Methylenedioxymethamphetamine, Morphine, Nortriptyline, Oxazepam, Oxycodone, Phencyclidine, d-Propoxyphene, Secobarbital, Tramadol in human urine. The test cut-off concentrations and the compounds the THC-COOH and tests are calibrated to are as follows:
| Test | Calibrator | Cut-off (ng/mL) |
|---|---|---|
| 6-MAM | 6-Monoacetylmorphine | 10 |
| AMP | d-Amphetamine | 500 / 1000 |
| BAR | Secobarbital | 300 |
| BUP | Buprenorphine | 10 |
| BZO | Oxazepam | 300 |
| COC | Benzoylecgonine | 150 / 300 |
| EDDP | 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrolidine | 300 |
| FEN or FYL | Norfentanyl | 5 |
| MDMA | Methylenedioxymethamphetamine | 500 |
| MET | d-Methamphetamine | 500 / 1000 |
| MTD | d/l-Methadone | 300 |
| OPI | Morphine | 300 / 2000 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | d-Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | 11-nor-Δ9-THC-COOH | 50 |
{7}------------------------------------------------
| TRA or TML | Tramadol | 100 |
|---|---|---|
| ------------ | ---------- | ----- |
The single or multi-test cups can consist of up to eighteen (18) of the above listed analytes in any combination with or without on-board adulteration/specimen validity tests (SVT).
The tests provide only a preliminary result. A more specific alternative chemical must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (CC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
The Healgen® Accurate Home Multi-Drug Urine Test Cup is a rapid qualitative immunoassay.
| CODE | SUBSTANCE | CUT-OFF(ng/mL) |
|---|---|---|
| AMP | Amphetamine | 1000 or 500 |
| BUP | Buprenorphine | 10 |
| BAR | Secobarbital | 300 |
| BZO | Oxazepam | 300 |
| COC | Cocaine | 300 or 150 |
| EDDP | EDDP | 300 |
| FYL | Norfentanyl | 5 |
| MET/mAMP | Methamphetamine | 1000 or 500 |
| MDMA | Ecstasy | 500 |
| OPI | Morphine | 2000 or 300 |
| MTD | Methadone | 300 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | Marijuana | 50 |
| TML | Tramadol | 100 |
| 6-MAM | 6-Monoacetylmorphine | 10 |
The device provides preliminary results for the detection of one or more of the following drugs.
This drug test cup may contain any combination of the drug tests listed in the table above. This test provides only preliminary result. A more specific alternative chemical method must be used to obtain a confirmed positive result. Gas Chromatography-Mass Spectrometry (GC-MS), Liquid Chromatography-Mass Spectrometry (LC-MS), and their tandem massspectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
8. Device Description
Healgen® Accurate Home Muti-Drug Urine Test Cup and Healgen® Accurate Muti-Drug Urine
{8}------------------------------------------------
Drug Screen Cup are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The device is a cup format. Each test device is sealed with two sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.
| Similarities | ||
|---|---|---|
| Item | Device | Predicate(K240686) |
| Intended use | Qualitative detection of drugs of abuse in urine. For prescription use or over-the-counter use | Same. |
| Methodology | Competitive binding, lateral flowimmunochromatographic assay based on antigen-antibody reaction | Same |
| Type of Test | Qualitative | Same |
| Specimen Type | Human urine | Same |
| Target Drugand Cut OffValues | Target Drugs | Cutoff(ng/mL) |
| Amphetamine(AMP) | 1000 or 500 | |
| Secobarbital (BAR) | 300 | |
| Buprenorphine (BUP) | 10 | |
| Oxazepam (BZO) | 300 | |
| Cocaine (COC) | 300 or 150 | |
| 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine (EDDP) | 300 | |
| Norfentanyl | 5 | |
| Methylenedioxymethamphetamine(MDMA) | 500 | |
| Methamphetamine (MET) | 1000 or 500 | |
| Morphine (MOP300/OPI2000) | 2000 or 300 | |
| Methadone (MTD) | 300 | |
| Oxycodone (OXY) | 100 | |
| Phencyclidine (PCP) | 25 | |
| Propoxyphene(PPX) | 300 | |
| Nortriptyline (TCA) | 1000 | |
| Same except noNorfentanyl andTramadol |
Substantial Equivalence Information 9.
{9}------------------------------------------------
| Cannabinoids (THC) | 50 |
|---|---|
| 6-Monoacetylmorphine(6-MAM) | 10 |
| Tramadol | 100 |
| Differences | |
| Configurations | |
| Test cup | Test Dip Card |
10. Standard/Guidance Document Reference (if applicable)
None referenced.
11. Test Principle
Healgen® Accurate Home Muti-Drug Urine Test Cup or Healgen® Accurate Muti-Drug Urine Drug Screen Cup is a competitive immunoassay that is used to screen for the presence of various drugs and drug metabolites in urine. It is chromatographic absorbent device in which, drugs within a urine sample, competitively combined to a limited number of drug monoclonal antibody (mouse) conjugate binding sites.
When the test is activated, the urine is absorbed into each test strip by capillary action, mixes with the respective drug monoclonal antibody conjugate, and flows across a pre-coated membrane. When drug within the urine sample is below the detection level of the test, respective drug monoclonal antibody conjugate binds to the respective drug-protein conjugate immobilized in the Test Region (T) of the test strip. This produces a colored Test line in the Test Region (T) of the strip, which, regardless of its intensity, indicates a negative test result.
When sample drug levels are at or above the detection level of the test, the free drug in the sample binds to the respective drug monoclonal antibody conjugate, preventing the respective drug monoclonal antibody conjugate from binding to the respective drug-protein conjugate immobilized in the Test Region (T) of the device. This prevents the development of a distinct colored band in the test region, indicating a preliminary positive result.
To serve as a procedure control, a colored line will appear at the Control Region (C) of each strip, if the test has been performed properly.
12. Performance Characteristics
A. Analytical performance
a. Precision/Reproducibility:
Precision studies were carried out for samples with concentrations of +100% cutoff, +75% cutoff, +50% cutoff, +25% cutoff, cutoff, -25% cutoff, -50% cutoff, -75% cut off and -100% cutoff. Other samples were prepared by spiked target drug in drug-free urine samples. Each drug concentration was confirmed by LC-MS/MS. For each concentration, tests were performed two runs per day for 25 days using three lots of test cups with one operator per lot. The results obtained are summarized in the following tables:
{10}------------------------------------------------
| Drug | LotNumber | +100%cutoff | +75%cutoff | +50%cutoff | +25%cutoff | Cutoff | -25%cutoff | -50%cutoff | -75%cutoff | -100%cut-off |
|---|---|---|---|---|---|---|---|---|---|---|
| AMP1000 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 25-/25+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 26-/24+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| BAR300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 25-/25+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 25-/25+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 25-/25+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| BUP10 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 26-/24+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 27-/23+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| BZO300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 26-/24+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 22-/28+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| COC300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 25-/25+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 25-/25+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| EDDP300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| FYL5 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 24-/26+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 25-/25+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 24-/26+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| MDMA500 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 22-/28+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 25-/25+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 23-/27+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| MET1000 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| OPI300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 25-/25+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| MTD300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 23-/27+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 25-/25+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| OXY100 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| PCP25 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 23-/27+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 23-/27+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | ||||||||||
| PPX | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| 300 | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 23-/27+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| TCA | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| 1000 | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| THC | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| 50 | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 22-/28+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 23-/27+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| TML | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| 100 | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 25-/25+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 26-/24+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| 6-MAM | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 26-/24+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| 10 | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 24-/26+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| AMP | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| 500 | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 25-/25+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| COC | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 25-/25+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| 150 | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 22-/28+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 2-/48+ | 23-/27+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| MET | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| 500 | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 25-/25+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 25-/25+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| OPI | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 26-/24+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| 2000 | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 25-/25+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
{11}------------------------------------------------
b. Linearity/assay reportable range:
Not applicable. This device is intended for qualitative use only.
c. Stability:
The device is stable at 2-30°C for 36 months based on real time stability study.
d. Analytical specificity/Interference:
To test the specificity, drug metabolites and other components that are likely to cross-react in urine samples were spiked into drug-free urine samples were tested using three lots of the device.
Percent cross-reactivity, provided in the below table, was calculated as the cutoff concentration divided by the concentration of analyte tested that yielded a positive result, multiplied by 100.
{12}------------------------------------------------
| Drug/Cutoff | Compound | Minimumconcentrationrequired to obtaina positive result(ng/mL) | % Cross-Reactivity |
|---|---|---|---|
| AMP 1000 | Hydroxyamphetamine | 8000 | 12.5% |
| (+/-)- | 400 | 250% | |
| Methylenedioxyamphetamine(MDA) | |||
| D,L-Amphetamine | 1,000 | 100% | |
| D-Amphetamine | 1,000 | 100% | |
| Diethylstilbestrol | 5,000 | 20% | |
| L-Amphetamine | 50,000 | 2% | |
| Phentermine | 8,000 | 12.5% | |
| β-Phenylethylamine | 100,000 | 1% | |
| Tyramine | 100,000 | 1% | |
| p-Hydroxynorephedrine | 100,000 | 1% | |
| D,L-Norephedrine | 100,000 | 1% | |
| p-Hydroxyamphetamine | 100,000 | 1% | |
| D-Methamphetamine | 100,000(Negative) | Not detected | |
| L-Methamphetamine | 100,000(Negative) | Not detected | |
| Ephedrine hydrochloride | 100,000(Negative) | Not detected | |
| (+/-)3,4- | |||
| Methylenedioxymethamphetamine(MDMA) | 100,000(Negative) | Not detected | |
| Phenylpropanolamine | 100,000(Negative) | Not detected | |
| Benzphetamine | 100,000(Negative) | Not detected | |
| L-Ephedrine | 100,000(Negative) | Not detected | |
| L-Epinephrine | 100,000(Negative) | Not detected | |
| D,L-Epinephrine | 100,000(Negative) | Not detected | |
| (+/-)3,4- | |||
| Methylenedioxyethylamphetamine(MDEA) | 100,000(Negative) | Not detected | |
| Alphenal | 150 | 200% | |
| BAR 300 | Amobarbital | 300 | 100% |
| Aprobarbital | 250 | 120% | |
| Butabarbital | 2,500 | 12% | |
| Butethal | 100 | 300% | |
| Cyclopentobarbital | 600 | 50% | |
| Pentobarbital | 250 | 120% | |
| Phenobarbital | 250 | 120% | |
| Secobarbital | 300 | 100% | |
| Butalbital | 2,500 | 12% | |
| Buprenorphine | 10 | 100% | |
| Buprenorphine -3-D-Glucuronide | 160 | 6.25% | |
| Norbuprenorphine | 10 | 100% | |
| BUP 10 | Norbuprenorphine-3-D-Glucuronide | 200 | 5% |
| Morphine | 100000 (Negative) | Not Detected | |
| Oxymorphone | 100000 (Negative) | Not Detected | |
| Hydromorphone | 100000 (Negative) | Not Detected | |
| a-Hydroxyalprazolam | 1,260 | 23.8% | |
| Alprazolam | 200 | 150% | |
| Bromazepam | 1,560 | 19.2% | |
| Chlordiazepoxide | 1,560 | 19.2% | |
| Clobazam | 100 | 300% | |
| Clonazepam | 2,500 | 12% | |
| Clorazepate Dipotassium | 200 | 150% | |
| Desalkylflurazepam | 400 | 75% | |
| Diazepam | 200 | 150% | |
| BZO 300 | Estazolam | 2,500 | 12% |
| Flunitrazepam | 400 | 75% | |
| D,L-Lorazepam | 1,560 | 19.2% | |
| Midazolam | 12,500 | 2.4% | |
| Nitrazepam | 100 | 300% | |
| Norchlordiazepoxide | 200 | 150% | |
| Nordiazepam | 400 | 75% | |
| Oxazepam | 300 | 100% | |
| Oxazepam glucuronide | 500 | 60% | |
| R,S-Lorazepam glucuronide | 160 | 187.5% | |
| Temazepam | 100 | 300% | |
| Triazolam | 2,500 | 12% | |
| Demoxepam | 2,000 | 15% | |
| Flurazepam | 500 | 60% | |
| Delorazepam | 100,000(negative) | Not detected | |
| COC 300 | Benzoylecogonine | 300 | 100% |
| Cocaethylene | 300 | 100% | |
| Cocaine hydrochloride | 300 | 100% | |
| COC 300 | Ecgonine | 50,000 | 0.6% |
| Norcocaine | 100,000 | 0.3% | |
| Ecgonine methyl ester | 100,000(negative) | Not detected | |
| 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine | 300 | 100% | |
| Methadone | 100,000(negative) | Not detected | |
| EMDP | 100,000(negative) | Not detected | |
| EDDP 300 | Doxylamine | 100,000(negative) | Not detected |
| Levacetylmethadol (LAAM) | 100,000(negative) | Not detected | |
| Disopyramide | 100,000(negative) | Not detected | |
| Alpha Methadol | 100,000(negative) | Not detected | |
| Norfentanyl | 5 | 100% | |
| Fentanyl | 10 | 50% | |
| Acetyl fentanyl | 150 | 3.3% | |
| Acetyl Norfentanyl | 200 | 2.5% | |
| (+)-β-Hydroxythiofentanyl HCl | 2,500 | 0.2% | |
| Acryl Fentanyl | 2,500 | 0.2% | |
| Butyryl Fentanyl | 5,000 | 0.1% | |
| FYL 5 | Cis-d, I 3-Methylfentanyl | 50,000 | 0.01% |
| Furanyl Fentanyl | 10,000 | 0.05% | |
| Para-fluoro butyrl Fentanyl (P-FBF) | 80,000 | 0.01% | |
| Para-fluoro Fentanyl | 40,000 | 0.005% | |
| 9-HydroxyRisperidone | 10,000 | 0.05% | |
| Alfentanil | 20,000 | 0.03% | |
| Isobutyryl Fentanyl | 5,000 | 0.1% | |
| Trans-d. 1-3-Methylfentanyl | 50 | 10% | |
| Norcarfentail Oxalate | 50,000 | 0.01% | |
| Remifentanil | 15,000 | 0.03% | |
| Valeryl Fentanyl | 20,000 | 0.03% | |
| Thienyl Fentanyl | 50 | 10% | |
| 4-Fluoro-isobutyryl Fentanyl | 20,000(Negative) | Not detected | |
| Despropionyl fentanyl (4-ANPP) | 20,000(Negative) | Not detected | |
| MT-45 | 100,000(Negative) | Not detected | |
| Ocfentanil | 100,000(Negative) | Not detected | |
| Risperidone | 100,000(Negative) | Not detected | |
| Sufentanil | 100,000(Negative) | Not detected | |
| Carfentanil Oxalate | 100,000(Negative) | Not detected | |
| Labetalol Hydrochloride | 100,000(Negative) | Not detected | |
| Trazodone | 100,000(Negative) | Not detected | |
| U-47700 | 100,000(Negative) | Not detected | |
| ω-1-Hydroxyfentanyl | 20,000(Negative) | Not detected | |
| 6-Acetyl morphine | 100,000(Negative) | Not detected | |
| Amphetamine | 100,000(Negative) | Not detected | |
| Buprenorphine | 100,000(Negative) | Not detected | |
| Buprenorphine -3-D-Glucuronide | 100,000(Negative) | Not detected | |
| Codeine | 100,000(Negative) | Not detected | |
| Dextromethorphan | 100,000(Negative) | Not detected | |
| Dihydrocodeine | 100,000(Negative) | Not detected | |
| EDDP | 100,000(Negative) | Not detected | |
| EMDP | 100,000(Negative) | Not detected | |
| Fluoxetine | 100,000(Negative) | Not detected | |
| Heroin | 100,000(Negative) | Not detected | |
| Hydrocodone | 100,000(Negative) | Not detected | |
| Hydromorphone | 100,000(Negative) | Not detected | |
| Ketamine | 100,000(Negative) | Not detected | |
| Levorphanol tartrate | 100,000(Negative) | Not detected | |
| Meperidine | 100,000(Negative) | Not detected | |
| Methadone | 100,000(Negative) | Not detected | |
| Morphine | 100,000(Negative) | Not detected | |
| Morphine-3-β-D-Glucuronide | 100,000(Negative) | Not detected | |
| Naloxone hydrochloride | 100,000(Negative) | Not detected | |
| Naltrexone hydrochloride | 100,000(Negative) | Not detected | |
| Norbuprenorphine | 100,000(Negative) | Not detected | |
| Norcodeine | 100,000(Negative) | Not detected | |
| Norketamine | 100,000(Negative) | Not detected | |
| Normeperidine | 100,000(Negative) | Not detected | |
| Normorphine | 100,000(Negative) | Not detected | |
| Noroxycodone | 100,000(Negative) | Not detected | |
| Oxycodone | 100,000(Negative) | Not detected | |
| Oxymorphone | 100,000(Negative) | Not detected | |
| Pentazocine (Talwin) | 100,000(Negative) | Not detected | |
| Pipamperone | 100,000(Negative) | Not detected | |
| Tapentadol hydrochloride | 100,000(Negative) | Not detected | |
| Thioridazine | 100,000(Negative) | Not detected | |
| Tilidine | 100,000(Negative) | Not detected | |
| Tramadol | 100,000(Negative) | Not detected | |
| o-Desmethyl Tramadol | 100,000(Negative) | Not detected | |
| n-Desmethyl Tramadol | 100,000(Negative) | Not detected | |
| MDMA 500 | (+/-)3,4-MethylenedioxymethamphetamineHCI(MDMA) | 500 | 100% |
| (+/-)3,4-MethylenedioxyamphetamineHCI (MDA) | 3,000 | 16.7% | |
| (+/-)3,4-Methylenedioxyethylamphetamine(MDEA) | 300 | 167% | |
| L-Methamphetamine | 50,000 | 1% | |
| 7-Aminoclonazepam | 100,000 | 0.5% | |
| d-methamphetamine | 100,000(Negative) | Not detected | |
| d-amphetamine | 100,000(Negative) | Not detected | |
| l-amphetamine | 100,000(Negative) | Not detected | |
| MET 1000 | (+/-)3,4-Methylenedioxy-n-ethylamphetamine (MDEA) | 20,000 | 5% |
| (+/-)3,4- | 2,500 | 40% | |
| Methylenedioxymethamphetamine(MDMA) | |||
| D-Methamphetamine | 1,000 | 100% | |
| L-Methamphetamine | 25,000 | 4% | |
| Fenfluramine | 50,000 | 2% | |
| p-Hydroxymethamphetamine | 10,000 | 10% | |
| D,L-Methamphetamine | 1,000 | 100% | |
| β-Phenylethylamine | 50,000 | 2% | |
| Mephetermine | 50,000 | 2% | |
| Methoxyphenamine hydrochloride | 50,000 | 2% | |
| L-Amphetamine | 75,000 | 1.33% | |
| D-Amphetamine | 100,000(Negative) | Not detected | |
| D,L-Amphetamine | 100,000(Negative) | Not detected | |
| Chloroquine | 100,000(Negative) | Not detected | |
| Ephedrine hydrochloride | 100,000(Negative) | Not detected | |
| (+/-)3,4-Methylenedioxyamphetamine(MDA) | 100,000(Negative) | Not detected | |
| Trimethobenzamide | 100,000(Negative) | Not detected | |
| l-phenylephrine | 100,000(Negative) | Not detected | |
| (1R,2S)-(-)-Ephedrine | 100,000(Negative) | Not detected | |
| Procaine hydrochloride | 100,000(Negative) | Not detected | |
| 6-acetylmorphine | 2,500 | 80% | |
| Codeine | 1,000 | 200% | |
| Dihydrocodeine | 1,500 | 133.3% | |
| OPI 2000 | EthylMorphine | 2,500 | 80% |
| Codeine-6-β-D-glucuronide | 2,500 | 80% | |
| Heroin | 5,000 | 40% | |
| Hydrocodone | 5,000 | 40% | |
| Hydromorphone | 2,500 | 80% | |
| Levorphanol tartrate | 10,000 | 20% | |
| Morphine | 2,000 | 100% | |
| Nalorphine hydrochloride | 5,000 | 40% | |
| Norcodeine | 4,000 | 50% | |
| Normorphine | 5.000 | 40% | |
| Oxymorphone | 75,000 | 2.7% | |
| s-Monoacetylmorphine | 2,000 | 100% | |
| Morphine-6-β-D-glucuronide | 2,000 | 100% | |
| Thebaine | 13,000 | 15.4% | |
| Morphine 3-β-D-glucuronide | 2,000 | 100% | |
| 6-Monoacetylmorphine (6-MAM) | 1,500 | 133% | |
| 6-Acetylcodeine | 10,000 | 20% | |
| Oxycodone | 100,000(Negative) | Not detected | |
| Procaine | 100,000(Negative) | Not detected | |
| Norpropoxyphene | 100,000(Negative) | Not detected | |
| MTD 300 | EDDP | 100,000(Negative) | Not detected |
| Doxylamine | 100,000(Negative) | Not detected | |
| Levacetylmethadol (LAAM) | 100,000(Negative) | Not detected | |
| EMDP | 100,000(Negative) | Not detected | |
| Alpha Methadol | 100,000(Negative) | Not detected | |
| OXY 100 | Ethyl Oxycodone | 75,000 | 0.13% |
| Hydrocodone | 5,000 | 2% | |
| Hydromorphone | 25,000 | 0.4% | |
| Levorphanol tartrate | 25,000 | 0.4% | |
| Naloxone hydrochloride | 10,000 | 1% | |
| Naltrexone hydrochloride | 50,000 | 0.2% | |
| Oxycodone | 100 | 100% | |
| Oxymorphone | 200 | 50% | |
| Oxymorphone-3β-D-glucuronide | 500 | 20% | |
| Noroxycodone | 1,500 | 6.7% | |
| Noroxymorphone | 3,000 | 3.3% | |
| Dihydrocodeine | 100,000(Negative) | Not detected | |
| Codeine | 100,000(Negative) | Not detected | |
| Morphine | 100,000(Negative) | Not detected | |
| Buprenorphine | 100,000(Negative) | Not detected | |
| Ethylmorphine | 100,000(Negative) | Not detected | |
| Thebaine | 100,000(Negative) | Not detected | |
| 6-acetylmorphine | 100,000(Negative) | Not detected | |
| PCP 25 | Phencyclidine | 25 | 100% |
| 4-Hydroxy Phencyclidine | 12,500 | 0.2% | |
| PPX 300 | d-Propoxyphene | 300 | 100% |
| d-Norpropoxyphene | 300 | 100% | |
| TCA 1000 | Amitriptyline | 1,500 | 66.7% |
| Chlorpheniramine | 50,000 | 2% | |
| Clomipramine | 10,000 | 10% | |
| Cyclobenzaprine Hydrochloride | 5,000 | 20% | |
| Desipramine | 1,000 | 100% | |
| Doxepine | 2,000 | 50% | |
| Duloxetine | 10,000 | 10% | |
| Imipramine | 1,000 | 100% | |
| Norclomipramine | 12,500 | 8% | |
| Nordoxepine | 1,000 | 100% | |
| Nortriptyline | 1,000 | 100% | |
| Promazine | 50,000 | 2% | |
| Trimipramine | 10,000 | 10% | |
| Maprotiline | 100,000(Negative) | Not detected | |
| Promethazine hydrochloride | 100,000(Negative) | Not detected | |
| THC 50 | 11-nor-Δ8-THC -9-COOH | 30 | 166.7% |
| (-)-11-nor-9-carboxy-Δ9-THC | 50 | 100% | |
| (±)-11-nor-9-Carboxy-Δ9-THC | 50 | 100% | |
| 11-nor-Δ9-THC -carboxy glucuronide | 100 | 50% | |
| 11-hydroxy-Δ9-Tetrahydrocannabinol | 5,000 | 1% | |
| Δ8- Tetrahydrocannabinol | 1,300 | 3.8% | |
| Δ9- Tetrahydrocannabinol | 5,000 | 1% | |
| Cannabinol | 20,000 | 0.25% | |
| Cannabidiol | 100,000(Negative) | Not detected | |
| (±)-11-Hydroxy-Δ9-THC | 100,000(Negative) | Not detected | |
| TML 100 | Tramadol | 100 | 100% |
| n-Desmethyl Tramadol | 400 | 25% | |
| o-Desmethyl Tramadol | 1,000 | 10% | |
| o-Desmethyl Venlafaxine | 10,000(Negative) | Not detected | |
| Venlafaxine HCl | 100,000(Negative) | Not detected | |
| 6-MAM 10 | 6-acetylmorphine | 10 | 100% |
| 6-Monoacetylmorphine (6-MAM) | 10 | 100% | |
| Heroin | 60 | 16.7% | |
| Morphine | 75,000 | 0.01% | |
| s-Monoacetylmorphine | 10 | 100% | |
| Normorphine | 100,000(Negative) | Not detected | |
| Nalorphine HCl | 100,000(Negative) | Not detected | |
| Hydrocodone | 100,000(Negative) | Not detected | |
| Hydromorphone | 100,000(Negative) | Not detected | |
| Chlordiazepoxide | 100,000(Negative) | Not detected | |
| Clobazam | 100,000(Negative) | Not detected | |
| D-Amphetamine | 100,000(Negative) | Not detected | |
| (±)-Amphetamine | 100,000(Negative) | Not detected | |
| Levorphanol tartrate | 100,000(Negative) | Not detected | |
| Codeine | 100,000(Negative) | Not detected | |
| Ethylmorphine | 100,000(Negative) | Not detected | |
| Morphine3-β-D-glucuronide | 100,000(Negative) | Not detected | |
| Norcodeine | 100,000(Negative) | Not detected | |
| Oxycodone | 100,000(Negative) | Not detected | |
| Oxymorphone | 100,000(Negative) | Not detected | |
| Procaine hydrochloride | 100,000(Negative) | Not detected | |
| Thebaine | 100,000(Negative) | Not detected | |
| 6-Acetylcodeine | 100,000(Negative) | Not detected | |
| Buprenorphine | 100,000(Negative) | Not detected | |
| Dihydrocodeine | 100,000(Negative) | Not detected | |
| Dextromethorphan | 100,000(Negative) | Not detected | |
| Imipramine hydrochloride | 100,000(Negative) | Not detected | |
| Meperidine | 100,000(Negative) | Not detected | |
| (±)-Methadone | 100,000(Negative) | Not detected | |
| Mitragynine(kratom) | 100,000(Negative) | Not detected | |
| Morphine-6-B-D-glucuronide | 100,000(Negative) | Not detected | |
| Naloxone hydrochloride | 100,000(Negative) | Not detected | |
| Naltrexone hydrochloride | 100,000(Negative) | Not detected | |
| Naproxen | 100,000(Negative) | Not detected | |
| Norbuprenorphine | 100,000(Negative) | Not detected | |
| Norbuprenorphine-3-D-Glucuronide | 100,000(Negative) | Not detected | |
| Noroxycodone HCL | 100,000(Negative) | Not detected | |
| Noroxymorphone HCL | 100,000(Negative) | Not detected | |
| (+)-Norpropoxyphene maleate | 100,000(Negative) | Not detected | |
| Oxymorphone-3β-D-glucuronide | 100,000(Negative) | Not detected | |
| Tapentadol HCl | 100,000(Negative) | Not detected | |
| Tramadol | 100,000(Negative) | Not detected | |
| AMP 500 | Hydroxyamphetamine | 4,000 | 12.5% |
| (+/-)-Methylenedioxyamphetamine(MDA) | 200 | 250% | |
| D,L-Amphetamine | 500 | 100% | |
| D-Amphetamine | 500 | 100% | |
| Diethylstilbestrol | 2,500 | 20% | |
| L-Amphetamine | 25,000 | 2% | |
| Phentermine | 4,000 | 12.5% | |
| β-Phenylethylamine | 50,000 | 1% | |
| Tyramine | 50,000 | 1% | |
| p-Hydroxynorephedrine | 50,000 | 1% | |
| p-Hydroxyamphetamine | 50,000 | 1% | |
| D,L-Norephedrine | 50,000 | 1% | |
| (+/-)3,4-Methylenedioxy-n-ethylamphetamine(MDEA) | 100,000 | 0.5% | |
| d-Methamphetamine | 100,000(Negative) | Not detected | |
| l-Methamphetamine | 100,000(Negative) | Not detected | |
| (+/-)3,4-Methylenedioxymethamphetamine(MDMA) | 100,000(Negative) | Not detected | |
| Ephedrine hydrochloride | 100,000(Negative) | Not detected | |
| Phenylpropanolamine | 100,000(Negative) | Not detected | |
| Benzphetamine | 100,000(Negative) | Not detected | |
| L-Ephedrine | 100,000(Negative) | Not detected | |
| L-Epinephrine | 100,000(Negative) | Not detected | |
| D,L-Epinephrine | 100,000(Negative) | Not detected | |
| COC 150 | Benzoylecogonine | 150 | 100% |
| Cocaethylene | 150 | 100% | |
| Cocaine hydrochloride | 150 | 100% | |
| Ecgonine | 25,000 | 0.6% | |
| Norcocaine | 50,000 | 0.3% | |
| Ecgonine methyl Ester | 100,000(negative) | Not detected | |
| MET 500 | (+/-)3,4-Methylenedioxy-n-ethylamphetamine (MDEA) | 10,000 | 5% |
| (+/-)3,4-methylenedioxumethamphetamine(MDMA) | 1,250 | 40% | |
| D-Methamphetamine | 500 | 100% | |
| L-Methamphetamine | 12,500 | 4% | |
| Fenfluramine | 25,000 | 2% | |
| p-Hydroxymethamphetamine | 5,000 | 10% | |
| D,L-Methamphetamine | 500 | 100% | |
| β-Phenylethylamine | 25,000 | 2% | |
| Mephetermine | 25,000 | 2% | |
| Methoxyphenamine hydrochloride | 25,000 | 2% | |
| L-Amphetamine | 40,000 | 1.25% | |
| Ephedrine hydrochloride | 100,000 | 0.5% | |
| (1R,2S)-(-)-Ephedrine | 100,000 | 0.5% | |
| D-Amphetamine | 100,000(negative) | Not detected | |
| Chloroquine | 100,000(negative) | Not detected | |
| (+/-)3,4-Methylenedioxyamphetamine(MDA) | 100,000(negative) | Not detected | |
| L-Phenylephrine | 100,000(negative) | Not detected | |
| Trimethobenzamide | 100,000(negative) | Not detected | |
| Procaine hydrochloride | 100,000(negative) | Not detected | |
| d/l-Amphetamine | 100,000(negative) | Not detected | |
| OPI 300 | 6-acetylmorphine | 400 | 75% |
| Codeine | 300 | 100% | |
| Dihydrocodeine | 1,000 | 30% | |
| EthylMorphine | 100 | 300% | |
| Heroin | 600 | 50% | |
| Hydrocodone | 500 | 60% | |
| Hydromorphone | 1,000 | 30% | |
| Levorphanol tartrate | 10,000 | 3% | |
| Morphine | 300 | 100% | |
| Nalorphine hydrochloride | 50,000 | 0.6% | |
| Thebaine | 6,240 | 4.8% | |
| s-Monoacetylmorphine | 300 | 100% | |
| Morphine-3-β-d-glucuronide | 1,000 | 30% | |
| 6-Monoacetylmorphine (6-MAM) | 150 | 200% | |
| Codeine-6-ß-D-glucuronide | 150 | 200% | |
| Morphine-6-β-D-glucuronide | 150 | 200% | |
| 6-Acetylcodeine | 900 | 33.3% | |
| Normorphine | 100,000(negative) | Not detected | |
| Oxycodone | 100,000(negative) | Not detected | |
| Oxymorphone | 100,000(negative) | Not detected | |
| Norcodeine | 100,000(negative) | Not detected | |
| Procaine | 100,000(negative) | Not detected | |
| Norpropoxyphene | 100,000(negative) | Not detected |
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
{17}------------------------------------------------
{18}------------------------------------------------
{19}------------------------------------------------
{20}------------------------------------------------
{21}------------------------------------------------
{22}------------------------------------------------
{23}------------------------------------------------
To evaluate potential interference, non-structurally related compounds were added to drug-free urine and to urine samples containing the target drugs at 50% below and 50% above each corresponding cutoff.
Compounds that show no interference at a concentration of 100µg/mL are summarized in the following table.
| (-) Cotinine | Diclofenac sodium | Nitroglycerin |
|---|---|---|
| 3-Hydroxytyramine | Diflunisal | Norethindrone |
| 7-Aminoclonazepam (except MDMA test) | Digoxin | Norpropoxyphene |
| 7-Aminoflunitrazepam | Dimethyl-aminoantipyrine | Norpseudoephedrine |
| 7-Aminonitrazepam | Diphenhydramine HCl | Nortriptyline (except TCA test) |
| Acetaminophen | Diphenylhydantoin | Noscapine |
| Acetone (1000 mg/dL) | Disopyramide | Octopamine |
| Acetophenetidin | Dopamine HCl | O-Hydroxyhippuric acid |
| Acetylsalicylic acid | Doxepine (except TCA test) | Olanzapine |
| Acyclovir | Doxylamine | Omeprazole |
| Albumin(100mg/dL) | D-Pseudoephedrine | Oxalic acid (100mg/dL) |
| Albuterol | Duloxetine (except TCA test) | Oxazepam (except BZO test) |
{24}------------------------------------------------
| Albuterol sulfate(Proair HFA) | Ecgonine methyl ester | Oxazepam Glucuronide (exceptBZO test) |
|---|---|---|
| Alpha Methadol | EMDP | Oxalinic acid |
| Aminophylline | Ephedrine hydrochloride(except MET test) | Oxymetazoline |
| Aminopyrine | Erythromycin | Paliperidone |
| Amitriptyline(except TCA test) | Esomeprazole Magnesium | Papaverine |
| Amlodipine besylate | Estradiol | Penicillin-G |
| Amobarbital(except BAR test) | Estrone | PenicillinV Potassium |
| Amoxicillin | Ethanol(1%) | Perphenazine |
| Ampicillin | Fenfluramine(except MET test) | Phenacetin |
| Apomorphine | Fenofibrate | Phencyclidine (except PCP test) |
| Aripiprazole | Fenoprofen | Phenelzine |
| Ascorbic acid | Fluoxetine Hydrochloride | Phenylethylamine |
| Aspartame | Fluphenazine | Phenobarbital (except BAR test) |
| Aspirin | Fotemustine | Phentermine (except AMP test) |
| Atomoxetine | Furosemide | Phenylpropanolamine |
| Atorvastatin Calcium | Gabapentin | Prednisone |
| Atropine | Galactose | Pregablin |
| Azithromycin | Gatifloxacin | Procaine |
| Baclofen | Gemfibrozil | Promazine (except TCA test) |
| Benzilic acid | Gentisic acid | Promethazine |
| Benzocaine | Glucose (3000mg/dL) | Propoxyphene (except PPX test) |
| Benzoic acid | Guaiacolglyceryl ether | Propranolol |
| Benzoylecgonine (except COC test) | Hemoglobin | Pseudoephedrine |
| Benzphetamine | Hexobarbital | Pyridoxine |
| Benzylpiperiazine | Hydralazine | Pyrilamine |
| Bilirubin | Hydrochlorothiazide | Pyrogallol |
| Boric Acid (1%) | Hydrocortisone | Quetiapine |
| Bromo-2,5,Dimethoxyphenethylamine | Hydroxybutyric Acid | Quinidine |
| Bupropion | Ibuprofen | Quinine |
| Caffeine | Imipramine(except TCA test) | Quinolinic Acid |
| Cannabidiol | Isoproterenol | Ranitidine |
| Captopril | Isoxsuprine | Riboflavin |
| Carbamazepine | Ketamine | Rifampicin |
| Carfentanil | Ketoprofen | Salicylic acid |
| Carisoprodol | LAAM HC1 | Secobarbital (except BAR test) |
| Cefradine | Labetalol | Serotonin |
| Cephalexin | Lamotrigine | Serotonin (5-Hydroxytyramine) |
| Cetirizine | L-Ephedrine | Sertraline |
| Chloral hydrate | L-Epinephrine | Sildenafil Citrate |
| Chloramphenicol | Levofloxacin Hydrochloride | Simvastatin |
| Chlordiazepoxide(except BZO test) | Levonorgestrel | Sodium Azide |
| Chloroquine | Levothyroxine Sodium | Sulfamethazine |
| Chlorothiazide | Lidocaine Hydrochloride | Sulindac |
| Chlorpheniramine(except TCA test) | Lisinopril | Telmisartan |
| Chlorpromazine | Loperamide | Tetracycline |
| Cholesterol | Loratadine | Tetrahydrocortisone 3-(β-D-glucuronide) |
| Ciprofloxacin Hydrochloride | Lorazepam Glucuronide(except BZO test) | Tetrahydrocortisone, 3-acetate |
| Citalopram | L-phenylephrine | Tetrahydrozoline |
| Clarithromycin | LSD | THC (except THC test) |
| Clofibrate | L-thyroxine | Theophylline |
| Clomipramine(except TCA test) | Magnesium | Thiamine |
| Clonidine | Maprotiline | Thioridazine |
| Clozapine | Meperidine | Trazodone Hydrochloride |
| Conjugated Estrogens | Meprobamate | Triamterene |
| Cortisone | Metformin | Trifluoperazine |
| Creatine Hydrate | Methapyrilene | Trifluoromethylphenyl-piperazine |
| Creatinine | Methaqualone | Trimethobenzamide |
| Cyclobenzaprine(except TCA test) | Methoxyphenamine (exceptMET test) | Trimethoprim |
| Cyproheptadine | Methylphenidate | Tryptamine |
| D,L-Epinephrine | Metoprolol Tartrate | Tyramine (except AMP test) |
| D,L-Isoproterenol | Metronidazole | Urea (2000 mg/dL) |
| D,L-Lorazepam (except BZO test) | Mifepristone | Uric acid |
| D,L-Octopamine | N-Acetylprocainamide | Valproic acid (250 µg/mL) |
| D,L-Propranolol | NaCl (4000 mg/dL) | Venlafaxine HCl |
| D,L-Tryptophan | Nalidixic acid | Verapamil |
| D,L-Tyrosine | Naloxone hydrochloride(except OXY test) | Vitamin B2 |
| Delorazepam | Naltrexonehydrochloride(except OXYtest) | Vitamin C |
| Demoxepam(except BZO test) | Naproxen | Zaleplon |
| Deoxycorticosterone | N-desmethyl Tapentadol | Zolpidem |
| Desloratadine | Niacinamide | Zomepirac |
| Desipramine(except TCA test) | Nicotine | β-Estradiol |
| Dextromethorphan | Nicotinic Acid | γ-Cyclodextrin |
| Diclofenac | Nifedipine | γ-Globulin (500mg/dL) |
{25}------------------------------------------------
Interference by pH and specific gravity were also evaluated using pooled urine specimens with concentrations of 0 (drug-free), at 50% below and 50% above each corresponding cutoff. The
{26}------------------------------------------------
results demonstrated that pH levels of 4 to 9 and specific gravity levels of 1.000 to 1.035 do not affect the results of the assays.
B. Method comparison study
The method comparison studies for the device were performed in-house with three operators. Operators ran 80 (40 negative and 40 positive) unaltered urine clinical samples for each drug. The samples were blind labeled and compared to LC-MS/MS results. The results are presented in the table below:
| Drug test | Test Result | Drug-Free | LowNegative byLC-MS/MS(less than -50%) | Near CutoffNegative byLC-MS/MS(Between -50% and theCutoff) | Near CutoffPositive byLC-MS/MS(Betweenthe cutoffand +50%) | High Positiveby LC-MS/MS(greater than+50%) | |
|---|---|---|---|---|---|---|---|
| AMP(1000) | Operator | + | 0 | 0 | 0 | 17 | 22 |
| A | - | 10 | 14 | 16 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 17 | 22 | |
| B | - | 10 | 14 | 15 | 1 | 0 | |
| Operator | + | 0 | 0 | 0 | 17 | 22 | |
| C | - | 10 | 14 | 16 | 1 | 0 | |
| AMP(500) | Operator | + | 0 | 0 | 1 | 18 | 21 |
| A | - | 10 | 14 | 15 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 18 | 21 | |
| B | - | 10 | 14 | 15 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 18 | 21 | |
| C | - | 10 | 14 | 15 | 1 | 0 | |
| BAR | Operator | + | 0 | 0 | 1 | 11 | 28 |
| A | - | 10 | 15 | 14 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 12 | 28 | |
| B | - | 10 | 15 | 14 | 0 | 0 | |
| Operator | + | 0 | 0 | 1 | 11 | 28 | |
| C | - | 10 | 15 | 14 | 1 | 0 | |
| BUP | Operator | + | 0 | 0 | 2 | 18 | 20 |
| A | - | 10 | 16 | 12 | 2 | 0 | |
| Operator | + | 0 | 0 | 1 | 18 | 20 | |
| B | - | 10 | 16 | 13 | 2 | 0 | |
| Operator | + | 0 | 0 | 2 | 18 | 20 | |
| C | - | 10 | 16 | 12 | 2 | 0 | |
| BZO | Operator | + | 0 | 0 | 1 | 22 | 16 |
| A | - | 10 | 17 | 12 | 2 | 0 | |
| Operator | + | 0 | 0 | 1 | 22 | 16 | |
| B | - | 10 | 17 | 12 | 2 | 0 | |
| Operator | + | 0 | 0 | 1 | 22 | 16 | |
| C | - | 10 | 17 | 12 | 2 | 0 | |
| COC | Operator | + | 0 | 0 | 1 | 16 | 23 |
| (300) | A | - | 10 | 16 | 13 | 1 | 0 |
| Operator | + | 0 | 0 | 2 | 16 | 23 | |
| B | - | 10 | 16 | 12 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 16 | 23 | |
| C | - | 10 | 16 | 13 | 1 | 0 | |
| COC | Operator | + | 0 | 0 | 1 | 21 | 18 |
| (150) | A | - | 10 | 16 | 13 | 1 | 0 |
| Operator | + | 0 | 0 | 1 | 20 | 18 | |
| B | - | 10 | 16 | 13 | 2 | 0 | |
| Operator | + | 0 | 0 | 1 | 21 | 18 | |
| C | - | 10 | 16 | 13 | 1 | 0 | |
| EDDP | Operator | + | 0 | 0 | 1 | 16 | 23 |
| A | - | 10 | 16 | 13 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 15 | 23 | |
| B | - | 10 | 16 | 13 | 2 | 0 | |
| Operator | + | 0 | 0 | 0 | 17 | 23 | |
| C | - | 10 | 16 | 14 | 0 | 0 | |
| FYL | Operator | + | 0 | 0 | 2 | 15 | 24 |
| A | - | 10 | 16 | 12 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 14 | 24 | |
| B | - | 10 | 16 | 13 | 2 | 0 | |
| Operator | + | 0 | 0 | 1 | 15 | 24 | |
| C | - | 10 | 16 | 13 | 1 | 0 | |
| MDMA | Operator | + | 0 | 0 | 2 | 16 | 23 |
| A | - | 10 | 15 | 13 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 16 | 23 | |
| B | - | 10 | 15 | 14 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 16 | 23 | |
| C | - | 10 | 15 | 14 | 1 | 0 | |
| MET | Operator | + | 0 | 0 | 0 | 19 | 20 |
| (1000) | A | - | 10 | 17 | 13 | 1 | 0 |
| Operator | + | 0 | 0 | 1 | 19 | 20 | |
| B | - | 10 | 17 | 12 | 1 | 0 | |
| Operator | + | 0 | 0 | 0 | 19 | 20 | |
| C | - | 10 | 17 | 13 | 1 | 0 | |
| MET | Operator | + | 0 | 0 | 1 | 23 | 17 |
| (500) | A | - | 10 | 15 | 14 | 0 | 0 |
| Operator | + | 0 | 0 | 1 | 22 | 17 | |
| B | - | 10 | 15 | 14 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 22 | 17 | |
| C | - | 10 | 15 | 14 | 1 | 0 | |
| OPI (300) | Operator | + | 0 | 0 | 1 | 20 | 19 |
| A | - | 10 | 15 | 14 | 1 | 0 | |
| Operator | + | 0 | 0 | 0 | 20 | 19 | |
| B | - | 10 | 15 | 15 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 20 | 19 | |
| C | - | 10 | 15 | 14 | 1 | 0 | |
| OPI(2000) | Operator | + | 0 | 0 | 0 | 18 | 22 |
| A | - | 10 | 16 | 14 | 0 | 0 | |
| Operator | + | 0 | 0 | 1 | 17 | 22 | |
| B | - | 10 | 16 | 13 | 1 | 0 | |
| Operator | + | 0 | 0 | 0 | 17 | 22 | |
| C | - | 10 | 16 | 14 | 1 | 0 | |
| MTD | Operator | + | 0 | 0 | 1 | 15 | 24 |
| A | - | 10 | 16 | 13 | 1 | 0 | |
| Operator | + | 0 | 0 | 0 | 14 | 24 | |
| B | - | 10 | 16 | 14 | 2 | 0 | |
| Operator | + | 0 | 0 | 0 | 15 | 24 | |
| C | - | 10 | 16 | 14 | 1 | 0 | |
| OXY | Operator | + | 0 | 0 | 1 | 13 | 25 |
| A | - | 10 | 15 | 14 | 2 | 0 | |
| Operator | + | 0 | 0 | 1 | 14 | 25 | |
| B | - | 10 | 15 | 14 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 13 | 25 | |
| C | - | 10 | 15 | 14 | 2 | 0 | |
| PCP | Operator | + | 0 | 0 | 1 | 13 | 25 |
| A | - | 10 | 16 | 13 | 2 | 0 | |
| Operator | + | 0 | 0 | 1 | 13 | 25 | |
| B | - | 10 | 16 | 13 | 2 | 0 | |
| Operator | + | 0 | 0 | 1 | 14 | 25 | |
| C | - | 10 | 16 | 13 | 1 | 0 | |
| PPX | Operator | + | 0 | 0 | 0 | 15 | 24 |
| A | - | 10 | 16 | 14 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 15 | 24 | |
| B | - | 10 | 16 | 13 | 1 | 0 | |
| Operator | + | 0 | 0 | 0 | 15 | 24 | |
| C | - | 10 | 16 | 14 | 1 | 0 | |
| TCA | Operator | + | 0 | 0 | 1 | 19 | 20 |
| A | - | 10 | 15 | 14 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 19 | 20 | |
| B | - | 10 | 15 | 14 | 1 | 0 | |
| Operator | + | 0 | 0 | 0 | 19 | 20 | |
| C | - | 10 | 15 | 15 | 1 | 0 | |
| THC | Operator | + | 0 | 0 | 1 | 16 | 22 |
| A | - | 10 | 16 | 13 | 2 | 0 | |
| Operator | + | 0 | 0 | 2 | 16 | 22 | |
| B | - | 10 | 16 | 12 | 2 | 0 | |
| Operator | + | 0 | 0 | 1 | 16 | 22 | |
| C | - | 10 | 16 | 13 | 2 | 0 | |
| TML | Operator | + | 0 | 0 | 1 | 14 | 24 |
| A | - | 10 | 16 | 13 | 2 | 0 | |
| Operator | + | 0 | 0 | 1 | 15 | 24 | |
| B | - | 10 | 16 | 13 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 15 | 24 | |
| C | - | 10 | 16 | 13 | 1 | 0 | |
| 6-MAM | Operator | + | 0 | 0 | 1 | 18 | 20 |
| A | - | 10 | 16 | 13 | 2 | 0 | |
| Operator | + | 0 | 0 | 1 | 18 | 20 | |
| B | - | 10 | 16 | 13 | 2 | 0 | |
| Operator | + | 0 | 0 | 2 | 19 | 20 | |
| C | - | 10 | 16 | 12 | 1 | 0 |
{27}------------------------------------------------
{28}------------------------------------------------
{29}------------------------------------------------
Discordant Results are summarized below.
| Drug | Operator | Sample Number | LC/MS/MS Result(ng/mL) | Accurate Result |
|---|---|---|---|---|
| 6-MAM 10 | Operator A, C | TF-6-MAM06071 | 8.13 | + |
| Operator C | TF-6-MAM06001 | 8.54 | + | |
| Operator B | TF-6-MAM06030 | 9.85 | + | |
| Operator B | TF-6-MAM06018 | 10.84 | - | |
| Operator A, B, C | TF-6-MAM06037 | 11.04 | - | |
| Operator A | TF-6-MAM06060 | 11.31 | - | |
| AMP 1000 | Operator B | TF-AMP06041 | 968.58 | + |
| Operator A, B | TF-AMP06035 | 1008.47 | - | |
| Operator C | TF-AMP06027 | 1038.84 | - | |
| BAR 300 | Operator A, C | TF-BAR06026 | 274.49 | + |
| Operator B | TF-BAR06024 | 291.27 | + | |
| Operator A, C | TF-BAR06037 | 316.44 | - | |
| BUP 10 | Operator A, C | TF-BUP06058 | 9.22 | + |
| Operator B, C | TF-BUP06065 | 9.58 | + | |
| Operator A | TF-BUP06032 | 9.79 | + | |
| Operator A | TF-BUP06071 | 10.18 | - | |
| Operator B, C | TF-BUP06002 | 10.33 | - | |
| Operator B | TF-BUP06059 | 10.77 | - | |
| Operator C | TF-BUP06039 | 11.02 | - | |
| Operator A | TF-BUP06057 | 11.22 | - | |
| BZO 300 | Operator A, C | TF-BZO06049 | 276.48 | + |
| Operator B | TF-BZO06004 | 289.45 | + | |
| Operator A, B | TF-BZO06014 | 302.46 | - | |
| Operator A, C | TF-BZO06018 | 318.53 | - | |
| Operator B | TF-BZO06006 | 322.77 | - | |
| Operator C | TF-BZO06071 | 336.40 | - | |
| COC 300 | Operator A | TF-COC06076 | 294.37 | + |
| Operator B | TF-COC06061 | 296.67 | + | |
| Operator B | TF-COC06010 | 288.91 | + | |
| Operator C | TF-COC06066 | 291.80 | + | |
| Operator A | TF-COC06036 | 331.25 | - | |
| Operator B | TF-COC06002 | 302.88 | - | |
| Operator C | TF-COC06029 | 311.54 | - | |
| EDDP 300 | Operator B | TF-EDDP06008 | 278.15 | + |
| Operator A | TF-EDDP06065 | 294.41 | + | |
| Operator B | TF-EDDP06042 | 308.29 | - | |
| Operator B | TF-EDDP06040 | 314.11 | - | |
| Operator A | TF-EDDP06025 | 326.51 | - | |
| FYL | Operator A | TF-FYL06017 | 4.28 | + |
| Operator C | TF-FYL06062 | 4.54 | + | |
| Operator B | TF-FYL06038 | 4.61 | + | |
| Operator A | TF-FYL06078 | 4.83 | + | |
| Operator B | TF-FYL06049 | 5.11 | - | |
| Operator A | TF-FYL06063 | 5.48 | - | |
| Operator C | TF-FYL06073 | 5.63 | - | |
| Operator B | TF-FYL06022 | 5.77 | - | |
| MDMA 500 | Operator A | TF-MDMA06047 | 434.68 | + |
| Operator C | TF-MDMA06080 | 443.57 | + | |
| Operator B | TF-MDMA06072 | 483.40 | + | |
| Operator A | TF-MDMA06046 | 491.32 | + | |
| Operator C | TF-MDMA06006 | 505.64 | - | |
| Operator A, B | TF-MDMA06002 | 519.63 | - | |
| MET 1000 | Operator B | TF-MET06052 | 946.52 | + |
| Operator B | TF-MET06031 | 1064.20 | - | |
| Operator A | TF-MET06007 | 1139.47 | - | |
| Operator C | TF-MET06006 | 1162.76 | - | |
| OPI 300 | Operator A, C | TF-MOP06065 | 291.05 | + |
| Operator B | TF-MOP06062 | 302.36 | - | |
| Operator A | TF-MOP06031 | 317.77 | - | |
| MTD 300 | Operator C | TF-MOP06032 | 351.52 | - |
| MTD 300 | Operator A | TF-MTD06025 | 280.13 | + |
| Operator B | TF-MTD06078 | 306.14 | - | |
| Operator C | TF-MTD06056 | 321.60 | - | |
| Operator A | TF-MTD06006 | 336.27 | - | |
| Operator B | TF-MTD06060 | 351.49 | - | |
| OXY 100 | Operator A, C | TF-OXY06058 | 95.88 | + |
| Operator B | TF-OXY06003 | 97.96 | + | |
| Operator A | TF-OXY06078 | 100.28 | - | |
| Operator A, C | TF-OXY06073 | 100.77 | - | |
| Operator B | TF-OXY06024 | 102.28 | - | |
| Operator C | TF-OXY06060 | 115.34 | - | |
| PCP 25 | Operator A, C | TF-PCP06039 | 20.19 | + |
| Operator B | TF-PCP06002 | 23.68 | + | |
| Operator A, C | TF-PCP06018 | 25.19 | - | |
| Operator A, B | TF-PCP06075 | 25.88 | - | |
| Operator B | TF-PCP06043 | 28.42 | - | |
| PPX 300 | Operator B | TF-PPX06066 | 277.71 | + |
| Operator B | TF-PPX06037 | 304.35 | - | |
| Operator A | TF-PPX06012 | 321.47 | - | |
| Operator C | TF-PPX06045 | 335.28 | - | |
| TCA 1000 | Operator B | TF-TCA06015 | 892.55 | + |
| Operator A | TF-TCA06008 | 951.60 | + | |
| Operator B | TF-TCA06020 | 1013.28 | - | |
| Operator C | TF-TCA06068 | 1120.01 | - | |
| Operator A | TF-TCA06080 | 1120.63 | - | |
| THC 50 | Operator B | TF-THC06021 | 42.33 | + |
| Operator A | TF-THC06073 | 44.93 | + | |
| Operator C | TF-THC06060 | 47.61 | + | |
| Operator B | TF-THC06050 | 49.77 | + | |
| Operator A, C | TF-THC06062 | 53.28 | - | |
| Operator B | TF-THC06014 | 57.74 | - | |
| Operator B | TF-THC06040 | 58.69 | - | |
| Operator A, C | TF-THC06064 | 59.33 | - | |
| TML | Operator A | TF-TML06052 | 84.31 | + |
| Operator C | TF-TML06043 | 90.15 | + | |
| Operator B | TF-TML06024 | 94.68 | + | |
| Operator A | TF-TML06054 | 101.37 | - | |
| Operator B, C | TF-TML06056 | 108.45 | - | |
| Operator A | TF-TML06046 | 114.31 | - | |
| AMP 500 | Operator B | TF-SAMP06031 | 447.35 | + |
| Operator C | TF-SAMP06034 | 465.44 | + | |
| Operator A | TF-SAMP06006 | 479.87 | + | |
| Operator B | TF-SAMP06055 | 502.90 | - | |
| Operator A | TF-SAMP06075 | 541.26 | - | |
| Operator C | TF-SAMP06054 | 577.15 | - | |
| COC 150 | Operator A, B, C | TF-SCOC06030 | 133.67 | + |
| Operator B | TF-SCOC06062 | 155.27 | - | |
| Operator A, C | TF-SCOC06053 | 160.33 | - | |
| Operator B | TF-SCOC06019 | 165.57 | - | |
| MET 500 | Operator B | TF-SMET06029 | 428.49 | + |
| Operator A, C | TF-SMET06042 | 481.16 | + | |
| Operator B, C | TF-SMET06035 | 512.71 | - | |
| OPI 2000 | Operator B | TF-SMOP06029 | 1972.07 | + |
| Operator B | TF-SMOP06014 | 2108.15 | - | |
| Operator C | TF-SMOP06065 | 2194.62 | - |
{30}------------------------------------------------
{31}------------------------------------------------
{32}------------------------------------------------
C. Lay person study
A lay user study was performed at three intended user sites with 280 lay persons. 76 male and 64 female tested Healgen® Accurate Muti-Drug Urine Drug Screen Cup Configuration 1; 73 male and 67 female tested Healgen® Accurate Muti-Drug Urine Drug Screen Cup Configuration 2. They had diverse educational and professional backgrounds and their age range from 21 to > 50. Urine samples were prepared at the following concentrations; -100%, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC-MS/MS. Each sample was aliquoted into individual containers and blindlabeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below.
| Drug | Concentration | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Cutoff(ng/mL) | Results | -100%cutoff | -75%cutoff | -50%cutoff | -25%cutoff | +25%cutoff | +50%cutoff | +75%cutoff | |
| AMP | 1000 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% | ||
| BAR | 300 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 95% | 100% | 100% | ||
| BUP | 10 | Negative | 20 | 20 | 20 | 18 | 2 | 0 | 0 |
| Positive | 0 | 0 | 0 | 2 | 18 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 90% | 100% | 100% | ||
| BZO | 300 | Negative | 20 | 20 | 20 | 18 | 2 | 0 | 0 |
| Positive | 0 | 0 | 0 | 2 | 18 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 90% | 100% | 100% | ||
| Negative | 20 | 20 | 20 | 19 | 2 | 0 | 0 | ||
| Positive | 0 | 0 | 0 | 1 | 18 | 20 | 20 | ||
| COC | 300 | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | 100% | ||
| Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 | ||
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | ||
| EDDP | 300 | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% | ||
| Negative | 20 | 20 | 20 | 18 | 2 | 0 | 0 | ||
| Positive | 0 | 0 | 0 | 2 | 18 | 20 | 20 | ||
| FYL | 5 | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
| Agreement (%) | 100% | 100% | 100% | 90% | 90% | 100% | 100% | ||
| Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 | ||
| 500 | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | |
| MDMA | Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |
| Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% | ||
| 1000 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 | |
| MET | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | |
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% | ||
| 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 | |
| MOP | Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | |
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% | ||
| 300 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 | |
| MTD | Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 | |
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 95% | 100% | 100% | ||
| Negative | 20 | 20 | 20 | 18 | 2 | 0 | 0 | ||
| OXY | 100 | Positive | 0 | 0 | 0 | 2 | 18 | 20 | 20 |
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 90% | 100% | 100% | ||
| Negative | 20 | 20 | 20 | 19 | 2 | 0 | 0 | ||
| PCP | 25 | Positive | 0 | 0 | 0 | 1 | 18 | 20 | 20 |
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | 100% | ||
| Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 | ||
| PPX | 300 | Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 |
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 95% | 100% | 100% | ||
| TCA | 1000 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 100% | 100% | 100% | ||
| THC | 50 | Negative | 20 | 20 | 20 | 18 | 2 | 0 | 0 |
| Positive | 0 | 0 | 0 | 2 | 18 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 90% | 100% | 100% | ||
| TML | 100 | Negative | 20 | 20 | 20 | 19 | 2 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 18 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | 100% | ||
| 6-MAM | 10 | Negative | 20 | 20 | 20 | 18 | 2 | 0 | 0 |
| Positive | 0 | 0 | 0 | 2 | 18 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90% | 90% | 100% | 100% |
Result of Healgen® Accurate Muti-Drug Urine Drug Screen Cup Configuration 1:
{33}------------------------------------------------
{34}------------------------------------------------
Result of Healgen® Accurate Muti-Drug Urine Drug Screen Cup Configuration 2:
| Drug | Cutoff(ng/mL) | Results | -100%cutoff | -75%cutoff | -50%cutoff | -25%cutoff | +25%cutoff | +50%cutoff | +75%cutoff | |
|---|---|---|---|---|---|---|---|---|---|---|
| AMP | 500 | Negative | 20 | 20 | 20 | 19 | 2 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 1 | 18 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | 100% | |||
| BAR | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 95% | 95% | 100% | 100% | |||
| BUP | 10 | Negative | 20 | 20 | 20 | 18 | 2 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 2 | 18 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 90% | 90% | 100% | 100% | |||
| BZO | 300 | Negative | 20 | 20 | 20 | 19 | 2 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 1 | 18 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | 100% | |||
| COC | 150 | Negative | 20 | 20 | 20 | 19 | 2 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 1 | 18 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | 100% | |||
| EDDP | 300 | Negative | 20 | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 2 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100% | 90% | 95% | 100% | 100% | ||
| FYL | 5 | Negative | 20 | 20 | 20 | 20 | 19 | 2 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 1 | 18 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100% | 95% | 90% | 100% | 100% | ||
| MDMA | 500 | Negative | 20 | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 1 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100% | 95% | 95% | 100% | 100% | ||
| MET | 500 | Negative | 20 | 20 | 20 | 20 | 20 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 0 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100% | 100% | 95% | 100% | 100% | ||
| MOP | 2000 | Negative | 20 | 20 | 20 | 20 | 19 | 0 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 1 | 20 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100% | 95% | 100% | 100% | 100% | ||
| MTD | 300 | Negative | 20 | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 1 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100% | 95% | 95% | 100% | 100% | ||
| OXY | 100 | Negative | 20 | 20 | 20 | 20 | 19 | 2 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 1 | 18 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100% | 95% | 90% | 100% | 100% | ||
| PCP | 25 | Negative | 20 | 20 | 20 | 20 | 18 | 2 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 2 | 18 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100% | 90% | 90% | 100% | 100% | ||
| PPX | 300 | Negative | 20 | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 1 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100% | 95% | 95% | 100% | 100% | ||
| TCA | 1000 | Negative | 20 | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 1 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100% | 95% | 95% | 100% | 100% | ||
| THC | 50 | Negative | 20 | 20 | 20 | 20 | 19 | 2 | 0 | 0 |
| Positive | 0 | 0 | 0 | 0 | 1 | 18 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 100% | 95% | 90% | 100% | 100% | ||
| Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | 100% | |||
| TML | 100 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 95% | 90% | 100% | 100% | |||
| 6-MAM | 10 | Negative | 20 | 20 | 20 | 18 | 2 | 0 | 0 | |
| Positive | 0 | 0 | 0 | 2 | 18 | 20 | 20 | |||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | |||
| Agreement (%) | 100% | 100% | 100% | 90% | 90% | 100% | 100% |
{35}------------------------------------------------
{36}------------------------------------------------
Participants were given surveys on the ease of understanding the instruction for use. All participants indicated that the device instruction is easy to understand and follow. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.
Clinical Studies: Not applicable.
13. Conclusion
Based on the test principle and performance characteristics of the device including precision, cut-off, interference, specificity, method comparison and lay-user studies of the devices, it's concluded that Healgen® Accurate Home Muti-Drug Urine Test Cup and Healgen® Accurate Muti-Drug Urine Drug Screen Cup are substantially equivalent to the predicate devices.
§ 862.3650 Opiate test system.
(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).