(79 days)
IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/or facial hair.
IPL Hair Removal Device is a personal, light-based, home-use hair reduction device intended to be sold over-the-counter. The device provides hair reduction by using Intense Pulsed Light (IPL) technology. It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain.
The device is only powered by the external power adapter (100-240V, 50/60Hz) and its IPL emission activation is by finger switch. It contains a skin sensor to detect appropriate skin contact, if the light outlet of the device is not in full contact with the skin, the device cannot emit the treatment light pulses.
The device includes JP1, JR3, JR9 four models. There is difference in product appearance, physical product dimension, indicator display, spot size and output energy, but their intended use, performance, structure design and operation are basically identical.
This FDA 510(k) summary describes the substantial equivalence of the "IPL Hair Removal Device" to predicate devices, focusing on the safety and efficacy for its intended use. It does not contain information about acceptance criteria and performances of an AI/ML powered device.
Therefore, I cannot provide a table of acceptance criteria and reported device performance related to AI/ML, nor details about sample sizes, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies for an AI component.
The document primarily addresses:
- Device Description: An over-the-counter IPL device for hair removal.
- Comparison to Predicate Devices: Demonstrates similar intended use, technology, and operational characteristics to previously cleared devices.
- Performance Data (Non-Clinical):
- Biocompatibility Testing: Conducted in accordance with ISO 10993 standards (ISO 10993-5, ISO 10993-10, ISO 10993-23) to assess cytotoxicity, irritation, and skin sensitization. All tests were passed.
- Electrical Safety and EMC: Tested against IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 60601-2-83 standards. All tests were passed.
- Eye Safety: Tested against IEC 62471.
- Software Verification and Validation: Submitted documentation consistent with a moderate level of concern, demonstrating that all software requirements are met and hazards mitigated.
- Usability: Evaluated and verified according to FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices."
The conclusion states that based on this non-clinical data, the device is as safe and effective as the legally marketed predicate devices.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 7, 2023
Shenzhen Oianyu Technology Co., Ltd. % Rilev Chen Registration engineer Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90 Qianhai Road Shenzhen, Guangdong 518052 China
Re: K231800
Trade/Device Name: IPL Hair Removal Device, Model(s): JP1, JR3, JR8, JR9 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: OHT Dated: June 19, 2023 Received: June 20, 2023
Dear Riley Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tanisha L. Hithe -S Tanisha L. 2023.09.07 Hithe -S 12:21:30 -04'00'
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K231800
Device Name IPL Hair Removal Device Model(s): JP1, JR3, JR8, JR9
Indications for Use (Describe)
IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/or facial hair.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary - K231800
"510(k) Summary" as required by 21 CFR Part 807.92.
I. Submitter
Shenzhen Qianyu Technology Co., Ltd. Room 1002, Building A2, KeXing Science Park, No.15 Keyuan Road, Nanshan District, Shenzhen, Guangdong Province, China Post code: 518057
Guoyang Li Certification Engineer Tel: +86 15099998872 Email: liguoyang(@jovs-beauty.com
II. Device
Name of Device: IPL Hair Removal Device Model(s): JP1, JR3, JR8, JR9 Common or Usual Name: Light Based Over-The-Counter Hair Removal Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810
III. Predicate Device
Primary predicate device:
| Manufacturer | Predicate Device | 510(k) Number | Approval Date |
|---|---|---|---|
| SHENZHENMEDICALINSTRUMENTS CO., LTD. | JVKHand-held IPL device (JOVSGraphene Hair RemovalDevice) | K214113 | April 5, 2022 |
Secondary predicate device:
| Manufacturer | Predicate Device | 510(k) Number | Approval Date |
|---|---|---|---|
| Shenzhen Ulike SmartElectronics Co.,Ltd | IPL Hair Removal Device,Model(s): UI06 PN, UI06 PL,UI06 JL, UI06 BR, UI06 DB,UI06 PR, UI06 OG, UI06 RD | K223618 | Feb. 28, 2023 |
Reference device:
{4}------------------------------------------------
| Manufacturer | Reference Device | 510(k) Number | Approval Date |
|---|---|---|---|
| ShenzhenJunbobeautyTechnology Co., Ltd. | IPL HAIR REMOVALHANDSET, Model: IPL-666 | K220669 | May 16, 2022 |
IV. Device Description
IPL Hair Removal Device is a personal, light-based, home-use hair reduction device intended to be sold over-the-counter. The device provides hair reduction by using Intense Pulsed Light (IPL) technology. It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain.
The device is only powered by the external power adapter (100-240V, 50/60Hz) and its IPL emission activation is by finger switch. It contains a skin sensor to detect appropriate skin contact, if the light outlet of the device is not in full contact with the skin, the device cannot emit the treatment light pulses.
The device includes JP1, JR3, JR9 four models. There is difference in product appearance, physical product dimension, indicator display, spot size and output energy, but their intended use, performance, structure design and operation are basically identical.
V. Indications for Use
IPL Hair Removal Device is an over-the-counter device intended for removal of unwanted body and/or facial hair.
VI. Materials
| Model | Material of Component | Body Contact Category | Contact Duration |
|---|---|---|---|
| JP1 | Plastic | ||
| JR3 | ABS, PC | ||
| JR8 | ABS, PC, Sapphire, Zinc Alloy,Epoxy | Surface-contactingdevice: Intact skin | Less than 24 hours |
| JR9 | ABS, PC, Sapphire |
We have tested the device for biocompatibility by a reliable third-party lab. For details, please refer to "Biocompatibility Discussion".
VII.Comparison of Technological Characteristics With the Predicate Device
The IPL Hair Removal Device has the same intended use, mode of action and similar operational characteristics as the predicate devices. Any minor differences between the subject device and the listed predicate devices and reference device do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate devices for its intended use. Therefore, the IPL Hair Removal Device may be found substantially equivalent to its predicate devices.
{5}------------------------------------------------
| ComparisonElements | Subject Device | Primary Predicate Device | Secondary Predicate Device | Reference Device | Remark |
|---|---|---|---|---|---|
| 510(k) Number | Pending | K214113 | K223618 | K220669 | / |
| Trade name | IPL Hair Removal Device(model: JP1, JR3, JR8, JR9) | Hand-held IPL device(JOVS Graphene HairRemoval Device) | IPL Hair Removal Device,Model(s): UI06 PN, UI06 PL, UI06JL, UI06 BR, UI06 DB, UI06 PR,UI06 OG, UI06 RD | IPL HAIR REMOVALHANDSETModel: IPL-666 | / |
| Regulationnumber | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Product code | OHT | OHT | OHT | OHT | Same |
| Deviceclassification | Class II | Class II | Class II | Class II | Same |
| Indicationforuse/Intendeduse | IPL Hair Removal Device isan over-the-counter deviceintended for removal ofunwanted body and/or facialhair. | The Hand-held IPL device(JOVS Graphene HairRemoval Device) is an over-the-counter device intendedfor removal of unwantedbody and/or facial hair. | IPL Hair Removal Device isindicated for the removal ofunwanted hair. The device is alsoindicated for the permanentreduction in hair regrowth, definedas the long-term, stable reduction inthe number of hairs regrowingwhen measured at 6, 9 and 12months after the completion of atreatment regime. | IPL HAIR REMOVALHANDSET is an over the-counter device intended forremoval of unwanted bodyand/or facial hair. | Same |
| Prescription orOTC | OTC | OTC | OTC | OTC | Same |
| Source energy | Supplied by external adapter | Supplied by external adapter | Supplied by external adapter | External Power supply | Same |
| Power supply | 100-240V, 50/60Hz | AC 100-240 V, 50/60 Hz | 100~240V, 50/60Hz | 100~240V AC Input12V3A DC Output | Same |
| Dimension | JP1: 188.15×58×70.6mmJR3: 159×58×58mm | Unknown | 58×34×179mm(W x H x D) | 1248348.5mm | Different |
| ComparisonElements | Subject Device | Primary Predicate Device | Secondary Predicate Device | Reference Device | Remark |
| JR8: 158.6×58×39mmJR9: 163.8×78.2×44.6mm | |||||
| Weight | JP1 & JR3: 265gJR8: 256gJR9: 260g | 275g | Unknown | 186g | Different |
| Sterilization | Not required | Not required | Not required | Not required | Same |
| Light source | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light | Intense Pulsed Light | Same |
| Energy medium | Xenon Flashlamp | Xenon lamp | Xenon Arc Flashlamp | Xenon Quartz Tube | Same |
| Wavelengthrange | JP1 & JR3: 590-1200nmJR8 & JR9: 560-1200nm | 590nm~1200nm | 560-1200nm | 470nm ~1100nm | Same |
| Energy density | JP1: 1.4 | 2.9~5.4 J/cm2 | 3~6 J/cm2 | 1.3~2.49J/cm2 | Similar |
| Spot size | JP1: 1.6cm2JR3: 3cm2JR8&JR9: 3.42cm2±0.2cm2 | 3.4cm2 | 3.3cm2 | 3cm2 | Similar |
| Pulse duration | JP1 & JR3: 7.5ms±2msJR8 & JR9: 8.2±2ms | 5.5~9.5ms | 1ms~7ms | 11.5~15ms | Similar |
| Pulsing control | Finger switch | Finger switch | Finger switch | Finger switch | Same |
| Delivery device | Direct illumination to tissue | Direct illumination to tissue | Direct illumination to tissue | Direct illumination to tissue | Same |
| Output intensitylevel | JP1: 3 levelsJR3: 6 levelsJR8: 6 levelsJR9: 5 levels | 6 levels | 3 levels | 5 levels | Same |
| Software/Firmware/ | Yes | Yes | Yes | Yes | Same |
| ComparisonElements | Subject Device | Primary Predicate Device | Secondary Predicate Device | Reference Device | Remark |
| MicroprocessorControl? | |||||
| Electrical safety | IEC 60601-1IEC 60601-1-2IEC 60601-1-11IEC 60601-2-83 | IEC 60601-1IEC 60601-1-2IEC 60601-1-11IEC 60601-2-83 | ANSI AAMI ES60601-1IEC 60601-1-2IEC 60601-1-11IEC 60601-2-57IEC 60601-2-83 | IEC 60601-1IEC 60601-1-2IEC 60601-1-11IEC 60601-2-83 | Same |
| Eye safety | IEC 62471 | IEC 62471 | IEC 62471 | IEC 62471 | Same |
| Biocompatibility | ISO 10993-5ISO 10993-10ISO 10993-23 | ISO 10993-5ISO 10993-10 | ISO 10993-5ISO 10993-10 | ISO 10993-5ISO 10993-10 | Same |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
VIII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
1) Biocompatibility Testing
The biocompatibility evaluation for the body-contacting components of the IPL Hair Removal Device was conducted in accordance with the "Use of International Standard ISO 10993-1. 'Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020", as recognized by FDA. The following testing was performed to, and passed, including:
-
ISO 10993-5: 2009, Biological Evaluation of Medical Devices -Part 5: Tests for In Vitro Cytotoxicity
-
ISO 10993-10: 2010, Biological Evaluation of Medical Devices -Part 10: Tests for Irritation and Skin Sensitization
-
ISO 10993-10: 2021, Biological evaluation of medical devices Part 10: Tests for skin sensitization
-
ISO 10993-23: 2021, Biological evaluation of medical devices Part 23: Tests for irritation
2) Electrical Safety and EMC
Electrical safety and EMC testing was performed to, and passed, as per the following standards:
-
IEC 60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
-
IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility
-
IEC 60601-1-11 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance-Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
-
IEC 60601-2-83 Medical electrical equipment Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment
3) Eye Safety
-
IEC 62471 Photobiological safety of lamps and lamp systems
4) Software Verification and Validation
Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
5) Usability
The product usability has been evaluated and verified according to the FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices", issued on FEBRUARY 2016.
IX. Conclusions
{9}------------------------------------------------
Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device IPL Hair Removal Device is as safe, as effective, and performs as well as the legally marketed predicate devices.
§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.
(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.