K Number
K240653
Manufacturer
Date Cleared
2024-10-31

(238 days)

Product Code
Regulation Number
870.1025
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SmartCardia 7L Platform is a prescribed device used for the purpose of identifying non-lethal arrhythmias. The SmartCardia 7L Platform is intended for continuous external electrocardiogram (ECG) information to support Holter Monitoring, Extended Holter Monitoring, and Outpatient Cardiac Telemetry (OCT) commonly called Mobile Cardiac Telemetry (MCT) monitoring.

The SmartCardia 7L Platform is intended for:

  1. Patients who experience transient symptoms that may suggest cardiac arrhythmia.

  2. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation).

  3. Patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c) dyspnea (shortness of breath).

  4. Patients recovering from cardiac surgery or interventional procedures who are indicated for outpatient arrhythmia monitoring.

  5. ECG data recorded by the device can be analyzed by other processing systems to provide Holter style reports Measurements include: electrocardiogram (ECG signal), R-R interval, Heart Rate. Notification alerts can be set for one or more of these measures.

The SmartCardia 7L Platform is indicated for use on patients who are 18 years of age or older to provide monitoring of physiological information. It is intended for use in a physician office, outpatient facility, or in the patient's home.

SmartCardia 7L Platform contraindications:

  1. The SmartCardia 7L Platform is contraindicated for use for detection of hemodynamically unstable or life-threatening arrhythmias or cardiac events requiring urgent medical response. It is not intended for monitoring patients during cardiac rehabilitation outside of healthcare facilities.

  2. The SmartCardia 7L Platform is contraindicated for use during external defibrillation.

Device Description

The SmartCardia 7L Platform MCT is a body worn monitoring product that is designed with a disposable adhesive 7L Patch and reusable 7L Sensor. It is designed to be worn by the patient for up to 14 days. If longer monitoring is necessary, the 7L Patch is removed from the body, the 7L Sensor is removed from the worn 7L Patch which is discarded. The 7L Sensor is inserted into a new 7L Patch and the new assembly is placed on the patient for monitoring to continue.

The 7L Sensor/7L Patch assembly communicates to the SmartCardia Phone via Bluetooth technology and shows the patient's heart rate and ECG on its display and allows the patient to input symptoms (Mark Event) which are shown in the patient record. The SmartCardia Phone has a medical grade mains powered charger and uses its cellular technology to act as a gateway to the SmartCardia Cloud Service provided by Amazon Web Services. The SmartCardia Phone is pre-configured by SmartCardia and placed in a kiosk mode. Data stored in the SmartCardia cloud can be viewed in the SmartCardia Web Browser application by a clinician or healthcare provider. The SmartCardia 7L Platform incorporates three modes of monitoring:

  1. Holter Monitoring (up to 48 hours) and Extended Holter Monitoring (>48 hours and up to 14 days),
  2. Event Monitoring (up to 48 hours, and >48 hours up to 14 days)
  3. Cardiac Outpatient Telemetry (OCT) commonly called Mobile Cardiac Telemetry (MCT) (>48 hours up to 30 days when changing the 7L Patch)

During any of the selected modes, a clinician or healthcare provider can use the SmartCardia Web Browser to view the continuously streaming patient's ECG.
The SmartCardia 7L Platform provides alarm notifications for heart rate and atrial fibrillation. As stated in the Intended Use statement, the system is contraindicated for use in a critical care setting where an immediate response to life threatening conditions such as lethal arrhythmias is required. There are no alarm signals presented to the clinician for conditions other than heart rate and atrial fibrillation. The SmartCardia 7L Platform performs retrospective analysis and identifies events which it shows the clinician when they are reviewing historical data. The clinician must then review these events and determine if they are indeed valid and should be included in a physician's report. These events are things like "Ventricular Bigeminy', 'Supraventricular Couplet', etc. A full list is included in the Clinician Instructions for Use.

AI/ML Overview

The provided text describes the SmartCardia 7L Platform, an ECG monitoring device, and its evaluation for FDA 510(k) clearance. However, the document does not contain specific details regarding acceptance criteria for the algorithm's performance, nor does it provide a full study report with detailed performance metrics (like sensitivity, specificity for various arrhythmias), sample sizes, ground truth establishment methods, or expert qualifications for the algorithm testing.

The document primarily focuses on:

  • General device description and intended use.
  • Comparison to predicate devices for substantial equivalence.
  • Non-clinical bench performance testing for electrical safety, EMC, wireless coexistence, and general ECG monitoring standards (IEC 60601-1, IEC 60601-2, IEC 60601-2-47, ANSI/AAMI EC12).
  • Mention of testing the ECG detection algorithm per ANSI/AAMI EC57, but explicitly states it does not claim ST analysis and developed its own protocol and database for evaluating the algorithm.

Therefore, many of the requested details about the algorithm's performance study and acceptance criteria are NOT present in the provided text.

Here's a summary of what can be extracted and what is explicitly missing:

Acceptance Criteria and Reported Device Performance

The document states: "The ECG detection algorithm was tested per ANSI/AAMI EC57 standard for testing and reporting performance results of cardiac rhythm and ST-segment measurement algorithms. The SmartCardia 7L Platform does not claim ST analysis or measurements. Therefore, the sections of the EC57 standard relating to ST measurement is not applicable. Because of the nature of the SmartCardia 7L Platform as a body worn sensor and patch device with the electrodes in close proximity to each other, and not utilizing ECG lead wires such as a traditional ECG device, SmartCardia developed a protocol and database for evaluating the analysis algorithm. The results show that the SmartCardia 7L Platform algorithm detection and reported results for the stated output arrhythmia were satisfactory and met industry norms."

This statement vaguely indicates that the algorithm's performance met "industry norms" and was "satisfactory." However, a specific table of acceptance criteria with numerical targets and the device's reported performance against those targets is NOT provided in this document.

Acceptance Criteria (Example - Not explicitly stated in document)Reported Device Performance (Example - Not explicitly stated in document)
Arrhythmia Detection (General):"satisfactory and met industry norms" (Vague statement, no specific metrics)
e.g., AFib Sensitivity > 90%Not provided
e.g., AFib Specificity > 95%Not provided
e.g., Ventricular Bigeminy detection accuracy"satisfactory and met industry norms" (Vague statement, no specific metrics)
e.g., Supraventricular Couplet detection accuracy"satisfactory and met industry norms" (Vague statement, no specific metrics)

Details of the Study Proving Device Meets Criteria

  1. Sample size used for the test set and the data provenance:

    • Test set sample size: NOT provided. The document mentions that SmartCardia "developed a protocol and database for evaluating the analysis algorithm" but does not give the size or composition of this test set.
    • Data provenance: NOT provided. The country of origin and whether the data was retrospective or prospective are not mentioned.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • NOT provided. The document does not mention the number or qualifications of experts involved in establishing ground truth for any test set for the algorithm. It only notes that the clinician "must then review these events and determine if they are indeed valid and should be included in a physician's report." This refers to clinical use, not the validation study methodology.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • NOT provided. No information on an adjudication method for establishing ground truth is given.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    • NO. There is no mention of an MRMC study or any comparison of human readers with vs. without AI assistance. The device description mentions a "clinician or qualified cardiac technician" can view continuous ECG streams and the device performs "retrospective analysis" to identify events which the clinician "must then review... and determine if they are indeed valid." This implies the AI is assisting but a formal comparative effectiveness study like an MRMC is not described.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • YES, implied. The statement "The ECG detection algorithm was tested per ANSI/AAMI EC57 standard... The results show that the SmartCardia 7L Platform algorithm detection and reported results for the stated output arrhythmia were satisfactory and met industry norms" indicates that the algorithm itself was evaluated in a standalone manner against a "protocol and database." However, specific performance metrics are not given.
  6. The type of ground truth used:

    • Implied Expert Consensus/Annotation for algorithm training/testing. The text mentions "SmartCardia developed a protocol and database for evaluating the analysis algorithm" and that the algorithm identifies "events" that clinicians later review. While not explicitly stated for the validation test set, it implies that the "ground truth" for the algorithm's performance evaluation would stem from expert annotation of ECG data. It is not pathology or direct outcomes data.
  7. The sample size for the training set:

    • NOT provided. The document does not specify the size of the training set used for the algorithm. It only refers to a "protocol and database."
  8. How the ground truth for the training set was established:

    • NOT provided explicitly. It is implied that for any "database" used, ground truth would be established through expert interpretation or annotation of ECG recordings, but the exact methodology (e.g., number of experts, adjudication) is not detailed for the training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 31, 2024

SmartCardia SA % Robert Steurer Consultant Steurer Consulting Group LLC 800 Blue Quail Rd Keller, Texas 76248

Re: K240653

Trade/Device Name: SmartCardia 7L Platform Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: QYX, DSI, DRG Dated: October 1, 2024 Received: October 2, 2024

Dear Robert Steurer:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Jennifer W. Shih -S

Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240653

Device Name SmartCardia 7L Platform

Indications for Use (Describe)

The SmartCardia 7L Platform is a prescribed device used for the purpose of identifying non-lethal arrhythmias. The SmartCardia 7L Platform is intended for continuous external electrocardiogram (ECG) information to support Holter Monitoring, Extended Holter Monitoring, and Outpatient Cardiac Telemetry (OCT) commonly called Mobile Cardiac Telemetry (MCT) monitoring.

The SmartCardia 7L Platform is intended for:

  1. Patients who experience transient symptoms that may suggest cardiac arrhythmia.

  2. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation).

  3. Patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c)

dyspnea (shortness of breath).

  1. Patients recovering from cardiac surgery or interventional procedures who are indicated for outpatient arrhythmia monitoring.

  2. ECG data recorded by the device can be analyzed by other processing systems to provide Holter style reports Measurements include: electrocardiogram (ECG signal), R-R interval, Heart Rate. Notification alerts can be set for one or more of these measures.

The SmartCardia 7L Platform is indicated for use on patients who are 18 years of age or older to provide monitoring of physiological information. It is intended for use in a physician office, outpatient facility, or in the patient's home.

SmartCardia 7L Platform contraindications:

  1. The SmartCardia 7L Platform is contraindicated for use for detection of hemodynamically unstable or life-threatening arrhythmias or cardiac events requiring urgent medical response. It is not intended for monitoring patients during cardiac rehabilitation outside of healthcare facilities.

  2. The SmartCardia 7L Platform is contraindicated for use during external defibrillation.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

Date Prepared:October 28, 2024
Submitter:SmartCardia SAEPFL Innovation ParkBuilding C1015, Lausanne, Switzerland
Srinivasan MuraliSmartCardia SAPhone: +41 7887 50864Fax: N/AEmail: srinivasan.murali@smartcardia.com
Application Correspondent:Robert Steurer
Principal ConsultantSteurer Consulting Group800 Blue Quail Rd.Keller, TX 76248steurerbob@gmail.com+1 (425) 358-1072
Manufacturing Site:SmartCardia SAEPFL Innovation ParkBuilding C1015, Lausanne, Switzerland
Common Name:SmartCardia 7L Platform
Classification Name:Arrhythmia detector and alarm (including ST-segment measurementand alarm)
Device Class:Class II
Primary ClassificationRegulation:21 CFR 870.1025
Primary Product Code:QYX
Secondary Codes:DSI,DRG
Primary Predicate Device:InfoBionics MoMe Kardia, K160004
Reference Predicate Device:SmartCardia 7L Platform, K231276
Device Description:
The SmartCardia 7L Platform MCT is a body worn monitoring product that is designed with a disposable adhesive 7L Patch and reusable 7L Sensor. It is designed to be worn by the patient for up to 14 days. If longer monitoring is necessary, the 7L Patch is removed from the body, the 7L Sensor is removed from the worn 7L Patch which is discarded. The 7L Sensor is inserted into a new 7L Patch and the new assembly is placed on the patient for monitoring to continue.The 7L Sensor/7L Patch assembly communicates to the SmartCardia Phone via Bluetooth technology and shows the patient's heart rate and ECG on its display and allows the patient to input symptoms (Mark Event) which are shown in the patient record. The SmartCardia Phone has a medical grade mains powered charger and uses its cellular technology to act as a gateway to the SmartCardia Cloud Service provided by Amazon Web Services. The SmartCardia Phone is pre-configured by SmartCardia and placed in a kiosk mode. Data stored in the SmartCardia cloud can be viewed in the SmartCardia Web Browser application by a clinician or healthcare provider. The SmartCardia 7L Platform incorporates three modes of monitoring:1. Holter Monitoring (up to 48 hours) and Extended Holter Monitoring (>48 hours and up to 14 days),2. Event Monitoring (up to 48 hours, and >48 hours up to 14 days)3. Cardiac Outpatient Telemetry (OCT) commonly called Mobile Cardiac Telemetry (MCT) (>48 hours up to 30 days when changing the 7L Patch)During any of the selected modes, a clinician or healthcare provider can use the SmartCardia Web Browser to view the continuously streaming patient's ECG.The SmartCardia 7L Platform provides alarm notifications for heart rate and atrial fibrillation. As stated in the Intended Use statement, the system is contraindicated for use in a critical care setting where an immediate response to life threatening conditions such as lethal arrhythmias is required. There are no alarm signals presented to the clinician for conditions other than heart rate and atrial fibrillation. The SmartCardia 7L Platform performs retrospective analysis and identifies events which it shows the clinician when they are reviewing historical data. The clinician must then review these events and determine if they are indeed valid and should be included in a physician's report. These events are things like "Ventricular Bigeminy', 'Supraventricular Couplet', etc. A full list is included in the Clinician Instructions for Use.
Intended Use / Indications for Use:
The SmartCardia 7L Platform is a prescribed device used for the purpose of identifying non-lethal arrhythmias. The SmartCardia 7L Platform is intended for continuous external electrocardiogram (ECG) information to support Holter Monitoring, Extended Holter Monitoring, and Outpatient Cardiac Telemetry (OCT) commonly called Mobile Cardiac Telemetry (MCT) monitoring.The SmartCardia 7L Platform is intended for:1. Patients who experience transient symptoms that may suggest cardiac arrhythmia.2. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation).3. Patients with symptoms that may be due to cardiac arrhythmias. These

{6}------------------------------------------------

{7}------------------------------------------------

Summary of Substantial Equivalence:Substantial Equivalence – Predicate Device: InfoBionics MoMe Kardia, K160064
Intended Use:
MoMe Kardia is intended to be used for: Patients who experience transient symptoms that may suggest cardiac arrhythmia.Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation)Patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c) dyspnea (shortness of breath)Patients recovering from cardiac surgery or interventional procedures who are indicated for outpatient arrhythmia monitoring.ECG data recorded by the device can be analyzed by other processing systems to provide Holter style reports.
MoMe® Kardia is contraindicated for:
MoMe® Kardia is contraindicated for those patients requiring attended, in-hospital monitoring for life threatening arrhythmias.
Note: MoMe® Kardia does not provide interpretive statements. Interpretation and diagnosis is the responsibility of a physician.

may include but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c) dyspnea (shortness of breath).

  1. Patients recovering from cardiac surgery or interventional procedures who are indicated for outpatient arrhythmia monitoring.

  2. ECG data recorded by the device can be analyzed by other processing systems to provide Holter style reports. Measurements include: electrocardiogram (ECG signal), R-R interval, Heart Rate. Notification alerts can be set for one or more of these measures.

The SmartCardia 7L Platform is indicated for use on patients who are 18 years of age or older to provide monitoring of physiological information. It is intended for use in a physician office, outpatient facility, or in the patient's home.

SmartCardia 7L Platform contraindications:

  1. The SmartCardia 7L Platform is contraindicated for use for detection or notification of hemodynamically unstable or life-threatening arrhythmias or cardiac events requiring urgent medical response. It is not intended for monitoring patients during cardiac rehabilitation outside of healthcare facilities.

  2. The SmartCardia 7L Platform is contraindicated for use during external defibrillation.

{8}------------------------------------------------

Substantial Equivalence - Intended Use Comparison Statement:The indications for use of the InfoBionics MoMe Kardia predicate device andthe SmartCardia 7L Platform MCT device are substantially equivalent. TheSmartCardia 7L Platform MCT adds Outpatient Cardiac Telemetry, commonlycalled Mobile Cardiac Telemetry as a mode of operation in its indications foruse. The user population and environment of operation are the same. Theproduct code is changed to the newly released QYX. The regulation numberand special controls are the same as the predicate product code DSI, 870.1025.
The subject device SmartCardia 7L Platform MCT is an update to the ReferencePredicate Device SmartCardia 7L Web Browser software. The data acquisitionuses the same technology as the predicate device.
1) The same SmartCardia 7L Sensor and 7L Patch communicate wirelessly tothe same SmartCardia Phone that transmits the data to the sameSmartCardia 7L cloud based server for storage and further analysis.
2) The SmartCardia 7L Platform MCT has three modes of operation forHolter, Extended Holter and Event monitoring. All of these modes ofoperation provide continuous streaming of the ECG for viewing by a clinicianor qualified cardiac technician in the same manner as the PredicateDevice. The subject SmartCardia 7L Platform MCT updates the platform toinclude Outpatient Cardiac Telemetry (OCT) which is commonly calledMobile Cardiac Telemetry (MCT) as a mode of operation. The use of theSmartCardia 7L Platform MCT for Mobile Cardiac Telemetry is manged bySmartCardia through its supplier control requirements, personnel accesslimitations, monitoring settings, and timely notifications by an independentdiagnostic testing facility (IDTF) personnel.
Non-Clinical BenchPerformance TestingSummary:The SmartCardia 7L Platform was tested for compliance with FDAGuidance "Class II Special Controls Guidance Document: ArrhythmiaDetector and Alarm" released October 28, 2003.
Additional non- clinical bench performance testing was done perthe following:
• Basic Safety and Essential Performance per IEC 60601-1• Electromagnetic Compatibility per IEC 60601-2• Wireless Coexistence testing per FDA Guidance document "RadioFrequency Wireless Technology in Medical Devices, August 14, 2013",and exposure to RFID devices per AIM 7351731.• Wireless Coexisting testing per ANSI C63.27: 2017• ECG Monitoring for Ambulatory ECG devices per IEC 60601-2-47• Life Cycle testing per IEC 60601-11 Home Healthcare• ECG electrode testing per ANSI/AAMI EC12• Cleaning and disinfection methods as defined by SmartCardia SA• ECG Waveform Quality and Wearability• Effects of Sensor Positioning on ECG Performance• DVR-66 SCDB ECG Test Results (algorithm testing and reporting) perEC57
Objective(s) of the Test(s):To test the claimed ranges of measurement provided in thespecification and labeling that support the intended use and threemonitoring modes of operation: Holter and Extended Holter(Ambulatory electrocardiographic monitoring, Event Monitoring).

{9}------------------------------------------------

Test Methods

Bench testing was performed on an end-to-end system using electronic patient simulators and/or electronic test equipment (signal generators, etc.). This was performed per the applicable standards using production equivalent units.

Pass/Fail Criteria

Pass criteria is compliance with the claimed range and precision provided in the labeling for the SmartCardia 7L Platform. Deviations from the recognized standard criteria are identified in the applicable test report.

Results Summary

Testing showed that the SmartCardia 7L Platform met the pass/fail criteria established by the appropriate standards. The SmartCardia 7L Platform is contraindicated for use during defibrillation and is visibly labeled to indicate removal during defibrillation. Defibrillator protection testing was performed to ensure there was not excessive energy shunted away from the patient during defibrillation.

To verify the reproduction of the ECG and ensure that there are not significant morphological changes that could impact the

measurements and/or classifications, SmartCardia SA performed verification testing using a standard 12 lead ECG machine and comparing the morphology and ECG characteristics. The result of this testing is included in the document, DVR-67-01 ECG Waveform Quality and Wearability Report is included in this section.

The ECG detection algorithm was tested per ANSI/AAMI EC57 standard for testing and reporting performance results of cardiac rhythm and ST-segment measurement algorithms. The SmartCardia 7L Platform does not claim ST analysis or measurements. Therefore, the sections of the EC57 standard relating to ST measurement is not applicable. Because of the nature of the SmartCardia 7L Platform as a body worn sensor and patch device with the electrodes in close proximity to each other, and not utilizing ECG lead wires such as a traditional ECG device, SmartCardia developed a protocol and database for evaluating the analysis algorithm. The results show that the SmartCardia 7L Platform algorithm detection and reported results for the stated output arrhythmia were satisfactory and met industry norms.

Discussion/Conclusion: Based on a review of the test results and a comparison to the predicate devices characteristics and known specifications, the results show that the SmartCardia 7L Platform is substantially equivalent to the predicate and reference devices.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.