K Number
K242625

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-10-25

(52 days)

Product Code
Regulation Number
878.4350
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Skin Clinic Freeze Point product is intended for the treatment of common and plantar warts by OTC consumers. May be used with children 4 years of age or older under adult supervision.

Device Description

The Skin Clinic Freeze Point device utilizes extreme cold to facilitate the removal of warts and by freezing. Each kit contains a container of cryogen gas in a plastic holder and instructions for use. The device is for OTC use and utilizes difluoroethane cryogen delivered from the canister into a foam wrapped tip which acts as a reservoir for the cryogen gas. The gas rapidly evaporates and cools the applicator to as low as -50°C. The applicator is then placed against the wart for up to 40 seconds which freezes the targeted tissue. The frozen wart falls away over time and new epidermis grows in its place. One to four treatments with intervals of two weeks may be required.

AI/ML Overview

This document describes the FDA's 510(k) clearance for the Skin Clinic Freeze Point device. However, it does not contain the detailed information necessary to fully answer your request regarding acceptance criteria and the comprehensive study that proves the device meets those criteria, particularly for a medical device that relies on AI/ML or complex diagnostic capabilities.

The provided text focuses on demonstrating substantial equivalence to predicate devices for a cryosurgical unit, which is typically cleared based on bench testing (temperature comparison, cell destruction), biocompatibility, and human factors usability for an OTC product. It does not involve performance metrics or study designs typically associated with AI/ML devices or diagnostic accuracy studies (e.g., sensitivity, specificity, MRMC studies, detailed ground truth establishment, or specific acceptance criteria for algorithm performance).

Therefore, I cannot provide a complete answer with all the requested details (especially points 1, 2, 3, 4, 5, 6, 7, 8, 9) because the information is not present in this kind of 510(k) summary for this type of device.

However, I can extract what is provided and explain why other points are not applicable or present:


Based on the provided document for the Skin Clinic Freeze Point device:

1. A table of acceptance criteria and the reported device performance:

This 510(k) summary does not present specific quantitative acceptance criteria or detailed performance results in the format of a table as one would expect for a diagnostic or AI/ML device (e.g., sensitivity, specificity thresholds). Instead, "performance" is demonstrated by:

  • Bench Testing:
    • Acceptance Criteria (Implied): Temperatures attained by the Skin Clinic Freeze Point product are comparable to the Wartie and Freeze n Clear Skin Clinic predicate devices.
    • Reported Performance: "Bench testing compared the temperatures attained by the Wartie and Freeze n Clear Skin Clinic predicate devices to the Skin Clinic Freeze Point product to demonstrate that they were comparable."
    • Acceptance Criteria (Implied): Ability to freeze and destroy target cells is comparable to predicate devices.
    • Reported Performance: "Additional bench testing using an in vitro method compared all the same products and showed they were comparable in their ability to freeze and destroy target cells."
  • Biocompatibility:
    • Acceptance Criteria (Implied): Meets the requirements of ISO 10993.
    • Reported Performance: "product was tested for biocompatibility including cytotoxicity, sensitivity, irritation according to FDA's biocompatibility 2020 guidance and meets the requirements of ISO 10993."
  • Human Factors Usability:
    • Acceptance Criteria (Implied): Consumers can comprehend and understand the labeling and use of the product for OTC indication.
    • Reported Performance: "The human factors usability testing evaluated consumers ability to comprehend and understand the labelling and use of the product."

2. Sample size used for the test set and the data provenance:

  • Bench Testing: Not specified as "sample size" in the context of clinical data. It refers to a comparison of devices rather than patient data. Data provenance is implied to be laboratory testing.
  • Human Factors Usability Study: The specific sample size for the usability study is not provided in this summary. Data provenance would be prospective usability testing with human participants.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable / Not provided. This device is a cryosurgical unit, not a diagnostic device requiring expert interpretation for ground truth. Its performance is evaluated through physical characteristics (temperature, cell destruction) and user comprehension.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable / None specified. This concept is relevant for diagnostic accuracy studies involving multiple readers or assessors to establish a reference standard, which is not the primary mode of evaluation for this device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, not done. This is a physical device (cryosurgical unit), not an AI-assisted diagnostic or imaging device. Therefore, MRMC studies and AI assistance are irrelevant to its evaluation.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device does not have an "algorithm" in the sense of AI/ML or image processing that performs standalone tasks.

7. The type of ground truth used:

  • Bench Testing:
    • Temperature Measurement: The "ground truth" would be the measured temperature of the applicator or the frozen tissue, compared against the predicate devices' performance.
    • Cell Destruction: The "ground truth" would be the observed destruction of target cells in an in vitro model, compared against the predicate devices' performance.
  • Human Factors Usability: The "ground truth" would be the observed user interactions and comprehension of instructions, typically assessed against pre-defined usability success criteria (e.g., successful task completion, correct answers to comprehension questions).

8. The sample size for the training set:

  • Not applicable / Not provided. This device does not use a "training set" in the context of machine learning or algorithms. Its design and manufacturing are based on established engineering principles and safety standards for medical devices.

9. How the ground truth for the training set was established:

  • Not applicable. See point 8.

Summary:

The provided 510(k) summary focuses on demonstrating substantial equivalence for a cryosurgical unit based on its physical characteristics, biocompatibility, and user-friendliness for over-the-counter use, not on complex diagnostic or AI-driven performance metrics. Therefore, many of the questions related to AI/ML device evaluation are not applicable or answered by this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG' and 'ADMINISTRATION' in blue text.

October 25, 2024

CryoConcepts LP Sam Niedbala, Ph.D. CEO 1100 Conroy Place Easton, Pennsylvania 18040

Re: K242625

Trade/Device Name: Skin Clinic Freeze Point Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: GEH Dated: August 30, 2024 Received: September 3, 2024

Dear Sam Niedbala:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Digitally signed by Long H. Chen Long H. Chen-S Date: 2024.10.25 09:30:41 -04'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K242625

Device Name Skin Clinic Freeze Point

Indications for Use (Describe)

The Skin Clinic Freeze Point product is intended for the treatment of common and plantar warts by OTC consumers. May be used with children 4 years of age or older under adult supervision.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K242625

Premarket Notification

510(K) SUMMARY

(As Required by 21 CFR 807.92)

I. SUBMITTER

CryoConcepts LP, 1100 Conroy Place Easton, PA 18040 Phone: 855-355-2796 Contact: Sam Niedbala, Ph.D. October 24, 2024

II. DEVICE

Name of Device: Skin Clinic Freeze Point

Usual Name - Cryosurgical unit and accessories

Classification Name - General & Plastic Surgery

Regulatory Class: II

Code of Federal Regulation: 878.4350

Product Code: GEH

III. PREDICATE DEVICES

Skin Clinic Freeze n' Clear for Skin Tags & Warts K211099

Wartie Wart Remover K140314

IV. DEVICE DESCRIPTION

The Skin Clinic Freeze Point device utilizes extreme cold to facilitate the removal of warts and by freezing. Each kit contains a container of cryogen gas in a plastic holder and instructions for use. The device is for OTC use and utilizes difluoroethane cryogen delivered from the canister into a foam wrapped tip which acts as a reservoir for the cryogen gas. The gas rapidly

Page 1 of 4

{5}------------------------------------------------

evaporates and cools the applicator to as low as -50°C. The applicator is then placed against the wart for up to 40 seconds which freezes the targeted tissue. The frozen wart falls away over time and new epidermis grows in its place. One to four treatments with intervals of two weeks may be required.

V. INDICATIONS FOR USE

The Skin Clinic Freeze Point for Warts product is intended for the OTC treatment of common warts and plantar warts. May be used with children 4 years of age or older under adult supervision.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

A summary of the technological characteristics for the Skin Clinic Freeze Point for Warts vs Freeze 'n Clear Skin Clinic for Warts/Skin Tags and the Wartie are provided in the following table. Each of these products utilize the same technological characteristics as shown in the table below.

TechnologicalCharacteristicsFreeze 'n Clear SkinClinic forWarts (K211099)Wartie(K2130314)Skin Clinic FreezePoint for Warts(Subject Device)
-Intended UseThe Freeze 'n Clear SkinClinic for Warts andSkin Tags product isintended for the OTCtreatment of commonwarts, plantar warts, andskin tags.Wartie Wart Remover isintended for the over thecounter treatment ofcommon and plantarwarts to be used inadults and children 4years of age and olderThe Skin Clinic FreezePoint for Warts productis intended for the OTCtreatment of commonwarts and plantar warts.May be used withchildren 4 years of ageor older under adultsupervision.
-CryogenMixture of DMEPMixture of DMEPDifluoroethane
TechnologicalCharacteristicsFreeze 'n Clear SkinClinic forWarts (K211099)Wartie(K2130314)Skin Clinic FreezePoint for Warts(Subject Device)
-Materials-canister containingcryogen-Foam Applicators-canister containingcryogen-Metal Tip Applicator-canister containingcryogen-Foam and Metal TipApplicator
-Mode of UseCryogen dispensed intofoam applicator which isthen applied to wartCryogen dispensed ontometal tip which is thenapplied to wartCryogen dispensed ontofoam surrounding themetal tip which is thenapplied to wart
-Mechanism ofactionExtreme cold destroysthe target tissueExtreme cold destroysthe target tissueExtreme cold destroysthe target tissue
-Storage & SafetyConditions-Keep away from fire orflame-Do not smoke whileusing the product-Do not puncture orincinerate canister-Do not expose to heator store at temperaturesabove 120°F.-Store at roomtemperature away fromheat-Keep away from fire orflame-Do not smoke whileusing the product-Do not puncture orincinerate canister-Do not expose to heator store at temperaturesabove 120°F.-Store at roomtemperature away fromheat-Keep away from fire orflame-Do not smoke whileusing the product-Do not puncture orincinerate canister-Do not expose to heator store at temperaturesabove 120°F.-Store at roomtemperature away fromheat
-TreatmentProcedureSpray the cryogen intothe applicator and thenplace it directly onto thewart for a specifiednumber of secondsSpray the cryogen intothe applicator and thenplace it directly onto thewart for a specifiednumber of secondsSpray the cryogen intothe applicator and thenplace it directly onto thewart for a specifiednumber of seconds
TechnologicalCharacteristicsFreeze 'n Clear SkinClinic forWarts (K211099)Wartie(K2130314)Skin Clinic FreezePoint for Warts(Subject Device)
-DisposalEntire unit is disposableafter emptied ofcryogen.Entire unit is disposableafter emptied ofcryogen.Entire unit is disposableafter emptied ofcryogen.
-DefinedOperatorsOTC for consumer useOTC for consumer useOTC for consumer use
-Service / RepairNoneNoneNone

{6}------------------------------------------------

{7}------------------------------------------------

VII. PERFORMANCE DATA

As part of the submission, the product was tested for biocompatibility including cytotoxicity, sensitivity, irritation according to FDA's biocompatibility 2020 guidance and meets the requirements of ISO 10993.

Bench testing compared the temperatures attained by the Wartie and Freeze n Clear Skin Clinic predicate devices to the Skin Clinic Freeze Point product to demonstrate that they were comparable. Additional bench testing using an in vitro method compared all the same products and showed they were comparable in their ability to freeze and destroy target cells.

A human factors usability study was performed in support of the over-the-counter indication for Skin Clinic Freeze Point product to treat common and plantar warts. The human factors usability testing evaluated consumers ability to comprehend and understand the labelling and use of the product.

VIII. CONCLUSIONS

The Skin Clinic Freeze Point product is substantially equivalent to the Freeze n Clear Skin Clinic Wart/Skin Tag product (K211099) and the Wartie wart treatment product (K140314) predicate devices. All are intended for OTC treatment of common and plantar warts. The combination of studies and performance data presented demonstrates the subject device is as safe and effective as the predicate device(s) for the intended use.

§ 878.4350 Cryosurgical unit and accessories.

(a)
Identification —(1)Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold.(2)
Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold.(3)
Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.(b)
Classification. Class II.