K Number
K243287
Manufacturer
Date Cleared
2025-02-24

(129 days)

Product Code
Regulation Number
870.1250
Panel
NE
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The APRO 70 Swift Catheter with an aspiration pump and the Alembic Aspiration Tubing is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for thrombolytic drug therapy or who failed thrombolytic drug therapy are candidates for treatment.

The Alembic Aspiration Tubing is intended to connect the APRO 70 Swift Catheter to the aspiration pump.

Device Description

The APRO 70 Swift Catheter is a single-lumen, braid and coil reinforced catheter. The APRO 70 Swift Catheter is designed to remove thrombus from the vasculature using aspiration. The APRO 70 Swift Catheter targets aspiration from a suction pump directly to the thrombus to remove thrombus from an occluded vessel. The APRO 70 Swift Catheter is introduced through a guide catheter or long femoral sheath and into the intracranial vasculature and guided over a neurovascular guidewire under fluoroscopic visualization to the site of the primary occlusion. The distal shaft has a hydrophilic coating to aid navigation through the vasculature. A radiopaque marker is located at the distal end of the catheter for visualization under fluoroscopy. For the aspiration source, the APRO 70 Swift Catheter is used in conjunction with an aspiration pump with pre-specified performance parameters that is connected using the Alembic Aspiration Tubing, along with a legally marketed canister and accessories kit. The APRO 70 Swift Catheter is available in lengths of 125 cm. 132 cm, and 135 cm and is provided with an introducer sheath.

The Alembic Aspiration Tubing connects the APRO 70 Swift Catheter to the aspiration pump. The flow control valve allows control of the aspiration flow using an ON/OFF switch. It is available in one size.

AI/ML Overview

The provided text describes the acceptance criteria and the study results for the APRO 70 Swift Catheter and Alembic Aspiration Tubing. It is important to note that this document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a legally marketed predicate device through non-clinical performance data, rather than a clinical trial report. Therefore, some of the requested information, particularly regarding clinical studies, human readers, and effect sizes, is not applicable or available in this document.

Here's the breakdown of the information based on your request:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria and performance data are primarily from non-clinical bench testing.

Test CategoryAcceptance CriteriaReported Device Performance
Visual and Dimensional CharacteristicsCatheter meets visual and dimensional specifications.
Introducer Sheath meets visual and dimensional specifications.The APRO 70 Swift Catheter met the acceptance criteria.
The Introducer Sheath met the acceptance criteria.
ParticulateCatheter meets the acceptance criteria.
Subject device was evaluated with a predicate device under the same test conditions.The APRO 70 Swift Catheter particulates were comparable to the predicate device.
Vacuum IntegrityCatheter is free from collapse and loss of vacuum between aspiration source and catheter tip.The APRO 70 Swift Catheter met the acceptance criteria.
Kink ResistanceCatheter shaft shall not kink at clinically relevant radii.The APRO 70 Swift Catheter met the acceptance criteria.
Catheter Hub LeakageCatheter does not leak into hub assembly during aspiration, with methods specified in ISO 10555-1, Annex D.The APRO 70 Swift Catheter met the acceptance criteria.
Hub CompatibilityHub meets the requirements per EN ISO 80369-7.The APRO 70 Swift Catheter met the acceptance criteria.
Catheter Torque StrengthCatheter must withstand the minimum required number of rotations without breakage and without kinking compared to legally marketed devices.The APRO 70 Swift Catheter met the acceptance criteria.
Dynamic Burst PressureNo damage to catheter with dynamic pressure.The APRO 70 Swift Catheter met the acceptance criteria.
Fluid LeakageCatheter must withstand pressure with methods per ISO 10555-1 Annex C.The APRO 70 Swift Catheter met the acceptance criteria.
Static BurstCatheter must withstand pressures anticipated for clinical use.The APRO 70 Swift Catheter met the acceptance criteria.
Tensile Strength of Catheter Hub and ShaftCatheter hub and shaft must meet tensile strength specification.The APRO 70 Swift Catheter met the acceptance criteria.
Tensile Strength of Catheter TipCatheter tip must meet tip tensile strength specification.The APRO 70 Swift Catheter met the acceptance criteria.
Delivery and Retrieval ForceCatheter delivery and retrieval force must be acceptable. Forces were compared to a predicate.The APRO 70 Swift Catheter met the acceptance criteria.
Tip Buckling ForceCatheter tip buckling force must be acceptable. Forces were compared to a predicate.The APRO 70 Swift Catheter met the acceptance criteria.
Simulated Use including Clot RetrievalWhen used per the Instructions for Use with accessory devices in an anatomical neurovascular model, the Catheter and Aspiration Tubing must meet functionality specifications. Clot retrieval from various locations of the anatomical model was performed using the subject device system.The APRO 70 Swift Catheter and Alembic Aspiration Tubing met the acceptance criteria.
Biocompatibility (Sensitization)Did not elicit a sensitization response.Non-sensitizing
Biocompatibility (Irritation)Demonstrated no evidence of irritation.Non-irritant
Biocompatibility (Cytotoxicity)Did not elicit a cytotoxic response at 24 hours and 48 hours.Non-cytotoxic
Biocompatibility (Hemolysis – Indirect)No significant differences between the test article extract and negative control article results.Non-hemolytic
Biocompatibility (Hemolysis - Direct)No differences between the hemolytic index of the test article and the negative control.Non-hemolytic
Biocompatibility (Thrombogenicity - PTT)Average clotting time of the test article was greater than vehicle control and negative control.Acceptable clotting times
Biocompatibility (Thrombogenicity - Platelet Leukocyte Count)Performed similar to comparator and negative controls.Non-thrombogenic
Biocompatibility (Thrombogenicity - Comparative Surface and Geometry Assessment)Comparative surface assessment using 40X optical microscopy at representative locations of the subject device and the reference device APRO 70 Catheter did not find any differences in roughness or presence of any defects. The subject device has the same geometry as the reference device APRO 70 Catheter.Acceptable results
Biocompatibility (SC5b9 Complement Activation)Sc5b9 concentration of the test article was statistically less than the positive control and was not statistically higher than the negative control.Acceptable
Biocompatibility (Acute Systemic Toxicity)No weight loss, mortality, or evidence of systemic toxicity from the extract exposure to the mice.Non-toxic
Biocompatibility (Material-Mediated Pyrogenicity)All individual rabbits for both the test article and negative control showed a total rise in temperature of

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).