K Number
K241444

Validate with FDA (Live)

Date Cleared
2025-01-21

(244 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Biliary Pancreaticobiliary Scope System is indicated for use in diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts. The Biliary Pancreaticobiliary Scope System comprises two components: the Single-use Biliary Pancreaticobiliary Scope and the Image Processor. The Single-use Biliary Pancreaticobiliary Scope is intended to provide direct visualization, illumination and to guide accessory devices for diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts. The Image Processor is intended to receive, process, and output images from the Single-use Biliary Pancreaticobiliary Scope for diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts.

Device Description

The Biliary Pancreaticobiliary Scope System comprises two components: the Single-use Biliary Pancreaticobiliary Scope (model:CL-A, CL- B), and the Image Processor (model: HUV-02). The Single-use Biliary Pancreaticobiliary Scope is a flexible scope, introduced into the pancreatico-biliary system via a duodenoscope. The Single-use Biliary Pancreaticobiliary Scope comprises a handle, an insertion tube, and a connection cable.

AI/ML Overview

I am sorry, but the provided text does not contain specific acceptance criteria or a detailed study description with performance metrics for a device to meet those criteria. The document is a 510(k) clearance letter from the FDA for a "Biliary Pancreaticobiliary Scope System."

Here's what I can tell you based on the provided text, and what is missing:

Information Present in the Document:

  • Device Name: Biliary Pancreaticobiliary Scope System
  • Intended Use/Indications for Use: The system is indicated for use in diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system, including the hepatic ducts. It provides direct visualization, illumination, and guidance for accessory devices.
  • Non-Clinical Tests Summary: The document lists various non-clinical tests performed, including:
    • Biocompatibility (Cytotoxicity, Intracutaneous reactivity, Sensitization, Acute systemic toxicity, Pyrogen)
    • Sterilization and shelf life testing (ISO 11135:2014, ISO 10993-7:2008, ASTM F1980-21, ISO 11607-1.2019, ASTM F 1886/F1886M-16, ASTM F88F88M-15, ASTM F 1929-15)
    • Electrical safety and electromagnetic compatibility (EMC) (IEC 60601-1, IEC60601-2-18, IEC 60601-1-2)
    • Software Verification and Validation Testing (according to FDA guidance)
    • Bench performance testing:
      • Optical performance (ISO 8600 series)
      • Color performance (color reproduction), optical performance (resolution, depth of field, image intensity uniformity), SNR, dynamic range, frame frequency, and system delay test compared with the predicate device.
      • Mechanical testing (water resistance, deflection, working channel system leakage, articulation reliability, attachment to duodenoscope reliability, irrigation pump compatibility)
    • Animal testing (referring to FDA Guidance-General Considerations for Animal Studies Intended to Evaluate Medical Devices)
  • Clinical Testing: "Not Applicable." This explicitly states that clinical testing was not performed for this submission.
  • Predicate Device: SpyGlass DS Direct Visualization System (K142922)

Information NOT Present in the Document (and therefore cannot be provided by me):

  1. A table of acceptance criteria and the reported device performance: While various tests are listed, the document does not provide specific quantitative acceptance criteria (e.g., "resolution must be X lp/mm") or the measured performance results of the device against these criteria. It only states that the device "complies with" certain standards or that tests were "performed."
  2. Sample size used for the test set and the data provenance: This information is not provided for any of the tests, nor is there a "test set" in the context of clinical or AI performance. The animal test simply states it was conducted, but no details on size or provenance.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This is irrelevant as no clinical or AI performance study with human expert review is described.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Irrelevant due to no clinical or AI performance study.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI-assisted device, and no MRMC study is mentioned.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is not an AI algorithm.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable for the non-clinical tests described. For the animal study, the ground truth would be based on observations during the animal procedure, but no details are given.
  8. The sample size for the training set: Not applicable, as this is not an AI device with a training set.
  9. How the ground truth for the training set was established: Not applicable.

In summary, the provided FDA clearance letter focuses on establishing substantial equivalence through non-clinical performance, safety, and compatibility testing, rather than a detailed clinical effectiveness study with acceptance criteria and performance data. The phrase "Performance testing and compliance with voluntary standards demonstrate that the proposed Biliary Pancreaticobiliary Scope System is substantially equivalent to the predicate device" serves as the overall conclusion for meeting FDA's requirements for this 510(k) pathway.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 21, 2025

Shenzhen HugeMed Medical Technical Development Co., Ltd. Cathy Shi Regulatory Engineer 401, 501, Building 4, Haizhi Technology Park, Fortis, No. 17 Bulan Road,Xialilang Community, Nanwan Street, Longgang Dist Shenzhen, Guangdong 518112 China

Re: K241444

Trade/Device Name: Biliary Pancreaticobiliary Scope System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FBN, KQM, NTN, FET Dated: May 15, 2024 Received: December 20, 2024

Dear Cathy Shi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen -S

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K241444

Device Name

Biliary Pancreaticobiliary Scope System

Indications for Use (Describe)

The Biliary Pancreaticobiliary Scope System is indicated for use in diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts.

The Biliary Pancreaticobiliary Scope System comprises two components: the Single-use Biliary Pancreaticobiliary Scope and the Image Processor.

The Single-use Biliary Pancreaticobiliary Scope is intended to provide direct visualization,

illumination and to quide accessory devices for diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts.

The Image Processor is intended to receive, process, and output images from the Single-use Biliary Pancreaticobiliary Scope for diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary
510(k) #:K241444Prepared on: 2025-01-17
Contact Details21 CFR 807.92(a)(1)
Applicant NameShenzhen HugeMed Medical Technical Development Co., Ltd.
Applicant Address401, 501, Building 4, Haizhi Technology Park, Fortis, No. 17, Bulan Road Xialilang Community, Nanwan Street, Longgang District Shenzhen Guangdong 518112 China
Applicant Contact Telephone+86 13534156592
Applicant ContactMs. Cathy Shi
Applicant Contact Emailcathy.shi@hugemed.cn
Device Name21 CFR 807.92(a)(2)
Device Trade NameBiliary Pancreaticobiliary Scope System
Common NameEndoscope and accessories
Classification NameCholedochoscope And Accessories, Flexible/Rigid
Regulation Number876.1500
Product Code(s)FBN, KQM, NTN , FET
Legally Marketed Predicate Devices21 CFR 807.92(a)(3)
Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K142922SpyGlass DS Direct Visualization SystemFBN
Device Description Summary21 CFR 807.92(a)(4)
The Biliary Pancreaticobiliary Scope System comprises two components: the Single-use Biliary Pancreaticobiliary Scope (model:CL-A, CL- B), and the Image Processor (model: HUV-02). The Single-use Biliary Pancreaticobiliary Scope is a flexible scope, introduced into the pancreatico-biliary system via a duodenoscope. The Single-use Biliary Pancreaticobiliary Scope comprises a handle, an insertion tube, and a connection cable.
Intended Use/Indications for Use21 CFR 807.92(a)(5)
The Biliary Pancreaticobiliary Scope System is indicated for use in diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts. The Biliary Pancreaticobiliary Scope System comprises two components: the Single-use Biliary Pancreaticobiliary Scope and the Image Processor. The Single-use Biliary Pancreaticobiliary Scope is intended to provide direct visualization, illumination and to guide accessory devices for diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts. The Image Processor is intended to receive, process, and output images from the Single-use Biliary Pancreaticobiliary Scope for diagnostic and therapeutic applications during endoscopic procedures in the pancreatico-biliary system including the hepatic ducts.
Indications for Use Comparison21 CFR 807.92(a)(5)
The indications for use of the subject and predicate device are the same.

{5}------------------------------------------------

Technological Comparison

The devices have the same sterilization method, energy source, principle of operation and similar design with same parameters such as FOV, DOV, max. outer diameter of insertion, deflection angle, minimum instrument channel width, work length etc. They have similar technological characteristics as below. These new question on safety and effectiveness of the proposed device.

  1. The proposed image processor has more types of signal output than the predicate device to give physicians more usage options. Therefore, this difference will not raise new question on safety and effectiveness of the proposed device.

Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Biocompatibility testing

Biocompatibility of the Single-use Biliary Pancreaticobiliary Scope was evaluated in accordance with the FDA quidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The following tests were performed, as recommended:

Cytotoxicity

Intracutaneous reactivity

& Sensitization

& Acute systemic toxicity

& Pyrogen

Sterilization and shelf life testing

The Single-use Biliary Pancreaticobiliary Scope is provided sterile and its shelf-life is 2 years.

Sterilization Process has been validated accordance with ISO 11135:2014.

EO/ECH residual test was performed according to ISO 10993-7:2008.

The shelf life is determined based on optical testing and product performance testing after according to ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.

Package validation was conducted according to ISO 11607-1.2019 and ASTM F 1886/F1886M-16, ASTM F88F88M-15, ASTM F 1929-15.

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Biliary Scope System. The system complies with the IEC 60601-1 and IEC60601-2-18 for safety and the IEC 60601-1-2 for EMC.

Software Verification and Validation Testing

Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Device Software Functions".

Bench performance testing

The following bench tests were performed:

  1. Optical performance testing according to ISO 8600 series.

  2. Color performance (color reproduction), optical performance (resolution, depth of field and image intensity uniformity), SNR and dynamic range frame frequency and system delay test compared with the predicate device.

  3. Mechanical testing including water resistance, deflection and working channel system leakage, articulation reliability,

attachment to duodenoscope reliability, irigation pump compatibility, irigation pump compatibility.

Animal testing

Animal testing has been conducted referring to FDA Guidance-General Considerations for Animal Studies Intended to Evaluate Medical Devices.

Clinical testing Not Applicable.

Conclusion

Performance testing and compliance with voluntary standards demonstrate that the proposed Biliary Pancreaticobiliary Scope System is substantially equivalent to the predicate device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.