K Number
K232274
Manufacturer
Date Cleared
2023-09-28

(59 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The product is an over-the-counter device intended for removal of unwanted body and facial hair.

Device Description

IPL hair removal device is a personal, light-based, hair reduction device intended to be sold over-thecounter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology. It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains a Xenon lamp and a skin tone sensor to detect appropriate skin contact. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot be triggered a pulse emitting.

The IPL hair removal device includes LS101 and LS102 two models. Their intended use, performance, structure design and operation are identical, with the different color of grip part. The difference does not affect or change the intended use of the device.

AI/ML Overview

The provided document is a 510(k) Premarket Notification for an IPL hair removal device. It focuses on demonstrating substantial equivalence to predicate devices based on technical specifications and safety standards, rather than clinical performance or AI/software-based diagnostics. Therefore, much of the information requested, particularly regarding acceptance criteria for diagnostic performance, AI study details, expert review, and ground truth establishment in a medical imaging context, is not present in this document.

However, I can extract the information that is available:

1. Table of Acceptance Criteria and Reported Device Performance:

This document does not define explicit "acceptance criteria" in the sense of a clinical trial or AI model performance metrics (like sensitivity, specificity, AUC). Instead, the acceptance is based on demonstrating substantial equivalence to legally marketed predicate devices by confirming the subject device meets the same safety and performance standards.

The table below summarizes the comparison elements used to demonstrate substantial equivalence, which implicitly act as performance criteria for the device's technical specifications. The "Reported Device Performance" column reflects the subject device's specifications.

Elements of Comparison (Implicit Acceptance Criteria)Reported Device Performance (Subject Device: IPL hair removal device, Model: LS101, LS102)Predicate Device 1: iPulse SmoothSkin Gold Hair Removal DevicePredicate Device 2: PerfectSmoothRemarks (from document)
Indications for UseRemoval of unwanted body and facial hair (OTC)Removal of unwanted hair (OTC)Removal of unwanted hair (OTC), e.g., underarm, facial hair below chin, legs.SE (Substantially Equivalent). Different wording, but similar actual use.
Wavelength range523nm ~ 1100nm510-1100nm≥510nmSE. Both comply with IEC 60601-1, IEC 60601-2-57.
Energy mediumXenon Arc FlashlampXenon Arc FlashlampXenon Arc FlashlampSE
Energy density3.2-6.5J/cm²3-6J/cm²4.7 J/cm²SE. Complies with IEC 60601-1, IEC 60601-2-57. Subject device range overlaps with Predicate 1.
Spot sizeStandard care head: 3cm² (30mm10mm)Precision care head: 1.2cm² (12mm10mm)3cm² (3cm by 1cm)4.5 cm²SE. Both comply with IEC 60601-1, IEC 60601-2-57.
Pulse duration<10 ms2ms to 10ms11-13 msSE. Complies with IEC 60601-1, IEC 60601-2-57. Subject device duration almost identical to Predicate 1.
Pulsing controlFinger switchFinger switchFinger switchSE
Delivery deviceDirect illumination to tissueDirect illumination to tissueDirect illumination to tissueSE
Location for useOTC (Over-The-Counter)OTCOTCSE
Biocompatibility (Standards met)ISO 10993-5, ISO 10993-10ISO 10993-5, ISO 10993-10ISO 10993-5, ISO 10993-10SE
Electrical Safety (Standards met)IEC 60601-1, IEC 60601-1-11, IEC 60601-2-57IEC 60601-1, IEC 60601-1-11, IEC 60601-2-57IEC 60601-1, IEC 60601-1-11, IEC 60601-2-57SE
EMC (Standard met)IEC 60601-1-2IEC 60601-1-2IEC 60601-1-2SE
Eye Safety (Standard met)IEC 62471IEC 62471IEC 62471SE

2. Sample size used for the test set and the data provenance:

  • OTC Usability Study: A total of 20 participants were enrolled in this study.
  • Data Provenance: Not specified, but likely associated with the manufacturer (Lumenessa Corp) and its R&D activities, potentially in China (where the correspondent is located). The study is described as a "Company" study. It's a prospective study designed to gather user feedback.
  • Other Testing (Biocompatibility, Electrical Safety, Eye Safety): These were standard lab bench tests against recognized international standards (ISO, IEC). The "sample size" here refers to the materials or prototypes tested, which is not numerically specified but implied as adequate to meet the standard requirements.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • OTC Usability Study: This study focused on user understanding and operation, not on establishing a "ground truth" for a medical condition. It mentions gathering data from "layperson" participants. No medical experts were explicitly involved in this usability study for ground truth establishment.
  • Other Testing: These involved adherence to established engineering and safety standards. Experts would be engineers and lab technicians ensuring compliance, but not in the context of clinical "ground truth" for disease.

4. Adjudication method for the test set:

  • OTC Usability Study: The document states the study "was completed successfully and the results were no issues or negative comments." This implies a qualitative assessment of user feedback rather than a formal adjudication process (like 2+1 or 3+1 consensus).
  • Other Testing: Adjudication is not applicable as these are compliance tests against objective standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This device is an IPL hair removal device, not an AI-assisted diagnostic tool. Therefore, an MRMC study and AI-related performance metrics are not applicable and were not conducted or reported.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • No. This is not an AI algorithm device. Performance is based on mechanical, electrical, optical and safety specifications.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • For the OTC Usability Study, the "ground truth" was about user comprehension and successful operation of the device by laypersons, rather than a clinical diagnosis. The outcome (no issues or negative comments) served as the "ground truth" regarding usability.
  • For technical performance and safety tests, the "ground truth" was adherence to established international and national standards (ISO, IEC) for medical devices. This is an engineering ground truth, not a clinical one.

8. The sample size for the training set:

  • Not applicable. This device does not involve a "training set" in the machine learning sense.

9. How the ground truth for the training set was established:

  • Not applicable. As there is no training set, there is no ground truth establishment for it.

In summary: The provided document is a 510(k) submission for a non-AI, over-the-counter IPL hair removal device. The acceptance criteria and "studies" described focus on demonstrating the device's technical specifications comply with recognized safety standards and that it is substantially equivalent to existing predicate devices, along with a basic usability study for laypersons. It does not provide information relevant to AI diagnostic performance, clinical ground truth establishment, or multi-reader studies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 28, 2023

Lumenessa Corp % Xuexia Ren Director Chengdu Office Delta Technology Service (Shenzhen) Co., Ltd. 1-01, Building 1&1-01 A-01, Building3, No.15 Jinhui Rd. Kengzi Subdistrict, Pingshan District Shenzhen, 518118 China

Re: K232274

Trade/Device Name: IPL hair removal device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: July 28, 2023 Received: July 31, 2023

Dear Xuexia Ren:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Tanisha L. Tanisha L. Hithe -S 2023.09.28 Hithe -S 10:57:51 -04'00'

Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232274

Device Name IPL hair removal device

Indications for Use (Describe)

The product is an over-the-counter device intended for removal of unwanted body and facial hair.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Chapter 6, 510(k) Summary

510(k) Summary K232274

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

1. Submitter's Information

  • 510(k) Owner's Name: Lumenessa Corp �
  • Establishment Registration Number: Applying �
  • Address: 10606 Shoemaker Ave #A, Santa Fe Springs, CA 90670 �
  • Tel: 1-408-5623072121 �
  • � Fax: /
  • � Contact Person: Peter Kim (Vice President)
  • Email: peter@lumenessa.com
  • � Date Prepared: September 26, 2023

Application Correspondent:

  • Contact Person: Ms. Ren Xuexia �
  • � Delta Technology Service(Shenzhen) Co., Ltd.
  • Address: 1-01, Building 1&1-01 A-01, Building3, No.15 Jinhui Rd., Kengzi Subdistrict, Pingshan � District, Shenzhen, 518118, China
  • � Tel: +86-18030755441
  • � Email: renxuexia@deltates.com

2. Subject Device Information

  • � Common Name: Light Based Over-The-Counter Hair Removal
  • Trade Name: IPL hair removal device (Model: LS101, LS102) �
  • � Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
  • � Review Panel: General & Plastic Surgery
  • Product Code: OHT �
  • � Regulation Number: 878.4810
  • � Requlation Class: II

3. Predicate Device Information Predicate Device 1 Information

  • � Common Name: Light based over the counter hair removal system
  • 510(k) Number: K160968
  • � Sponsor: CyDen Limited.
  • Trade Name: iPulse SmoothSkin Gold Hair Removal Device �
  • Classification Name: Laser surgical instrument for use in general and plastic surgery and in � dermatology

{5}------------------------------------------------

  • � Review Panel: General & Plastic Surgery
  • � Product Code: OHT, GEX
  • Requlation Number: 878.4810 �
  • � Regulation Class: II

Predicate Device 2 Information

  • � Common Name: Light Based Over-The-Counter Hair Removal
  • 510(k) Number: K161428
  • � Sponsor: Shen Zhen CosBeauty Co., Ltd
  • � Trade Name: PerfectSmooth
  • Classification Name: Laser surgical instrument for use in general and plastic surgery and in � dermatology
  • � Review Panel: General & Plastic Surgery
  • � Product Code: OHT
  • � Regulation Number: 878.4810
  • � Regulation Class: II

4. Device Description

IPL hair removal device is a personal, light-based, hair reduction device intended to be sold over-thecounter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology. It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains a Xenon lamp and a skin tone sensor to detect appropriate skin contact. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot be triggered a pulse emitting.

The IPL hair removal device includes LS101 and LS102 two models. Their intended use, performance, structure design and operation are identical, with the different color of grip part. The difference does not affect or change the intended use of the device.

5. Intended Use / Indications for Use

The product is an over-the-counter device intended for removal of unwanted body and facial hair.

6. Test Summary

IPL hair removal device, Model: LS101, LS102 has been evaluated for the safety and performance by lab bench testing as following:

1) Biocompatibility Testing

The biocompatibility evaluation for the body-contacting components of the subject device was conducted in accordance with the "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", as recognized by FDA, including:

  • ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for In Vitro Cytotoxicity

  • A ISO 10993-10:2010 Biological evaluation of medical devices-Part 10:Tests for irritation and skin sensitization
  • A ISO 10993-10:2021 Biological evaluation of medical devices-Part 10:Tests for skin sensitization

{6}------------------------------------------------

Sponsor:Lumenessa Corp
Subject Device:IPL hair removal device, Model: LS101, LS102
Document Name:FDA 510(k) Submission Report

2) Electrical Safety and Eye Safety

  • A IEC 60601-1 Medical electrical equipment –Part 1: General requirements for basic safety and essential performance
  • IEC 60601-1-2 Medical electrical equipment –Part 1-2: General requirements for basic safety and A essential performance –Collateral standard: electromagnetic compatibility
  • A IEC 60601-1-11 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
  • A IEC 60601-2-57 Medical electrical equipment –Part 2-57: Particular requirements for the basic safety and essential performance of non-laser source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use

3) Eye Safety

  • IEC 62471 Photobiological safety of lamps and lamp systems

4) Software Verification and Validation

Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.

5) OTC Usability Study

An OTC Usability Study was enrolled a total of 20 participant and was conducted to provide a basis for the Company to gather data from the layperson relative to level of labeling understanding of what the device is, what the device does and how to operate the device. This Study was completed successfully and the results were no issues or negative comments.

7. Comparison to predicate device and conclusion

The technological characteristics, features, specifications, materials, and intended use of IPL hair removal device (Model: LS101, LS102) is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new question of safety or effectiveness.

Elements ofComparisonSubject DevicePredicate Device 1Predicate Device 2Remark
Device NameIPL hair removal deviceModel: LS101, LS102iPulse SmoothSkin GoldHair Removal DevicePerfectSmooth--
510(k) NumberK232274K160968K161428--
Indications for UseThe product is an over-the-counter deviceintended for removal ofunwanted body andfacial hair.The iPulse SmoothSkinGold Hair Removal Systemis an over the counterdevice intended for theremoval of unwanted hair.The PerfectSmooth is anover-the-counter deviceintended for removal ofunwanted hair such as butnot limited to small areassuch as underarm andfacial hair below the chinline and large areas suchas legs.SENote 1

{7}------------------------------------------------

Sponsor:Lumenessa Corp
Subject Device:IPL hair removal device, Model: LS101, LS102
Document Name:FDA 510(k) Submission Report
Elements ofComparisonSubject DevicePredicate Device 1Predicate Device 2Remark
Wavelength range523nm ~ 1100nm510-1100nm≥510nmSENote 2
Energy mediumXenon Arc FlashlampXenon Arc FlashlampXenon Arc FlashlampSE
Energy density3.2-6.5J/cm²3-6J/cm²4.7 J/cm²SENote 3
Spot sizeStandard care head:3cm²(30mm10mm)Precision care head:1.2cm²(12mm10mm)3cm² (3cm by 1cm)4.5 cm²SENote 2
Pulse duration<10 ms2ms to 10ms11-13 msSENote 4
Pulsing controlFinger switchFinger switchFinger switchSE
Delivery deviceDirect illumination totissueDirect illumination to tissueDirect illumination totissueSE
Location for useОТСОТСОТСSE
BiocompatibilityISO 10993-5,ISO 10993-10ISO 10993-5,ISO 10993-10ISO 10993-5,ISO 10993-10SE
Electrical SafetyIEC 60601-1,IEC 60601-1-11,IEC 60601-2-57IEC 60601-1,IEC 60601-1-11,IEC 60601-2-57IEC 60601-1,IEC 60601-1-11,IEC 60601-2-57SE
EMCIEC 60601-1-2IEC 60601-1-2IEC 60601-1-2SE
Eye SafetyIEC 62471IEC 62471IEC 62471SE

Comparison in Detail(s):

Note 1:

The "Indications for Use" of subject device is described in a different way than the predicate devices, but they show that the actual use of the device is similar, namely "removal of unwanted hair". So the differences of description will not raise any safety or effectiveness issue.

Note 2:

Although the "Wavelength range" and "Spot size" of subject device are a little different from the predicate devices, they all comply with IEC 60601-1, IEC 60601-2-57 requirement. So the differences of function specification will not raise any safety or effectiveness issue.

Note 3:

Although the "Energy density" of subject device is a little different from the predicate devices, they all comply with IEC 60601-1, IEC 60601-2-57 requirement. And the Energy density of subject device is in the range of the predicate device 1.

So the differences of function specification will not raise any safety or effectiveness issue.

Note 4:

{8}------------------------------------------------

Sponsor:Lumenessa Corp
Subject Device:IPL hair removal device, Model: LS101, LS102
Document Name:FDA 510(k) Submission Report

Although the "Pulse duration" of subject device is a little different from the predicate devices, they all comply with IEC 60601-1, IEC 60601-2-57 requirement. And the Pulse duration of subject device is almost identical to the predicate device 1.

So the differences of function specification will not raise any safety or effectiveness issue.

Summary:

The subject device "IPL hair removal device, model: LS101, LS102" is Substantial Equivalent to the predicate devices.

8. Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the subject device "IPL hair removal device, model: LS101, LS102" is to be concluded substantial equivalent to its predicate devices.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.