K Number
K233995
Date Cleared
2024-02-23

(67 days)

Product Code
Regulation Number
872.3760
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Complete edentulism Areas of application: Denture bases in removable complete denture prosthetics

Device Description

Ivotion® Base Print is a light-curing resin material for stereolithography 3D printing. It is suitable for the fabrication of permanent denture bases in removable complete denture prosthetics. The material can be processed with approved 3D printing system. The dental professional starts with preparation of the printing procedure, cleaning, post-curing, finishing and polishing. A denture fabricated with Ivotion Base Print can be repaired and relined using a conventional self-curing denture base material.

AI/ML Overview

The provided text is a 510(k) summary for the medical device "Ivotion Base Print." It describes the device, its intended use, and its comparison to a predicate device to establish substantial equivalence.

Based on the information provided, this document does not describe a study involving an AI/ML powered device and therefore does not contain the requested information regarding acceptance criteria, study design for AI/ML performance, ground truth establishment, or human-in-the-loop studies.

The device "Ivotion Base Print" is a light-curing resin material for stereolithography 3D printing. The studies mentioned are primarily non-clinical performance testing to assess physical properties and biocompatibility testing according to ISO standards, which are standard for dental materials, not AI/ML algorithms.

Therefore, I cannot extract the information requested in your prompt because the provided text is not about an AI/ML powered medical device and the associated performance studies you are asking for.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 23, 2024

Ivoclar Vivadent, Inc Anderjeet Gulati Sr Manager of QA & Regulatory Affairs 175 Pineview Drive Amherst, New York 14228

Re: K233995

Trade/Device Name: Ivotion Base Print Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI Dated: December 5, 2023 Received: December 18, 2023

Dear Anderjeet Gulati:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}------------------------------------------------

Sincerely,

Michael E. Adjodha -S

Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

Submission Number (if known)

K233995

Device Name

Ivotion Base Print

Indications for Use (Describe)

Complete edentulism

Areas of application: Denture bases in removable complete denture prosthetics

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K233995 510(K) SUMMARY

1

Contact:Anderjeet Gulati, Sr. Manager QA & Regulatory AffairsIvoclar Vivadent, Inc.175 Pineview DriveAmherst, New York 14228716-264-2046Anderjeet.gulati@ivoclar.com
Company:Ivoclar Vivadent, AGBendererstrasse 2, Schaan, FL-9494, Liechtenstein+423-235-3535
Date Prepared:February 5, 2024
Proprietary Name:Ivotion® Base Print
Classification Name:Resin, Denture, Relining, Repairing, Rebasing (872.3760)(Classification Code EBI)
Common Name:Denture Material
Predicate Device:FotoDent denture (K200580) by Dreve Dentamid GmbH, Germany
Reference Device:NexDent Denture/E-Denture (K162572) by Vertex-Dental BV

Device Description: Ivotion® Base Print is a light-curing resin material for stereolithography 3D printing. It is suitable for the fabrication of permanent denture bases in removable complete denture prosthetics. The material can be processed with approved 3D printing system. The dental professional starts with preparation of the printing procedure, cleaning, post-curing, finishing and polishing. A denture fabricated with Ivotion Base Print can be repaired and relined using a conventional self-curing denture base material.

Indications for Use:

Complete edentulism

Areas of application:

Denture Bases in removable complete denture prosthetics

Comparison to Predicate: The primary predicate device to which Ivotion Base Print has been compared is Dreve Dentamid GmbH- FotoDent denture (K200580).

{5}------------------------------------------------

510(K) SUMMARY

ivoclar

Predicate Device:Proposed Device:Deviation
DeviceDreve Dentamid GmbH: FotoDent denture (K200580)Ivoclar Vivadent AG: Ivotion Base PrintYesNo
Indications for UseIndications for Use:FotoDent denture is a light curing resin intended for manufacturing of full and partial removable denturesIndications for Use:Complete edentulismAreas of application:Denture bases in removable complete denture prosthetics
Precaution Measures/Contraindications/ Processing restrictions/Side effectsContraindications:FotoDent denture is contraindicated for individuals with known allergy to methacrylates and acrylates.Contraindications:The use of the product is contraindicated if the patient is known to be allergic to any of its ingredients.
Summary of Indications, Precaution Measures/Contraindications/ Processing restrictions/Side effectsThe new device is indicated for removable complete dentures in contrast to the predicate, which is used for full and partial removable dentures.The Contraindications are the same.
Working PrincipleDevice description: FotoDent denture is a light curable resin for 3D printing of full and partial dentures.Technological characteristics:FotoDent denture is a light-curing resin for 3D printingIvotion Base Print is a light-curing resin material for stereolithography 3D printing. It is suitable for the fabrication of permanent denture bases in removable complete denture prosthetics.The material can be processed with approved 3D printing systems.
Summary of Working PrincipleThe working principles for the predicate and new device are basically the same. Both devices are used to 3D print removable dentures.
Delivery forms/dosageSupplied in liquid form, 1 kgBottle of liquid, 1000ml
Summary of Delivery forms/dosageThe new device is available in less shades and translucencies compared to the predicate device. The new device is available with a color gradient ("Multi") which is a new feature compared to the predicate.
Storage Conditions2 years at 18-28 °C24 months at 2-28 °C / 36-82 °F
Summary of Storage ConditionsNo difference in shelf life. Storage conditions differ between the 2 devices.
Principles of OperationProcessing:- Preparation for printing procedure- Cleaning- Post curing- Temper (Finish)Step-by-step:- Preparation of the printing procedure- Cleaning- Post-curing- Finishing- Polishing- Repair/Relining
Summary Principles of operationThe principles of operation are basically the same. Both devices are used to 3D print dentures (full or partial).

{6}------------------------------------------------

510(K) SUMMARY

Summary ofChemicalCompositionBiocompatibility has been thoroughly assessed for Ivotion Base Print, see BiocompatibilityAssessment for Ivotion Base Print.
Finished DeviceSpecificationISO 20795-1:2013 – Dentistry – Basepolymers - Part 1: Denture basepolymersISO 20795-1:2013 – Dentistry – Basepolymers - Part 1 Denture basepolymers
DeviceSpecificationPerformanceValueUnitValueUnit
SpecificationFlexural strength> 80 (predicatedevice)MPa≥ 65MPa
Flexuralmodulus> 2000 (predicatedevice)MPa≥ 2000MPa
WaterabsorptionNo values availablefor predicate orreference device---≤ 32µg/mm3
Water solubiltyReference device:water sorption meetstherequirement of Type 2materials acc to ISO20795-1.---≤ 8.0µg/mm3
MMA residualmonomercontentNo values availablefor predicate orreference device---≤ 2.2%
Summary ofFinished DeviceSpecificationCurrently, there is no ISO standard that defines the physical requirements for denture basematerials processed with 3D printing systems. The product complies with the requirements ofISO 20795-1 for photo-polymerizable denture base resins.The water solubility value fulfils the specification defined in ISO 20795-1 for self-curingmaterials.The FDA Guidance is fulfilled by Ivotion Base Print.
SterilizationNot applicable. No sterilizationrecommendation.Not applicable. No sterilizationrecommendation.
Single useConsumable materialConsumable material
Summary ofPerformanceSpecificationThe performance is compliant with the requirements of ISO 20795-1 and substantiallyequivalent for both devices.

Substantial Equivalence to the predicate:

The working principle is the same for both devices. There is no detailed information regarding the chemical composition available for the predicate device. Therefore, detailed biocompatibility testing has been assessed for Ivotion Base Print. The device performance is considered to be equivalent for the new and the predicate device.

Therefore, Ivotion Base Print is substantially equivalent to the predicate device.

Differences:

Indication differs in that it has been reduced from full and partial for the predicate to only be for complete dentures for Ivotion Base Print.

The temperature range for storage conditions differ. The storage conditions are a result of the storage stability testing performed.

{7}------------------------------------------------

510(K) SUMMARY

1

There are some deviations in the principles of operation, but they are basically the same. The new device includes the additional steps of finishing, polishing and repair/relining in the principles of operation.

There is no detailed composition available for the predicate device, but it is known that FotoDent denture is a light-curing methacrylate-based material for denture fabrication. The ingredients that have not been used before in Ivoclar Vivadent AG's products have a CAS number identified and biocompatibility has been thoroughly assessed for Ivotion base Print.

There is a difference between the flexural strength values of the predicate and the new device. but the other device specifications differences were a result of no values available for the predicate. The device specifications-flexural strength, flexural modulus, water absorption and MMA residual monomer content are compliant to the requirements defined in ISO 20795-1 Type 4 (light activated).

Non-clinical performance testing:

Bench testing was performed to test the physical properties included in the Finished Device Specification for the subject device including: flexural modulus, water absorption, water solubility, and MMA residual monomer content according to EN ISO 20795-1:2013- Dentistry- Base polymers- Part 1 Denture base polymers (ISO 207951:2013).

Biocompatibility:

The subject device was also evaluated for Biocompatibility according to ISO 10993. ISO 7405. ISO 21726:2019 and ISO 14971:2019. The biological evaluation was performed based on toxicological data on relevant component materials / compounds, information on prior use of relevant component materials / compounds, data from biological tests, data on the history of clinical use or human exposure. Based on the information included in the biological evaluation report it can be concluded that the product under evaluation is acceptable in relation to its clinical benefit. The product under assessment does not release leachable substances in cytotoxic concentrations; The product is considered to have no sensitizing properties; the product is not irritative to the oral mucosa; The product does not contain biologically derived materials. Consequently, no material mediated pyrogenicity is expected; The product shows no acute systemic toxic characteristics; The risk for sub-acute, sub-chronic systemic toxicity induced by the product under assessment is acceptable; the finished product is considered to be non-mutagenic; The novel raw materials contained in the product are considered to be nonmutagenic; and the carcinogenic risk induced by the finished product is negligible.

It can be concluded that the product does not represent a toxicological risk for the patient and the user.

Based on the toxicological evaluation of the product and the longstanding worldwide clinical use of similar materials it can be concluded that the benefits provided by the final product will exceed any potential risks produced by device materials providing that instructions for use have been followed.

{8}------------------------------------------------

Conclusion:

The new device Ivotion Base Print and the predicate device FotoDent Denture are 3D printable materials to fabricate denture bases for full (new device) or partial removable dentures (predicate device). The working principle is the same for both devices. There is no detailed information reqarding the chemical composition available for the predicate device. Therefore, detailed biocompatibility testing has been assessed for Ivotion Base Print. The device performance is considered to be equivalent for the new and the predicate device.

Therefore, Ivotion Base Print is substantially equivalent to the predicate device.

510K Summary -5

§ 872.3760 Denture relining, repairing, or rebasing resin.

(a)
Identification. A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.(b)
Classification. Class II.