K Number
K241380
Device Name
FETOLY-HEART
Manufacturer
Date Cleared
2024-09-11

(119 days)

Product Code
Regulation Number
892.1550
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
FETOLY-HEART is intended to analyse fetal ultrasound images and clips using machine learning techniques to automatically detect heart views and quality criteria within the views. The device is intended for use as a concurrent reading aid during the acquisition and interpretation of fetal ultrasound images. FETOLY-HEART is indicated for use during routine fetal heart examination of 2nd and 3rd trimester pregnancy (gestational age: from 17 to 40 weeks).
Device Description
FETOLY-HEART is a software that aims at helping sonographers, obstetricians, radiologists, maternal-fetal medicine specialists, and pediatric cardiologists (designated as healthcare professionals i.e. HCPs) to perform fetal ultrasound examinations of the fetal heart in real-time. FETOLY-HEART can be used by HCPs during fetal ultrasound examinations in the second and third trimesters (gestational age window: from 17 to 40 weeks). The software is intended to assist HCPs in the completeness assessment of the fetal heart ultrasound examination in accordance with national and international guidelines. To utilize FETOLY-HEART, the software needs to be installed on a hardware device which is connected to an Ultrasound Machine through an HDMI connection. The software receives ultrasound images captured by the connected Ultrasound Machine in real-time. The software's frozen deep learning algorithm, which was trained by supervised learning, analyzes images of this ultrasound image stream to detect heart views and quality criteria within those views. The software provides the following user-accessible information: - . Examination completeness: the software displays in real-time which heart views and quality criteria are verified by the software during the examination. It is the main and principal output of the FETOLY-HEART device. The verified heart views and quality criteria are accessible by clinicians at any moment of the ultrasound examination, in real-time. - . Completeness illustration: the software selects an image subset that illustrates the verified views and quality criteria. These images can be reviewed by clinicians to verify the views and criteria's presence. This is a secondary output of the FETOLY-HEART device. Optionally, clinicians can display detected quality criteria localization on selected images.
More Information

Yes
The intended use and device description explicitly state the use of "machine learning techniques" and a "frozen deep learning algorithm" for image analysis.

No

The device is intended to analyze fetal ultrasound images and clips to detect heart views and quality criteria, acting as a reading aid during acquisition and interpretation. It is not designed to provide or deliver therapy.

No

The device is described as a "concurrent reading aid" and is "intended to assist HCPs in the completeness assessment of the fetal heart ultrasound examination," rather than providing a diagnosis or clinical assessment itself. Its primary output is "Examination completeness," indicating which heart views and quality criteria are verified by the software, not a diagnostic finding.

Yes

The device is explicitly described as "a software" and its function is to analyze images received from an ultrasound machine via an HDMI connection. It does not include or require any specific hardware components beyond the standard computing device it is installed on and the connection to the ultrasound machine.

Based on the provided information, no, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body (like blood, urine, tissue) to provide information about a person's health. They are used outside the body.
  • FETOLY-HEART's Function: FETOLY-HEART analyzes ultrasound images and clips of the fetal heart. These are images generated by an external medical device (the ultrasound machine) and are not samples taken from the body.
  • Intended Use: The intended use is to analyze images and act as a concurrent reading aid during the acquisition and interpretation of fetal ultrasound images. This is focused on image analysis and interpretation, not on testing biological samples.

While FETOLY-HEART is a medical device that uses AI and image processing, its function and intended use clearly fall outside the scope of In Vitro Diagnostics. It is a software intended to assist healthcare professionals in interpreting medical imaging data.

Yes
The letter explicitly states, "FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP)." This language directly satisfies the condition for a "Yes" answer according to the Key Decision Rules.

Intended Use / Indications for Use

Fetoly-Heart is intended to analyze fetal ultrasound image sequences using machine learning techniques to automatically detect heart views and quality criteria within the views. The device is intended for use as a concurrent reading aid during the acquisition and interpretation of fetal ultrasound images.

Fetoly-Heart is indicated for use during routine fetal heart examination of 2nd and 3rd trimester pregnancy (gestational age: from 17 to 40 weeks).

Product codes (comma separated list FDA assigned to the subject device)

IYN, IYO, QIH

Device Description

FETOLY-HEART is a software that aims at helping sonographers, obstetricians, radiologists, maternal-fetal medicine specialists, and pediatric cardiologists (designated as healthcare professionals i.e. HCPs) to perform fetal ultrasound examinations of the fetal heart in real-time. FETOLY-HEART can be used by HCPs during fetal ultrasound examinations in the second and third trimesters (gestational age window: from 17 to 40 weeks). The software is intended to assist HCPs in the completeness assessment of the fetal heart ultrasound examination in accordance with national and international guidelines.

To utilize FETOLY-HEART, the software needs to be installed on a hardware device which is connected to an Ultrasound Machine through an HDMI connection. The software receives ultrasound images captured by the connected Ultrasound Machine in real-time. The software's frozen deep learning algorithm, which was trained by supervised learning, analyzes images of this ultrasound image stream to detect heart views and quality criteria within those views. The software provides the following user-accessible information:

  • . Examination completeness: the software displays in real-time which heart views and quality criteria are verified by the software during the examination. It is the main and principal output of the FETOLY-HEART device. The verified heart views and quality criteria are accessible by clinicians at any moment of the ultrasound examination, in real-time.
  • . Completeness illustration: the software selects an image subset that illustrates the verified views and quality criteria. These images can be reviewed by clinicians to verify the views and criteria's presence. This is a secondary output of the FETOLY-HEART device. Optionally, clinicians can display detected quality criteria localization on selected images.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Yes

Input Imaging Modality

Ultrasound

Anatomical Site

Fetal heart

Indicated Patient Age Range

Gestational age: from 17 to 40 weeks (2nd and 3rd trimester pregnancy)

Intended User / Care Setting

Sonographers, obstetricians, radiologists, maternal-fetal medicine specialists, and pediatric cardiologists (healthcare professionals) during routine fetal heart examination.

Description of the training set, sample size, data source, and annotation protocol

Not Found. The document states "This testing dataset originated from distinct clinical sites from which the data used during model development (training/validation) was sourced, ensuring testing independence." but does not provide details on the training/validation data.

Description of the test set, sample size, data source, and annotation protocol

Description of the test set: Dataset of 2,288 fetal ultrasound images across 480 patient cases, including full examination still images, cardiac clip frames and full examination video frames.
Sample size: 2,288 images from 480 patient cases.
Data source: 7 clinical sites in the United States. Cases are representative of the intended use population. Patient cases were retrospectively collected in reverse chronological order until at least 20 patient files per subgroup and an overall of 275 patient files were reached. One image maximum per view per patient case was selected by categorizing 12,934 images from the patient cases into heart views and randomly picking one image maximum per view, resulting in a total of 2,288 images.
Annotation protocol:
For view classification: A 2+1 ground truth procedure was used. Six annotators (3 sonographers and 3 OB/GYN doctors) were paired and assigned uniformly distributed batches of images. Each image was annotated by a pair of annotators as belonging to one of 6 views. Images with annotator agreement were considered ground truth. Images in which the pair of annotators disagreed were reviewed by an adjudicator, who made the final decision.
For quality criteria classification and localization: Each image was annotated by a pair of annotators who drew bounding boxes on present criteria. If their boxes had at least 50% overlap, their coordinates were averaged to form the ground truth. If the overlap was lower or there was a disagreement on the criterion presence, an adjudicator reviewed the boxes. The final decision regarding the presence was based on majority consensus among the adjudicator and annotators. The final decision for the criteria localization was based on the adjudicator's decision to either keep one of the annotator's boxes or draw a new one.

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Study type: Standalone performance testing.
Sample size: 2,288 fetal ultrasound images across 480 patient cases.
AUC: Not mentioned.
MRMC: Not mentioned.
Standalone performance:

  • Fetal heart view detection: Sensitivity >= 85% (acceptance criterion) and specificity >= 85% (acceptance criterion).
  • Quality criteria within heart views detection: Sensitivity >= 90% (acceptance criterion) and specificity >= 90% (acceptance criterion).
  • Localization of bounding boxes of quality criteria: Mean Intersection over Union (IoU) of >= 50% (acceptance criterion).
    Key results:
  • All acceptance criteria were met.
  • The lowest 95% CI for sensitivity for all fetal heart views was 0.960 for Abdomen view. The lowest 95% CI for specificity for all fetal heart views was 0.977 for "Other view".
  • The lowest 95% CI for sensitivity for all individual quality criteria was 0.860 (Right pulmonary vein). The lowest 95% CI for specificity for all individual quality criteria was 0.985 (Trachea / bronchi).
  • The lowest 95% CI for mean IoU for all individual quality criteria was 0.491 (Trachea / bronchi).
  • Performance metrics were analyzed for each subgroup (maternal age, gestational age, territory, BMI, scanner manufacturer, heart abnormality, races and ethnicities, clinical sites) and confounder (image digital quality, image type) to validate the model's robustness and generalizability.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Fetal heart view detection:

  • Overall Sensitivity: Lowest 95% CI for sensitivity among all views was 0.960 (Abdomen view). Point estimates range from 0.976 to 0.987.
  • Overall Specificity: Lowest 95% CI for specificity among all views was 0.977 ("Other view"). Point estimates range from 0.983 to 1.00.

Quality criterion detection:

  • Overall Sensitivity: Lowest 95% CI for sensitivity among all quality criteria was 0.860 (Right pulmonary vein). Point estimates range from 0.903 to 0.990.
  • Overall Specificity: Lowest 95% CI for specificity among all quality criteria was 0.985 (Trachea / bronchi). Point estimates range from 0.990 to 1.00.
  • Mean Intersection over Union (mIoU): The lowest 95% CI for mIoU among all quality criteria was 0.491 (Trachea / bronchi). Point estimates range from 0.512 to 0.792.

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K240406

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K220358

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Presence / Absence: Included.
Scope granted / cleared under the PCCP: PCCP is for machine learning (ML) algorithm modifications, specifically:

  1. Modification of training and/or validation datasets: Increase or recovery (in case of data drift) of FETOLY-HEART's performance.
  2. Modification of model training hyperparameters: Improvement and optimization of FETOLY-HEART's performance.
  3. Heart quality criteria addition/removal: Maintain alignment with quality criteria recommended for fetal heart screening in state-of-the-art international guidelines.

Restrictions:

  • All algorithm modifications will be adequately trained, tuned, and locked prior to release of the software with the modified ML model.
  • The PCCP does not include the implementation of adaptive algorithms that will continuously learn in the field.
  • Implemented modifications to the FETOLY-HEART algorithm will be communicated to users via the software update notifications and through updated labelling.
  • Modified models will be tested for superiority on the performance study test dataset which will contain new unseen data (for dataset and hyperparameter modifications).
  • New quality criteria will be controlled using the same acceptance criteria as defined by secondary endpoints, with proper performance testing to ensure no decrease in test performance (for quality criteria addition/removal). This change only pertains to quality criteria belonging to one of the 5 heart views already included in FETOLY-HEART.

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.

0

September 11, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and reads "FDA U.S. FOOD & DRUG ADMINISTRATION".

Diagnoly % Nima Akhlaghi Associate Director, Digital Health Regulatory Affairs MCRA, LLC 505 Park Avenue, 14th Floor New York, NY 10022

Re: K241380

Trade/Device Name: FETOLY-HEART Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, QIH Dated: August 9, 2024 Received: August 9, 2024

Dear Nima Akhlaghi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an

1

established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rele"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

2

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Yanna S. Kang -S

Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

3

Indications for Use

Submission Number (if known)

K241380

Device Name

FETOLY-HEART

Indications for Use (Describe)

Fetoly-Heart is intended to analyze fetal ultrasound image sequences using machine learning techniques to automatically detect heart views and quality criteria within the views. The device is intended for use as a concurrent reading aid during the acquisition and interpretation of fetal ultrasound images.

Fetoly-Heart is indicated for use during routine fetal heart examination of 2nd and 3rd trimester pregnancy (gestational age: from 17 to 40 weeks).

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

4

Image /page/4/Picture/0 description: The image shows a logo for a company called DIAGNOLY. The logo features a blue magnifying glass with a stack of three horizontal lines inside the lens. The word "DIAGNOLY" is written in blue, sans-serif font below the magnifying glass. The background of the image is white.

In accordance with 21 CFR 807.92 the 510(k) summary for FETOLY-HEART is provided below.

510(k) owner 1

| Owner | Diagnoly
60 Avenue Rockefeller
69008 Lyon, France
+33(0)4.78.76.85.75 |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Primary contact person | Ivan Voznyuk
Chief Executive Officer
Diagnoly
Phone: +33(0)6.95.87.04.55
Email: ivan@diagnoly.com |
| Secondary contact person | Nima Akhlaghi
Associate Director, Digital Health Regulatory Affairs
MCRA, LLC
Phone: 202.742.3889
Email: nakhlaghi@mcra.com |
| Date prepared | 2024-08-30 |

2 Device

Trade NameFETOLY-HEART
Classification nameAccessory to Ultrasonic Pulsed Doppler Imaging System, 21 CFR 892.1550
Accessory to Ultrasonic Pulsed Echo Imaging System, 21 CFR 892.1560
Medical image management and processing system, 21 CFR 892.2050
ClassII
Product codeIYN (Primary)
IYO, QIH (secondary)

Predicate device identification ന

The predicate device used for FETOLY-HEART is the cardiac-related component of Sonio Detect (K240406).

Additionally, a reference device was chosen for FETOLY-HEART based on its substantially equivalent technical characteristics of automatic extraction of views from a sequence of images. This reference

5

Image /page/5/Picture/0 description: The image shows a logo for a company called DIAGNOLY. The logo features a blue magnifying glass with a stylized image of a stack of objects inside the lens. The text "DIAGNOLY" is written in blue below the magnifying glass.

device corresponds to the SonoLystLive view suggestion feature within the Voluson Expert Series 22/20/18 (K220358).

Device description 4

FETOLY-HEART is a software that aims at helping sonographers, obstetricians, radiologists, maternal-fetal medicine specialists, and pediatric cardiologists (designated as healthcare professionals i.e. HCPs) to perform fetal ultrasound examinations of the fetal heart in real-time. FETOLY-HEART can be used by HCPs during fetal ultrasound examinations in the second and third trimesters (gestational age window: from 17 to 40 weeks). The software is intended to assist HCPs in the completeness assessment of the fetal heart ultrasound examination in accordance with national and international guidelines.

To utilize FETOLY-HEART, the software needs to be installed on a hardware device which is connected to an Ultrasound Machine through an HDMI connection. The software receives ultrasound images captured by the connected Ultrasound Machine in real-time. The software's frozen deep learning algorithm, which was trained by supervised learning, analyzes images of this ultrasound image stream to detect heart views and quality criteria within those views. The software provides the following user-accessible information:

  • . Examination completeness: the software displays in real-time which heart views and quality criteria are verified by the software during the examination. It is the main and principal output of the FETOLY-HEART device. The verified heart views and quality criteria are accessible by clinicians at any moment of the ultrasound examination, in real-time.
  • . Completeness illustration: the software selects an image subset that illustrates the verified views and quality criteria. These images can be reviewed by clinicians to verify the views and criteria's presence. This is a secondary output of the FETOLY-HEART device. Optionally, clinicians can display detected quality criteria localization on selected images.
Heart viewsQuality criteria within the views
Image: Heart view
(A) ABD
Abdomen view
(n = 8)(A1) Sp Spine
(A2) IRb Left rib
(A3) rRb Right rib
(A4) Ao Descending aorta
(A5) VC Inferior vena cava
(A6) St Stomach
(A7) Uv Umbilical vein
(A8) Ap Thorax apex
(B1) Sp Spine
(B2) IRbLeft rib
(B3) rRbRight rib
(B4) AoDescending aorta
(B5) IPVLeft pulmonary vein
(B6) rPVRight pulmonary vein
(B7) LALeft atrium
(B8) RARight atrium
(B) 4CH(B9) FOPForamen Ovale flap (Vieussens valve)
Four chamber view
(n = 19)(B10) FOOpen Foramen Ovale
(B11) MVMitral valve
(B12) TVTricuspid valve
(B13) bCrConnection between crux and atrial septum (vestibular septum)
(B14) CrAtrioventricular valve offset in crux
(B15) tCrConnection between interventricular septum and crux
(B16) IVSInterventricular septum
(B17) LVLeft ventricle
(B18) RVRight ventricle
(B19) StrSternum
(C1) LALeft atrium
(C2) aAoProximal ascending aorta
(C) LVOT(C3) SVSemilunar valves
(C4) LVLeft ventricle
(C5) IVSInterventricular septum
Left Ventricular Outflow Tract view
(n = 6)(C6) RVRight ventricle
(D1) dAoDescending aorta
(D2) TrTrachea / bronchi
(D3) IPALeft pulmonary artery
(D4) DuDuctus arteriosus
(D5) rPARight pulmonary artery
(D6) OrOrigin of pulmonary arteries
(D) RVOT(D7) SSeptum between pulmonary artery trunk and ascending aorta
Right Ventricular Outflow Tract view
(n = 10)(D8) aAoAscending aorta
(D9) SVCSuperior vena cava
(D10) PAPulmonary trunk
(E1) SpSpine
(E2) TrTrachea
(E3) ESSide space on the left of ductus / pulmonary artery
(E4) PAMain pulmonary artery
(E5) DuDuctus (Ductal arch)
(E6) aAoAscending aorta
(E) 3VX
(E7) aArAortic arch
Three vessels view
(n = 9)(E8) SVCSuperior vena cava
(E9) ThThymus / sternum

Definition of a complete examination 4.1

6

Image /page/6/Picture/0 description: The image shows a logo for a company called DIAGNOLY. The logo features a blue magnifying glass with a stylized image of a stack of three cones inside. The company name, DIAGNOLY, is written in blue text below the magnifying glass.

7

Image /page/7/Picture/0 description: The image shows a logo for a company called "Diagnoly". The logo features a blue magnifying glass with a stack of three cone-shaped objects inside the lens. The word "Diagnoly" is written in blue, sans-serif font below the magnifying glass.

Table 1. List of 52 quality criteria defining the 5 views recommended by the International Society of Ultrasound in Obstetrics and Gynecology for the foetal heart screening of 2nd and 3rd trimesters of pregnancy. Abbreviations from this table are corresponding to the abbreviations used within the software.

International and national guidelines- recommend 5 foetal cardiac views for routine ultrasound examination of 2nd and 3rd trimesters: (A) Abdomen view, (B) Four chamber view, (C) Left Ventricular Outflow Tract view, (D) Right Ventricular Outflow Tract view, (E) Three vessels view. The quality of these 5 heart views depends on the presence of 52 anatomical quality criteria within the views (Table 1). Thus, an examination can be defined as complete when all 5 heart views and their quality criteria are obtained by the HCP.

Functionality 1: completeness overview 4.2

The software assesses the completeness of the foetal ultrasound examination. It verifies whether all the information corresponding to the recommended guidelines for the foetal heart examination has been acquired. This information corresponds to the presence of 5 main foetal cardiac views and 52 quality criteria, detailed in the above section, allowing for compliance in cardiac screening.

Functionality 2: completeness illustration 4.3

This functionality was developed to enhance the security of the completeness evaluation which is done by the first module. It enables clinicians to verify the examination completeness overview, i.e. verified views and quality criteria by the software through the gallery page of the software interface. Operating continuously, it evaluates each image processed by the first module in real-time, retaining an image set as visual evidence of the verified heart views and quality criteria up to that moment.

FETOLY-HEART does not aim to select the 'best' or of 'high' diagnostic quality according to a given qualitative scale. Rather, FETOLY-HEART aims to select images illustrating the examination quantitative completeness in terms of verified views and quality criteria.

Indications for use ഗ

FETOLY-HEART is intended to analyse fetal ultrasound images and clips using machine learning techniques to automatically detect heart views and quality criteria within the views. The device is intended for use as a concurrent reading aid during the acquisition and interpretation of fetal ultrasound images.

FETOLY-HEART is indicated for use during routine fetal heart examination of 2nd and 3rd trimester pregnancy (gestational age: from 17 to 40 weeks).

1 Carvalho JS, Axt-Fliedner R, Chaoui R, Copel JA, Cuneo BF, Goff D, Gordin Kopylov L, Hecher K, Lee W, Moon-Grady A, Mousa HA, Munoz H, Paladini D, Prefumo F, Quarello E, Rychik J, Tutschek B, Wiechec M, Yagel S. ISUOG Practice Guidelines (updated): fetal cardiac screening. Ultrasound Obstet Gynecol 2023; 61: 788-803.

8

Image /page/8/Picture/0 description: The image shows the text "510(k) Summary - K241380". The text is written in a bold, sans-serif font. The words "510(k) Summary" are followed by a dash and then the code "K241380". The text is centered and appears to be the title or heading of a document.

Image /page/8/Picture/1 description: The image shows a logo for a company called DIAGNOLY. The logo features a blue magnifying glass with a stack of three objects inside. The word "DIAGNOLY" is written in blue letters below the magnifying glass.

6 Summary of the technological characteristics of FETOLY-HEART compared to the predicate device

| Aspect | Predicate device:
Sonio Detect
(Cardiac Features)
K240406 | Reference device:
Voluson Expert
18/20/22 (Cardiac-
related SonolystLive
Feature)
K220358 | Proposed device:
FETOLY-HEART | Comparison between
Proposed and
Predicate device |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General | | | | |
| Manufacturer
name | Sonio | GE Healthcare | Diagnoly | NA |
| Device name | Cardiac-related
features in Sonio
Detect | Cardiac-related
SonolystLive in the
Voluson Expert
18/20/22 | FETOLY-HEART | NA |
| Product
code(s) | IYN (Primary)
IYO, QIH (secondary) | IYN (Primary)
IYO, ITX (Secondary) | IYN (Primary)
IYO, QIH (secondary) | Substantially
equivalent
Primary codes are the
same for all devices |
| Regulation
number | - Accessory to
Ultrasonic Pulsed
Doppler Imaging
System, 21 CFR
892.1550

  • Accessory to
    Ultrasonic Pulsed
    Echo Imaging System,
    21 CFR 892.1560 | - Ultrasonic Pulsed
    Doppler Imaging
    System, 21 CFR
    892.1550
  • Ultrasonic Pulsed
    Echo Imaging System,
    21 CFR 892.1560
  • Diagnostic
    Ultrasound
    Transducer, 21 CFR
    892.1570, 90-ITX | - Accessory to
    Ultrasonic Pulsed
    Doppler Imaging
    System, 21 CFR
    892.1550
  • Accessory to
    Ultrasonic Pulsed
    Echo Imaging System,
    21 CFR 892.1560
  • Medical image
    management and
    processing system,
    21 CFR 892.2050 | Substantially
    equivalent
    All devices are class II
    devices subject to
    510(k) regulatory
    pathway. |
    | Brief
    description | The predicate device
    is a software that
    aims at helping
    sonographers,
    OB/GYNs, MFMs and
    Fetal surgeons (all
    three designated as
    healthcare
    professionals i.e.
    HCP) to perform their
    routine fetal heart
    ultrasound | The reference device
    is a software that
    aims at helping
    sonographers,
    OB/GYNs, MFMs and
    Fetal surgeons (all
    three designated as
    healthcare
    professionals i.e.
    HCP) to perform their
    routine fetal heart
    ultrasound | FETOLY-HEART is a
    software that aims at
    helping
    sonographers,
    OB/GYNs, MFMs and
    Fetal surgeons (all
    three designated as
    healthcare
    professionals i.e.
    HCPs) to perform
    their routine fetal
    heart ultrasound | Substantially
    equivalent
    The subject device
    and the predicate
    devices have the
    same objective. |
    | | examinations in real-time. | examinations in real-time. | examinations in real-time. | |
    | Indications for
    use | The predicate device
    is intended to analyze
    fetal ultrasound
    images and clips
    using machine
    learning techniques
    to automatically
    detect heart views,
    detect anatomical
    structures within the
    views and verify
    quality criteria of the
    views.
    The device is
    intended for use as a
    concurrent reading
    aid during the
    acquisition and
    interpretation of fetal
    ultrasound images. | The device is a
    general purpose
    ultrasound system
    intended for use by
    qualified and trained
    healthcare
    professionals. | FETOLY-HEART is
    intended to analyse
    fetal ultrasound
    images and clips
    using machine
    learning techniques
    to automatically
    detect heart views
    and quality criteria
    within the views. The
    device is intended for
    use as a concurrent
    reading aid during
    the acquisition and
    interpretation of fetal
    ultrasound images.

FETOLY-HEART is
indicated for use
during routine fetal
heart examination of
2nd and 3rd
trimester pregnancy
(gestational age:
from 17 to 40 weeks). | Substantially
equivalent
Indications for Use
are the same
between predicate
and subject devices. |
| Targeted
population | Pregnant women
during the 2nd and
3rd trimester of
pregnancy | Pregnant women
during the 2nd
trimester of
pregnancy | Pregnant women
during the 2nd and
3rd trimester of
pregnancy | Substantially
equivalent
Subject device has
the same intended
patient population
than the predicate
devices. |
| Clinical
outcome | - Images labeled with
correct fetal heart
view | - Images labeled
with correct fetal
heart view | - Images labeled with
correct fetal heart
view for patient cases | Substantially
equivalent
Performance testing
has successfully
validated the clinical
outcomes |
| | - Quality criteria
identified as
"Verified" when
detected and "Not
verified" when not
detected | - Quality criteria
identified as "Found"
when detected and
"Not found" when
not detected | - Quality criteria
identified as
"Verified" when
detected and "Not | Substantially
equivalent
In the subject device,
quality criteria
bounding box
localization can be |
| | | | verified" when not
detected

  • Images labeled with
    the localization of
    quality criteria | optionally displayed
    on post-acquisition
    images to enhance
    explainability of the
    Al model.
    Performance testing
    has been performed
    and does not
    introduce new
    questions of safety
    and effectiveness. |
    | Intended user | Qualified healthcare
    professional
    specialized in
    prenatal ultrasound
    imaging | Qualified healthcare
    professional
    specialized in
    prenatal ultrasound
    imaging | Qualified healthcare
    professional
    specialized in
    prenatal ultrasound
    imaging | Substantially
    equivalent
    Subject device has
    the same intended
    users as the predicate
    devices. |
    | Clinical
    applications | Fetal/Obstetrics | Fetal/Obstetrics | Fetal/Obstetrics | Substantially
    equivalent
    Clinical application is
    the same for subject
    and predicate
    devices. |
    | Inclusion of a
    PCCP | N/A | N/A | Included | Different
    The PCCP in the
    subject device
    includes proposed
    modifications related
    to modifying model
    training
    hyperparameters,
    additional retraining
    with new training and
    validation datasets
    collected, and
    addition/removal of
    heart quality criteria. |
    | Functionality 1: completeness overview | | | | |
    | Automatically
    detect views | Detection of 4ch, 3vx,
    LVOT, RVOT and Abd
    views (complete
    implementation of
    ISUOG
    recommendations) | Detection of 4ch, 3vx,
    LVOT, RVOT and Abd
    views (complete
    implementation of
    ISUOG
    recommendations) | Detection of 4ch, 3vx,
    LVOT, RVOT and Abd
    views (complete
    implementation of
    ISUOG
    recommendations) | Substantially
    equivalent
    The subject device
    includes the
    detection of the same
    views than the
    predicate device |
    | Automatically
    detect quality
    criteria | Detection of 28 heart
    quality criteria. The
    quality also
    incorporates the
    zoom level of the
    view. | Detection of NA
    heart quality criteria. | Detection of 52 heart
    quality criteria (see
    Table 1). The quality
    also incorporates the
    zoom level of the
    view. | Substantially
    equivalent
    The subject device
    includes the
    detection of new
    quality criteria when
    compared to the
    predicate device. This
    quantitative
    enhancement has
    been tested and does
    not raise any new
    question of safety
    and effectiveness. |
    | Functionality 2: completeness illustration | | | | |
    | Automatically
    selects views | NA | Automatic suggestion
    of views from a
    sequence of images. | Automatic extraction
    of views from a
    sequence of images. | Substantially
    equivalent
    This image selection
    functionality is a
    feature absent in the
    predicate device but
    present in the
    reference device.
    Software testing has
    been performed to
    validate its use and
    does not introduce
    new questions of
    safety and
    effectiveness. |
    | Technical characteristics | | | | |
    | Data input | Accepts images and
    image sequences
    from ultrasound
    machines | Accepts images and
    image sequences
    from ultrasound
    machines | Accepts images and
    image sequences
    from ultrasound
    machines | Substantially
    equivalent
    The input data is the
    same for the subject
    device and the
    predicate and
    reference devices. |
    | Algorithm
    Methodology | Artificial Intelligence:
    Utilizes computer
    vision algorithms to
    analyze ultrasound
    images and provides
    visualization of
    detected landmarks | Artificial Intelligence:
    Utilizes computer
    vision algorithms to
    analyze ultrasound
    images and provides
    visualization of
    detected landmarks | Artificial Intelligence:
    Utilizes computer
    vision algorithms to
    analyze ultrasound
    images and provides
    visualization of
    detected landmarks | Substantially
    equivalent
    The subject device
    and the primary
    predicate device use
    both artificial
    intelligence. |
    | Platform | Operates in a cloud-
    based environment
    functioning
    independently from
    the ultrasound
    equipment. | Operates as a local
    software embedded
    in the ultrasound
    equipment. | Operates as a local
    software functioning
    independently from
    the ultrasound
    equipment. | Substantially
    equivalent
    The edge-based
    approach reduces
    exposure to potential
    cloud-related
    vulnerabilities and
    latency issues.
    Therefore, this
    difference does not
    raise any safety or
    effectiveness
    concerns. |
    | Ultrasound
    Machine
    compatibility | Compatible with
    ultrasound system
    from GE Medical,
    Samsung, Canon and
    Philips | NA | Compatible with
    ultrasound system
    from GE Medical,
    Samsung and Canon | Substantially
    equivalent
    This compatibility has
    been tested and
    validated as part of
    device
    generalizability in the
    performance testing
    study. |
    | User
    interaction | The user can interact
    with the software to
    override the
    software's outputs.
    The user has the
    ability to review and
    edit/override the
    matching at any time
    during or at the end
    of the exam. | NA | The user can interact
    with the software to
    override the
    software's outputs.
    The user has the
    ability to review and
    edit/override the
    matching at any time
    during or at the end
    of the exam. | Substantially
    equivalent
    User interactions are
    the same between
    primary predicate
    and subject devices. |

9

Image /page/9/Picture/0 description: The image shows a logo for a company called DIAGNOLY. The logo features a blue magnifying glass with a stack of three curved lines inside, suggesting a focus on analysis or investigation. The company name, "DIAGNOLY," is written in a stylized blue font below the magnifying glass, emphasizing the company's diagnostic or analytical services.

10

Image /page/10/Picture/0 description: The image shows a blue magnifying glass with a symbol inside. The symbol consists of several horizontal lines that are stacked on top of each other, resembling a radar or sonar display. Below the magnifying glass, the word "DIAGNOLY" is written in blue, with the letters in a slightly rounded font.

11

Image /page/11/Picture/0 description: The image shows a logo for a company called DIAGNOLY. The logo features a blue magnifying glass with a stylized image of a stack of objects inside the lens. The word "DIAGNOLY" is written in blue letters below the magnifying glass.

12

Image /page/12/Picture/0 description: The image shows a blue logo for a company called DIAGNOLY. The logo features a magnifying glass with a stylized image of three stacked, truncated cones inside the lens. The word "DIAGNOLY" is written in blue below the magnifying glass.

Non-clinical performance data 7

7.1 FETOLY-HEART testing strategy

The following V&V testing were included into the development of the system:

  • Software verification testing per IEC 62304 standard
  • Tablet compatibility testing
  • Cybersecurity verification testing ●
  • Software Al model validation

13

Image /page/13/Picture/1 description: The image shows a logo for a company called "DIAGNOLY". The logo features a blue magnifying glass with a stylized image of a stack of three horizontal lines inside the lens. The word "DIAGNOLY" is written in blue below the magnifying glass.

FETOLY-HEART uses a machine learning (ML) algorithm for detection of heart views and quality criteria within these views in ultrasound images. Modifications to FETOLY-HEART will be made in accordance with the guiding principles on predetermined change control plans (PCCP) for machine learning-enabled medical devices. This PCCP provides a description of the device's planned modifications, and those modifications will be triggered and implemented in a controlled manner that ensures the continued safety and efficacy on the performance testing dataset, mitigating risks associated with changes to the ML model to not adversely impact the device's performance, safety, or effectiveness associated with its indications for use, and an impact assessment of the planned modifications.

In accordance with the PCCP, all algorithm modifications will be adequately trained, tuned, and locked prior to release of the software with the modified ML model. The PCCP does not include the implementation of adaptive algorithms that will continuously learn in the field. Implemented modifications to the FETOLY-HEART algorithm will be communicated to users via the software update notifications and through updated labelling. The modifications outlined in the PCCP are summarized in the table below. The PCCP in the subject device with the proposed modifications related to modifying model training hyperparameters, additional retraining and validation datasets collected, and addition/removal of heart quality criteria do not raise different questions of safety and effectiveness from the predicate device (see table below).

ModificationRationaleTesting MethodsImpact Assessment
Modification of
training and/or
validation datasetsIncrease or recovery (in
case of data drift) of
FETOLY-HEART's
performance.Re-training of the FETOLY-
HEART model with new
data to optimize its
performance followed by
internal testing and a
comparison of the initial
model to the modified
model using performance
metrics on the test
dataset.Increased performance metrics of the
modified model for view or quality
criteria detection.
Benefits: Increase or recovery of
performance; generalization for
diverse cases.
Risks: Performance decrease
(overfitting, unintended bias).
Risk mitigation: The modified model
will be tested for superiority on the
performance study test dataset which
will contain new unseen data.
Modification of
model training
hyperparametersImprovement and
optimization of FETOLY-
HEART's performanceRe-training of the FETOLY-
HEART model with new
parameters to optimize its
performance followed by
internal testing and a
comparison of the initial
model to the modified
model using performanceIncreased performance metrics of the
modified model for view or quality
criteria detection.
Benefits: Increased performance;
generalization for diverse cases.
Risks: Performance decrease
(overfitting, unintended bias).

Summary of changes to FETOLY-HEART per the PCCP:

14

Image /page/14/Picture/0 description: The image shows a blue logo for a company called "DIAGNOLY". The logo features a blue magnifying glass with a stack of three cone-shaped objects inside. The text "DIAGNOLY" is written in blue below the magnifying glass.

| | | metrics on the test
dataset. | Risk mitigation: The modified model
will be tested for superiority on the
performance study test dataset which
will contain new unseen data. |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Heart quality
criteria
addition/removal | Maintain alignment
with quality criteria
recommended for fetal
heart screening in
state-of-the-art
international
guidelines. | New quality criteria will be
controlled using the same
acceptance criteria as
defined by secondary
endpoints. | Enhanced compliance with standard
international guidelines.
Benefits: Keeping the device relevant
by aligning with the updated list of
heart quality criteria.
Risks: Performance decrease and user
confusion.
Risk mitigation: Proper performance
testing with no decrease in test
performance. This change only
pertains to quality criteria belonging
to one of the 5 heart views already
included in FETOLY-HEART. |

FETOLY-HEART performance study 7.2

Diagnoly conducted a standalone performance testing on a dataset of 2,288 fetal ultrasound images across 480 patient cases, including full examination still images, cardiac clip frames and full examination video frames from 7 clinical sites in the United States and cases are representative of the intended use population. This testing dataset originated from distinct clinical sites from which the data used during model development (training/validation) was sourced, ensuring testing independence.

The results of the standalone performance testing demonstrate that FETOLY-HEART automatically detects fetal heart ultrasound views with a sensitivity ≥ 85% (acceptance criterion) and specificity ≥ 85% (acceptance criterion), detects quality criteria within heart views with a sensitivity ≥ 90% (acceptance criterion) and a specificity ≥ 90% (acceptance criterion), and localizes bounding boxes of quality criteria with a mean intersection over union (loU) of ≥ 50% (acceptance criterion). Sensitivity and specificity were evaluated individually for each view, and the performance goal to exceed 85% as the lower bound of the corresponding 95% Confidence Interval (Cl) was met. The Cl was estimated using bootstrap resampling at the subject level based on 1,000 samples, with traditional bootstrap Cl confidence limits derived as the 2.5th and 97.5th percentiles of the distribution of bootstrap estimates. The results are summarized in the tables below:

SensitivitySpecificity
Fetal heart viewN(positive)Point
estimateBootstrap Cl
(95%)N(negative)Point
estimateBootstrap Cl
(95%)
Abdomen view4280.976(0.960,0.990)18600.998(0.996,1.000)

15

Image /page/15/Picture/0 description: The image shows a logo for a company called "Diagnoly". The logo features a blue magnifying glass with a stylized image of a sound wave inside. The company name "Diagnoly" is written in blue, sans-serif font below the magnifying glass.

Four chamber view3910.987(0.974,0.997)18971.00(1.000,1.000)
Left ventricular outflow
tract view3600.983(0.969,0.994)19280.999(0.998,1.000)
Right ventricular outflow
tract view3130.987(0.974,0.996)19750.998(0.996,1.000)
Three vessels view3160.981(0.965,0.993)19720.998(0.997,1.000)
Other view4800.985(0.972,0.995)18080.983(0.977,0.989)
SensitivitySpecificitymloU
Quality criterionNPoint estimateBootstrap Cl (95%)NPoint estimateBootstrap CI (95%)Point estimateBootstrap CI (95%)
Abdomen view
Spine4170.966(0.949,0.981)18710.995(0.992,0.998)0.739(0.728,0.750)
Left rib4010.903(0.873,0.933)18870.997(0.995,0.999)0.651(0.631,0.673)
Right rib3880.910(0.882,0.940)19000.997(0.994,0.999)0.624(0.602,0.643)
Descending aorta4170.947(0.925,0.969)18710.998(0.996,0.999)0.528(0.515,0.542)
Inferior vena cava3630.915(0.884,0.941)19250.992(0.988,0.996)0.512(0.496,0.528)
Stomach4140.976(0.961,0.988)18740.995(0.992,0.998)0.734(0.720,0.746)
Umbilical vein3450.962(0.942,0.980)19430.991(0.987,0.994)0.676(0.660,0.693)
Thorax apex3970.919(0.893,0.944)18910.993(0.989,0.997)0.571(0.549,0.592)
Four chamber view
Spine3390.965(0.944,0.982)19490.999(0.997,1.000)0.768(0.754,0.779)
Left rib3080.945(0.917,0.970)19800.996(0.993,0.999)0.744(0.726,0.764)
Right rib3090.958(0.934,0.980)19790.996(0.993,0.998)0.749(0.730,0.768)
Descending aorta3610.981(0.965,0.994)19270.996(0.994,0.999)0.646(0.632,0.659)
Left pulmonary vein1270.921(0.871,0.965)21610.998(0.995,1.000)0.628(0.601,0.653)
Right pulmonary vein1980.904(0.860,0.943)20900.996(0.993,0.999)0.601(0.578,0.623)
Left atrium3870.990(0.979,0.997)19010.998(0.996,1.000)0.759(0.747,0.771)
Right atrium3910.987(0.975,0.997)18971.00(1.000,1.000)0.774(0.763,0.786)
Foramen ovale flap1130.929(0.883,0.971)21751.00(0.999,1.000)0.530(0.505,0.561)
Open Foramen Ovale3480.951(0.928,0.972)19400.996(0.993,0.999)0.616(0.598,0.634)
Mitral valve2070.908(0.866,0.948)20810.998(0.995,1.000)0.676(0.654,0.695)
Tricuspid valve2430.959(0.931,0.983)20450.995(0.992,0.998)0.718(0.701,0.735)
Connection
between
crux and atrial septum2990.943(0.915,0.969)19890.990(0.986,0.994)0.587(0.567,0.603)
Atrioventricular valve
offset in crux1150.939(0.897,0.980)21730.992(0.988,0.995)0.648(0.623,0.673)
Connection between
interventricular
septum and crux1770.910(0.868,0.951)21110.996(0.993,0.999)0.565(0.544,0.585)

16

Image /page/16/Picture/0 description: The image shows a blue logo for a company called Diagnoly. The logo features a magnifying glass with a stack of three objects inside. The text "DIAGNOLY" is written in blue below the magnifying glass.

| Interventricular

septum216 0.917(0.877,0.949)2072 0.996(0.993,0.999)0.71(0.689,0.728)
Left ventricle389 0.987(0.975,0.997)1899 1.00(1.000,1.000)0.748(0.734,0.761)
Right ventricle387 0.990(0.979,0.997)1901 0.999(0.997,1.000)0.758(0.746,0.770)
Sternum322 0.972(0.952,0.988)1966 0.991(0.986,0.995)0.681(0.661,0.700)
Left ventricular outflow tract view
Left atrium356 0.98(0.966,0.992)1932 0.997(0.995,0.999)0.656(0.633,0.676)
Proximal ascending
aorta360 0.983(0.969,0.994)1928 0.999(0.998,1.000)0.716(0.702,0.730)
Semilunar valves165 0.909(0.866,0.950)2123 0.998(0.995,1.000)0.527(0.503,0.552)
Left ventricle358 0.978(0.961,0.992)1930 0.999(0.998,1.000)0.745(0.732,0.758)
Interventricular
septum199 0.925(0.888,0.959)2089 0.995(0.992,0.998)0.663(0.638,0.690)
Right ventricle358 0.980(0.964,0.992)1930 0.999(0.997,1.000)0.698(0.675,0.719)
Right ventricular outflow tract view
Descending aorta294 0.949(0.924,0.971)1994 0.997(0.994,0.999)0.617(0.595,0.638)
Trachea / bronchi180 0.922(0.882,0.960)2108 0.990(0.985,0.994)0.520(0.491,0.548)
Left pulmonary artery125 0.960(0.921,0.992)2163 0.995(0.992,0.998)0.615(0.582,0.646)
Ductus arteriosus177 0.944(0.907,0.974)2111 0.997(0.995,0.999)0.631(0.606,0.656)
Right
pulmonary
artery302 0.980(0.965,0.994)1986 0.994(0.991,0.997)0.654(0.633,0.675)
Origin of
pulmonary
arteries258 0.922(0.889,0.953)2030 0.997(0.994,0.999)0.589(0.568,0.610)
Septum
between
pulmonary
artery
trunk and
ascending
aorta218 0.945(0.911,0.973)2070 0.993(0.989,0.996)0.692(0.672,0.710)
Ascending aorta312 0.974(0.957,0.990)1976 0.998(0.996,1.000)0.633(0.613,0.650)
Superior vena cava218 0.940(0.910,0.971)2070 0.994(0.991,0.998)0.577(0.549,0.603)
Pulmonary trunk312 0.971(0.952,0.990)1976 0.999(0.998,1.000)0.760(0.744,0.775)
Three vessels views
Spine276 0.953(0.929,0.975)2012 0.998(0.996,1.000)0.792(0.779,0.805)
Trachea206 0.932(0.899,0.964)2082 0.997(0.994,0.999)0.539(0.517,0.562)
Side space on the left
of ductus270 0.933(0.901,0.961)2018 0.991(0.987,0.995)0.784(0.767,0.800)
Main
pulmonary
artery169 0.947(0.910,0.976)2119 0.999(0.998,1.000)0.733(0.708,0.755)
Ductus141 0.972(0.944,0.993)2147 0.998(0.996,1.000)0.749(0.724,0.771)
Ascending aorta181 0.978(0.954,0.995)2107 0.999(0.997,1.000)0.621(0.599,0.643)
Aortic arch133 0.97(0.942,0.993)2155 0.997(0.995,0.999)0.724(0.699,0.747)
Superior vena cava303 0.964(0.942,0.983)1985 0.996(0.994,0.999)0.528(0.508,0.547)
Thymus/sternum212 0.915(0.876,0.953)2076 0.998(0.995,1.000)0.716(0.693,0.737)

17

Image /page/17/Picture/1 description: The image shows a logo for a company called DIAGNOLY. The logo features a blue magnifying glass with a stack of three cone-shaped objects inside the lens. The word "DIAGNOLY" is written in blue below the magnifying glass.

The performance validation dataset included the following clinical subgroups: maternal age, gestational age, territory, BMI, scanner manufacturer, heart abnormality. It also comprised races and ethnicities and various clinical sites, ensuring representation across the intended use population. Image digital quality (bad, average, good) and image type (still image from entire examination, cardiac clip frame, full examination frame) were identified as potential controlled for. Performance metrics were analyzed for each subgroup and confounder to validate the model's robustness and generalizability. The subgroups distribution is summarized in the table below:

Number of casesNumber of images
Subgroup
(total=480)(total=2,288)
Center 199 (20.6%)505 (22.1%)
CenterCenter 263 (13.1%)248 (10.8%)
Center 384 (17.5%)378 (16.5%)
Center 455 (11.5%)264 (11.5%)
Center 554 (11.2%)275 (12.0%)
Center 655 (11.5%)265 (11.6%)
Center 770 (14.6%)353 (15.4%)
TerritoryUS234 (48.8%)1157 (50.6%)
EU246 (51.2%)1131 (49.4%)
Gestational age2nd trimester233 (48.5%)1189 (52.0%)
3rd trimester243 (50.6%)1075 (47.0%)
Unknown4 (0.8%)24 (1.0%)
Maternal age