K Number
K233967

Validate with FDA (Live)

Manufacturer
Date Cleared
2024-05-09

(146 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DePuy Synthes MatrixSTERNUM Fixation System is indicated for use in adults with normal bone quality.

MatrixSTERNUM Sternal Body Plates are indicated for internal fixation of bone discontinuities following sternotomy.

MatrixSTERNUM Thoracotomy Plates are indicated for internal fixation of bone and/or cartilage discontinuities following thoracotomy.

Device Description

The DePuy Synthes MatrixSTERNUM Fixation System consists of two families of plates, namely the MatrixSTERNUM Sternal Body Plates and MatrixSTERNUM Thoracotomy Plates, and two families of screws, namely the MatrixSTERNUM Self-Drilling Locking and Non-Locking Screws.

The DePuy Synthes MatrixSTERNUM Fixation System is intended for the stabilization and fixation of bones in the anterior chest wall. The subject plates are available in different shapes and sizes and are made from Titanium Alloy or commercially pure Titanium. The system also consists of non-implantable dedicated use screw guides, a screw guide handle, trays and modules, as well as general use instruments to be used as accessories with the subject implants.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the DePuy Synthes MatrixSTERNUM Fixation System, which is a metallic bone fixation appliance.

It is crucial to understand that this document describes a submission for a medical device (hardware), not an Artificial Intelligence (AI) or software-as-a-medical-device (SaMD) product.

Therefore, the information regarding acceptance criteria and studies that prove the device meets acceptance criteria are focused on the mechanical, material, and biological performance of physical implants, not on the performance of an AI algorithm in tasks like image interpretation or diagnosis.

Because the request is specifically about AI/SaMD performance criteria (e.g., sample size for test set, expert ground truth establishment, MRMC studies, standalone performance, training set data), the provided text does not contain the information needed to answer these questions for an AI/SaMD product.

The document states:

  • "Clinical testing was not necessary for the determination of substantial equivalence." This directly indicates that no clinical performance study, multi-reader multi-case (MRMC) study, or standalone algorithm performance study (as would be typical for AI/SaMD) was conducted.
  • The performance data provided is under "Non-clinical Performance Data" and includes:
    • Mechanical performance evaluation and testing: Comparing the subject plates to predicate devices.
    • Magnetic resonance compatibility assessment.
    • Endotoxin testing.
    • Biocompatibility evaluation and testing (in accordance with ISO 10993-1).

These are standard non-clinical tests for orthopedic implants to demonstrate safety and effectiveness and substantial equivalence to existing devices. They are not metrics or study designs relevant to AI/SaMD performance.

Therefore, I cannot populate the requested table and answer the subsequent questions in the context of an AI/SaMD product, as the provided text pertains to a physical medical device.

{0}------------------------------------------------

May 9, 2024

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized eagle and text encircling it. To the right is the FDA logo, with the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

Synthes GmbH Georgina Mueller, PhD Sr. Regulatory Affairs Program Lead Luzernstrasse 21 Zuchwil. SO 4528 Switzerland

Re: K233967

Trade/Device Name: DePuy Synthes MatrixSTERNUM Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: April 9, 2024 Received: April 9, 2024

Dear Dr. Mueller:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali -S

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K233967

Device Name

DePuy Synthes MatrixSTERNUM Fixation System

Indications for Use (Describe)

The DePuy Synthes MatrixSTERNUM Fixation System is indicated for use in adults with normal bone quality.

MatrixSTERNUM Sternal Body Plates are indicated for internal fixation of bone discontinuities following sternotomy.

MatrixSTERNUM Thoracotomy Plates are indicated for internal fixation of bone and/or cartilage discontinuities following thoracotomy.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 1 of 1

{3}------------------------------------------------

510(k) Summary

SponsorSynthes GmbHGeorgina MuellerLuzernstasse 214528 Zuchwil, SwitzerlandPhone: +41 61 965 63 14
Date PreparedMay 3, 2024
Proprietary NameDePuy Synthes MatrixSTERNUM Fixation System
ClassificationNameSingle/multiple component metallic bone fixation appliances and accessoriesSmooth or threaded metallic bone fixation fastener
ClassificationPlatesClass IIRegulation Number: 21 CFR 888.3030 (Primary)Product Code: HRS
ClassificationScrewsClass IIRegulation Number: 21 CFR 888.3040Product Code: HWC
Predicate /ReferencedevicesPredicate for MatrixSTERNUM Sternal Body Plates and MatrixSTERNUM Screws:Biomet Microfixation Sternal Closure System (SternaLock Blu Primary ClosureSystem), K161896
Predicate /Referencedevices
Predicate /Referencedevices
DeviceDescriptionThe DePuy Synthes MatrixSTERNUM Fixation System consists of two families ofplates, namely the MatrixSTERNUM Sternal Body Plates and MatrixSTERNUMThoracotomy Plates, and two families of screws, namely the MatrixSTERNUMSelf-Drilling Locking and Non-Locking Screws.
The DePuy Synthes MatrixSTERNUM Fixation System is intended for thestabilization and fixation of bones in the anterior chest wall. The subject plates areavailable in different shapes and sizes and are made from Titanium Alloy orcommercially pure Titanium. The system also consists of non-implantablededicated use screw guides, a screw guide handle, trays and modules, as well asgeneral use instruments to be used as accessories with the subject implants.
Indications forUseThe DePuy Synthes MatrixSTERNUM Fixation System is indicated for use inadults with normal bone quality.MatrixSTERNUM Sternal Body Plates are indicated for internal fixation of bonediscontinuities following sternotomy.MatrixSTERNUM Thoracotomy Plates are indicated for internal fixation of boneand/or cartilage discontinuities following thoracotomy.
TechnologicalCharacteristicsThe design, features, and specifications of the subject and predicate devices arecompared below.MatrixSTERNUM Sternal Body Plates The range of plate shapes, sizes, and thicknesses of the subject devices are similar to those offered in the predicate devices. The materials of both the subject and predicate devices are the same with plates offered in commercially pure titanium. Subject and predicate plates and screws feature Locking Technology. The subject plates include locking holes that mate with locking screw heads, whereas the predicate plates have locking thread geometry cut into the plates by the screws during insertion. The subject and predicate plates have varying screw holes; in number, type and diameter. The subject plate screw holes accommodate a 2.7mm diameter screw whereas the predicate plate screw holes accommodate both 2.4mm diameter and 2.7mm diameter screws. MatrixSTERNUM Thoracotomy Plates The plate screw hole geometry and thickness of the subject devices are similar to those offered in the predicate devices. The materials of both subject and predicate devices are the same with plates offered in titanium alloy. Both subject and predicate devices have locking holes that mate with
Non-clinicalPerformanceDataAdditional features are found in the subject plates that are not found in the predicate plates. Subject plates include a straight bar region to span the bone discontinuity. Subject T-plates include a perpendicular region to span a bone discontinuity from sternum to rib. MatrixSTERNUM Self-Drilling Locking and Non-Locking Screws The subject locking and non-locking self-drilling screws are of the same geometry and offered in the same lengths as the predicate locking and non-locking self-drilling screws. The materials of both subject and predicate devices are the same with screws offered in titanium alloy. The anodization color varies between subject and predicate. Subject and predicate locking screws have the same locking head geometry to mate with the locking plate threads. Subject and predicate non-locking screws have the same head geometry to lag and temporarily fixate the plate to bone. Mechanical performance evaluation and testing of constructs has been performed to compare: The subject MatrixSTERNUM Sternal Body Plates to the predicate Biomet Microfixation Sternal Closure System; and The subject MatrixSTERNUM Thoracotomy Plates to the predicate DePuy Synthes MatrixRIB Fixation System. This data supports that the subject devices do not raise any new issues of safety and effectiveness in terms of mechanical performance compared to the respective predicate devices.A magnetic resonance compatibility assessment has been performed to establish MR Conditional parameters for the subject DePuy Synthes MatrixSTERNUM Fixation System Plates and Screws.Endotoxin testing has been performed according to the LAL test method to establish that the sterile subject DePuy Synthes MatrixSTERNUM Fixation System Implants (Plates and Screws) meet the specified endotoxin requirement of 20EU/device.
Biocompatibility evaluation and testing has been performed in accordance withISO 10993-1 and it is concluded that the subject plates, screws, screw guides,
screw guide handle, trays and modules part of the DePuy Synthes
MatrixSTERNUM Fixation System are biologically safe when used as intended.
ClinicalPerformanceDataClinical testing was not necessary for the determination of substantialequivalence.
MatrixSTERNUM Sternal Body Plates
The subject and the predicate devices share the same intended use, which is forthe stabilization and fixation of the anterior chest wall, and are both intended tobe used by surgeons within a health care facility.The subject device indications are a subset of the predicate Biomet MicrofixationSternal Closure System (K161896) indications.
MatrixSTERNUM Thoracotomy Plates
SubstantialEquivalenceThe subject devices' intended use is the stabilization and fixation of the anteriorchest wall, whereas the predicate's intended use is the stabilization and fixationof the chest wall. The subject devices' intended use, therefore, falls within thesame intended use as the predicates' intended use, as it is a subset thereof.Both subject and predicate devices are intended to be used by surgeons within ahealth care facility.The subject device indications are a subset of the predicate DePuy SynthesMatrixRIB Fixation System (K161590 / K190409) indications.
The non-clinical performance data as well as the comparison of design featuresincluded in this premarket notification demonstrate that any differences intechnological characteristics of the subject devices compared to the predicatedevices do not raise any new issues of safety and effectiveness.
It is concluded that the information provided herein supports substantialequivalence of the subject devices.

{4}------------------------------------------------

K233967

{5}------------------------------------------------

{6}------------------------------------------------

K233967

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.