K Number
K232968
Date Cleared
2024-06-06

(259 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Grande Medical Masks is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. The Grande Medical Masks are single use, disposable device, provided non-sterile.

Device Description

Grande Medical Masks are 4-layer surgical masks that covers the user's nose and mouth and provides a physical barrier to fluids and particulate materials. This device is non-sterile and for single use only. The mask is constructed of nonwoven fabric, including the bottom layer, surface layer 1, surface layer 2 and middle layer, and is provided with ear loops and nose clip.

AI/ML Overview

The provided text describes the acceptance criteria and the study that proves the device meets the acceptance criteria for the "Grande Medical Masks, Model SF-600".

1. Table of Acceptance Criteria and Reported Device Performance:

TestAcceptance Criteria (F2100-19 Level 2)Reported Device PerformanceResult
Bacteria Filtration Efficiency (ASTM F2101)≥ 98 %Pass at 99 %Pass
Particle Filtration Efficiency (ASTM F2299)≥ 98 %Pass at 99%Pass
Differential Pressure (EN 14683:2019)< 6.0 mmH2O/cm²Less than 4.5 mmH2O/cm²Pass
Resistance to penetration by synthetic blood (ASTM F1862)120 mmHg (≥ 29/32 show passing results per lot)120 mmHgPass
Flammability (16 CFR Part 1610)Class I (Burn time ≥3.5 s, IBE, or DNI)Class IPass
In vitro Cytotoxicity (ISO 10993-5)Non-cytotoxicNon-cytotoxicPass
Skin sensitization Test (ISO 10993-10)Not a sensitizerNot a sensitizerPass
Skin Irritation Test (ISO 10993-10)Not an irritantNot an irritantPass

2. Sample size used for the test set and the data provenance:

The document does not explicitly state the sample sizes used for each specific non-clinical test. However, it indicates standard test methodologies like ASTM F2101, F2299, F1862, EN 14683:2019, 16 CFR Part 1610, ISO 10993-5, and ISO 10993-10, which inherently involve specific sample sizes as part of their standardized procedures. The data provenance is derived from these non-clinical tests, which are conducted in a controlled environment as per international and national standards. The country of origin of the data is not specified beyond the reference to the manufacturer, Golden Grande Corp. in Taiwan. The data is retrospective in the sense that it's generated from lab tests, not from observations of real-world use over time.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable. For medical masks, ground truth is established by objective pass/fail criteria of standardized performance tests, not by expert consensus on subjective interpretations. The "experts" involved are the certified laboratories and technicians performing the tests according to the specified methodologies.

4. Adjudication method for the test set:

Not applicable. The tests conducted for medical masks (e.g., bacterial filtration, particle filtration, differential pressure) yield objective, quantifiable results that are directly compared against predefined numerical acceptance criteria. There is no subjective interpretation or multi-reader adjudication process involved.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a medical mask, a physical barrier device, not an AI-powered diagnostic or assistive tool. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

Not applicable. As stated above, this is a physical medical device, not an algorithm or software. No standalone algorithm-only performance testing was performed because there is no algorithm.

7. The type of ground truth used:

The ground truth used is based on objective, quantifiable results derived from standardized in vitro and in vivo (for biocompatibility) performance tests. These tests have established pass/fail criteria defined by international and national standards (e.g., ASTM F2100-19 Level 2 criteria, ISO 10993 standards). This is not expert consensus, pathology, or outcomes data in the usual sense, but rather a direct measurement of the device's physical and biological properties against established safety and performance benchmarks.

8. The sample size for the training set:

Not applicable. This device is a medical mask, not a machine learning model. Therefore, there is no "training set" in the context of AI or software development.

9. How the ground truth for the training set was established:

Not applicable. As there is no training set for a medical mask, there is no ground truth established for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 6, 2024

Golden Grande Corp. % Tyra Chiu Regulatory Consultant VOLER Biotech Consulting Co., Ltd. 1F., No. 3-1, Ln. 58, Hejiang St., Zhongshan Dist. Taipei City, 104 Taiwan

Re: K232968

Trade/Device Name: Grande Medical Masks, Model SF-600 Regulatory Class: Not Classified Product Code: FXX Dated: July 8, 2023 Received: September 21, 2023

Dear Tyra Chiu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

AllanıG

For Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices

{2}------------------------------------------------

Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232968

Device Name Grande Medical Masks, Model SF-600

Indications for Use (Describe)

Grande Medical Masks is intended to be worn to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. The Grande Medical Masks are single use, disposable device, provided non-sterile.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary (K232968)

(As required by 21 CFR 807.92)

DateJune 4, 2024
Manufacturer/510(k) OwnerGolden Grande Corp.NO. 622-6, GUANGHUA LI, XUEJIA DIST., TAINAN CITY,72651, TAIWAN
Contact PersonHsin Hui TsengPhone:E-mail: cindytseng@grande.com.tw
Device Trade NameGrande Medical Masks
Common NameSurgical Masks
Classification NameMasks, Surgical
Device ClassII
Medical SpecialtyGeneral & Plastic Surgery
Regulation Number878.4040
Product CodeFXX
Device Descriptionand TechnologyCharacteristicsGrande Medical Masks are 4-layer surgical masks that covers theuser's nose and mouth and provides a physical barrier to fluidsand particulate materials. This device is non-sterile and for singleuse only.The mask is constructed of nonwoven fabric, including the bottomlayer, surface layer 1, surface layer 2 and middle layer, and isprovided with ear loops and nose clip.
ModelsSF-600
Indications for UseGrande Medical Masks is intended to be worn to protect both thepatient and healthcare personnel from the transfer ofmicroorganisms, body fluids and particulate material. These facemasks are intended for use in infection control practices to reducethe potential exposure to blood and body fluids. The GrandeMedical Masks are single use, disposable device, provided non-sterile.

Predicate Device(s) K160269 Surgical Face Masks, SAN-M PACKAGE CO., LTD.

Proposed DevicePredicate DeviceComparison
Device NameGrande Medical MasksSurgical Face MasksDifferent
510(k) #K232968K160269Different
ApplicantGolden Grande Corp.SAN-M PACKAGE CO., LTD.Different
Product codeFXX (21 CFR 878.4040)FXX (21 CFR 878.4040)Same

comparison of technological characteristics with the predicate device

{5}------------------------------------------------

ClassificationIIIISame
OTC useYESYESSame
Intended UseGrande Medical Masks is intended tobe worn to protect both the patient andhealthcare personnel from the transferof microorganisms, body fluids andparticulate material. These face masksare intended for use in infection controlpractices to reduce the potentialexposure to blood and body fluids. TheGrande Medical Masks are single use,disposabledevice, providednon-sterile.The surgical face masks are intendedto be worn to protect both the patientand healthcare personnel fromtransfer of microorganisms, bodyfluids, and particulate material. Theseface masks are intended for use ininfection control practices to reducethe potential exposure to blood andbody fluids. This is a single-use,disposable device, provided non-sterile.Same
Dimensions175±3 mm x 95±3 mm90mm±3175±5mm,90mm±3180±5mmSimilar
Layers44Same
Mask StyleFlat-pleatedFlat-pleatedSame
Design FeatureEar loopsEar loopsSame
SterilityNon-SterileNon-SterileSame
UseSingle UseSingle UseSame
Material:Surface Layer 1: PolypropylenespunbondSurface Layer 2: PolypropylenespunbondMiddle Layer: PolypropyleneMeltblownBottom Layer: PolypropyleneSpunbondNose Clip: malleable aluminum wirecovered with polyethyleneEar Loops: PolyamideOuter Layer: Polypropylene.Filter Layer 1: PolypropylenespunbondFilter Layer 2: PolypropylenemeltblownInner Layer: polypropylene.Earloops: Polyester, polyurethaneSide tapes: Polyester spunbond (earloops mask only)Tie tapes: Polypropylene spunbondor polyester spunbondNose clamp: Polyethylene coatedsteel wireSimilar
ASTM Level21,2,3Similar
Fluid ResistanceASTM F1862120 mmHg80 mmHg ,120 mmHg or 160 mmHgSimilar
Bacterial FiltrationEfficiencyASTM F2101Pass at 99 %Pass at 98% (Level 1)Pass at 98% (Level 2)Pass at 99% (Level 3)Similar
DifferentialPressureMil-M-36954CLess than 4.5 mmH2O/cm²Less than 2.0 mmH2O/cm2 (Level 1)Less than 1.6 mmH2O/cm2 (Level 2)Less than 2.5 mmH2O/cm2 (Level 3)Similar
Particle FiltrationEfficiencyASTM F2299Pass at 99%Pass at 99.6% (Level 1)Pass at 99.6% (Level 2)Pass at 99.7% (Level 3)Similar
Flammability16 CFR 1610Class IClass ISame
Biocompatibilitynon- cytotoxic, non-irritating,non-sensitizing (ISO 10993-5,ISO 10993-10)non- cytotoxic, non-irritating, non-sensitizing (ISO 10993-5, ISO10993-10)Same

{6}------------------------------------------------

Discussion on The proposed devices were tested and conformed to the following Performance Data standards and the requirements stated in the Guidance for Non-Clinical Tests Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submission.

Non-Clinical Testing Summary
Test MethodologyPurpose of the testF2100-19Level 2 CriteriaResult
Bacteria Filtration EfficiencyDetermine the bacterial filtration efficiency asdirected in Test Method ASTM F2101.≥ 98 %Pass
Particle Filtration EfficiencyDetermine particulate filtration efficiency asdirected in Test Method F2299.≥ 98 %Pass
Differential Pressure (Delta-P)Determine breathing resistance or differentialpressure as directed in EN 14683:2019, AnnexC.<6.0 mmH2O/cm²Pass
Resistance to penetrationby synthetic blood,minimum pressure in mmHg for pass resultDetermine synthetic blood penetrationresistance as specified in Test Method F1862120 mmHg(≥ 29/32 showpassing results perlot)Pass
FlammabilityDetermine flammability as specified in 16 CFRPart 1610Class I(Burn time ≥3.5 s,IBE, or DNI)Pass

Non-Clinical Testing Summary

Biocompatibility Testing Summary:

Test MethodologyPurpose of the testCriteriaResults
In vitro CytotoxicitytestDetermine the effects on cells following ISO 10993-5Under the conditions ofthe test, non-cytotoxicPass
Skin sensitization TestEstimate the potential for contactsensitization following ISO 10993-10Under the conditions ofthe test, not asensitizerPass
Skin Irritation TestEstimate the irritation potential ofmedical device following ISO 10993-10Under the conditions ofthe test, not an irritantPass

A shelf-life evaluation was conducted, and the test results demonstrate that the device maintains its performance in its claimed 3-year shelf life.

Discussion on No clinical testing was performed in support of this submission.

Clinical Test Performed

Conclusion The conclusions drawn from the non-clinical tests demonstrate

{7}------------------------------------------------

that the subject device, Grande Medical Masks is as safe, as effective, and performs as well as or better than the legally marketed predicate device K160269.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.