AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MissLan® Early Detection Pregnancy Test Strip is used for qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. This test is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the day of the missed period, (i.e. as early as five (5) days before the day of the expected period). It is intended for use by people who would like to find out whether they are pregnant in a home environment. Only for use outside the body. For over-the-counter use.

Important note regarding positive results:

Because this test detects low levels of hCG, it is possible that this test may give positive results even if you are not pregnant. All results should be confirmed by your healthcare provider, especially when making decisions about future medical care.

This device is intended for home-use only.

MissLan® Early Detection Pregnancy Test Cassette is used for qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. This test is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the day of the missed period, (i.e. as early as five (5) days before the day of the expected period). It is intended for use by people who would like to find out whether they are pregnant in a home environment. Only for use outside the body. For over-the-counter use.

Important note regarding positive results:

Because this test detects low levels of hCG, it is possible that this test may give positive results even if you are not pregnant. All results should be confirmed by your healthcare provider, especially when making decisions about future medical care.

This device is intended for home-use only.

MissLan® Early Detection Pregnancy Test Midstream is used for qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. This test is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the day of the missed period, (i.e. as early as five (5) days before the day of the expected period). It is intended for use by people who would like to find out whether they are pregnant in a home environment. Only for use outside the body. For over-the-counter use.

Important note regarding positive results:

Because this test detects low levels of hCG, it is possible that this test may give positive results even if you are not pregnant. All results should be confirmed by your healthcare provider, especially when making decisions about future medical care.

This device is intended for home-use only.

Device Description

MissLan® Early Detection Pregnancy Test will be sold in Strip. Cassette and Midstream format. The Strip format is a single test strip. The Cassette format consists of a single test strip assembled in a plastic housing. The Midstream format consists of a single test strip assembled in a plastic housing with an absorbent tip, and is designed to be tested in dip or midstream mode.

MissLan® Early Detection Pregnancy Test Strip. MissLan® Early Detection Pregnancy Test Cassette and MissLan® Early Detection Pregnancy Test Midstream each contains a pouch with the device and instructions, and in addition, cassette format is packaged with pipette dropper and urine collection cup.

The devices utilize a combination of antibodies to detect hCG in urine as well as to serve as a run control. Each device contains mouse monoclonal anti-ß-hCG antibody colloidal gold conjugate pre-dried on the sample pad. Mouse monoclonal anti-a-hCG antibody (on the Test Line) and goat anti mouse IgG polyclonal antibody (on the Control Line) are coated and immobilized on a nitrocellulose membrane. The result is displayed to the user in the test window as two lines for a 'Pregnant' positive result and one line for a 'Not Pregnant' negative result.

AI/ML Overview

The provided FDA 510(k) summary describes the acceptance criteria and study results for the MissLan® Early Detection Pregnancy Test Strip, Cassette, and Midstream devices.

Here's a breakdown of the requested information:

1. Table of acceptance criteria and reported device performance

The document does not explicitly state "acceptance criteria" as a separate section with numerical targets for each performance metric. However, the performance characteristics studies demonstrate that the device meets the necessary performance for its intended use, implying that the observed results constitute the criteria it should achieve. The predicate device's performance often forms the implicit benchmark for substantial equivalence.

Based on the provided data, the implicit acceptance criteria and reported performance are as follows:

Acceptance Criterion (Implicit)Reported Device Performance
Analytical Sensitivity (Limit of Detection): Ability to detect hCG at 10 mIU/mL.100% detection at 10 mIU/mL for all formats (Strip, Cassette, Midstream - both dip and in-stream methods) across all operators and lots.
Reproducibility: Consistent results across operators, lots, and days.Demonstrated high reproducibility across varying hCG concentrations, operators, and lots. For example, 100% positive agreement at 10, 15, and 25 mIU/mL, and 100% negative agreement at 0 and 3 mIU/mL.
Hook Effect: No false negative results at very high hCG concentrations.No hook effect observed up to 500,000 mIU/mL for all formats.
Interfering Substances: No interference from common substances in urine.No interference observed from 23 listed endogenous and exogenous substances (e.g., Albumin, Bilirubin, Glucose, Hemoglobin, common drugs) at specified concentrations.
Cross-reactivity: No false positives from structurally similar hormones (hLH, hFSH, hTSH).No cross-reactivity observed with 500 mIU/mL hLH, 1000 mIU/mL hFSH, and 1000 µIU/mL hTSH.
hCG ß-core fragment effect: Performance not affected by hCG ß-core fragments.Performance not affected by hCG ß-core fragment concentrations up to 500,000 pmol/L.
Urine pH effect: Performance robust across physiological urine pH range.Urine pH values between 4 and 9 do not affect performance.
Urine Density effect: Performance robust across physiological urine density range.Urine with relative density range of 1.000 to 1.035 does not affect performance.
Method Comparison with Predicate Device: High concordance with the legally marketed predicate.100% conformity between the candidate device (all formats) and the predicate device across 400 clinical urine samples (100 for each format/method).
Lay Person Readability/Accuracy: High agreement with professional results when used by laypersons.First Study: 100% positive and 100% negative conformity with professional results for all formats (total 400 laypersons, 100 for each format/method, using clinical samples). Second Study: Percent agreement with professional results ranges from 95% to 100% for various hCG spike concentrations (300 laypersons, 100 for each format, testing spiked urine samples). Specifically, 100% agreement at 0 and 10 mIU/mL, and 95-97% agreement at 3.5 and 6.5 mIU/mL.
Early Pregnancy Detection Rate: Ability to detect pregnancy early, especially before a missed period.Detection Rate at EMP-5: 76.92% (5 days before expected period). Detection Rate at EMP-4: 96.92%. Detection Rate at EMP-3 and later: 100% (3 days before expected period and thereafter).
False-Positive Rate (Specificity): No false positives in non-pregnant individuals.0% false positive results among 900 non-pregnant female urine samples (300 pre-menopausal, 300 peri-menopausal, 300 post-menopausal, across various formats and lots).

2. Sample size used for the test set and the data provenance

Analytical Performance (Precision/Reproducibility/Sensitivity):

  • Sample Size: 3 lots x 3 operators x 10 replicates x 9 hCG concentrations = 810 tests per device format. Since there are 4 effective formats (strip, cassette, midstream in-stream, midstream dip), this would be 810 x 4 = 3240 individual test results for this section.
  • Data Provenance: Retrospective. Negative female urine was spiked with hCG standard.

Method Comparison Study:

  • Sample Size: 400 women. Samples were divided: 100 for strip, 100 for cassette, 100 for midstream (dip), 100 for midstream (in-stream).
  • Data Provenance: Prospective, collected from women aged 18 to 51 at three clinical sites. Approximately half were early pregnant. Samples were collected randomly throughout the day. Country of origin not specified, but typically US or manufacturer's country due to regulatory context.

Lay Person Study (First Study - Clinical Samples):

  • Sample Size: 400 females (100 for each device format/method).
  • Data Provenance: Prospective, from "diverse educational and occupational backgrounds and ages ranging from 18 to 51 years old" at three sites. Country not specified.

Lay Person Study (Second Study - Spiked Samples):

  • Sample Size: 300 laypersons (100 for each device format). Each tested 4 blind-labeled spiked samples, so 300 * 4 = 1200 tests.
  • Data Provenance: Retrospective, using negative urine samples spiked with hCG. Country not specified.

Early Pregnancy Detection Study:

  • Sample Size: 65 pregnant women, contributing 650 early pregnancy urine samples (from day -8 to day +1 relative to the day of expected menstrual period).
  • Data Provenance: Prospective, collected from pregnant women. Country not specified.

Specificity Study (False-Positive Results Rate):

  • Sample Size: 900 urine samples from non-pregnant females (300 pre-menopausal, 300 peri-menopausal, 300 post-menopausal). These were divided further by device format (e.g., for each age group, 100 for strip, 100 for cassette, 50 for midstream dip, 50 for midstream in-stream).
  • Data Provenance: Clinical, collected from non-pregnant females at three sites. Country not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Analytical Performance (Precision/Reproducibility/Sensitivity): 3 operators performed the tests for the device, but the "ground truth" for hCG concentration was based on spiking with a standard traceable to the 5th WHO.
  • Method Comparison Study: "Three professionals at each site" tested the candidate device, and "one professional at each site" tested the predicate device. Their specific qualifications are not detailed beyond "professional." It's implied they are trained laboratory personnel.
  • Lay Person Study (First Study - Clinical Samples): "Professional" testing was used as ground truth for comparison with layperson results. The number of professionals is not specified, nor their detailed qualifications, but it implies a single assessment for ground truth by trained personnel.
  • Lay Person Study (Second Study - Spiked Samples): "Professional" testing was used as ground truth, but the ultimate ground truth was the known hCG concentration of the spiked samples.
  • Early Pregnancy Detection Study: The ground truth for pregnancy status was established by B-ultrasound and HCG levels for the 65 pregnant women/650 samples. This refers to established clinical diagnostic methods, implying expert medical interpretation of ultrasound and quantitative lab HCG results.
  • Specificity Study: The samples were collected from "non-pregnant females," implying a clinical determination of non-pregnancy as ground truth.

4. Adjudication method for the test set

  • The document does not explicitly describe an adjudication method (such as 2+1 or 3+1 consensus) for establishing ground truth from multiple experts.
  • In studies involving comparisons (e.g., method comparison, layperson study), results from "professionals" or "known spiked concentrations" served as the reference standard (ground truth). For the early pregnancy detection study, B-ultrasound and HCG levels were the ground truth, which are typically conclusive diagnostic methods.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a multi-reader, multi-case (MRMC) comparative effectiveness study was not conducted, nor is this device an AI-assisted device. This device is a rapid in-vitro diagnostic (IVD) test for qualitative detection of HCG. The "readers" are either laypersons or professionals interpreting a visual line on a test strip.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • This question is not applicable as the device is a manual, visually interpreted rapid diagnostic test, not an algorithm or software. It is inherently human-in-the-loop (for reading the result).
  • However, the analytical performance studies (e.g., precision, hook effect, interfering substances) could be considered "standalone" in the sense that they assess the device's chemical and physical characteristics independent of human interpretation variability, but still involve human technicians performing the tests.

7. The type of ground truth used

  • Analytical Performance Studies (Precision/Reproducibility, Hook Effect, Interfering Substances, Cross-Reactivity, HCG ß-core fragment, pH, density): Ground truth was based on known spiked concentrations of hCG (traceable to WHO standard) or interfering substances into negative urine.
  • Method Comparison Study: Ground truth was established by comparison to a legally marketed predicate device (Wondfo One Step HCG Urine Pregnancy Test) and/or potentially by clinical reference methods by professionals.
  • Lay Person Study (First Study): Ground truth was established by professional testing results of the same clinical urine samples.
  • Lay Person Study (Second Study): Ground truth was based on known spiked concentrations of hCG and confirmed by professional testing.
  • Early Pregnancy Detection Study: Ground truth was established by clinical B-ultrasound and HCG levels to confirm pregnancy status and gestational age.
  • Specificity Study: Ground truth was the clinical diagnosis of non-pregnant females.

8. The sample size for the training set

  • The document does not specify a separate "training set" for the device. This is typical for traditional rapid diagnostic tests, which are validated against established performance metrics rather than trained via machine learning. The studies described are validation and verification studies.

9. How the ground truth for the training set was established

  • As there's no specified "training set" in the context of machine learning, this question is not directly applicable. The ground truth for validation/verification studies was established as described in point 7.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 21, 2024 Guangzhou Decheng Biotechnology Co., Ltd. % Joe Shia Regulatory Consultant LSI International Inc. 504 East Diamond Ave., Suite H Gaithersburg, Maryland 20877

Re: K240643

Trade/Device Name: MissLan® Early Detection Pregnancy Test Strip; MissLan® Early Detection Pregnancy Test Cassette; MissLan® Early Detection Pregnancy Test Midstream Regulation Number: 21 CFR 862.1155 Regulation Name: Human Chorionic Gonadotropin (HCG) Test System Regulatory Class: Class II Product Code: LCX Dated: May 20, 2024 Received: May 20, 2024

Dear Joe Shia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Paula V. Caposino -S

Paula Caposino, Ph.D. Deputy Division Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K240643

Device Name

MissLan® Early Detection Pregnancy Test Strip MissLan® Early Detection Pregnancy Test Cassette MissLan® Early Detection Pregnancy Test Midstream

Indications for Use (Describe)

MissLan® Early Detection Pregnancy Test Strip is used for qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. This test is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the day of the missed period, (i.e. as early as five (5) days before the day of the expected period). It is intended for use by people who would like to find out whether they are pregnant in a home environment. Only for use outside the body. For over-the-counter use.

Important note regarding positive results:

Because this test detects low levels of hCG, it is possible that this test may give positive results even if you are not pregnant. All results should be confirmed by your healthcare provider, especially when making decisions about future medical care.

This device is intended for home-use only.

MissLan® Early Detection Pregnancy Test Cassette is used for qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. This test is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the day of the missed period, (i.e. as early as five (5) days before the day of the expected period). It is intended for use by people who would like to find out whether they are pregnant in a home environment. Only for use outside the body. For over-the-counter use.

Important note regarding positive results:

Because this test detects low levels of hCG, it is possible that this test may give positive results even if you are not pregnant. All results should be confirmed by your healthcare provider, especially when making decisions about future medical care.

This device is intended for home-use only.

MissLan® Early Detection Pregnancy Test Midstream is used for qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. This test is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the day of the missed period, (i.e. as early as five (5) days before the day of the expected period). It is intended for use by people who would like to find out whether they are pregnant in a home environment. Only for use outside the body. For over-the-counter use.

Important note regarding positive results:

Because this test detects low levels of hCG, it is possible that this test may give positive results even if you are not pregnant. All results should be confirmed by your healthcare provider, especially when making decisions about future medical care.

This device is intended for home-use only.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

K240643

1.Date:June 21, 2024
2.Submitter:Guangzhou Decheng Biotechnology Co., Ltd.Floor 3/4/5/7, Building A1, No.12, Nanyun 1stRoad, Science City, Huangpu District,Guangzhou, Guangdong, 510663, P.R. China
3.Contact person:Joe ShiaLSI International Inc.504 East Diamond Ave., Suite HGaithersburg, MD 20877Telephone: 240-505-7880Email: shiajl@yahoo.com
4.Device Name:MissLan® Early Detection Pregnancy Test StripMissLan® Early Detection Pregnancy Test CassetteMissLan® Early Detection Pregnancy Test Midstream
Classification:Product Code:CFR:Class IILCX862.1155
5.Predicate Devices:Wondfo One Step HCG Urine Pregnancy Test Midstream,Wondfo One Step HCG Urine Pregnancy Test Strip,Wondfo One Step HCG Urine Pregnancy Test Cassette(K150022)

6. Intended Use

MissLan® Early Detection Pregnancy Test Strip is used for qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. This test is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the day of the missed period, (i.e. as early as five (5) days before the day of the expected period). It is intended for use by people who would like to find out whether they are pregnant in a home environment. Only for use outside the body. For over-the-counter use.

Important note regarding positive results:

Because this test detects low levels of hCG, it is possible that this test may give positive results even if you are not pregnant. All results should be confirmed by your healthcare provider, especially when making decisions about future medical care.

{5}------------------------------------------------

This device is intended for home-use only.

MissLan® Early Detection Pregnancy Test Cassette is used for qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. This test is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the day of the missed period. (i.e. as early as five (5) days before the day of the expected period). It is intended for use by people who would like to find out whether they are pregnant in a home environment. Only for use outside the body. For over-the-counter use.

Important note regarding positive results:

Because this test detects low levels of hCG, it is possible that this test may give positive results even if you are not pregnant. All results should be confirmed by your healthcare provider, especially when making decisions about future medical care.

This device is intended for home-use only.

MissLan® Early Detection Pregnancy Test Midstream is used for qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. This test is intended for use as an aid in early detection of pregnancy, in some cases as early as six (6) days before the day of the missed period, (i.e. as early as five (5) days before the day of the expected period). It is intended for use by people who would like to find out whether they are pregnant in a home environment. Only for use outside the body. For over-the-counter use.

Important note regarding positive results:

Because this test detects low levels of hCG, it is possible that this test may give positive results even if you are not pregnant. All results should be confirmed by your healthcare provider, especially when making decisions about future medical care.

This device is intended for home-use only.

7. Device Description

MissLan® Early Detection Pregnancy Test will be sold in Strip. Cassette and Midstream format. The Strip format is a single test strip. The Cassette format consists of a single test strip assembled in a plastic housing. The Midstream format consists of a single test strip assembled in a plastic housing with an absorbent tip, and is designed to be tested in dip or midstream mode.

MissLan® Early Detection Pregnancy Test Strip. MissLan® Early Detection Pregnancy Test Cassette and MissLan® Early Detection Pregnancy Test Midstream each contains a pouch with the device and instructions, and in addition, cassette format is packaged with pipette dropper and urine collection cup.

The devices utilize a combination of antibodies to detect hCG in urine as well as to serve as a run control. Each device contains mouse monoclonal anti-ß-hCG antibody colloidal gold conjugate pre-dried on the sample pad. Mouse monoclonal

{6}------------------------------------------------

anti-a-hCG antibody (on the Test Line) and goat anti mouse IgG polyclonal antibody (on the Control Line) are coated and immobilized on a nitrocellulose membrane. The result is displayed to the user in the test window as two lines for a 'Pregnant' positive result and one line for a 'Not Pregnant' negative result.

Similarities
ItemCandidate devicePredicate device
Intended useEarly detection of pregnancySame
SpecimenUrineSame
Assay technicalImmunochromatographicassaySame
Sensitivity10 mIU/mLSame
ResultsQualitativeSame
Device formatStrip, Cassette, MidstreamSame
Differences
ItemDevicePredicate
Target userFor over-the-counter usePrescription use and OTC use
Read time3-10 minutes5 minutes

Substantial Equivalence Information 8.

9. Test Principle

MissLan® Early Detection Pregnancy Test Strip, MissLan® Early Detection Pregnancy Test Cassette and MissLan® Early Detection Pregnancy Test Midstream use lateral flow immunoassay for in vitro qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. If hCG is present in the sample, it will reach the Test Zone (T) of the membrane and form a colored line. When the test is performed properly, a colored line will always appear in the Control Zone (C). The test result is shown in the result window and read visually in 3 minutes of urine addition. Two distinct colored lines, one in the Test Zone (T) and another in the Control Zone (C) indicate a positive test result (pregnant). Absence of a colored line in the Test Zone (T) and only a colored line in the Control Zone (C) indicates a negative test result (not pregnant). Absence of a colored line in the Control Zone(C) even in the presence of a colored line in the Test Zone (T) indicates an invalid test result.

10. Performance Characteristics

A. Analytical performance

a. Precision/Reproducibility/Sensitivity

Negative female urine was spiked with hCG standard (Traceable to the 5th WHO) to hCG concentrations of 0, 3, 3.5 5, 6.5, 7.5, 10, 15 and 25 mIU/mL. Each sample

{7}------------------------------------------------

was tested in 10 replicates per day for 5 days for each device lot. Total of three device lots for each format were tested and each operator tested one lot separately. Tests were performed by three different operators for each sample concentration. The midstream format was performed using both dip and in-stream methods. The results are summarized in the table below:

hCGConcentration(mIU/mL)Operator 1Lot IOperator 2Lot IIOperator 3Lot IIITotal result%Negative%Positive
-+-+-+-+
05005005001500100%0%
35005005001500100%0%
3.54464554551341689.33%10.67%
53614371336141094172.67%27.33%
6.5272323272624767450.67%49.33%
7.51040133713373611424.00%76.00%
1005005005001500%100%
1505005005001500%100%
2505005005001500%100%

Strip format

Cassette format

hCGConcentration(mIU/mL)Operator 1Lot IOperator 2Lot IIOperator 3Lot IIITotal result%Negative%Positive
-+-+-+-+
05005005001500100%0%
35005005001500100%0%
3.54554554551351590.00%10.00%
53713351535151074371.33%28.67%
6.5232726242426737748.67%51.33%
7.51139133710403411622.67%77.33%
1005005005001500%100%
1505005005001500%100%
2505005005001500%100%

Midstream format (in-stream method)

hCGConcentration(mIU/mL)Operator 1Lot IOperator 2Lot IIOperator 3Lot IIITotal result% Negative% Positive
-+-+-+-+
05005005001500100%0%
35005005001500100%0%
3.54554554551351590.00%10.00%
53713361437131104073.33%26.67%
6.5242626242426747649.33%50.67%
7.51139123814363711324.67%75.33%
1005005005001500%100%
1505005005001500%100%
2505005005001500%100%

Midstream format (dip method)

{8}------------------------------------------------

hCGConcentration(mIU/mL)Operator 1Lot IOperator 2Lot IIOperator 3Lot IIITotal result% Negative% Positive
-+-+-+-+
05005005001500100%0%
35005005001500100%0%
3.54554464551341689.33%10.67%
53713361435151084272.00%28.00%
6.5232724262624737748.67%51.33%
7.5123814369413511523.33%76.67%
1005005005001500%100%
1505005005001500%100%
2505005005001500%100%

MissLan® Early Detection Pregnancy Test exhibited reproducibility of results. Based on the above results, the sensitivity of MissLan® Early Detection Pregnancy Test is demonstrated to be 10 mIU/mL.

b. Linearity/assay reportable range:

Linearity is not applicable since this is a qualitative test. The test device was evaluated for high dose or hook effect.

Hook effect test:

Negative urine samples were spiked with varying hCG concentrations (6,250 mIU/mL, 12,500 mIU/mL, 25,000 mIU/mL, 50,000 mIU/mL, 100,000 mIU/mL, 200,000 mIU/mL and 500,000 mIU/mL). The results demonstrated that no hook effect was observed at hCG concentration up to 500,000 mIU/mL.

c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability:

MissLan® Early Detection Pregnancy Test is calibrated against reference material traceable to WHO International Standard 5th edition, NIBSC code 07/364.

Stability:

The shelf-life of the MissLan® Early Detection Pregnancy Test at 2~30℃ is 36 months based on real time stability data.

d. Analytical specificity

Interfering substance:

To evaluate potential interfering substances of the MissLan® Early Detection Pregnancy Test, urine samples containing 0 mIU/mL, 3 mIU/mL and 10 mIU/mL hCG were spiked with interfering substances to the concentrations listed below. No interference effect was observed at the tested concentrations.

SubstanceConcentration
Albumin2000 mg/dL

{9}------------------------------------------------

SubstanceConcentration
Bilirubin40 mg/dL
Glucose2000 mg/dL
Hemoglobin1000 mg/dL
Uric acid23.5 mg/dL
Ketone20 mg/dL
β-hydroxybutyrate2000 mg/dL
Pregnanediol1.5 mg/dL
Acetaminophen20 mg/dL
Acetylsalicylic acid80 mg/dL
Amoxicillin20 mg/dL
Ampicillin20 mg/dL
Ascorbic acid20 mg/dL
Atropine20 mg/dL
Benzoylecgonine10 mg/dL
Caffeine20 mg/dL
Cannabinol10 mg/dL
Codeine6 µg/dL
EDTA80 mg/dL
Ephedrine20 mg/dL
Ethanol1%
Folic acid0.03 mg/dL
Salicylic acid20 mg/dL
Gentisic acid20 mg/dL
Ibuprofen40 mg/dL
Phenothiazine20 mg/dL
Phenylpropanolamine20 mg/dL
Tetracycline20 mg/dL
Thiophene20 mg/dL
Vitamin B180 mg/dL

Cross-reactivity :

To evaluate cross-reactivity, negative and positive urine samples (0 mIU/mL, 3 mIU/mL and 10 mIU/mL hCG) were spiked with potential cross reactants (500 mIU/mLhLH, 1000 mIU/mL hFSH, 1000 µIU/mL hTSH). No cross-reactivity was observed at tested concentration.

Effects of hCG ß-core fragment:

To evaluate the effect of the hCG ß-core fragment, negative urine samples (0 mIU/mL and 3 mIU/mL hCG) and positive urine samples (10 mIU/mL and 20,000 mIU/mL hCG) were spiked with hCG ß-core fragment (hCGBcf) at concentrations of 50,000 pmol/L, 125,000 pmol/L, 250,000 pmol/L and 500,000 pmol/L. The performance of MissLan® Early Detection Pregnancy Test was not affected by hCG ß-core fragment concentrations up to 500,000 pmol/L.

{10}------------------------------------------------

Effects of urine pH:

To evaluate the effect of urine pH on the results of MissLan® Early Detection Pregnancy Test, urine samples containing 0 mIU/mL, 3 mIU/mL and 10 mIU/mL hCG were tested at pH values of 4, 5, 6, 7, 8 and 9. The results indicated that urine pH ranges between 4 and 9 does not affect the performance of MissLan® Early Detection Pregnancy Test.

Effects of urine density:

To evaluate the effect of urine density on the results of MissLan® Early Detection Pregnancy Test, urine samples containing 0 mIU/mL, 3 mIU/mL and 10 mIU/mL hCG were tested at density values of 1.000, 1.005, 1.010, 1.015, 1.020, 1.025, 1.030 and 1.035. The results indicated that urine with a relative density rang of 1.000 to 1.035 does not affect the performance of MissLan® Early Detection Pregnancy Test.

B. Method comparison study

Method comparison with predicate device:

The performance of the candidate device was compared to the predicate device. Urine samples were collected from 400 women aged 18 to 51 at three clinical sites. Approximately half of the subjects were pregnant in the early stage of less than 5 weeks. Samples were randomly collected at various time throughout the day.

All samples were tested by three professionals at each site using the candidate device and by one professional at each site using the predicate device. 100 samples were tested for strip format, and 100 samples were tested for cassette format; for the midstream format, both dip testing (100 samples) and in-stream testing (100 samples) were evaluated. The samples were blinded and randomized before being tested by professionals. All results are summarized in the table below.

Candidate devicePredicate device
PositiveNegativeTotal
Strip formatPositive49049
Negative05151
Total4951100
Candidate devicePredicate device
PositiveNegativeTotal
Cassette formatPositive51051
Negative04949
Total5149100
Candidate devicePredicate device
PositiveNegativeTotal
Midstream formatPositive48048

{11}------------------------------------------------

dip method)Negativet The Children the Children the Children the Children the Children the Children the Children the Children the Children the Children the Children the Children the Children the(ﺎ ﻟﻠﺘﻌﻠﻴﻤﺎﺕ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍ
Totalit LN I100
Predicate device
Candidate devicePositiveNegativeTotal
Midstream format(in-stream method)Positive49049
Negative05151
Total4951100

The conformity between MissLan® Early Detection Pregnancy Test (strip / cassette / midstream) and the predicate device is 100%.

C. Clinical studies

    1. Clinical Sensitivity:
      Not applicable.
    1. Clinical Specificity:
      Not applicable.
    1. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
  • 3.1 Lay person study

First study:

A lay user study was performed at three sites with a total of 400 females with diverse educational and occupational backgrounds and ages ranging from 18 to 51 years old. Each subject tested her own urine sample using the device according to the instructions for use and provided a sample for professional testing. 100 subjects tested the strip format, 100 subjects tested the cassette format, 100 subjects tested the midstream format using dip method, and 100 subjects tested the midstream format using in-stream method.

The results are summarized in the table below.

Strip format

Professional
Strip formatPositiveNegativeTotal
LaypersonPositive49049
Negative05151
Total4951100

Cassette format

Professional
Cassette formatPositiveNegativeTotal
LaypersonPositive51051
Negative04949

{12}------------------------------------------------

Total5149100
--------------------

Midstream format

Midstream format(dip method)Professional
PositiveNegativeTotal
LaypersonPositive48048
Negative05252
Total4852100
Midstream format(in-stream method)Professional
PositiveNegativeTotal
LaypersonPositive49049
Negative05151
Total4951100

From the above tables, the lay person results showed 100% positive and 100% negative conformity with the professional results.

Second study:

Negative urine sample pools were spiked with 3 mIU/mL, 3.5 mIU/mL, 6.5 mIU/mL and 10 mIU/mL hCG. All aliquots were blind labeled by the person who prepared the samples and didn't take part in the sample testing. A total of 300 laypersons performed the testing, of which 100 using the strip format, 100 using the cassette format and 100 using the midstream format. Each layperson tested 4 blind-labeled samples (3 mIU/mL, 3.5 mIU/mL, 6.5 mIU/mL and 10 mIU/mL), and an aliquot of each of the spiked urine samples was also tested by the professional.

Results of spiked samples are summarized in the table below.

Strip format:

hCGConcentration(mIU/mL)Layperson resultProfessional resultPercentAgreement
No. ofPositiveNo. ofNegative%positiveNo. ofPositiveNo. ofNegative%positive
301000%01000%100%
3.59919%109010%95%
6.5495149%514951%96%
101000100%1000100%100%

Cassette format:

hCGConcentration(mIU/mL)Layperson resultProfessional resultPercentAgreement
No. ofPositiveNo. ofNegative%positiveNo. ofPositiveNo. ofNegative%positive

{13}------------------------------------------------

31000%1000%100%
3.5109010%118911%તે તે જેની જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામમાં મુખ્યત્વે ખેત
6.5485248%224852%96%
10100100%100100%100%

Midstream format:

hCGConcentration(mIU/mL)Layperson resultProfessional resultPercentAgreement
No. ofPositiveNo. ofNegative%positiveNo. ofPositiveNo. ofNegative%positive
301000%01000%100%
3.5109010%118911%95%
6.5495149%524852%97%
101000100%1000100%100%

The results above show that MissLan® Early Detection Pregnancy Test can be used by the laypersons and get the correct results.

3.2 Early pregnancy detection study

A study was performed to evaluate the early pregnancy detection rate of MissLan® Early Detection Pregnancy Test. A total of 650 early pregnancy urine samples from day -8 to +1 relative to the day of the expected menstrual period were collected from 65 pregnant women. Each sample was tested with each format of the MissLan® Early Detection Pregnancy Test. For the midstream format, both dip and in-stream methods were evaluated.

The specific detection rates per day are summarized in the table below. Strip format:

Test dateEMP-8EMP-7EMP-6EMP-5EMP-4EMP-3EMP-2EMP-1EMPEMP+1
# of positiveresults of B-ultrasound65656565656565656565
# of positiveresults of HCG182550636565656565
Detection rate1.54%12.31%38.46%76.92%96.92%100%100%100%100%100%

Cassette format:

Test dateEMP-8EMP-7EMP-6EMP-5EMP-4EMP-3EMP-2EMP-1EMPEMP+1
# of positiveresults of B-ultrasound65656565656565656565
# of positiveresults of HCG182550636565656565
Detection rate1.54%12.31%38.46%76.92%96.92%100%100%100%100%100%

Midstream format (in-stream method):

Test dateEMP-8EMP-7EMP-6EMP-5EMP-4EMP-3EMP-2EMP-1EMPEMP+1
# of positive65656565656565656565

{14}------------------------------------------------

results of B-ultrasound
# of positiveresults of HCG182550636565656565
Detection rate1.54%12.31%38.46%76.92%96.92%100%100%100%100%100%

Midstream format (dip method):

Test dateEMP-8EMP-7EMP-6EMP-5EMP-4EMP-3EMP-2EMP-1EMPEMP+1
# of positiveresults of B-ultrasound65656565656565656565
# of positiveresults of HCG182550636565656565
Detection rate1.54%12.31%38.46%76.92%96.92%100%100%100%100%100%

The detection rate of MissLan® Early Detection Pregnancy Test (strip/cassette/ midstream) in EMP-5 (i.e. five days before the expected menstrual period, 6 days before the missed menstrual period) was 76.92%, in EMP-4 was 96.92%, and in EMP-1 was 100%.

3.3 Specificity Study to Determine False-Positive Results Rate

900 urine samples were collected from non-pregnant females in pre-menopausal (18-40 years old), peri-menopausal (41-55 years old) and post-menopausal (>55 years old) age groups at three sites. 300 participants for each age group. In each age group, 100 participants test strip, 100 participants tested test cassette, 50 participants tested test midstream using dip method, and 50 participants tested test midstream using in-stream method. Three batches of each format were used in the study. No false positive results were observed for all age groups.

Strip format:

GroupLot ILot IILot IIITotal result
Pre-menopausal0+/34-0+/33-0+/33-0+/100-
Peri-menopausal0+/33-0+/34-0+/33-0+/100-
Post-menopausal0+/33-0+/33-0+/34-0+/100-

Cassette format:

GroupLot ILot IILot IIITotal result
Pre-menopausal0+/34-0+/33-0+/33-0+/100-
Peri-menopausal0+/33-0+/34-0+/33-0+/100-
Post-menopausal0+/33-0+/33-0+/34-0+/100-

Midstream format:

GroupLot ILot IILot IIITotal result
Pre-menopausal0+/34-0+/33-0+/33-0+/100-
Peri-menopausal0+/33-0+/34-0+/33-0+/100-
Post-menopausal0+/33-0+/33-0+/34-0+/100-

{15}------------------------------------------------

11. Conclusion

Based on the test principle and performance characteristics of the device including precision, cut-off, interference, specificity, method comparison, lay-user study and early pregnancy detection study of the devices, it can be concluded that MissLan® Early Detection Pregnancy Test is substantially equivalent to the predicate device.

§ 862.1155 Human chorionic gonadotropin (HCG) test system.

(a)
Human chorionic gonadotropin (HCG) test system intended for the early detection of pregnancy —(1)Identification. A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class II.(b)
Human chorionic gonadotropin (HCG) test system intended for any uses other than early detection of pregnancy —(1)Identification. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class III.(3)
Date PMA or notice of completion of a PDP is required. As of the enactment date of the amendments, May 28, 1976, an approval under section 515 of the act is required before the device described in paragraph (b)(1) may be commercially distributed. See § 862.3.