K Number
K240482
Date Cleared
2024-05-17

(87 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Intense Pulsed Light System is intended for medical use in the treatment of the following conditions:

  1. Moderate inflammatory acne vulgaris;
  2. Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles);
  3. Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial truncal and leg telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations;
  4. Permanent hair reduction-long-term stable reduction in number of hairs re-growing after a treatment regimen.
Device Description

Intense Pulsed light (IPL) System is a type of intensive, broadband, coherent light source which has a wavelength spectrum of 420 nm -1200 nm. There are six optical filters that can be using in this system listed in table. With these special properties, the IPL System has a wide application in non-ablative therapies based on theory of human skin tissue's selective absorption.

AI/ML Overview

The provided FDA 510(k) summary (K240482) for the Smedtrum Medical Technology Co., Ltd. Intense Pulsed Light System (ST-690) indicates that the device's acceptance criteria are based on bench testing for optical energy output and electrical safety, and its performance is demonstrated by compliance with relevant IEC standards.

Here's an analysis of the requested information:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Electrical Safety and Electromagnetic Compatibility: Compliance with international safety and EMC standards.Complies with: - IEC 60601-1: 2005+corr.1:2006+Corr.2.2007+A1:2012 (General requirements for basic safety and essential performance) - IEC 60601-1-2:2020 (Electromagnetic disturbances) - IEC 62471 First edition 2006-07 (Photobiological safety) - IEC 60601-2-57 Edition 1.0 2011-01 (Particular requirements for non-laser light source equipment)
Bench Testing (Optical Energy Output): Spatial variation of the LS equipment output over the treatment area shall not deviate from the average irradiance or radiant exposure by more than ±20%.The product fulfills the requirements of IEC 60601-2-57 (which includes parameters for optical energy output and spatial variation). Specifically, the document states "The Intense Pulsed Light system... has been determined through engineering testing to verify optical energy output... The product fulfills the requirements of IEC 60601-2-57."
Intended Use and Indications for Use: Equivalence to the predicate device.The device has the same intended use and indications for use as the predicate device (K231394).
Technological Characteristics and Operating Principles: Equivalence to the predicate device.The device has the same technological characteristics, energy used, and operating principles as the predicate device (K231394), with minor differences in form factor (tabletop vs. standing, single handpiece vs. potentially more on predicate).

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify a patient-based test set sample size or data provenance (country, retrospective/prospective). The performance data cited is based on engineering and bench testing to verify hardware and software functionality and safety, not clinical trial data with human subjects.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable to this submission. The ground truth relies on engineering and safety standards compliance, not expert consensus on clinical data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable. There was no clinical test set requiring human adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is an Intense Pulsed Light System, a physical medical device for treatment, not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a physical therapeutic system, not an algorithm. Performance assessment is based on physical and electrical output specifications.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth used for this submission is compliance with established international engineering and electrical safety standards (IEC 60601-1, IEC 60601-1-2, IEC 62471, IEC 60601-2-57) and verification of optical energy output specifications.

8. The sample size for the training set

This is not applicable. There is no "training set" in the context of an algorithm or AI model for this type of medical device submission. The device is evaluated based on its physical and electrical performance against predefined engineering standards.

9. How the ground truth for the training set was established

This is not applicable as there is no training set. The "ground truth" for the device's performance relies on engineering measurements and adherence to specified performance ranges outlined in the electrical and optical standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym and name on the right. The FDA acronym and name are in blue, with the acronym in a square and the name in a sans-serif font.

May 17, 2024

Smedtrum Medical Technology Co., Ltd. Crimson Wu Senior Regulatory Engineer 1F., No. 8, Ln. 97, Wugong Rd., Xinzhuang Dist. New Taipei City, 248016 Taiwan

Re: K240482

Trade/Device Name: Intense Pulsed Light System (ST-690) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: ONF Dated: February 16, 2024 Received: February 20, 2024

Dear Crimson Wu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).

{2}------------------------------------------------

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tanisha L. Digitally signed by Hithe -S - Date: 2024.05.17
15:10:25 -04'00'

Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240482

Device Name Intense Pulsed Light System (ST-690)

Indications for Use (Describe)
----------------------------------

The Intense Pulsed Light System is intended for medical use in the treatment of the following conditions:

  1. Moderate inflammatory acne vulgaris;

  2. Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles);

  3. Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial truncal and leg telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations;

  4. Permanent hair reduction-long-term stable reduction in number of hairs re-growing after a treatment regimen.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a logo for a medical company. The logo features a stylized blue and gray geometric shape resembling a stylized letter "S" or a medical symbol. To the right of the symbol, there is text in Chinese characters followed by the word "Smart" in a modern sans-serif font. Below "Smart", the word "MEDICAL" is written in smaller letters.

510(k) Summary: K240482 Intense Pulsed Light System

SUBMITTER I.

Smedtrum Medical Technology Co., Ltd. 1F., No. 8, Ln. 97, Wugong Rd., Xinzhuang Dist., New Taipei City 248016, Taiwan (R.O.C.) Contact Person

Crimson Wu Position: Senior Regulatory Engineer Tel: +886-2-2298-9578, Ext. 301 Fax: +886-2-2298-9426 E-mail: crimsonwu@smedtrum.com Date of preparation: May 17th, 2024

II. DEVICE

Trade Name:Intense Pulsed Light System
Common or Usual Name:Intense Pulsed Light System
Product code:ONF
Classification Name:Powered Light Based Non-Light Surgical InstrumentWith Thermal Effect

21 C.F.R. § 878.4810, Device Class II

PREDICATE DEVICE III.

ManufacturerSmedtrum Medical Technology Co., Ltd.
Trade Name:Intense Pulsed Light System
Common or Usual Name:Laser surgical instrument for use in general and plasticsurgery and in dermatology
Classification Name:Powered Light Based Non-Light Surgical InstrumentWith Thermal Effect21 C.F.R. § 878.4810, Device Class II
Premarket Notification:K231394 Aug 9th, 2023

IV. DEVICE DESCRIPTION

Intense Pulsed light (IPL) System is a type of intensive, broadband, coherent light source which has a wavelength spectrum of 420 nm -1200 nm. There are six optical filters that can be using in this system listed in table. With these special properties, the IPL System has a wide application in non-ablative therapies based on theory of human

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains a logo for Smedtrum Medical Technology Co., LTD. The logo features a stylized "S" shape in shades of blue and gray on the left side. To the right of the shape is the company name, "Smedtrum," in a light gray sans-serif font, with the words "MEDICAL TECHNOLOGY CO.,LTD." in smaller letters below. Above the company name is the company name in Chinese.

skin tissue's selective absorption.

SpectraApplication areasExpected Working Life
420 -1200mmAcne100,000 shots
510 -1200mmAcne, vascular, pigment80,000 shots
560 -1200nmVascular, pigmentation70,000 shots
610 -1200mmHair removal60,000 shots
640 -1200nmHair removal50,000 shots
610 – 980nm (SHR)Hair removal30,000 shots

V. INDICATIONS FOR USE

The Intense Pulsed Light System is intended for medical use in the treatment of the following conditions:

    1. Moderate inflammatory acne vulgaris;
  1. Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles);

  2. Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial truncal and leg telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations;

  3. Permanent hair reduction-long-term stable reduction in number of hairs re-growing after a treatment regimen.

COMPARISON OF TECHNOLOGICAL CHARATERISTICS WITH THE PREDICATE DEVICE

FeatureProposed devicePredicate device(K231394)
Device NameIntense Pulsed Light SystemIntense Pulsed Light System
Product CodeONFONF
Regulation No.21 CFR 878.481021 CFR 878.4810
Device ClassClass IIClass II
FeatureProposed devicePredicate device(K231394)
Indication forUseThe Intense Pulsed LightSystem is intended for medicaluse in the treatment of thefollowing conditions:The Intense Pulsed Light System isintended for medical use in thetreatment of the followingconditions:
1. Moderate inflammatory acnevulgaris;1. Moderate inflammatory acnevulgaris;
2. Benign pigmented epidermallesions including dyschromia,hyperpigmentation, melasma,ephelides (freckles);2. Benign pigmented epidermallesions including dyschromia,hyperpigmentation, melasma,ephelides (freckles);
3. Benign cutaneous vascularlesions including port winestains, hemangiomas, facialtruncal and leg telangiectasias,erythema of rosacea, leg veins,spider angiomas and venousmalformations;3. Benign cutaneous vascularlesions including port wine stains,hemangiomas, facial truncal andleg telangiectasias, erythema ofrosacea, leg veins, spider angiomasand venous malformations;
4. Permanent hair reduction-long-term stable reduction innumber of hairs re-growingafter a treatment regimen.4. Permanent hair reduction-long-term stable reduction in numberof hairs re-growing after atreatment regimen.
Light SourceIntense Pulsed light (XenonIntense Pulsed light (Xenon Flash
Flash Lamp)Lamp)
Wavelength420 – 1200 nm420 – 1200 nm
Light DeliverySystemHandpieceHandpiece
Filter420 -1200nm: Acne;510 -1200nm: Acne, vascular,pigment;560 -1200nm: Vascular,pigment;610-1200nm: Hair removal;640-1200nm: Hair removal;610-980nm: Hair removal;420 -1200nm: Acne;510 -1200nm: Acne, vascular,pigment;560 -1200nm: Vascular, pigment;610-1200nm: Hair removal;640-1200nm: Hair removal;610-980nm: Hair removal;
Fluence420 -1200nm: 4.1-34.8 J/cm²;510 -1200nm: 3.8-31.6 J/cm²;560 -1200nm: 3.9-30.2 J/cm²;610-1200nm: 3.9-27.8 J/cm²;640-1200nm: 3.3-24.9 J/cm²;610-980nm: 3.2-23.3 J/cm²;420 -1200nm: 4.1-34.8 J/cm²;510 -1200nm: 3.8-31.6 J/cm²;560 -1200nm: 3.9-30.2 J/cm²;610-1200nm: 3.9-27.8 J/cm²;640-1200nm: 3.3-24.9 J/cm²;610-980nm: 3.2-23.3 J/cm²;
Pulsed Energydensity4.135 J/cm² (1050 Level)4.135 J/cm² (1050 Level)
Pulsed width5~20 ms5~20 ms
FeatureProposed devicePredicate device(K231394)
Pulsedduration5~50 ms5~50 ms
Spot Size12 mm × 35 mm &15 mm × 50 mm12 mm × 35 mm &15 mm × 50 mm
Coolingwater + air + TECwater + air + TEC
AppearanceHeight 451 mmHeight 1190 mm
Width 468 mmWidth 590 mm
Depth 630 mmDepth 550 mm
Weight Approx. 45 kgWeight Approx. 68 kg

{6}------------------------------------------------

Image /page/6/Picture/6 description: The image contains the logo for Smedtrum Medical Technology Co. LTD. The logo features the company name in both Chinese and English. The English name "Smedtrum" is prominently displayed in a large, modern font, while the Chinese name, 巨興醫學科技股份有限公司, is above it in a smaller font. To the left of the text is an abstract graphic of teal-colored geometric shapes.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image contains a logo for Smedtrum Medical Technology Co., Ltd. The logo features a stylized geometric shape in shades of blue and gray on the left. To the right of the shape is the company name, "Smedtrum," in a light blue sans-serif font, with the words "MEDICAL TECHNOLOGY CO., LTD." in smaller, gray font below it.

VI. PERFORMANCE DATA

The Intense Pulsed Light System has been determined through engineering testing to verify optical energy output and electrical safety.

Electrical safety and electromagnetic compatibility

The test results demonstrated that the proposed device complies with the following standards:

IEC 60601-1: 2005+corr.1:2006+Corr.2.2007+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

IEC 60601-1-2:2020 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances - Requirements and tests

IEC 62471 First edition 2006-07 Photobiological safety of lamps and lamp systems

IEC 60601-2-57 Edition 1.0 2011-01 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use

Bench testing

Intense Pulsed Light system is an Light source equipment. The spatial variation of the LS equipment output over the treatment area shall not deviate from the average irradiance or radiant exposure by more than ±20%

The product fulfills the requirements of IEC 60601-2-57.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Smedtrum Medical Technology Co., LTD. The logo consists of a geometric shape in shades of blue and gray on the left, followed by the company name in Chinese and English. The word "Smedtrum" is in a larger font size and is colored in light blue and gray.

VII. CONCLUSION

The Intense Pulsed Light System (ST-690) has the same intended use, same indications for use, the same technological characteristics, the same energy used, and the same operating principles as its predicates. The non-clinical data and performance testing reports in this submission demonstrate that Intense Pulsed Light System meets the expected performance requirements. ST-690 is designed in Tabletop using without caster, but predicate device is designed in Standing using with moving by caster. ST-690 can only setup one handpiece on the device. There is no big difference in intended use or indications for use for the Intense Pulsed Light System. Any difference between the subject and predicate device do not raise new issues of safety or effectiveness. Based on above analysis, the Intense Pulsed Light System is as safe, as effective, and performs as well as the cited predicate device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.