K Number
K232888

Validate with FDA (Live)

Date Cleared
2023-12-21

(94 days)

Product Code
Regulation Number
874.1820
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The disposable laryngeal electrodes are intended to be used as a disposable, self-adhesive electrodes attached to an endotracheal tube and positioned for continuous EMG monitoring of the larynx during surgical procedures.

Device Description

The Disposable Laryngeal Electrodes are single used electrodes constructed from an medical-grade ink with a conductive polymer film covered by an insulating coating, a connector made of polypropylene and a cable assembly. The electrodes are designed to monitor the recurrent nerve during thyroid, anterior cervical, carotid endartecrectomy surgery and vagus nerve monitoring during brain surgery. The monitoring is performed with a surface electrode that is attached to an endotracheal tube placed on the vocal cord. A neutral adhesive electrode is placed on the patient's shoulder. Both adhesive electrodes are connected to a reusable recording cable and to the nerve monitor.

AI/ML Overview

The provided document is a 510(k) summary for a medical device called "Disposable Laryngeal Electrodes." It describes the device, its intended use, and the testing performed to demonstrate its substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and the study conducted, structured according to your request:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Adhesive propertyThe electrode should be able to be firmly adhered.Pass
Electrode ImpedanceEach path of the electrode should be able to conduct, impedance: ≤100 Ω.Pass
Security (Electrical)Normal state (μΑ): AC: ≤0.01mA (10μA) and DC: ≤0.1mA(100μA); Application part pressurized state (mA): ≤5mAPass
Sterility testingThe products were sterile both before and after aged.Pass
In vitro cytotoxicityNon-cytotoxic (as per ISO 10993-5)Non-cytotoxic
Skin sensitizationNon-sensitive (as per ISO 10993-10)Non-sensitive
Intracutaneous reactivityNon-irritation (as per ISO 10993-23)Non-irritation
Acute systemic toxicityNon-acute systemic toxicity (as per ISO 10993-11)Non-acute systemic toxicity
PyrogenNon-pyrogenic (as per ISO 10993-11)Non-pyrogenic
Sterilization ETOSterility Assurance Level (SAL) of 10^(-6) demonstrated per ISO 11135:2014; Meets EO residuals per ISO 10993-7Demonstrated and met
Shelf-life3 years (based on accelerated aging testing)Established 3 years
Electrical safetyMeets requirements of IEC 60601-1Met requirements

2. Sample size used for the test set and the data provenance

The document does not specify the sample sizes for the individual tests (adhesive property, electrode impedance, security, sterility, biocompatibility tests). It only states that testing was performed.

The data provenance is from non-clinical laboratory testing conducted by Suzhou Haishen United Medical Device Associates Co., Ltd. The document does not specify the country of origin of the data beyond the manufacturer's location (Suzhou, Jiangsu Province, China). The studies are by nature prospective as they are conducted specifically for the purpose of demonstrating device performance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable to this type of device and study. The testing performed is bench testing (e.g., electrical, mechanical, biological) against established engineering and biological standards, not clinical performance requiring expert interpretation of diagnostic outputs or ground truth establishment by clinical experts.

4. Adjudication method for the test set

Not applicable. As the studies are bench tests against objective criteria and standards, there is no need for expert adjudication methods like 2+1 or 3+1. The results are quantitative measurements or pass/fail determinations based on predefined criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done. This device is a passive electrode for EMG monitoring during surgery, not an AI-powered diagnostic or assistive tool for human readers. Therefore, this section is not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is an electrode, which is a hardware component. There is no algorithm or AI component to be evaluated in a standalone manner.

7. The type of ground truth used

The "ground truth" for the non-clinical tests are established international standards and internal requirements for medical devices. Specifically:

  • Performance (electrical/physical): Internal requirements, and likely general engineering principles for electrodes.
  • Biocompatibility: ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-11, ISO 10993-23, and FDA guidance on ISO 10993-1.
  • Sterilization: ISO 11135:2014 and ISO 10993-7.
  • Electrical safety: IEC 60601-1.

8. The sample size for the training set

Not applicable. This device is a hardware component and does not involve AI or machine learning models that require a training set.

9. How the ground truth for the training set was established

Not applicable. As there is no training set, there is no ground truth to establish for it.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".

December 21, 2023

Suzhou Haishen Medical Device Associates Co., Ltd % Xiaoqing Xue Registration Engineer Sinow Medical AS Høyteknologisenteret Thormøhlens Gate 55 Bergen, 5008 Norway

Re: K232888

Trade/Device Name: Disposable Laryngeal Electrodes Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: ETN Dated: November 24, 2023 Received: November 24, 2023

Dear Xiaoqing Xue:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shuchen Peng -S

Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K232888

Device Name Disposable Laryngeal Electrode

Indications for Use (Describe)

The disposable laryngeal electrodes are intended to be used as a disposable, self-adhesive electrodes attached to an endotracheal tube and positioned for continuous EMG monitoring of the larynx during surgical procedures.

Type of Use (Select one or both, as applicable)

For activities under 21 CFR 312.2, Sponsor-InvestigatorCompassionate Use, including Treatment Use
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

K232888

1. Submitter

Submitter: Suzhou Haishen United Medical Device Associates Co., Ltd Address: 3F, 4F 401, 4F 402, 5F, Building 10, No. 168 Majian Road, Suzhou, Jiangsu Province, China Telephone: +86 18921623557 Email: 785445240@gq.com Contact Person: Leyi Dai Date Prepared: Dec.15, 2023

2. Designated Submission Correspondent

Company: Sinow Medical AS Address: Vestre Fantoftåsen 44, 5072, Bergen, Norway Contact Person: Xiaoqing Xue Telephone: +86 15161196032 Email: xue(@bergemed.com

3. Subject Device

Trade name: Disposable Laryngeal Electrodes 510(K) number: K232888 Regulatory Class: II Regulation Number: 21 CFR 874.1820 Product Codes: ETN Classification Panel: Ear Nose & Throat Classification Name: Stimulator, Nerve

4. Primary Predicate Device

Manufacturer: Inomed Medizintechnik GmbH Trade name: Inomed Adhesive Laryngeal Electrode 510(K) Number: K091874 Regulation Number: 21 CFR 874.1820 Product Codes: ETN Classification Panel: Ear Nose & Throat Classification Name: Stimulator, Nerve

5. Device Description

The Disposable Laryngeal Electrodes are single used electrodes constructed from an medical-grade ink with a conductive polymer film covered by an insulating coating, a connector made of polypropylene and a cable assembly. The electrodes are designed to monitor the recurrent nerve during thyroid, anterior cervical, carotid endartecrectomy surgery and vagus nerve monitoring during brain surgery. The monitoring is performed with a surface electrode that is attached to an endotracheal tube placed on the vocal cord. A neutral adhesive electrode is placed on the patient's

{4}------------------------------------------------

shoulder. Both adhesive electrodes are connected to a reusable recording cable and to the nerve monitor.

6. Intended Use/Indications for Use

The Disposable Laryngeal Electrodes are intended to be used as a disposable, self-adhesive electrode attached to an endotracheal tube and positioned for continuous EMG monitoring of the larynx during surgical procedures.

AttributeDisposable LaryngealElectrode (SubjectDevice)Inomed AdhesiveLaryngeal Electrode(Predicate deviceK091874)Comment
RegulationNumber21 CFR 874.182021 CFR 874.1820Same
Regulation NameSurgical nervestimulator/locatorSurgical nervestimulator/locatorSame
Regulatory ClassClass IIClass IISame
Product CodeETNETNSame
Indications forUseThe Disposable LaryngealElectrodes are intended tobe used as a disposable,self-adhesive electrodeattached to an endotrachealtube and positioned forcontinuous EMGmonitoring of the larynxduring surgical procedures.The Inomed AdhesiveLaryngeal Electrodes areintended to be used as adisposable, self-adhesiveelectrode attached to anendotracheal tube andpositioned for continuousEMG monitoring of thelarynx during surgicalprocedures.Same
Monitoring siteTrachea/larynxTrachea/larynxSame
Monitoring typeContinuous EMGmonitoringContinuous EMGmonitoringSame
May be used withall commercialEMG unitsYesYesSame
Method ofelectrodeattachmentAttached to the surface ofthe endotracheal tubeAttached to the surface ofthe endotracheal tubeSame
Number ofelectrodesutilized22Same
Number ofchannels2, 42Different 1
Device design-Medical grade inkssuspended in a polyestersubstrate-Medical grade inkssuspended in a polyestersubstrateSame
-polypropylene connector-cable assembly-polypropylene connector-cable assembly
ElectricalinsulationElectrical insulation on allsurfaces until the head ofthe electrodeElectrical insulation on allsurfaces until the head ofthe electrodeSame
Single use onlyYesYesSame
SafetycharacteristicsNon-invasiveNon-invasiveSame
SterilizationETOETOSame

7. Comparison of Technological Characteristics

{5}------------------------------------------------

Discussion of difference !: The subject device included two and four channels type.The predicate device only have two channels type. Although differences between numbers of channel, the subject two and four channel device have the same design, materials and intended performance, only increasing two identical channels. Furthermore, the four channel device is selected as representative sample to performed electrical safety testing according to IEC 60601-1 to ensure meet the electrical safety requirement. Therefore, we considered that this difference does not raise new questions of safety or effectiveness.

Non-Clinical Testing 8.

8.1. Performance Testing-Bench

Suzhou Haishen United Medical Device Associates Co., Ltd has performed the following non-clinical laboratory testing to determine substantial equivalence.

TestAcceptable criteriaResult
Adhesive propertyThe electrode should be able to be firmly adhered.Pass
Electrode ImpedanceEach path of the electrode should be able to conduct,impedance: ≤100 Ω .Pass
SecurityNormal state (μΑ):AC: ≤0.01mA (10μA) and DC: ≤0.1mA(100μA)Application part pressurized state (mA): ≤5mAPass
Sterility testingThe products were sterile both before and after aged.Pass

The essential performance and safety of Disposable Laryngeal Electrode was tested for performance in accordance with internal requirements. The adhesive property, electrode Impedance and security were evaluated for subject and predicate device to ensure that the subject device performed as intended and supported substantial equivalence to the predicate devices.

8.2. Biocompatibility testing

The contact classification for the Disposable Laryngeal Electrode of the subject device is a surface contact device that directly contact with the mucosal membrane ( ≤ 24h). The biocompatibility evaluation for the subject device was conducted in accordance with ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and in accordance with FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", Guidance for Industry and Food and Drug Administration Staff. The results of biocompatibility

{6}------------------------------------------------

testing included in the table below demonstrate that the device meets biological safety requirements per ISO 10993-1 for mucosal membrane contact device.

TestStandardResult
In vitro cytotoxicityISO 10993-5Non-cytotoxic
Skin sensitizationISO 10993-10Non-sensitive
Intracutaneous reactivityISO 10993-23Non- irritation
Acute systemic toxicityISO 10993-11Non-acute systemic toxicity
PyrogenISO 10993-11Non-pyrogenic

8.3. Sterilization and shelf-life testing

The subject device is sterilized using Ethylene Oxide. The sterilization validation has been performed in accordance with the principles of ISO 11135:2014 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices. A sterility assurance level (SAL) of 10th has been demonstrated. The device meets EO residuals per ISO 10993-7.

A shelf-life of 3 years has been established based on accelerated aging testing.

8.4. Electrical safety testing

Suzhou Haishen United Medical Device Associates Co., Ltd has performed the electrical safety testing according to IEC 60601-1 Medical Electrical Equipment Part 1: General requirements for basic safety and essential performance. Testing were performed and confirmed that the subject device is meet the requirement.

9. Conclusion

The Disposable Laryngeal Electrode is substantially equivalent to the predicate device when evaluating intended use and technological characteristics. The subject device has the identical intended use as the predicate device and only has a minor technological difference that does not raise different questions of safety or effectiveness. The nonclinical performance data submitted in the 510(k) demonstrates the subject device is substantially equivalent to the predicate device.

§ 874.1820 Surgical nerve stimulator/locator.

(a)
Identification. A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.(b)
Classification. Class II.