K Number
K232240
Date Cleared
2023-11-13

(108 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Microwave Ablation System is indicated for the coagulation (ablation) of soft tissue. The Microwave Ablation System is not intended for cardiac use.

Device Description

The Microwave Ablation System is a microwave power (energy) source that can meet the needs of clinical soft tissue microwave thermal ablation therapy. The microwave energy is transmitted to the disposable microwave ablation needle through coaxial transmission cable, and the disposable microwave ablation needle is directly inserted into the targeted soft tissue, and the electromagnetic wave (microwave) is radiated from the front end of the disposable microwave ablation needle to cause the soft tissue to produce heat and rise. When it reaches the temperature of tissue coagulation and necrosis, in-situ inactivation is formed, that is, the purpose of microwave ablation therapy is achieved.

The Disposable Thermometer Needle and Foot Switch are accessories for the proposed device. And the proposed device is used in conjunction with the Disposable Microwave Ablation Needle made by our company.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a Microwave Ablation System. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study to prove the device meets specific acceptance criteria in the way a clinical trial or performance study for a diagnostic device would.

Therefore, much of the requested information regarding acceptance criteria, reported performance, sample sizes, expert ground truth, adjudication, MRMC studies, standalone performance, and training set details are not applicable or available from this 510(k) summary. This document primarily focuses on engineering and regulatory compliance rather than clinical performance metrics as might be found in a diagnostic AI device submission.

Here's an attempt to answer the questions based on the provided text, indicating where information is not available:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not present specific quantitative performance acceptance criteria (e.g., accuracy, sensitivity, specificity) for a clinical outcome in the way an AI diagnostic device might. Instead, the "acceptance criteria" are implied by the conformance to various safety, performance, and software standards. The reported "performance" focuses on meeting these standards and demonstrating equivalence to a predicate device.

Acceptance Criteria CategorySpecific Criteria (Implied from standards)Reported Device Performance/Compliance
Electrical SafetyConformance to IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020Designed and tested in accordance with the specified IEC 60601-1 standards.
EMCConformance to EN 60601-1-2: 2015+A1: 2021/ IEC 60601-1-2: 2014+AMD1:2020Designed and tested in accordance with the specified EN/IEC 60601-1-2 standards.
Device PerformanceConformance to IEC 60601-2-6: 2012, AMD1:2016 (Particular requirements for microwave therapy equipment)Designed and tested in accordance with IEC 60601-2-6. The overall performance test of the proposed device was conducted "as same as the predicate device," addressing main functions (power module, control feedback, microwave transmitting, water circulation).
Software Life-cycleConformance to ANSI AAMI IEC 62304:2006/A1:2016 and FDA Guidance: Content of Premarket Submissions for Device Software Functions.Software verification was conducted on the proposed device according to FDA software guidance, and the test result proves the functions in each operational mode can be achieved, and the differences in operational modes do not raise new risks.
Thermal EffectsConformance to FDA Guidance: Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery (Thermal Effects test and Temperature monitoring test).Thermal Effects test and Temperature monitoring test were conducted in accordance with the FDA Guidance. The device includes temperature monitoring features to ensure system safety and utilizes a pumped cooling system (distilled water or saline) for the ablation needle (consistent with predicate).
Intended UseIndicated for coagulation (ablation) of soft tissue, not intended for cardiac use.The stated indication for use is identical to the predicate device.
Safety and EffectivenessNo new questions of safety or effectiveness are raised compared to the predicate device.Differences from the predicate device (e.g., operational modes, specific design of components) were analyzed and determined to not raise new risks or affect substantial equivalence.

2. Sample size used for the test set and the data provenance

The document explicitly states: "No clinical study is included in this submission." This means there isn't a "test set" of patient data in the context of evaluating a diagnostic or AI-driven device's clinical performance. The testing mentioned (electrical safety, EMC, thermal effects) would involve laboratory setups and devices, not patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable, as no clinical study or test set using patient data for ground truth establishment was conducted or reported.

4. Adjudication method for the test set

Not applicable, as no clinical study or test set requiring expert adjudication was conducted or reported.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a Microwave Ablation System, which is a therapeutic device, not a diagnostic device that would typically involve human readers or AI assistance for interpretation.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This device is a therapeutic system, not an algorithm or AI product. Its performance is related to its physical functions and safety, not an AI's standalone interpretation capability.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the engineering and safety tests, the "ground truth" would be the specifications and requirements defined by the relevant international standards (e.g., IEC 60601 series) and FDA guidance documents. For instance, an electrical safety test's ground truth is whether the device does or does not meet the leakage current limits. This is based on objective measurements against engineering standards, not clinical ground truth like pathology.

8. The sample size for the training set

Not applicable. There is no mention of a training set as this is not an AI/machine learning device.

9. How the ground truth for the training set was established

Not applicable. There is no mention of a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

November 13, 2023

Betters (Suzhou)Medical Co., Ltd. % Ada Wang Operation Manager APlus Healthcare Technology (Shanghai) Co., Ltd Room 223, Building 17, JY-WISDOMBAY Huging Road 158, Baoshan District Shanghai, Shanghai China

Re: K232240

Trade/Device Name: Microwave Ablation System (BD-GT) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: NEY Dated: July 28, 2023 Received: October 16, 2023

Dear Ada Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See DICE website -(https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2023.11.13 11:12:52 -05'00'

Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical

{2}------------------------------------------------

and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K232240

Device Name Microwave Ablation System

Indications for Use (Describe)

The Microwave Ablation System is indicated for the coagulation (ablation) of soft tissue. The Microwave Ablation System is not intended for cardiac use.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

I Submitter

Betters(Suzhou) Medical Co., Ltd

Room 101, 201, 501, Building 7, 52 Yingang Road, Taicang Port Economic and Technological Development Zone Suzhou, Jiangsu Province, China, 215412

Contact person:

Lei Liu Operation Manager Tel.: +86-15261671606 E-mail: liulei(@baidemed.com

Preparation date: July 28, 2023

II Proposed Device

Trade Name of Device:Microwave Ablation System
Regulation Name:Electrosurgical Cutting and Coagulation Device and Accessories
Regulation Number:21 CFR 878.4400
Regulatory Class:Class II
Product code:NEY
Review PanelGeneral & Plastic Surgery

III Predicate Devices

510(k) Number:K201262
Trade name:Microwave Therapeutic System
Regulation Name:Electrosurgical Cutting and Coagulation Device and Accessories
Classification:Class II
Product code:NEY
Manufacturer:Nanjing ECO Microwave System Co., Ltd

IV Device Description

The Microwave Ablation System is a microwave power (energy) source that can meet the needs of clinical soft tissue microwave thermal ablation therapy. The microwave energy is transmitted to the disposable microwave ablation needle through coaxial transmission cable, and the disposable microwave ablation needle is directly inserted into the targeted soft tissue, and the electromagnetic wave (microwave) is radiated from the front end of the disposable microwave ablation needle to cause the soft tissue to produce heat and rise. When it reaches the temperature of tissue coagulation and necrosis, in-situ inactivation is formed, that is, the purpose of microwave ablation therapy is achieved.

1

{5}------------------------------------------------

The Disposable Thermometer Needle and Foot Switch are accessories for the proposed device. And the proposed device is used in conjunction with the Disposable Microwave Ablation Needle made by our company.

The supported models and specifications of the Disposable Microwave Ablation Needle and Disposable Thermometer Needle are shown in Table 1.

Table 1 Specifications of Disposable Thermometer Needle and Supported Disposable Microwave Ablation Needle

ModelNameDiameter(mm)Length(mm)Maximumoutput powerType of handle
T-1408DisposableMicrowaveAblation Needle1.48060W
T-14101.410060WStraight handle
T-16081.68060W
T-16101.610060W
L-1815DisposableMicrowaveAblation Needle1.815070W
L-18181.818070WCurved handle
L-20152.015070W
L-20182.018070W
CWZ-1620DisposableThermometerNeedle1.6200--
CWZ-16151.6150--

V Indication for use

The Microwave Ablation System is indicated for the coagulation) of soft tissue. The Microwave Ablation System is not intended for cardiac use.

VI Comparison of technological characteristics with the predicate devices

The comparison and discussion between the proposed device and the predicate device are listed in Table 2.

ItemProposed devicePredicate device(K201262)Discussion
Device/Trade nameMicrowave Ablation SystemMicrowave Therapeutic System--
ManufacturerBetters(Suzhou) Medical Co., LtdNanjing ECO Microwave SystemCo., Ltd--
Product CodeNEYNEYSame
Regulation No.21 CFR 878.440021 CFR 878.4400Same
ClassIIIISame
Indication for useThe Microwave Ablation System is indicated for the coagulation (ablation) of soft tissue. The Microwave Ablation System is not intended for cardiac use.The Microwave Therapeutic System is indicated for the coagulation (ablation) of soft tissue. The Microwave Therapeutic System is not intended for cardiac use.Same
Intended purposeCoagulation and ablation of tissueCoagulation and ablation of tissueSame
Operating principleThe Microwave Ablation System is a microwave power (energy) source that can meet the needs of clinical soft tissue microwave thermal ablation therapy. The microwave energy is transmitted to the disposable microwave ablation needle through coaxial transmission cable, and the disposable microwave ablation needle is directly inserted into the targeted soft tissue, and the electromagnetic wave (microwave) is radiated from the front end of the disposable microwave ablation needle to cause the soft tissue to produce heat and rise. When it reaches the temperature of tissue coagulation and necrosis, in-situ inactivation is formed, that is, the purpose of microwave ablation therapy is achieved.Microwave oscillating signals are generated by a microwave transistor and amplified by a microwave power amplifier. Generator delivers microwave energy to the applicator tip to thermally target tissue, resulting in coagulation and ablation.A microwave therapeutic system refers to equipment for treating diseases with microwave energy at a working frequency of 2450 MHz. The liquid in the silicone tube is driven by the rotation of the motor to flow through the radiator to realize the purpose of radiator cooling.Equivalent1
DesignOne-way ringSingle channelSame
The device is mainly composed ofthe microwave ablation instrumentwith its built-in peristaltic pumpand external foot switch, as well ascoaxial transmission cable,disposable thermometer needle andwater cooling system.cooling-water cycle, thermalablation, the probe of TEMP, footswitch.
Main FunctionThe host is mainly composed ofsolid state source and powersupply, general control system,temperature control system, displaysystem and LCD touch screen, footswitch, peristaltic pump and controlsystem, etc.The generator of the microwavetherapeutic system is composed ofa casing, a power module, acontrol feedback unit, a displaymodule, a microwave transmittingmodule, a water circulationmodule, and a temperaturemeasurement module.Similar2
AC input Voltage100-240V~, 50/60HzAC100-240V, 50/60HzSame
Output Impedance50Ω nominal50Ω nominalSame
Output parameters2450MHz±25 MHz2450MHz±20MHzSame
DeviceTemperatureMonitoringTemperature monitoring featuresused to ensure system safetyTemperature monitoring featuresused to ensure system safetySame
Device coolingPumped, distilled water or saline isused to cool the ablation needlePumped, normal saline is used tocool the antennaSame
Operational modeThree modes: continuous mode,pulse mode, foot modeThree modes can be selected bythe userDifferent3

Table 2 General Comparison of Microwave Ablation System

{6}------------------------------------------------

{7}------------------------------------------------

1 The operating principle description of the proposed device is different from the predicate device. However, the proposed device and the predicate device have working frequency, deliver microwave energy to the target tissue for coagulation and ablation, and have a device temperature monitor and cooling component. The operating principle of the devices are equivalent.

2 The main function description of the proposed device is different from the predicate device. However, the performance test of proposed device has been conducted as same as the predicate device. The general control system of the proposed device has the same function as the power module, control feedback unit and microwave transmitting module of the predicate device; the peristaltic pump and the control system of the proposed device have the same function as the water circulation module of the predicate device. Therefore, the difference does not affect substantially equivalence.

{8}------------------------------------------------

3 The operational mode of the proposed device is different from the predicate device. The predicate device has three modes, and the main difference between these three modes is that timing method is countdown or forward timing, foot switch control or manual control, output is continuous or intermittent. The proposed device has three modes; continuous mode and foot mode. The different operational modes are designed to give users more operating options, and do not change the intended use and function of the device. Software verification has been conducted on the proposed device according to the FDA software guidance, and the test result proves the functions in each operational mode can be achieved. The difference does not raise a new risk.

VII Non-Clinical Testing

The proposed device and the predicate device are substantially equivalent in design concepts, technologies, which have been designed and tested in accordance with:

Electrical Safety:

  • IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance Electromagnetic Compatibility:

  • EN 60601-1-2: 2015+A1: 2021/ IEC 60601-1-2: 2014+AMD1:2020 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbance - Requirements and tests

Performance:

  • IEC 60601-2-6: 2012, AMD1:2016 for use in conjunction with IEC 60601-1:2005 and with IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012 Medical electrical equipment - Part 2-6: Particular requirements for the basic safety and essential performance of microwave therapy equipment Software life-cycle:

-ANSI AAMI IEC 62304:2006/A1:2016 Medical device software life cycle processes [Including Amendment 1 (2016)]

  • FDA Guidance: Content of Premarket Submissions for Device Software Functions.

Thermal Effects test and Temperature monitoring test:

  • FDA Guidance: Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery

VIII Clinical Testing

No clinical study is included in this submission.

IX Conclusion

The proposed device is equivalent with respect to the basic system design and function of the predicate device. The proposed device isn't the implants and high-risk device. And it doesn't have new intended purpose, new medical indication, new target population, or new intended user. The differences between the predicate device and proposed device do not raise new questions of safety or effectiveness. Based on the comparison and analysis above, the proposed device is determined to be substantially equivalent to the predicate device.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.