K Number
K240095
Date Cleared
2024-10-09

(271 days)

Product Code
Regulation Number
886.5928
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

"CAREUS Contact Lens Case" is a device intended for the storage of soft (hydrophilic), rigid gas permeable, and/or hard contact lenses. Used for storage during chemical disinfection only. Not for use during heat disinfection.

Device Description

"CAREUS Contact Lens Case" is a medical device for the storage of soft (hydrophilic) and rigid gas permeable contact lenses. The applicant device of Contact Lens Case consists of two parts: case body and case lid. The case body is based with adjoining dual wells for the containment of fluid, and the two lids are designed for screwing. "CAREUS Contact Lens Cases" have six different models, differ in dimension, case color, and packaging. The exterior of the cases total have 9 colors: white, pink, yellow, green, light green, red, purple, blue, and transparent (no color). Contact lenses can be fully immersed into the well from both models accommodates all lenses currently being sold in the market. The bottom of each well is marked with L (left) or R (right). CAREUS Contact Lens Cases are made of polypropylene.

AI/ML Overview

The provided document is a 510(k) Summary for the "CAREUS Contact Lens Case." It describes a medical device, not a device that relies on algorithms for its function. Therefore, most of the requested information regarding acceptance criteria, study details, and AI/algorithm-related aspects is not applicable to this submission.

Here is the information that can be extracted from the document:

1. A table of acceptance criteria and the reported device performance:

The document broadly states that "All tests were verified to meet acceptance criteria" for bench testing and biocompatibility. However, it does not provide specific numerical acceptance criteria or detailed reported performance figures for the physical and chemical specifications.

Acceptance Criteria CategoryReported Device PerformanceStudy Type
Bench TestingMet physical and chemical specifications.Internal procedures
BiocompatibilityDemonstrated biocompatibility.ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Irritation & Sensitization), ISO 10993-11 (Systematic Toxicity)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

The document does not specify the sample sizes for the bench testing or biocompatibility tests. It only states that testing was performed. The manufacturing country of origin for the device is China (Shanghai Care Us Medical Product Co., Ltd., Shanghai, China). The testing itself is not explicitly stated as retrospective or prospective, but bench and biocompatibility testing are generally prospective in nature for a new device submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

Not applicable. This device is a physical contact lens case and does not involve expert interpretation or ground truth establishment in the context of diagnostic accuracy, as would be relevant for an AI/algorithm-based device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable. This concept applies to studies where expert interpretation is being evaluated, which is not the case for a contact lens storage device.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device does not involve human readers or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This device does not have an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

For the bench testing, the "ground truth" would be the predetermined physical and chemical specifications. For biocompatibility, the "ground truth" is adherence to the standards defined by ISO 10993. These do not fall into the categories of expert consensus, pathology, or outcomes data in the usual sense.

8. The sample size for the training set:

Not applicable. This device is not an algorithm and does not have a training set.

9. How the ground truth for the training set was established:

Not applicable. This device is not an algorithm and does not have a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font below that.

October 09, 2024

Shanghai Care Us Medical Product Co., Ltd. Charles Shen Director Manton Business and Technology Services 37 Winding Ridge Oakland, NJ 07436

Re: K240095

Trade/Device Name: CAREUS Contact Lens Case Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (hydrophilic) contact lens care products Regulatory Class: Class II Product Code: LRX Dated: September 03, 2024 Received: September 03, 2024

Dear Charles Shen:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

J Angelo Green -S

J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K240095

Device Name CAREUS Contact Lens Case

Indications for Use (Describe)

"CAREUS Contact Lens Case" is a device intended for the storage of soft (hydrophilic), rigid gas permeable, and/or hard contact lenses. Used for storage during chemical disinfection only. Not for use during heat disinfection.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary: K240095

This summary of 510k safety and effectiveness information is being submitted In accordance with the requirements of 21CFR 807.92

1 Submitter & Foreign Manufacture Identification

Shanghai Care Us Medical Product Co., Ltd. No.2, Lane 10, West Chenchuan Road Baoshan District Shanghai CN-31 Shanghai 200949 China Tel: 86- 15026598135 Submitter's FDA Registration Number: N/A

2 US Agent and Contact Person

Image /page/4/Picture/5 description: The image shows a logo with a stylized letter 'M' in blue, overlaid with a crescent shape in light blue. Above the crescent, there are orange Chinese characters. The 'M' is textured with small dots, giving it a slightly grainy appearance. The overall design is simple and modern, likely representing a company or organization.

Charles Shen Manton Business and Technology Services 37 Winding Ridge, Oakland, NJ 07436 Tel: 608-217-9358 Email: cyshen@aol.com

3 Date of Summary: October 9, 2024

4 Device Name:

Proprietary Name:"CAREUS Contact Lens Case"
Common Name:Contact Lens Case
Classification Name:Soft (Hydrophilic) Contact Lens Care Products
Device Classification:2
Regulation Number:21 CFR 886.5928
Panel: GeneralOphthalmic
Product Code:LRX

5 Predicate Device Information:

K120904, "The Polaris Dial-a-Date Contact Lens Case", manufactured by "Reliance Design & Manufacture Corp. " in Tainan, Taiwan

6 Device description:

"CAREUS Contact Lens Case" is a medical device for the storage of soft (hydrophilic) and rigid gas permeable contact lenses. The applicant device of Contact Lens Case consists of two parts: case body and case lid. The case body is based with adjoining dual wells for the containment of fluid, and the two lids are designed for screwing.

{5}------------------------------------------------

"CAREUS Contact Lens Cases" have six different models, differ in dimension, case color, and packaging.

The exterior of the cases total have 9 colors: white, pink, yellow, green, light green, red, purple, blue, and transparent (no color).

Contact lenses can be fully immersed into the well from both models accommodates all lenses currently being sold in the market. The bottom of each well is marked with L (left) or R (right).

CAREUS Contact Lens Cases are made of polypropylene.

7 Indications for Use:

"CAREUS Contact Lens Case" is a device intended for the storage of soft (hydrophilic), rigid gas permeable, and/or hard contact lenses. Used for storage during chemical disinfection only. Not for use during heat disinfection.

Technological Comparison with Predicate Device 8

"CAREUS Contact Lens Case" is compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance.

  • K120904. "The Polaris Dial-a-Date Contact Lens Case", manufactured by (1) "Reliance Design & Manufacture Corp." in Tainan, Taiwan
    The following table shows similarities and differences of use, design, and material between our device and the predicate devices.
DescriptionOur DevicePredicate Device (K120904)SE Comparison
Indicationfor Use"CAREUS Contact Lens Case"is a device intended for thestorage of soft (hydrophilic),rigid gas permeable, and/or hardcontact lenses. Used for storageduring chemical disinfectiononly. Not for use during heatdisinfection.The Polaris Dial-a-DateContact Lens Case isintended for the storage ofsoft (hydrophilic), rigid gaspermeable, and/or hardcontact lenses. Used forstorage during chemicaldisinfection only. Not foruse during heat disinfection.SE
Basic DesignTwo adjoining wells with screwdown lidsTwo adjoining wells withscrew down lidsSE
MaterialsPolypropylenePolypropyleneSE
SizeVarious68 x 32 x 21.5 mm (length xwidth x height)Similar

Table 5.1: Comparison of Intended Use, Design, and Material

{6}------------------------------------------------

Volume3.8-4,5 mL4.4 mLSimilar
ColorsNine different colorsFour different colorsSimilar

Our devices are the same with the predicate device in indications for use and basic design, similar to predicate devices in geometry and dimension. Our devices and the predicate device have some minor differences in dimension and color, which do not affect the safety and performance of the devices.

9 Summary of Non-Clinical Testing:

Bench testing was performed per internal procedures to ensure that the "CAREUS Contact Lens Case" met its physical and chemical specifications. All tests were verified to meet acceptance criteria.

Biocompatibility testing was performed following current version of ISO 10993-5 and ISO 10993-10:

Cytotoxicity (ISO 10993-5) Irritation & Sensitization (ISO 10993-10) Systematic Toxicity (ISO 10993-11)

10 Summary of Clinical Study:

Clinical Study is not performed for this device.

11 Substantial Equivalence Conclusion

It has been shown in this 510(k) submission that "CAREUS Contact Lens Case" and its predicate devices have similar indications for use, similar composition, and biocompatibility, similar manufacturing process, and similar performance.

The difference between the "CAREUS Contact Lens Case" and their predicate device do not raise any question regarding its equivalence.

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, the subject device is respectively substantially equivalent to the predicate device.

§ 886.5928 Soft (hydrophilic) contact lens care products.

(a)
Identification. A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.(b)
Classification. Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”