K Number
K240943
Device Name
LungVision
Date Cleared
2024-10-01

(179 days)

Product Code
Regulation Number
892.2050
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LungVision System is intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.

Device Description

The Lung Vision System is designed to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.

The System is intended to assist the guidance of endobronchial tools to areas of interest inside a patient's lungs. The System allows the user to mark lesion locations and pathways to marked lesions using a patient's CT scan. During the endoscopic procedure, the System overlays planning information on real-time fluoroscopic images to guide endobronchial tool navigation. The System also provides tomographic images for lesion identification, as well as 3D views for understanding tool and lesion proximity and orientation. The System is designed to be integrated with fluoroscopic imaging systems and external displays.

The Lung Vision system includes a main unit and a tablet. Image processing algorithms are executed on the main unit and the tablet is used as a primary method of interacting with the system.

AI/ML Overview

The provided text describes specific acceptance criteria and the studies conducted to demonstrate the subject device's performance.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance (Mean Accuracy ± STD)
AI-Tomo Accuracy (Simulated cases)3.15 mm ± 3.2 mm (N=500)
AI-Tomo Accuracy (Rigid body model cases)3.64 mm ± 1.57 mm (N=93)
AI-Tomo Accuracy (CBCT scans cases)5.34 mm ± 3.32 mm (N=191)
Lesion marking accuracy (AI-Tomo vs. CABT)Comparable

2. Sample Size and Data Provenance for Test Set

  • Simulated Cases: 500 cases, based on "real human's CT scans," synthetically simulated.
  • Rigid Body Model Cases: 93 cases, using "only one lung anatomy." Data provenance is laboratory testing in a simulated environment.
  • CBCT Scans Cases: 191 cases, based on "real human's CT scans and real human's CBCT scans taken in real procedures." The images are CABTs collected from real cases performed by physicians using the LungVision system in the USA, Italy, and Israel. This represents retrospective real-world data.

3. Number of Experts and Qualifications for Ground Truth Establishment

The document mentions "clinical validation was performed by physicians" but does not specify the exact number of experts or their qualifications (e.g., years of experience, specific medical specialty).

4. Adjudication Method for Test Set

The document does not explicitly state an adjudication method (e.g., 2+1, 3+1). It states that the output (from AI-Tomo post-processing) is "evaluated using the ground truthing methodology." The ground truth methods are listed as "Geometry test" and "Lesion contrast on real data tests (with and without tool)," implying a direct comparison against a defined standard rather than an expert consensus adjudication protocol.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The document states, "The clinical validation of AI Tomo confirmed that lesion marking accuracy was comparable when using CABT and AI Tomo images." This implies a comparison between human performance with CABT and human performance with AI-Tomo (assisted by physicians), but it does not explicitly detail an MRMC study comparing human readers with and without AI assistance (i.e., effect size of human improvement with AI assistance). It focuses on the comparison of the output of two imaging modalities for similar accuracy.

6. Standalone (Algorithm Only) Performance

Yes, standalone (algorithm only) performance was conducted for AI-Tomo. The accuracy metrics (mean accuracy and standard deviation) presented for simulated, rigid body model, and CBCT scan cases refer to the performance of the AI-Tomo algorithm itself in reconstructing images or guiding, based on input. The statement "The images are CABTs collected from real cases ... and the output is evaluated using the ground truthing methodology" further supports this, indicating the AI's output was evaluated against a ground truth.

7. Type of Ground Truth Used

The ground truth used includes:

  • Geometry test: This likely refers to a predefined geometric standard or measurement.
  • Lesion contrast on real data tests (with and without tool): This suggests evaluation of the AI's ability to delineate lesions and maintain contrast.
  • For the simulated and rigid body model cases, the ground truth would be precisely known due to the controlled nature of these environments (e.g., in simulated cases, the intended registration error, or in rigid body models, the pre-defined transform).
  • For CBCT cases, the ground truth methodology is stated, but the ultimate true "lesion location" reference is implied to be established through the "Geometry test" and "Lesion contrast" evaluations.

8. Sample Size for Training Set

The document does not specify the sample size for the training set used to develop the AI-Tomo feature. It mentions that "CABTs are used as input data to the AI-Tomo post processing," implying these real CABT cases were used for some phase of development or validation, but it doesn't separate training from testing data clearly in terms of counts.

9. How Ground Truth for Training Set Was Established

The document does not explicitly state how the ground truth for the training set was established. It describes ground truth methods for evaluation, but not specifically for the data used to train the AI model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Body Vision Ltd. % Paul Dryden Consultant ProMedic Consulting LLC 131 Bay Point Dr. NE St. Petersburg, Florida 33704

October 1, 2024

Re: K240943

Trade/Device Name: LungVision Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: August 30, 2024 Received: August 30, 2024

Dear Paul Dryden:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page

2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Re"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

{2}------------------------------------------------

the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K240943

Device Name

LungVision

Indications for Use (Describe)

The LungVision System is intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Official Contact:Benny Krauz – VP RA/QABody Vision Medical Ltd.7 Hamada St.Herzliya 4673341 IsraelTel: +972-54-208-098
Submission Correspondent:Paul DrydenProMedic, LLCSt. Petersburg, FL 33704
Proprietary or Trade Name:LungVision System
Common/Usual Name:System, image processing, radiological
Classification Name:Automated radiological image processing softwareQIH, CFR 892.2050Picture archiving and communications systemLLZ, CFR 892.2050
Predicate Device:BodyVision - LungVision - K183593Picture archiving and communications systemLLZ, CFR 892.2050

Description:

The Lung Vision System is designed to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.

The System is intended to assist the guidance of endobronchial tools to areas of interest inside a patient's lungs. The System allows the user to mark lesion locations and pathways to marked lesions using a patient's CT scan. During the endoscopic procedure, the System overlays planning information on real-time fluoroscopic images to guide endobronchial tool navigation. The System also provides tomographic images for lesion identification, as well as 3D views for understanding tool and lesion proximity and orientation. The System is designed to be integrated with fluoroscopic imaging systems and external displays.

The Lung Vision system includes a main unit and a tablet. Image processing algorithms are executed on the main unit and the tablet is used as a primary method of interacting with the system.

System Components Overview

The following is a list of the LungVision System's main components: LungVision Main Unit LungVision Tablet LungVision Router LungVision Board (passive device) LungVision Software

The submission includes claims of:

  • Transform 2D X-ray images into intraoperative tomographic scans ●
  • AI-driven intraoperative tomographic imaging

{5}------------------------------------------------

  • We find that this reflects the functionality of AI TOMO feature. The system receives CABT O output and transform them into tomographic imaging.
  • . CABT is a limited angle tomography based on the SIRT algorithm

Indications for Use:

The LungVision System is intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.

Contraindications:

None

Device Comparison

Table 1 compares the subject device to the predicate

FeatureProposed DeviceLungVisionPrimary PredicateK183593Body Vision LungVision
Indications for UseThe LungVision System is intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.The LungVision System is intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.
ClassificationAutomated radiological image processing softwareQIH, CFR 892.2050Picture archiving and communications system LLZ, CFR 892.2050Picture archiving and communications system LLZ, CFR 892.2050
Target anatomyLungsLungs
Anatomy accessBronchial airwaysBronchial airways
Windows OSWindows 10Windows 10
Medical imaging softwareYesYes
General Image 2D/3D reviewYesYes
3D rendering viewYesYes
Multi-modality SupportYesYes
Image registrationYesYes
Multi-planar reformatting (MPR)YesYes
DICOM importYesYes
Fluoroscopic videoYesYes
Standard Image viewing toolsYesYes
Segmentation toolYesYes
Video captureYesYes
Live Image overlaysYesYes
Import prior plansYesYes

{6}------------------------------------------------

510(k) Summarv

30-Aug-24
FeatureProposed DeviceLungVisionPrimary PredicateK183593Body Vision LungVision
Pointmarking/ taggingYesYes
Navigation typeVisualVisual
FeatureProposed DeviceLungVisionPrimary PredicateK183593Body Vision LungVision
Modifications
Virtual bronchoscopyYesYes
C-Arm based CTYesYes
Multi-view set-upYesYes
Real-timecompensationN/AYes
3D GuidanceYesYes

Substantial Equivalence Discussion

We will discuss the table above.

Indications for Use / Patient Population / Environment of Use: As in comparison of Indications For Use above, we can conclude that the indications for use for the LungVision and the predicate are substantially equivalent.

Discussion: The differences in proposed indications for use are minor. The minor differences do not raise new risk or safety concerns, and the subject device can be found substantially equivalent.

Prescriptive: Both the LungVision and predicate are prescription devices. Discussion: There are no differences.

Design and Technology: The LungVision utilizes similar technology as the predicate. This includes:

  • software applications that provide 2D and 3D medical image acquisition including real-time video . image acquisition and visualization of the anatomy
  • allow co-registration of real-time images to previously created 3D image sets based on previously collected DICOM CT images
  • include image enhancements such as contrast and brightness, zoom and pan capabilities ●
  • . addition of a Tablet to the PC but not to be used for diagnostic purposes

Discussion: The software modifications including virtual bronchoscopy, C-arm based Tomography, 3D Guidance, and Real-time compensation are similar to the reference device.

Performance and Specifications:

The performance and specifications demonstrate that the Lung Vision and predicate devices perform the same functions using the same technologies thus can be found substantially equivalent.

Discussion: There are no differences, thus the subject device can be found substantially equivalent.

Compliance with Standards:

LungVision includes hardware and complies with ANSI/AAMI/ES 60601-1:2005(2012) and IEC 60601-1-2:2014.

{7}------------------------------------------------

Discussion: The proposed device complies with the latest standards and thus can be found substantially equivalent.

Non-clinical / Bench Performance Testing:

We have performed bench tests and found that the Lung Vision met all requirements specifications and was found to be equivalent in comparison to the predicate. Testing includes verification testing of the requirements, testing of hazards mitigations and performance testing of the system.

Testing has also been performed on physical and simulated lung models representing deformable tissue.

For AI-Tomo the clinical validation was performed by physicians with data that was recorded from historical LungVision clinical cases. Cases were selected to include a representative range of lesion sizes, types, and locations, as well as a range of tools.

Simulated cases: synthetically simulated test cases, based on real human's CT scans. With this method we can simulate a lot of different use cases, and use diverse data. This method uses simulated data in a simulated environment, and therefore measures the registration error only, without additional real-world inaccuracies: CT to body divergence, jig estimation errors, fluoroscope distortion etc.

The mean accuracy calculated for 500 cases is 3.15 mm with 3.2 mm std.

Rigid body model cases: Laboratory test cases recordings and analysis in a patient simulated environment with a rigid lung model under real-time fluoroscopy.

This method provides a lower amount of test cases, as it uses only one lung anatomy with one ground truth transform. It is closer to a real-world scenario as it is influenced by jig estimation errors and fluoroscope distortion, but still does not include CT to body divergence.

The mean accuracy calculated for 93 cases is 3.64 mm with 1.57 mm std.

CBCT scans cases: test cases based on real human's CT scans and real human's CBCT scans taken in real procedures. This method is based on real cases data and therefore represents a real-world scenario and influenced by real-world inaccuracies: CT to body divergence, jig estimation errors, fluoroscope distortion etc.

The mean accuracy calculated for 191 cases is 5.34 mm with 3.32 mm std

The clinical validation of AI Tomo confirmed that lesion marking accuracy was comparable when using CABT and AI Tomo images. Therefore, regardless of the imaging modality used, the accuracy of the registration process remains consistent.

The images are CABTs collected from real cases performed by physicians using the LungVision system in the USA, Italy, and Israel. The CABTs are used as input data to the AI-Tomo post processing, and the output is evaluated using the ground truthing methodology.

Ground Truth methods included:

  • Geometrv test .
  • Lesion contrast on real data tests (with and without tool) .

Biocompatibility: There are no patient contacting parts of the LungVision System

{8}------------------------------------------------

Substantial Equivalence Conclusion

Based upon the foregoing performance testing and comparison to the legally marketed predicate devices for indications for use, technology, and performance we believe we have demonstrated that the LungVision System is substantially equivalent in safety and effectiveness to the predicate device.

N/A