K Number
K240423
Manufacturer
Date Cleared
2024-04-08

(55 days)

Product Code
Regulation Number
888.3060
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TDM Anterior Cervical Plate System is intended for anterior fixation of the cervical spine (C2 to C7). The system is to be used to provide stabilization of the anterior cervical spine as an adjunct to fusion for the treatment ve disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Device Description

The TDM Anterior Cervical Plate System consists of cervical plates, screws and the instruments necessary to implant this specific system. All implant components are made from a titanium alloy (Ti-6AL-4V). The fixation construct consists of a cervical plate that is attached to the vertebral body of the cervical spine with self-tapping and self-drilling bone screws using an anterior approach. Cervical Plate System is provided as non-sterile. The implants are intended for single-use only, while the instruments are reusable.

AI/ML Overview

I am sorry, but the provided text does not contain information about acceptance criteria and a study that proves a device meets those criteria.

The document is a 510(k) premarket notification clearance letter from the FDA for the "TDM Anterior Cervical Plate System." It confirms that the device is substantially equivalent to a legally marketed predicate device.

The section titled "Performance Data" mentions:

  • Non-clinical testing: static compression, dynamic compression, and static torsion.
  • Standard: ASTM F1717.
  • Purpose: to demonstrate substantial equivalence to the predicate device.

However, it does not provide specific acceptance criteria values for these tests, nor does it list detailed results or a study report with sample sizes, data provenance, ground truth establishment, or any of the other specific information requested in your prompt regarding acceptance criteria and a study proving their fulfillment. The document focuses on regulatory clearance based on substantial equivalence, not on a detailed performance study with explicit acceptance criteria.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

April 8, 2024

TDM Co., Ltd. % Jeena Mathai President Eerkie Corporation 4027 Runnymeade Dr Collegeville, Pennsylvania 19426

Re: K240423

Trade/Device Name: TDM Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: February 9, 2024 Received: February 13, 2024

Dear Jeena Mathai:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin
O'neill -S FDA

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K240423

Device Name TDM Anterior Cervical Plate System

Indications for Use (Describe)

The TDM Anterior Cervical Plate System is intended for anterior fixation of the cervical spine (C2 to C7). The system is to be used to provide stabilization of the anterior cervical spine as an adjunct to fusion for the treatment ve disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

TDM's TDM Anterior Cervical Plate System

Sponsor:TDM Co. Ltd.
Manufacturer69, Cheomdan Venture So-ro, 37 beon-gil, Buk-gu Gwangju, 61003, Republic of Korea
Official ContactJungwook Choi
Phone:+82-62-971-7460
Fax:+82-62-971-7461
Contact PersonJeena Mathai
mgsharemg@gmail.com
Date:March 30, 2024
Device Name:TDM Anterior Cervical Plate System
Common Name:Anterior Cervical Plate System
Classification Name:Spinal Intervertebral Body Fixation Orthosis
Classification Number:21 CFR 888.3060
Product Code/Classification:KWQ, class II
Description:The TDM Anterior Cervical Plate System consists of cervical plates, screws and the instruments necessary to implant this specific system. All implant components are made from a titanium alloy (Ti-6AL-4V). The fixation construct consists of a cervical plate that is attached to the vertebral body of the cervical spine with self-tapping and self-drilling bone screws using an anterior approach. Cervical Plate System is provided as non-sterile. The implants are intended for single-use only, while the instruments are reusable.
Indications For Use:The TDM Anterior Cervical Plate System is intended for anterior fixation of the cervical spine (C2 to C7). The system is to be used to provide stabilization of the anterior cervical spine as an adjunct to fusion for the treatment of degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or

{4}------------------------------------------------

curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor,pseudoarthrosis, and failed previous fusion.
Performance Data:Non-clinical testing was performed to demonstrate that thesubject TDM Anterior Cervical Plate System is substantiallyequivalent to the predicate device. The following testing wasperformed in accordance with ASTM F1717:- Static compression- Dynamic compression- Static Torsion
Primary PredicateDevice:Synthes Spine - Synthes Anterior CSLP System (K000742)
Reference Devices:K171808 & K190391 - TDM Plate and Screw System
TechnologicalCharacteristics:The TDM Anterior Cervical Plate System has substantially equivalenttechnological characteristics to its predicate and reference devicesthrough comparison in areas including design, labeling/intendeduse, material composition, function, range of sizes, and packaging.
Conclusion:The TDM Anterior Cervical Plate System has been demonstratedto be substantially equivalent to the predicate system(s) withrespect to technical characteristics, performance, and intendeduse. The information provided within this premarket notificationsupports substantial equivalence of the subject device to thepredicate device(s).

N/A