K Number
K240744

Validate with FDA (Live)

Device Name
uMR 680
Date Cleared
2024-04-10

(23 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The uMR 680 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and spectroscopic images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan.

Device Description

The uMR 680 is a 1.5T superconducting magnetic resonance diagnostic device with a 70cm size patient bore. It consists of components such as magnet, RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, computer, equipment cabinets, power distribution system, internal communication system, and vital signal module etc. The uMR 680 Magnetic Resonance Diagnostic Device is designed to conform to NEMA and DICOM standards. uMR 680 has been previously cleared by FDA via K222755. The modification performed on uMR 680 (K222755) in this submission is due to the addition of - Breast Coil -24 - epi_se_mre - MRE (Magnetic Resonance Elastography) The modification does not affect the intended use or alter the fundamental scientific technology of the device.

AI/ML Overview

The provided text describes the acceptance criteria and the study results for the Shanghai United Imaging Healthcare Co., Ltd. uMR 680 Magnetic Resonance Diagnostic Device.

Here's the breakdown of the information requested:

1. Table of Acceptance Criteria and Reported Device Performance

ItemAcceptance CriteriaReported Device Performance
Breast Coil - 24
Surface heatingThe maximum temperature of all temperature probes shall not exceed 41 °C.Pass
General electrical/mechanical safetyConform with ANSI/AAMI ES60601-1Pass
SNR and UniformitySNR and Uniformity shall fulfill with the design specification.Pass
BiocompatibilityMaterials of construction and manufacturing materials exempt from testing according to the Biocompatibility guidance (Attachment G), the 510(k) numbers for devices where these materials have been previously approved, or full biocompatibility report (assessment of sensitization, irritation and cytotoxicity risks) for components that have direct contact with the patient.All the materials of patient-contacting components for the Breast Coil - 24 are identical to uMR Omega which was cleared in K230152 in formulation, processing, sterilization, and geometry, and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents).
EMC-immunity, electrostatic dischargeConform with IEC 60601-1-2 and IEC 60601-4-2Pass
Clinical image qualityImage quality is sufficient for diagnostic use.The U.S. Board Certified radiologist approves that image quality is sufficient for diagnostic use.
MRE/epi_se_mre
General electrical/mechanical safetyConform with ANSI/AAMI ES60601-1Pass
EMCConform with IEC 60601-1-2 and IEC 60601-4-2Pass
PerformanceBias of accuracy, repeatability, reproducibility and parameter sensitivity shall fulfill with the design specification.Pass
Clinical image qualityImage quality is sufficient for diagnostic use.The U.S. Board Certified radiologist approves that image quality is sufficient for diagnostic use.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not explicitly stated for each test beyond "phantom and volunteer test" for MRE performance and "the image generated by Breast Coil-24" for clinical image quality.
  • Data Provenance: Not specified. It does not mention country of origin or if the data was retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Number of Experts: For clinical image quality for both Breast Coil - 24 and MRE, it states "The U.S. Board Certified radiologist approves." This implies at least one, but the exact number of radiologists is not specified.
  • Qualifications of Experts: "U.S. Board Certified radiologist."

4. Adjudication method for the test set:

  • Adjudication Method: Not explicitly stated. The statement "The U.S. Board Certified radiologist approves" suggests a single expert's approval for clinical image quality assessment, rather than a multi-expert consensus method like 2+1 or 3+1.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned as being done for this submission. The studies detailed focus on system performance parameters and clinical image quality approval by a radiologist.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • This device is a Magnetic Resonance Diagnostic Device (MRDD), which is a hardware system that produces images and physical parameters to be interpreted by a trained physician. The performance tests described (surface heating, electrical/mechanical safety, SNR, uniformity, biocompatibility, EMC, performance, and clinical image quality) are for the device itself and its components/functionalities, not for an AI algorithm working in a standalone capacity. So, this question is not directly applicable in the terms of an AI algorithm, but the system performance was evaluated standalone from human interpretation in most of the tests.

7. The type of ground truth used:

  • Clinical Image Quality: Expert consensus (or approval by a U.S. Board Certified radiologist) that the image quality is sufficient for diagnostic use.
  • Biocompatibility: Demonstrated by using materials identical to a previously cleared device (uMR Omega, K230152) and meeting regulatory guidance.
  • Other performance metrics (Surface heating, electrical/mechanical safety, SNR, Uniformity, EMC, MRE Performance): Based on compliance with established engineering and safety standards (NEMA MS 14, ANSI/AAMI ES60601-1, NEMA MS 1, NEMA MS 3, NEMA MS 6, NEMA MS 9, IEC 60601-1-2, IEC 60601-4-2) and design specifications.

8. The sample size for the training set:

  • The document does not detail any "training set" as it is describing a hardware medical device with specific new coils and pulse sequences rather than a machine learning or AI-driven diagnostic algorithm that would typically require a training set.

9. How the ground truth for the training set was established:

  • Not applicable, as no training set for an AI/ML algorithm is mentioned.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in large, bold, blue letters, and the word "ADMINISTRATION" in smaller, blue letters below.

April 10, 2024

Shanghai United Imaging Healthcare Co., Ltd. % Xin Gao RA Manager No.2258 Chengbei Rd. Jiading District Shanghai, 201807 China

Re: K240744 Trade/Device Name: uMR 680 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH, MOS Dated: March 18, 2024 Received: March 18, 2024

Dear Xin Gao:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)K240744
Device NameuMR 680
Indications for Use (Describe)The uMR 680 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and spectroscopic images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained

physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image contains the logo for United Imaging. The words "UNITED" and "IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The logo is simple and modern.

510 (k) SUMMARY

1. Date of Prepared

March 18, 2024

Sponsor Identification 2.

Shanghai United Imaging Healthcare Co., Ltd. No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China

Contact Person: Xin GAO Position: Regulatory Affairs Specialist Tel: +86-021-67076888-5386 Fax: +86-021-67076889 Email: xin.gao@united-imaging.com

Identification of Proposed Device(s) 3.

Trade Name: uMR 680 Common Name: Magnetic Resonance Diagnostic Device Model: uMR 680

Regulatory Information Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH, MOS Review Panel: Radiology

4. Identification of Predicate Devices(s)

Predicate Device 510(k) Number: K222755 Model: uMR 680 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Review Panel: Radiology

Reference Device 510(k) Number: K230152 Model: uMR Omega

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for United Imaging. The words "UNITED" and "IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the words is a stylized "U" shape, which is also in a bold font. The logo is simple and modern, and the colors are muted.

Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Review Panel: Radiology

5. Device Description

The uMR 680 is a 1.5T superconducting magnetic resonance diagnostic device with a 70cm size patient bore. It consists of components such as magnet, RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, computer, equipment cabinets, power distribution system, internal communication system, and vital signal module etc. The uMR 680 Magnetic Resonance Diagnostic Device is designed to conform to NEMA and DICOM standards.

uMR 680 has been previously cleared by FDA via K222755. The modification performed on uMR 680 (K222755) in this submission is due to the addition of

  • Breast Coil -24
  • epi_se_mre
  • MRE (Magnetic Resonance Elastography)

The modification does not affect the intended use or alter the fundamental scientific technology of the device.

6. Indications for Use

The uMR 680 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and spectroscopic images, and that display internal anatomical structure and/or function of the head, body and extremities.

These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan.

7. Comparison of Technological Characteristics with the Predicate Device

The differences from the predicate device are discussed in the comparison table in this submission as below.

ITEMProposed DeviceuMR 680Predicate DeviceuMR 680 (K222755)Remark
IndicationsFor UseThe uMR 680 system is indicatedfor use as a magnetic resonanceThe uMR 680 system is indicatedfor use as a magnetic resonanceSame

Table 1 Comparison to Predicate Device

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized "U" symbol, which is dark gray. The logo is simple and modern in design.

ITEMProposed DeviceuMR 680Predicate DeviceuMR 680 (K222755)Remark
diagnostic device (MRDD) thatproduces sagittal, transverse,coronal, and oblique crosssectional images, andspectroscopic images, and thatdisplay internal anatomicalstructure and/or function of thehead, body and extremities.These images and the physicalparameters derived from theimages when interpreted by atrained physician yieldinformation that may assist thediagnosis. Contrast agents may beused depending on the region ofinterest of the scan.diagnostic device (MRDD) thatproduces sagittal, transverse,coronal, and oblique cross sectionalimages, and spectroscopic images,and that display internal anatomicalstructure and/or function of thehead, body and extremities.These images and the physicalparameters derived from the imageswhen interpreted by a trainedphysician yield information thatmay assist the diagnosis. Contrastagents may be used depending onthe region of interest of the scan.
BreastCoil - 24YesNoThe intended use ofBreast coil -24 isequivalent to previouslycleared Breast Coil-10.The difference betweenthem is the number ofchannels of the receivercoil.The difference did notraise new safety andeffectiveness concerns.

Table 2 Comparison to Reference Device

ITEMProposed DeviceuMR 680Predicate DeviceuMR Omega (K230152)Remark
MREYesYesSame

Based on the comparison and analysis above, the proposed device has same intended use, similar performance, equivalence safety and effeteness in hardware and software as the predicate device. The differences above between the proposed device and predicate device do not affect the intended use, technology characteristics, safety and effectiveness. And no issues are raised regarding to safety and effectiveness.

Performance Data 8.

The following testing was conducted on uMR 680 and were provided in support of the substantial determination.

ItemVerification/Validation Method(s)Acceptance CriteriaSummary of Results
Breast Coil - 24
Surface heatingPerform test according to NEMA MS 14The maximum temperature of all temperature probes shall not exceed 41 °C.Pass
General electrical/mechanical safetyPerform test according to ANSI/AAMI ES60601-1Conform with ANSI/AAMI ES60601-1Pass
SNR and UniformityPerform test according to NEMA MS 1, NEMA MS 3, NEMA MS 6 and NEMA MS 9SNR and Uniformity shall fulfill with the design specification.Pass
BiocompatibilityBiocompatibility evaluation in agreement with recommendations in Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"Materials of construction and manufacturing materials exempt from testing according to the Biocompatibility guidance (Attachment G), the 510(k) numbers for devices where these materials have been previously approved, or full biocompatibility report (assessment of sensitization, irritation and cytotoxicity risks) for components that have direct contact with the patient.All the materials of patient- contacting components for the Breast Coil - 24 are identical to uMR Omega which was cleared in K230152 in formulation, processing, sterilization, and geometry, and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents).
EMC-immunity, electrostatic dischargePerform test according to IEC 60601-1-2 and IEC 60601-4-2Conform with IEC 60601-1-2 and IEC 60601-4-2Pass
Clinical image qualityEvaluate the image generated by Breast Coil-24 with the same method as the predicate deviceImage quality is sufficient for diagnostic use.The U.S. Board Certified radiologist approves that image quality is sufficient for diagnostic use.
MRE/epi_se_mre
General electrical/mechanical safetyPerform test according to ANSI/AAMI ES60601-1Conform with ANSI/AAMI ES60601-1Pass
EMCPerform test according to IEC 60601-1-2 and IEC 60601-4-2Conform with IEC 60601-1-2 and IEC 60601-4-2Pass
PerformancePerform a phantom and volunteer test under the proposed device with MREBias of accuracy, repeatability, reproducibility and parameter sensitivity shall fulfill with the design specification.Pass
ClinicalqualityimageEvaluate the image generated by MREImage quality is sufficientfor diagnostic use.The U.S. Board Certifiedradiologist approves thatimage quality is sufficientfor diagnostic use.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the word "UNITED" stacked on top of the word "IMAGING", both in a bold, sans-serif font. To the right of the text is a stylized "U" symbol, which is dark teal in color and has a vertical white line running through the center.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the word "UNITED" stacked on top of the word "IMAGING". To the right of the words is a stylized letter "U" that is dark gray. The logo is simple and modern.

The test results demonstrated that the device performed as expected and thus, it is substantial equivalent to the predicate devices to which it has been compared.

Conclusion 9.

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, we conclude that uMR 680 Magnetic Resonance Diagnostic Device is substantially equivalent to the predicate device. It does not introduce new indications for use, and performance test results support that the technical characteristic change of the Breast Coil - 24 (i.e., the number of channels in the coil) and incorporating MRE/epi_se_mre do not introduce new potential hazards or safety risks.

N/A