(83 days)
Electronic Blood Pressure Monitor intended for use in measuring blood pressure and pulse rate in adult patient population. It is intended to measure the diastolic, systolic blood pressures and pulse rate through an inflatable cuff wrapped around the arm. It can be used by medical professionals or at home. The cuff circumference is limited to 22-42 cm.
Electronic Blood Pressure Monitor mainly consist of the main body (include screen display, air tube connector, memory button and start/stop button), cuff, USB cable, air tube, and AA batteries. Electronic Blood Pressure Monitor that uses the oscillometric principle to measure your blood pressure and pulse rate. The radial artery in the arm changes from blocked to open as the pressure in the cuff tied around the arm changes from high to low, causing the pressure in the cuff to be superimposed on a series of small pressure pulses. The sphygmomanometer senses these signals and, after certain calculations, finds the systolic and diastolic pressures of the radial artery in the body. Electronic Blood Pressure Monitor can be divided into three models (YJ320, YJ321E, YJ326E) according to their appearance and functions.
The provided text is for an FDA 510(k) clearance for an Electronic Blood Pressure Monitor. It details non-clinical and clinical testing performed to demonstrate substantial equivalence to a predicate device.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
The document provides a comprehensive table of non-clinical performance and safety acceptance criteria, along with the reported "Conclusion" (Pass/Fail). For the clinical validation, it states the acceptance criteria are met based on deviations from a reference standard.
Table of Acceptance Criteria and Reported Device Performance (Non-Clinical):
Inspecting Item | Acceptance Standard (Simplified) | Reported Device Performance (Conclusion) |
---|---|---|
1. Identification requirement | Meets IEC 60601-1:2005 in 7.2. | Pass |
2. Life Span | After at least 10,000 full scale cycles, the sphygmomanometer should still meet safety and performance requirements. A full scale cycle: pressure rises from = 0.267kPa/s (2mmHg/s). 5.2.3 Venting: Rapid venting from 34.67kPa (260mmHg) to 2kPa (15mmHg) in |
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).