K Number
K243259
Manufacturer
Date Cleared
2025-01-07

(84 days)

Product Code
Regulation Number
888.3070
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Pedicle Screw System: The MASADA Modular Spinal Fixation System is intended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease ( defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), tumor, pseudoarthrosis, and/or failed previous fusion.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the MASADA pedicle screw implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The MASADA pedicle screw is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Navigated Instrument System: The MASADA navigated instruments are intended to be used in the preparation and placement of MASADA screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medtronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. Refer to the Astura Navigated Instrument system Instructions For Use (INS-00006) regarding the use of these instruments.

Sublaminar Band System: The MASADA sublaminar band is a temporary implant for use in orthopedic surgery. The band system is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures. The indications for use include the following applications: - Spinal trauma surgery, used in sublaminar or facet wiring techniques. - Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as idiopathic and neuromuscular scoliosis in patients 8 years of age or older, adult scoliosis, kyphosis and spondylolisthesis. - Spinal degenerative surgery, as an adjunct to spinal fusions. The MASADA sublaminar band may also be used in conjunction with other medical grade implants made of similar metals whenever "wiring" may help secure the attachment of the other implants.

Device Description

The MASADA Modular Spinal Fixation System is intended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion. The system is composed of modular polyaxial screws, monoaxial screws, rods, cross connectors, rod connectors, sublaminar band cerclages, band connectors, and hooks. The system is supported by a comprehensive set of instruments to install the implants within the system. All implant components are manufactured from the materials listed in the table below.

AI/ML Overview

The provided text is an FDA 510(k) clearance letter for a medical device called the "MASADA Modular Spinal Fixation System." This document is a regulatory approval, not a scientific study report describing the performance of an AI/ML device. Therefore, it does not contain the information requested regarding acceptance criteria, device performance metrics, sample sizes, ground truth establishment, expert qualifications, or MRMC studies for an AI/ML system.

The "Non-clinical Test Summary" explicitly states: "An engineering analysis determined that no new worst case conditions have been introduced since the previously cleared submission (K231694)." and the "Clinical Test Summary" explicitly states: "No clinical studies were performed."

This means the clearance was based on substantial equivalence to predicate devices through design, materials, and engineering analysis, rather than new performance data from a clinical or AI/ML-specific study.

Therefore, I cannot provide the requested table and details about acceptance criteria and study findings for an AI/ML device based on the given input.

{0}------------------------------------------------

January 7, 2025

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Astura Medical Parker Kelch Quality and Regulatory Manager 4949 W Royal Ln Irving, Texas 75063

Re: K243259

Trade/Device Name: MASADA Modular Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, KWP, OWI, OLO Dated: October 11, 2024 Received: October 15, 2024

Dear Parker Kelch:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Image /page/2/Picture/3 description: The image contains the text "Colin O'neill -S FDA". The text is arranged in two lines, with "Colin" on the first line and "O'neill -S" on the second line. The letters "FDA" are in a larger font and a lighter color, and they are positioned to the right of the other text.

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K243259

Device Name MASADA Modular Spinal Fixation System

Indications for Use (Describe)

Pedicle Screw System: The MASADA Modular Spinal Fixation System is intended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease ( defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), tumor, pseudoarthrosis, and/or failed previous fusion.

When used for posterior non-cervical pediale screw fixation in pediatric patients, the MASADA pedicle screw implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The MASADA pedicle screw is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Navigated Instrument System: The MASADA navigated instruments are intended to be used in the preparation and placement of MASADA screws during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. These instruments are designed for use with the Medronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. Refer to the Astura Navigated Instrument system Instructions For Use (INS-00006) regarding the use of these instruments.

Sublaminar Band System: The MASADA sublaminar band is a temporary implant for use in orthopedic surgery. The band system is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures. The indications for use include the following applications: - Spinal trauma surgery, used in sublaminar or facet wiring techniques. - Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as idiopathic and neuromuscular scoliosis in patients 8 years of age or older, adult scoliosis, kyphosis and spondylolisthesis. - Spinal degenerative surgery, as an adjunct to spinal fusions. The MASADA sublaminar band may also be used in conjunction with other medical grade implants made of similar metals whenever "wiring" may help secure the attachment of the other implants.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

510(k) Summary: MASADA Modular Spinal Fixation System

Date PreparedJanuary 6, 2025
Submitted ByAstura Medical4949 W Royal Ln.Irving, TX 75063Phone: 469-501-5530
ContactParker Kelch4949 W Royal Ln.Irving, TX 75063Phone: 469-501-5530 x503Email: parker@asturamedical.com
Trade NameMASADA Modular Spinal Fixation System
Common NamePosterior spinal fixation system
Classification NameThoracolumbosacral pedicle screw systemSpinal interlaminal fixation orthosisOrthopedic stereotaxic instrument
ClassII
Product CodeNKB, KWP, OWI, OLO
CFR Section21 CFR section 888.3070, 888.3050, 888.3010, 882.4560
Device PanelOrthopedic
Primary Predicate DeviceMASADA Modular Spinal Fixation System (K231694)
Additional PredicateDevice(s)OLYMPIC MIS (K182239)OLYMPIC Navigated Instruments (K172166)Globus CREO Stabilization System (K124058)
Device DescriptionThe MASADA Modular Spinal Fixation System is intended forimmobilization and stabilization of the posterior, non-cervical spine inskeletally mature patients as an adjunct to fusion.The system is composed of modular polyaxial screws, monoaxial screws,rods, cross connectors, rod connectors, sublaminar band cerclages, bandconnectors, and hooks. The system is supported by a comprehensive setof instruments to install the implants within the system. All implantcomponents are manufactured from the materials listed in the tablebelow.
MaterialsTi-6Al-4V ELI (ASTM F136)CP Titanium Grade 4 (ASTM F67)CoCrMo alloy (ASTM F1537)Elgiloy CoCrNi alloy (ASTM F1058)Nitinol #1 (ASTM F2063)PET (ISO 10993)
Substantial EquivalenceClaimed to PredicateDevicesThe MASADA system is substantially equivalent to the predicate devicesin terms of intended use, design, materials used, mechanical safety andperformances.
Indications for UsePedicle Screw System: The MASADA Modular Spinal Fixation System isintended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion for thefollowing indications: degenerative disc disease ( defined as back pain ofdiscogenic origin with degeneration of the disc confirmed by history andradiographic studies), spondylolisthesis, trauma (i.e., fracture ordislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, orlordosis), tumor, pseudoarthrosis, and/or failed previous fusion.
When used for posterior non-cervical pedicle screw fixation in pediatricpatients, the MASADA pedicle screw implants are indicated as an adjunctto fusion to treat adolescent idiopathic scoliosis. The MASADA pediclescrew is intended to be used with autograft and/or allograft. Pediatricpedicle screw fixation is limited to a posterior approach.
Navigated Instrument System: The MASADA navigated instruments areintended to be used in the preparation and placement of MASADA screwsduring spinal surgery to assist the surgeon in precisely locatinganatomical structures in either open or minimally invasive procedures.These instruments are designed for use with the MedtronicStealthStation® System, which is indicated for any medical condition inwhich the use of stereotactic surgery may be appropriate, and wherereference to a rigid anatomical structure, such as a skull, a long bone, orvertebra, can be identified relative to a CT or MR based model,fluoroscopy images, or digitized landmarks of the anatomy. Refer to theAstura Navigated Instrument system Instructions For Use (INS-00006)regarding the use of these instruments.
Sublaminar Band System: The MASADA sublaminar band is a temporaryimplant for use in orthopedic surgery. The band system is intended toprovide temporary stabilization as a bone anchor during the developmentof solid bony fusion and aid in the repair of bone fractures. Theindications for use include the following applications: - Spinal traumasurgery, used in sublaminar or facet wiring techniques. - Spinalreconstructive surgery, incorporated into constructs for the purpose ofcorrection of spinal deformities such as idiopathic and neuromuscularscoliosis in patients 8 years of age or older, adult scoliosis, kyphosis andspondylolisthesis. - Spinal degenerative surgery, as an adjunct to spinalfusions. The MASADA sublaminar band may also be used in conjunctionwith other medical grade implants made of similar metals whenever"wiring" may help secure the attachment of the other implants.
Non-clinical TestSummaryAn engineering analysis determined that no new worst case conditionshave been introduced since the previously cleared submission (K231694).
Clinical Test SummaryNo clinical studies were performed.
Conclusions: Non-Clinicaland ClinicalAstura Medical considers MASADA Modular Spinal Fixation System to beequivalent to the predicate devices listed above. This conclusion is based

In accordance with 21 CFR 807.92 of the Federal Code of Regulations

{6}------------------------------------------------

{7}------------------------------------------------

upon the devices' similarities in principles of operation, technology, materials and indications for use.

{8}------------------------------------------------

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.