K Number
K241178
Date Cleared
2024-05-24

(25 days)

Product Code
Regulation Number
880.5730
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DEKA ACE Pump System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, ages six and above. The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, and confirm commands from these devices. The pump is intended for single patient, home use and requires a prescription.

The bolus calculator is indicated for use for aiding the user in determining the bolus insulin dosage for management of diabetes mellitus based on consumed carbohydrates, operator entered blood glucose, insulin sensitivity, insulin to carbohydrate ratio, target glucose values, and current insulin on board.

Device Description

The DEKA ACE Pump System described herein contains a modification of the labeling of the previously cleared DEKA ACE Pump System (K233952) to add Novolog U-100 (insulin aspart) as a compatible insulin. The device is a wearable alternate controller enabled (ACE) insulin infusion pump intended to subcutaneously deliver insulin. at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The pump is able to reliably and securely communicate with compatible, digitally connected devices to receive, confirm, and execute commands. The pump is intended for single patient, ambulatory use and requires a prescription. The DEKA ACE Pump System is intended for the management of diabetes mellitus in persons six years of age and greater.

The pump was previously cleared and is indicated for use with Humalog U-100 insulin. This Special 510(k) utilizes the same methodology and acceptance criteria used to obtain the clearance and indication for Humalog U-100 to add the indication for Novolog U-100.

The DEKA ACE Pump System, consistent with the predicate K233952, consists of the following durable and disposable components:

  1. Pump: A durable pump that incorporates fluid delivery algorithms and interfaces to a cassette, external wireless user interface, and iCGM. The pump is powered by a rechargeable lithium ion battery.
  2. Cassette: A single-use pumping cassette that combines microfluidic valves, a pump chamber, insulin reservoir, and Acoustic Volume Sensing (AVS) measurement chamber. The cassette interfaces to a pump and off-the-shelf infusion sets.
  3. DEKA Loop App: An iOS mobile application that serves as the primary user interface for the system. The DEKA Loop app can be downloaded onto the user's personal iPhone.

Also consistent with the predicate, the DEKA ACE Pump system includes the following electronic interfaces:

  • Dexcom G6 iCGM: The DEKA ACE Pump System is compatible with Dexcom G6 iCGMs. Communication between the Dexcom G6 and DEKA ACE Pump is unchanged from the predicate device. The BLE central role on the ACE Pump radio processor connects directly to the Dexcom iCGM transmitter using the sensor's medical slot. The ACE Pump communicates with the iCGM transmitter at a regular interval to provide iCGM sensor data to the iAGC.
  • Halo Cloud: Halo Cloud is a digital platform that connects the DEKA ACE Pump System to a variety of cloud-related services. These services include:
  • Patient onboarding and device pairing via secure key transfer
  • Prescription setting downloads to the ACE Pump
  • Event log uploads from the ACE pump to the cloud
  • Remote (OTA) software updates

No changes to the hardware or software of the system from that of the predicate are necessary to add the Novolog U-100 compatibility claim.

AI/ML Overview

The provided text describes a 510(k) submission for the DEKA ACE Pump System, specifically seeking to add Novolog U-100 (insulin aspart) as a compatible insulin. The core of this submission is to demonstrate substantial equivalence to a previously cleared version of the same pump (K233952), which was cleared for use with Humalog U-100 insulin.

Based on the provided text, the device in question is an insulin pump, not an AI/Software as a Medical Device (SaMD) for image analysis or diagnosis. Therefore, many of the typical acceptance criteria and study aspects related to AI/SaMD (like MRMC studies, expert ground truth establishment for diagnostic images, sample sizes for training/test sets for AI models, etc.) are not applicable to this type of medical device submission.

The "acceptance criteria" in this context refer to the demonstration that the modified device (with Novolog U-100 compatibility) is substantially equivalent to the predicate device (with Humalog U-100 compatibility) in terms of safety and effectiveness, and that it meets all applicable Special Controls for Alternate Controller Enabled Infusion Pumps (21 CFR 880.5730, Product Code QFG).

Here's an attempt to answer the questions based on the provided document, highlighting what is and isn't applicable:

  1. A table of acceptance criteria and the reported device performance

    The document doesn't present a table of quantitative acceptance criteria and reported numerical performance for a new study, but rather asserts that the same methodology and acceptance criteria used for Humalog U-100 were applied to Novolog U-100, implying the new performance matches the previously accepted thresholds. The comparison focuses on demonstrating equivalence across various characteristics.

    Implicit Acceptance Criteria (based on equivalence to predicate and Special Controls):

    • Indications for Use: No change.
    • Prescription Use: Yes.
    • Intended Population: Persons with Diabetes Mellitus ages 6 and above.
    • Environment of Use: Professional healthcare facilities and home healthcare environments.
    • Patient Environment: On-body wearable ambulatory pump.
    • Delivery Method: Microprocessor controlled Micro-dosing pump mechanism supplemented with acoustic volume sensor (AVS) feedback for monitoring delivery accuracy.
    • Insulin Basal Rate Delivery Range: 0 units/hour - 30 units/hour.
    • Insulin Bolus Delivery Range: Programmable from 0.05 - 25.00 Units in 0.01 Unit increments.
    • Basal Accuracy: Unchanged from K213536 (predicate's reference).
    • Bolus Accuracy: Unchanged from K213536 (predicate's reference).
    • Bolus Volume after Occlusion Release: No more than 0.74 units.
    • Time to occlusion alarm: 10 min (Bolus); 3 hours (Basal, 1 U/h); 6 hours (Basal, 0.1 U/hr).
    • Material Biocompatibility: Compliant with ISO-10993.
    • Cartridge/Cassette Shelf Life: 1 year.
    • Ingress Protection: IP28.
    • Applicable Safety Standards: Compliance with listed IEC and ISO standards (e.g., IEC 60601-1, ISO 14971).
    • Power Source: Rechargeable Lithium Ion Battery.
    • Pump Storage Conditions: Temperatures of -25 °C to 70 °C, Non-condensing humidity 15% to 90%.
    • Operating Conditions: Temperatures of 5 °C to 40 °C, Non-condensing humidity of 15% to 90%.
    • System User Feedback: Visual, audible, and vibratory.
    • Battery Operating Time: 72 hours.
    • New "Acceptance": The addition of Novolog U-100 compatibility does not impact the safety and effectiveness, and testing with Novolog U-100 is equivalent to that performed with Humalog U-100.

    Reported Device Performance (based on assertion of equivalence):
    For all the criteria listed above, the "Subject Device" (DEKA ACE Pump with Novolog U-100 compatibility) is stated to have the exact same characteristics and performance as the "Predicate Device" (DEKA ACE Pump with Humalog U-100 compatibility), or that its performance is equivalent to the predicate's established performance for the relevant aspects (e.g., Novolog U-100 testing being equivalent to Humalog U-100 testing). The key is that "No new or modified risks" are introduced.

  2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document states: "No clinical data was obtained in support of this premarket submission."
    It also states that "Performance testing was performed in order to establish substantial equivalence for the compatibility claim of Novolog U-100 in comparison to the previously cleared compatibility claim for Humalog U-100 in terms of both safety and effectiveness..."

    This heavily implies in vitro or bench testing rather than human subject testing. The "test set" in this context would likely refer to the number of tests/replicates performed with Novolog U-100 to demonstrate equivalent performance to Humalog U-100, not a dataset for an AI model. This information (specific sample sizes for bench testing, e.g., number of pump cycles, number of insulin lots tested) is not provided in this summary document. Data provenance for such non-clinical testing is typically internal lab data, not human patient data with country of origin.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is not an AI/SaMD for diagnostic image analysis. Ground truth would be established through engineering specifications, validated test methods, and measurement standards for pump performance, not expert consensus on medical images.

  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not an AI/SaMD for diagnostic image analysis.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/SaMD for diagnostic image analysis. The device is an insulin pump.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device has algorithms (fluid delivery, bolus calculator) but it's a physical pump with user interaction, not a standalone diagnostic algorithm being evaluated. Its performance is inherent to the pump's mechanical and electronic design working with insulin.

  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the performance of the insulin pump (e.g., basal/bolus accuracy, occlusion detection, battery life) would be established through engineering specifications, standardized test methods, and quantifiable measurements (e.g., precision volumetric measurements, pressure sensor readings, time-to-event logging). For this submission, the "ground truth" for Novolog U-100 compatibility is that its chemical and physical properties allow the pump to perform identically or equivalently to how it performs with Humalog U-100, which has already been established as safe and effective.

  8. The sample size for the training set

    Not applicable. This device does not use machine learning that requires a "training set" in the conventional AI sense for diagnostic tasks. The "training" of the pump would be its design, development, and validation testing based on engineering principles and regulatory standards.

  9. How the ground truth for the training set was established

    Not applicable. As above, there isn't a "training set" in the AI sense. The "ground truth" for the pump's design and function comes from established medical device standards, diabetes management requirements, and rigorous engineering testing.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 24, 2024

DEKA Research and Development Paul Smolenski Director of Regulatory Affairs 340 Commercial Street Manchester, New Hampshire 03101

Re: K241178

Trade/Device Name: DEKA ACE Pump System Regulation Number: 21 CFR 880.5730 Regulation Name: Alternate Controller Enabled Infusion Pump Regulatory Class: Class II Product Code: QFG, NDC Dated: April 26, 2024 Received: April 29, 2024

Dear Paul Smolenski:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joshua Balsam -S

Joshua M. Balsam, Ph.D. Branch Chief for Diabetes Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Submission Number (if known)

K241178

Device Name

DEKA ACE Pump System

Indications for Use (Describe)

The DEKA ACE Pump System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, ages six and above. The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, and confirm commands from these devices. The pump is intended for single patient, home use and requires a prescription.

The bolus calculator is indicated for use for aiding the user in determining the bolus insulin dosage for management of diabetes mellitus based on consumed carbohydrates, operator entered blood glucose, insulin sensitivity, insulin to carbohydrate ratio, target glucose values, and current insulin on board.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitter Information

510(k) Sponsor: DEKA Research & Development 340 Commercial Street Manchester, NH 03101

Contact Person: Paul Smolenski Regulatory Affairs DEKA Research & Development Corporation Phone: (603) 669-5139 Fax: (603) 624-0573 psmolenski@dekaresearch.com

Date Prepared: April 26, 2024

Proposed Device

Common/Usual Name: ACE Pump Trade/Proprietary Name: DEKA ACE Pump System Classification Name: Alternate Controller Enabled ACE Pump Device Classification: 880.5730 Product Code: QFG Class: II Device Panel: Clinical Chemistry

Predicate Device

The predicate device for this submission is the DEKA ACE Pump System cleared on 3/13/2024 under 510(k) K233952.

Device Description

The DEKA ACE Pump System described herein contains a modification of the labeling of the previously cleared DEKA ACE Pump System (K233952) to add Novolog U-100 (insulin aspart) as a compatible insulin. The device is a wearable alternate controller enabled (ACE) insulin infusion pump intended to subcutaneously deliver insulin. at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The pump is able to reliably and securely communicate with compatible, digitally connected devices to receive, confirm, and execute commands. The pump is intended for single patient, ambulatory use and requires a prescription. The DEKA ACE Pump System is intended for the management of diabetes mellitus in persons six years of age and greater.

The pump was previously cleared and is indicated for use with Humalog U-100 insulin. This Special 510(k) utilizes the same methodology and acceptance criteria used to obtain the clearance and indication for Humalog U-100 to add the indication for Novolog U-100.

{4}------------------------------------------------

The DEKA ACE Pump System, consistent with the predicate K233952, consists of the following durable and disposable components:

    1. Pump: A durable pump that incorporates fluid delivery algorithms and interfaces to a cassette, external wireless user interface, and iCGM. The pump is powered by a rechargeable lithium ion battery.
    1. Cassette: A single-use pumping cassette that combines microfluidic valves, a pump chamber, insulin reservoir, and Acoustic Volume Sensing (AVS) measurement chamber. The cassette interfaces to a pump and off-the-shelf infusion sets.
    1. DEKA Loop App: An iOS mobile application that serves as the primary user interface for the system. The DEKA Loop app can be downloaded onto the user's personal iPhone.

Also consistent with the predicate, the DEKA ACE Pump system includes the following electronic interfaces:

  • Dexcom G6 iCGM: The DEKA ACE Pump System is compatible with Dexcom G6 iCGMs. Communication between the Dexcom G6 and DEKA ACE Pump is unchanged from the predicate device. The BLE central role on the ACE Pump radio processor connects directly to the Dexcom iCGM transmitter using the sensor's medical slot. The ACE Pump communicates with the iCGM transmitter at a regular interval to provide iCGM sensor data to the iAGC.
  • . Halo Cloud: Halo Cloud is a digital platform that connects the DEKA ACE Pump System to a variety of cloud-related services. These services include:
    • Patient onboarding and device pairing via secure key transfer o
    • Prescription setting downloads to the ACE Pump o
    • Event log uploads from the ACE pump to the cloud о
    • Remote (OTA) software updates o

No changes to the hardware or software of the system from that of the predicate are necessary to add the Novolog U-100 compatibility claim.

Information is being supplied in this 510(k) premarket submission to demonstrate that the device is substantially equivalent in safety and effectiveness through comparison of indications for use and technological characteristics to the predicate DEKA ACE Pump System cleared on 3/13/2024 under K233952. As described throughout this submission, the subject DEKA ACE Pump System meets the product definition and all of the Special Controls defined in 21CFR 880.5730 for Alternate Controller Enabled Insulin Infusion Pumps, Product Code QFG.

Substantial Equivalence Discussion

Intended Use Comparison

The table below compares the intended use of the subject device to that of the predicate device:

{5}------------------------------------------------

CharacteristicPredicate DeviceSubject Device
Indications for UseThe DEKA ACE Pump System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, ages 6 and above. The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices. The pump is intended for single patient, home use and requires a prescription.The bolus calculator is indicated for use for aiding the user in determining the bolus insulin dosage for management of diabetes mellitus based on consumed carbohydrates, operator entered blood glucose, insulin sensitivity, insulin to carbohydrate ratio, target glucose values, and current insulin on board.The DEKA ACE Pump System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, ages 6 and above. The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices. The pump is intended for single patient, home use and requires a prescription.The bolus calculator is indicated for use for aiding the user in determining the bolus insulin dosage for management of diabetes mellitus based on consumed carbohydrates, operator entered blood glucose, insulin sensitivity, insulin to carbohydrate ratio, target glucose values, and current insulin on board.
Prescription UseYesYes
Intended PopulationPersons with Diabetes Mellitus ages 6 and abovePersons with Diabetes Mellitus ages 6 and above
Environment of UseIn professional healthcare facilities and home healthcare environmentsIn professional healthcare facilities and home healthcare environments

Discussion of differences in Indications for Use statements

The indications for use are unchanged for this submission, and the statements for both the predicate and subject devices are identical.

{6}------------------------------------------------

The indications for use for the predicate and subject device are as follows:

The DEKA ACE Pump System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin, ages 6 and above. The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, and confirm commands from these devices. The pump is intended for single patient, home use and requires a prescription.

The bolus calculator is indicated for use for aiding the user in determining the bolus insulin dosage for management of diabetes mellitus based on consumed carbohydrates, operator entered blood glucose, insulin sensitivity, insulin to carbohydrate ratio, target glucose values, and current insulin on board.

{7}------------------------------------------------

Discussion of differences in intended populations

Both the predicate and subject device are intended for use in persons with Diabetes Mellitus ages 6 and above. There are no differences in the intended populations.

Discussion of differences in environments of use

Both the predicate and subject devices are intended to be used in professional healthcare facilities and home healthcare environments. There are no differences in the environments of use.

Comparison of Technological Characteristics

The table below compares the intended use and technological characteristics of the subject device with that of the predicate device:

{8}------------------------------------------------

CharacteristicPredicate DeviceSubject DeviceEquivalence
DeviceClassificationRegulation andProduct CodeAlternate Controller EnabledInfusion Pump cleared under 21CFR 880.5730, Procode QFGAlternate Controller EnabledInfusion Pump submitted under 21CFR 880.5730, Procode QFGSame
Indications for UseThe DEKA ACE Pump System isintended for the subcutaneousdelivery of insulin, at set andvariable rates, for the managementof diabetes mellitus in personsrequiring insulin, ages six andabove. The pump is able toreliably and securely communicatewith compatible, digitallyconnected devices, includingautomated insulin dosing software,to receive, execute, and confirmcommands from these devices.The pump is intended for singlepatient, home use and requires aprescription.The DEKA ACE Pump System is Sameintended for the subcutaneousdelivery of insulin, at set andvariable rates, for the managementof diabetes mellitus in personsrequiring insulin, ages six andabove. The pump is able toreliably and securely communicatewith compatible, digitallyconnected devices, includingautomated insulin dosing software,to receive, execute, and confirmcommands from these devices.The pump is intended for singlepatient, home use and requires aprescription.Same
Prescription UseYesYesSame
Intended PopulationPersons with Diabetes Mellitusages 6 and abovePersons with Diabetes Mellitusages 6 and aboveSame
PatientEnvironmentOn-body wearable ambulatorypumpOn-body wearable ambulatorypumpSame
Environment ofUseIn professional healthcare facilitiesand home environmentsIn professional healthcare facilitiesand home environmentsSame
Delivery MethodMicroprocessor controlled Micro-dosing pump mechanismsupplemented with acousticMicroprocessor controlled Micro-dosing pump mechanismsupplemented with acousticSame
CharacteristicPredicate DeviceSubject DeviceEquivalence
Insulin Basal RateDelivery Rangevolume sensor (AVS) feedback formonitoring delivery accuracy.0 units/hour - 30 units/hourvolume sensor (AVS) feedback formonitoring delivery accuracy.0 units/hour - 30 units/hourSame
Insulin BolusDelivery RangeProgrammable from 0.05 - 25.00Units in 0.01 Unit increments.Programmable from 0.05 - 25.00Units in 0.01 Unit increments.Same
Basal AccuracyUnchanged from K213536Unchanged from K213536Same
Bolus AccuracyUnchanged from K213536Unchanged from K213536Same
Bolus Volume afterOcclusion ReleaseNo more than 0.74 units.No more than 0.74 units.Same
Time to occlusionalarm10 min (Bolus); 3 hours (Basal, 1U/h); 6 hours (Basal, 0.1 U/hr)10 min (Bolus); 3 hours (Basal, 1U/h); 6 hours (Basal, 0.1 U/hr)Same
MaterialBiocompatibilityCompliant with ISO-10993Compliant with ISO-10993Same
Cartridge/CassetteShelf Life1 year1 yearSame
Ingress ProtectionIP28, indicating protection fromcontinuous immersion in water.The pump can tolerate immersionto depths of up to 12 feet (3.7 m)for 1 hour.IP28, indicating protection fromcontinuous immersion in water.The pump can tolerate immersionto depths of up to 12 feet (3.7 m)for 1 hour.Same
Applicable SafetyStandards• IEC 60601-1• IEC 60601-1-2• IEC 60601-1-6• IEC 60601-1-8• IEC 60601-1-10• IEC 60601-1-11• IEC 60601-2-24• IEC 60601-1• IEC 60601-1-2• IEC 60601-1-6• IEC 60601-1-8• IEC 60601-1-10• IEC 60601-1-11• IEC 60601-2-24Same
CharacteristicPredicate DeviceSubject DeviceEquivalence
• ISO 11137-1 (Sterilized via Gamma Radiation)• ISO 10993-1• ISO 14971• ISO 11137-1 (Sterilized via Gamma Radiation)• ISO 10993-1• ISO 14971
Power SourceRechargeable Lithium Ion BatteryRechargeable Lithium Ion BatterySame
Pump Storage ConditionsTemperatures of -25 °C (-13 °F) to 70 °C (158 °F)Non-condensing humidity 15% to 90%Temperatures of -25 °C (-13 °F) to 70 °C (158 °F)Non-condensing humidity 15% to 90%Same
Operating ConditionsTemperatures of 5 °C (41 °F) to 40 °C (104 °F)Non-condensing humidity of 15% to 90%Temperatures of 5 °C (41 °F) to 40 °C (104 °F)Non-condensing humidity of 15% to 90%Same
System User FeedbackVisual, audible, and vibratoryVisual, audio, and vibratorySame
Battery Operating Time72 hours72 hoursSame
Compatible InsulinsHumalog U-100Humalog U-100Novolog U-100The addition of a Novolog U-100 compatibility claim to the labeling does not impact the safety and effectiveness of the device.Testing and results performed with Novolog U-100 in this submission are equivalent to that performed with Humalog U-100 in the predicate clearance. No new or modified risks.

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

Non-Clinical / Performance Testing:

Performance testing was performed in order to establish substantial equivalence for the compatibility claim of Novolog U-100 in comparison to the previously cleared compatibility claim for Humalog U-100 in terms of both safety and effectiveness, and to ensure the subject device met all applicable Special Controls. Performance testing from the predicate submission, K233952, is applicable to the subject submission.

Clinical Study

No clinical data was obtained in support of this premarket submission.

Design Control

The DEKA ACE Pump System was specified and developed by DEKA. DEKA complies with the FDA Quality System Regulation as specified in 21 CFR 820, as well as ISO 13485.

Conclusion

The DEKA ACE Pump System is substantially equivalent to the predicate cleared on 3/13/2024 in K233952. The addition of the claim for use with Novolog U-100 does not raise different questions of safety or effectiveness. The performance of the device is supported by DEKA's design control process which included non-clinical testing and risk management activities. The device meets all Special Controls for this product type as required by 21 CFR 880.5730 for Alternate Controller Enabled Insulin Infusion Pumps, Product Code QFG.

§ 880.5730 Alternate controller enabled infusion pump.

(a)
Identification. An alternate controller enabled infusion pump (ACE pump) is a device intended for the infusion of drugs into a patient. The ACE pump may include basal and bolus drug delivery at set or variable rates. ACE pumps are designed to reliably and securely communicate with external devices, such as automated drug dosing systems, to allow drug delivery commands to be received, executed, and confirmed. ACE pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of drug delivery.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Evidence demonstrating that device infusion delivery accuracy conforms to defined user needs and intended uses and is validated to support safe use under actual use conditions.
(A) Design input requirements must include delivery accuracy specifications under reasonably foreseeable use conditions, including ambient temperature changes, pressure changes (
e.g., head-height, backpressure, atmospheric), and, as appropriate, different drug fluidic properties.(B) Test results must demonstrate that the device meets the design input requirements for delivery accuracy under use conditions for the programmable range of delivery rates and volumes. Testing shall be conducted with a statistically valid number of devices to account for variation between devices.
(ii) Validation testing results demonstrating the ability of the pump to detect relevant hazards associated with drug delivery and the route of administration (
e.g., occlusions, air in line, etc.) within a clinically relevant timeframe across the range of programmable drug delivery rates and volumes. Hazard detection must be appropriate for the intended use of the device and testing must validate appropriate performance under the conditions of use for the device.(iii) Validation testing results demonstrating compatibility with drugs that may be used with the pump based on its labeling. Testing must include assessment of drug stability under reasonably foreseeable use conditions that may affect drug stability (
e.g., temperature, light exposure, or other factors as needed).(iv) The device parts that directly or indirectly contact the patient must be demonstrated to be biocompatible. This shall include chemical and particulate characterization on the final, finished, fluid contacting device components demonstrating that risk of harm from device-related residues is reasonably low.
(v) Evidence verifying and validating that the device is reliable over the ACE pump use life, as specified in the design file, in terms of all device functions and in terms of pump performance.
(vi) The device must be designed and tested for electrical safety, electromagnetic compatibility, and radio frequency wireless safety and availability consistent with patient safety requirements in the intended use environment.
(vii) For any device that is capable of delivering more than one drug, the risk of cross-channeling drugs must be adequately mitigated.
(viii) For any devices intended for multiple patient use, testing must demonstrate validation of reprocessing procedures and include verification that the device meets all functional and performance requirements after reprocessing.
(2) Design verification and validation activities must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following:
(i) Risk control measures shall be implemented to address device system hazards and the design decisions related to how the risk control measures impact essential performance shall be documented.
(ii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design.
(3) The device shall include validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following:
(i) Secure authentication (pairing) to external devices.
(ii) Secure, accurate, and reliable means of data transmission between the pump and connected devices.
(iii) Sharing of necessary state information between the pump and any digitally connected alternate controllers (
e.g., battery level, reservoir level, pump status, error conditions).(iv) Ensuring that the pump continues to operate safely when data is received in a manner outside the bounds of the parameters specified.
(v) A detailed process and procedure for sharing the pump interface specification with digitally connected devices and for validating the correct implementation of that protocol.
(4) The device must include appropriate measures to ensure that safe therapy is maintained when communications with digitally connected alternate controller devices is interrupted, lost, or re-established after an interruption (
e.g., reverting to a pre-programmed, safe drug delivery rate). Validation testing results must demonstrate that critical events that occur during a loss of communications (e.g., commands, device malfunctions, occlusions, etc.) are handled appropriately during and after the interruption.(5) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the system must, at a minimum, include:
(i) A record of all drug delivery
(ii) Commands issued to the pump and pump confirmations
(iii) Device malfunctions
(iv) Alarms and alerts and associated acknowledgements
(v) Connectivity events (
e.g., establishment or loss of communications)(6) Design verification and validation must include results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use.
(7) Device labeling must include the following:
(i) A prominent statement identifying the drugs that are compatible with the device, including the identity and concentration of those drugs as appropriate.
(ii) A description of the minimum and maximum basal rates, minimum and maximum bolus volumes, and the increment size for basal and bolus delivery, or other similarly applicable information about drug delivery parameters.
(iii) A description of the pump accuracy at minimum, intermediate, and maximum bolus delivery volumes and the method(s) used to establish bolus delivery accuracy. For each bolus volume, pump accuracy shall be described in terms of the number of bolus doses measured to be within a given range as compared to the commanded volume. An acceptable accuracy description (depending on the drug delivered and bolus volume) may be provided as follows for each bolus volume tested, as applicable: Number of bolus doses with volume that is 250 percent of the commanded amount.
(iv) A description of the pump accuracy at minimum, intermediate, and maximum basal delivery rates and the method(s) used to establish basal delivery accuracy. For each basal rate, pump accuracy shall be described in terms of the amount of drug delivered after the basal delivery was first commanded, without a warmup period, up to various time points. The information provided must include typical pump performance, as well as worst-case pump performance observed during testing in terms of both over-delivery and under-delivery. An acceptable accuracy description (depending on the drug delivered) may be provided as follows, as applicable: The total volume delivered 1 hour, 6 hours, and 12 hours after starting delivery for a typical pump tested, as well as for the pump that delivered the least and the pump that delivered the most at each time point.
(v) A description of delivery hazard alarm performance, as applicable. For occlusion alarms, performance shall be reported at minimum, intermediate, and maximum delivery rates and volumes. This description must include the specification for the longest time period that may elapse before an occlusion alarm is triggered under each delivery condition, as well as the typical results observed during performance testing of the pumps.
(vi) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device.
(vii) For any infusion pumps intended for multiple patient reuse, instructions for safely reprocessing the device between uses.