K Number
K242718

Validate with FDA (Live)

Date Cleared
2025-02-28

(171 days)

Product Code
Regulation Number
868.5440
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Portable Oxygen Concentrator P2-TOC is intended to provide supplemental low flow oxygen.

The device is not intended for life support, nor does it provide any patient monitoring capabilities. This device is for adults only.

Device Description

The Portable Oxygen Concentrator P2-TOC is a device that uses the principle of molecular sieve pressure swing adsorption to increase oxygen concentration by adsorption of nitrogen and other gas components. The device needs to be used with a nasal oxygen cannula, which can provide oxygen supplementation to the user.

The Portable Oxygen Concentrator P2-TOC has two oxygen supply modes, namely continuous oxygen supply mode and pulse oxygen supply mode. Hereinafter referred to as "continuous mode" and "pulse mode" .

In the continuous mode, the Portable Oxygen Concentrator P2-TOC can continuously deliver oxygen at a fixed flow rate. In the pulse mode, the Portable Oxygen Concentrator P2-TOC is able to deliver oxygen only when the user inhales by detecting the human respiratory rate.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a Portable Oxygen Concentrator (P2-TOC). It details the device's characteristics and compares it to predicate devices to demonstrate substantial equivalence. However, the document does not contain information related to an AI/ML-enabled device or a study involving human experts, ground truth establishment, or comparative effectiveness with AI assistance.

Therefore, I cannot fulfill most of the requested points as the provided document is for a physical medical device (portable oxygen concentrator) and not an AI/ML diagnostic or therapeutic device.

The document primarily focuses on:

  • Device Description and Indications for Use: What the portable oxygen concentrator does.
  • Principle of Operation: How it separates oxygen from air.
  • Comparison with Predicate Devices: Demonstrating similarity in function and safety to already-marketed oxygen concentrators.
  • Non-Clinical Tests Performed: This section lists various engineering and safety standards (electrical safety, EMC, software verification, biocompatibility) that the device complies with. These are standard tests for medical devices to ensure quality and safety, not AI model performance.
  • Absence of Clinical Accuracy Testing: Explicitly states "There was no clinical testing performed." This is common for 510(k) submissions where substantial equivalence can be demonstrated through non-clinical means.

Here's what I can extract from the provided text regarding acceptance criteria and performance, focusing on what is relevant for a physical medical device:

The "acceptance criteria" for a physical device like a Portable Oxygen Concentrator are typically defined by compliance with recognized consensus standards and performance specifications. The "study that proves the device meets the acceptance criteria" refers to the non-clinical testing performed to show compliance with these standards and performance specifications.

1. Table of Acceptance Criteria and Reported Device Performance (Based on "Comparison with predicate device" and "Non-Clinical Tests Performed"):

Acceptance Criteria CategorySpecific Acceptance Standard/Criterion (Implied or Explicit)Reported Device Performance / Compliance
Safety & ElectricalANSI AAMI ES 60601-1 (General Safety)Complies with AAMI ANSI ES60601-1
IEC 60601-1-11 (Home Healthcare)Complies with IEC 60601-1-11
Electromagnetic Comp.IEC 60601-1-2 (EMC)Complies with IEC 60601-1-2
Performance (Oxygen Conc.)ISO 80601-2-69 (Oxygen Concentrator Equ.)Complies with ISO 80601-2-69
ISO 80601-2-67 (Oxygen-conserving Equ.)Complies with ISO 80601-2-67
SoftwareIEC 62304 (Software Verification)Complies; Software V&V performed per FDA Guidance
UsabilityIEC 60601-1-6 (Usability)Complies with IEC 60601-1-6
AlarmsIEC 60601-1-8 (Alarm Systems)Complies with IEC 60601-1-8
BiocompatibilityISO 18562-1, -2, -3 (Breathing Gas Pathways)Complies with ISO 18562-1, -2, -3
ISO 10993-1, -5, -10, -23 (Biological Evaluation - Patient Contact)Complies with ISO 10993-1, -5, -10, -23
Battery SafetyIEC 62133-2 (Lithium Batteries)Complies with IEC 62133-2
Oxygen Concentration90%-3%/+6% at all settings (P2-TOC)"90%-3%/+6% at all settings" (vs. Predicates 90% +/- 3%) - Risk mitigated by ISO 80601-2-67: 2020 testing.
Acoustic Noise≤ 60dB (A) (Gear 10 of pulse mode/ Gear 5 of continuous mode)"litter bigger than predicate device Model 1000" but "standard requirements have been met" (ISO 80601-2-69, IEC 60601-1)
Physical Dimensions/Weight(Not a specific standard, but compared to predicates)Different from predicates but risks mitigated by ISO 80601-2-69, EC60601-1, ISO 80601-2-67 testing.
Flow Rates(Not a specific standard, but compared to predicates)Continuous: 1.0-3.0 LPM (Same as predicates). Pulse: 5-200 ml Pulse Volumes (differs from predicates, max flow rate is smaller) - risks mitigated by various ISO/IEC testing.

Regarding the points specific to AI/ML devices, I must state that the document does not provide the requested information for the following reasons:

2. Sample size used for the test set and the data provenance: Not applicable. This document describes testing for a physical device's compliance with engineering standards, not an AI model trained on data. There is no concept of a "test set" from a data perspective. The "testing" refers to bench testing of the physical hardware and embedded software. Data provenance would not be a relevant concept as it pertains to clinical data for AI model training/testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No clinical test set or ground truth established by experts is mentioned. The ground truth for device performance is based on physical measurements against established engineering and medical device standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. No clinical test set requiring expert adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a physical oxygen concentrator, not an AI-assisted diagnostic or therapeutic device that would involve human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This refers to a physical device, not a standalone AI algorithm. While it has embedded software, its "performance" is measured by its physical output (oxygen concentration, flow rate) and adherence to safety standards, not by an algorithm's classification or detection accuracy.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable. The "ground truth" for this device's performance is determined by direct measurement of physical parameters (e.g., oxygen purity, flow volume, noise levels) and compliance with specified engineering and safety standards, as detailed in the "Non-Clinical Tests Performed" section.

8. The sample size for the training set: Not applicable. This is for an AI/ML model, not a physical medical device. The device itself is manufactured; it doesn't "learn" from a training set in the AI sense.

9. How the ground truth for the training set was established: Not applicable. See point 8.

In summary, the provided document is a regulatory submission for a physical Portable Oxygen Concentrator, and therefore, the information requested, which pertains to AI/ML device performance and validation studies, is not present. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" for this device are based on demonstrating compliance with a comprehensive set of recognized consensus engineering and safety standards through non-clinical bench testing.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. The words "FOOD & DRUG" are on the top line, and the word "ADMINISTRATION" is on the bottom line.

February 28, 2025

Qingdao Kingon Medical Science and Technology Co., Ltd Benrong Zhang Official Correspondent Room 301-302(B), No.15 Hancheng Road,Qingdao Free Trade Zone 24th Building, NO. 252 Yanhe Road, Huangdao Qingdao, Shandong 266510 China

Re: K242718

Trade/Device Name: Portable Oxygen Concentrator (P2-TOC) Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: Class II Product Code: CAW Dated: January 27, 2025 Received: January 27, 2025

Dear Benrong Zhang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical

{2}------------------------------------------------

devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bradley Q. Quinn -S

Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242718

Device Name Portable Oxygen Concentrator (P2-TOC)

Indications for Use (Describe)

The Portable Oxygen Concentrator P2-TOC is intended to provide supplemental low flow oxygen.

The device is not intended for life support, nor does it provide any patient monitoring capabilities. This device is for adults only.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary as required by section 807.92(c)

I. Date of the summary prepared: 20/01/2025

II. Administrative Information

ManufacturerinformationEstablishment registration number3014777423
Owner/Operator Number10061814
NameQingdao Kingon Medical Science andTechnology Co., Ltd.
AddressRoom 301-302, No.15 Hancheng Road,Qingdao Free Trade Zone,Shandong, China, 266555
Contact PersonName: Benrong ZhangAddress: Room 301-302, No.15Hancheng Road, Qingdao FreeTrade Zone, Shandong, China, 266555TEL: +86-18565833539FAX: +86 532 58792324Email: augus@kingonmed.com

III. Device Information

Type of 510(k)Traditional 510K
Prior submissionK240354
Common NameGenerator, Oxygen, Portable
Classification namePortable oxygen generator
Trade NamePortable Oxygen Concentrator (Model:P2-TOC)
Review panelAnesthesiology
Product codeCAW
Regulation Number868.5440
Regulation Class2

IV. Predicate Device Information

1 Product namePortable Oxygen Concentrator. Model:P2-S4,P2-S3,P2-K4,P2-K3
510(k) numberK230702
2 Product namePortable Oxygen Concentrator. Model: P2-E7,P2-E

{5}------------------------------------------------

510(k) numberK223379
3Product namePortable Oxygen Concentrator. Model: P2-E6
510(k) numberK210371
4Product nameKingon P2 Oxygen Concentrator
510(k) numberK190304

V. Device description and Indications for Use

Device description: The Portable Oxygen Concentrator P2-TOC is a device that uses the principle of molecular sieve pressure swing adsorption to increase oxygen concentration by adsorption of nitrogen and other gas components. The device needs to be used with a nasal oxygen cannula, which can provide oxygen supplementation to the user.

The Portable Oxygen Concentrator P2-TOC has two oxygen supply modes, namely continuous oxygen supply mode and pulse oxygen supply mode. Hereinafter referred to as "continuous mode" and "pulse mode" .

In the continuous mode, the Portable Oxygen Concentrator P2-TOC can continuously deliver oxygen at a fixed flow rate. In the pulse mode, the Portable Oxygen Concentrator P2-TOC is able to deliver oxygen only when the user inhales by detecting the human respiratory rate.

Indications for Use: The Portable Oxygen Concentrator P2-TOC is intended to provide supplemental low flow oxygen.

The device is not intended for life support, nor does it provide any patient monitoring capabilities. This device is for adult only.

VI. Principle of operation

The Portable Oxygen Concentrator P2-TOC works by getting use of the molecular sieves character that the internal pressure of a sealed container containing of molecular sieve will increase when injecting air into it. At this time, the molecular sieve will absorb a lot of nitrogen in the air with the increasing of ambient pressure, while the oxygen in the air is still existed in gaseous form, then the oxygen are collected in the Air receiver. When the nitrogen absorption process in the container reaches a certain level, then exhaust of the vacuum container and nitrogen will be released from molecular sieve with the ambient pressure decreases. It will detect when the user begins to take a breath and then delivers a pulsed volume of oxygen during the inhalation period. The volume of the oxygen pulse is dependent on the setting value.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "kingo'n" in a dark blue font. The letters are all lowercase, and there is a small circle above the "o" in "kingo'n". The word is centered and takes up most of the frame.

VII. Comparison with predicate device

Model 1000 Portable Oxygen Concentrator and Model 4000 that manufactured by CAIRE was selected for comparison. See the following table for details of comparison:

IDItemsPredicate Device (K013931)Reference Device (K120785)Device to be submitted for 501KComparison
1Device NameOmni Portable Oxygen ConcentratoreQuinox Oxygen SystemPortable Oxygen ConcentratorSimilar
2ModelModel 1000Model 4000P2-TOCDifferent modenames
3Indication forUseThe Omni portable oxygen concentratoris intended for the administration ofsupplemental oxygen. The device is notintended for life support nor does itprovide any patient monitoringcapabilities.The eQuinox is intended for theadministration of supplemental oxygen. Thedevice is not intended for life support, nordoes it provide any patient monitoringcapabilities.The Portable Oxygen Concentrator P2-TOC isintended to provide supplemental low flowoxygen..The device is not intended for life support, nordoes it provide any patient monitoringcapabilities.Similar(See below noteID_3)
4Environment ofUseHome, outside the home and institutionsHome, outside the homeHome, outside the homeSimilar(See below noteID_4)
5PrescriptiveYesYesYesSame
6Patient InterfaceNasal CannulaNasal CannulaNasal CannulaSame
7TechnologyPressure Swing Adsorption withmolecular sievePressure Swing Adsorption with molecularsievePressure Swing Adsorption with molecularsieveSame
8Dimensions19.3 high x 12.3 wide x 7.1 deep(inches)49.0 height x 31.2 wide x 18.0 deep15.5 high x10.8 wide x 7.3 deep (inches)11.1 length x 6.7 width x15 height (inch)28.2 length x 17.1 width x 38.2 height (cm)Different (See belownote ID_8)
(cm)
9Weight17.9 pounds (8.1 kg) with PowerCartridge12lbs18.8 lbs / 8.5kg ( with battery)Different (See belownote ID_9)
10OxygenConcentration90% +/- 3% for all flow settings90%+3%/-3% at all settings90%-3%/+6% at all settingsDifferent (See belownoteID_10)
11Equivalent FlowRatesContinuous Flow Mode: 1.0-3.0 LPMContinuous Flow Mode: 1.0-3.0 LPMContinuous Flow Mode: 1.0-3.0 LPMDifferent (See belownoteID_11)
Pulse Dose Mode: 16-96 ml PulseVolumesPulse Dose Mode: 16-192 ml Pulse VolumesPulse Dose Mode: 5-133 ml Pulse VolumesBreathing rate is 15 times a minute
6 to 25 breaths per minute average,sustainableBreathing rate is 15 times a minutePulse Dose Mode: 5-200 ml Pulse Volumes
Breathing rate is 10 times a minute
12FiltersAir Inlet FilterAir Inlet FilterAir Inlet FilterSame
13User InterfaceButtons, LCD DisplayButtons, LCD DisplayButtons, LCD DisplaySame
14Power SupplyAC Power (100-240 VAC, 50-60 Hz)AC Power (100VAC, 50-60 Hz)AC Adaptor (Input: 100-240 VAC, 50-60 Hz)Similar (See belownoteID_14)
DC Power (12V nominal)DC Power (12V nominal)DC Adaptor (Input: 11-16 VDC Output: 19V)
Battery (Lithium Ion)Battery (Lithium Ion)Battery (Lithium Ion)
15SoftwareEmbeddedEmbeddedEmbeddedSame
16Acoustic Noise48 dBA at 3.0 LPM in Continuous FlowMode<40 dBA≤ 60dB (A) (Gear 10 of pulse mode/ Gear 5 ofcontinuous mode)Different (See belownoteID_16)
40 dBA at 3.0 LPM in Pulse Flow Mode
17AlarmsLoss of powerLoss of powerLoss of powerDifferent (See belownoteID_17)
Low BatteryLow BatteryLow Battery
Low Therapeutic O2 OutputLow Therapeutic O2 OutputLow O2 Concentration
O2 flow outside normal limitsO2 flow outside normal limitsGas Delivery Fail
No Inspiration detected in Pulse Dose ModeNo Inspiration detected in Pulse Dose ModeAbsenceof Breath (Only in Pulse Mode)
Unit MalfunctionUnit MalfunctionSystem Startup Fail
18Status IndicatorTool iconTool iconTool iconSame
Battery Status GaugeBattery Status GaugeBattery Status GaugeSame
External Power IndicatorExternal Power IndicatorExternal Power IndicatorSame
Flow Control SettingFlow Control SettingFlow Control SettingSame
Low Priority Technical AlarmLow Priority Technical AlarmLow Priority Technical AlarmSame
BuzzerBuzzerBuzzerSame
19Battery DurationUp to 5.4 hours (Pulse setting of 1.0 at 12 BPM )Up to 5.94 hours (Pulse setting of 1.0 at 12 BPM )
20Operating EnvironmentTemperature: 50° F to 104° F (10° C to 40° C)Relative humidity: 10% - 95% at an 82.4° F (28°C) dew pointAltitude: 0 – 13,123 feet (0 - 4,000 meters)0 °C tot 40 °C (32 °F tot 104 °F)Temperature: 41 to 104°F (5 to 40°C)Humidity: 10% to 90%, non-condensingAltitude: 0 to 10,000 ft. (0 to 3048 meters, 70kPa to 106kPa)Different (See below note ID_20)
21Shipping Storage Environment-4° F to 140° F (-20° C to 60° C)Relative humidity: Up to 95% Non-Condensing-25 °C tot 70 °C (-13 °F tot 158 °F)Vochtigheid: tot 95% niet condenserend-20 °C tot 50 °C (-4 °F tot 122 °F) voor accu'sTemperature: -4 to 158°F (-20 to 70°C)Humidity: 5% to 90%, non-condensingStore in a dry environmentAltitude range: 0 to 10,000 ft (0 to 3048 meters,70kPa to 106 kPa)Different (See below note ID 21)
22Principle of OperationPulse Flow Mode is designed to deliver 16 ml boluses of oxygen per incrementalPulse Flow Mode is designed to deliver 16 ml boluses of oxygen per incremental flowThe Portable Oxygen Concentrator P2-TOC works by getting use of the molecular sievesSimilar.Both use pressure
flow setting for each patient inspirationdetected.Bolus volume is held constantregardless of breath rate within a givensetting. Delivery of the bolus is triggeredby detection of the end of inspirationfollowed by the start of a subsequentinspiration. Sensitivity is dealer/provideradjustable. Based on this inspirationtrigger signal, a "pulse" of oxygen isreleased to the patient by anelectronically controlled flow valve.There is no mixing of oxygen with roomair.setting for each patient inspirationdetected.Bolus volume is held constantregardless of breath rate within a givensetting. Delivery of the bolus is triggered bydetection of the end of inspiration followed bythe start of a subsequent inspiration.Sensitivity is dealer/provider adjustable.Based on this inspiration trigger signal, a"pulse" of oxygen is released to the patient byan electronically controlled flow valve. Thereis no mixing of oxygen with room air.character that the internal pressure of a sealedcontainer containing of molecular sieve willincrease when injecting air into it. At thistime, the molecular sieve will absorb a lot ofnitrogen in the air with the increasing ofambient pressure, while the oxygen in the airis still existed in gaseous form, then theoxygen are collected through some pipelines.When the nitrogen absorption process in thecontainer reaches a certain level, then exhaustof the vacuum container and nitrogen will bereleased from molecular sieve with theambient pressure decreases. It will detectwhen the user begins to take a breath and thendelivers a pulsed volume of oxygen during theinhalation period. The volume of the oxygenpulse is dependent on the setting value.swing adsorptionwith molecular sieve
23DeviceComponentsOxygen concentrator, Universal Cart,AC Power Supply, DC Power Supply,Battery.Oxygen concentrator, Cart, AC Power SupplyDC Power Supply, Battery.Oxygen concentrator, Cart, AC Power Supply,AC Power Supply, Battery, Backpack.Similar.A backpackis provided forpatient convenience
24Materials forfiltersChemical fiberChemical fiberChemical fiberSame
25Material ofPatient contactcomponentsButton panel: PETMain housing: PC+ABSButton panel: PETMain housing: PC+ABSButton panel: PETMain housing: PC+ABSSame
26PatientAdultAdultAdultSame
Population
27Single Patient,YesYesYesSame
multi-use
28Electrical SafetyClass IIClass IIClass IISame
Classification
29Electrical SafetyIEC 60601-1IEC 60601-1IEC 60601-1Same
30ElectromagneticIEC 60601-1-2IEC 60601-1-2IEC 60601-1-2Same
compatibility
31BiocompatibilityPM2.5,PM10, VOC's less than ambientPM2.5,PM10, VOC's less than ambientPM2.5,PM10, VOC's less than ambientSame
32Output5.0 psi nominal5.0 psi nominal5 PSISame
pressure
33Classification21CFR 868.544021CFR 868.544021CFR 868.5440Same
34Product CodeCAWCAWCAWSame
35FDA ClassIIIIIISame
36Heat ex-changerCooling fanCooling fanCooling fanSame

{7}------------------------------------------------

kıngon

{8}------------------------------------------------

kıngon

{9}------------------------------------------------

kıngon

{10}------------------------------------------------

kıngo'n

Note:

ID3,ID4:

The Indication for Use description of P2-TOC is similar to the description of Model 4000. We added a "low flow" reference to the intended use to prevent user misunderstanding and reduce risk.

The Environment of Use of P2-TOC is basically the same as the description of Model 4000.Therefore the difference does not raise new questions of safety and effectiveness.

ID8:

The dimensions of Model 1000and Model 4000 compared and theP2-TOC devices differ in dimensions due to their different designs. The risk is mitigated by tests tested according to ISO 80601-2-69: 2020,EC60601-1and ISO 80601-2-67: 2020, therefore the difference does not raise new questions of safety and effectiveness.

{11}------------------------------------------------

kınqo'n

ID9, ID14, ID20 and ID21:

Although the weight, power supply, operating and shipping storage environment conditions of P2-TOC is different from the predicate devices (Model 1000 and Model 4000), but the subject device has passed the differences don't raise new questions of safety and effectiveness.

ID10:

The oxygen concentration of predicate devices Model 4000 both are 90% +- 3% for all 110w settings and of subject device P2-TOC is 90%-3% at all settings. The risk is mitigated by tests tested according to ISO 80601-2-67: 2020, therefore the difference does not raise new questions of safety and effectiveness.

ID11:

In continuous mode, both P2-TOC and predicate devices (Model 4000) have same ranges of 1.0-3.0L/min.

In pulse mode, the flow range of the Model1000 is 16-96 mL. The breathing rate is 6-25 beats per minute, so the maximum flow rate in pulse mode is 2400ml (25*96mL).

In pulse mode, the flow range of the Model4000 is 16-192mL when the breathing rate is 15 times per minute, so the maximum flow rate in pulse mode is 2880mL (15*192mL).

In pulse mode, the maximum flow rate of P2-TOC is 2000mL. The flow rate is smaller than that predicate devices. (Model 1000 and Model 4000).

The risk is mitigated by tests tested according to ISO 80601-2-67: 2020, ISO 18562-2: 2017, ISO 18562-3: 2017, ISO 18562-3: 2017, ISO 10993-5: 2009, ISO10993-10: 2021 and ISO10993-23: 2021, therefore the difference does not raise new questions of safety and effectiveness.

ID16:

The acoustic noise of subject device P2-TOC is litter bigger than predicate device Model 1000. Since the subject device has been tested against ISO 80601-2-69: 2020 and IEC60601- 1: 2005 /A1:2012 and A2:2020, the standard requirements have been met.The difference of subject device do not raise new questions of safety and effectiveness.

ID17:

The subject device and the predicate device have information displayed. However, the subject device meets the requirements of the standard ISO80601-2-69 and IEC60601-1-8. In addition, performance testing, risk analysis has been conducted on the

{12}------------------------------------------------

kıngo'n

subject device. Hence, the difference of subject device do not raise new questions of safety and effectiveness. ID19:

The subject device P2-TOC can be equipped with the battery, which can up to 9.0 hours (in pulse setting of 1 at 20 BPM). The risk is mitigated by test tested according to IEC 62133: 2017, therefore the difference does not raise new questions of safety and effectiveness.

{13}------------------------------------------------

VIII. Discussion of Non-Clinical Tests Performed for Safety and

effectiveness are as follows

The recognized consensus standards for safety of medical electrical equipment: ANSI AAMI ES60601- 1, IEC 60601-1-11 for safety, IEC 60601-1-2 for electromagnetic compatibility, ISO 80601-2-69 and ISO 80601-2-67 for performance and IEC 62304 for software verification are complied. See below table for details:

StandardsStandards Name
ANSI AAMI ES 60601- 1:2005 /A1:2012 andA2:2020Medical Electrical Equipment - Part 1: General RequirementsFor Basic Safety And Essential Performance
IEC 60601-1-2: 2020Medical Electrical Equipment -- Part 1-2: GeneralRequirements For Basic Safety And Essential Performance --Collateral Standard: Electromagnetic Disturbances --Requirements And Tests
IEC 60601-1-11:2015/A1:2020Medical Electrical Equipment - Part 1-2: GeneralRequirements For Basic Safety And Essential Performance -Collateral Standard: Electromagnetic Disturbances -Requirements And Tests
IEC 60601-1-8: 2006/A1:2012 /A2:2020Medical electrical equipment - Part 1-8: General requirementsfor basic safety and essential performance - CollateralStandard: General requirements, tests and guidance for alarmsystems in medical electrical equipment and medical electricalsystems
IEC 60601-1-6: 2010/A1:2013 /A2:2020Medical electrical equipment - Part 1-6: General requirementsfor basic safety and essential performance - Collateralstandard: Usability
ISO 80601-2-69: 2020Medical electrical equipment. Particular requirements for thebasic safety and essential performance of oxygen concentratorequipment
ISO 80601-2-67: 2020Medical electrical equipment. Particular requirements forbasic safety and essential performance of oxygen-conservingequipment
ISO 18562-1: 2017Biocompatibility evaluation of breathing gas pathways inhealthcare applications - Part 1 Evaluation and testing within arisk management process
ISO 18562-2: 2017Biocompatibility evaluation of breathing gas pathways inhealthcare applications - Part 2 Tests for emissions ofparticulate matter
ISO 18562-3: 2017Biocompatibility evaluation of breathing gas pathways in
healthcare applications - Part 3: Tests for emissions of volatile
organic compounds
IEC 62133-2: 2017Secondary cells and batteries containing alkaline or othernon-acid electrolytes - Safety requirements for portable sealedsecondary cells, and for batteries made from them, for use inportable applications-Part2: Lithium systems
ISO 10993-1:2018Biological evaluation of medical devices - Part 1: Evaluationand testing within a risk management process
ISO 10993-5:2009Biological evaluation of medical devices - Part 5: Tests for invitro cytotoxicity
ISO 10993-10:2021Biological evaluation of medical devices - Part 10: Tests forskin sensitization
ISO 10993-23:2021Biological evaluation of medical devices - Part 23:Tests forirritation

{14}------------------------------------------------

· Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the subject device model P2-TOC. The system complies with the AAMI ANSI ES60601-1, IEC 60601-1-8, IEC 60601-1-6, IEC 60601-1-11, ISO 80601-2-67, and ISO 80601-2-69 standards for electrical safety and the IEC 60601-1-2 standard for EMC.

·Software Verification and Validation Testing

Software verification and validation was performed for the subject device in accordance with Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff.

·Biocompatibility Testing

Biocompatibility testing were conducted on the subject device model P2-TOC. The system complies with the ISO 18562-2, ISO 18562-3, ISO 18562-1, ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-23 for biocompatibility safety.

IX. Discussion of Clinical Accuracy Testing Performed

There was no clinical testing performed.

X. Conclusions

The Portable Oxygen Concentrator (Model: P2-TOC) have the same intended use and similar characteristics as the cleared predicate device Portable Oxygen Concentrator, model 1000 and model 4000. Moreover, bench testing contained in this submission supplied demonstrate that the differences existed between Portable Oxygen

{15}------------------------------------------------

kıngo'n

Concentrator (Model: P2-TOC) and Portable Oxygen Concentrator, model 1000 and model 4000 do not raise any new questions of safety or effectiveness.

The non-clinical tests support the safety of the device and the hardware and software verification and validation demonstrate that the Portable Oxygen Concentrator (Model: P2-TOC) performs as intended in the specified use conditions are same with predicate device. The performance tests demonstrate that the Portable Oxygen Concentrator (Model: P2-TOC) performs comparably to the predicate device that is currently marketed for the same intended use. Thus, Portable Oxygen Concentrator (Model: P2-TOC) is Substantially Equivalent (SE) to the predicate device.

§ 868.5440 Portable oxygen generator.

(a)
Identification. A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).(b)
Classification. Class II (performance standards).