K Number
K242651
Device Name
GM85
Date Cleared
2024-10-01

(27 days)

Product Code
Regulation Number
892.1720
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GM85 Digital Mobile X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

Device Description

The GM85 Digital Mobile X-ray Imaging System is used to capture images by transmitting X-ray to a patient's body. The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process on the S-Station, which is the Operation Software (OS) of Samsung Digital Diagnostic X-ray System, and save in DICOM file, a standard for medical imaging. The captured images are tuned up by an Image Post-processing Engine (IPE) which is exclusively installed in S-Station, and send to the Picture Archiving & Communication System (PACS) sever for reading images.

The GM85 Digital Mobile X-ray imaging System was previously cleared with K222353, and through this premarket notification, we would like to add more configurations in the previously cleared GM85 as a detector, accessories are newly added and software is updated for user convenience.

The new detector added in the proposed device is designed to achieve a higher IP rating of Dust and Water and reduce weight while maintaining durability, functionality and operation like the detector of the predicate device. The new detector and predicate device's detector are both an x-ray conversion device using an amorphous silicon flat panel and absorb incident x-rays, converts it to a digital signal, and then transmits it to the Samsung Digital X-ray System like that of the predicate device.

AI/ML Overview

The provided text is a 510(k) Premarket Notification for the GM85 Digital Mobile X-ray Imaging System. It describes changes made to a previously cleared device (K222353) and argues for substantial equivalence.

Based on the provided text, the device is the GM85 Digital Mobile X-ray Imaging System. It is an X-ray imaging system, and the study focuses on the performance of a new detector (F4343-AW) and other accessories and software updates compared to the predicate device, also named GM85.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state "acceptance criteria" in a quantitative format for specific imaging metrics. Instead, it focuses on demonstrating that the proposed device, with its new detector, accessories, and software, is substantially equivalent to the predicate device (GM85, K222353). The performance is assessed by comparing technical specifications and qualitative evaluation by experts.

The key performance characteristics and comparisons are as follows:

AttributeAcceptance Criteria (Implied by Substantial Equivalence Goal)Reported Device Performance (Proposed Device)Comparison to Predicate (GM85, K222353)
Detector CharacteristicsEquivalent or improved
Detector TypeSame as predicate (CsI Indirect)CsI IndirectSame
Detector AreaSame as predicate (17"X17")17"X17" (425mmX425mm)Same
Number of pixelsSame as predicate (3036X3040)3036X3040Same
Pixel Pitch (um)Same as predicate (140)140Same
High Contrast Limiting Resolution (LP/mm)Same as predicate (3.57)3.57Same
CommunicationSame as predicate (Wired / Wireless)Wired / WirelessSame
Dust/Water-resistanceEquivalent or improved (IP54 for predicate)IP57Difference (Improved)
Max. load capacitySame as predicate (400 kg/200 kg)400 kg/200 kgSame
DQE (0lp/mm, Typical)Same as predicate (76%)76%Same
MTF (0.5lp/mm, Typical)Same as predicate (86%)86%Same
Weight (w/o Battery Set)Equivalent or improved (Approx. 3.4 kg for predicate)Approx. 2.5 kgDifference (Improved/Lighter)
Image Quality (Phantom)Equivalent to predicate"Equivalent to the predicate devices""No significant difference in the average score of image quality evaluation"
Safety and EffectivenessNo adverse impactVerified by standards and testing"does not contribute any adverse impact"

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The document refers to "Anthropomorphic phantom images" but does not specify the number of phantom images used for evaluation. It also notes that clinical data was not required.
  • Data Provenance: The study is non-clinical. The "Anthropomorphic phantom images" would have been generated in a controlled testing environment, likely at the manufacturer's facility. The country of origin of the data is implicitly South Korea, where SAMSUNG ELECTRONICS Co., Ltd. is located. The study is a prospective evaluation of the new detector and modified system against the predicate.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: The document states that phantom images "were evaluated by professional radiologists." It does not specify the number of radiologists who participated in this evaluation.
  • Qualifications of Experts: The experts are described as "professional radiologists." No further details on their experience level (e.g., years of experience, subspecialty) are provided.

4. Adjudication Method for the Test Set

The document states that phantom images "were evaluated by professional radiologists and found to be equivalent to the predicate devices" and that there was "no significant difference in the average score of image quality evaluation." This suggests a comparative scoring or assessment. However, the specific adjudication method (e.g., consensus, majority vote, independent reads with statistical comparison) is not described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly done, nor was there a "human readers improve with AI vs without AI assistance" component. The study for substantial equivalence focused on comparing the image quality of the proposed device (with new detector) against the predicate device using phantom images, evaluated by radiologists. The purpose was to show equivalence of the device's output, not the improvement of human readers with AI assistance.

6. Standalone Performance Study (Algorithm Only)

The provided text describes the device as a "Digital Mobile X-ray Imaging System," which includes hardware (X-ray generator, detector) and software for image processing (IPE, S-Station). The evaluation primarily focuses on the entire system's ability to generate radiographic images with equivalent quality to the predicate, particularly with the new detector.

While software features are mentioned (S-Share, S-Enhance, SimGrid, PEM, QAP, Bone Suppression, Remote View, Mirror View, RFID, Value-up Package), and software was updated for user convenience, the study does not report a standalone algorithm-only performance (without human-in-the-loop performance) in terms of diagnostic accuracy or reader improvement for specific diagnostic tasks. The "phantom image evaluation" evaluates the quality of the images produced by the overall system, not an AI algorithm's diagnostic output.

7. Type of Ground Truth Used

The ground truth for the phantom image evaluation was established based on expert consensus/evaluation of image quality metrics. The "professional radiologists" evaluated the anthropomorphic phantom images. This is not pathology, nor outcomes data.

8. Sample Size for the Training Set

The document does not provide information on the sample size for the training set. Training data would typically be applicable if this were an AI/CADe device with a specific machine learning model for diagnostic tasks, which is not the primary focus of this 510(k) for a basic X-ray imaging system with incremental changes. The "Image Post-processing Engine (IPE)" and features like "Bone Suppression" would have been developed using some form of training data previously, but details are absent here for this particular submission.

9. How the Ground Truth for the Training Set Was Established

As no training set sample size is provided, the method for establishing ground truth for a training set is not mentioned in this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 1, 2024

SAMSUNG ELECTRONICS Co., Ltd. % Jaesang Noh Regulatory Affairs, Principal Professional 129, Samsung-ro, Yeongtong-gu Suwon-si, Gyeonggi-do 16677 SOUTH KOREA

Re: K242651

Trade/Device Name: Gm85 Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: Class II Product Code: IZL Dated: August 30, 2024 Received: September 4, 2024

Dear Jaesang Noh:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242651

Device Name GM85

Indications for Use (Describe)

The GM85 Digital Mobile X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

Type of Use (Select one or both, as applicable)

Transactional Use (Permitted under SEC Rule 15c3-5, as amended)
Non-Transactional Use

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

SAMSUNG

510(k) Premarket Notification - Special

K242651

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted accordance with requirements of 21 CFR 807.92

    1. Date: August 30, 2024

2. Submitter

  • A. Company Name: SAMSUNG ELECTRONICS Co., Ltd.
  • B. Address: 129, Samsung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 16677, Republic of Korea

3. Primary Contact Person

  • A. Name: JAESANG NOH
  • B. Title: Regulatory Affairs, Principal Professional
  • C. Phone Number: +82-2-2193-2444
  • D. FAX Number: +82-2-2194-0284
  • E-Mail: jaesang.noh@samsung.com ய்

4. Secondary Contact Person

  • A. Name: Ninad Gujar
  • B. Title: Vice President, Regulatory Affairs & Quality Control
  • C. Phone Number: 978-564-8503
  • D. FAX Number: 978-560-0602
  • E. E-Mail: ngujar@neurologica.com

5. Proposed Device

  • A. Trade Name: GM85
  • B. Device Name: GM85
  • C. Common Name: Digital Diagnostic Mobile X-ray System
  • D. Classification Name: Mobile X-ray System
  • E. Product Code: IZL
  • ட் Regulation: 21 CFR 892.1720

6. Predicate Devices

Predicate Device
Device NameGM85
Classification NameMobile X-ray system.
Product CodeIZL
Regulation21 CFR 892.1720
510(k)#K222353
510(K) Decision DateSeptember 29, 2022

{5}------------------------------------------------

510(k) Premarket Notification - Special

7. Device Description

The GM85 Digital Mobile X-ray Imaging System is used to capture images by transmitting X-ray to a patient's body. The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process on the S-Station, which is the Operation Software (OS) of Samsung Digital Diagnostic X-ray System, and save in DICOM file, a standard for medical imaging. The captured images are tuned up by an Image Post-processing Engine (IPE) which is exclusively installed in S-Station, and send to the Picture Archiving & Communication System (PACS) sever for reading images.

The GM85 Digital Mobile X-ray imaging System was previously cleared with K222353, and through this premarket notification, we would like to add more configurations in the previously cleared GM85 as a detector, accessories are newly added and software is updated for user convenience.

The new detector added in the proposed device is designed to achieve a higher IP rating of Dust and Water and reduce weight while maintaining durability, functionality and operation like the detector of the predicate device. The new detector and predicate device's detector are both an x-ray conversion device using an amorphous silicon flat panel and absorb incident x-rays, converts it to a digital signal, and then transmits it to the Samsung Digital X-ray System like that of the predicate device.

8. Intended Use

The GM85 Digital Mobile X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

9. Summary of Technological characteristic of the proposed device compared with the predicate device

The proposed device, GM85, has the same technological characteristics and hardware as its original predicate device, GM85 (K222353). There are some differences but, they do not have significant changes in energy source or technological characteristics compared to the predicate device.

Comparisons of technological characteristics were executed and demonstrate the substantial equivalence to the predicate device.

AttributeProposed Device(GM85)Predicate Device(GM85, K222353)ComparisonResults
Device NameGM85GM85-
ManufacturerSAMSUNGELECTRONICSSAMSUNGELECTRONICS-
510(k) Number-K222353-
  • A. Comparing with Predicate Device

{6}------------------------------------------------

SAMSUNG

510(k) Premarket Notification - Special

AttributeProposed Device(GM85)Predicate Device(GM85, K222353)ComparisonResults
AppearancesImage: [C-Type*][C-Type*]*Collapsible columntype withanautomaticcollimator(C-Type)Image: [F-Type**][F-Type**]**Fixedcolumntype witha manualcollimator(F-Type)Image: [Fit-Type***][Fit-Type***]***Collapsiblecolumntype amanualcollimator(Fit-Type)Image: [C-Type*][C-Type*]*Collapsible columntype withanautomaticcollimator(C-Type)Image: [F-Type**][F-Type**]**Fixedcolumntype witha manualcollimator(F-Type)Image: [Fit-Type***][Fit-Type***]***Collapsiblecolumntype amanualcollimator(Fit-Type)Same
Indications for UseThe GM85 Digital MobileX-ray Imaging System isintended for use ingenerating radiographicimages of human anatomyby a qualified/traineddoctor or technician. Thisdevice is not intended formammographicapplicationsThe GM85 Digital MobileX-ray Imaging System isintended for use ingenerating radiographicimages of human anatomyby a qualified/traineddoctor or technician. Thisdevice is not intended formammographicapplicationsSame
AttributeProposed Device(GM85)Predicate Device(GM85, K222353)ComparisonResults
(1) Detector
NameS4335-WS4343-WS3025-WS4335-AWS4343-AWS4335-AWMS4343-AWMS3025-AWS3025-AWMF4335-AWF4343-AWF4343-AWS4335-WS4343-WS3025-WS4335-AWS4343-AWS4335-AWMS4343-AWMS3025-AWS3025-AWMF4335-AWS4343-AWDifference (1)
Detector TypeCslIndirectCslIndirectSame
Detector Area17"X17"(425mmX425mm)17"X17"(425mmX425mm)Same
Number of pixels3036X30403036X3040Same
Pixel Pitch(um)140140Same
High Contrast LimitingResolution (LP/mm)3.573.57Same
CommunicationWired / WirelessWired / WirelessSame
Dust/Water-resistanceIP57IP54Difference (1)-1
Max.load capacity(uniform load / localload,,400 kg/200 kg400 kg/200 kgSame

{7}------------------------------------------------

SAMSUNG

510(k) Premarket Notification - Special

AttributeProposed Device(GM85)Predicate Device(GM85, K222353)ComparisonResults
(1) Detector
at the center)
DQE(Detective QuantumEfficiency)76%(0lp/mm, Typical)76%(0lp/mm, Typical)Same
MTF(Modulation TransferFunction)86%(0.5lp/mm, Typical)86%(0.5lp/mm, Typical)Same
Weight(w/o Battery Set)Approx. 2.5 kgApprox. 3.4 kgDifference (1)-2
AttributeProposed Device(GM85)Predicate Device(GM85, K222353)ComparisonResults
(2) Accessories
WirelessExposureSwitchNameTS_CES(SAM)C2UW-LP-ISDR-OGRC30KCAPE H301W-S1TS_CES(SAM)C2UW-LP-ISDR-OGRC30KDifference (2)-1
TypesInfra-red BluetoothInfra-red BluetoothSame
RFIDreaderNamepcProxRDR-80031BKU V2pcProxDifference (2)-2
TypeUSB TypeUSB TypeSame
TransmitFrequency125kHz, 13.56 MHz125kHz, 13.56 MHzSame
AttributeProposed Device(GM85)Predicate Device(GM85, K222353)ComparisonResults
(3) Software Features
S-ShareS-ShareS-ShareSame
S-EnhanceS-EnhanceS-EnhanceSame
SimGridSimGridSimGridSame
PEMPEMPEMSame
QAPQAPQAPSame
Bone SuppressionBone SuppressionBone SuppressionSame
Remote ViewRemote ViewRemote ViewSame
Mirror ViewMirror ViewMirror ViewSame

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the word "SAMSUNG" in blue, block letters. The font is bold and sans-serif. The letters are evenly spaced and the word is horizontally oriented.

510(k) Premarket Notification - Special

AttributeProposed Device (GM85)Predicate Device (GM85, K222353)Comparison Results
RFIDRFIDRFIDSame
Value-up PackageValue-up Package-Difference (3)
NoComparison ResultsDiscussion
(1)DetectorA new detector (F4343-AW) is added to the GM85 device.This change does not contribute any adverse impact to thedevice's safety and effectiveness.
(1)-1Dust/Water-resistanceThe IP rating of the new detector is higher than that of thepredicate device because the process to block any potentialwater leak paths was added such as adding seals aroundvulnerable areas. This change does not contribute anyadverse impact to the device's safety and diagnosticeffectiveness.
(1)-2Weight(w/o Battery Set)The new detectors which are added to the GM85 device islighter than that of the predicate device because a non-glass substrate was used, instead of a glass substrate usedin the detector of the predicate device. The non-glasssubstrate is the same one used for the detector, F4335-AW,cleared with K222353. This change does not contribute anyadverse impact to the device's safety and diagnosticeffectiveness.
(2)-1Wireless ExposureSwitchA new Wireless Exposure Switch is added optionally to theGM85 device which is the same specification as those ofthe predicate device. These changes do not contribute anyadverse impact to the device's safety and effectiveness.
(2)-2RFID readerA RFID reader is added optionally to the GM85 device whichis the same specification as those of the predicate device.These changes do not contribute any adverse impact to thedevice's safety and effectiveness.
(3)Value-up PackageSome software features which is an option to provideconvenience for use, called as the Value-up Package, isapplied to the GM85 device and these changes do notcontribute any adverse impact to the device's safety andeffectiveness.

B. Safety, EMC and Performance Data

Electrical, mechanical, environmental safety and performance testing according to standard ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO 14971, 21 CFR1020.30 and 21 CFR1020.31 were performed, and EMC testing was conducted in accordance with standard IEC 60601-1-2. Wireless function was tested and verified followed by guidance, Radio frequency Wireless Technology in

{9}------------------------------------------------

510(k) Premarket Notification - Special

Medical Devices. All test results were satisfying the standards.

  • C. Non-clinical data
    Non-clinical data was provided in conformance to the FDA "Guidance for the Submission of 510(k)'s for Solid-State X-ray Imaging Devices", which includes MTF and DQE measurements as tested by IEC 62220-1.

The proposed device has a new detector that has equivalent image characteristics as the existing ones. Specific description is added to make it clear with the non-clinical data and phantom image evaluation report. And this detector is evaluated by Software System Test Case for verification and validation.

Phantom image evaluation of the new detector was performed in accordance with FDA quidance for the submission of 510(k)'s for Solid State X-ray Imaging Devices. Anthropomorphic phantom images were provided; these images were not necessary to establish substantial equivalence based on the modifications to the device (note X-ray flat-panel detector similar to the predicate detector) but they provide further evidence in addition to the performance data to show that the complete system works as intended. They were evaluated by professional radiologists and found to be equivalent to the predicate devices. There is no significant difference in the average score of image quality evaluation between the proposed device and the predicate device. Therefore, this change does not affect either the safety or the effectiveness, compared to the predicate device.

  • D. Clinical data
    This submission does not required clinical data.

  • E. Summary of the Standards and Guidance Compliance

      1. ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012 C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical Electrical Equipment -Part 1: General requirements for basic safety and essential performance
    • IEC 60601-1-2 Edition 4.1 Medical Electrical Equipment Part 1-2: General 2. Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic compatibility - Requirements and Tests
      1. IEC 60601-1-3 Edition 2.2 Medical Electrical Equipment – Part 1-3: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Radiation protection in Diagnostic X-ray Equipment
      1. IEC 60601-2-28 Edition 3 Medical Electrical Equipment - Part 2-28: Particular Requirements for the Basic Safety and Essential Performance of X-ray Tube Assemblies for Medical Diagnosis
    • IEC 60601-2-54 Edition 2.0 Medical Electrical Equipment Part 2-54: particular 5. Requirements for the Basic Safety and Essential Performance of X-ray Equipment for Radiography and Radioscopy
    • IEC 62220-1-1 Edition 1.0 Medical electrical Equipment Characteristics of digital 6. X-ray imaging devices Part 1-1: Determination of the detective quantum efficiency Detectors used in radiographic imaging
    • Cybersecurity in Medical Device: Quality System Considerations and Content of 7. Premarket Submissions issued on September 27, 2023

{10}------------------------------------------------

510(k) Premarket Notification - Special

    1. Content of Premarket Submissions for Device software Functions issued on June 14, 2023
    1. Guidance for the Submission for 510(k) for Solid State X-ray Imaging Devices Guidance for Industry and Food and Drug Administration Staff issued on September 1, 2016
    1. Pediatric Information for X-ray Imaging Device Premarket Notifications Guidance for Industry and Food and Drug Administration Staff issued on November 28, 2017
  • F. Conclusions

The non-clinical data demonstrates that the proposed device is as safe, as effective, and performs as well as the legally marketed devices.

§ 892.1720 Mobile x-ray system.

(a)
Identification. A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.