(133 days)
Not Found
No
The document describes standard medical image viewing, manipulation, and post-processing software without mentioning AI or ML capabilities. The performance studies focus on specific image processing functionalities like biplane views and registration, not on AI/ML model performance.
No
The device is described as medical software for viewing, manipulating, and storing medical images for diagnostic purposes and treatment planning, not for directly treating a disease or condition.
Yes
Explanation: The "Intended Use / Indications for Use" section explicitly states, "The syngo Application Software is a medical software for real-time viewing, image manipulation, 3Dvisualization, communication, and storage of medical images and data on exchange media. It is used for diagnostic image viewing and postprocessing and postprocessing during interventional procedures." Additionally, the "Device Description" refers to it as "medical diagnostic software."
No
The device description explicitly states that the software can be deployed on independent hardware and configured within a network, and also deployed on systems of the Siemens Healthineers Angiography system family, fluoroscopy systems, or radiographic systems. This indicates it is software that runs on or is integrated with hardware, not a standalone software-only medical device.
Based on the provided information, the syngo Application Software is not an In Vitro Diagnostic (IVD).
Here's why:
- IVDs analyze samples from the human body (like blood, urine, tissue) to provide information about a person's health. The syngo Application Software deals with medical images and data from imaging modalities (digital radiographic, fluoroscopic, interventional, and angiographic systems). It processes and displays these images for diagnostic review, post-processing, and guidance during procedures.
- The intended use and device description clearly state its purpose is for viewing, manipulating, and analyzing medical images and data. There is no mention of analyzing biological samples.
- The input modalities are imaging systems, not laboratory equipment for sample analysis.
Therefore, the syngo Application Software falls under the category of medical imaging software, not an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The syngo Application Software is a medical software for real-time viewing, image manipulation, 3Dvisualization, communication, and storage of medical images and data on exchange media. It is used for diagnostic image viewing and postprocessing and postprocessing during interventional procedures.
The syngo Application Software can be deployed on independent hardware such as a stand-alone diagnostic review, postprocessing, and reporting workstation. It can also be configured within a network to send and receive DICOM data.
Furthermore, the syngo Application Software can be deployed on systems of the Siemens Healthineers Angiography system family.
It provides image guided solutions in the operating room, for image guided surgery, by image fusion and by navigation systems, image guided solutions in interventional cardiology and electrophysiology and image guided solutions for interventional oncology, interventional radiology, and interventional neuroradiology.
The syngo Application Software can also be combined with fluoroscopy systems or radiographic systems.
The syngo Application Software can be configured with a variety of syngo or Windows-based software options, which are intended to assist the physician in diagnosis, treatment planning and treatment control. It includes commercially available postprocessing techniques and OEM options.
Procedures that can be performed include: minimally invasive surgical procedures and minimally invasive tumor treatment.
Product codes
LLZ
Device Description
The "syngo Application Software" is medical diagnostic software for real-time viewing, diagnostic review, post-processing image manipulation, communication, reporting, and storage of medical images and data on exchange media. It provides imageguided solutions in the operating room, for image-quided surgery, by Image Fusion and by navigation systems, image-guided solutions in interventional cardiology and electrophysiology, and image-guided solutions for interventional oncology, interventional radiology, and interventional neuroradiology. It can be deployed with a variety of syngo or Windows-based software options, which are intended to assist the physician in the evaluation of digital radiographic examinations, including diagnosis and/or treatment planning.
Siemens "syngo Application Software" is designed to work with digital radiographic, fluoroscopic, interventional, and angiographic systems.
Siemens Medical Solutions USA, Inc. hereby submits this Traditional 510(k) to request clearance to market the following modifications:
-
- Updated Software VE30A to VE40C
- A. Updated myNeedle Guide "Progress Needle" steps with an additional "couple A+B" button (Software feature for ARTIS icono biplane only)
- B. Updated syngo Biplane 3D Roadmap feature
-
- Product Claims List
The "syngo Application Software" may be installed either on Siemens released PC hardware, on Siemens X-ray systems, or on Siemens angiography systems. The "syngo Application Software" is within the same classification and the intended use and the general Indications for Use Statement for Siemens' Medical Image Management and Processing System remains the same except for a minor grammatical change (inclusion of "Healthineers").
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found (The text mentions digital radiographic, fluoroscopic, interventional, and angiographic systems but does not specify them as input modalities.)
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
physician (for diagnosis, treatment planning and treatment control); operating room (for image guided surgery), interventional cardiology and electrophysiology, interventional oncology, interventional radiology, and interventional neuroradiology.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-clinical tests were conducted for the "syngo Application Software" during product development.
The modifications described in this Premarket Notification are supported with verification and validation testing.
Software documentation for a Basic Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Device Software Functions" issued on June 14, 2023, and "Off-The-Shelf Software Use in Medical Devices" is also included as part of this submission. Non-clinical tests were conducted on "syngo Application Software" during product development.
The Risk analysis was completed, and risk control was implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.
The Human Factor Usability Validation showed no safety-relevant functions that need to be validated with a summative usability validation according to the IEC and FDA Guidelines. "syngo Application Software" is safe and effective for intended users, uses, and use environments through the design control verification and validation process. No further risk mitigations are necessary.
Tests were conducted in support of the new/updated functionalities for the "syngo Application Software": Updated myNeedle Guide testing encompassed Biplane-Couple-A-B, Invoke Progress Needle Step: Only Messages Progress Needle No Needle or Registration; Switch Off Drive To BE-PV; and Switch Off Drive To BE-PV-Biplane Bulls Eye View-Biplane. Updated syngo Biplane 3D Roadmap testing encompassed: Overlay internal datasets biplane; Biplane Overlay Contour from VRT on Monitor of the Biplane Overlay 3DGraphics on Monitor; Alternative Flow Biplane; 2D/2D 3D Registration OV Basic Flow Overlay Assist Biplane; Overlay Assist Set Active Plane; and 2D2D3D Biplane Registration Modes Segment Selection.
Performance tests were conducted to test the functionality of the "syngo Application Software. These tests have been performed to assess the functionality of the Subject Device. The results of all conducted testing were found acceptable and did not raise any new issues of safety or effectiveness.
All software validation data demonstrates that the Subject Device is as safe and effective when compared to the Predicate Device that is currently marketed for the same intended use.
All updated and new syngo Application software features performance functionality has been tested and validated.
In summary, All conducted test results passed. Software validation data demonstrates that the Subject Device is as safe and effective when compared to the Predicate Device.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
0
October 11, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Siemens Medical Solutions USA, Inc % Patricia Jones Regulatory Affairs Professional 40 Liberty Boulevard MALVERN, PA 19355
Re: K241569
Trade/Device Name: syngo Application Software Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: May 31, 2024 Received: May 31, 2024
Dear Patricia Jones:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
1
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
2
Image /page/2/Figure/2 description: The image shows the word "Sincerely," followed by a signature. The signature appears to be "Sunita Kakar". The signature is written in black ink and is on a white background.
for
Lu Jiang Assistant Director Diagnostic X-ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
3
Indications for Use
Submission Number (if known)
Device Name
syngo Application Software
Indications for Use (Describe)
The syngo Application Software is a medical software for real-time viewing, image manipulation, 3Dvisualization, communication, and storage of medical images and data on exchange media. It is used for diagnostic image viewing and postprocessing and postprocessing during interventional procedures.
The syngo Application Software can be deployed on independent hardware such as a stand-alone diagnostic review, postprocessing, and reporting workstation. It can also be configured within a network to send and receive DICOM data.
Furthermore, the syngo Application Software can be deployed on systems of the Siemens Healthineers Angiography system family.
It provides image guided solutions in the operating room, for image guided surgery, by image fusion and by navigation systems, image guided solutions in interventional cardiology and electrophysiology and image guided solutions for interventional oncology, interventional radiology, and interventional neuroradiology.
The syngo Application Software can also be combined with fluoroscopy systems or radiographic systems.
The syngo Application Software can be configured with a variety of syngo or Windows-based software options, which are intended to assist the physician in diagnosis, treatment planning and treatment control. It includes commercially available postprocessing techniques and OEM options.
Procedures that can be performed include: minimally invasive surgical procedures and minimally invasive tumor treatment.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4
Image /page/4/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.
510(k) Summary: syngo Application Software 510(k) #: K241569
Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard. Malvern, PA 19355
Date of Preparation: October 11, 2024
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
-
- General Information: lmporter / Distributor: Siemens Medical Systems USA, Inc. 40 Liberty Boulevard, Malvern, PA 19355 Establishment Registration Number: 2240869
Manufacturing Site: Siemens Healthcare GmbH Siemensstr. 1 91301 Forchheim, Germany Establishment Registration Number: 3004977335
- General Information: lmporter / Distributor: Siemens Medical Systems USA, Inc. 40 Liberty Boulevard, Malvern, PA 19355 Establishment Registration Number: 2240869
-
- Contact Person:
Ms. Patricia D Jones Regulatory Affairs Professional Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Phone: (678) 575-8832 Email: patricia.jones@siemens-healthineers.com
- Contact Person:
3. Device Name and Classification:
Trade Name: Classification Name: Classification Panel: Classification Regulation: Device Class: Product Code:
syngo Application Software Medical Image Management and Processing System Radiology 21 CFR §892. 2050 Class II LLZ
-
- Legally Marketed Predicate Device Trade Name: syngo Application Software 510(k) Clearance K230955 Clearance Date 12/20/2023
5
Image /page/5/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the text is a graphic of orange dots arranged in a circular pattern.
Classification Name: Classification Panel: Classification Regulation: Device Class: Product Code: Recall Information:
Medical Image Management and Processing System Radiology 21 CFR §892. 2050 Class II LLZ This predicate device has not been the subject of any design-related recalls.
5. Device Description:
The "syngo Application Software" is medical diagnostic software for real-time viewing, diagnostic review, post-processing image manipulation, communication, reporting, and storage of medical images and data on exchange media. It provides imageguided solutions in the operating room, for image-quided surgery, by Image Fusion and by navigation systems, image-guided solutions in interventional cardiology and electrophysiology, and image-guided solutions for interventional oncology, interventional radiology, and interventional neuroradiology. It can be deployed with a variety of syngo or Windows-based software options, which are intended to assist the physician in the evaluation of digital radiographic examinations, including diagnosis and/or treatment planning.
Siemens "syngo Application Software" is designed to work with digital radiographic, fluoroscopic, interventional, and angiographic systems.
Siemens Medical Solutions USA, Inc. hereby submits this Traditional 510(k) to request clearance to market the following modifications:
-
- Updated Software VE30A to VE40C
- A. Updated myNeedle Guide "Progress Needle" steps with an additional "couple A+B" button (Software feature for ARTIS icono biplane only)
- B. Updated syngo Biplane 3D Roadmap feature
-
- Product Claims List
The "syngo Application Software" may be installed either on Siemens released PC hardware, on Siemens X-ray systems, or on Siemens angiography systems. The "syngo Application Software" is within the same classification and the intended use and the general Indications for Use Statement for Siemens' Medical Image Management and Processing System remains the same except for a minor grammatical change (inclusion of "Healthineers").
6. Indications for Use:
The syngo Application Software is a medical software for real-time viewing, image manipulation, 3D visualization, communication, and storage of medical images and data on exchange media. It is used for diagnostic image viewing and postprocessing and for viewing and postprocessing during interventional procedures.
6
Image /page/6/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern.
The syngo Application Software can be deployed on independent hardware such as a stand-alone diagnostic review, postprocessing, and reporting workstation. It can also be configured within a network to send and receive DICOM data.
Furthermore, the syngo Application Software can be deployed on systems of the Siemens Healthineers Angiography system family. It provides image-quided solutions in the operating room, for image-quided surgery, by image fusion and by navigation systems, image-quided solutions in interventional cardiology and electrophysiology, and imageguided solutions for interventional oncology, interventional radiology, and interventional neuroradiology.
The syngo Application Software can also be combined with fluoroscopy systems or radiographic systems.
The syngo Application Software can be configured with a variety of syngo or Windowsbased software options, which are intended to assist the physician in the diagnosis, treatment planning, and treatment control. It includes commercially available postprocessing techniques and OEM options.
Procedures that can be performed include minimally invasive surgical procedures and minimally invasive tumor treatment.
7. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device:
The subject device has the same intended use as the predicate device. Besides the proposed device modifications, the Subject Device has the same functionality and technology. Therefore the subject device is considered substantially equivalent to the commercially available Siemens' "syngo Application Software" (VE30).
All software components of the subject device are the same as the ones from the predicate device except for the new optional software applications as presented in the table below:
| Modifications | Subject Device
syngo Application Software | Predicate Device
syngo Application Software (VE30A)
K230955 | Comparison Results |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| New System
Software/
Changes | 1. Updated syngo Application
Software from VE30A to
VE40C
A. Updated myNeedle
Guide “Progress
Needle” steps with an
additional “couple A+B”
button (Software feature | syngo Application
Software
syngo myNeedle Guide | Modified:
Software VE40C was
updated to support
modifications 1A &B. All
Software modifications
conform to “Guidance
for the Content of
Premarket Submissions
for Device Software |
7
Image /page/7/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a cluster of orange dots.
| Modifications | Subject Device
syngo Application Software | Predicate Device
syngo Application
Software (VE30A)
K230955 | Comparison
Results |
|---------------|------------------------------------------------|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | for ARTIS icono biplane
only) | | Functions"". Provided in
the software section of
eStar is all required
software testing per
software
guidance
"Basic Level". |
| | Updated syngo
Biplane 3D Roadmap
feature | syngo 3D Roadmap | Provided in the Non-
Clinical Testing Section
of eStar are Bench Test
Summaries.
Bench
tests were conducted
and were found
acceptable and did not
raise any new issues of
safety or effectiveness. |
| | | | All software validation
data demonstrates that
the Subject device is as
safe and effective when
compared to the
Predicate Device that is
currently marketed for
the same intended use.
All test results met all
acceptance criteria. |
The Subject Device modifications do not alter the fundamental scientific technology from the 510(k) cleared predicate device Siemens' "syngo Application Software" (VE30A), K230955.
8. Nonclinical Performance Testing:
Non-clinical tests were conducted for the "syngo Application Software" during product development.
| Recognition
Number | Product
Area | Title of Standard | Reference
Number
and Date | Publication
Date | Standards
Development
Organization |
|-----------------------|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|---------------------|------------------------------------------|
| 5-125 | General | Medical devices - Application of
risk management to medical
devices | 14971 | 2019 | ISO |
| 5-134 | General | Medical devices - Symbols to
be used with medical device
labels, labeling, and | 15223-1 | 2021 | ISO |
| Recognition
Number | Product
Area | Title of Standard | Reference
Number
and Date | Publication
Date | Standards
Development
Organization |
| | | information to be supplied -
Part 1: General requirements | | | |
| 5-134 | General | Medical devices - Symbols to
be used with medical device
labels, labeling, and
information to be supplied -
Part 1: General requirements | 15223-1 | 2021 | ISO |
| 13-97 | Software/
Informatics | Health software - Part 1:
General requirements for
product safety | 82304-1 | 2016 | IEC |
| 13-79 | Software/
Informatics | Medical Device Software -
Software life-cycle process | 62304 | 2015 | IEC |
| 5-129 | General | Medical devices - Part 1
Application of usability
engineering to medical devices | 62366-1 | 2020 | IEC |
| 13-38 | Software/
Informatics | Application of risk management
for IT - networks incorporating
medical devices - Part 1:
Roles, responsibilities, and
activities | 80001-1 | 2010 | IEC |
| 13-122 | Software/
Informatics | Health software and health IT
systems safety, effectiveness
and security - Part 5-1: Security
- Activities in the product life
cycle | 81001-5-1 | 2021 | IEC |
| 12-349 | Radiology | Digital Imaging and
Communications in Medicine
(DICOM) Set | PS 3.1 | 2022 | NEMA |
| 13-96 | Software/
Informatics | Standard for Safety, Standard
for Software Cybersecurity
Network-Connectable
Products, Part 1: General
Requirements | 2900-1 | 2017 | ANSI UL |
| 13-104 | Software/
Informatics | Standard for Safety, Software
Cybersecurity for Network-
Connectable Products, Part 2-
1: Particular Requirements for
Network Connectable | 2900-2-1 | 2017 | ANSI UL |
Siemens claims conformance to the following performance standards:
8
Image /page/8/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" are a series of orange dots arranged in a circular pattern.
9
Image /page/9/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots in a circular pattern.
| Recognition
Number | Product
Area | Title of Standard | Reference
Number
and Date | Publication
Date | Standards
Development
Organization |
|-----------------------|-----------------|--------------------------------------------------|---------------------------------|---------------------|------------------------------------------|
| | | Components of Healthcare and
Wellness Systems | | | |
The modifications described in this Premarket Notification are supported with verification and validation testing.
Software Verification and Validation:
Software documentation for a Basic Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Device Software Functions" issued on June 14, 2023, and "Off-The-Shelf Software Use in Medical Devices" is also included as part of this submission. Non-clinical tests were conducted on "syngo Application Software" during product development.
The Risk analysis was completed, and risk control was implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.
The Human Factor Usability Validation showed no safety-relevant functions that need to be validated with a summative usability validation according to the IEC and FDA Guidelines. "syngo Application Software" is safe and effective for intended users, uses, and use environments through the design control verification and validation process. No further risk mitigations are necessary.
Siemens conforms to the cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse, or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Provided in this submission is a cybersecurity statement that considers IEC 80001-1:2010. The responsibility for compliance with IEC 80001-1:2010 is the hospital. Provided in the Software Section is the required cybersecurity information.
Summary of Performance Testing Data:
Tests were conducted in support of the new/updated functionalities for the "syngo Application Software": Updated myNeedle Guide testing encompassed Biplane-Couple-A-B, Invoke Progress Needle Step: Only Messages Progress Needle No Needle or Registration; Switch Off Drive To BE-PV; and Switch Off Drive To BE-PV-Biplane Bulls Eye View-Biplane. Updated syngo Biplane 3D Roadmap testing encompassed: Overlay internal datasets biplane; Biplane Overlay Contour from VRT on Monitor of the Biplane Overlay 3DGraphics on Monitor; Alternative Flow Biplane; 2D/2D 3D Registration OV Basic Flow Overlay Assist Biplane; Overlay Assist Set Active Plane; and 2D2D3D Biplane Registration Modes Segment Selection. After transfer and integration, the SW/System
10
Image /page/10/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a cluster of orange dots.
will be carried out within the framework of SOP406 "Software Development" and cover compliance with IEC 62304.
Performance tests were conducted to test the functionality of the "syngo Application Software. These tests have been performed to assess the functionality of the Subject Device. The results of all conducted testing were found acceptable and did not raise any new issues of safety or effectiveness.
All software validation data demonstrates that the Subject Device is as safe and effective when compared to the Predicate Device that is currently marketed for the same intended use.
Performance Verification/Validation & Test Report Summaries:
All updated and new syngo Application software features performance functionality has been tested and validated.
In summary, All conducted test results passed. Software validation data demonstrates that the Subject Device is as safe and effective when compared to the Predicate Device.
The comparison of technological characteristics, non-clinical performance data, Human Factor Usability data, and software validation data demonstrates that the Subject Device is as safe, and effective when compared to the Predicate Device that is currently marketed for the same intended use.
9. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device safely and effectively.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification, and validation testing. Furthermore, the operators are healthcare professionals familiar with and responsible for the evaluation and post-processing of medical images.
10. Conclusion as to Substantial Equivalence:
The predicate device "syngo Application Software" (VE30A), K230955, was cleared based on non-clinical supportive information. Similar non-clinical test results demonstrate that the subject device's "syngo Application Software" acceptance criteria are adequate for the intended use of the device. The comparison of technological characteristics, nonclinical performance data, clinical images, and software validation data demonstrates that the subject device is as safe, and effective when compared to the predicate device that is currently marketed for the same intended use.