K Number
K242057
Date Cleared
2024-08-14

(30 days)

Product Code
Regulation Number
882.4400
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

COOLIEF* Radiofrequency Generator (CRG) System is indicated for use to create lesions during neurological lesion procedures and for the coagulation and decompression of disc material to treat symptomatic patients with contained herniated discs. The CRG is to be used with only AVANOS* and Diros RF Probes and accessories, such as Cooled / Standard RF Pain Management Probes, Cannulas, and Introducers. Use of the CRG System is limited to the indications cleared under these devices and accessories.

Device Description

COOLIEF* Radiofrequency Generator (CRG-BASIC, CRG-BASIC-R, CRG-ADVANCED, CRG-ADVANCED-R)

AI/ML Overview

The provided text is an FDA 510(k) clearance letter for the COOLIEF* Radiofrequency Generator. It does not contain information about the acceptance criteria or a study that proves the device meets specific acceptance criteria in the manner requested (e.g., performance metrics, sample sizes, ground truth establishment, or expert qualifications).

The letter primarily:

  • Acknowledges receipt and review of the 510(k) premarket notification.
  • Determines substantial equivalence to legally marketed predicate devices.
  • Outlines general controls provisions and other regulatory requirements applicable to the device.
  • Specifies the "Indications for Use" for the device.

Therefore, I cannot provide a response with the information requested in your bullet points as it is not present in the given document.

{0}------------------------------------------------

August 14, 2024

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration".

Avanos Medical, Inc. % Wendy Laird Avanos Consultant, Regulatory Affairs Specialist 5405 Windward Parkway Alpharetta, Georgia 30004

Re: K242057

Trade/Device Name: COOLIEF* Radiofrequency Generator (CRG-BASIC, CRG-BASIC-R, CRG-ADVANCED, CRG-ADVANCED-R) Regulation Number: 21 CFR 882.4400 Regulation Name: Radiofrequency lesion generator Regulatory Class: Class II Product Code: GXD Dated: July 12, 2024 Received: July 15, 2024

Dear Wendy Laird:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Image /page/1/Picture/5 description: The image shows a digital signature. The signature is from Adam D. Pierce -S. The date of the signature is August 14, 2024. The time of the signature is 20:55:24 -04'00'.

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices

{2}------------------------------------------------

Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K242057

Device Name

COOLIEF* Radiofrequency Generator (CRG-BASIC, CRG-BASIC-R, CRG-ADVANCED, CRG-ADVANCED-R)

Indications for Use (Describe)

COOLIEF* Radiofrequency Generator (CRG) System is indicated for use to create lesions during neurological lesion procedures and for the coagulation and decompression of disc material to treat symptomatic patients with contained herniated discs. The CRG is to be used with only AVANOS* and Diros RF Probes and accessories, such as Cooled / Standard RF Pain Management Probes, Cannulas, and Introducers. Use of the CRG System is limited to the indications cleared under these devices and accessories.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 882.4400 Radiofrequency lesion generator.

(a)
Identification. A radiofrequency lesion generator is a device used to produce lesions in the nervous system or other tissue by the direct application of radiofrequency currents to selected sites.(b)
Classification. Class II (performance standards).