(58 days)
This Product is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.
RadiForce MX317W is a color LCD monitor for viewing medical images other than those of mammography. The color panel employs in-plane switching (IPS) technology allowing wide viewing angles and the matrix size (or resolution) is 4,096 x 2,160 pixels (8MP) with a pixel pitch of 0.1674 mm.
Since factory calibrated display modes, each of which is characterized by a specific tone curve (including DICOM GSDF), a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor, the tone curve is e.g. DICOM compliant regardless of the display controller used. This helps ensure tone curves even if a display controller or workstation must be replaced or serviced.
RadiCS is application software to be installed in each workstation offering worry-free quality control of diagnostic monitors including the RadiForce MX317W based on the OC standards and guidelines and is capable of quantitative tests and visual tests defined by them. The RadiCS is included in this 510(k) submission as an accessory to the RadiForce MX317W.
RadiCS is Basic Documentation level and that it's being used unchanged from the predicate software. RadiCS supports the functions of the monitor RadiForce MX317W and it's not a medical imaging software.
The provided text describes the RadiForce MX317W, a color LCD monitor for displaying radiological images. It is a Class II device (Product Code: PGY).
Here's an analysis of the acceptance criteria and the study performed, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not provide a specific table of acceptance criteria with numerical targets and the device's numerical performance against those targets. Instead, it states that the device's characteristics "meet the pre-defined criteria when criteria are set" and that the test results showed "display characteristics equivalent to those of the predicate device, RadiForce MX315W."
However, it lists the types of performance tests conducted, which implies general acceptance criteria for these characteristics:
| Acceptance Criterion (Implied) | Reported Device Performance |
|---|---|
| Spatial resolution (Modulation Transfer Function - MTF) | Met pre-defined criteria; equivalent to predicate device. |
| Maximum number of pixel defects/faults | Met pre-defined criteria; equivalent to predicate device. |
| Absence of miscellaneous artifacts (based on TG18 guideline) | Met pre-defined criteria; equivalent to predicate device. |
| Temporal response | Met pre-defined criteria; equivalent to predicate device. |
| Luminance | Met pre-defined criteria; equivalent to predicate device. |
| Conformance to DICOM GSDF (based on TG18 guideline) | Met pre-defined criteria; equivalent to predicate device. |
| Color tracking | Met pre-defined criteria; equivalent to predicate device. |
2. Sample size used for the test set and the data provenance
The document describes bench testsperformed on the RadiForce MX317W. It does not mention a "test set" in the context of clinical data, images, or patient cases. Therefore, information about sample size for a test set and data provenance (country of origin, retrospective/prospective) is not applicable to this type of device and study. The testing was performed on the physical monitor itself.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable. The ground truth for the performance of a display device is established through technical measurements using specialized equipment and adherence to industry standards/guidelines (e.g., AAPM TG18), rather than expert consensus on medical images.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable as there was no clinical test set involving human interpretation of images. The performance was assessed through objective physical measurements.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This type of study is relevant for AI-powered diagnostic software, not for a display device like the RadiForce MX317W. The document explicitly states: "No animal or clinical testing was performed on the RadiForce MX317W."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable. The RadiForce MX317W is a display device, not an algorithm or AI system. Its performance is assessed as a standalone hardware unit for displaying images, not as an algorithm's standalone performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth for the display characteristics (e.g., MTF, luminance, DICOM GSDF conformance) was established based on technical standards and guidelines, specifically "Guidance for Industry and FDA Staff: Display Devices for Diagnostic Radiology" issued on September 28, 2022, and the "Assessment of Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 guideline)". This involves objective physical measurements against defined technical specifications.
8. The sample size for the training set
This information is not applicable. The RadiForce MX317W is a hardware display device and does not involve machine learning algorithms that require a "training set."
9. How the ground truth for the training set was established
This information is not applicable, as there was no training set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the agency name in a sans-serif font.
EIZO Corporation Hiroaki Hashimoto Senior Manager of Regulatory Compliance and Safety Department 153 Shimokashiwano Hakusan, Ishikawa, 924-8566 Japan
Re: K233465
December 21, 2023
Trade/Device Name: RadiForce MX317W Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: PGY Dated: October 24, 2023 Received: October 24, 2023
Dear Hiroaki Hashimoto:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb
Jessica Lamb Assistant Director Imaging Software Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
Device Name
RadiForce MX317W
Indications for Use (Describe)
This Product is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for EIZO. On the left side of the logo is a gray, geometric shape that resembles a flag or banner. The right side of the logo shows the company name, "EIZO", in a bold, sans-serif font. There is a registered trademark symbol next to the "O" in "EIZO".
EIZO Corporation, 153 Shimokashiwano, Hakusan, Ishikawa
U.S. Food and Drug Administration Center for Devices and Radiological Health Document Mail Center - WO66-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002
| Name | Hiroaki Hashimoto |
|---|---|
| Department | Regulatory Compliance and Safety |
| Telephone | +81 (76) 274-2468 |
| hiroaki.hashimoto@eizo.com |
510(k) Summary
1. Submitter
EIZO Corporation 153 Shimokashiwano, Hakusan, Ishikawa 924-8566 Japan Phone: +81 (76) 274-2468 Contact Person: Hiroaki Hashimoto Date of Prepared: November 17th, 2023
2. Device
- Name of Device: RadiForce MX317W
- Common or Usual Name: 30.5 inch (77.5 cm) Color LCD Monitor
- Classification Name: Display, Diagnostic Radiology
- (21 CFR 892.2050)
- Regulatory Class: II
- . Product Code: PGY
3. Predicate Device
- Name of Device: RadiForce MX315W (K170949) ●
- Common or Usual Name: 79.0 cm (31.1 inch) class Color LCD Monitor
- Classification Name: Display, Diagnostic Radiology ●
- (21 CFR 892.2050) Regulatory Class: ● II
- Product Code: PGY
- Name of Device: RadiForce RX240 (K113844) ●
- Common or Usual Name: 2MP Color LCD Monitor
- Classification Name: System, Image Processing, Radiological (21 CFR 892.2050) ● Regulatory Class: II
- Product Code: LLZ
{4}------------------------------------------------
4. Device Description
RadiForce MX317W is a color LCD monitor for viewing medical images other than those of mammography. The color panel employs in-plane switching (IPS) technology allowing wide viewing angles and the matrix size (or resolution) is 4,096 x 2,160 pixels (8MP) with a pixel pitch of 0.1674 mm.
Since factory calibrated display modes, each of which is characterized by a specific tone curve (including DICOM GSDF), a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor, the tone curve is e.g. DICOM compliant regardless of the display controller used. This helps ensure tone curves even if a display controller or workstation must be replaced or serviced.
RadiCS is application software to be installed in each workstation offering worry-free quality control of diagnostic monitors including the RadiForce MX317W based on the OC standards and guidelines and is capable of quantitative tests and visual tests defined by them. The RadiCS is included in this 510(k) submission as an accessory to the RadiForce MX317W.
RadiCS is Basic Documentation level and that it's being used unchanged from the predicate software. RadiCS supports the functions of the monitor RadiForce MX317W and it's not a medical imaging software.
5. Indications for use
This Product is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.
{5}------------------------------------------------
6. Comparison of Technological Characteristics with the predicate device
The comparison table below enumerates information derived from the product brochure and measured values of the each device and different technological characteristics are discussed in it:
| Attributes | Proposed Device:RadiForce MX317W | Predicate Device:RadiForce MX315W |
|---|---|---|
| Indications for Use | This Product is indicated for usein displaying radiological imagesfor review, analysis, anddiagnosis by trained medicalpractitioners. The display is notintended for mammography. | This product is intended to be usedin displaying and viewing digitalimages for review, analysis anddiagnosis by trained medicalpractitioners. It does not supportthe display of mammographyimages for diagnosis. |
| Display Technology | TFT ColorLCD Panel (IPS) | TFT ColorLCD Panel (IPS) |
| Screen size | 77.5cm / 30.5"Aspect ratio: 17 : 9 | 79.0cm / 31.1"Aspect ratio: 17 : 9 |
| Backlight type | LED | LED |
| Frame rate and refresh rate | ||
| Digital Scanning Frequency(H / V) | USB Type-C: 31 - 134 kHz / 59- 61 HzDisplayPort: 31 - 134 kHz / 59 -61 HzHDMI: 31 - 136 kHz / 59 - 61Hz | 31 - 134 kHz / 14 - 61 Hz |
| Display Interface | ||
| Input videosignals | USB Type-C (DisplayPort AltMode) x 1,DisplayPort x 2,HDMI x 1 | DVI-D x 1,DisplayPort x 2 |
| Output videosignals | USB Type-C (daisy chain) x 1 | DisplayPort x 1 (daisy chain) |
| Video bandwidth | ||
| HDMI : 25-600MHzDisplayPort : 25-570MHzDisplayPort Alt Mode Input(onUSB Type-C) : 25-570MHz | DP1 Port Single Display: 25-570MHzPbyP/PinP: 25-300MHzDP2 Port Single Display: NotsupportedPbyP/PinP: 25-300MHzUp to 288MHz: 8/10bit color | |
| Ambient light sensing | ||
| Ambient light sensor | Photo DiodePosition: In the upper bezel of the screenSoftware tool: RadiCS | Photo DiodePosition: In the upper bezel of the screenSoftware tool: RadiCS |
| Luminance calibration tools | ||
| Integrated optical sensorExternal optical sensorCalibration software:RadiCS | Integrated optical sensorExternal optical sensorCalibration software:RadiCS | |
| Quality-control procedures | ||
| Software: RadiCS | Software: RadiCS |
{6}------------------------------------------------
It is clear that the technological characteristics differences discussed above do not affect the safety and the effectiveness of the MX317W.
7. Performance Testing
The bench tests below were performed on the RadiForce MX317W following the instructions in "Guidance for Industry and FDA Staff: Display Devices for Diagnostic Radiology" issued on September 28,2022:
- Measurement of spatial resolution expressed as modulation transfer function (MTF) ●
- The maximum number allowed for each type of pixel defects/faults ●
- . Visual check of presence or absence of miscellaneous artifacts on the display screen as specified in TG18 guideline
- Measurement of temporal response ●
- Measurement of Luminance .
- Verification of the conformance to DICOM GSDF as specified in Assessment of . Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 guideline)
- Measurement of Color tracking
The test results showed that the RadiForce MX317W has display characteristics equivalent to those of the predicate device, RadiForce MX315W.
Besides, the display characteristics of the RadiForce MX317W meet the pre-defined criteria when criteria are set.
No animal or clinical testing was performed on the RadiForce MX317W.
{7}------------------------------------------------
8. Conclusion
The RadiForce MX317W was determined to be substantially equivalent to the predicate device due to the following reasons:
- The stated intended use is substantially the same as that of the predicate device. ●
- It was confirmed that the technological characteristics differences from those of the . predicate device do not affect the safety or the effectiveness.
- The bench tests demonstrated that the display characteristics are equivalent to those of the . predicate device.
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).