K Number
K231862
Manufacturer
Date Cleared
2023-07-21

(28 days)

Product Code
Regulation Number
882.4560
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TruDi® Navigation system is intended for use during surgical procedures in ENT and ENT skull base surgery to support navigation of instruments to targeted anatomy, where reference to a rigid anatomical structure can be identified relative to a CT or MR based model.

Device Description

The TruDi® Navigation System V3 is intended to be used during surgical procedures in ENT and ENT skull base surgery to support navigation of instruments to the targeted anatomy, where reference to a rigid anatomical structure can be identified relative to a CT or MR based model. The TruDi® Navigation System V3 enables ENT physicians to access sphenoid, frontal, and maxillary sinuses, as well as the skull base, by using the systems magnetic tracking technology, which is the same technology used by the predicate device. The system incorporates a Navigation Console, Emitter Pad, Instrument Hub, Patient Tracker, Registration Probe, Workstation and peripherals. A magnetic field generated by the Emitter Pad induces a current in the magnetic sensor embedded in the tip of the navigated tool, which helps to accurately calculate the tool tip position. A CT or MR image is imported and registered to the patient coordinates and the navigated tool tip icon is displayed on top of the registered image, indicating the position of the tool tip in reference to the patient anatomy. A Patient Tracker is fixed to the patient forehead to compensate for the head movement during the surgical procedure.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Specification)Reported Device Performance (Subject Device V3)
Accuracy specification: 2mm RMSBench Location Accuracy: 0.55mm (STD 0.2)
Simulated Use Location Accuracy: 1.2mm (STD 0.2mm)

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a distinct "test set" in the context of human data or patient cases. The performance data is derived from non-clinical bench testing and simulated use.

  • Bench Location Accuracy: This test compared the TruDi® Navigation System's electromagnetic locations to those provided by a highly accurate robot system over its entire navigation volume. While a sample size of "locations" would have been involved in this measurement, the specific number is not provided.
  • Simulated Use Location Accuracy: This involved performing a complete CT image registration and instrument navigation workflow. The specific sample size for "simulated uses" is not provided.

The data provenance is not from human patients; it's from laboratory (bench) and simulated environments. Therefore, country of origin is not applicable in the typical sense.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This information is not applicable as the 'test set' for accuracy was based on objective measurements (robot system, simulated workflow) and not on expert interpretation of medical images or patient outcomes.

4. Adjudication Method for the Test Set

This information is not applicable as the ground truth was established through objective measurement tools and processes (robot system, simulated workflow) rather than human adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical data was not necessary to determine that the subject TruDi® Navigation System V3 performs as intended." This indicates that the evaluation was purely technical and non-clinical.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, a standalone (algorithm only without human-in-the-loop performance) evaluation of the device's core accuracy functions was done. The "Location Accuracy tests... where the TruDi® Navigation System electromagnetic locations were compared to the locations provided by a highly accurate robot system" directly assesses the system's inherent accuracy in a controlled, non-human environment. The "Simulated use accuracy test" also falls under this, evaluating the system's accuracy within a simulated workflow.

7. The Type of Ground Truth Used

The ground truth used for accuracy assessments was based on:

  • Highly accurate robot system measurements for Location Accuracy tests.
  • Known/defined parameters within a simulated use workflow for Simulated Use Location Accuracy tests.

8. The Sample Size for the Training Set

The document does not provide any information regarding a training set size. This device is a navigation system primarily relying on electromagnetic tracking technology, not a machine learning model that typically requires a separate training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no mention of a training set or its ground truth establishment within the provided document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 21, 2023

Acclarent, Inc. % Dikla Dayan Regulatory Affairs Manager Biosense Webster (Israel) Ltd. 4 Hatnufa St Yokneam, 2066717 Israel

Re: K231862

Trade/Device Name: TruDi® Navigation System V3 Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: June 23, 2023 Received: June 23, 2023

Dear Dikla Dayan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joyce C. Lin -S

for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231862

Device Name TruDi® Navigation System V3

Indications for Use (Describe)

The TruDi® Navigation system is intended for use during surgical procedures in ENT and ENT skull base surgery to support navigation of instruments to targeted anatomy, where reference to a rigid anatomical structure can be identified relative to a CT or MR based model.

Type of Use (Select one or both, as applicable) Registration Use (Part 21 CFR 201 Subpart D) Over-The-Counter Use (21 CFR 201 Subpart C) Registration Use (Part 21 CFR 201 Subpart D) Over-The-Counter Use (21 CFR 201 Subpart C)
Registration Use (Part 21 CFR 201 Subpart D) Over-The-Counter Use (21 CFR 201 Subpart C)

Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Acclarent, a medical technology company. The word "Acclarent" is written in a combination of black and purple font, with a purple arc above the "ent" portion of the name. Below the name, in smaller font, it says "PART OF THE Johnson-Johnson FAMILY OF COMPANIES".

Acclarent, Inc 31 Technology Drive
Main 650.687.5888 Fax 650.687.5889 www.acclarent.com

510(k) Summary - K231862

[807.92(a)(1)] Submitter Information
Applicant:Biosense Webster, Inc.31 Technology Drive, Suite200Irvine, CA 92618, USATel.: 419-233-2611Fax: 949-450-6886
Contact person:Kamrie SarnoskySenior Regulatory SpecialistAcclarent, Inc.
Authored by:Dikla DayanRA ManagerBiosense Webster (Israel) Ltd
Date summary prepared:June 23, 2023
[807.92(a)(2)] Name of Device
Device Trade NameTruDi® Navigation System V3
Device Common NameImage Guided Surgery System
Device ClassificationClass II, 21 CFR 882.4560Ear, Nose, and Throat Manual Surgical Instrument (21 CFR 882.4560)
Product CodePGW
[807.92(a)(3)] Legally Marketed Devices
Predicate Device:TruDi® Navigation System V2 (K192397)
[807.92(a)(4)] Device Description
Device Description:The TruDi® Navigation System V3 is intended to be used during surgicalprocedures in ENT and ENT skull base surgery to support navigation ofinstruments to the targeted anatomy, where reference to a rigid anatomicalstructure can be identified relative to a CT or MR based model.The TruDi® Navigation System V3 enables ENT physicians to accesssphenoid, frontal, and maxillary sinuses, as well as the skull base, by usingthe systems magnetic tracking technology, which is the same technologyused by the predicate device.The system incorporates a Navigation Console, Emitter Pad, InstrumentHub, Patient Tracker, Registration Probe, Workstation and peripherals. Amagnetic field generated by the Emitter Pad induces a current in themagnetic sensor embedded in the tip of the navigated tool, which helps toaccurately calculate the tool tip position. A CT or MR image is importedand registered to the patient coordinates and the navigated tool tip icon isdisplayed on top of the registered image, indicating the position of the tooltip in reference to the patient anatomy. A Patient Tracker is fixed to thepatient forehead to compensate for the head movement during the surgicalprocedure.
[807.92(a)(5)] Intended Use
Intended useThe TruDi® Navigation System V3 is intended for use during surgicalprocedures in ENT and skull based surgery to support navigation ofinstruments to targeted anatomy, where reference to rigid anatomicstructure can be identified relative to a CT or MR based model.
Difference in Indicationsfrom Predicate Device:There is no difference in intended use between the subject device andpredicate device.
[807.92(a)(6)] Technical Characteristics
Predicate overview andcomparison table:The substantial equivalence of the TruDi® Navigation System V3 to thepredicate is shown by similarity in intended use, indications for use,technology, and performance.The TruDi® Navigation System V3 is an image-guided navigation systemdesigned for use during ENT and skull based surgical procedures. ENTphysicians can track and display the real-time location of the tip ofnavigated instruments relative to pre-acquired CT/MR images. Like thepredicate device, the subject system utilizes electromagnetic trackingtechnology for navigation, uses anatomical reference points on the patientanatomy for intraoperative registration to the image-based model of theanatomy, and uses CT/MR image sets as reference images for the image-based model. The table below provides a comparison of the technologicalcharacteristics between the subject device and the predicate device.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Acclarent, a company that is part of the Johnson & Johnson family of companies. The word "Acclarent" is written in a bold, sans-serif font, with the first part of the word in black and the "ent" in purple. Above the purple part of the word is a purple arc. Below the word "Acclarent" is the phrase "PART OF THE Johnson & Johnson FAMILY OF COMPANIES".

AttributesPREDICATE DEVICE:TRUDI NAVIGATION SYSTEM V 2.0SUBJECT DEVICE:TRUDI NAVIGATION SYSTEM V 3.0
Fundamental ScientificTechnologyElectromagnetic locationSame
TechnologicalCharacteristicsRadiation of low intensity electromagnetic field.Acquisition of magnetic sensor inducedvoltages. Electromagnetic location in referenceto registered CT/MR background of patient headSame
TechnologySoftware embedded deviceSame
PlatformPC BasedSame
Operating SystemWindows 10 (64 bits)Same
Programming LanguagesVisual C++, C# and C, MATLAB, HTMLSame
Accuracy specification2mm RMSSame
Bench Location Accuracy0.55mm (standard Deviation 0.7 mm)0.55 (STD 0.2)
Simulated Use LocationAccuracy1.1 mm (Standard deviation 0.2 mm)1.2 mm (Standard deviation 0.2mm)
AttributesPREDICATE DEVICE:TRUDI NAVIGATION SYSTEM V 2.0SUBJECT DEVICE:TRUDI NAVIGATION SYSTEM V 3.0
Main FeatureDisplays the instruments position in relation topreoperative scans utilizing electromagnetictracking technology.Same
Location Update Rate10HzSame
Supported NavigationInstrumentsFlexible and Rigid instruments with magneticsensor on instrument tipSame
Supported PreoperativeImagesDICOM CT, MRSame
Supported CT/MR ImportMediaCD-ROM, USB flash drive, LAN network(PACs)Same
Patient Data ArchivingNot supportedSupported in local disk, flash memoryand hospital network PACS (newSoftware Function)
PackagingProvided non-sterileSame
HW Main components
Instrument HubEnables the user to connect components andnavigated tool to the systemAdded connection flexibility.Verification testing has demonstratedthat this change does not raise newquestions on the safety andeffectiveness of the device.
WorkstationDell 5810 or Dell5820Enables data processingSame
Mouse and KeyboardStandard mouse and keyboardedAdd touchpad mouse
ConsoleDrives the magnetic signals from Emitter Pad toWorkstationSame
Emitter PadGenerates magnetic fieldSame
Patient trackerCompensates on patient movement (reusable ordisposable)Same
Registration ProbeEnables correlation between patient's face andCT imageSame
Isolation TransformerAdjusting power input to the systemSame

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Acclarent, a company that is part of the Johnson & Johnson family of companies. The word "Acclarent" is written in a bold, sans-serif font, with the first part of the word in black and the last part in purple. Above the purple part of the word is a purple arc. Below the word "Acclarent" is the text "PART OF THE Johnson & Johnson FAMILY OF COMPANIES".

Acclarent, Inc 31 Technology in the line
www.acclarent.com

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Acclarent, a medical technology company. The word "Acclarent" is written in a combination of black and purple font, with a purple arc above the "ent" portion of the word. Below the logo, the text "PART OF THE Johnson & Johnson FAMILY OF COMPANIES" is written in red and gray.

Acclarent, Inc 31 Technology Drive Irvine, CA 92618 Main 650.687.5888 Fax 650.687.5889 www.acclarent.com

Summary of Technological Patient Data Archive Characteristics ● Note: The Patient Data Archive is a new Software Function, although, being a Medical Device Data System (MDDS), it is not regulated by the FDA. ● Instrument Hub improvement ● Support of additional navigated tools Software enhancements . Segmentation manual editing o Path to Target 3D visualization O ● Cables robustness improvement Optional touchpad mouse ● The TruDi® Navigation System V3 was tested to ensure that it functions in Non-clinical Performance accordance with the system design specifications related to substantial Data: equivalence in terms of device safety and effectiveness. The following non-clinical tests were performed: 1. Location Accuracy tests, where the TruDi® Navigation System electromagnetic locations were compared to the locations provided by a highly accurate robot system over its entire navigation volume 2. Software functional tests, covering the complete system functionality, and including error handling, usability and time performance (latency). 3. Safety, and EMC tests were performed by a nationally recognized testing laboratory to verify compliance with safety and EMC standards for medical devices. 4. Simulated use accuracy test, in which a complete CT image registration and instrument navigation workflow was performed, to verify the overall accuracy of the system. The proposed TruDi® Navigation System V3 passed all tests in accordance with specified test criteria and standards, and the modified device did not raise new questions of safety or effectiveness. Clinical data was not necessary to determine that the subject TruDi® Clinical performance data: Navigation System V3 performs as intended. Conclusion: The modified TruDi® Navigation System V3 is substantially equivalent to the currently cleared TruDi® Navigation System V2 based on the completion of non-clinical bench testing as well as similar principles of design, operation and indications for use.

The subject device has the same technological characteristics as the predicate device, with addition of:

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).