K Number
K233107

Validate with FDA (Live)

Date Cleared
2024-08-30

(338 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is a digital radiography/fluoroscopy system used interventional angography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen, and lower extremities.

Device Description

The Alphenix, INFX-8000V/B, INFX-8000V/S, V9.5, is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in an interventional setting. The system consists of a C-arm/ Ω-arm which is equipped with an X-ray tube, beam limiter and X-ray receptor, X-ray controller, computers with system and processing software, and a patient radiographic table.

AI/ML Overview

The provided text is a 510(k) summary for the Alphenix, INFX-8000V/B, INFX-8000V/S, V9.5. This document focuses on demonstrating substantial equivalence to a predicate device (Alphenix, V9.1) due to software modifications (V9.1 to V9.5) and integrating previously cleared functions/components.

Crucially, the document explicitly states there are no new indications for use or intended uses of the device, and that the basic system configuration, method of operation, base software, and manufacturing process remain unchanged.

Because the submission is for software changes to an existing device, and not for a new AI/ML-driven diagnostic or assistive device that would require performance validation against clinical endpoints and specific acceptance criteria related to diagnostic accuracy, the provided text does not contain the detailed information necessary to answer points 1-9 of your request.

The study described is not a performance study in the sense of evaluating the diagnostic accuracy of an AI algorithm. Instead, it is a verification/validation testing to demonstrate that the established specifications for the modified device (software update) have been met, and that the device remains as safe and effective as its predicate.

Here's a breakdown of why the requested information is not present and what is stated:

Information Not Present in the Provided Text:

  • 1. A table of acceptance criteria and the reported device performance: This would be relevant for a device with a new diagnostic or assistive function whose performance needs to be validated (e.g., sensitivity, specificity, AUC for an AI algorithm detecting disease). For software updates to an existing interventional fluoroscopic X-ray system, the "performance" validated is typically functional and safety-related, not diagnostic accuracy.
  • 2. Sample size used for the test set and the data provenance: Not applicable as it's not a diagnostic performance study.
  • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
  • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI-assisted diagnostic device submission.
  • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable.
  • 8. The sample size for the training set: Not applicable, as it's not an AI/ML device being developed and trained.
  • 9. How the ground truth for the training set was established: Not applicable.

What is stated regarding testing and acceptance:

The document indicates that the acceptance criteria are related to safety and functionality of the X-ray system after the software update, ensuring it remains substantially equivalent to the predicate device.

  • Testing Method: "Risk analysis and verification/validation testing conducted through bench testing demonstrate that the established specifications for the device have been met."
  • Conformance to Standards: "Testing of the modified system was conducted in accordance with the applicable standards published by the International Electromechanical Commission (IEC) for Medical Devices and XR Systems."
    • Applicable standards listed include: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-28, IEC 60601-2-43, IEC 62304, IEC 62366-1, IEC 81001-5-1, ISO 17664-2, IEC TR 60601-4-2.
  • Grounds for Conclusion of Substantial Equivalence: "Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls presented in this submission it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate device."
  • Software Validation: "A Basic Documentation Level was determined, per the FDA guidance document, 'Content of Premarket Submissions for Device Software Functions' issued on June 14, 2023, is also included as part of this submission." This indicates that software validation was performed according to FDA guidance for medical device software.

In summary, the provided document is a 510(k) for a software update to an existing fluoroscopic X-ray system, asserting substantial equivalence based on safety, functionality, and compliance with general medical device and X-ray specific standards, rather than a clinical performance study of a novel AI diagnostic algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 30, 2024

Canon Medical Systems Corporation % Orlando Tadeo Jr. Sr. Manager, Regulatory Affairs Canon Medical Systems, USA 2441 Michelle Drive TUSTIN, CA 92780

Re: K233107

Trade/Device Name: Alphenix, INFX-8000V/B, INFX-8000V/S, V9.5 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB Dated: August 2, 2024 Received: August 2, 2024

Dear Orlando Tadeo Jr.:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233107

Device Name Alphenix, INFX-8000V/B, INFX-8000V/S, V9.5

Indications for Use (Describe)

This device is a digital radiography/fluoroscopy system used interventional angography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen, and lower extremities.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K233107

EDICAL SYSTEMS USA, INC.

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92

CLASSIFICATION and DEVICE NAME 1.

Classification NameImage-Intensified Fluoroscopic X-ray System
Product CodeOWB
Regulation Number21 CFR 892.1650
Regulatory ClassClass II
Trade Proprietary NameAlphenix, INFX-8000V/B, INFX-8000V/S, V9.5

2. SUBMITTER'S NAME

Canon Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550

3. OFFICIAL CORRESPONDENT

Fumiaki Teshima Senior Manager, Quality Assurance Department

4. CONTACT PERSON, U.S. AGENT and ADDRESS

Contact Person

Orlando Tadeo, Jr. Sr. Manager, Regulatory Affairs Canon Medical Systems USA 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 483-1551 Fax: (714) 730-1310 otadeo@us.medical.canon

5. MANUFACTURING SITE

Canon Medical Systems Corporation (CMSC) 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan

6. ESTABLISHMENT REGISTRATION 9614698

7. DATE PREPARED September 26, 2023

Official Correspondent/U.S. Agent

Paul Biggins Sr. Director, Regulatory Affairs Canon Medical Systems USA 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 669-7808 Fax: (714) 730-1310 pbiggins@us.medical.canon

{5}------------------------------------------------

MEDICAL SYSTEMS USA, INC.

8. TRADE NAME(S)

Alphenix, INFX-8000V/B, INFX-8000V/S, V9.5

9. DEVICE NAME

Interventional Fluoroscopic X-ray System

10. CLASSIFICATION PANEL

Radiology

11. DEVICE CLASSIFICATION

Class II (per 21 CFR 892.1650)

12. PRODUCT CODE / DESCRIPTION

Product Code: OWB - Image-Intensified Fluoroscopic X-ray System

13. PERFORMANCE STANDARD

This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR Subchapter J, Federal Diagnostic X-ray Equipment Standard].

14. PREDICATE DEVICE

Alphenix, INFX-8000V/B, INFX-8000V/S, V9.1 (K203551)

TABLE 1: Predicate Device
-----------------------------------
Trade Proprietary NameAlphenix, INFX-8000V/B, INFX-8000V/S, V9.1
Marketed byCanon Medical Systems USA, Inc.
510(k) NumberK203551
Clearance DateDecember 18, 2020
Classification NameImage-Intensified Fluoroscopic X-ray System
Product CodeOWB
Regulation Number21 CFR 892.1650
Regulatory ClassClass II

15. REASON FOR SUBMISSION

Modification of a cleared device

16. SUBMISSION TYPE

Traditional 510(k)

17. DEVICE DESCRIPTION

The Alphenix, INFX-8000V/B, INFX-8000V/S, V9.5, is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in an interventional setting. The system consists of a C-arm/ Ω-arm which is equipped with an X-ray tube, beam limiter and X-ray receptor, X-ray controller, computers with system and processing software, and a patient radiographic table.

{6}------------------------------------------------

18. INDICATIONS FOR USE

This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

19. SAFETY

The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC60601-1 standards, its collateral standards and particular standards; IEC 60601-2-43, IEC60601-2-28, and IEC TR 60601-4-2. All requirements of the Federal Diagnostic Equipment Standard, as outlined in 21 CFR §1020, that apply to this device, will be met and reported via product report.

LIST OF APPLICABLE STANDARDS

  • IEC 60601-1:2005+A1:2012+A2:2020
  • IEC 60601-1-2:2014 + A1:2020
  • IEC 60601-1-3:2008+A1:2013+A2:2021 ●
  • IEC 60601-1-6:2010+A1:2013+A2:2020
  • IEC 60601-2-28:2017
  • IEC 60601-2-43:2010+A1:2017+A2:2019
  • IEC 62304:2006+A1:2015
  • IEC 62366-1:2015 + A1:2020
  • IEC 81001-5-1:2021
  • ISO 17664-2:2021 ●
  • IEC TR 60601-4-2:2016

20. TESTING

Risk analysis and verification/validation testing conducted through bench testing demonstrate that the established specifications for the device have been met. Testing of the modified system was conducted in accordance with the applicable standards published by the International Electromechanical Commission (IEC) for Medical Devices and XR Systems.

A Basic Documentation Level was determined, per the FDA guidance document, "Content of Premarket Submissions for Device Software Functions" issued on June 14, 2023, is also included as part of this submission.

Additionally, the design controls used for this device included risk management and all known risks were mitigated to an acceptable level.

{7}------------------------------------------------

21. SUBSTANTIAL EQUIVALENCE

The Alphenix, INFX-8000V/B, INFX-8000V/S, V9.5 is substantially equivalent to the Alphenix, INFX-8000V/B, INFX-8000V/S, V9.1, which received premarket clearance under K203551, marketed by Canon Medical Systems. Alphenix, INFX-8000V/B, INFX-8000V/S, V9.5, includes modifications to the cleared device consisting of software changes from V9.1 to V9.5, integrating previously cleared functions/components.

The basic system configuration, method of operation, base software and manufacturing process remain unchanged from the cleared device. There are no new indications for use or intended uses of the device.

22. CONCLUSION

The Alphenix, INFX-8000V/B, INFX-8000V/S, V9.5, performs in a manner similar to and is intended for the same use as the predicate device, as indicated in product the labeling. The modifications incorporated do not change the indications for use or the intended use of the device. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls presented in this submission it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate device. It is the contention of Canon Medical Systems Corporation that all new safety issues have been addressed in the design of this change and that adequate evidence of this is presented with this submission.

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.