K Number
K232249
Date Cleared
2023-12-01

(126 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Gas/Water Feeding Valve OE-B14 is intended to be attached in place of Air/Water Feeding Valve (OE-B12) to PENTAX Medical GI endoscopes and used to feed the gas/water to the body (via the air/water channel).

Device Description

The Gas/Water Feeding Valve OE-B14 is intended to be attached in place of Air/Water Feeding Valve (OE-B12) to PENTAX Medical GI endoscopes and used to feed the gas/water to the body (via the air/water channel).

The OE-B14 is attached onto the air/water feeding cylinder of the control body of PENTAX Medical endoscope. The endoscope is linked to non-flammable gas cylinder and the water bottle assembly via air/water feeding circuit.

The device descriptions for the PENTAX Medical EPK-i8020c Video Imaging System remain unchanged from K231249. They include the PENTAX Medical Video Processor EPK-i8020c, PENTAX Medical Video Upper GI Scope EG29-i20c, and PENTAX Medical Video Colonoscope EC38-i20cL.

AI/ML Overview

The provided text describes the regulatory clearance of the PENTAX Medical Gas/Water Feeding Valve OE-B14, an accessory for GI endoscopes. The device's clearance is based on its substantial equivalence to a predicate device. The information provided focuses on non-clinical performance data rather than a clinical study evaluating diagnostic or therapeutic efficacy with human subjects.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Given that this is an accessory device and the review focuses on substantial equivalence based on non-clinical performance, the "acceptance criteria" here refer to meeting standards for reprocessing, sterilization, biocompatibility, and system/mechanical performance. There are no specific quantifiable clinical performance metrics like sensitivity/specificity for a diagnostic device.

Acceptance Criteria CategorySpecific Criteria/Tests PerformedReported Device Performance
Reprocessing ValidationCleaning, High-Level Disinfecting, Rinsing (after cleaning and HLD) validation studies; assessed against AAMI TIR 30: 2011/(R)2016 for residual soil accumulation and extraction efficiency.All acceptance criteria were satisfied, confirming effectiveness of reprocessing procedures.
Sterilization & Shelf LifeSteam sterilization validation (conducted with Nelson Laboratories, LCC).Validated for steam sterilization. Device is unsterile, so shelf-life is not applicable.
BiocompatibilityAssessment of cytotoxicity, sensitization, and intracutaneous reactivity in accordance with ISO 10993-1: 2018.Risk levels of local toxicity determined as "acceptable" based on risk evaluation criteria.
Software & CybersecurityNot applicable; device does not contain software.No software testing was performed as the device does not contain software.
Electrical Safety & EMCNot applicable; device is not an electrical accessory.No testing of electrical safety or electromagnetic compatibility was performed.
System PerformanceBench testing to demonstrate performance for gas/water feeding procedure.Bench testing results "demonstrated the performance enough to conduct the gas/water feeding procedure in a clinical use."
Mechanical PerformanceVerification by comparing to predicate device (OF-B194) in combination with a compatible endoscope.Mechanical performance was verified through comparison.

2. Sample Size Used for the Test Set and Data Provenance

The document describes non-clinical bench testing and validation studies, not a clinical trial with human subjects. Therefore, the concept of a "test set" in the context of clinical data or patient samples, sample size, or data provenance (country of origin, retrospective/prospective) is not applicable here as it would be for a diagnostic or therapeutic device. The "test set" would refer to the physical devices subjected to the various non-clinical tests (reprocessing, mechanical, etc.), for which specific numerical sample sizes are not provided in this summary.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This information is not applicable. The device underwent non-clinical performance testing and validation against recognized standards and internal criteria. There was no "ground truth" derived from expert clinical opinion on patient data as would be for an AI/diagnostic device.

4. Adjudication Method for the Test Set

This information is not applicable. There was no clinical test set requiring expert adjudication for ground truth.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. The device is an accessory that feeds gas/water, not a diagnostic or AI-assisted tool that would involve human readers interpreting cases.

6. If a Standalone (i.e. algorithm only, without human-in-the-loop performance) was done

No, a standalone performance study in the context of an algorithm or AI was not done. The device is a physical accessory.

7. The Type of Ground Truth Used

For this device, the "ground truth" for proving its safety and effectiveness was established through:

  • Compliance with recognized consensus standards: (e.g., AAMI TIR 30: 2011/(R)2016 for reprocessing, ISO 10993-1: 2018 for biocompatibility).
  • Achievement of pre-defined acceptance criteria for bench testing of functions, performance, and safety (e.g., residual soil levels, successful steam sterilization, acceptable toxicity levels, sufficient gas/water feeding performance, verified mechanical performance).
  • Comparison to a legally marketed predicate device (PENTAX Gas/Water Feeding Valve OF-B194) to demonstrate substantial equivalence, particularly regarding intended use and technological characteristics.

8. The Sample Size for the Training Set

This information is not applicable. The device is a physical medical accessory, not an AI or machine learning algorithm that requires a training set.

9. How the Ground Truth for the Training Set was Established

This information is not applicable as there was no training set for an AI/ML algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 1, 2023

PENTAX of America, Inc. Gurvinder Singh Nanda Senior Director, Regulatory Affairs and Quality 3 Paragon Drive Montvale, New Jersey 07645-1782

Re: K232249

Trade/Device Name: PENTAX Medical Gas/Water Feeding Valve OE-B14 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: ODC, OCX Dated: July 27, 2023 Received: July 28, 2023

Dear Gurvinder Singh Nanda:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Sivakami Venkatachalam -S

for

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K232249

Device Name

PENTAX Medical Gas/Water Feeding Valve OE-B14

Indications for Use (Describe)

The Gas/Water Feeding Valve OE-B14 is intended to be attached in place of Air/Water Feeding Valve (OE-B12) to PENTAX Medical GI endoscopes and used to feed the gas/water to the body (via the air/water channel).

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

PENTAX Medical Gas/Water Feeding Valve OE-B14 Traditional 510(k) Submission

Image /page/3/Picture/2 description: The image shows the Pentax Medical logo. The word "PENTAX" is in large, red, bold letters. Below that, separated by a gray line, the word "MEDICAL" is in gray, bold letters. At the top of the image, the text "K232249, Page 1 of 5" is visible.

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 221 CFR 807.92. All data included in this document is accurate and complete to the best of PENTAX Medical's knowledge.

1. SUBMITTER

Applicant: PENTAX Medical HOYA Corporation PENTAX Division 3 Paragon Drive Montvale, New Jersey 07645-1782

Contact: Gurvinder Singh Nanda Senior Director, Regulatory Affairs and Quality PENTAX Medical 3 Paragon Drive Montvale, New Jersey 07645-1782 Telephone: 650-722-4189 FAX: (201)391-4189 Email: gurvinder.nanda@pentaxmedical.com

Date Prepared: Nov 27th, 2023

{4}------------------------------------------------

PENTAX Medical Gas/Water Feeding Valve OE-B14 Traditional 510(k) Submission

Image /page/4/Picture/2 description: The image shows the Pentax Medical logo. The word "PENTAX" is in large, bold, red letters. Below "PENTAX" is a gray line. Below the line, the word "MEDICAL" is in large, bold, gray letters. At the top of the image, the text "K232249, Page 2 of 5" is visible.

2. SUBJECT DEVICE

PENTAX Medical is seeking clearance of a new product of PENTAX Medical Gas/Water Feeding Valve OE-B14 with the compatible PENTAX Medical Video Upper GI Gastroscope EG29-i20c and PENTAX Medical Video Colonoscope EC38-i20cL.This PENTAX Medical Gas/Water Feeding Valve OE-B14 is an accessory and is part of the Pentax Medical EPK-i8020c Video Imaging System (K231249).

DeviceNamePENTAX Medical Gas/Water Feeding Valve OE-B14
CommonNameEndoscope Channel AccessoryEndoscopic Irrigation/SuctionSystem
ClassificationNameEndoscope and accessories
Regulation No.876.1500
Device ClassII
Product CodeODCOCX
ClassificationPanelGastroenterology / Urology

3. PREDICATE DEVICE

The predicate device for this submission is PENTAX Gas/Water Feeding Valve OF-B194 which is compatible with PENTAX Medical Video Upper GI Scope EG29-i10c and PENTAX Medical Video Colonoscope EC34-i10cL (K190805).

Subject DevicePredicate Device
PENTAX Medical Gas/Water FeedingValveOE-B14PENTAX Medical Gas/Water FeedingValveOF-B194 (K190805)
PENTAX Medical Video ProcessorEPK-i8020c (K231249)PENTAX Medical EPK-i7010 VideoProcessor (K150618)
PENTAX Medical Video Upper GIScopeEG29-i20c (K231249)PENTAX Medical Video Upper G.I.ScopeEG29-i10 (K131902)
PENTAX Medical Video ColonoscopeEC38-i20cL (K231249)PENTAX Medical Video ColonoscopeEC38-i10L (K131855)

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the logo for PENTAX Medical. The word "PENTAX" is in large, red, bold letters. Below the word "PENTAX" is a gray line. Below the gray line, the word "MEDICAL" is in large, gray, bold letters. At the top of the image, the text "K232249, Page 3 of 5" is visible.

4. DEVICE DESCRIPTION

PENTAX Medical Gas/Water Feeding Valve OE-B14

The Gas/Water Feeding Valve OE-B14 is intended to be attached in place of Air/Water Feeding Valve (OE-B12) to PENTAX Medical GI endoscopes and used to feed the gas/water to the body (via the air/water channel).

The OE-B14 is attached onto the air/water feeding cylinder of the control body of PENTAX Medical endoscope. The endoscope is linked to non-flammable gas cylinder and the water bottle assembly via air/water feeding circuit.

The device descriptions for the PENTAX Medical EPK-i8020c Video Imaging System remain unchanged from K231249. They include the PENTAX Medical Video Processor EPK-i8020c, PENTAX Medical Video Upper GI Scope EG29-i20c, and PENTAX Medical Video Colonoscope EC38-i20cL.

5. INTENDED USE AND INDICATIONS FOR USE

Intended use and Indications for use for PENTAX Medical Gas/Water Feeding Valve OE-B14

The Gas/Water Feeding Valve OE-B14 is intended to be attached in place of Air/Water Feeding Valve (OE-B12) to PENTAX Medical GI endoscopes and used to feed the gas/water to the body (via the air/water channel).

6. COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE

The subject device OE-B14 is technologically and functionally equivalent to the predicate device OF-B194, and the difference between the two devices are the compatible endoscopes.

The changes in the subject device have been evaluated through performance testing including air tightness and gas / water feeding volume and raised no issues of safety and effectiveness of the device. These differences have no effect on the performance, function or general intended use of the device.

{6}------------------------------------------------

PENTAX Medical Gas/Water Feeding Valve OE-B14 Traditional 510(k) Submission

Image /page/6/Picture/2 description: The image shows the logo for Pentax Medical. The logo is in red and gray. The word "PENTAX" is in red, and the word "MEDICAL" is in gray. There is a gray line separating the two words. The text "K232249, Page 4 of 5" is at the top of the image.

7. NON-CLINICAL PERFORMANCE DATA

The PENTAX Medical Gas/Water Feeding Valve OE-B14 has been successfully tested for their functions, performance, and safety as per FDA recognized consensus standards. The following performance data are provided in support of the substantial equivalence determination.

  • i. Reprocessing Validation
    The PENTAX Medical Gas/Water Feeding Valve OE-B14 is a reusable semi-critical device (Spaulding Classification System). Since it is packaged non-sterile, it must be properly cleaned and sterilized before use.

As result of the assessment, simulated use testing, cleaning, high-level disinfecting and rinsing (after cleaning and after HLD) validation studies of the OE-B14 were conducted and confirmed the effectiveness of reprocessing procedures in accordance with FDA's 2015 Final Guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling. " Acceptance criteria were established in accordance with AAMI TIR 30: 2011/(R)2016 for amount of residual soil accumulation and extraction efficiency. All acceptance criteria were satisfied.

  • Sterilization and Shelf Life ii.
    PENTAX Medical coordinated with Nelson Laboratories, LCC to validate the use of steam sterilization for the sterilization of the OE-B14. The device is not provided sterile, therefore, shelf-life is not applicable.

  • iii. Biocompatibility
    Biocompatibility of the OE-B14 on the indirect contact materials was confirmed by assessing the cytotoxicity, sensitization, and intracutaneous reactivity in accordance with ISO 10993-1: 2018 "Biological evaluation of medical devices". The risk levels of local toxicity were determined as "acceptable" as a result of applying the risk level of local toxicity to the risk evaluation criteria.

  • iv. Software and Cybersecurity
    The PENTAX Medical Gas/Water Feeding Valve OE-B14 does not contain software. Therefore, no software testing was repeated.

  • Electrical Safety and Electromagnetic compatibility v. The OE-B14 is not an electrical accessory. Therefore, no testing of electrical safety or electromagnetic compatibility testing were repeated.

{7}------------------------------------------------

Image /page/7/Picture/2 description: The image shows the Pentax Medical logo. The word "PENTAX" is in large, red, bold letters. Below the word "PENTAX" is a gray line. Below the gray line is the word "MEDICAL" in gray, bold letters. At the top of the image, the text "K232249, Page 5 of 5" is visible.

vi. System Performance

The bench testing results of the OE-B14 demonstrated the performance enough to conduct the gas/water feeding procedure in a clinical use.

No optical characteristics bench testing or animal image capture study were conducted since the OE-B14 is not an optical accessory.

vii. Mechanical Performance

The mechanical performance of the PENTAX Medical Gas/Water Feeding Valve OE-B14 was verified by comparing it to the OF-B194 in combination with the PENTAX Medical Video Colonoscope EC38-i0cL (K190805).

Substantial Equivalence Discussion:

After analyzing the intended use, indications for use, technological characteristics (including fundamental operating principle, energy source, scientific technology, functional characteristics, design features, performance characteristics, and constituent materials), labeling, and sterilization method, PENTAX Medical concludes that the subject device is as safe and effective as the predicate device. There are no differences in indications for use and intended use between the subject and predicate device and are therefore substantially equivalent. The technological differences in terms of design features, performance characteristics and constituent materials are not substantive.

8. CONCLUSION

PENTAX Medical Gas/Water Feeding Valve OE-B14 with the compatible PENTAX Medical Video Upper GI Scope EG29-i20c and PENTAX Medical Video Colonoscope EC38-i20cL does not raise any different questions of safety and effectiveness. Therefore, the subject device is substantially equivalent to the identified predicate, the PENTAX Gas/Water Feeding Valve OF-B194 with the compatible PENTAX Medical Video Upper GI Scope EG29-i10c and PENTAX Medical Video Colonoscope EC34i10cL (K190805).

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.