K Number
K240737

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-04-05

(18 days)

Product Code
Regulation Number
888.3070
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The G Surgical Marksman MIS System is intended to provide immobilization of the posterior noncervical spine (T1-S2/llium) in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis and/or lordosis), tumor, pseudarthrosis, and/or failed previous fusion. The G Surgical Marksman MIS System can be used in an open approach or a percutaneous approach with MIS instrumentation.

Device Description

The G Surgical Marksman MIS System consists of longitudinal members (rods), anchors (screws) and fasteners in a variety of sizes to accommodate differing anatomic requirements. The devices are sold nonsterile.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the G Surgical Marksman MIS System. It details the device's indications for use, materials, and references mechanical testing. However, it does not provide specific acceptance criteria or the full study details in the format requested for an AI/software device.

The G Surgical Marksman MIS System is a medical device (a thoracolumbosacral pedicle screw system), not an AI/software device. The performance data mentioned refers to mechanical testing (ASTM F1717) to demonstrate structural integrity and durability, which is standard for orthopedic implants. It does not involve AI performance metrics like accuracy, sensitivity, or specificity, nor does it involve human readers or ground truth established by experts in the context of diagnostic or interpretive tasks.

Therefore, most of the requested information regarding AI/software device performance, such as sample sizes for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance, and ground truth types for AI, is not applicable to this submission.

Here is the information that can be extracted or derived from the provided text, acknowledging the nature of the device:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Type of Test)Reported Device Performance (Summary)
ASTM F1717 Mechanical TestingStatic and dynamic compression bending, and static torsion tests were performed on worst-case G Surgical Marksman MIS System constructs. (No specific numerical acceptance values or pass/fail results are provided in this summary, but the implication of clearance is that these tests met predefined standards.)

2. Sample size used for the test set and the data provenance:

  • Sample size: Not explicitly stated, as this refers to mechanical test constructs, not patient data or images for AI. The text mentions "worst-case G Surgical Marksman MIS System constructs."
  • Data provenance: Not applicable in the context of patient data for AI. The "data" refers to the results of mechanical laboratory tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. Ground truth in this context would refer to material science and mechanical engineering standards, not expert medical opinion for diagnostic accuracy.

4. Adjudication method for the test set:

  • Not applicable. Adjudication is relevant for expert disagreement in AI ground truth establishment. Mechanical testing follows standardized protocols.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This is not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is not an AI algorithm.

7. The type of ground truth used:

  • Ground truth: Adherence to ASTM F1717 standards for mechanical safety and performance, as assessed through laboratory testing.

8. The sample size for the training set:

  • Not applicable. This is not an AI device that requires a training set.

9. How the ground truth for the training set was established:

  • Not applicable.

In summary: The provided document is a 510(k) clearance letter for a physical medical device (spinal implant components), not for an AI/software device. The performance data refers to mechanical testing for physical properties, not diagnostic accuracy or human-AI interaction. Therefore, many of the requested details are not relevant to this specific submission.

{0}------------------------------------------------

April 5, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.

G Surgical LLC % Karen Warden President BackRoads Consulting Inc. PO Box 566 Chesterland. Ohio 44026

Re: K240737

Trade/Device Name: G Surgical Marksman MIS System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB Dated: March 18, 2024 Received: March 18, 2024

Dear Karen Warden:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'neill -S

Image /page/1/Picture/7 description: The image shows the text 'n O'neill -S' on the left side of the image. On the right side of the image is the FDA logo. The FDA logo is in a light blue color. The text is in black.

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K240737

Device Name G Surgical Marksman MIS System

Indications for Use (Describe)

The G Surgical Marksman MIS System is intended to provide immobilization of the posterior noncervical spine (T1-S2/llium) in skeletally mature patients as an adjunct to fusion in the following acute and chronic instabilities or deformities: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis and/or lordosis), tumor, pseudarthrosis, and/or failed previous fusion. The G Surgical Marksman MIS System can be used in an open approach or a percutaneous approach with MIS instrumentation.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Date:18 March 2024
Sponsor:G Surgical LLC
9433 Bee Cave Road
Building 3, Suite 101-A
Austin, Texas 78733 USA
Tel.: 512.494.4749
Sponsor Contact:Don Grafton, Managing Director
510(k) Contact:Karen E. Warden, PhD
BackRoads Consulting Inc.
PO Box 566
Chesterland, OH 44026
Office: 440.729.8457
Proposed Trade Name:G Surgical Marksman MIS System
Common Name:Posterior pedicle screw system
Device Classification:Class II
Regulation Name,Regulation Number,Product Code:Thoracolumbosacral pedicle screw system, 888.3070, NKB
Submission purpose:The subject 510(k) adds long arm polyaxial screws and a larger diameterscrew to the cleared components.
Device Description:The G Surgical Marksman MIS System consists of longitudinal members(rods), anchors (screws) and fasteners in a variety of sizes to accommodatediffering anatomic requirements. The devices are sold nonsterile.
Indications for Use:The G Surgical Marksman MIS System is intended to provide immobilizationand stabilization of the posterior non-cervical spine (T1-S2/Ilium) inskeletally mature patients as an adjunct to fusion in the treatment of thefollowing acute and chronic instabilities or deformities: degenerative discdisease (defined as back pain of discogenic origin with degeneration of thedisc confirmed by history and radiographic studies), spondylolisthesis,trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e.,scoliosis, kyphosis and/or lordosis), tumor, pseudarthrosis, and/or failedprevious fusion. The G Surgical Marksman MIS System can be used in anopen approach or a percutaneous approach with MIS instrumentation.
Materials:G Surgical Marksman MIS System is manufactured from Ti-6Al-4V ELItitanium alloy (ASTM F136) and Cobalt Chrome (per ASTM F1537).
Primary Predicate:G Surgical Marksman MIS System (G Surgical LLC - K161516)
Performance Data:ASTM F1717 mechanical testing of worst case G Surgical Marksman MISSystem constructs was relied upon in support of this G Surgical MarksmanMIS clearance. The testing included static and dynamic compressionbending and static torsion.
TechnologicalCharacteristics:The G Surgical Marksman MIS System Is identical to itself as the primarypredicate with respect to the following technological characteristics: basicdesign, material and dimensional features.
Conclusion:The G Surgical Marksman MIS System possesses the same intended useand technological characteristics as the predicate devices. Therefore GSurgical Marksman MIS System is substantially equivalent for its intendeduse.

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.