K Number
K233254
Manufacturer
Date Cleared
2024-06-14

(260 days)

Product Code
Regulation Number
882.4560
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TELIGEN Navigation Ready Instruments:

The TELIGEN System is indicated to provide minimally invasive access, visualization, illumination, magnification and discectomy of the surgical area of the spine.

The TELIGEN Access Probe and TELIGEN Clear are Navigation Ready Instruments and when used with the compatible Universal Navigation Adaptor Set (UNAS) are intended to assist the surgeon in locating anatomical structures in either open or percutaneous procedures. The navigation feature is used in surgical spinal procedures, in which:

· the use of stereotactic surgery may be appropriate, and

· reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system and associated tracking arrays.

These procedures include but are not limited to spinal fusion.

TELIGEN Clear and TELIGEN Access Probe, when used with UNAS Navigation Rings, can be precalibrated with:

  • · the VELYS Spine System using the VELYS Spine Instrument Arrays,
  • · the Brainlab Navigation System using the UNAS Arrays.

TELIGEN Access Probe in conjunction with UNAS can also be manually calibrated with other navigation systems, using tracking arrays supplied by the navigation system manufacturer.

The TELIGEN Access Probe is indicated for stimulation of peripheral motor nerves, including spinal nerve roots, for location and identification during surgery.

Discectomy Navigation Ready Instruments:

The Discectorny Navigation Ready Instruments when used with the compatible Universal Navigation Adaptor Set are intended to assist the surgeon in locating anatomical structures to facilitate disc space preparation, including discectomy or bony resection. These are indicated for use in surgical spinal procedures, in which:

  • · the use of stereotactic surgery may be appropriate, and
    · reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data-based model of the anatomy using a navigation system and associated navigation arrays.

These procedures include but are not limited to spinal fusion. The Discectomy Navigation Ready Instruments can be pre-calibrated with the VELYS Spine System (only for TELIGEN Graft Delivery Cannula and the curettes) as well as the Brainlab Navigation System.Universal Navigation Adaptor Set:

The Universal Navigation Adaptor Set (UNAS) is intended for use with the compatible DePuy Synthes Navigation Ready Instruments to assist the surgeon in locating anatomical structures in either open or percutaneous procedures. These are indicated for use in surgical spinal procedures, in which:

· the use of stereotactic surgery may be appropriate, and

· reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system and associated tracking arrays.

These procedures include but are not limited to spinal fusion. The DePuy Synthes Navigation Ready Instruments, when used with UNAS, can be:

· pre-calibrated with the VELYS Spine System using VELYS Spine System Instrument Arrays,

  • · pre-calibrated and/or manually calibrated with the Brainlab Navigation System,
    where other navigation systems require manual calibration and tracking arrays supplied by the navigation system manufacturer.
Device Description

TELIGEN Navigation Ready Instruments

TELIGEN Navigation Ready Instruments are part of TELIGEN Kits.

The TELIGEN Kits are sterile, single use kits intended for use in surgical spinal procedures allowing for access, visualization, discectomy, graft delivery, navigation and peripheral motor nerve stimulation.

The TELIGEN Kits include a camera, ports and port holder, TELIGEN Clear, a soft tissue retractor, a port cutter cartridge and bone graft delivery instruments. Additionally, the TELIGEN Procedure Kit Pro includes an Access Probe.

TELIGEN Access Probe and TELIGEN Clear are part of the DePuy Synthes Navigation Ready Instruments Portfolio and are designed for navigated and non-navigated use. Navigation of these instruments is achieved using the DePuy Synthes Universal Navigation Adaptor Set (UNAS). For further details on UNAS, refer to the UNAS labeling.

Discectomy Navigation Ready Instruments

The Discectomy Navigation Ready Instruments are reusable instruments used for disc space preparation, including discectomy or bony resection. These instruments are designed for navigated and non-navigated use. Navigation of these instruments is achieved using the DePuy Synthes Universal Navigation Adaptor Set (UNAS) and associated navigation arrays. For further details on UNAS, refer to the UNAS labeling.

Universal Navigation Adaptor Set

The Universal Navigation Adaptor Set (UNAS) contains reusable spine surgical instruments used to aid in determining the correct location and trajectory of spinal instruments and implants. The UNAS has an interface between third-party navigation systems and the DePuy Synthes Navigation Ready Instruments. The UNAS can only be used with the VELYS Spine System as well as Brainlab and Medtronic StealthStation® navigation systems. The UNAS includes:

  • Brainlab compatible UNAS Navigation Arrays,
  • . VELYS Spine/Brainlab compatible Navigation Rings and
  • Medtronic compatible Navigation Ring ST. ●

The Navigation Rings and Navigation Ring ST mate with compatible DePuy Synthes Navigation Ready Instruments. These instruments include implant site preparation and implant insertion instruments as well as access and discectomy instruments.

When the VELYS Spine/Brainlab compatible Navigation Ring is attached to the Navigation Ready Instrument:

  • . VELYS Spine System Instrument Array can be attached and the instrument can be used with the VELYS Spine System as pre-calibrated instrument, or
  • . UNAS Navigation Array can be attached and the instrument can be used with the Brainlab Navigation System as either a manually calibrated and/or pre-calibrated instrument.

When the Navigation Ring ST is attached to the Navigation Ready Instrument, a Medtronic SureTrak® II Universal Tracker Fighter array (SureTrak II array) can be attached, and the instrument can be manually calibrated only with the Medtronic StealthStation Navigation System.

AI/ML Overview

This submission pertains to the TELIGEN System Navigation Ready Instruments, which include the TELIGEN Access Probe, TELIGEN Clear, TELIGEN Graft Delivery Cannula, and associated curettes, along with the Universal Navigation Adaptor Set (UNAS). The primary purpose of this 510(k) submission is to expand the compatibility of these instruments with the VELYS Spine System.

1. Table of Acceptance Criteria and Reported Device Performance:

The provided document does not explicitly list quantitative acceptance criteria in a tabular format. However, it states the following performance aspects were evaluated:

Acceptance Criteria CategoryReported Device Performance (as described)
Accuracy VerificationVerified through engineering analysis.
Navigation System Instrument Accuracy RequirementsFulfilled navigation systems instrument accuracy requirements as stated by the navigation manufacturer.
CAD Model EvaluationEvaluation performed. (Implies satisfactory evaluation)
Simulated Use EvaluationEvaluation performed. (Implies satisfactory evaluation)
Technological Characteristics (Design, Materials, Performance)Consistent with predicate devices.
Intended UseConsistent with predicate devices.
Safety and EffectivenessDoes not raise new questions of safety and effectiveness based on application of recognized consensus standards and design controls.

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the exact sample size or number of "test cases" for the evaluations (Accuracy Verification, CAD Model Evaluation, Simulated Use Evaluation). It refers to "engineering analysis" and "simulated use evaluation" which typically involve a set of defined tests rather than a statistical sample size of patient data.

The data provenance is not explicitly stated in terms of country of origin or whether it's retrospective or prospective. Given the nature of the device (surgical instruments with navigation capabilities), the evaluations would likely be laboratory-based and simulated environments.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

The document does not mention the use of experts to establish ground truth for the test set. The evaluations described (engineering analysis, meeting manufacturer's requirements, CAD model, simulated use) suggest a focus on design validation and performance specifications rather than clinical evaluation with expert-defined ground truth on patient data.

4. Adjudication method for the test set:

The document does not describe any adjudication method. This type of evaluation, for surgical instruments, typically relies on direct measurement and adherence to specifications rather than subjective expert assessment requiring adjudication.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

No mention of an MRMC comparative effectiveness study, or any studies involving human readers, is found in the provided text. The device is a set of surgical instruments, not an AI diagnostic tool that would typically involve human reader studies.

6. Standalone (algorithm only without human-in-the-loop performance) Study:

Not applicable. The device is a set of physical surgical instruments with navigation capabilities, not a standalone algorithm. Its performance is intrinsically tied to its use by a surgeon within a navigation system.

7. Type of Ground Truth Used:

Based on the evaluation types:

  • For Accuracy Verification and Navigation Systems Instrument Accuracy Requirements, the ground truth would be the established engineering specifications, dimensional tolerances, and the accuracy claims/requirements published by the navigation system manufacturers (e.g., VELYS Spine System, Brainlab, Medtronic StealthStation®).
  • For CAD Model Evaluation, the ground truth would be the design specifications and theoretical performance derived from the CAD models.
  • For Simulated Use Evaluation, the ground truth would be the defined objective criteria for successful surgical simulation (e.g., precise instrument placement, proper interaction with simulated anatomy).

8. Sample Size for the Training Set:

Not applicable. The device is a set of physical surgical instruments, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

June 14, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of the HHS logo is the FDA logo, which consists of a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

DePuy Synthes Marina Minnock Regulatory Affairs Specialist Eimattstrasse 3 Oberdorf, 4436 Switzerland

Re: K233254

Trade/Device Name: TELIGEN System Navigation Ready Instruments Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: October 11, 2023 Received: June 7, 2024

Dear Marina Minnock:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Jesse
Jesse Muir -S Muir-S
Date: 2024.06.14 17:26:02
-04'00'

For Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Submission Number (if known)

K233254

Device Name

TELIGEN System Navigation Ready Instruments

Indications for Use (Describe)

TELIGEN Navigation Ready Instruments:

The TELIGEN System is indicated to provide minimally invasive access, visualization, illumination, magnification and discectomy of the surgical area of the spine.

The TELIGEN Access Probe and TELIGEN Clear are Navigation Ready Instruments and when used with the compatible Universal Navigation Adaptor Set (UNAS) are intended to assist the surgeon in locating anatomical structures in either open or percutaneous procedures. The navigation feature is used in surgical spinal procedures, in which:

· the use of stereotactic surgery may be appropriate, and

· reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system and associated tracking arrays.

These procedures include but are not limited to spinal fusion.

TELIGEN Clear and TELIGEN Access Probe, when used with UNAS Navigation Rings, can be precalibrated with:

  • · the VELYS Spine System using the VELYS Spine Instrument Arrays,
  • · the Brainlab Navigation System using the UNAS Arrays.

TELIGEN Access Probe in conjunction with UNAS can also be manually calibrated with other navigation systems, using tracking arrays supplied by the navigation system manufacturer.

The TELIGEN Access Probe is indicated for stimulation of peripheral motor nerves, including spinal nerve roots, for location and identification during surgery.

Discectomy Navigation Ready Instruments:

The Discectorny Navigation Ready Instruments when used with the compatible Universal Navigation Adaptor Set are intended to assist the surgeon in locating anatomical structures to facilitate disc space preparation, including discectomy or bony resection. These are indicated for use in surgical spinal procedures, in which:

  • · the use of stereotactic surgery may be appropriate, and
    · reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data-based model of the anatomy using a navigation system and associated

{3}------------------------------------------------

navigation arrays.

These procedures include but are not limited to spinal fusion. The Discectomy Navigation Ready Instruments can be pre-calibrated with the VELYS Spine System (only for TELIGEN Graft Delivery Cannula and the curettes) as well as the Brainlab Navigation System.Universal Navigation Adaptor Set:

The Universal Navigation Adaptor Set (UNAS) is intended for use with the compatible DePuy Synthes Navigation Ready Instruments to assist the surgeon in locating anatomical structures in either open or percutaneous procedures. These are indicated for use in surgical spinal procedures, in which:

· the use of stereotactic surgery may be appropriate, and

· reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system and associated tracking arrays.

These procedures include but are not limited to spinal fusion. The DePuy Synthes Navigation Ready Instruments, when used with UNAS, can be:

· pre-calibrated with the VELYS Spine System using VELYS Spine System Instrument Arrays,

  • · pre-calibrated and/or manually calibrated with the Brainlab Navigation System,
    where other navigation systems require manual calibration and tracking arrays supplied by the navigation system manufacturer.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

Submitter Information A.

510(k) Sponsor:Medos International, SARL
Contact Person:Marina MinnockRegulatory Affairs SpecialistEimattstrasse 34436 Oberdorf BLSwitzerland
Telephone:+41 76 706 0799
Email:mminnock@its.jnj.com
B. Date Prepared05 June 2024
C. Device Name
Trade/Proprietary Name:TELIGEN System Navigation Ready Instruments
Common/Usual Name:Orthopedic Stereotaxic Instrument
Device Classification andRegulation:Class IIOLO - 21 CFR §882.4560HRX - 21 CFR §888.1100PDQ - 21 CFR §874.1820
Classification Product andPanel CodeOLO - NeurologyHRX - OrthopedicPDQ - Neurology
D. Predicate Device Names
Primary Predicate Device:

TELIGEN System Navigation Ready Indications – K223108 – OLO

Additional Predicate Device:

Discectomy Navigation Ready Instruments and Universal Navigation Adaptor Set – K212756 – OLO

{5}------------------------------------------------

E. Device Description

TELIGEN Navigation Ready Instruments

TELIGEN Navigation Ready Instruments are part of TELIGEN Kits.

The TELIGEN Kits are sterile, single use kits intended for use in surgical spinal procedures allowing for access, visualization, discectomy, graft delivery, navigation and peripheral motor nerve stimulation.

The TELIGEN Kits include a camera, ports and port holder, TELIGEN Clear, a soft tissue retractor, a port cutter cartridge and bone graft delivery instruments. Additionally, the TELIGEN Procedure Kit Pro includes an Access Probe.

TELIGEN Access Probe and TELIGEN Clear are part of the DePuy Synthes Navigation Ready Instruments Portfolio and are designed for navigated and non-navigated use. Navigation of these instruments is achieved using the DePuy Synthes Universal Navigation Adaptor Set (UNAS). For further details on UNAS, refer to the UNAS labeling.

Discectomy Navigation Ready Instruments

The Discectomy Navigation Ready Instruments are reusable instruments used for disc space preparation, including discectomy or bony resection. These instruments are designed for navigated and non-navigated use. Navigation of these instruments is achieved using the DePuy Synthes Universal Navigation Adaptor Set (UNAS) and associated navigation arrays. For further details on UNAS, refer to the UNAS labeling.

Universal Navigation Adaptor Set

The Universal Navigation Adaptor Set (UNAS) contains reusable spine surgical instruments used to aid in determining the correct location and trajectory of spinal instruments and implants. The UNAS has an interface between third-party navigation systems and the DePuy Synthes Navigation Ready Instruments. The UNAS can only be used with the VELYS Spine System as well as Brainlab and Medtronic StealthStation® navigation systems. The UNAS includes:

  • Brainlab compatible UNAS Navigation Arrays,
  • . VELYS Spine/Brainlab compatible Navigation Rings and
  • Medtronic compatible Navigation Ring ST. ●

The Navigation Rings and Navigation Ring ST mate with compatible DePuy Synthes Navigation Ready Instruments. These instruments include implant site preparation and implant insertion instruments as well as access and discectomy instruments.

When the VELYS Spine/Brainlab compatible Navigation Ring is attached to the Navigation Ready Instrument:

{6}------------------------------------------------

  • . VELYS Spine System Instrument Array can be attached and the instrument can be used with the VELYS Spine System as pre-calibrated instrument, or
  • . UNAS Navigation Array can be attached and the instrument can be used with the Brainlab Navigation System as either a manually calibrated and/or pre-calibrated instrument.

When the Navigation Ring ST is attached to the Navigation Ready Instrument, a Medtronic SureTrak® II Universal Tracker Fighter array (SureTrak II array) can be attached, and the instrument can be manually calibrated only with the Medtronic StealthStation Navigation System.

F. Indications for Use

TELIGEN Navigation Ready Instruments:

The TELIGEN System is indicated to provide minimally invasive access, visualization, magnification and discectomy of the surgical area of the spine.

The TELIGEN Access Probe and TELIGEN Clear are Navigation Ready Instruments and when used with the compatible Universal Navigation Adaptor Set (UNAS) are intended to assist the surgeon in locating anatomical structures in either open or percutaneous procedures. The navigation feature is used in surgical spinal procedures, in which:

  • . the use of stereotactic surgery may be appropriate, and
  • . reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system and associated tracking arrays.

These procedures include but are not limited to spinal fusion.

TELIGEN Clear and TELIGEN Access Probe, when used with UNAS Navigation Rings, can be precalibrated with:

  • . the VELYS Spine System using the VELYS Spine Instrument Arrays,
  • . the Brainlab Navigation System using the UNAS Arrays.

TELIGEN Access Probe in conjunction with UNAS can also be manually calibrated with other navigation systems, using tracking arrays supplied by the navigation system manufacturer.

The TELIGEN Access Probe is indicated for stimulation of peripheral motor nerves, including spinal nerve roots, for location and identification during surgery.

Discectomy Navigation Ready Instruments:

The Discectomy Navigation Ready Instruments when used with the compatible Universal Navigation Adaptor Set are intended to assist the surgeon in locating anatomical structures to

{7}------------------------------------------------

facilitate disc space preparation, including discectomy or bony resection. These are indicated for use in surgical spinal procedures, in which:

  • . the use of stereotactic surgery may be appropriate, and
  • . reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data-based model of the anatomy using a navigation system and associated navigation arrays.

These procedures include but are not limited to spinal fusion. The Discectomy Navigation Ready Instruments can be pre-calibrated with the VELYS Spine System (only for TELIGEN Graft Delivery Cannula and the curettes) as well as the Brainlab Navigation System.

Universal Navigation Adaptor Set:

The Universal Navigation Adaptor Set (UNAS) is intended for use with the compatible DePuy Synthes Navigation Ready Instruments to assist the surgeon in locating anatomical structures in either open or percutaneous procedures. These are indicated for use in surgical spinal procedures, in which:

  • . the use of stereotactic surgery may be appropriate, and
  • . reference to a rigid anatomical structure, such as the pelvis or a vertebrae can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy using a navigation system and associated tracking arrays.

These procedures include but are not limited to spinal fusion. The DePuy Synthes Navigation Ready Instruments, when used with UNAS, can be:

  • . pre-calibrated with the VELYS Spine System using VELYS Spine System Instrument Arrays,
  • . pre-calibrated and/or manually calibrated with the Brainlab Navigation System,

where other navigation systems require manual calibration and tracking arrays supplied by the navigation system manufacturer.

G. Summary of Similarities and Differences in Technological Characteristics, Performance, and Intended Use

The technological characteristics, including design, material, and performance as well as intended use of the TELIGEN System Navigation Ready Instruments are consistent with those of the predicate devices.

Compared to the predicate devices, the subject devices expand the scope of the TELIGEN System Navigation Ready Instruments to establish compatibility with an additional navigation system, namely the VELYS Spine System. TELIGEN System Navigation Ready Instruments include TELIGEN Access Probe, TELIGEN Clear, TELIGEN Graft Delivery Cannula, and the associated curettes, in conjunction with UNAS Navigation Rings, previously cleared for compatibility with different

{8}------------------------------------------------

navigation systems. This does not raise new questions of safety and effectiveness based on application of recognized consensus standards and design controls.

H. Materials

The TELIGEN System Navigation Ready Instruments are manufactured as follows:

TELIGEN Access Probe

Stainless steel with-Parylene C coating Stainless steel Plastic (Polycarbonate)

TELIGEN Clear

Stainless steel Plastic (polycarbonate, polyethylene, thermoplastic urethane (TPU)) Nitrile rubber

TELIGEN Graft Delivery Cannula and the curettes

Stainless steel, TiAIN (Titanium Aluminum Nitride Coating) Radel® (Polyphenylsulfone) Silicone rubber

UNAS Navigation Rings:

Stainless steel Titanium alloy

l. Performance Data

The performance data for the subject devices consists of the following evaluations:

  • . Accuracy Verification through engineering analysis,
  • . Fulfillment of navigation systems instrument accuracy requirements as stated by the navigation manufacturer,
  • . CAD Model Evaluation,
  • . Simulated Use Evaluation.

J. Conclusion

The indications for use of the TELIGEN System Navigation Ready Instruments are consistent with those of the predicate devices. The technological characteristics of the TELIGEN System Navigation Ready Instruments in terms of design, materials and performance are consistent with those of the predicate devices. The TELIGEN System Navigation Ready Instruments are substantially equivalent to the predicate devices.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).