(221 days)
Water Solubility Human Body Lubricant is a personal lubricant for penile, anal and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.
Water Solubility Human Body Lubricant is a clear, colorless, odorless and semi-viscous personal lubricant that is compatible with condoms made of natural rubber latex and polyisoprene, and is not compatible with polyurethane condoms. This device is a non-sterile personal lubricant for penile, anal, and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is sold as an over-the-counter (OTC) product in 120 mL size provided in clear polyethylene terephthalate (PET) bottle. The bottles are capped with silver disc tops. The individual bottles are hermetically sealed during the production process. The device is composed of Hyaluronic acid, Lubrajel CG, Lubrajel oil, Panthenol, Trimethyl glycine, Pentanediol, Hydroxyacetophenone, Ethylene glycol, Glycerol, Purified water.
This FDA clearance letter is for a Water Solubility Human Body Lubricant, which is regulated as a condom accessory (personal lubricant). The information provided is standard for a 510(k) submission for such a device and focuses on biocompatibility, shelf-life, and condom compatibility.
Therefore, the concept of "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of an AI/algorithm-based medical device (like those requiring MRMC studies, ground truth establishment by experts, training/test sets) does not apply to this product.
The document discusses performance testing, but these tests are for physical, chemical, and biological properties of the lubricant itself, not for an AI algorithm's diagnostic or assistive performance.
Here's why the AI-related questions cannot be answered based on the provided text:
- No AI/Algorithm: The device is a "Water Solubility Human Body Lubricant," a physical product, not an AI software or algorithm.
- No Diagnostic/Clinical Performance: The device's function is lubrication, not diagnosis, image analysis, or any task typically performed by an AI in healthcare.
- No Human Readers/Experts for AI Ground Truth: Since there's no AI, there's no need for human readers to establish ground truth or conduct MRMC studies.
Instead, I can summarize the performance criteria and the studies conducted for this specific medical device (the lubricant):
Device Category: Personal Lubricant (regulated under Condom, Product Code NUC)
The acceptance criteria and studies for the Water Solubility Human Body Lubricant are focused on its physical, chemical, microbiological properties, biocompatibility, shelf-life, and compatibility with condoms, as is standard for this type of device.
1. Table of Acceptance Criteria and Reported Device Performance
| Property | Acceptance Criteria (Specification) | Reported Device Performance (as tested) |
|---|---|---|
| Physical/Chemical Properties | ||
| Appearance | Semi-viscous liquid | Semi-viscous liquid |
| Color | Clear | Clear |
| Odor | Odorless | Odorless |
| Viscosity (per USP<911>) | 3500 – 4500 cps | Within range (implied by meeting specification) |
| pH (per USP<791>) | 5.6 – 6.4 | Within range (implied by meeting specification) |
| Specific gravity (per USP<831>) | 0.950 – 1.100 | Within range (implied by meeting specification) |
| Osmolality (per USP<785>) | 100 – 150 mOsm/kg (1:5 dilution) | Within range (implied by meeting specification) |
| Microbiological Properties | ||
| Total Aerobic Microbial Count (TAMC, USP<61>) | <100 cfu/g | <100 cfu/g (meets specification) |
| Total Yeast and Mold Count (TYMC, USP<61>) | <10 cfu/g | <10 cfu/g (meets specification) |
| Pseudomonas aeruginosa (USP<62>) | Absent | Absent (meets specification) |
| Staphylococcus aureus (USP<62>) | Absent | Absent (meets specification) |
| Candida albicans (USP<62>) | Absent | Absent (meets specification) |
| Escherichia coli (USP<62>) | Absent | Absent (meets specification) |
| Antimicrobial Effectiveness (USP<51>) | Meets USP<51> criteria for category 2 | Meets USP<51> criteria for category 2 (per categories) |
| - Bacteria: _>_2.0 log reduction at 14 days, no increase from 14-day count at 28 days | (Implied by "Meets USP <51> criteria") | |
| - Yeast/Molds: No increase from initial at 14 & 28 days | (Implied by "Meets USP <51> criteria") | |
| Biocompatibility | ||
| Cytotoxicity (ISO 10993-5:2009) | Non-cytotoxic | Non-cytotoxic |
| Sensitization (ISO 10993-10:2021) | Non-sensitizing | Non-sensitizing |
| Vaginal Irritation (ISO 10993-23:2021) | Non-irritating | Non-irritating |
| Acute Systemic Toxicity (ISO 10993-11:2017) | Not acutely, systemically toxic | Not acutely, systemically toxic |
| Shelf-Life | ||
| Stability for 3 years | Maintains specifications in Table 1 | Maintains specifications (demonstrated by accelerated aging) |
| Condom Compatibility | ||
| Natural Rubber Latex Condoms | Compatible | Compatible |
| Polyisoprene Condoms | Compatible | Compatible |
| Polyurethane Condoms | Not compatible | Not compatible |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify the exact sample sizes used for each physical, chemical, or microbiological test (e.g., how many units were tested for viscosity or pH, or how many animals for biocompatibility). These are typically standard laboratory sample sizes for quality control and testing.
- Data Provenance: The tests were conducted according to recognized international and US standards (USP, ISO, ASTM). The manufacturer is "Dongguan Yanxuan Biotechnology Co., Ltd." in China, and the studies were performed as part of their 510(k) submission to the US FDA. The studies are prospective tests performed on the manufactured product to demonstrate compliance.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
N/A - This device is a physical product (lubricant), not an AI/algorithmic device that requires expert-established ground truth for performance evaluation. The "ground truth" here is the result of the standardized laboratory tests meeting predefined physical, chemical, and biological specifications.
4. Adjudication Method for the Test Set
N/A - Not applicable for this type of product testing. Results are quantitative measurements or qualitative observations based on direct testing according to established standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
N/A - MRMC studies are for evaluating the performance of diagnostic devices, especially those involving human interpretation (e.g., radiologists reading images) sometimes with AI assistance. This device is a lubricant.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
N/A - There is no algorithm for this device.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance is established by objective laboratory measurements and tests against pre-defined specifications from recognized standards (e.g., USP monographs, ISO biocompatibility standards, ASTM condom compatibility standards).
8. The Sample Size for the Training Set
N/A - Not applicable. There is no AI model to train.
9. How the Ground Truth for the Training Set Was Established
N/A - Not applicable. There is no AI model to train.
U.S. Food & Drug Administration FDA Clearance Letter
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.07.05
May 9, 2025
Dongguan Yanxuan Biotechnology Co., Ltd.
℅ Joyce Yang
Consultant
Shenzhen Joyantech Consulting Co., Ltd.
1713A, 17th Floor, Block A, Zhongguan Times Square
Nanshan District
Shenzhen, 518000
CHINA
Re: K243107
Trade/Device Name: Water Solubility Human Body Lubricant
Regulation Number: 21 CFR 884.5300
Regulation Name: Condom
Regulatory Class: II
Product Code: NUC
Received: April 7, 2025
Dear Joyce Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
Page 2
K243107 - Joyce Yang Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
Page 3
K243107 - Joyce Yang Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K243107
Device Name
Water Solubility Human Body Lubricant
Indications for Use (Describe)
Water Solubility Human Body Lubricant is a personal lubricant for penile, anal and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
Page 5
510(k) Summary
K243107
Water Solubility Human Body Lubricant
1. Submitter Information
Applicant: Dongguan Yanxuan Biotechnology Co., Ltd.
Contact: Mr. Wang Long
Address: Room 203, No. 509, Shenyuan 1st Lane, Qiaotou Town, Dongguan City, Guangdong Province Dongguang China
Phone: +86-15318330732
Email: 3175281971@qq.com
2. Correspondent Information
Applicant: Shenzhen Joyantech Consulting Co., Ltd.
Contact: Ms. Joyce Yang
Address: 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District, Shenzhen Shenzhen 518000 China
Phone: +86-755-8606919
Email: joyce@cefda.com
3. Date prepared: May 5, 2025
4. Subject Device Information
Device Trade Name: Water Solubility Human Body Lubricant
Common Name: Personal Lubricant
Regulation Number: 21 CFR 884.5300
Regulation Name: Condom
Product Code: NUC (lubricant, personal)
Device Class: Class II
5. Predicate Device Information
Device Name: Agape Warming Personal Lubricant
510(k) Number: K200208
Manufacturer: CC Wellness LLC
The predicate device has not been subject to a design-related recall.
6. Device Description
Water Solubility Human Body Lubricant is a clear, colorless, odorless and semi-viscous personal lubricant that is compatible with condoms made of natural rubber latex and polyisoprene, and is not compatible with polyurethane condoms. This device is a non-sterile personal lubricant for penile, anal, and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication.
This product is sold as an over-the-counter (OTC) product in 120 mL size provided in clear polyethylene terephthalate (PET) bottle. The bottles are capped with silver disc tops. The individual bottles are hermetically sealed during the production process.
K243107
Page 1 of 4
Page 6
The device is composed of Hyaluronic acid, Lubrajel CG, Lubrajel oil, Panthenol, Trimethyl glycine, Pentanediol, Hydroxyacetophenone, Ethylene glycol, Glycerol, Purified water.
Device specifications for the Water Solubility Human Body Lubricant are listed in Table 1 below.
Table 1: Device Specifications for Water Solubility Human Body Lubricant
| Property | Specification |
|---|---|
| Appearance | Semi-viscous liquid |
| Color | Clear |
| Odor | Odorless |
| Viscosity (per USP<911>) | 3500 – 4500 cps |
| pH (per USP<791>) | 5.6-6.4 |
| Specific gravity (per USP<831>) | 0.950-1.100 |
| Osmolality (per USP<785>) | 100-150 mOsm/kg (1:5 dilution) |
| Total Aerobic Microbial Count (TAMC, per USP <61>) | <100 cfu/g |
| Total Yeast and Mold Count (TYMC, per USP <61>) | <10 cfu/g |
| Presence of Pathogens (per USP <62>) | Specification |
|---|---|
| Pseudomonas aeruginosa | Absent |
| Staphylococcus aureus | Absent |
| Candida albicans | Absent |
| Escherichia coli | Absent |
| Antimicrobial Effectiveness Testing (per USP <51>) | Specification |
|---|---|
| Bacteria | Meets USP <51> criteria for category 2. No less than 2.0 log reduction from initial count at 14 days and no increase from the 14-day count at 28 days |
| Yeast and Molds | No increase from the initial calculated count at 14 and 28 days |
7. Indications for Use
Water Solubility Human Body Lubricant is a personal lubricant for penile, anal and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.
8. Comparison of Intended Use and Technological Characteristics with the Predicate Device
The table below compares the intended use and technological characteristics of the subject and predicate device.
Table 2: Intended Use and Technological Characteristics Comparison of the Subject and Predicate Device
| Water Solubility Human Body Lubricant K243107 Subject Device | Agape Warming Personal Lubricant K200208 Predicate Device |
|---|
K243107
Page 2 of 4
Page 7
| Indications for Use | Water Solubility Human Body Lubricant is a personal lubricant for penile, anal and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. | Agape Warming Personal Lubricant is a personal lubricant for penile, anal and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms. |
|---|---|---|
| Base type | Water | Water |
| Primary ingredients | Purified water, Hyaluronic acid, Lubrajel CG, Lubrajel oil, Panthenol, Trimethyl glycine, Pentanediol, Hydroxyacetophenone, Ethylene glycol, Glycerol | water, Propanediol, Gluconolactone, Hydroxyethylcellulose, Sodium Benzoate, Polysorbate 20, Citric Acid and Capsicum Oleoresin |
| Rx/OTC | OTC | OTC |
| Sterile | No | No |
| Appearance | Semi-viscous liquid | Semi-viscous liquid |
| Odor | Odorless | Odorless |
| Viscosity per USP <911> | 3500 – 4500 cps | 2150 – 4000 cps |
| pH per USP <791> | 5.6 – 6.4 | 3.4 – 4.5 |
| Osmolality per USP <785> | 100 – 150 mOsm/Kg (1:5 dilution) | 575 – 750 mOsm/kg |
| Total Aerobic Microbial count (TAMC) per USP <61> | <100 cfu/g | <100 cfu/g |
| Total Yeast and Mold Count (TYMC) per USP <61> | <10 cfu/g | <10 cfu/g |
| Absence of Pathogenic Organisms per USP <62> | Yes | Yes |
| Antimicrobial Effectiveness Tested per USP <51> | Yes | Yes |
| Condom Compatibility | Compatible with natural rubber latex and polyisoprene condoms | Compatible with natural rubber latex and polyisoprene condoms |
| Biocompatibility Tested | Yes | Yes |
The subject and predicate device have similar indications for use and the same intended use – to provide lubrication for intimate sexual activity. The subject and predicate device have different technological characteristics, including different formulations, and specifications for viscosity, pH, and osmolality. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness.
9. Summary of Non-Clinical Performance Testing
K243107
Page 3 of 4
Page 8
Biocompatibility
Biocompatibility testing was performed in accordance with the 2023 FDA guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process." The following testing was conducted:
- Cytotoxicity (ISO 10993-5:2009)
- Sensitization (ISO 10993-10:2021)
- Vaginal Irritation (ISO 10993-23:2021)
- Acute Systemic Toxicity (ISO 10993-11:2017)
The results of testing demonstrate that the subject device is non-cytotoxic, non-irritating, non-sensitizing, and not acutely, systemically toxic.
Shelf-Life
The subject device has a shelf-life of 3 years. Results from accelerated age testing demonstrated that the device maintains its specifications (as shown in Table 1) over the duration of its shelf-life.
Condom Compatibility
The compatibility of Water Solubility Human Body Lubricant with condoms was evaluated in accordance with ASTM D7661-2023 "Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms." The results of this test showed Water Solubility Human Body Lubricant is compatible with natural rubber latex and polyisoprene condoms. Results showed Water Solubility Human Body Lubricant is not compatible with polyurethane condoms.
10. Conclusion
The results of the performance testing described above demonstrate that Water Solubility Human Body Lubricant is as safe and effective as the predicate device and supports a determination of substantial equivalence.
K243107
Page 4 of 4
§ 884.5300 Condom.
(a)
Identification. A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.(b)
Classification. (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.