(235 days)
Not Found
Yes
The device description explicitly states that Clarify DL is a "deep learning-based image reconstruction method." Deep learning is a subset of machine learning.
No
The device is an image reconstruction method intended to reduce noise and improve image quality for diagnostic purposes, not to provide therapy.
No
The device is described as an image reconstruction method intended to reduce noise and improve image quality for Nuclear Medicine bone SPECT scintigraphy images. Its purpose is to enhance the images for review, not to directly provide a medical diagnosis on its own.
Yes
The device description explicitly states "Clarify DL is a software-only device". It is deployed within the software of existing predicate systems and performs image reconstruction, which is a software function.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is to "reduce noise and improve image quality of Nuclear Medicine bone SPECT scintigraphy images." This is a function related to image processing and enhancement, not the analysis of biological samples (blood, urine, tissue, etc.) to diagnose or monitor a disease or condition.
- Device Description: The device is described as "software-only" for "reconstruction of NM bone SPECT scintigraphy images." This further reinforces its role in image manipulation rather than in vitro analysis.
- Input: The input is "Nuclear Medicine bone SPECT scintigraphy" images, which are generated from a medical imaging procedure, not from a biological sample.
IVD devices are specifically designed to perform tests on specimens taken from the human body. This device operates on medical images, which are a result of a medical imaging procedure, not a biological sample in the traditional sense of an IVD.
Yes
The letter explicitly states, "FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP)." This phrasing directly meets the criteria for "cleared the PCCP for this device."
Intended Use / Indications for Use
Clarify DL is a deep learning-based image reconstruction method intended to reduce noise and improve image quality of Nuclear Medicine bone SPECT scintigraphy images. Clarify DL may be used for patients of all ages.
Product codes (comma separated list FDA assigned to the subject device)
LLZ, KPS
Device Description
Clarify DL is a software-only device for the reconstruction of NM bone SPECT scintigraphy images obtained using supported GE HealthCare SPECT and SPECT-CT systems. Clarify DL is designed to reduce image noise while maintaining contrast to enable increased Contrast-to-Noise Ratio (CNR) and Contrast Recovery Coefficient (CRC). Clarify DL is deployed within the software of the predicate Xeleris V Processing and Review System and StarGuide devices.
Clarify DL includes a Predetermined Change Control Plan (PCCP) with one planned modification. The modification is for making Clarify DL available for imaging data acquired using a SPECT-CT system that incorporates a different CT. Clarify DL remains unchanged and maintains the same intended use, indications for use and equivalent performance. Performance evaluation for the modification includes testing to confirm equivalence of the input CT data and that Clarify DL's overall IQ performance is maintained and substantially equivalent to its performance with imaging data from a supported SPECT-CT system. The user documentation will be updated to reflect the modification (i.e. description of new source of data, summary of release and performance evaluation).
Mentions image processing
Clarify DL is a software-only device for the reconstruction of NM bone SPECT scintigraphy images obtained using supported GE HealthCare SPECT and SPECT-CT systems. Clarify DL is designed to reduce image noise and improve image quality of Nuclear Medicine bone SPECT scintigraphy images.
Mentions AI, DNN, or ML
Clarify DL is a deep learning-based image reconstruction method intended to reduce noise and improve image quality of Nuclear Medicine bone SPECT scintigraphy images.
OSEM-based Iterative Image Reconstruction with integrated deep learning model trained to reduce noise
OSEM-based and BSREM-based Iterative Image Reconstruction with integrated deep learning model trained to reduce noise
Input Imaging Modality
Nuclear Medicine bone SPECT scintigraphy images
Anatomical Site
bone
Indicated Patient Age Range
all ages
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
The clinical exams included in Clarify DL's clinical evaluation came from 11 hospitals, including 5 US hospitals from 4 different states (i.e Ohio, Kansas, Rhode Island, and California) and 6 non-US hospitals from Europe and Asia. The clinical reader evaluation included review by expert NM physicians to assess overall Image Quality, Image Resolution, and Noise Level.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Summary of Non-Clinical Testing
CDL has successfully completed the design control testing per GEHC's quality system. No additional hazards were identified, and no unexpected test results were observed. CDL was designed under the Quality System Regulations of 21CFR 820 and ISO 13485. GEHC believes that the extensive bench testing and the physician evaluations performed are sufficient for FDA's substantial equivalence determination. The following quality assurance measures have been applied to the development of the system:
- Requirement Definition
- Risk Analysis
- Technical Design Reviews
- Formal Design Reviews
- Software Development Lifecycle
- Testing on unit level (Module verification)
- Integration testing (System verification)
- System Testing:
- Safety Testing (Verification)
- System and Image Performance Testing (Verification)
- Simulating Use Testing (Validation)
Additional engineering bench testing included testing that demonstrates performance and substantiates the product claims. The testing confirmed substantial equivalent performance compared to the predicate devices' cleared image reconstruction for several acquisition settings. The testing included testing for image and lesion level characteristics, as follows:
- Image level testing included testing for Structure Similarity Index Measure (SSIM), Mean Squared Error (MSE), and Peak Signal-to-Noise Ratio (PSNR)
- Lesion level testing included testing for Contrast-to-Noise ratio (CNR) and Contrast Recovery Coefficient (CRC).
The additional engineering bench testing substantiated all performance claims. All the testing and results did not raise new or different questions of safety and effectiveness than associated with predicate devices. GEHC considers that the proposed Clarify DL device is of comparable type and substantially equivalent to the predicate devices.
Summary of Clinical Testing
The clinical testing included a clinical reader evaluation of clinical bone SPECT and SPECT-CT exams obtained using supported SPECT CT systems. The clinical exams included in Clarify DL's clinical evaluation came from 11 hospitals, including 5 US hospitals from 4 different states (i.e Ohio, Kansas, Rhode Island, and California) and 6 non-US hospitals from Europe and Asia. The clinical reader evaluation included review by expert NM physicians to assess overall Image Quality, Image Resolution, and Noise Level. The results of the clinical reader evaluation support the determination of substantial equivalence. All readers attested that their assessments of Clarify DL demonstrated acceptable diagnostic results.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
- Structure Similarity Index Measure (SSIM)
- Mean Squared Error (MSE)
- Peak Signal-to-Noise Ratio (PSNR)
- Contrast-to-Noise ratio (CNR)
- Contrast Recovery Coefficient (CRC)
- Overall Image Quality
- Image Resolution
- Noise Level
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Clarify DL includes a Predetermined Change Control Plan (PCCP) with one planned modification. The modification is for making Clarify DL available for imaging data acquired using a SPECT-CT system that incorporates a different CT. Clarify DL remains unchanged and maintains the same intended use, indications for use and equivalent performance. Performance evaluation for the modification includes testing to confirm equivalence of the input CT data and that Clarify DL's overall IQ performance is maintained and substantially equivalent to its performance with imaging data from a supported SPECT-CT system. The user documentation will be updated to reflect the modification (i.e. description of new source of data, summary of release and performance evaluation).
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device. or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 3, 2025
GE Medical Systems Israel, Functional Imaging % Maysana Mousa Elias Senior Regulatory Affairs Leader GE HealthCare 3000 N Grandview Blvd W 440 Waukesha. Wisconsin 53188
Re: K241350
Trade/Device Name: Clarify DL Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ, KPS Dated: May 13, 2024 Received: December 3, 2024
Dear Maysana Mousa Elias:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device. or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these
2
requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D. Rayfield
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
3
Indications for Use
Submission Number (if known)
Device Name
Clarify DL
Indications for Use (Describe)
Clarify DL is a deep learning-based image reconstruction method intended to reduce noise and improve image quality of Nuclear Medicine bone SPECT scintigraphy images. Clarify DL may be used for patients of all ages.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4
Image /page/4/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a purple circle with the letters "GE" in a stylized font in the center. There are decorative swirls around the letters and the edge of the circle. The logo is simple and recognizable.
K241350
510(k) SUMMARY OF SAFETY AND EFFECTIVNESS
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h):
Date: | January 03, 2025 |
---|---|
Submitter: | GE Medical Systems Israel, Functional Imaging (GE HealthCare) |
4 Hayozma Street | |
Tirat Hacarmel, 30200, Israel | |
Primary Contact: | Maysana Mousa Elias |
Senior Regulatory Affairs Leader | |
GE Medical Systems Israel, Functional Imaging | |
Tel: +972-4-8563666 | |
Mobile: +972-54-7850710 | |
Email: Maysana.MousaElias@gehealthcare.com | |
Secondary Contact: | John Jaeckle |
Chief Regulatory Affairs Strategist | |
GE Healthcare | |
Tel: 262-424-9547 | |
Email: john.jaeckle@gehealthcare.com | |
Device Trade Name: | Clarify DL |
Device Classification: | Class II |
Regulation Number: | 21CFR 892.1200 & 21CFR 892.2050 |
Product Codes: | KPS & LLZ |
Predicate Device Information
Predicate Device 1 | |
---|---|
Device Name: | Xeleris V Processing and Review System |
Manufacturer: | GE Medical Systems Israel, Functional Imaging |
510(k) Number: | K221680 |
Regulation Number/ Product Code: | 21CFR 892.2050 / LLZ |
Predicate Device 2 | |
Device Name: | StarGuide |
Manufacturer: | GE Medical Systems Israel, Functional Imaging |
510(k) Number: | K210173 |
Regulation Number/ Product Code: | 21CFR 892.1200 / KPS |
5
Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a purple circle with the letters "GE" in white inside. There are decorative white swirls around the letters. The logo is simple and recognizable.
21CFR 892.1750 / JAK
Device Description
Clarify DL is a software-only device for the reconstruction of NM bone SPECT scintigraphy images obtained using supported GE HealthCare SPECT and SPECT-CT systems. Clarify DL is designed to reduce image noise while maintaining contrast to enable increased Contrast-to-Noise Ratio (CNR) and Contrast Recovery Coefficient (CRC). Clarify DL is deployed within the software of the predicate Xeleris V Processing and Review System and StarGuide devices.
Clarify DL includes a Predetermined Change Control Plan (PCCP) with one planned modification. The modification is for making Clarify DL available for imaging data acquired using a SPECT-CT system that incorporates a different CT. Clarify DL remains unchanged and maintains the same intended use, indications for use and equivalent performance. Performance evaluation for the modification includes testing to confirm equivalence of the input CT data and that Clarify DL's overall IQ performance is maintained and substantially equivalent to its performance with imaging data from a supported SPECT-CT system. The user documentation will be updated to reflect the modification (i.e. description of new source of data, summary of release and performance evaluation).
Intended Use
Clarify DL is a deep learning-based image reconstruction method for Nuclear Medicine SPECT images.
Indications For Use
Clarify DL is a deep learning-based image reconstruction method intended to reduce noise and improve image quality of Nuclear Medicine bone SPECT scintigraphy images. Clarify DL may be used for patients of all ages.
Technology Characteristics
Clarify DL is an additional image reconstruction option for the predicate Xeleris V Processing and Review System and StarGuide devices. Clarify DL produces substantially equivalent images to those produced using the predicate devices' cleared image reconstruction. Because Clarify DL is deployed within the predicate Xeleris V and StarGuide devices software, it utilizes the same hardware and software platform as their cleared image reconstruction. The tables below summarize the substantive feature / technological differences between the predicate device and the proposed device:
Specification | Predicate Device 1 | Proposed Device |
---|---|---|
Xeleris V Processing and Review | ||
System (K221680) | Clarify DL | |
on Xeleris V | ||
Technology | OSEM-based Iterative Image | |
Reconstruction | OSEM-based Iterative Image | |
Reconstruction with integrated deep | ||
learning model trained to reduce noise |
6
GE Healthcare 510(k) Premarket Notification Submission for Clarify DL
Image /page/6/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a purple circle with the letters "GE" in white inside. There are four white swirls around the circle. The logo is simple and recognizable.
Specification | Predicate Device 2 | Proposed Device |
---|---|---|
StarGuide (K221680) | Clarify DL | |
including Smart Console | on StarGuide's Smart Console | |
Technology | OSEM-based and BSREM-based | OSEM-based and BSREM-based Iterative |
Iterative Image Reconstruction | Image Reconstruction with integrated deep | |
learning model trained to reduce noise |
Clarify DL's technological characteristics do not raise new questions of safety or effectiveness, and did not introduce any new risks/hazards, warnings, or limitations.
Determination of Substantial Equivalence
Summary of Non-Clinical Testing
CDL has successfully completed the design control testing per GEHC's quality system. No additional hazards were identified, and no unexpected test results were observed. CDL was designed under the Quality System Regulations of 21CFR 820 and ISO 13485. GEHC believes that the extensive bench testing and the physician evaluations performed are sufficient for FDA's substantial equivalence determination. The following quality assurance measures have been applied to the development of the system:
- । Requirement Definition
- Risk Analysis ।
- -Technical Design Reviews
- । Formal Design Reviews
- Software Development Lifecycle ।
- Testing on unit level (Module verification) -
- Integration testing (System verification) ।
- System Testing: ।
- Safety Testing (Verification) ı
- System and Image Performance Testing (Verification)
- Simulating Use Testing (Validation) i
Additional engineering bench testing included testing that demonstrates performance and substantiates the product claims. The testing confirmed substantial equivalent performance compared to the predicate devices' cleared image reconstruction for several acquisition settings. The testing included testing for image and lesion level characteristics, as follows:
- Image level testing included testing for Structure Similarity Index Measure (SSIM), Mean Squared Error (MSE), and Peak Signal-to-Noise Ratio (PSNR)
- Lesion level testing included testing for Contrast-to-Noise ratio (CNR) and Contrast Recovery ı Coefficient (CRC).
7
Image /page/7/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized font, enclosed within a circle. The circle is purple and has a white border with decorative swirls.
The additional engineering bench testing substantiated all performance claims. All the testing and results did not raise new or different questions of safety and effectiveness than associated with predicate devices. GEHC considers that the proposed Clarify DL device is of comparable type and substantially equivalent to the predicate devices.
Summary of Clinical Testing
The clinical testing included a clinical reader evaluation of clinical bone SPECT and SPECT-CT exams obtained using supported SPECT CT systems. The clinical exams included in Clarify DL's clinical evaluation came from 11 hospitals, including 5 US hospitals from 4 different states (i.e Ohio, Kansas, Rhode Island, and California) and 6 non-US hospitals from Europe and Asia. The clinical reader evaluation included review by expert NM physicians to assess overall Image Quality, Image Resolution, and Noise Level. The results of the clinical reader evaluation support the determination of substantial equivalence. All readers attested that their assessments of Clarify DL demonstrated acceptable diagnostic results.
Substantial Equivalence Conclusion
The changes associated with Clarify DL do not change the indications for use and represent technological characteristics that produce images of equivalent diagnostic quality.
GEHC's quality system's design verification, and risk management processes did not identify any new questions of safety or effectiveness, hazards, unexpected results, or adverse effects stemming from the changes to the predicate.
Design verification, along with the bench testing and the clinical reader evaluation included in the "Performance Testing" section of this submission, demonstrate that Clarify DL is substantially equivalent to, and hence as safe and as effective for its intended use as the legally marketed predicate devices.
All the testing and results did not raise new or different questions of safety and effectiveness than associated with predicate device. GEHC considers the proposed device is substantially equivalent to the predicate devices.