(32 days)
AllTest Multi-Drug Urine Test Panel tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana and Fentanyl in human urine at the cutoff concentrations of:
| Drug (Identifier) | Cut-off leve |
|---|---|
| Amphetamine (AMP) | 500 ng/mL |
| Buprenorphine (BUP) | 10 ng/mL |
| Secobarbital (BAR) | 300 ng/mL |
| Oxazepam (BZO) | 300 ng/mL |
| Cocaine (COC) | 150 ng/mL |
| Methamphetamine (MET) | 500 ng/mL |
| Methylenedioxymethamphetamine (MDMA) | 500 ng/mL |
| Morphine (MOP/OPI) | 300 ng/mL |
| Methadone (MTD) | 300 ng/mL |
| Oxycodone (OXY) | 100 ng/mL |
| Phencyclidine (PCP) | 25 ng/mL |
| Nortriptyline (TCA) | 1000 ng/mL |
| Marijuana (THC) | 50 ng/mL |
| Fentanyl(FYL) | 1 ng/mL |
AllTest Multi-Drug Urine Test Panel can be a single drug test panel or used for any combination of the above listed analytes. It is for in vitro diagnostic use only. It is intended for OTC use.
The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.
AllTest Multi-Drug Rapid Urine Test Panel tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, Methamphetamine, Methylenedioxymethamphetamine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana and Fentanyl in human urine at the cutoff concentrations of:
| Drug (Identifier) | Calibrator | Cut-off (ng/mL) |
|---|---|---|
| Amphetamine (AMP) | d-Amphetamine | 500 |
| Buprenorphine (BUP) | Buprenorphine | 10 |
| Secobarbital (BAR) | Secobarbital | 300 |
| Oxazepam (BZO) | Oxazepam | 300 |
| Cocaine (COC) | Benzoylecgonine | 150 |
| Methamphetamine (MET) | d-Methamphetamine | 500 |
| Methylenedioxymethamphetamine (MDMA) | d,l-Methylenedioxymethamphetamine | 500 |
| Morphine (MOP/OPI) | Morphine | 300 |
| Methadone (MTD) | Methadone | 300 |
| Oxycodone (OXY) | Oxycodone | 100 |
| Phencyclidine (PCP) | Phencyclidine | 25 |
| Nortriptyline (TCA) | Nortriptyline | 1000 |
| Marijuana (THC) | 11-nor-Δ9-THC-9 COOH | 50 |
| Fentanyl(FYL) | Fentanyl | 1 |
AllTest Multi-Drug Rapid Urine Test Panel can be a single drug test panel or used for any combination of the above listed analytes. It is for in vitro diagnostic use only.
The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinquish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.
AllTest Multi-Drug Urine Test Panel and AllTest Multi-Drug Rapid Urine Test Panel are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The devices are a panel format. Each test device is sealed with sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.
The provided document describes the acceptance criteria and study that proves the device (AllTest Multi-Drug Urine Test Panel; AllTest Multi-Drug Rapid Urine Test Panel) meets the acceptance criteria.
Note: This device is a qualitative in-vitro diagnostic test, not an AI-based medical device. Therefore, questions related to AI-specific aspects (e.g., number of experts for ground truth, MRMC study, training set, effect size of AI assistance) are not applicable to this submission. The "study" described focuses on analytical performance, method comparison (against a gold standard chemical method), and a lay-person usability study.
1. A table of acceptance criteria and the reported device performance
The document defines performance in terms of precision (reproducibility) and method comparison with LC-MS/MS. The acceptance criteria for these qualitative tests are implicitly demonstrated by the reported agreement with LC-MS/MS results, particularly around the cutoff concentrations. For precision, the goal is consistent detection above the cutoff and consistent non-detection below the cutoff. For method comparison, the aim is high agreement with the confirmatory method, especially for samples near the cutoff.
Table of Performance (Based on "Precision/Reproducibility" and "Method Comparison Study" sections):
| Drug | Cut-off Level (ng/mL) | Precision Study Performance (Examples) | Method Comparison Study Performance (Agreement with LC-MS/MS) |
|---|---|---|---|
| AMP | 500 | At Cutoff (500 ng/mL): 21-24 negative / 26-29 positive results out of 50 tests across 3 lots. At +25% Cutoff (625 ng/mL): 0-1 negative / 49-50 positive results. At -25% Cutoff (375 ng/mL): 49-50 negative / 0-1 positive results. | Overall: Good agreement. Examples of Discordant Results (Individual Operator): - Operator A: 1 sample (494.481 ng/mL, expected positive) was reported positive. - Operator B: 1 sample (491.559 ng/mL, expected positive) was reported positive, 1 sample (619.861 ng/mL, expected negative) was reported negative. - Operator C: 1 sample (491.559 ng/mL, expected positive) was reported positive, 1 sample (619.861 ng/mL, expected negative) was reported negative. (The document states "Accurate result" as + or - based on whether it tested above or below the cutoff, implying the device's output. The discordance is when the device's output ("accurate result") doesn't match the LC-MS/MS interpretation.) |
| BAR | 300 | At Cutoff (300 ng/mL): 21-23 negative / 27-29 positive results out of 50 tests across 3 lots. At +25% Cutoff (375 ng/mL): 0-1 negative / 49-50 positive results. At -25% Cutoff (225 ng/mL): 50 negative / 0 positive results. | Overall: Good agreement. Examples: - Operator A: 1 sample (304.564 ng/mL, expected negative) was reported negative. - Operator B: 1 sample (294.112 ng/mL, expected positive) was reported positive. - Operator C: 1 sample (271.093 ng/mL, expected positive) was reported positive, 1 sample (304.564 ng/mL, expected negative) was reported negative. |
| BUP | 10 | At Cutoff (10 ng/mL): 24-25 negative / 25-26 positive results out of 50 tests across 3 lots. At +25% Cutoff (12.5 ng/mL): 0-1 negative / 49-50 positive results. At -25% Cutoff (7.5 ng/mL): 49-50 negative / 0-1 positive results. | Overall: Good agreement. Examples: - Operator A: 1 sample (11.971 ng/mL, expected negative) was reported negative, 1 sample (9.635 ng/mL, expected positive) was reported positive. - Operator B: 1 sample (10.385 ng/mL, expected negative) was reported negative. - Operator C: 1 sample (9.149 ng/mL, expected positive) was reported positive. |
| ... | ... | (Similar patterns observed for all other drugs listed: BZO, COC, MDMA, MET, MOP/OPI, MTD, OXY, PCP, TCA, THC, FYL) | (Similar patterns observed for all other drugs listed, with specific discordant samples detailed in the raw tables. The "Accurate Result" column appears to denote the manual interpretation of the device's result, implying it was designed to align with LC/MS/MS around the cutoff. The discordance is where the device's result does not align with the LC-MS/MS value relative to the cutoff.) |
Key Acceptance Criteria (Implicit from Study Design):
- Precision/Reproducibility: The device should consistently return positive results for samples significantly above the cutoff concentration and consistently negative results for samples significantly below the cutoff. Near the cutoff, a mix of positive and negative results is expected due to inherent assay variability.
- Analytical Specificity/Interference: The device should accurately detect the target drug without significant cross-reactivity from other structurally similar compounds or interference from common endogenous/exogenous substances.
- Method Comparison: A high level of agreement between the device's qualitative results and the quantitative LC-MS/MS confirmatory method, especially for samples near the cutoff.
- Lay Person Usability: Intended lay users should be able to understand and follow instructions and obtain correct results.
2. Sample sizes used for the test set and the data provenance
-
Analytical Performance (Precision/Reproducibility):
- Sample Size: For each drug, 8 concentrations (+100%, +75%, +50%, +25%, cutoff, -25%, -50%, -75%, -100% cutoff). Each concentration was tested two runs per day for 25 days using three lots of test panels.
- This means (8 concentrations * 2 runs/day * 25 days * 3 lots) = 1200 tests per drug.
- Data Provenance: Samples were prepared by spiking target drugs into drug-free urine samples. The concentrations were confirmed by LC-MS/MS. This suggests controlled laboratory conditions. The country of origin is not specified but the submitter is Hangzhou AllTest Biotech Co.,Ltd (China) and contact person is in Maryland (USA), so the testing location could be either. Retrospective, as samples were prepared.
- Sample Size: For each drug, 8 concentrations (+100%, +75%, +50%, +25%, cutoff, -25%, -50%, -75%, -100% cutoff). Each concentration was tested two runs per day for 25 days using three lots of test panels.
-
Method Comparison Study:
- Sample Size: For each drug, 80 unaltered urine clinical samples (40 negative and 40 positive).
- Data Provenance: "Unaltered urine clinical samples." The source of these clinical samples (e.g., country of origin, prospective or retrospective collection) is not explicitly stated, but "clinical samples" generally implies collection from human subjects. As they were "unaltered," they were likely retrospective, though collection parameters are not detailed.
-
Lay Person Study:
- Sample Size: 140 lay persons.
- Sample Types: Urine samples created by spiking drug(s) into drug-free pooled urine specimens at various concentrations (-100%, +/-75%, +/-50%, +/-25% of the cutoff). Each participant tested 1 blind-labeled sample. This means 20 samples were used for each of the 7 concentrations (+/-100%, +/-75%, +/-50%, +/-25% of the cutoff for each drug).
- Data Provenance: Controlled laboratory setting using spiked, blind-labeled samples. The study was performed at "three intended user sites," implying locations where lay persons would typically use such a device. Country not specified. Retrospective setup using prepared samples.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- This device is not an AI-based medical device. Ground truth for the analytical and method comparison studies was established by LC-MS/MS (Liquid Chromatography-Mass Spectrometry/Mass Spectrometry), which is considered the gold standard for quantitative drug concentration measurement in urine. LC-MS/MS does not rely on expert readers for interpretation in the same way imaging or pathology might.
- For the lay person study, the ground truth was the known concentration of the spiked samples.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Analytical Study (Precision) & Method Comparison: The ground truth was established by LC-MS/MS. The device's result was then compared to this LC-MS/MS reference. There was no "adjudication" in the sense of multiple human readers resolving discrepancies; the LC-MS/MS result served as the definitive reference.
- Lay Person Study: The ground truth was the known concentration of the spiked samples. The lay person's result was compared directly to this known concentration.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study was not done. This device is an in-vitro diagnostic test, not an AI-assisted diagnostic tool. Its performance is evaluated as a standalone test, and for lay-user interpretation against a ground truth.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Yes, the "Precision/Reproducibility" and "Method Comparison Study" sections represent the standalone performance of the device (which is an immunochromatographic assay, not an algorithm) without human interpretation variability, as results were compared directly to LC-MS/MS.
- The "Lay Person Study" then evaluated human-in-the-loop performance (i.e., how lay users interpret the device's results).
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The primary ground truth for the analytical and method comparison studies was LC-MS/MS quantitative analysis, which is a highly accurate chemical method for measuring drug concentrations.
- For the lay person study, the ground truth was the known, spiked concentrations of the drug in the urine samples.
8. The sample size for the training set
- This document describes a 510(k) premarket notification for an in-vitro diagnostic device, not an AI/ML device. Therefore, there is no "training set" in the context of machine learning model development.
9. How the ground truth for the training set was established
- As stated above, this is not an AI/ML device, so there is no "training set." The performance studies rely on LC-MS/MS as the reference method and known spiked concentrations for establishing ground truth for evaluation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Hangzhou Alltest Biotech Co.,Ltd % Jenny Xia, Director LSI International Inc 504E Diamond Ave., Suite H Gaithersburg, Maryland 20877
Re: K242540
Trade/Device Name: AllTest Multi-Drug Urine Test Panel; AllTest Multi-Drug Rapid Urine Test Panel Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Product Code: NFT, PTH, NGL, NFV, NFY, PTG, NGG, NGM, OAW, NFW Dated: August 23, 2024 Received: August 26, 2024
Dear Jenny Xia:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely, Joseph A. Digitally signed by Kotarek -S Date: 2024.09.27
Kotarek -S Date: 2024.09.20 Joseph Kotarek Branch Chief for Toxicology Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
Submission Number (if known)
Device Name
AllTest Multi-Druq Urine Test Panel; AllTest Multi-Drug Rapid Urine Test Panel
Indications for Use (Describe)
AllTest Multi-Drug Urine Test Panel tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana and Fentanyl in human urine at the cutoff concentrations of:
| Drug (Identifier) | Cut-off leve |
|---|---|
| Amphetamine (AMP) | 500 ng/mL |
| Buprenorphine (BUP) | 10 ng/mL |
| Secobarbital (BAR) | 300 ng/mL |
| Oxazepam (BZO) | 300 ng/mL |
| Cocaine (COC) | 150 ng/mL |
| Methamphetamine (MET) | 500 ng/mL |
| Methylenedioxymethamphetamine (MDMA) | 500 ng/mL |
| Morphine (MOP/OPI) | 300 ng/mL |
| Methadone (MTD) | 300 ng/mL |
| Oxycodone (OXY) | 100 ng/mL |
| Phencyclidine (PCP) | 25 ng/mL |
| Nortriptyline (TCA) | 1000 ng/mL |
| Marijuana (THC) | 50 ng/mL |
| Fentanyl(FYL) | 1 ng/mL |
| AllTost Multi-Drug Urine Test Panel can be a single drug tes |
rug Urine Test Panel can be a single druq test panel or used for any combination of the above listed analytes. It is for in vitro diagnostic use only. It is intended for OTC use.
The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.
AllTest Multi-Drug Rapid Urine Test Panel tests are competitive binding, lateral flow
immunochromatographic assays for qualitative and simultaneous detection of Amphetamine,
Buprenorphine, Secobarbital, Oxazepam, Cocaine, Methamphetamine,
Methylenedioxymethamphetamine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana and Fentanyl in human urine at the cutoff concentrations of:
| Drug (Identifier) | Calibrator | Cut-off (ng/mL) |
|---|---|---|
| Amphetamine (AMP) | d-Amphetamine | 500 |
| Buprenorphine (BUP) | Buprenorphine | 10 |
| Secobarbital (BAR) | Secobarbital | 300 |
| Oxazepam (BZO) | Oxazepam | 300 |
{3}------------------------------------------------
| Cocaine (COC) | Benzoylecgonine | 150 |
|---|---|---|
| Methamphetamine (MET) | d-Methamphetamine | 500 |
| Methylenedioxymethamphetamine (MDMA) | d,l-Methylenedioxymethamphetamine | 500 |
| Morphine (MOP/OPI) | Morphine | 300 |
| Methadone (MTD) | Methadone | 300 |
| Oxycodone (OXY) | Oxycodone | 100 |
| Phencyclidine (PCP) | Phencyclidine | 25 |
| Nortriptyline (TCA) | Nortriptyline | 1000 |
| Marijuana (THC) | 11-nor-Δ9-THC-9 COOH | 50 |
| Fentanyl(FYL) | Fentanyl | 1 |
AllTest Multi-Drug Rapid Urine Test Panel can be a single drug test panel or used for any combination of the above listed analytes. It is for in vitro diagnostic use only.
The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinquish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS or LC/MS is the recommended confirmatory method.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) SUMMARY K242540
- August 23, 2024 Date: 1. Hangzhou AllTest Biotech Co., Ltd. 2. Submitter: Plant Bldg. 3, 4, 5, No. 550 Yinhai Street, Baiyang Street, Hangzhou ETDZ, Jianggan District 3. Contact person: Jenny Xia LSI International Inc. 504 East Diamond Ave., Suite H Gaithersburg, MD 20877 Telephone: 301-525-6856 Email: jxia@lsi-consulting.org
-
- Device Name: AllTest Multi-Drug Urine Test Panel AllTest Multi-Drug Rapid Urine Test Panel Classification: Class II ਨ.
| Product CodeTarget Drug | Regulation Section | Panel |
|---|---|---|
| NFTAmphetamine (AMP) | 862.3100, Amphetamine Test System | Toxicology |
| PTHSecobarbital (BAR) | 862.3150, Barbiturate Test System | Toxicology |
| NGLBuprenorphine (BUP)Fentanyl (FYL)Morphine (MOP/OPI)Oxycodone (OXY) | 862.3650, Opiate Test System | Toxicology |
| NFVOxazepam (BZO) | 862.3170, Benzodiazepine Test System | Toxicology |
| NFYCocaine (COC) | 862.3250, Cocaine and cocainemetabolite test system | Toxicology |
| PTGMethadone (MTD) | 862.3620, Methadone Test System | Toxicology |
| NGGMethylenedioxymethamphetamine(MDMA)Methamphetamine (MET) | 862.3610,Methamphetamine Test System | Toxicology |
| NGMPhencyclidine (PCP) | Unclassified | Toxicology |
| QAWNortriptyline (TCA) | 862.3910 Tricyclic antidepressant drugstest system | Toxicology |
{5}------------------------------------------------
| NFWCannabinoids (THC) | 862.3870, Cannabinoids Test System | Toxicology |
|---|---|---|
| --------------------------- | ------------------------------------ | ------------ |
6. Predicate Devices:
AllTest Multi-Drug Urine Test Cup (K241428)
7. Intended Use
AllTest Multi-Drug Urine Test Panel tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana and Fentanyl in human urine at the cutoff concentrations of:
| Drug (Identifier) | Cut-off level |
|---|---|
| Amphetamine (AMP) | 500 ng/mL |
| Buprenorphine (BUP) | 10 ng/mL |
| Secobarbital (BAR) | 300 ng/mL |
| Oxazepam (BZO) | 300 ng/mL |
| Cocaine (COC) | 150 ng/mL |
| Methamphetamine (MET) | 500 ng/mL |
| Methylenedioxymethamphetamine (MDMA) | 500 ng/mL |
| Morphine (MOP/OPI) | 300 ng/mL |
| Methadone (MTD) | 300 ng/mL |
| Oxycodone (OXY) | 100 ng/mL |
| Phencyclidine (PCP) | 25 ng/mL |
| Nortriptyline (TCA) | 1000 ng/mL |
| Marijuana (THC) | 50 ng/mL |
| Fentanyl(FYL) | 1 ng/mL |
AllTest Multi-Drug Urine Test Panel can be a single drug test panel or used for any combination of the above listed analytes. It is for in vitro diagnostic use only. It is intended for OTC use.
The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS is the recommended confirmatory method.
{6}------------------------------------------------
AllTest Multi-Drug Rapid Urine Test Panel tests are competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Buprenorphine, Secobarbital, Oxazepam, Cocaine, Methamphetamine,
Methylenedioxymethamphetamine, Morphine, Methadone, Oxycodone, Phencyclidine, Nortriptyline, Marijuana and Fentanyl in human urine at the cutoff concentrations of:
| Drug (Identifier) | Calibrator | Cut-off(ng/mL) |
|---|---|---|
| Amphetamine (AMP) | d-Amphetamine | 500 |
| Buprenorphine (BUP) | Buprenorphine | 10 |
| Secobarbital (BAR) | Secobarbital | 300 |
| Oxazepam (BZO) | Oxazepam | 300 |
| Cocaine (COC) | Benzoylecgonine | 150 |
| Methamphetamine(MET) | d-Methamphetamine | 500 |
| Methylenedioxymethamphetamine (MDMA) | d,l-Methylenedioxymethamphetamine | 500 |
| Morphine (MOP/OPI) | Morphine | 300 |
| Methadone (MTD) | Methadone | 300 |
| Oxycodone (OXY) | Oxycodone | 100 |
| Phencyclidine (PCP) | Phencyclidine | 25 |
| Nortriptyline (TCA) | Nortriptyline | 1000 |
| Marijuana (THC) | 11-nor-Δ9-THC-9 COOH | 50 |
| Fentanyl (FYL) | Fentanyl | 1 |
AllTest Multi-Drug Rapid Urine Test Panel can be a single drug test panel or used for any combination of the above listed analytes. It is for in vitro diagnostic use only.
The tests may yield positive results for the prescription drugs when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result.
The tests provide only preliminary results. To obtain a confirmed analytical result, a more specific alternate chemical method must be used. GC/MS is the recommended confirmatory method.
Device Description 8.
AllTest Multi-Drug Urine Test Panel and AllTest Multi-Drug Rapid Urine Test Panel are immunochromatographic assays that use a lateral flow system for the qualitative detection of single or multiple drugs in human urine.
The devices are a panel format. Each test device is sealed with sachets of desiccant in an aluminum pouch. The device is in a ready-to-use format and no longer requires assembly before use.
{7}------------------------------------------------
| Similarities | ||
|---|---|---|
| Item | Candidate Device(K242540) | Predicate(K241428) |
| Intended use | Qualitative detection of drugs of abuse in urine. For prescription use or over-the-counter use | Same |
| Methodology | Competitive binding, lateral flow immunochromatographic assay based on antigen-antibody reaction | Same |
| Type of Test | Qualitative | Same |
| Specimen Type | Human urine | Same |
| Target Drugand Cut OffValues | Target Drugs | Cutoff (ng/mL) Same |
| Amphetamine(AMP) | 500 | |
| Secobarbital (BAR) | 300 | |
| Buprenorphine (BUP) | 10 | |
| Oxazepam (BZO) | 300 | |
| Cocaine (COC) | 150 | |
| Methylenedioxymethamphetamine (MDMA) | 500 | |
| Methamphetamine (MET) | 500 | |
| Morphine (MOP/OPI) | 300 | |
| Methadone (MTD) | 300 | |
| Oxycodone (OXY) | 100 | |
| Phencyclidine (PCP) | 25 | |
| Nortriptyline (TCA) | 1000 | |
| Cannabinoids (THC) | 50 | |
| Fentanyl (FYL) | 1 | |
| Differences | ||
| Configuration | Test Panel | Test Cup |
Substantial Equivalence Information 0
10. Standard/Guidance Document Reference (if applicable)
None referenced.
11. Test Principle
AllTest Multi-Drug Urine Test Panel or AllTest Multi-Drug Rapid Urine Test Panel is a competitive immunoassay that is used to screen for the presence of various drugs and drug metabolites in urine. It is chromatographic absorbent device in which, drugs within a urine sample, competitively combined to a limited number of drug monoclonal antibody (mouse) conjugate binding sites.
When the test is activated, the urine is absorbed into each test strip by capillary action, mixes with the respective drug monoclonal antibody conjugate, and flows across a pre-coated membrane. When a drug within the urine sample is below the detection level of the test, respective drug monoclonal antibody conjugate binds to the respective drug-protein conjugate immobilized in the Test Region (T) of the test strip. This produces a colored Test line in the Test
{8}------------------------------------------------
Region (T) of the strip, which, regardless of its intensity, indicates a negative test result.
When sample drug levels are at or above the detection level of the tree drug in the sample binds to the respective drug monoclonal antibody conjugate, preventing the respective drug monoclonal antibody conjugate from binding to the respective drug-protein conjugate immobilized in the Test Region (T) of the device. This prevents the development of a distinct colored band in the test region, indicating a preliminary positive result.
To serve as a procedure control, a colored line will appear at the Control Region (C) of each strip, if the test has been performed properly.
12. Performance Characteristics
A. Analytical performance
a. Precision/Reproducibility:
Precision studies were carried out for samples with concentrations of +100% cutoff, +50% cutoff, +25% cutoff, cutoff, -25% cutoff, -50% cutoff, -75% cut off and -100% cutoff. Samples were prepared by spiking target drugs in drug-free urine samples. Each drug concentration was confirmed by LC-MS/MS. For each concentration, tests were performed two runs per day for 25 days using three lots of test panels. The results obtained are summarized in the following tables:
| Drug | LotNumber | +100%cutoff | +75%cutoff | +50%cutoff | +25%cutoff | Cutoff | -25%cutoff | -50%cutoff | -75%cutoff | -100%cut-off |
|---|---|---|---|---|---|---|---|---|---|---|
| AMP500 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 22-/28+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 21-/29+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| BAR300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 21-/29+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| BUP 10 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 50+/0- | 25-/25+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 50+/0- | 24-/26+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| BZO300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| MDMA500 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 22-/28+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 25-/25+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 22-/28+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| MET500 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 22-/28+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 25-/25+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
{9}------------------------------------------------
| MOP/ | Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 22-/28+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
|---|---|---|---|---|---|---|---|---|---|---|
| OPI300 | Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| MTD300 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 22-/28+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 25-/25+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| OXY100 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 21-/29+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| PCP 25 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 25-/25+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| TCA1000 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 22-/28+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 24-/26+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| THC50 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 25-/25+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 1-/49+ | 25-/25+ | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| COC150 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 50+/0- | 28+/22- | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 50+/0- | 29+/21- | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 49+/1- | 26+/24- | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| FYL 1 | Lot 1 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 22-/28+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ |
| Lot 2 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 20-/30+ | 49-/1+ | 50-/0+ | 50-/0+ | 50-/0+ | |
| Lot 3 | 0-/50+ | 0-/50+ | 0-/50+ | 0-/50+ | 23-/27+ | 48-/2+ | 50-/0+ | 50-/0+ | 50-/0+ |
b. Linearity/assay reportable range:
Not applicable. This device is intended for qualitative use only.
c. Stability:
The device is stable at 2-30°C for 24 months based on real time stability study.
d. Analytical specificity/Interference:
To test the specificity, drug metabolites and other components that are likely to cross-react in urine samples were spiked into drug-free urine samples were tested using three lots of the device.
Percent cross-reactivity, provided in the below table, was calculated as the cutoff concentration divided by the concentration of analyte tested that yielded a positive result, multiplied by 100.
| Drug/Cutoff | Compound | Minimum concentration required to obtain a positive result | % Cross-Reactivity |
|---|---|---|---|
| (ng/mL) | |||
| Hydroxyamphetamine | 5000 | 10% | |
| (+/-)-Methylenedioxyamphetamine(MDA) | 25 | 2000% | |
| D,L-Amphetamine | 500 | 100% | |
| D-Amphetamine | 500 | 100% | |
| Diethylstilbestrol | >100000 | <0.5% | |
| L-Amphetamine | 500 | 100% | |
| Phentermine | 500 | 100% | |
| β-Phenylethylamine | >100000 | <0.5% | |
| Tyramine | >100000 | <0.5% | |
| p-Hydroxynorephedrine | >100000 | <0.5% | |
| D,L-Norephedrine | >100000 | <0.5% | |
| AMP 500 | p-Hydroxyamphetamine | 5000 | 10% |
| D-Methamphetamine | >100000 | <0.5% | |
| L-Methamphetamine | >100000 | <0.5% | |
| Ephedrine hydrochloride | >100000 | <0.5% | |
| (+/-)3,4-Methylenedioxymethamphetamine(MDMA) | 100000 | 0.5% | |
| Phenylpropanolamine | >100000 | <0.5% | |
| Benzphetamine | >100000 | <0.5% | |
| L-Ephedrine | >100000 | <0.5% | |
| L-Epinephrine | >100000 | <0.5% | |
| D,L-Epinephrine | >100000 | <0.5% | |
| (+/-)3,4-Methylenedioxyethylamphetamine(MDEA) | 100000 | 0.5% | |
| Alphenal | 300 | 100% | |
| Amobarbital | 300 | 100% | |
| BAR 300 | Aprobarbital | 300 | 100% |
| Butabarbital | 500 | 60% | |
| Butethal | 200 | 150% | |
| Cyclopentobarbital | 500 | 60% | |
| Pentobarbital | 200 | 150% | |
| Phenobarbital | 200 | 150% | |
| Secobarbital | 300 | 100% | |
| Butalbital | 2000 | 15% | |
| Barbital | 300 | 100% | |
| Buprenorphine | 10 | 100% | |
| Buprenorphine-3-D-Glucuronide | 20 | 50% | |
| Norbuprenorphine | 20 | 50% | |
| Norbuprenorphine-3-D- | 50% | ||
| Glucuronide | 20 | ||
| BUP 10 | Morphine | >100000 | <0.01% |
| Oxymorphone | >100000 | <0.01% | |
| Hydromorphone | >100000 | <0.01% | |
| Codeine | >100000 | <0.01% | |
| Nalorphine | >100000 | <0.01% | |
| a-Hydroxyalprazolam | 10000 | 3% | |
| Alprazolam | 200 | 150% | |
| Bromazepam | 1000 | 30% | |
| Chlordiazepoxide | 300 | 100% | |
| Clobazam | 200 | 150% | |
| Clonazepam | 200 | 150% | |
| Clorazepate Dipotassium | 300 | 100% | |
| Desalkylflurazepam | >100000 | <0.3% | |
| Diazepam | 300 | 100% | |
| BZO 300 | Estazolam | 100 | 300% |
| Flunitrazepam | 300 | 100% | |
| D,L-Lorazepam | 1000 | 30% | |
| Midazolam | 1000 | 30% | |
| Nitrazepam | 200 | 150% | |
| Norchlordiazepoxide | 200 | 150% | |
| Nordiazepam | 200 | 150% | |
| Oxazepam | 300 | 100% | |
| R.S-Lorazepam glucuronide | 1000 | 30% | |
| Temazepam | 100 | 300% | |
| Triazolam | 300 | 100% | |
| Demoxepam | 2000 | 15% | |
| Flurazepam | 300 | 100% | |
| Delorazepam | 2000 | 15% | |
| Lormetazepam | 100 | 300% | |
| COC 150 | Benzoylecogonine | 150 | 100% |
| Cocaethylene | 250 | 60% | |
| Cocaine hydrochloride | 125 | 120% | |
| Ecgonine | >100000 | <0.2% | |
| Norcocaine | >100000 | <0.2% | |
| Ecgonine methyl ester | >100000 | <0.2% | |
| (+/-)3,4-Methylenedioxy-n-ethylamphetamine(MDEA) | 250 | 200% | |
| (+/-)-Methylenedioxyamphetamine(MDA) | 1000 | 50% | |
| (±)-MDMA | 500 | 100% | |
| L-Methamphetamine | >100000 | <0.5% | |
| MDMA 500 | d-methamphetamine | >100000 | <0.5% |
| d-amphetamine | >100000 | <0.5% | |
| l-amphetamine | >100000 | <0.5% | |
| d,l-Amphetamine | >100000 | <0.5% | |
| d,l-Methamphetamine | >100000 | <0.5% | |
| Phentermine | >100000 | <0.5% | |
| Phenylephrine | >100000 | <0.5% | |
| Ephedrine | >100000 | <0.5% | |
| Pseudoephedrine | >100000 | <0.5% | |
| MET 500 | (+/-)3,4-Methylenedioxy-n-ethylamphetamine(MDEA) | 6000 | 8.3% |
| (±)-MDMA | 1000 | 50% | |
| D-Methamphetamine | 500 | 100% | |
| L-Methamphetamine | 10000 | 5% | |
| Fenfluramine | 50000 | 1% | |
| p-Hydroxymethamphetamine | 250 | 200% | |
| D,L-Methamphetamine | 250 | 200% | |
| β-Phenylethylamine | >100000 | <0.5% | |
| Mephetermine | 25000 | 2% | |
| Methoxyphenamine hydrochloride | 75000 | 0.7% | |
| L-Amphetamine | 100000 | 0.5% | |
| D-Amphetamine | 100000 | 0.5% | |
| D,L-Amphetamine | 250 | 200% | |
| Chloroquine | 25000 | 2.0% | |
| Ephedrine hydrochloride | >100000 | <0.5% | |
| (+/-)3,4-Methylenedioxyamphetamine(MDA) | >100000 | <0.5% | |
| Trimethobenzamide | >100000 | <0.5% | |
| l-phenylephrine | >100000 | <0.5% | |
| (1R,2S)-(-)-Ephedrine | >100000 | <0.5% | |
| Procaine hydrochloride | >100000 | <0.5% | |
| Phentermine | >100000 | <0.5% | |
| Pseudoephedrine | >100000 | <0.5% | |
| 6-acetylmorphine | 300 | 100% | |
| Codeine | 250 | 120% | |
| Dihydrocodeine | 5000 | 6% | |
| EthylMorphine | 200 | 150% | |
| Heroin | 500 | 60% | |
| Hydrocodone | 1000 | 30% | |
| MOP/OPI300 | Hydromorphone | 10000 | 3% |
| Levorphanol tartrate | 1500 | 20% | |
| Morphine | 300 | 100% | |
| Nalorphine hydrochloride | 300 | 100% | |
| Thebaine | 20000 | 1.5% | |
| s-Monoacetylmorphine | 300 | 100% | |
| Morphine-3-β-d-glucuronide | 1000 | 30% | |
| 6-Monoacetylmorphine (6-MAM) | 300 | 100% | |
| Codeine-6-β-D-glucuronide | 25000 | 1.2% | |
| Morphine-6-β-D-glucuronide | 1000 | 30% | |
| 6-Acetylcodeine | 25000 | 1.2% | |
| Normorphine | 50000 | 0.6% | |
| Oxycodone | >100000 | <0.3% | |
| Oxymorphone | >100000 | <0.3% | |
| Norcodeine | 25000 | 1.2% | |
| Procaine hydrochloride | >100000 | <0.3% | |
| Norpropoxyphene | >100000 | <0.3% | |
| MTD 300 | (±)-Methadone | 300 | 100% |
| EDDP | >100000 | <0.3% | |
| EMDP | >100000 | <0.3% | |
| LAAM | >100000 | <0.3% | |
| Alpha Methadol | >100000 | <0.3% | |
| Doxylamine | >100000 | <0.3% | |
| Disopyramide | 6000 | 5.0% | |
| Esomeprazole | 10000 | 3.0% | |
| Pheniramine | 100000 | 0.3% | |
| OXY 100 | Ethyl Oxycodone | 100 | 100% |
| Hydrocodone | 50000 | 0.2% | |
| Hydromorphone | 25000 | 0.4% | |
| Levorphanol tartrate | 50000 | 0.2% | |
| Naloxone hydrochloride | 50000 | 0.2% | |
| Naltrexone hydrochloride | >100000 | <0.1% | |
| Oxycodone | 100 | 100% | |
| Oxymorphone | 100 | 100% | |
| Oxymorphone-3β-D-glucuronide | >100000 | <0.1% | |
| Noroxycodone | 1000 | 10% | |
| Noroxymorphone | 25000 | 0.4% | |
| Dihydrocodeine | 12500 | 0.8% | |
| Codeine | 100000 | 0.1% | |
| Morphine | >100000 | <0.1% | |
| Acetylmorphine | >100000 | <0.1% | |
| Buprenorphine | >100000 | <0.1% | |
| Ethylmorphine | >100000 | <0.1% | |
| Thebaine | >100000 | <0.1% | |
| 6-acetylmorphine | >100000 | <0.1% | |
| PCP 25 | PCP (Phencyclidine) | 25 | 100% |
| 4-Hydroxyphencyclidine | 15 | 167% | |
| Pheniramine | 50000 | 0.05% | |
| TCA 1000 | Amitriptyline | 1500 | 66.7% |
| Chlorpheniramine | >100000 | <1% | |
| Clomipramine | 1000 | 100% | |
| Cyclobenzaprine Hydrochloride | >100000 | <1% | |
| Desipramine | 300 | 333.3% | |
| Doxepine | 10000 | 10% | |
| Duloxetine | >100000 | <1% | |
| Imipramine | 500 | 200% | |
| Norclomipramine | 250 | 400% | |
| Nordoxepine | >100000 | <1% | |
| Nortriptyline hydrochloride | 1000 | 100% | |
| Promazine | 500 | 200% | |
| Trimipramine | 1500 | 66.7% | |
| Maprotiline | >100000 | <1% | |
| Promethazine hydrochloride | 50000 | 2% | |
| THC 50 | 11-nor-Δ8-THC-9-COOH | 20000 | 0.25% |
| (-)-11-nor-9-carboxy-Δ9-THC | 50 | 100% | |
| (±)-11-nor-9-Carboxy-Δ9-THC | 50 | 100% | |
| 11-nor-Δ9-THC -carboxyglucuronide | 50 | 100% | |
| 11-hydroxy-Δ9-Tetrahydrocannabinol | 5000 | 1.0% | |
| Δ8- Tetrahydrocannabinol | 100000 | 0.05% | |
| Δ9- Tetrahydrocannabinol | 100000 | 0.05% | |
| Cannabinol | >100000 | <0.05% | |
| Cannabidiol | >100000 | <0.05% | |
| FYL 1 | Fentanyl | 1 | 100% |
| Acetyl fentanyl | 1 | 100% | |
| Acrylfentanyl | 1 | 100% | |
| ω-1-Hydroxyfentanyl | 20000 | 0.005% | |
| Isobutyryl fentanyl | 1 | 100% | |
| Ocfentanil | 2.5 | 40% | |
| Butyryl fentanyl | 2.5 | 40% | |
| Furanyl fentanyl | 5 | 20% | |
| Valeryl fentanyl | 10 | 10% | |
| (±) β-hydroxythiofentanyl | 2 | 50% | |
| 4-Fluoro-isobutyrylfentanyl | 50 | 2% | |
| Para-fluorobutyryl fentanyl | 4 | 25% | |
| Para-fluoro fentanyl | 3 | 33.3% | |
| (±)-3-cis-methyl fentanyl | 50 | 2% | |
| Carfentanil | 2 | 50% | |
| Sufentanil | 10 | 10% | |
| Alfentanil | 5000 | 0.02% | |
| Despropionyl fentanyl (4-ANPP) | 2500 | 0.04% | |
| Cyclopropylfentanyl | 3 | 33.3% | |
| Remifentanil | >100000 | <0.001% | |
| Norfentanyl | >100000 | <0.001% | |
| Acetyl norfentanyl | >100000 | <0.001% | |
| Norcarfentanil | >100000 | <0.001% | |
| 6-Acetyl morphine | >100000 | <0.001% | |
| Amphetamine | >100000 | <0.001% | |
| Buprenorphine | >100000 | <0.001% | |
| Buprenorphineglucuronide | >100000 | <0.001% | |
| Codeine | >100000 | <0.001% | |
| Dextromethorphan | >100000 | <0.001% | |
| Dihydrocodeine | >100000 | <0.001% | |
| EDDP | >100000 | <0.001% | |
| EMDP | >100000 | <0.001% | |
| Fluoxetine | >100000 | <0.001% | |
| Heroin | >100000 | <0.001% | |
| Hydrocodone | >100000 | <0.001% | |
| Hydromorphone | >100000 | <0.001% | |
| Ketamine | >100000 | <0.001% | |
| Levorphanol | >100000 | <0.001% | |
| Meperidine | >100000 | <0.001% | |
| Methadone | >100000 | <0.001% | |
| Morphine | >100000 | <0.001% | |
| Morphine-3-glucuronide | >100000 | <0.001% | |
| Naloxone | >100000 | <0.001% | |
| Naltrexone | >100000 | <0.001% | |
| Norbuprenorphine | >100000 | <0.001% | |
| Norcodeine | >100000 | <0.001% | |
| Norketamine | >100000 | <0.001% | |
| Normeperidine | >100000 | <0.001% | |
| Normorphine | >100000 | <0.001% | |
| Noroxycodone | >100000 | <0.001% | |
| Oxycodone | >100000 | <0.001% | |
| Oxymorphone | >100000 | <0.001% | |
| Pentazocine (Talwin) | >100000 | <0.001% | |
| Pipamperone | >100000 | <0.001% | |
| Risperidone | >100000 | <0.001% | |
| Tapentadol | >100000 | <0.001% | |
| Thioridazine | >100000 | <0.001% | |
| Tilidine | >100000 | <0.001% | |
| Tramadol | >100000 | <0.001% | |
| Tramadol-O-Desmethyl | >100000 | <0.001% | |
| Tramadol-N-Desmethyl | >100000 | <0.001% | |
| Isotonitaze | >100000 | <0.001% |
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
{17}------------------------------------------------
To evaluate potential interference, non-structurally related compounds were added to drug-free urine and to urine samples containing the target drugs at 50% below and 50% above each corresponding cutoff.
Compounds that show no interference at a concentration of 100μ.g/mL are summarized in the following table.
| 3-Hydroxytyramine | Digoxin | Nitroglycerin |
|---|---|---|
| Acetaminophen | Diphenhydramine HCl | Nordoxepin |
{18}------------------------------------------------
| Acetone(1000mg/dL) | Disopyramide(except MTD test) | Norethindrone |
|---|---|---|
| Acetophenetidin | DL-Tryptophan | Norfentanyl |
| Acetylsalicylic acid | DL-Tyrosine | Noscapine |
| Acyclovir | Dopamine HCl | O-Hydroxyhippuric acid |
| Albumin(100mg/dL) | Doxepin(except TCA test) | Octopamine |
| Albuterol sulfate(Proair HFA) | Doxylamine | Olanzapine |
| Alpha Methadol | Duloxetine | Omeprazole |
| Aminophylline | Ecgonine methyl ester | Oxalic acid (100 mg/dL) |
| Aminopyrine | EMDP | Oxolinic acid |
| Amoxicillin | Ephedrine | Oxymetazoline |
| Ampicillin | Erythromycin | Paliperidone |
| Apomorphine | Esomeprazole | Papaverine |
| Aripiprazole | Magnesium(except MTD test) | Penicillin-G |
| Ascorbic acid | Ethanol(1%) | PenicillinV Potassium |
| Aspartame | Fenoprofen | Perphenazine |
| Aspirin | Fluoxetine Hydrochloride | Phenacetin |
| Atomoxetine | Fluphenazine | Phenelzine |
| Atorvastatin Calcium | Furosemide | Phenethylamine |
| Atropine | Gabapentin | Phenylpropanolamine |
| Azithromycin | Galactose (10mg/dL) | Prednisone |
| Benzilic acid | Gamma Globulin (500mg/dL) | Pregablin |
| Benzocaine | Gatifloxacin | Procaine |
| Benzoic acid | Gentisic acid | Promethazine(except TCA test) |
| Benzphetamine | Glucose (3000mg/dL) | Propoxyphene |
| Bilirubin | Hemoglobin | Propranolol |
| Boric Acid (1%) | Hydralazine | Pseudoephedrine |
| Bupropion | Hydrochlorothiazide | Quetiapine |
| Caffeine | Hydrocortisone | Quinine |
| Cannabidiol | Hydroxytyramine | Ranitidine |
| Captopril | Ibuprofen | Riboflavin (10mg/dL) |
| Carbamazepine | Isoproterenol | Rifampicin |
| Carfentanil(except FYL test) | Isoxsuprine | Risperidone |
| Cefradine | Ketamine | Salicylic acid |
| Cephalexin | Ketoprofen | Serotonin (5-Hydroxytyramine) |
| Chloralhydrate | L-Ephedrine | Sertraline |
| Chloramphenicol | L-Epinephrine | Sildenafil Citrate |
| Chloroquine(except MET test) | L-phenylephrine | Simvastatin |
| Chlorothiazide | LAAM HCl | Sulfamethazine |
| Chlorpheniramine | Labetalol | Sulindac |
| Chlorpromazine | Levonorgestrel | Telmisartan |
| Cholesterol | Levothyroxine Sodium | Tetrahydrocortisone 3-(β-Dglucuronide) |
| Ciprofloxacin Hydrochloride | Lidocaine Hydrochloride | Tetrahydrocortisone, 3-acetate |
| Citalopram | Lisinopril | Tetrahydrozoline |
| Clarithromycin | Loperamide | Theophylline |
| Clonidine | Loratadine | Thiamine |
| Clozapine | Magnesium | Thioridazine |
| Conjugated Estrogens | Maprotiline | Tramadol Hydrochloride |
| Cortisone | Meperidine | Triamterene |
| Cotinine | Meprobamate | Trifluoperazine |
| Creatinine | Methapyrilene | Trimethobenzamide |
| Cyclobenzaprine | Methaqualone | Trimethoprim |
| D-Pseudoephedrine | Methoxyphenamine (exceptAMP/MET test) | Tyramine |
| D,L-Epinephrine | Metoprolol Tartrate | Urea (2000mg/dL) |
| D,L-Isoproterenol | Metronidazole (300ug/ml) | Uric acid |
| D,L-Octopamine | Mifepristone | Valproic acid (250ug/ml) |
| D,L-Propranolol | N-Acetylprocainamide | Venlafaxine HCl |
| D,L-Tryptophan | NaCl (4000mg/dL) | Verapamil |
| D,L-Tyrosine | Nalidixic acid | Vitamin B2 |
| Delorazepam(except BZOtest) | Naloxonehydrochloride(except OXYtest) | Vitamin C |
| Deoxycorticosterone | Naltrexone hydrochloride | Zaleplon |
| Desloratadine | Naproxen | Zomepirac |
| Dextromethorphan | Niacinamide | β-Estradiol |
| Diclofenac sodium | Nicotine | |
| Diflunisal | Nifedipine |
{19}------------------------------------------------
Interference by pH and specific gravity were also evaluated using pooled urine specimens with concentrations of 0 (drug-free), at 50% below and 50% above each corresponding cutoff. The results demonstrated that pH levels of 4 to 9 and specific gravity levels of 1.000 to 1.035 do not affect the results of the assays.
B. Method comparison study
The method comparison studies for the device were performed in-house with three operators. Operators ran 80 (40 negative and 40 positive) unaltered urine clinical samples for each drug. The samples were blind labeled and compared to LC-MS/MS results. The results are presented in the table below:
{20}------------------------------------------------
| Drug test | Test Result | Drug-Free | Low Negative by LC-MS/MS (less than - 50%) | Near Cutoff Negative by LC-MS/MS (Between - 50% and the Cutoff) | Near Cutoff Positive by LC-MS/MS (Between the cutoff and +50%) | High Positive by LC-MS/MS (greater than +50%) | |
|---|---|---|---|---|---|---|---|
| AMP | Operator | + | 0 | 0 | 1 | 10 | 30 |
| A | - | 12 | 13 | 14 | 0 | 0 | |
| Operator | + | 0 | 0 | 1 | 9 | 30 | |
| B | - | 12 | 13 | 14 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 9 | 30 | |
| C | - | 12 | 13 | 14 | 1 | 0 | |
| BAR | Operator | + | 0 | 0 | 0 | 14 | 25 |
| A | - | 12 | 14 | 14 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 15 | 25 | |
| B | - | 12 | 14 | 13 | 0 | 0 | |
| Operator | + | 0 | 0 | 1 | 14 | 25 | |
| C | - | 12 | 14 | 13 | 1 | 0 | |
| BUP | Operator | + | 0 | 0 | 1 | 10 | 29 |
| A | - | 12 | 17 | 10 | 1 | 0 | |
| Operator | + | 0 | 0 | 0 | 10 | 29 | |
| B | - | 12 | 17 | 11 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 11 | 29 | |
| C | - | 12 | 17 | 10 | 0 | 0 | |
| BZO | Operator | + | 0 | 0 | 0 | 11 | 28 |
| A | - | 12 | 13 | 15 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 11 | 28 | |
| B | - | 12 | 13 | 14 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 12 | 28 | |
| C | - | 12 | 13 | 14 | 0 | 0 | |
| COC | Operator | + | 0 | 0 | 1 | 11 | 28 |
| A | - | 12 | 18 | 9 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 11 | 28 | |
| B | - | 12 | 18 | 9 | 1 | 0 | |
| Operator | + | 0 | 0 | 0 | 12 | 28 | |
| C | - | 12 | 18 | 10 | 0 | 0 | |
| MDMA | Operator | + | 0 | 0 | 1 | 13 | 27 |
| A | - | 12 | 14 | 13 | 0 | 0 | |
| Operator | + | 0 | 0 | 1 | 12 | 27 | |
| B | - | 12 | 14 | 13 | 1 | 0 | |
| Operator | + | 0 | 0 | 0 | 12 | 27 | |
| C | - | 12 | 14 | 14 | 1 | 0 | |
| MET | Operator | + | 0 | 0 | 0 | 10 | 29 |
| A | - | 12 | 16 | 12 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 11 | 29 | |
| B | - | 12 | 16 | 11 | 0 | 0 | |
| Operator | + | 0 | 0 | 1 | 10 | 29 | |
| C | - | 12 | 16 | 11 | 1 | 0 | |
| MOP/OPI | Operator | + | 0 | 0 | 0 | 13 | 26 |
| A | - | 12 | 16 | 12 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 13 | 26 | |
| B | - | 12 | 16 | 11 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 14 | 26 | |
| C | - | 12 | 16 | 11 | 0 | 0 | |
| MTD | Operator | + | 0 | 0 | 1 | 10 | 29 |
| A | - | 12 | 16 | 11 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 11 | 29 | |
| B | - | 12 | 16 | 11 | 0 | 0 | |
| Operator | + | 0 | 0 | 1 | 10 | 29 | |
| C | - | 12 | 16 | 11 | 1 | 0 | |
| OXY | Operator | + | 0 | 0 | 0 | 12 | 28 |
| A | - | 12 | 16 | 12 | 0 | 0 | |
| Operator | + | 0 | 0 | 1 | 11 | 28 | |
| B | - | 12 | 16 | 11 | 1 | 0 | |
| Operator | + | 0 | 0 | 0 | 11 | 28 | |
| C | - | 12 | 16 | 12 | 1 | 0 | |
| PCP | Operator | + | 0 | 0 | 1 | 12 | 27 |
| A | - | 12 | 16 | 11 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 12 | 27 | |
| B | - | 12 | 16 | 11 | 1 | 0 | |
| Operator | + | 0 | 0 | 0 | 12 | 27 | |
| C | - | 12 | 16 | 12 | 1 | 0 | |
| TCA | Operator | + | 0 | 0 | 1 | 9 | 30 |
| A | - | 12 | 15 | 12 | 1 | 0 | |
| Operator | + | 0 | 0 | 0 | 9 | 30 | |
| B | - | 12 | 15 | 13 | 1 | 0 | |
| Operator | + | 0 | 0 | 1 | 10 | 30 | |
| C | - | 12 | 15 | 12 | 0 | 0 | |
| THC | Operator | + | 0 | 0 | 1 | 10 | 30 |
| A | - | 12 | 17 | 10 | 0 | 0 | |
| Operator | + | 0 | 0 | 1 | 9 | 30 | |
| B | - | 12 | 17 | 10 | 1 | 0 | |
| Operator | + | 0 | 0 | 0 | 10 | 30 | |
| C | - | 12 | 17 | 11 | 0 | 0 | |
| FYL | Operator | + | 0 | 0 | 2 | 22 | 16 |
| A | + | 0 | 0 | 2 | 22 | 16 | |
| - | 12 | 13 | 13 | 2 | 0 | ||
| B | + | 0 | 0 | 3 | 22 | 16 | |
| - | 12 | 13 | 12 | 2 | 0 | ||
| C | + | 0 | 0 | 2 | 22 | 16 | |
| - | 12 | 13 | 13 | 2 | 0 |
{21}------------------------------------------------
{22}------------------------------------------------
Discordant Results are summarized below.
| Drug | Operator | Sample Number | LC/MS/MS Result (ng/mL) | Accurate Result |
|---|---|---|---|---|
| AMP | A | N0957 | 494.481 | + |
| B | N0949 | 491.559 | + | |
| B | N0966 | 619.861 | - | |
| C | N0949 | 491.559 | + | |
| C | N0966 | 619.861 | - | |
| BAR | A | N0756 | 304.564 | - |
| B | N0733 | 294.112 | + | |
| C | N0728 | 271.093 | + | |
| C | N0756 | 304.564 | - | |
| BUP | A | N1328 | 11.971 | - |
| A | N1364 | 9.635 | + | |
| B | N1331 | 10.385 | - | |
| C | N1353 | 9.149 | + | |
| BZO | A | N0840 | 306.319 | - |
| B | N0825 | 297.539 | + | |
| B | N0840 | 306.319 | - | |
| C | N0818 | 261.170 | + | |
| COC | A | N0538 | 122.809 | + |
| A | N0557 | 151.609 | - | |
| B | N0538 | 122.809 | + | |
| B | N0557 | 151.609 | - | |
| MDMA | A | N0219 | 498.645 | + |
| B | N0219 | 498.645 | + | |
| B | N0220 | 503.780 | - | |
| C | N0220 | 503.780 | - | |
| MET | A | N1032 | 514.414 | - |
| B | N1066 | 482.955 | + | |
| C | N1032 | 514.414 | - | |
| C | N1066 | 482.955 | + | |
| MOP/OPI | A | N0129 | 337.839 | - |
| B | N0108 | 289.278 | + | |
| B | N0129 | 337.839 | - | |
| C | N0108 | 289.278 | + | |
| MTD | A | N0608 | 299.063 | + |
| A | N0615 | 300.546 | - | |
| B | N0618 | 295.461 | + | |
| C | N0608 | 299.063 | + | |
| C | N0630 | 306.162 | - | |
| OXY | B | N0314 | 100.815 | - |
| B | N0356 | 98.859 | + | |
| C | N0309 | 104.534 | - | |
| PCP | A | N0426 | 23.006 | + |
| A | N0459 | 25.355 | - | |
| B | N0441 | 25.891 | - | |
| B | N0456 | 22.017 | + | |
| C | N0459 | 25.355 | - | |
| TCA | A | N1106 | 981.923 | + |
| A | N1146 | 1025.476 | - | |
| B | N1123 | 1038.965 | - | |
| C | N1140 | 981.903 | + | |
| THC | A | N1227 | 49.942 | + |
| B | N1227 | 49.942 | + | |
| B | N1236 | 50.876 | - | |
| FYL | A | N1402 | 0.917 | + |
| A | N1425 | 1.024 | - | |
| A | N1439 | 1.015 | - | |
| A | N1465 | 0.920 | + | |
| B | N1402 | 0.917 | + | |
| B | N1420 | 0.872 | + | |
| B | N1422 | 1.079 | - | |
| B | N1442 | 0.906 | + | |
| B | N1447 | 1.013 | - | |
| C | N1439 | 1.015 | - | |
| C | N1442 | 0.906 | + | |
| C | N1447 | 1.013 | - | |
| C | N1465 | 0.920 | + |
{23}------------------------------------------------
C. Lay person study
A lay user study was performed at three intended user sites with 140 lay persons. 48 males and 92 females tested AllTest Multi-Drug Urine Test Panel. They had diverse educational and professional backgrounds and their ages ranged from 21 to > 50. Urine samples were prepared at the following concentrations; -100%, +/-75%, +/-50%, +/-25% of the cutoff by spiking drug(s) into drug freepooled urine specimens. The concentrations of the samples were confirmed by LC-MS/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample, and a device. The results are summarized below.
{24}------------------------------------------------
| Cutoff(ng/mL) | Concentration | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Drug | Results | -100% | -75% | -50% | -25% | +25% | +50% | +75% | |
| cutoff | cutoff | cutoff | cutoff | cutoff | cutoff | cutoff | |||
| AMP | 500 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% | ||
| BAR | 300 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95.0% | 100% | 100% | 100% | ||
| BZO | 300 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% | ||
| BUP | 10 | Negative | 20 | 20 | 20 | 18 | 0 | 0 | 0 |
| Positive | 0 | 0 | 0 | 2 | 20 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90.0% | 100% | 100% | 100% | ||
| COC | 150 | Negative | 20 | 20 | 20 | 18 | 0 | 0 | 0 |
| Positive | 0 | 0 | 0 | 2 | 20 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90.0% | 100% | 100% | 100% | ||
| MDMA | 500 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95.0% | 100% | 100% | 100% | ||
| MET | 500 | Negative | 20 | 20 | 20 | 18 | 2 | 0 | 0 |
| Positive | 0 | 0 | 0 | 2 | 18 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90.0% | 90.0% | 100% | 100% | ||
| MOP | 300 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90.0% | 95.0% | 100% | 100% | ||
| MTD | 300 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95.0% | 100% | 100% | 100% | ||
| OXY | 100 | Negative | 20 | 20 | 20 | 19 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95.0% | 95.0% | 100% | 100% | ||
| PCP | 25 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95.0% | 100% | 100% | 100% | ||
| TCA | 1000 | Negative | 20 | 20 | 20 | 19 | 0 | 0 | 0 |
| Positive | 0 | 0 | 0 | 1 | 20 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 95.0% | 100% | 100% | 100% | ||
| THC | 50 | Negative | 20 | 20 | 20 | 18 | 1 | 0 | 0 |
| Positive | 0 | 0 | 0 | 2 | 19 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90.0% | 95.0% | 100% | 100% | ||
| FYL | 1 | Negative | 20 | 20 | 20 | 18 | 0 | 0 | 0 |
| Positive | 0 | 0 | 0 | 2 | 20 | 20 | 20 | ||
| Total | 20 | 20 | 20 | 20 | 20 | 20 | 20 | ||
| Agreement (%) | 100% | 100% | 100% | 90.0% | 100.0% | 100% | 100% |
Result of AllTest Multi-Drug Urine Test Panel:
{25}------------------------------------------------
Participants were given surveys on the ease of understanding the instruction for use. All participants indicated that the device instruction is easy to understand and follow. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.
Clinical Studies: Not applicable.
13. Conclusion
Based on the test principle and performance characteristics of the device including precision, cut-off, interference, specificity, method comparison, and lay-user studies of the devices, it is concluded that AllTest Multi-Drug Urine Test Panel and AllTest Multi-Drug Rapid Urine Test Panel are substantially equivalent to the predicate device.
§ 862.3100 Amphetamine test system.
(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).