K Number
K241431
Date Cleared
2025-01-13

(237 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood Pressure (systolic and diastolic) of adults and adolescents over 12 years of age with circumference ranging from 22cm to 24cm.

Device Description

The Arm-type Fully Automatic Digital Blood Pressure Monitor (BPM) series is automatic, non-invasive, blood pressure measurement system for over-the-counter (OTC) use at home and clinical environment . The systolic and diastolic pressures are determined using the oscillometric method, where the cuff is inflated with an integral controllable piezoelectric pump and deflates via an electric automatic rapid deflation valve. During inflation measurements, an electric pump within the main unit slowly inflates the arm cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the microprocessor to determine systolic pressure, and diastolic pressure. The cuff can measure pressure range from 0 to 299mmHg.

Meanwhile, some models whith bluetooth function can be used as a stand-alone unit to finish the blood pressure measurement or in conjunction with the APP through embed a 2.4GHz BLE module that allow users to connect with nearby BT receiving terminal. Once measurement is over, the LCD of the device displays results. And the device will start to transmit data to the pair-up terminal automatically. This app is only intended to display measurement results from the blood pressure monitor device, which does not provide any diagnostic or measurement functions, and does not interpret or analyze the data for medical decision making. Unlimited readings can be stored in the app for archiving and review by the user.

There are total 7 arm-type blood pressure monitor models we submitted: DBP-62D0L, DBP-62D0B, DBP-61D0, DBP-61D0L, DBP-6293L, DBP-6193 and they are both bodyworn medical devices.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided FDA 510(k) summary for the JOYTECH Arm-type Fully Automatic Digital Blood Pressure Monitor:

Acceptance Criteria and Reported Device Performance

The primary acceptance criteria for a non-invasive blood pressure monitor are its accuracy in measuring systolic and diastolic blood pressure. The relevant standard cited is ISO 81060-2:2018+AMD2020 (Non-invasive sphygmomanometers — Part 2: Clinical investigation of intermittent automated measurement type).

The document states: "The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020." While specific numerical performance values (e.g., mean difference and standard deviation for systolic and diastolic pressure) are not explicitly provided in this summary, the statement indicates that the device met the statistical requirements defined by this international standard for accuracy.

Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Based on ISO 81060-2:2018+AMD2020)Reported Device Performance
Accuracy of Systolic Blood Pressure MeasurementWithin acceptable scope specified in ISO 81060-2:2018+AMD2020
Accuracy of Diastolic Blood Pressure MeasurementWithin acceptable scope specified in ISO 81060-2:2018+AMD2020

Details of the Study Proving Device Meets Acceptance Criteria:

  1. Sample Size and Data Provenance:

    • Test Set Sample Size: A total of 90 subjects were included in the clinical validation study.
    • Data Provenance: The document does not explicitly state the country of origin of the data, but the manufacturer is based in China. The study appears to be prospective as it describes recruitment of subjects, their participation, and the measurement process.
  2. Number of Experts and Qualifications:

    • The document mentions "The manual Mercury Sphygmomanometer was used as a reference device." This implies measurements were taken by trained observers (experts) using this reference device. However, the exact number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience" is not applicable here) are not specified in the provided text. For blood pressure clinical validation, these would typically be trained clinicians or technicians.
  3. Adjudication Method for the Test Set:

    • The study used a "Same arm sequential method." This means the test device measurements and reference device measurements were taken sequentially on the same arm.
    • The document does not describe a specific "adjudication method" in the sense of multiple experts reviewing and reaching consensus on an interpretation (as would be common in image-based AI studies). Instead, the comparison is made between the device readings and the reference standard readings.
  4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for AI devices that assist human readers in interpreting medical images (e.g., AI for chest X-rays assisting radiologists). The device in question is a standalone blood pressure monitor, not an AI-assisted diagnostic tool that requires human interpretation of complex data.
  5. Standalone Performance:

    • Yes, a standalone performance study was done. The clinical validation detailed here assesses the accuracy of the blood pressure monitor itself (the algorithm and hardware) against a recognized gold standard (manual mercury sphygmomanometer). The device operates independently to provide blood pressure measurements.
  6. Type of Ground Truth Used:

    • The ground truth used was manual Mercury Sphygmomanometer measurements. This is a widely accepted reference standard for validating automated blood pressure monitors, considered an expert consensus/reference standard method in this context.
  7. Sample Size for the Training Set:

    • The document does not provide information regarding the sample size of a training set. This is typical for a traditional medical device like a blood pressure monitor where the "algorithm" is often based on well-established non-AI methods (oscillometric principles) and validated through clinical testing rather than trained on large datasets like deep learning models. If there are software algorithms involved, they are likely more deterministic or based on classical signal processing rather than machine learning that requires a separate "training set."
  8. How Ground Truth for Training Set Was Established:

    • As no "training set" in the context of typical AI/machine learning is mentioned, this information is not applicable and therefore not provided in the document. The device's underlying measurement principles (oscillometric method) use known physical principles, not a data-driven training process that requires a labeled ground truth for learning.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 13, 2025

JOYTECH Healthcare Co., Ltd. Cong Jing Official Correspondent No. 365. Wuzhou Road, Yuhang Economic Development Zone Hangzhou, 310000 China

Re: K241431

Trade/Device Name: Arm-type Fully Automatic Digital Blood Pressure Monitor (DBP-62D0L, DBP-62D0B, DBP-61D0, DBP-61D0L, DBP-6293L, DBP-6293B, DBP-6193) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN, Dated: April 22, 2024 Received: May 21, 2024

Dear Cong Jing:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Robert T. Kazmierski -S for

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K241431

Device Name

Arm-type Fully Automatic Digital Blood Pressure Monitor (DBP-62D0L, DBP-61D0, DBP-61D0L, DBP-6293L, DBP-6293B, DBP-6193)

Indications for Use (Describe)

The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood Pressure (systolic and diastolic) of adults and adolescents over 12 years of age with circumference ranging from 22cm to 24cm.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for JOYTECH. The logo is in blue and red. The word "JOYTECH" is in blue, except for the "T", which is in red. To the right of "JOYTECH" are two Chinese characters in blue.

510(k) Summary

K241431

The assigned 510(k) number is: K241431

2.1 Subiectter's Identification:

Name: JOYTECH Healthcare Co., Ltd.

Add .: No. 365. Wuzhou Road, Yuhang Economic Development Zone, Hangzhou City, 311100

Zhejiang, China.

Contact Person: Cong Jing

Phone: +86-571-81957767

Fax: +86-571-81957750

Email: Jingc(@sejoy.com

2.2 Name of the Device:

Trade Name: Fully Automatic Digital Blood Pressure Monitor

Arm-typeDBP-62D0L, DBP-62D0B, DBP-61D0, DBP-61D0L, DBP-6293L, DBP-6293B, DBP-6193
-------------------------------------------------------------------------------------

Common Name: Blood Pressure Monitor

Classification name: Non-invasive blood pressure measurement System

21 CFR 870-1130, Class II, 74-DXN.

2.3 Classification Information:

Regulation Number: 870.1130

Product Code: DXN

Device Class: II

Panel: 74 Cardiovascular

2.4 Predicate Device Information:

The Arm-Type Fully Automatic Digital Blood Pressure Monitors are substantially equivalent to the following predicate devices:

{5}------------------------------------------------

510(k) numberPredicate device modelProduct codeManufacturer
K162092Evolv Model Bp7000DXNOmron Healthcare, Inc.

2.5 Device Description:

The Arm-type Fully Automatic Digital Blood Pressure Monitor (BPM) series is automatic, non-invasive, blood pressure measurement system for over-the-counter (OTC) use at home and clinical environment . The systolic and diastolic pressures are determined using the oscillometric method, where the cuff is inflated with an integral controllable piezoelectric pump and deflates via an electric automatic rapid deflation valve. During inflation measurements, an electric pump within the main unit slowly inflates the arm cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms within the microprocessor to determine systolic pressure, and diastolic pressure. The cuff can measure pressure range from 0 to 299mmHg.

Meanwhile, some models whith bluetooth function can be used as a stand-alone unit to finish the blood pressure measurement or in conjunction with the APP through embed a 2.4GHz BLE module that allow users to connect with nearby BT receiving terminal. Once measurement is over, the LCD of the device displays results. And the device will start to transmit data to the pair-up terminal automatically. This app is only intended to display measurement results from the blood pressure monitor device, which does not provide any diagnostic or measurement functions, and does not interpret or analyze the data for medical decision making. Unlimited readings can be stored in the app for archiving and review by the user.

There are total 7 arm-type blood pressure monitor models we submitted: DBP-62D0L, DBP-62D0B, DBP-61D0, DBP-61D0L, DBP-6293L, DBP-6193 and they are both bodyworn medical devices.

2.5.1 The detail functions for DBP-62D0L, DBP-62D0B, DBP-61D0, DBP-61D0L, DBP-6293L, DBP-6293B, DBP-6193 are listed in table below:

ModelABCDEFGHIJKLMN
DBP-6193YYYYY2*150 memoriesYYNNYYYY
DBP-6293BYYYYY2*150 memoriesYYNYYYYY
DBP-6293LYYYYY2*150 memoriesYYYYYYYY

{6}------------------------------------------------

JOYTECHll建拓
---------------
DBP-62D0LYYYYY2*150 memoriesYNYYYYYY
DBP-62D0BYYYYY2*150 memoriesYNNYYYYY
DBP-61D0YYYYY2*150 memoriesYNNNYYYY
DBP-61D0LYYYYY2*150 memoriesYNYNYYYY

A = Systolic pressure, Diastolic pressure measuring function

B = Automatic Power-Off

C =-WHO (World Health Organization) Classification Indicator

D = Last 3 Tests Average

E = Low Battery Detection

F = Memory

G = Voice

H = Backlight

I = lithium battery

J = Bluetooth Function

K = Arm Shake Detection

L = Cuff Loose Detection

M = Triple measurement function

N =Cuff size(22cm-42cm)

Note:

Y= Yes

N= No

2.6 Indication for use/Intended Use:

DBP-62D0L,DBP-62D0B, DBP-61D0, DBP-61D0L, DBP-6293L, DBP-6293B, DBP-6193: The Arm-type Fully Automatic Digital Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) of adults and adolescents over 12 years of age with circumference ranging from 22cm to 42cm.

2.7 Comparison of Technological Characteristics with predicate device:

2.7.1 Arm-type

The arm-type blood pressure monitor for model DBP-62D0L,DBP-62D0B, DBP-61D0, DBP-61D0L, DBP-6293L, DBP-6293B, DBP-6193 manufactured by JOYTECH have the same arm cuff type, features and specifications with the Omron Evolv model Bp 7000 Upper Arm Blood

{7}------------------------------------------------

JOYTECH-健拓

Pressure Monitor which 510k number is K162092, therefore we choose the device act as the

predicate device. The detail comparison of technical characteristic as below:

Comparison itemSubject device for Model DBP-6193,DBP-6293B,DBP-6293L,DBP-62D0L,DBP-62D0B, DBP-61D0, DBP-61D0L, DBP-6293L, DBP-6293B, DBP-6193Predicate device for Bp7000(K162092)ComparisonresultExplanation
The trade nameArm-type Fully Automatic Digital BloodPressure MonitorEVOLV Upper Arm Blood PressureMonitor Model Bp7000/
ManufacturerJOYTECH Healthcare Co., Ltd.Omron Healthcare,Inc./
Recommendedclassificationregulation21CFR 870.1130,Noninvasive BloodPressure MeasurementSystem21CFR 870.1130,Noninvasive BloodPressure MeasurementSystemIdentical
Regulatory classIIIIIdentical
Panel74 Cardiovascular74 CardiovascularIdentical
Product codeDXNDXNIdentical
Indication foruse/Intended useThe Arm-type Fully Automatic BloodPressure Monitors are intended to measureblood pressure (systolic and diastolic) ofadults and adolescents over 12 years ofage with circumference ranging from22cm to 42cm.The device is a digital monitor intendedfor use in measuring blood pressure andpulse rate in adult patient populationwith arm circumference ranging from 9inches to 17 inches (22cm to 42cm).The device detects the appearance ofirregular heartbeats during measurementand gives a warning signal withreadings.Similar
MeasuringprincipleOscillometric methodOscillometric methodIdentical
MeasurementtypeDetermined during inflationDetermined during inflationIdentical
Classification ofinstallation andusebody-wornbody-wornIdentical
Cuff locationUpper armUpper armIdentical
Specification
Measuring rangeSystolic Pressure:60mmHg260 mmHg;Diastolic Pressure:40mmHg200 mmHg;Cuff pressure:0mmHg~299 mmHg;Pressure:0mmHg299 mmHg;Pressure: 40 to 260 mmHgPulse rate:40180 Beats/Minute;SimilarNote 1
AccuracyPressure deviation:±3 mmHg;Pressure deviation:±3 mmHg;Pulse deviation: ±5%Identical
Pressure releaseBy solenoid valveBy solenoid valveIdentical
OperatingTemp. &humidityTemp.: 10°C40°CHumidity: 1593%RHAtmosphericpressure: 80kPa~106kPaTemp.:10°C40°CRelative humidity :1590%RHAtmosphericpressure: 80kPa~106kPaSimilarNote 2
Storage Temp.& humidityTemp.: -25°C~55°CHumidity:<93% RHTemp.:-20°C60°CRelative Humidity:1090% RH(Non-condensing)SimilarNote 3
Cuffcircumference22-42 cm22-42 cmIdentical
SupplypowersourceDBP-6193:3* AAA battery or Medical ACAdaptor:DBP-6293B:3AAA battery or MedicalAC Adaptor;DBP-6293L:Lithium battery or MedicalAC Adaptor;DBP-62D0L: Lithium battery or MedicalAC Adaptor;DBP-62D0B: 3AAA battery or MedicalAC Adaptor;DBP-61D0L: Lithium battery or MedicalAC Adaptor;DBP-61D0: 3*AAA battery or MedicalAC Adaptor;3*AAA batterySimilarNote 4
BluetoothtransmissionfunctionYesYesIdentical
SterilizationNot applicableNot applicableIdentical
Features
Cuffloose indicatorYesYesIdentical
Memory2*120 Memories in Two GroupsMemory FunctionSimilar.Note 5

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for JOYTECH. The logo is in blue and red. The word "JOYTECH" is in blue, except for the "T", which is in red. To the right of the word "JOYTECH" are two Chinese characters in blue.

K241431

Note 1: The measuring range of the subject device have been verified by 60601-2-30 standard test. This difference will not bring any new risks.

Note 2: The operation environment of the subject device have been verified by IEC 60601-1 and 80601-2-30 standard test. This information is indicated in the instruction manual to remind the user and will not bring any new risks.

Note 3: The storage environment of the subject device have been verified by IEC 60601-1 and 80601-2-30 standard test.This information is indicated in the instruction manual to remind the user

Note 4: The Lithium battery has passed the standard IEC 62133-2 test and the AC adaptor also has passed the IEC60601-1 test.

Note 5: The quantity of storage readings is different between the subject device and predicate device. The change in the specification is documented, and the change does not affect the intended use or the fundamental scientific technology.

2.8. Performance Data:

{9}------------------------------------------------

Testing information demonstrating safety and effectiveness of the device in the intended environment of use is supported by testing that was conducted.

The following testing was conducted to prove safety and effectiveness as well as substantial equivalence to the predicate devices.

The following National and International Standards were utilized for testing the subject device.

Electrical Safety and performance requirements:

  • IEC 60601-1:2005+AMD1:2012+AMD2:2020 , Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance.
  • AAMI ES60601-1:2005+AMD1:2012+AMD2:2021, Medical Electrical Equipment.
  • IEC 80601-2-30:2018, medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers.

Home-used medical equipment requirements and environmental test:

  • IEC 60601-1-11:2015+AMD1:2020,General requirements for basic safety and essential . performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
    Electromagnetic compatibility requirements:

  • . IEC 60601-1-2:2014+AMD1:2020 , Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
    Bio-compatibility Evaluation for patient contacting components:

  • ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for in vitro . cytotoxicity

  • ISO10993-10:2010, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization

FCC Test

  • FCC Part15 Subpart C
  • RF Exposure Evaluation

Guidance Document:

The software/firmware verification and validation was provided in accordance with the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"

The test result all meet or exceed the requirement of these standards.

{10}------------------------------------------------

2.10. Discussion of Clinical Tests Performed:

Clinical Validation:

  • ISO81060-2:2018+AMD2020 Non-invasive sphygmomanometers —Part 2: Clinical investigation of intermittent automated measurement type.

Model DBP-6279B was selected as representative for testing. It was conducted according to ISO 81060-2 Non-Invasive Sphygmomanometers - Part2: Clinical Investigation Of Automated Measurement Type. The subject demographics include total of 90 subjects meeting the inclusion criteria, all 90 include subjects are all older than 12 years.

Same arm sequential method was adopted during the clinical testing. The manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect. The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2:2018+AMD2020.

The applicable part of subject devices in this submission and previous arm-type blood pressure monitor models submitted by JOYTECH are both upper-arm. The only difference is that current subject devices are body-worn but previous submitted are portable. All new design elements have been successfully validated and the results fully demonstrate that the device has good safety and effectiveness.

Meanwhile, through the verification of internal simulation tests, it prove that even if the appearance is different, but in the case of the same algorithm, the performance of the DBP-6279B and the current submitted models are the same. There is no need to carry out clinical research on the new design, and the clinical results of DBP-6279B can cover the current submitted models.

2.11. Conclusions:

The subject device and predicate devices have same Measuring principle, measuring method, are designed for the measurement of blood pressure in adult population for home use. The minor difference between the subject devices and the predicate devices have been evaluated and determined to not raise any new issues of safety or effectiveness.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the logo for JOYTECH. The logo is in blue and red. To the right of the word JOYTECH are three Chinese characters.

Therefore, the new models as mentioned on this submission are considered substantial equivalent to the predicate devices.

--- End of this section---

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).