K Number
K241672
Device Name
COPRO 3
Date Cleared
2024-09-06

(87 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Non-fractional mode is indicated for incision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otorhinolaryngology (ENT), gynecology, neurosurgery, dental and oral surgery and genitourinary surgery. Fractional mode is indicated for only ablative skin resurfacing.

Device Description

Not Found

AI/ML Overview

This FDA clearance letter for the Copro 3 provides limited information regarding the performance study that proves the device meets acceptance criteria. The letter primarily focuses on the substantial equivalence determination and regulatory compliance.

Therefore, I cannot provide a detailed response to all aspects of your request based solely on the provided text. Specifically, the document does not contain:

  • A table of acceptance criteria and reported device performance.
  • Sample size used for the test set or its provenance.
  • Number of experts used to establish ground truth or their qualifications.
  • Adjudication method for the test set.
  • Information about a multi-reader multi-case (MRMC) comparative effectiveness study or effect size.
  • Details about a standalone (algorithm only) performance study.
  • The type of ground truth used for performance evaluation.
  • The sample size for the training set.
  • How the ground truth for the training set was established.

The letter states, "We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices..." This heavily implies that the device's performance was compared to a predicate device, and its safety and effectiveness were demonstrated through bench testing, perhaps some animal studies, or a literature review rather than a large-scale clinical trial with human subjects and a detailed AI/software performance study as typically seen for AI/ML devices.

Based on the provided information, I can only state the following:

1. A table of acceptance criteria and the reported device performance:

  • Not provided in the document. The letter confirms substantial equivalence without detailing specific performance metrics or acceptance criteria for studies typically associated with AI/ML devices. The "Indications for Use" section describes the intended applications of the device (e.g., incision, ablation, vaporization, coagulation of soft tissues, and ablative skin resurfacing). The acceptance criteria would likely relate to the device's ability to perform these functions safely and effectively, possibly measured by parameters like power output accuracy, tissue damage characteristics, or procedural outcomes, which are not detailed here.

2. Sample size used for the test set and the data provenance:

  • Not provided in the document.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not provided in the document.

4. Adjudication method for the test set:

  • Not provided in the document.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable/Not mentioned. This device
    ("Laser Surgical Instrument") is a hardware device, not an AI/ML algorithm that assists human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable/Not mentioned. This is a hardware device, not a standalone AI algorithm.

7. The type of ground truth used:

  • Not provided in the document. For a laser surgical instrument, ground truth would typically refer to objective measurements of the device's physical output (e.g., laser power, beam profile) and its effects on biological tissues (e.g., depth of ablation, coagulation zone), potentially verified through histology or other objective means.

8. The sample size for the training set:

  • Not applicable/Not mentioned. As this is a hardware device, the concept of a "training set" as understood for AI/ML algorithms does not directly apply. Performance is typically established through engineering tests, bench tests, and possibly animal or cadaver studies.

9. How the ground truth for the training set was established:

  • Not applicable/Not mentioned.

In summary, the provided FDA letter confirms substantial equivalence for the Copro 3 laser surgical instrument, indicating that it meets the regulatory requirements for market approval. However, it does not include the detailed performance study information, particularly for AI/ML-specific criteria, that you requested. Such details would typically be found in the manufacturer's 510(k) submission, not in the FDA's clearance letter itself.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

September 6, 2024

AMT Engineering Co., Ltd % Young Chi C.E.O. Bio-Med USA Inc 27 New England Dr Ramsey, New Jersey 07446

Re: K241672

Trade/Device Name: Copro 3 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: June 10, 2024 Received: June 11, 2024

Dear Young Chi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Yan Fu -S
Yan Fu -S Date: 2024.09.06 11:47:04
-04'00'

for

Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices

{2}------------------------------------------------

OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241672

Device Name Copro 3

Indications for Use (Describe)

Non-fractional mode is indicated for incision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otorhinolaryngology (ENT), gynecology, neurosurgery, dental and oral surgery and genitourinary surgery. Fractional mode is indicated for only ablative skin resurfacing.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.