(426 days)
Not Found
No
The description mentions a microprocessor and sensor controlling the device's function, but there is no mention of AI, ML, or any learning or adaptive capabilities. The operation appears to be based on pre-programmed sequential inflation/deflation cycles.
Yes
The device is intended to temporarily relieve minor muscle aches and/or pains and increase circulation, which are therapeutic claims. It also references predicate devices that are medical in nature (e.g., Air Compression Therapy Device).
No
Explanation: The device description states its intended use is for temporary relief of muscle aches and pain and to increase circulation, which are therapeutic functions, not diagnostic.
No
The device description explicitly states it consists of hardware components such as an air pressure sensor, air pump, and sleeves, which work together as a unit. This indicates it is a physical device, not software-only.
No, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The Leg Massager described is a physical therapy device that applies air pressure to the legs to relieve muscle aches and improve circulation. It does not perform any tests on biological samples.
- Intended Use: The intended use clearly states it's for temporary relief of muscle aches and increasing circulation, not for diagnosing any condition.
The information provided about the device's function, intended use, and lack of mention of any testing on biological samples all point away from it being an IVD.
N/A
Intended Use / Indications for Use
Leg Massager (Models:RP-ALM070H, RP-ALM071H) is intended for home to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas in people who are in good health.
Product codes
IRP, IRT
Device Description
Leg Massager (Models:RP-ALM070H, RP-ALM071H) is consist of air pressure sensor, air pump, sleeves etc working together as one unit. The air pump is connected to the dedicated sleeves via a series of hoses. The compression massage direction is from limb end to body center by inflating the air chambers sequentially and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient. The sleeve works under the action of sensor and microprocessor.
Leg Massager (Models:RP-ALM070H, RP-ALM071H), in medical market, it has a sequential squeezing from distal to proximal, thus help to improve circulation to the treated areas.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
home
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Not Found
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 890.5650 Powered inflatable tube massager.
(a)
Identification. A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
November 20, 2024
Shenzhen Ruiyi Business Technology Co., Ltd. % Libray Zhang Official Correspondent Shanghai Spica Management Consulting Co., Ltd. 609 Room, No.133 Shengang Avenue, Pudong New District Shanghai. 201306 China
Re: K232965
Trade/Device Name: Leg Massager (Models: RP-ALM070H, RP-ALM071H) Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP. IRT Dated: September 30, 2024 Received: September 30, 2024
Dear Libray Zhang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
2
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather L. Dean -S
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
3
Indications for Use
510(k) Number (if known) K232965
Device Name
Leg Massager (Models:RP-ALM070H, RP-ALM071H)
Indications for Use (Describe)
Leg Massager (Models:RP-ALM070H, RP-ALM071H) is intended for home to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas in people who are in good health.
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4
K232965
Type of Submission | Traditional |
---|---|
Date Prepared | Jun 27, 2023 |
Submission Sponsor | |
Manufacturer Name | Shenzhen Ruiyi Business Technology Co., Ltd. |
Address | Qianhai Complex A201, Qianwan Road 1, Qianhai |
Shenzhen-Hong Kong Cooperation Zone, | |
Shenzhen, China | |
Tel | 86-18566621145 |
derric@renpho.com | |
Contact Person | derric bai |
Application Correspondent | |
Company Name | Shanghai Spica Management Consulting Co., Ltd. |
Address | 609 Room, No.133 Shengang Avenue, Pudong New |
District, Shanghai, China | |
Tel | 86-13020102321 |
Libray@spicagloble.com | |
Contact Person | Libray Zhang |
Device Identification | |
Trade Name | Leg Massager (Models:RP-ALM070H, RP-ALM071H) |
Regulation Number | 21 CFR 890.5650 |
Classification Name | Massager, Powered Inflatable Tube |
Device Classification | Class II |
Panel | Physical Medicine |
Product Code | IRP |
Previous Submissions | None |
Indications for Use
Leg Massager (Models:RP-ALM070H, RP-ALM071H) is intended for home to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas in people who are in good health.
Device Description
5
Leg Massager (Models:RP-ALM070H, RP-ALM071H) is consist of air pressure sensor, air pump, sleeves etc working together as one unit. The air pump is connected to the dedicated sleeves via a series of hoses. The compression massage direction is from limb end to body center by inflating the air chambers sequentially and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient. The sleeve works under the action of sensor and microprocessor.
Leg Massager (Models:RP-ALM070H, RP-ALM071H), in medical market, it has a sequential squeezing from distal to proximal, thus help to improve circulation to the treated areas.
Predicate Device Information
Sponsor | Shenzhen Dongjilian Electronics Co., Ltd. |
---|---|
Trade/Device Name | Air Compression Therapy Device |
510(K) number | K193354 |
Regulation Number | 21 CFR 890.5650 |
Sponsor | Medella Health Limited |
Trade/Device Name | FLOWpresso |
510(K) number | K223729 |
Regulation Number | 21 CFR 890.5650 |
6
Table 6A: Summary of Comparison | ||
---|---|---|
Subject Device | Predicate Device | Differences Discussion | |
---|---|---|---|
Device name | Leg Massager (Models:RP-ALM070H, | ||
RP-ALM071H) | Air Compression Therapy Device | N/A | |
510(k) | |||
number | K232965 | K193354 | N/A |
Manufacturer | Shenzhen Ruiyi Business Technology Co., | ||
Ltd. | Shenzhen Dongjilian Electronics | ||
Co., Ltd. | N/A | ||
Product | |||
regulation | 21 CFR 890.5650 | 21 CFR 890.5650 | Same |
Classification | |||
name | Massager, Powered Inflatable Tube | Powered | |
Massager, | |||
Inflatable | |||
Tube | Same | ||
Regulation | |||
class | 2 | 2 | Same |
Product code | IRP | IRP | Same |
Indications for | |||
use | Leg Massager (Models:RP-ALM070H, | ||
RP-ALM071H) is intended for home to | |||
temporarily relieve minor muscle | |||
aches and/or pains, and to temporarily | |||
increase circulation to the treated areas in | |||
people who are in good health. | The Air Compression Therapy | ||
Device is indicated for the | |||
temporary relief of minor muscle | |||
aches and pains and for | |||
temporary increase in circulation | |||
to the treated areas in people who | |||
are in good health. The Air | |||
Compression Therapy Device | |||
simulates kneading and stroking | |||
of tissues by using an inflatable | |||
garment. | Same | ||
Rx or OTC | OTC | OTC | Same |
Pressure range | 0~240mmHg | 0~240mmHg | Same |
Inflation Time | 3-30s | 3-30s | Same |
Deflation | |||
Time | 1-5s | 1-5s | Same |
Treatment | |||
time | 20 minutes | 20 minutes | Same |
Standard | ES 60601-1; | ||
IEC60601-1-2; | |||
ISO 10993-5: | |||
ISO 10993-10; | |||
IEC 60601-1-11 | ES 60601-1; | ||
IEC60601-1-2; | |||
ISO 10993-5: | |||
ISO 10993-10; | |||
IEC 60601-1-11 | Same | ||
Mode of | |||
compression | Sequential | Sequential/ Peristaltic | Same |
Power source | 100~240V 50/60Hz | 100~240V 50/60Hz | Same |
Power | |||
consumption | 24W | 12W | Similar |
Size and | |||
appearance of | Thighs: | Leg: | |
sleeves (leg | |||
part) | One size: 111624.9inch+8*33.5inch | One size: 73*26cm | |
Photo | Image: RP-ALM070H RP-ALM071H | Image | Similar |
Housing | |||
materials | Molded ABS enclosure | Molded ABS enclosure | Same |
Number of | |||
chambers | 3 | 3 | Same |
Work mode | RP-ALM070H: | ||
Six models (3 combine massage modes and | |||
3 Separate Massage Modes) | |||
C1: Massage full legs. | |||
C2: Massage feet and calves. | |||
(It can be used individually, and don't need | Mode 1: | ||
Starting with the foot chamber | |||
and progressing up the thigh | |||
chamber, each section | |||
compresses and the pressure | |||
gradually rises to the | Although the subject device provides 6 kinds of | ||
work mode, the Mode F1, F2, F3, M1, M2, M3 are | |||
the similar with predicate device (K193354), while | |||
the other work modes of subject device just have | |||
to connect the air hose of thighs wraps.) | |||
C3: Massage feet, calves, thighs by turn. It | |||
will be turn off When you press the Button | |||
C again. | |||
T: Massages thighs | |||
C: Massages calves | |||
F : Massages feet. It will be turn off when | |||
you press the Button S again. |
RP-ALM071H:
Five models(3 combine massage modes and
2 Separate Massage Modes):
Combine 1: Massage feet and calves
Combine 2: Massage from calves to feet
Combine 3: Massage from feet to calves.
Press it again to turn off combine message
function.
F: Massage feet.
C: Massage calves. Press it again to turn off
separate massage function. | pre-determined air pressure level,
then decompresses and the
air pressure drops.
Once the thigh section
decompresses, the cycle begins
again.
Mode 1 follows this pressure
sequence:
Image: Diagram of pressure sequence
Mode 2:
Starting with the foot chamber
and progressing up the thigh,
each section compresses and the
pressure gradually rises to the
pre-determined air pressure level,
holds the air until the entire
garment is compressed.
All three sections then
decompress simultaneously and
the air pressure drops, then cycle
begins again.
Mode 2 follows this pressure | same under different work modes, so the difference
of pressure range would not raise adversely impact
on safety and effectiveness. | |
| | | sequence: | |
| | | | |
| | | Mode 3:
include two stage, stage 1: it
work according to the method of
mode 1, after the stage 1 is
completed, it go to stage
2(working according to the
method of mode 2) without
interruption time until finish the
stage 2, then enter next cycle
without interruption.
Model1 Mode2
The pressure sequence of mode 3
combines mode 1 and mode 2 | |
| Safety feature | Button on display allows user to stop or
pause therapy session at any time | Button on display allows user to
stop or pause therapy session at
any time | Same |
| Technology | Compressor and valve system which
sequentially inflates inflatable chambers | Compressor and valve system
which sequentially inflates
inflatable chambers | Same |
7
8
9
10
11
| Operating
environment | Temperature: 5°C-40°C,
Humidity: 15%-90% | Temperature: 5°C -40°C,
Humidity: 5%-90%
non-condensing | Same |
|----------------------------------------------|--------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|---------|
| Transportation
and Storage
environment | Temperature: - 25°C70°C55°C;
Humidity:15%-90% non condensing
Atmospheric Pressure:75kPa-106kPa | Temperature: -20°C
Humidity: 5%-90% non
condensing
Atmospheric Pressure:75kPa-
106kPa | Similar |
Table 6B: Summary of Comparison
Subject Device | Predicate Device | Differences Discussion | |
---|---|---|---|
Device name | Leg Massager (Models:RP-ALM070H, | ||
RP-ALM071H) | FLOWpresso | N/A | |
510(k) number | K232965 | K223729 | N/A |
Manufacturer | Shenzhen Ruiyi Business Technology Co., | ||
Ltd. | Medella Health Ltd | N/A | |
Product | |||
regulation | 21 CFR 890.5650 | 21 CFR 890.5650 | Same |
Classification | |||
name | Massager, Powered Inflatable Tube | Massager, Powered Inflatable Tube | Same |
Regulation class | 2 | 2 | Same |
Product code | IRP, IRT | IRP, IRT | Same |
Indications | |||
for Use | Leg Massager (Models:RP-ALM070H, | ||
RP-ALM071H) is intended for home to | |||
temporarily relieve minor muscle aches | The Flowpresso Device combines heat | ||
and compression therapy. | |||
Flowpresso is intended to treat post | Similar, although the predicate device is | ||
intended to treat post traumatic and | |||
post-surgical medical and/or surgical | |||
and/or pains, and to temporarily increase | |||
circulation to the treated areas in people | |||
who are in good health. | traumatic and post-surgical medical | ||
and/or surgical conditions for which | |||
localized thermal therapy are indicated. | |||
Flowpresso is intended to be used by, or | |||
on the order of, licensed health care | |||
professionals based in rehabilitation | |||
facilities, outpatient clinics and athletic | |||
training settings. | conditions for which localized thermal | ||
therapy are indicated, the subject device | |||
is ,the working principles and expected | |||
outcomes of the two are consistent. All of | |||
them relieve pain by combining heat | |||
therapy with air compression on the | |||
affected area (subject device only covers | |||
the thighs, calves, and feet, predictive | |||
device also includes the back/hip), allowing | |||
patients to relax. | |||
Number of | |||
patients that | |||
can be treated | |||
at one time | One | One | Same |
Treatment | |||
Time | 20 minutes | 01-40 minutes | Different, Although the “Treatment |
Time” of the subject device is different | |||
than the predicate devices, the treatment | |||
time of the subject device is included in the | |||
predicate device. | |||
Heat Therapy | Not exceeding 45°C | Default: 86°F - 104°F | Similar, the maximum temperature of the |
subject device is slightly higher than that of | |||
the predicate device. The heating program | |||
of the subject device can be selected to be | |||
turned on or not, and all have passed IEC | |||
testing. Small differences do not affect | |||
safety and effectiveness. | |||
Compression | |||
Pressure | |||
Levels | 0〜240mmHg | 5-130mm/Hg | Different, pressure range is dependent on |
patient comfort levels and are adjusted in | |||
real-time. | |||
Static or | |||
Intermittent | |||
Pressure | Both | Both | Same |
Power Down | Available | Available | Same |
Heating | |||
Mechanism | Thermotherapy | Thermotherapy | Same |
User Interface | Touch Screen | Touch Screen | Same |
Line voltage | 100~240V | 110-220V | Similar |
Line | |||
Frequency | 50/60Hz | 50/60 Hz | Same |
Electrical | |||
Safety | |||
Standards | ES 60601-1; | ||
IEC60601-1-2; | |||
IEC 60601-1-11 | ANSI/AAMI ES60601- | ||
1:2005 + A1:2012 | |||
IEC 60601-1:2005 + A1:2012 | |||
EN60601-1:2006 +A1:2013 | Similar | ||
Operating | |||
Temperature | 5°C- 40°C | 60°F - 80°F (16°C - 27°C) | Different, Although the "Operating |
Temperature" | |||
of the subject device is | |||
different than the predicate devices, they | |||
comply with ANSI/AAMI ES60601-1, so | |||
the difference does not affect the safety and | |||
effectiveness. | |||
Storage | |||
Temperature | - 25°C~70°C | 33°F - 122°F (1°C - 50°C) | Different, Although the "Storage |
Temperature" | |||
of the subject device is | |||
different than the predicate devices, they | |||
comply with ANSI/AAMI ES60601-1, so | |||
Operating | |||
Humidity | 15%-90% | Below 60% Noncondensing | the difference does not affect the safety and |
effectiveness. | |||
Different, Although the "Operating | |||
Humidity" of the subject device is different | |||
than the predicate devices, they comply | |||
with ANSI/AAMI ES60601-1, so the | |||
difference does not affect the safety and | |||
effectiveness. | |||
Storage | |||
Humidity | 15%-90% | Below 60% Noncondensing | Different, Although the "Storage |
Humidity" of the subject device is different | |||
than the predicate devices, they comply | |||
with ANSI/AAMI ES60601-1, so the | |||
difference does not affect the safety and | |||
effectiveness. | |||
Operating | |||
Atmospheric | |||
Pressure and | |||
Altitude | 75kPa-106kPa | 700 hPa – 1060 hPa | |
(corresponds to a max. elevation | |||
of 9,842 ft. 6 in (3000m)) | Similar | ||
Types of | |||
Garments | Sleeves for thighs, calves, and feet | Various anatomical | |
thermal/compression garments | |||
for: Back/Hip, arm, leg, feet | Different, the type of garments of the | ||
predicate device exceeds that of the subject | |||
device, which does not include the back/ip | |||
part. So the risk of subject device is smaller | |||
than that of predicate device. | |||
Patient | |||
Contacting | |||
Material | All encompassed with a Nylon with a | ||
Polyurethane laminate material. | Needle cotton, non-conductive | ||
and non-woven lining. All | |||
encompassed with a PVC | |||
material. | Different, Although the material of the | ||
subject device is different than the predicate | |||
devices, they comply with ISO10993-5 and | |||
ISO10993-10, so the difference does not | |||
affect the safety and effectiveness. | |||
Multi-Patient | |||
Use and Single | |||
Patient Use | |||
Wraps | Multi-Patient Use | Multi-Patient Use | Same |
Biocompatibility | Cytotoxicity testing per ISO | ||
10993-5 | |||
Sensitization testing per ISO | |||
10993-10 | |||
Irritation testing per | |||
ISO 10993-10 | Cytotoxicity testing per ISO | ||
10993-5 | |||
Sensitization testing per ISO | |||
10993-10 | |||
Irritation testing per | |||
ISO 10993-10 | Same | ||
Sterile/Non | |||
Sterile | Non-sterile only | Non-sterile only | Same |
Cleaning | |||
Disinfection | |||
Validation of | |||
Labeling | Yes- for Multi-Patient use | ||
reusable wraps | Yes- for Multi-Patient use | ||
reusable wraps | Same | ||
Human Factors | |||
testing to | |||
confirm | |||
intended users | |||
have found | |||
instructions for | |||
cleaning and | |||
disinfection | |||
easy to use | Yes- for Multi-Patient use | ||
reusable wraps | Yes- for Multi-Patient use | ||
reusable wraps | Same | ||
Expected Life of | Based on frequency of use and | Based on frequency of use and | Same |
Garments | continued functional | ||
performance | continued functional | ||
performance | |||
Validation of | |||
repeated | |||
cleaning and | |||
disinfection for | |||
reusable | |||
garments | Yes- for Multi-Patient use | ||
reusable wraps | Yes- for Multi-Patient use | ||
reusable wraps | Same |
12
13
14
15
16
17
Summarv of the technological characteristics of the device
The device meets all the applicable technical requirements of : IEC 60601-1-11: 2015 - Medical electrical equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment IEC 60601-1-2: 2014 - Medical electrical equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 -Medical electrical equipment - Part 1: General requirements for basic safety and essential performance IEC TS 60601-4-2: Medical electrical equipment - Part 4-2: Guidance and interpretation -Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems ISO 10993-5: 2009 - Biological Evaluation of Medical Device - Part 5: Tests for in vitro Cytotoxicity ISO 10993-10: 2010 - Biological Evaluation of Medical Devices - Part 10: Tests for irritation and skin sensitization
Conclusion
Based on the indications for use, technological characteristics, and non-clinical performance data, "Leg Massager (Models:RP-ALM070H, RP-ALM071H)" is as safe, as effective, and performs as well as the legally marketed predicate devices. Therefore, the subject device is substantially equivalent to the predicate device.