K Number
K241565
Date Cleared
2025-02-24

(269 days)

Product Code
Regulation Number
880.6850
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Sterilization Package/Reel is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows: · Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 30 minutes. · Ethylene Oxide (EO) with a concentration of 735 mg/L at 55°C (131°F) and 50% to 80% relative humidity for 60 minutes. Aeration time of 8 hours at 60℃(140°F). Furthermore, the Sterilization Package/Reel maintains the enclosed devices up until 6 months post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the Sterilization Package/Reel has undergone either a steam or EO sterilization process. The Sterilization Package/Reel is offered 3 types in the following: Self -Seal Sterilization Pouch; Flat Sterilization Reel; Gusseted Sterilization Reel.

Device Description

The medical devices are inserted into the Sterilization Package/Reel,sealed, and then sterilized. After completion of the sterilization process, the Pouch/Roll main- tains the sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 6 months post Steam or EO gas sterilization. The Self-seal Sterilization Package permits the sealing of the pouch without the need of heat-sealing equipment, while the Heat-Seal Sterilization Reel are heat-sealed prior to processing in the steam/or EO Sterilization. The chemical indicators ink printed on the "medical dialysis paper" will exhibit a color change after the Package/Reel is exposed to steam or ethylene oxide gas. The color of the Chemical Indicator changes from blue to black when exposed to Steam. And the color changes from pink to yellow, when exposed to EO gas. The Chemical Indicator which is Type 1 Process Indicator as categorized by ISO 11140-1:2014 offers an addition way to verity processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. The Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to Steam/or EO gas.

AI/ML Overview

I am sorry, but the provided text describes a 510(k) premarket notification for a medical device called "Sterilization Package/Reel." This document focuses on the substantial equivalence of the device to existing predicate devices, primarily through non-clinical bench testing and comparisons of technological characteristics.

The information you requested, relating to acceptance criteria and studies proving an AI/Software as a Medical Device (SaMD)'s performance (such as sample size for test/training sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth establishment), is not present in the provided text.

The document discusses validation of sterilization processes (ISO 11135, ISO 17665-1), material compatibility, package integrity, and biocompatibility, which are standard tests for physical sterilization packaging, not for AI device performance.

Therefore, I cannot fulfill your request for this specific information based on the text provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format. The logo is simple and professional, and it is easily recognizable.

February 24, 2025

Anqing Clean Dental Instrument Technology Co., Ltd. % Wang Boyle General Manager Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai. 200120 China

Re: K241565

Trade/Device Name: Sterilization Package/Reel Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG, JOJ Dated: January 20, 2025 Received: January 24, 2025

Dear Wang Boyle:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Stephen A. Anisko -S

Digitally signed by Stephen A. Anisko -S Date: 2025.02.24 20:03:28 -05'00'

for: Christopher K. Dugard, M.S. Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241565

Device Name Sterilization Package/Reel

Indications for Use (Describe)

The Sterilization Package/Reel is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows: · Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 30 minutes.

· Ethylene Oxide (EO) with a concentration of 735 mg/L at 55°C (131°F) and 50% to 80% relative humidity for 60 minutes. Aeration time of 8 hours at 60℃(140°F).

Furthermore, the Sterilization Package/Reel maintains the enclosed devices up until 6 months post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the Sterilization Package/Reel has undergone either a steam or EO sterilization process.

The Sterilization Package/Reel is offered 3 types in the following: Self -Seal Sterilization Pouch; Flat Sterilization Reel; Gusseted Sterilization Reel.

The following table (Table 1) lists the model numbers of the Sterilization Package/Reel by type, model, dimensions, and content/max. load (lbs.):

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Specification Size (WXL)Content/Max Load(lbs)
Product NameItemInchDimensionMetalPlasticGauze/Linens
KLDT-901653.5"x6.5"90mmx165mm0.120.110.1
KLDT-30543012"x17"305mmx430mm1.101.090.66
KLDT-702302.75"x9"70mm×230mm0.090.080.07
KLDT-1352803⅛"x11"135mmx280mm0.500.470.46
KLDT-901353.5"x5.2590mmx135mm0.100.090.08
SelfSealingSterilizationPouchKLDT-702602.75"x10"70mm×260mm0.100.090.08
KLDT-902603.5"x10"90mmx260mm0.150.140.12
KLDT-1352603⅛"x10"135mmx260mm0.480.460.45
KLDT-25037010"x14.5"250mmx370mm1.061.050.62
KLDT-1903607.5"x14"190mmx360mm0.600.580.55
KLDT-571302.25"x5"57mm×130mm0.040.030.005
KLDT-571002.25"x4"57mm×100mm0.030.020.004
KLDT-1353303⅛"x13"135mmx330mm0.550.530.50
KLDT-902303.5"x9"90mmx230mm0.140.130.11
FlatSterilizationReelKLDT-2002008"200mm×200m0.600.580.55
KLDT-30020012"300mm×200m1.041.030.60
KLDT-752003"75mm×200m0.090.050.004
KLDT-1002004"100mm×200m0.140.130.11
KLDT-502002"50mm×200m0.020.010.003
KLDT-40020016"400mm×200m1.081.070.64
KLDT-25020010"250mm×200m0.800.780.58
KLDT-35020014"350mm×200m1.061.050.62
KLDT-552002.25"55mm×200m0.030.020.003
KLDT-1502006"150mm×200m0.550.530.50
GussetedSterilizationReelKLDT-751003"75mm×30mm×100m0.090.050.004
KLDT-35010014"350mm×80mm×100m1.061.050.62
KLDT-1501006"150mm×50mm×100m0.550.530.50
KLDT-40010016"400mm×80mm×100m1.081.070.64
KLDT-25010010"250mm×65mm×100m0.800.780.58
KLDT-30010012"300mm×75mm×100m1.041.030.60
KLDT-2001008"200mm×55mm×100m0.600.580.55

Table 1. The model numbers of Sterilization Package/Reel (Type, Model and Dimension and Content/Max. Load)

{5}------------------------------------------------

KLDT-100100A 11100mm×50mm×100mCALI0 40V. IVCara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara Cara CaraVill

{6}------------------------------------------------

510(k) Summary K241565

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.

1.0 Submitter's Information

Name:Anqing Clean Dental Instrument Technology Co.,LTD.
Address:Floor 2, D4-A,Jiabao Industrial Park.Yangtze River BridgeEconomic Development Zone ,246000 Anqing City, AnhuiProvince,China
Tel:0086-556-5577869
Contact:Wu Yiming

Designated Submission Correspondent

Contact:Mr. Boyle Wang
Name:Shanghai Truthful Information Technology Co., Ltd.
Address:Room 1801, No. 161 East Lujiazui Rd., Pudong Shanghai200120 China
Tel:+86-21-50313932
Email:Info@truthful.com.cn

Date Submitted: Jan.20,2025

2.0 Device Information

Trade/Device name:Sterilization Package/Reel
Common name:Sterilization Package/Reel
Classification name:Sterilization Wrap;
Classification Product Code:FRG,JOJ
Regulation number:21 CFR880.6850
Classification:Class II
Panel:General Hospital

3.0 Predicate Device Information

Predicate Device:

Manufacturer:Sterileright Packaging Mfg., Inc.
Trade/Device Name:SterileRight Sterilization Pouch and Roll

{7}------------------------------------------------

510(k) number: K212338

4.0 Device Description

The medical devices are inserted into the Sterilization Package/Reel,sealed, and then sterilized. After completion of the sterilization process, the Pouch/Roll main- tains the sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 6 months post Steam or EO gas sterilization.

The Self-seal Sterilization Package permits the sealing of the pouch without the need of heat-sealing equipment, while the Heat-Seal Sterilization Reel are heat-sealed prior to processing in the steam/or EO Sterilization.

The chemical indicators ink printed on the "medical dialysis paper" will exhibit a color change after the Package/Reel is exposed to steam or ethylene oxide gas. The color of the Chemical Indicator changes from blue to black when exposed to Steam. And the color changes from pink to yellow, when exposed to EO gas.

The Chemical Indicator which is Type 1 Process Indicator as categorized by ISO 11140-1:2014 offers an addition way to verity processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. The Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to Steam/or EO gas.

5.0 Indication for Use Statement

The Sterilization Package/Reel is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:

· Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 30 minutes.

• Ethylene Oxide (EO) with a concentration of 735 mg/L at 55° C (131° F) and 50% to 80% relative humidity for 60 minutes. Aeration time of 8 hours at 60°C(140°F).

Furthermore, the Sterilization Package/Reel maintains the enclosed devices up until 6 months post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the Sterilization Package/Reel has undergone either a steam or EO sterilization process.

The Sterilization Package/Reel is offered 3 types in the following: Self -Seal Sterilization Pouch; Flat Sterilization Reel; Gusseted Sterilization Reel.

The following table (Table 1) lists the model numbers of the Sterilization Package/Reel by type, model, dimensions, and content/max. load (Ibs.):

{8}------------------------------------------------

Product NameItemSpecification Size (WXL)Content/Max Load(lbs)
InchDimensionMetalPlasticGauze/Linens
KLDT-901653.5"x6.5"90mmx165mm0.120.110.1
KLDT-30543012"x17"305mmx430mm1.101.090.66
KLDT-702302.75"x9"70mm×230mm0.090.080.07
KLDT-1352803⅛"x11"135mmx280mm0.500.470.46
KLDT-901353.5"x5.25"90mmx135mm0.100.090.08
SelfSealingSterilizationPouchKLDT-702602.75"x10"70mm×260mm0.100.090.08
KLDT-902603.5"x10"90mmx260mm0.150.140.12
KLDT-1352603⅛"×10"135mmx260mm0.480.460.45
KLDT-25037010"x14.5"250mmx370mm1.061.050.62
KLDT-1903607.5"x14"190mmx360mm0.600.580.55
KLDT-571302.25"x5"57mm×130mm0.040.030.005
KLDT-571002.25"x4"57mm×100mm0.030.020.004
KLDT-1353303⅛"×13"135mmx330mm0.550.530.50
KLDT-902303.5"x9"90mmx230mm0.140.130.11
KLDT-2002008"200mm×200m0.600.580.55
KLDT-30020012"300mm×200m1.041.030.60
KLDT-752003"75mm×200m0.090.050.004
KLDT-1002004"100mm×200m0.140.130.11
FlatSterilizationReelKLDT-502002"50mm×200m0.020.010.003
KLDT-40020016"400mm×200m1.081.070.64
KLDT-25020010"250mm×200m0.800.780.58
KLDT-35020014"350mm×200m1.061.050.62
KLDT-552002.25"55mm×200m0.030.020.003
KLDT-1502006"150mm×200m0.550.530.50
GussetedSterilizationReelKLDT-751003"75mm×30mm×100m0.090.050.004
KLDT-35010014"350mm×80mm×100m1.061.050.62
KLDT-1501006"150mm×50mm×100m0.550.530.50
KLDT-40010016"400mm×80mm×100m1.081.070.64
KLDT-25010010"250mm×65mm×100m0.800.780.58
KLDT-30010012"300mm×75mm×100m1.041.030.60
KLDT-2001008"200mm×55mm×100m0.600.580.55
KLDT-1001004"100mmx50mmx100m0.140.130.11

Table 1. The model numbers of Sterilization Package/Reel (Type, Model and Dimension and Content/Max. Load)

{9}------------------------------------------------

6.0 Summary of Non-Clinical Testing

Summary of non-clinical and performance testing bench testing was performed to evaluate the performance and functionality of the subject device against requirement specification. The test results demonstrated the subject device is confirmed to be safe and effective for the intended use.

Test ItemTest MethodologyAcceptance Criteria or End PointTest Results
SterilizationProcess ValidationISO 11135:2014Ethylene oxide: 60 minutes at 55 °C; relativehumidity between 50%- 80%;ethylene oxide concentrationis 735 mg/L, 8 hours aerationtime at 60°C (140°F).SAL=10-6SAL=10-6Pass
ISO 17665-1:2006;ISO TS 17665-2:2009Steam; 4 minutes at 132°C;30 minutes dry time.SAL=10-6SAL=10-6Pass
EO/ECH ResidualsISO 10993-7:2008EO < 4mg/dECH < 9mg/dPass
Package Integrity /MaterialCompatibility/SterilityMaintenanceASTM F 2251-1352μm±10%Pass
ASTMF1140/F1140M-13Burst value > 3.2 Kpa or No BurstPass
ASTM F1929-15The dye solution is no anyleakage across the seal widthof sterile barrier system.(No Infiltration)No InfiltrationPass
ASTM F88/F88M-21Seal strength > 2.5 (N/15mm)Pass
ISO 1924-2:2008Tensile strength:CD $\ge$ 2.2KN/m,MD $\ge$ 4.4KN/mPass
ASTM F2096-11No LeakageNo LeakagePass
DIN 58953-6CFU = 0CFU = 0Pass
ISO 5636-310-50 mL/min·m²20-30 mL/min·m²Pass

Table 2: Performance testing summary – Bench

{10}------------------------------------------------

Chemical IndicatorEfficacy TestingISO 11140-1:2014Steam:The real time aging test
1. Remain stablebefore use based onits shelf life.2. Maintain theendpoint stability of thecolor change afterbeing in the presenceof the sterilant.The color of CI changes fromblue to black, when exposedto Steam;EO:The color of CI changes frompink to yellow, when exposedto EO gas.was carried out thatdemonstrates: the testdevice which exposed toSteam maintain the colorof black, the test groupwhich exposed to EOmaintain the color ofyellow, and the real timeaging test was carried outon the empty SterilizationPackage/Reel thatdemonstrates the testdevice maintain theoriginal color within thestated validity period of 2years.
Shelf-LifeValidation/Shelf Life:2 Years;Shelf Life after Sterilized:6 monthsPass

Biocompatibility Testing

The biocompatibility evaluation for the Sterilization Package/Reel was conducted in accordance with the FDA's Biocompatibility Guidance "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process"

Table 3 Summary test of Biocompatibility Testing

BiocompatibilitytestingISO 10993-5:2009Non-cytotoxicUnder conditions of the study, did notshow potential toxicity to L-929 cells.Pass
ISO 10993-23:2021Non-irritatingUnder the conditions of the study, not anirritant.Pass
ISO 10993-10:2021Non-sensitizingUnder conditions of the study, not asensitizer.Pass

{11}------------------------------------------------

7.0 Technological Characteristic Comparison Table

ItemSubject DeviceK241565Predicate DeviceK212338ComparisonAnalysis
Product NameSterilization Package/ReelSterileRight sterilization pouch and roll----
Product CodeFRG,JOJFRG,JOJSame
Regulation No.21 CFR 880.685021 CFR 880.685021 CFR 880.2800Same
ClassIIIISame
Indication forUseThe Sterilization Package/Reel is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 30 minutes. Ethylene Oxide (EO) with a concentration of 735 mg/L at 55° C (131 ° F) and 50% to 80% relative humidity for 60 minutes. Aeration time of 8 hours at 60°C (140°F). Furthermore, the Sterilization Package/Reel maintains the enclosed devices up until 6 months post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the Sterilization Package/Reel has undergone either a steam or EO sterilization process.The SterileRight Sterilization Pouch and Roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:Gravity steam at 121°C (250°F) for 30 minutes; Drying time of 25 minutes. Pre-vacuum steam at 132°C (270°F) for 4 minutes; Drying time of 20 minutes. Ethylene Oxide (EO) with a concentration of 735 mg/L at 55°C (131°F) and 50% to 80% relative humidity for 60 minutes. Aeration time of 8 hours at 60°C (140°F). The SterileRight provide sterilization pouch and roll made with Paper/Film or Tyvek® /Film. The SterileRight sterilization pouch and roll which are made with Paper maintains the sterility of the enclosed devices for up to 6 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 3 years from the date ofSimilar
The Sterilization Package/Reel isoffered 3 types in the following:Self -Seal Sterilization Pouch;Flat Sterilization Reel;Gusseted Sterilization Reel.manufacture. The pouch'sexternal chemical ink indicatorsare designed to indicate to theuser that the pouch hasundergone either a steam or EOsterilization process.The SterileRight sterilizationpouch and roll which are madewith Tyvek® is for EO gas
sterilization only. It also maintainsthe sterility of the encloseddevices for up to 6 monthspost EO gas sterilization, and
before sterilization hasamaximum shelf life of 3 yearsfrom the date of manufacture.
MaterialCompositionComposed of medical dialysispaper and transparent coloredPET/CPP composite film , EO andSteam Process Indicator, Print Ink,adhesiveMedical Grade Paper, CPP, PET,adhesive,EO and Steam ProcessIndicator, Print Ink.Tyvek®, PET, PE, adhesive.Similar
SterilizationCycles• Steam Sterilization at 132°C(270°F) for 4 minutes; Drying timeof 30 minutes.• Ethylene Oxide (EO) with aconcentration of 735 mg/L at 55°C (131 ° F) and 50% to 80%relative humidity for 60 minutes.Aeration time of 8 hours at 60°C(140°F).• Gravity steam at 121°C (250°F)for 30 minutes; Drying time of 25minutes.• Pre-vacuum steam at 132°C(270°F) for4 minutes; Drying timeof 20 minutes.• Ethylene Oxide (EO) with aconcentration of 735 mg/L at55°C (131°F) and 50% to 80%relative humidity for 60 minutes.Aeration time of 8hours at 60°C(140°F).Similar
Configuration/DimensionVarious Size, Heat Sealing andSelf SealingVarious Size, Heat Sealing andSelf SealingSame
SterilantPenetrationEfficacyThe test meet the requirement ofSAL 10-6The test meet the requirement ofSAL 10-6Same
ChemicalIndicator(CI)Functionalityand EndpointThe sterilant penetrated throughthe pouch configuration andaffected the CI color change tothe endpoint colorThe sterilant penetrated throughthe pouch configuration andaffected the CI color change to theendpoint colorSame
Device Designof Steam ClThe color of Chemical Indicatorchanges from Blue toThe color of Chemical Indicatorchanges from Pink toSimilar
Black, when exposed to SteamBrown/Black, when exposed to Steam
Device Design of EO gas CIThe color of Chemical Indicator changes from Pink to Yellow, when exposed to SteamThe color changes from Blue to Yellow/Brown, when exposed to EO gasSimilar
Thickness Variations (mm) ASTM F 2251PassedPassedSame
Tensile strength ISO 1924-2PassedPassedSame
Burst Strength (kPa) ASTM F1140PassedPassedSame
Bubble Leak Test ASTM D 3078 or ASTM-F 2096PassedPassedSame
Seal Peel Test (N/15mm) ASTM F88/F88M ; ISO 11607-1PassedPassedSame
Dye penetration Test ASTM F1929PassedPassedSame
Microbial Barrier Test DIN 58953-6PassedPassedSame
Air permeance testing ISO 5636-3PassedPassedSame
End point stability testing resultsThe color of chemical indicator for EO sterilization indicator ink is Pink, and the color of chemical indicator for steam sterilization indicator ink is Blue after 2 year shelf life before sterilization.The color of chemical indicator for EO sterilization indicator ink is Yellow, and the color of chemical indicator for steam sterilizationThe color of chemical indicator for EO sterilization indicator ink is Blue, and the color of chemical indicator for steam sterilization indicator ink is Pink after 2 year shelf life before sterilization.The color of chemical indicator for EO sterilization indicator ink is Yellow/Brown , and the color of chemical indicator for steamSimilar
indicator ink is Blue after EO sterilized and 6 months shelf life.sterilization indicator ink is Pink after EO sterilized and 6 months shelf life
The color of chemical indicator for EO sterilization indicator ink is Pink, and the color of chemical indicator for steam sterilization indicator ink is Black after steam sterilized and 6 months shelf life.The color of chemical indicator for EO sterilization indicator ink is Bue, and the color of chemical indicator for steam sterilization indicator ink is Brown/Black after steam sterilized and 6 months shelf life.Similar
Maintenance of Sterility6 months6 monthsSame
Shelf Life2 years from date of manufacture for EO and Steam Indicators3 years from date of manufacture for EO and Steam IndicatorsDifferent
BiocompatibilityConform with ISO10993-1 (ISO10993-5, ISO10993-10)Conform with ANSI/AAMI/ISO 10993-10Same

Table 4- Comparison of Technology Characteristics

{12}------------------------------------------------

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

The technological characteristics of the subject device are identical to those of predicate device. The subject device has the same basic design as the predicate device.

The comparison between the subject and predicate devices is based on the following:

  • Same intended use
  • · Same indications for use
  • · Similar material types that meet ISO 10993 biocompatibility requirements
  • · Same sterilization methods (EO and Steam sterilization process)
  • Same fundamental technology/principal of operation/user interface

The sterilization parameters of the subject device are little different with those of the predicate device, but the EO sterilization validation results demonstrate the subject device fully meet the requirements of ISO 11135 and the steam sterilization validation results fully meet the requirements of ISO 17665-1.

8.0 Summary of Clinical Testing

No clinical study is included in this submission.

9.0 Conclusion

The conclusions drawn from the non-clinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the predicate device K212338.

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).