K Number
K240902
Manufacturer
Date Cleared
2024-04-23

(21 days)

Product Code
Regulation Number
862.1355
Panel
SU
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dexcom G7 Continuous Glucose Monitoring System (Dexcom G7 CGM System or G7) is a real time, continuous glucose monitoring device indicated for the management of diabetes in persons 2 years and older.

The Dexcom G7 CGM System is intended to replace fingerstick BG testing for diabetes treatment decisions. Interpretation of the Dexcom G7 CGM System results should be based on the glucose trends and several sequential sensor readings over time. The Dexcom G7 CGM System also aids in the detection of episodes of hyperglycemia and hypodlycemia, facilitating both acute and long-term therapy adjustments.

The Dexcom G7 CGM System is also intended to autonomously communicate with digitally connected devices, including automated insulin dosing (AID) systems. The Dexcom G7 CGM System can be used alone or in conjunction with these digitally connected medical devices for the purpose of managing diabetes.

Device Description

The Dexcom G7 Continuous Glucose Monitoring System) is an interoperable continuous glucose monitoring (CGM) system intended to continuously measure the interstitial fluid, calculate the qucose reading and make this available to the user. The G7 System is intended for single patient use at home and requires a prescription.

The G7 System comprises the following primary components: a wearable, consisting of a G7 sensor and transmitter worn on the body and a display device, which can be a G7 Mobile App) on an iOS or Android OS smart device, or a G7 Receiver (Receiver).

To achieve the intended functions and performance of the G7 System, one sensor and at least one display device (App or Receiver) must be used together. The user must pair the display device(s) with each unique sensor to enable communication and start a sensor session. During an active session, the sensor reports new glucose data to the display device every 5-minutes. The displays glucose data and provides alerts and information signals to the user. The reportable glucose range for the G7 System is 40 mg/dL to 400 mg/dL. Glucose values below this range are reported as 'LOW' and glucose values above this range are reported as 'HIGH'. The sensor has an expected wear period of up to 10 days with an extended 12-hour grace period after the sensor session. The grace period allows additional time for the user to change the sensor at a convenient time.

The G7 is an interoperable connected device that can communicate glucose readings and other information wirelessly and securely to and from compatible electronic interfaces via the following secure wireless connections:

• Wireless communication from the transmitter directly to an interoperable device communicating through the same protocol

· The Mobile App communicates to another app on a single mobile platform

· The Mobile App communicates through the cloud to another software device

o Dexcom Partner Web APIs: The Dexcom Partner Web APIs enable secure and reliable communication of CGM data to authorized client software intended to receive the cloud. The Partner Web APIs is not intended to be used by automated insulin delivery systems (AID).

Principle of Operation:

The principles of operation for the Dexcom G7 System renain the same as prior generations of Dexcom CGM System uses a wire-type sensing mechanism that continuously measures interstitial glucose levels and uses a radio transmitter to wirelessly communicate glucose data to the display device for the user to see and use accordingly.

AI/ML Overview

The provided text is a 510(k) summary for the Dexcom G7 Continuous Glucose Monitoring System. It states that the device is substantially equivalent to a previously cleared Dexcom G7 Continuous Glucose Monitoring System (K234133). The key difference between the subject device and the predicate device is limited to "BLE communication and associated labeling updates." Crucially, it explicitly states: "No design change was required to support the extended BLE communication." and "No clinical data was necessary to determine substantial equivalence."

Therefore, the only acceptance criteria and study that proves the device meets the acceptance criteria mentioned in this document pertain to this specific change:


Acceptance Criteria and Study for Dexcom G7 Continuous Glucose Monitoring System (K240902)

Given that the submission is for an updated version of an already cleared device (Dexcom G7 K234133) and the only difference noted is related to BLE communication and associated labeling updates, with no design change, the acceptance criteria and study detailed here are specific to proving the extended BLE communication functionality.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
Communication RangeExtended BLE communication range specification.The subject device G7 CGM System met the extended BLE communication range specification.

2. Sample Size Used for the Test Set and Data Provenance:

The document states, "No design change was required to support the extended BLE communication." and "No clinical data was necessary to determine substantial equivalence." This implies that the testing for the BLE communication range was likely a non-clinical bench top or engineering test.

  • Sample Size: Not specified in the provided text, but it would pertain to the number of devices or test repetitions used in the communication range testing.
  • Data Provenance: Not specified, but generally, non-clinical engineering tests would be controlled laboratory environments. It is implicitly prospective testing conducted specifically for this regulatory submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • This information is not applicable for the specific testing described. The acceptance criteria for BLE communication range would be met through quantitative measurements (e.g., signal strength, data transmission success rate over distance) against a pre-defined engineering specification, not through human expert assessment or ground truth establishment in the traditional sense for medical device diagnostics.

4. Adjudication Method for the Test Set:

  • This information is not applicable for the specific testing described. Adjudication methods are typically used in clinical studies or image review settings involving human interpretation. The BLE communication range testing would rely on automated data collection and comparison to a specification.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and what was the effect size of how much human readers improve with AI vs without AI assistance:

  • This information is not applicable. This submission is for a continuous glucose monitoring system, not an AI-assisted diagnostic imaging device. The document explicitly states, "No clinical data was necessary to determine substantial equivalence."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • The primary "performance" discussed for this specific 510(k) is the extended BLE communication range. This would be assessed as a "standalone" technical capability of the device's communication module, without direct human intervention in the communication process itself. However, it's not an "algorithm only" performance assessment in the sense of AI diagnostics.

7. The Type of Ground Truth Used:

  • For the BLE communication range, the "ground truth" would be the engineering specification for the extended communication range itself, against which the measured performance of the device's BLE module is compared. It's a quantitative, pre-defined target.

8. The Sample Size for the Training Set:

  • This information is not applicable. There is no mention of a "training set" as this is not an AI/machine learning model where a distinct training phase is described.

9. How the Ground Truth for the Training Set Was Established:

  • This information is not applicable for the same reason as above.

§ 862.1355 Integrated continuous glucose monitoring system.

(a)
Identification. An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Robust clinical data demonstrating the accuracy of the device in the intended use population.
(ii) The clinical data must include a comparison between iCGM values and blood glucose values in specimens collected in parallel that are measured on an FDA-accepted laboratory-based glucose measurement method that is precise and accurate, and that is traceable to a higher order (
e.g., an internationally recognized reference material and/or method).(iii) The clinical data must be obtained from a clinical study designed to fully represent the performance of the device throughout the intended use population and throughout the measuring range of the device.
(iv) Clinical study results must demonstrate consistent analytical and clinical performance throughout the sensor wear period.
(v) Clinical study results in the adult population must meet the following performance requirements:
(A) For all iCGM measurements less than 70 milligrams/deciliter (mg/dL), the percentage of iCGM measurements within ±15 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 85 percent.
(B) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 70 percent.
(C) For all iCGM measurements greater than 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 80 percent.
(D) For all iCGM measurements less than 70 mg/dL, the percentage of iCGM measurements within ±40 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 98 percent.
(E) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(F) For all iCGM measurements greater than180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(G) Throughout the device measuring range, the percentage of iCGM measurements within ±20 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 87 percent.
(H) When iCGM values are less than 70 mg/dL, no corresponding blood glucose value shall read above 180 mg/dL.
(I) When iCGM values are greater than 180 mg/dL, no corresponding blood glucose value shall read less than 70 mg/dL.
(J) There shall be no more than 1 percent of iCGM measurements that indicate a positive glucose rate of change greater than 1 mg/dL per minute (/min) when the corresponding true negative glucose rate of change is less than −2 mg/dL/min as determined by the corresponding blood glucose measurements.
(K) There shall be no more than 1 percent of iCGM measurements that indicate a negative glucose rate of change less than −1 mg/dL/min when the corresponding true positive glucose rate of change is greater than 2 mg/dL/min as determined by the corresponding blood glucose measurements.
(vi) Data demonstrating similar accuracy and rate of change performance of the iCGM in the pediatric population as compared to that in the adult population, or alternatively a clinical and/or technical justification for why pediatric data are not needed, must be provided and determined by FDA to be acceptable and appropriate.
(vii) Data must demonstrate that throughout the claimed sensor life, the device does not allow clinically significant gaps in sensor data availability that would prevent any digitally connected devices from achieving their intended use.
(2) Design verification and validation must include a detailed strategy to ensure secure and reliable means of iCGM data transmission to provide real-time glucose readings at clinically meaningful time intervals to devices intended to receive the iCGM glucose data.
(3) Design verification and validation must include adequate controls established during manufacturing and at product release to ensure the released product meets the performance specifications as defined in paragraphs (b)(1) and (b)(2) of this section.
(4) The device must demonstrate clinically acceptable performance in the presence of clinically relevant levels of potential interfering substances that are reasonably present in the intended use population, including but not limited to endogenous substances and metabolites, foods, dietary supplements, and medications.
(5) The device must include appropriate measures to ensure that disposable sensors cannot be used beyond its claimed sensor wear period.
(6) Design verification and validation must include results obtained through a usability study that demonstrates that the intended user can use the device safely and obtain the expected glucose measurement accuracy.
(7) The labeling required under § 809.10(b) of this chapter must include a separate description of the following sensor performance data observed in the clinical study performed in conformance with paragraph (b)(1) of this section for each intended use population, in addition to separate sensor performance data for each different iCGM insertion or use sites (
e.g., abdomen, arm, buttock):(i) A description of the accuracy in the following blood glucose concentration ranges: less than 54 mg/dL, 54 mg/dL to less than 70 mg/dL, 70 to 180 mg/dL, greater than 180 to 250 mg/dL, and greater than 250 mg/dL.
(ii) A description of the accuracy of positive and negative rate of change data.
(iii) A description of the frequency and duration of gaps in sensor data.
(iv) A description of the true, false, missed, and correct alert rates and a description of the available glucose concentration alert settings, if applicable.
(v) A description of the observed duration of iCGM life for the device.