K Number
K232400

Validate with FDA (Live)

Device Name
VariSeed (v10)
Date Cleared
2023-09-08

(29 days)

Product Code
Regulation Number
892.5050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VariSeed is intended for use as a software application used by medical professionals to plan, guide, optimize, and document low dose rate brachytherapy and procedures based on template guided needle insertion.

VariSeed is indicated for use as a treatment planning software application used by medical professionals to plan, guide, optimize and document low-dose-rate brachytherapy procedures and for use as a biopsy procedure tracking software application used by medical professionals to plan, guide, and document biopsy procedures based on template guided needle insertion. VariSeed may be used on any patient considered suitable for this type of treatment and is intended to be used outside of the sterile field in an operating room environment or in a normal office environment.

Device Description

VariSeed 10 is a free-standing PC based treatment planning software designed for preoperative and intraoperative planning of LDR implants, tracking of the implant procedure, and postoperative evaluation of completed implants. VariSeed also provides tools for supporting intraoperative template guided biopsy and using those results to guide future treatment.

AI/ML Overview

The provided text is an FDA 510(k) clearance letter for Varian Medical Systems' VariSeed (v10) device. It states that the device is substantially equivalent to a legally marketed predicate device (VariSeed v9.0).

However, the core of your request is to describe the acceptance criteria and the study that proves the device meets those criteria, specifically concerning a study that evaluates device performance through quantitative measures, such as accuracy metrics, and details about the test set, expert involvement, and ground truth establishment.

The provided FDA letter explicitly states: "No clinical tests have been included in this pre-market submission." and "Verification testing was performed to demonstrate that the performance and functionality of the VariSeed v10 treatment planning software meets the design input requirements. Validation testing was performed on production equivalent devices, under clinically representative conditions and by qualified personnel."

This indicates that the clearance was based on non-clinical performance data and verification/validation testing against design requirements, rather than a clinical study evaluating the device's diagnostic performance (e.g., accuracy against ground truth in a clinical setting).

Therefore, I cannot provide the detailed information you requested about acceptance criteria and a study that proves the device meets the acceptance criteria in the context of clinical performance, sample sizes, expert involvement, or ground truth for a clinical test set, because such a study was not included in this submission as per the document.

The "acceptance criteria" referred to in the document are implicitly the design input requirements and the successful completion of verification and validation against those requirements.

Here's a breakdown of what can be inferred from the document and why most of your requested points cannot be answered:

1. A table of acceptance criteria and the reported device performance:

  • Acceptance Criteria (Implied): The device meets its design input requirements and performs its intended functions (planning, guiding, optimizing, and documenting low dose rate brachytherapy and biopsy procedures, including support for PET and MR images).
  • Reported Device Performance: "Verification testing was performed to demonstrate that the performance and functionality of the VariSeed v10 treatment planning software meets the design input requirements." And "Validation testing was performed on production equivalent devices, under clinically representative conditions and by qualified personnel."
  • Quantitative Metrics: The document does not provide specific quantitative performance metrics (e.g., accuracy, sensitivity, specificity, or error margins) often seen in studies evaluating AI or diagnostic devices. This implies the focus was on functional verification and validation against engineering specifications, not clinical outcomes or diagnostic accuracy.

2. Sample size used for the test set and the data provenance:

  • Not specified. Since no clinical tests were included, there is no information about a "test set" in the context of clinical data. The validation was likely performed on synthetic, phantom, or previously acquired (non-patient-identifiable) data to test software functionality and accuracy of calculations, rather than on a true patient "test set" for clinical performance evaluation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. No clinical ground truth establishment process is described given the lack of clinical study data.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not applicable.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. The document explicitly states "No clinical tests have been included in this pre-market submission." Therefore, no MRMC study or AI assistance evaluation was performed as part of this submission. The device is described as "treatment planning software," not an AI-driven diagnostic assistance tool in the way you might expect. The "AI" implied in your question (human readers improving with AI assistance) is not detailed or alluded to in this context.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • The software itself operates "stand-alone" in the sense that it performs its calculations and functions without constant human input for each step once initiated. However, its intended use is "by medical professionals to plan, guide, optimize, and document," implying it's a tool used by a human, not a fully autonomous diagnostic device. The document does not describe "algorithm only" performance in the context of clinical decision-making or diagnosis.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Not specified. For internal verification and validation of a treatment planning system, ground truth would typically refer to known geometric properties of phantoms, known dose distributions from physical measurements, or validated computational models. It would not typically involve pathology or outcomes data for this type of software clearance.

8. The sample size for the training set:

  • Not applicable. This section does not describe an AI/ML device that requires a distinct "training set" for model development. It's a software application following a traditional software development lifecycle, not a deep learning model requiring vast amounts of labeled training data.

9. How the ground truth for the training set was established:

  • Not applicable.

In summary, the FDA clearance for VariSeed (v10) was based on its substantial equivalence to its predicate device (VariSeed v9.0) and on "non-clinical data," specifically "verification and validation" against "design input requirements" under "clinically representative conditions." It was not based on a clinical study demonstrating performance against a diagnostic ground truth with human experts and patient data.

{0}------------------------------------------------

September 8, 2023

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Varian Medical Systems % Lynn Allman Director, Regulatory Affairs 3100 Hansen Way M/s E-110 PALO ALTO CA 94304

Re: K232400

Trade/Device Name: VariSeed (v10) Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: MUJ, KXK Dated: August 8, 2023 Received: August 10, 2023

Dear Lynn Allman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Locon Weidner

Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K232400

Device Name Variseed 10.0

Indications for Use (Describe)

VariSeed is intended for use as a software application used by medical professionals to plan, guide, optimize, and document low dose rate brachytherapy and procedures based on template guided needle insertion.

VariSeed is indicated for use as a treatment planning software application used by medical professionals to plan, guide, optimize and document low-dose-rate brachytherapy procedures and for use as a biopsy procedure tracking software application used by medical professionals to plan, guide, and document biopsy procedures based on template guided needle insertion. VariSed may be used on any patient considered suitable for this type of treatment and is intended to be used outside of the sterile field in an operating room environment or in a normal office environment.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

varian

Premarket Notification 510(k) Summary

K232400

The following information is provided according to 21 CFR 807.92.

Submitter:Varian Medical Systems, Inc.3100 Hansen WayPalo Alto, CA 94304Contact Name: Lynn AllmanE-mail: submissions.support@varian.comDate Prepared: Aug 8, 2023
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Trade/ProprietaryNames:VariSeed v10.0
Common/Usual Name:Predicate Device:
Brachytherapy treatment planning softwareVariSeed v9.0 (K150636)
Classification Name:
System, planning, radiation therapy treatment,21 CFR §892.5050
Product Code:
MUJ
Regulatory Class:
Class II
Device Description:VariSeed 10 is a free-standing PC based treatment planning software designed for preoperative and intraoperative planning of LDR implants, tracking of the implant procedure, and postoperative evaluation of completed implants. VariSeed also provides tools for supporting intraoperative template guided biopsy and using those results to guide future treatment.
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Intended/Indicationsfor UseStatement:Intended UseIndications for Use
VariSeed is intended for use asa software application used bymedical professionals to plan,guide, optimize, and documentlow dose rate brachytherapyand procedures based ontemplate guided needleinsertion.VariSeed is indicated for use as a treatmentplanning software application used by medicalprofessionals to plan, guide, optimize and documentlow-dose-rate brachytherapy procedures and foruse as a biopsy procedure tracking softwareapplication used by medical professionals to plan,guide, and document biopsy procedures based ontemplate guided needle insertion.VariSeed may be used on any patient consideredsuitable for this type of treatment and is intended tobe used outside of the sterile field in an operatingroom environment or in a normal office environment.

The purpose of this submission is to provide details on how the modified VariSeed v10.0 is similar to Varian's VariSeed v9.0 (K150636) for which we are claiming substantial equivalence.

The subject device Indications for Use is the same as the predicate device and the Intended Use is the same as the predicate device.

{4}------------------------------------------------

Comparison of Technological Characteristics with the Predicate Device

At a high level, the subject and predicate devices are based on the following similar elements:

  • Design, technological characteristics, operation and use as the Predicate device ●

Significant Difference

The significant differences (based on "Deciding When to Submit a 510(k) for a Software Change to an Existing Device. (October 2017)") compared to the predicate device are:

  • Provide support for PET images as secondary image volumes. ●
  • . Inclusion of MR Images in existing image import-based workflow for VariPath.

Performance Data

Verification and validation were conducted according to QSR §820.30 and ISO 13485:2016 design control requirements. Submission documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, Format for Traditional and Abbreviated 510(k)s.

Verification testing was performed to demonstrate that the performance and functionality of the VariSeed v10 treatment planning software meets the design input requirements. Validation testing was performed on production equivalent devices, under clinically representative conditions and by qualified personnel. International standards were incorporated into the device design and development.

No clinical tests have been included in this pre-market submission.

Standards Conformance

The subject device conforms in whole or in part with the following standards:

  • . ISO 14971:2019
  • IEC 62083:2009
  • IEC 62366-1:2015 +A1:2020 ●
  • . ISO 20417:2021
  • . ISO 15223-1:2021
  • IEC 62304-1:2006 +A1:2015 ●
  • IEC 82304-1:2016 ●

The subject device also complies with the following non-FDA recognized standard:

  • ISO 13485:2016

Conclusion

The non-clinical data for the VariSeed v10 treatment planning software supports the safety of the device compared to the predicate and the verification and validation demonstrate that the subject device should perform as intended in the specified use conditions. Varian considers the VariSeed v10 treatment planning software to be as safe and effective as the predicate, and therefore substantially equivalent to the predicate device.

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.