(57 days)
No
The device description and performance studies do not mention any AI or ML components. The device is a set of physical dilators.
Yes
The device is intended to relieve symptoms of vaginismus and related painful sex, which are therapeutic medical conditions.
No
Explanation: The device is used for treatment (dilation) rather than for diagnosing a condition. Its purpose is to relieve symptoms and prepare for procedures, not to identify or quantify a disease.
No
The device description clearly states the device is comprised of "8 progressively larger and color-coded medical-grade silicone dilators," which are physical hardware components.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body (like blood, urine, tissue) to provide information about a person's health. They are used outside the body (in vitro).
- Device Function: The Intimate Rose Vaginal Dilators are physical tools used within the body (in vivo) to mechanically dilate the vagina. They do not perform any tests on samples or provide diagnostic information based on biological markers.
- Intended Use: The intended use is for controlled dilation for examination, surgical preparation, and symptom relief of vaginismus and painful sex. This is a therapeutic and preparatory function, not a diagnostic one.
Therefore, the Intimate Rose Vaginal Dilators fall under the category of a medical device, but not an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The Intimate Rose Vaginal Dilators are tools intended for controlled dilation of the vagina. They can be used for dilation for an examination (by your doctor), in preparation for a surgical procedure, or to help relieve the symptoms of vaginismus (condition that involves tightening of the vaginal muscles) and related painful sex.
Product codes (comma separated list FDA assigned to the subject device)
HDX
Device Description
The Intimate Rose Vaginal Dilators are intended to treat women suffering from vaginismus (including related dyspareunia). Vaginismus is the involuntary spasm of the muscles in the vaginal wall which then inhibits sexual intercourse by making it painful or impossible. Dyspareunia is the pain experienced during sexual intercourse caused by physical and/or emotional problems. The device is used as a tool to dilate the vagina in controlled stages. It is a reusable device that comes into contact with the vaginal mucosal membrane for a prolonged (> 24 hours to
§ 884.3900 Vaginal stent.
(a)
Identification. A vaginal stent is a device used to enlarge the vagina by stretching, or to support the vagina and to hold a skin graft after reconstructive surgery.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
August 14, 2024
Plus EV Holdings dba Intimate Rose % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059
Re: K241748
Trade/Device Name: Intimate Rose Vaginal Dilators Regulation Number: 21 CFR 884.3900 Regulation Name: Vaginal Stent Regulatory Class: II Product Code: HDX Dated: July 30, 2024 Received: July 31, 2024
Dear Dave Yungvirt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
1
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
2
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason Roberts -S
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
3
Indications for Use
510(k) Number (if known) K241748
Device Name Intimate Rose Vaginal Dilators
Indications for Use (Describe)
The Intimate Rose Vaginal Dilators are tools intended for controlled dilation of the vagina. They can be used for dilation for an examination (by your doctor), in preparation for a surgical procedure, or to help relieve the symptoms of vaginismus (condition that involves tightening of the vaginal muscles) and related painful sex.
Type of Use (Select one or both, as applicable) | |
---|---|
------------------------------------------------- | -- |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4
K241748 Intimate Rose Vaginal Dilators 510(k) Summary
Submitter Name: Plus EV Holdings dba Intimate Rose
Submitter Address: 1419 Murray St North Kansas City MO 64116 United States
Submitter Contact Telephone: 816-805-6722
Submitter Contact: Mr. Aaron Wilt
Submitter Contact Email: aaron@plusevholdings.com
Correspondent Name: FDA Compliance Group
Correspondent Address: 33 Golden Eagle Lane Littleton CO 80127 United States
Correspondent Contact Telephone: 720-254-5756
Correspondent Contact: Mr. Kevin Walls
Correspondent Contact Email: kevin@reginsight.com
Date Prepared: August 14, 2024
Device Trade Name: Intimate Rose Vaginal Dilators
Common Name: Vaginal Dilators
Classification Name: Vaginal stent
Classification Number: 884.3900
Product Code(s): HDX
Predicate Device:
-
- K220035, Milli Vaginal Dilator, Product Code HDX
The predicate device has not been subject to a design-related recall.
- K220035, Milli Vaginal Dilator, Product Code HDX
Reference Device:
- K231430, Intimate Rose Vaginal Dilators, Product Code HDX
Device Description Summary: The Intimate Rose Vaginal Dilators are intended to treat women suffering from vaginismus (including related dyspareunia). Vaginismus is the involuntary spasm of the muscles in the vaginal wall which then inhibits sexual intercourse by making it painful or impossible. Dyspareunia is the pain experienced during sexual intercourse caused by physical and/or emotional problems. The device is used as a tool to dilate the vagina in controlled stages. It is a reusable device that comes into contact with the vaginal mucosal membrane for a prolonged (> 24 hours to