(98 days)
The diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Thoracic, Trans-esophageal (Cardiac) and Peripheral vessel.
It is intended for use by, or by the order of, and under the supervision of, an appropriately trained healthcare professional who is qualified for direct use of medical devices. It can be used in hospitals, private practices, clinics and similar care environment for clinical diagnosis of patients.
Modes of Operation: 2D mode. Color Doppler mode. Pulsed Wave (PW) Doppler mode. Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, Power Doppler (PD) mode, ElastoScan™ Mode, MV-Flow Mode, Multi Image mode (Dual Quad) Combined modes 3D/AD model
The HERA Z20, R20, HERA Z30, R30 diagnostic ultrasound system are a general purpose, mobile, software controlled, diagnostic ultrasound system. Their function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler mode, Power Doppler (PD) mode, M mode, Pulsed Wave (PW) Doppler mode, Continuous Wave (CW) Doppler mode, Tissue Doppler Imaging (TDI) mode, Tissue Doppler Wave (TDW) mode, ElastoScan Mode, Combined modes, MV-Flow mode, Multi-Image mode(Dual, Quad), 3D/4D mode.
The HERA Z20, R20, HERA Z30, R30 diagnostic ultrasound system also give the operator the ability to measure anatomical structures and offer analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HERA Z20, R20, HERA Z30, R30 diagnostic ultrasound system have a real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
Here's a summary of the acceptance criteria and the studies conducted for the AI-powered features of the HERA Z20, R20, HERA Z30, R30 Diagnostic Ultrasound System, based on the provided text.
This document describes several AI-powered features: Live ViewAssist, EzVolume, UterineContour, ViewAssist, HeartAssist, and BiometryAssist. Each feature has its own acceptance criteria and study findings.
1. A table of acceptance criteria and the reported device performance
Note: Some performance metrics were not explicitly stated as "acceptance criteria" but rather as "summary test statistics or other test results," indicating the device's measured performance against implicit or internal targets.
AI Feature | Metric | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
Live ViewAssist | Quality assessment (Cohen's kappa) | Threshold 0.7 | Average Cohen's kappa coefficient: 0.818 |
Time/duration (Frames per Second - FPS) | Threshold 20 FPS | Average speed: 30.06 FPS | |
EzVolume | Acceptance Rate (segmentation) | Higher than 70% for each label | 1st Trimester: Fluid 98%, Fetus 96%, Umbilical-cord 80%, Placenta 86%, Uterus 89% |
2nd/3rd Trimester: Fluid 92%, Head 94%, Body 84%, Limbs 83%, Umbilical-cord 82%, Placenta 85%, Uterus 87% | |||
Mean DSC (segmentation) | (No explicit numerical criterion provided, but correlation with acceptance rate indicates adequacy) | 1st Tri (Accepted): Fluid 0.96, Fetus 0.91, Umbilical-cord 0.68, Placenta 0.74, Uterus 0.93 | |
1st Tri (Rejected): Fluid 0.17, Fetus 0.55, Umbilical-cord 0.37, Placenta 0.33, Uterus 0.32 |
2nd/3rd Tri (Accepted): Fluid 0.78, Head 0.94, Body 0.68, Umbilical-cord 0.67, Limbs 0.66, Placenta 0.75, Uterus 0.80
2nd/3rd Tri (Rejected): Fluid 0.25, Head 0.46, Body 0.29, Umbilical-cord 0.38, Limbs 0.39, Placenta 0.32, Uterus 0.30 |
| UterineContour| Segmentation (uterus Dice-score) | Not explicitly stated as acceptance criteria | Average dice-score of uterus: 96% |
| | Segmentation (endometrium Dice-score) | Not explicitly stated as acceptance criteria | Average dice-score of endometrium: 92% |
| | 3D coronal view adaptation | Proportion of appropriateness evaluated as clinically diagnosable, with over 90% of all cases | Over 90% of all cases |
| ViewAssist | View recognition accuracy | Threshold 89% | Average recognition accuracy: 94.50% |
| | Anatomy annotation (Dice-score) | Threshold 0.8 | Average Dice-score: 0.892 |
| HeartAssist | View recognition accuracy | Threshold 89% | Average recognition accuracy: 95.00% |
| | Segmentation (Dice-score) | Threshold 0.8 | Average Dice-score: 0.876 |
| | Size measurement (Area error rate) | Not explicitly stated as acceptance criteria | 8% or less |
| | Size measurement (Angle error rate) | Not explicitly stated as acceptance criteria | 4% or less |
| | Size measurement (Circumference error rate) | Not explicitly stated as acceptance criteria | 11% or less |
| | Size measurement (Diameter error rate) | Not explicitly stated as acceptance criteria | 11% or less |
| BiometryAssist| Segmentation (Dice-score) | Threshold 0.8 | Average Dice-score: 0.928 |
| | Size measurement (Circumference error rate) | Not explicitly stated as acceptance criteria | 8% or less |
| | Size measurement (Distance error rate) | Not explicitly stated as acceptance criteria | 4% or less |
| | Size measurement (NT, NB, IT error rate)| Not explicitly stated as acceptance criteria | 1mm or less |
2. Sample size used for the test set and the data provenance
AI Feature | Test Set Sample Size | Data Provenance |
---|---|---|
Live ViewAssist | 3,900 fetal ultrasound images | Mix of retrospective and prospective data collection in clinical practice from Americans and Koreans (gender: female, reproductive age; BMI 17-45.4) |
EzVolume | 200 test volumes (100 in 1st trimester, 100 in 2nd/3rd trimester) | Mix of retrospective and prospective data collection in clinical practice from Koreans, Americans, Italians, and British (gender: female, reproductive age) |
UterineContour | 450 sagittal uterus images (for segmentation) and 30 sagittal images (for 3D coronal view) | Mix of retrospective and prospective data collection in clinical practice from three hospitals in Korea (gender: female, reproductive age) |
ViewAssist | 1,600 fetal ultrasound and fetal biometry images | Mix of retrospective and prospective data collection in clinical practice from two hospitals in America and Korea (gender: female, reproductive age; BMI 17-45.4) |
HeartAssist | 440 fetal heart images | Mix of retrospective and prospective data collection in clinical practice from America and Korea (gender: female, reproductive age; BMI 17-45.4) |
BiometryAssist | 360 fetal biometry images | Mix of retrospective and prospective data collection in clinical practice from two hospitals in America and Korea (gender: female, reproductive age; BMI 17-45.4) |
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
AI Feature | Number of Experts | Qualifications of Experts |
---|---|---|
Live ViewAssist | 3 primary experts, 1 supervising expert | An obstetrician with more than 20 years of experience (primary). Two sonographers with more than 10 years of experience, all in fetal cardiology (primary). Another obstetrician with more than 25 years of experience (supervisor). |
EzVolume | 4 primary experts, 1 supervising expert | An obstetrician with more than 20 years of experience (primary). Three examiners (clinical experts) with more than 10 years of experience, all in fetal diagnosis (primary). Another obstetrician with more than 25 years of experience (supervisor). |
UterineContour | 3 OB/GYN experts | Three participating OB/GYN experts with more than 10 years' experience. |
ViewAssist | 3 primary experts, 1 supervising expert | An obstetrician with more than 20 years of experience (primary). Two sonographers with more than 10 years of experience, all in fetal cardiology (primary). Another obstetrician with more than 25 years of experience (supervisor). |
HeartAssist | 3 primary experts, 1 supervising expert | An obstetrician with more than 20 years of experience (primary). Two sonographers with more than 10 years of experience, all in fetal cardiology (primary). Another obstetrician with more than 25 years of experience (supervisor). |
BiometryAssist | 3 primary experts, 1 supervising expert | An obstetrician with more than 20 years of experience (primary). Two sonographers with more than 10 years of experience, all in fetal cardiology (primary). Another obstetrician with more than 25 years of experience (supervisor). |
4. Adjudication method for the test set
AI Feature | Adjudication Method |
---|---|
Live ViewAssist | Ground truth established by consensus of 3 experts, supervised by 1. Exact method (e.g., 2+1, 3+1) not explicitly detailed, but implied by "manual drawing" and "classified into acceptable and not-acceptable views by three participating experts." |
EzVolume | Ground truths were drawn manually by four participating clinical experts, supervised by one. |
UterineContour | Each of the 3 experts delineated structures. Conflicts were resolved by a consensus of the three experts ("fixed the wrong part with consensus"). |
ViewAssist | Ground truth established by consensus of 3 experts, supervised by 1. |
HeartAssist | Ground truth established by consensus of 3 experts, supervised by 1. |
BiometryAssist | Ground truth established by consensus of 3 experts, supervised by 1. |
5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly stated as performed to compare human readers with and without AI assistance. The studies described focus on the standalone performance of the AI algorithms.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, standalone performance studies of the AI algorithms were done for all the described features: Live ViewAssist, EzVolume, UterineContour, ViewAssist, HeartAssist, and BiometryAssist. The reported metrics like Cohen's kappa, FPS, acceptance rates, Dice scores, and error rates are all measures of the algorithm's performance without human intervention during the assessment phase (though human experts were used to establish ground truth).
7. The type of ground truth used
For all features, the ground truth was established by expert consensus based on manual classification, delineation, or drawing by qualified clinical experts (obstetricians, sonographers, and examiners).
8. The sample size for the training set
The document explicitly states that "Data used for training, tuning and validation purpose are completely separated from the ones during training process and there is no overlap among the three." However, the exact sample size for the training set itself is not provided for any of the features. The sample sizes listed in Section 2 are for the test/validation sets.
9. How the ground truth for the training set was established
For all features, the ground truth for the training set (and validation/evaluation sets) was established through manual classification, delineation, or drawing by the same groups of qualified clinical experts mentioned in section 3, following similar expert consensus processes as described for the test sets. For UterineContour, initial delineations by 3 experts were then reviewed and fixed with consensus for unmatched results.
§ 892.1550 Ultrasonic pulsed doppler imaging system.
(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.