(68 days)
The ADI Sensinel CPM (Cardiopulmonary Management) System is a wireless remote monitoring system intended for use by healthcare professionals for spot checking of physiological data in home and healthcare settings. This can include:
• ECG
- · Heart Auscultation Sounds
- · Skin Temperature
- · Thoracic Impedance (including Changes in Thoracic Impedance)
- · Respiration Rate and relative changes in Tidal Volume
· Heart Rate - Diastolic Heart Sounds Strength
- · Body Posture (including Tilt Angle)
Data are transmitted wirelessly from the ADI CPM (Cardiopulmonary Management) Wearable and Base Station for storage and analysis. The device is intended for use on general care adult patients who are 18 years of age or older to provide physiological information. The data from the ADI CPM (Cardiopulmonary Management) System Platform are intended for use by healthcare professional as an aid to diagnosis and treatment.
The ADI CPM System is intended to be used by patients at rest and not performing any activities or movements. This system is for spot checking and does not have continuous monitoring capability. The device does not produce alarms and is not intended for active patient monitoring (real-time).
The ADI CPM System in contraindicated for those patients with life threatening arrhythmias requiring immediate medical intervention.
Not Found
The provided FDA 510(k) clearance letter and Indications for Use document for the Sensinel Cardiopulmonary Management (CPM) System (ADCP1100) do not contain the detailed information required to describe the acceptance criteria and the study that proves the device meets those criteria.
This document primarily states that the device has been found substantially equivalent to legally marketed predicate devices and outlines the regulatory requirements and indications for use. It does not include performance data, study designs, sample sizes, ground truth methodologies, or expert qualifications.
To answer your request, the following information would typically be found in the 510(k) submission's technical documentation (e.g., performance testing details, clinical study reports), which is not included in the provided text.
Therefore, I cannot extract the information required for the detailed answer as it is not present in the given text.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
December 10, 2024
Analog Devices, Inc. Phani Puppala Senior Strategist. Ouality & Regulatory 1 Analog Wav Wilmington, Massachusetts 01887
Re: K243216
Trade/Device Name: Sensinel Cardiopulmonary Management (CPM) System (ADCP1100) Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, DPS, DXH, DSB, DQD, BZQ, FLL Dated: October 2, 2024 Received: October 3, 2024
Dear Phani Puppala:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Jennifer W. Shih -S
Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{3}------------------------------------------------
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
Device Name
Sensinel Cardiopulmonary Management (CPM) System (ADCP1100)
Indications for Use (Describe)
The ADI Sensinel CPM (Cardiopulmonary Management) System is a wireless remote monitoring system intended for use by healthcare professionals for spot checking of physiological data in home and healthcare settings. This can include:
• ECG
- · Heart Auscultation Sounds
- · Skin Temperature
- · Thoracic Impedance (including Changes in Thoracic Impedance)
- · Respiration Rate and relative changes in Tidal Volume
· Heart Rate
- Diastolic Heart Sounds Strength
- · Body Posture (including Tilt Angle)
Data are transmitted wirelessly from the ADI CPM (Cardiopulmonary Management) Wearable and Base Station for storage and analysis. The device is intended for use on general care adult patients who are 18 years of age or older to provide physiological information. The data from the ADI CPM (Cardiopulmonary Management) System Platform are intended for use by healthcare professional as an aid to diagnosis and treatment.
The ADI CPM System is intended to be used by patients at rest and not performing any activities or movements. This system is for spot checking and does not have continuous monitoring capability. The device does not produce alarms and is not intended for active patient monitoring (real-time).
The ADI CPM System in contraindicated for those patients with life threatening arrhythmias requiring immediate medical intervention.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
he-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).
(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).